EXP 35 - High Performance Liquid Chromatography

download EXP 35 - High Performance Liquid Chromatography

of 13

Transcript of EXP 35 - High Performance Liquid Chromatography

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    1/13

    METU CHEMICAL ENGINEERING DEPARTMENT

    ChE420 CHEMICAL ENGINEERING LABORATORY III

    EXP. 35

    HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

    OBJECTIVE

    The purpose of this experiment is to learn the basic principles of high performance

    liquid chromatography (HPLC), which is one of the most common analysis techniques to

    identify, purify and quantify the compounds of multicomponent systems. In this experiment,

    you will use an Ion Exclusion type of HPLC with Ultraviolet Spectrophotometry Detector to

    separate and analyze the constituents of organic solutions.

    GENERAL CONCEPTS IN CHROMATOGRAPHY

    Chromatography is essentially a physical method of separation in which the

    components to be separated are distributed between two phases, one of which is stationary

    (stationary phase) while the other (the mobile phase) moves in a definite direction [1]. The

    chromatographic separations therefore have three distinct features: a) they are physical

    methods of separation; b) two distinct phases are involved; c) separation results from

    differences in the distribution constants of the individual sample components between the two

    phases. Chromatographic separations depend on a favorable contribution from

    thermodynamic and kinetic properties of the compounds to be separated. Molecular

    interactions leading to differential affinities of solutes for two phases are associated with polar

    forces (dipole-dipole and H-bonding) and dispersion forces (induced dipole) [2]. Useful

    chromatographic separations require an adequate difference in the strength of physical

    interactions for the sample components in the two phases, combined with a favorable

    contribution from system transport properties that control the movement within and between

    phases. Individual compounds are distinguished by their abilities to participate in common

    intermolecular interactions in the two phases, which can generally be characterized by

    equilibrium constant, and is thus a property that can be predicted from chemical

    thermodynamics.

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    2/13

    A convenient classification of chromatographic techniques can be made in terms of the

    physical state of the phases employed for the separation. When the mobile phase is a gas and

    the stationary phase is a solid or liquid, the separation technique is known as gas

    chromatography (GC). In GC, the analyte (the chemical constituent that is of interest in the

    analytical procedure) must be vaporized in order for migration through the capillary to occur.

    When the mobile phase is a supercritical fluid, and the stationary phase being either a solid or

    an immobilized liquid, the separation technique is called supercritical fluid chromatography

    (SFC). For GC and SFC, the dominant separation mechanisms are partitioning between bulk

    phases and interfacial adsorption.

    Transport of solute zones in column chromatography occurs entirely in the mobile

    phase. Transport is an essential mechanism of the chromatographic system since the common

    arrangement for the experiment employs a sample inlet and detector at opposite ends of the

    column with sample introduction and detection occurring in the mobile phase. In a typical

    elution chromatography, which is the most convenient method for analysis (and also used in

    this analysis), the mobile and stationary phases are normally at equilibrium. The sample is

    applied to the column as a discrete band by injection, and sample components are

    successively eluted from the column diluted by the mobile phase. The mobile phase must

    compete with the stationary phase for the sample components, and for a separation to occur,the distribution constants for the sample components resulting from the competition must be

    different.

    HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY

    Modern liquid chromatography is dominated by the use of inorganic oxides as

    stationary phases with organic functional groups chemically bonded to their surface, known as

    bonded phases, and to a lesser extent, porous polymers. High-performance liquid

    chromatography (HPLC) is unquestionably the most widely used of all of the analytical

    separation techniques due to its sensitivity, its ready adaptability to accurate quantitative

    determinations and its suitability for separating nonvolatile species or thermally fragile ones

    [1-3]. About an order of magnitude smaller particle sizes with respect to open column LC,

    allow HPLC to attain higher column efficiencies, while higher pumping pressures are required

    to overcome the large back-pressures due to the smaller packing particles. HPLC has awidespread applicability to substances that are of prime interest to industry, to many fields of

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    3/13

    science, and to the public. Examples of such materials include amino acids, proteins, nucleic

    acids (DNA and RNA), hydrocarbons, carbohydrates, drugs, terpenoids, pesticides,

    antibiotics, steroids, metal-organic species, and a variety of inorganic substances.

    Types of Liquid Chromatography

    Liquid Chromatographic modes can be categorized in five different groups according

    to their retention mechanism as can be seen in Figure 1:

    Figure 1: Liquid Chromatography Modes

    The mechanism of interaction between the analyte molecules and the stationary phase

    particles can be categorized as (a) adsorption interactions, (b) partitioning, and (c) ion-

    exchange. When the stationary phase is a solid, and interfacial adsorption is the dominant

    separation mechanism, the technique is referred to as liquid-solid chromatography (LSC).

    Polar silica, alumina or metal oxide particles are used as the stationary phase, while the

    solvent systems used are typically mixtures of relatively nonpolar liquids. The combination of

    a polar stationary phase and a non-polar mobile phase is called normal-phase HPLC, where

    the analyte is retained due to the interactions between its polar functional groups and the polar

    groups on the packing surface. Analytes elute in the order of increasing polarity, i.e. the least

    polar compound will elute from the column first, followed by each compound in the order of

    increasing polarity. This technique is preferred to separate analytes with low to intermediate

    polarity, and sufficient solubility in nonpolar solvents.

    Partition chromatography is based on solubility interactions between the analyte and

    the stationary phase. In most cases, the stationary phase is a porous silica particle to which an

    organic moiety (a functional chemical group) has been bonded. The organic substituent serves

    effectively as a bound liquid that can interact with analyte molecules. This type of stationary

    phase is called a bonded phase packing. The most popular bonded phase packings are based

    Liquid Chromatography

    Adsorption

    Competition

    between liquid

    mobile phase and

    solid adsorbent

    Partition

    Competitionbetween liquid

    mobile phase and

    liquid stationary

    phase

    Size Exclusion

    Molecular sieving

    Affinity

    Lock and Key

    mechanism

    Ion Exchange

    Competitionbetween liquid

    mobile phase and

    ionic stationary

    phase

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    4/13

    on C8 or C18-alkyl chains. This creates a non-polar stationary phase. Solvent systems used

    with these stationary phases are typically mixtures of polar liquids such as water, methanol,

    and acetonitrile. The use of a non-polar stationary phase and a polar mobile phase is called

    reversed-phase (RP) HPLC, which has become the most popular separation mode in

    modern liquid chromatography. Since the analytes are attracted to the surface by their non-

    polar functional groups, they elute (to be washed out of the column, within the mobile phase)

    in the order of increasing polarity, i.e. the most polar compound will elute from the column

    first, followed by each compound in the order of decreasing polarity. Reverse phase

    chromatography is preferred for separating compounds with moderate to high polarity, and

    sufficient solubility in polar solvents.

    Ion Exchange Chromatography (IEC) employs either a porous polymeric stationary

    phase or silica substrate with immobilized acidic or basic functional groups attached as ionic

    sites. The dominant separation mechanism is based on electrostatic interactions between ions

    in the mobile phase and those on the stationary phase, which can allow an ion exchange

    reaction to take place. The magnitude of the reaction equilibrium constant represents the

    affinity of an analyte ion for the stationary phase. The mobile phase is also an aqueous

    solution of ionic species, which compete with the analyte for exchange sites on the stationary

    phase.

    Affinity chromatography (AC) is a type of liquid chromatography that makes use of

    biological-like interactions for the separation and specific analyses of sample components [4].

    The stationary phase is a solid with immobilized molecular recognition sites which can

    establish a specific interaction with a specific solute molecule; for example, the immobilized

    molecule may be an antibody to some specific protein. In this specific example, when a solute

    containing a mixture of proteins is passed by this molecule, only this specific protein can react

    with this antibody, and so binds to the stationary phase. This protein can later be extracted by

    changing the ionic strength or pH of the mobile phase.

    If the stationary phase is a solid with a controlled pore size distribution, and solutes are

    separated by size differences, i.e. molecular sieving, then the technique is called size-

    exclusion chromatography (SEC). On the contrary to the previous chromatography modes

    explained above, this type of chromatography which is also known as gel permeation or gel

    filtration, lacks an attractive interaction between the stationary phase and the solute. Since the

    pores are small and exclude the larger solute molecules, only the smaller molecules can enter

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    5/13

    the gel and thus flow through a larger volume. Therefore, larger molecules have less retention

    times, passing through the column at a faster rate than the smaller ones.

    Instrumentation for HPLC

    The basic components of a high-performance liquid chromatographic system is shown

    in Figure 2. The instrument consists of (a) eluent containers for the mobile phase, (b) a pump

    to move the eluent and sample through the system, (c) an injection device to allow sample

    introduction, (d) a column(s) to provide solute separation, (e) a detector to visualize the

    separated components, (f) a waste container for the used solvent, and finally (g) a data

    collection device to assist in interpretation and storage of results [3].

    The pump, injector, column, and detector are connected with tubing of narrow innerdiameter. The choice of detector is based on intrinsic properties of the solute. Often more than

    one detector can be used to maximize sample information and conform peak identities.

    Figure 2. Schematic diagram of the basic modules of an HPLC system.

    The pumping system with reciprocating pistons can supply either an isocratic or a

    gradient elution depending on the number of pumps used in the system. In isocratic elution,

    the mobile phase composition is kept constant throughout the analysis, while in gradient

    pumping systems, more than one solvent can be used, and so the composition of the mobile

    phase can be altered, which requires multiple pumps. The sample injection of the system can

    be either a manual injection using 6-port valve, or an automated sample injection device. Two

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    6/13

    types of detectors are widely used for monitoring separations. An absorbance detector

    measures the absorbance of a particular wavelength of ultraviolet light, usually tuned for

    aromatic compounds. The alternative is a detector that monitors changes in the refractive

    index of the solvents being eluted. This allows for detection of compounds that may be

    transparent to UV radiation.

    One of the most critical factors in HPLC analysis is the quality of the solvents used

    [3]. In addition to being particulate and contaminant free, solvents must contain minimal

    levels of dissolved gases. Dissolved gases come out of solution when the eluents are pumped

    through the system, and the probability of gases bubbling out of solution increases when two

    or more liquids are mixed. The microbubbles frequently become attached to check valves, and

    interfere with accurate flow. In addition, they can become lodged in detector cells, creating

    baseline instability. Therefore, the mobile phase and all the solvents must be degassed prior to

    their use in the instrument.

    The Output of Chromatography-Chromatograms

    The information obtained from a chromatographic experiment is contained in the

    chromatogram [1]. When the elution mode is used, this consists of a plot of a detector

    response (y-axis) as a continuous function of time or volume of mobile phase passed through

    the column (x-axis). The chromatogram contains a number of peaks of various sizes rising

    from a baseline. Information readily extracted from the chromatogram includes an

    identification of a samples complexity from the number of observed peaks; qualitative

    identification of sample components from the accurate determination of the peak positions;

    quantitative assessment of the relative concentration or amount of each component from their

    peak areas; and the characteristic physical properties of either the solute or the

    chromatographic system from the peak positions and profiles. The position of a peak in a

    chromatogram is characterized by its retention time (tR), which is the time taken for a specific

    solute to reach the detector, or retention volume (V R), which is the volume of solution that is

    pumped through the detector before a specific peak is eluted (Figure 3).

    There are four major concepts, in terms of which chromatography is described:

    capacity, efficiency, selectivity and resolution [3]. The capacity and selectivity of the column

    are variables that are controlled largely by the column manufacturer, whereas efficiency and

    resolution can be controlled to some extent by the operator. To obtain the best possibleseparation, the efficiency of the chromatographic system must be optimized in order to

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    7/13

    minimize band broadening. The column should have the capacity to retain the solutes, and the

    ability to separate sample components efficiently. The capacity factor kR of a column is a

    direct measure of the strength of the interaction of the sample with the packing material:

    = 0

    0Eqn 1

    where tR is the retention time and t0 is the time taken for non-retained species to reach the

    detector (hold-up time).

    Figure 3: Chromatographic terms used to calculate column capacity

    The capacity factor of a column is mostly a function of the packing material, but can

    be manipulated to a degree by varying the solvent strength. The higher the kR, the greater is

    the ability of the column to retain solutes. However, this also means longer analysis times;

    therefore, a compromise between resolution and analysis time must be reached. A kR value

    between 2 and 5 represents a good balance between the analysis time and resolution.

    The selectivity () of chromatographic separation is a measure of the difference in

    retention times between two given peaks and describes how effectively the system can

    separate two compounds.

    =0

    0=

    Eqn 2

    The selectivity of a column is primarily a function of the packing material; although

    the chromatographer has some control using the mobile phase or temperature. An effective

    way to increase is to change the composition of the mobile phase.

    The resolution represents the degree of separation between neighboring solute peaks.

    It depends on the selectivity (), efficiency (N), and capacity (kR) of the column.

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    8/13

    =1

    4

    1

    (1 2 )

    1+Eqn 3

    Peak broadening occurs generally due to sample components interactions with the

    stationary phase, and the torturous path that the solute has to take through the packingmaterial. The efficiency of a column, N, is a number that describes peak broadening as a

    function of retention, and can be described simply in terms of the number of theoretical plates.

    In plate theory, the column is considered to consist of a number of thin plates, each of which

    allows a solute to equilibriate between the stationary and mobile phases. The greater the

    number of theoretical plates, i.e., the efficiency, N, the more efficient the column is

    considered to be. The movement of a solute along the column is viewed as a stepwise transfer

    from one theoretical plate to the next. N can be defined using the half-height method as:

    = 5.545(

    /)2

    where W1/2 is the peak width at the half height of the peak (in units of time) (Figure 4).

    Figure 4: Baseline width (Wb) and width-at-half-height (W1/2) of a chromatographic peak.

    The efficiency can be varied by changing the physical parameters such as length,

    diameter, and construction material of the container of the column. It can also be varied by

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    9/13

    changing chemical parameters such as the size of the particles constituting the packing

    material or the mobile phase velocity.

    Note: The efficiency of a column is described more realistically with the rate theory which

    avoids the assumption of an instantaneous equilibrium, and addresses the diffusional factors

    contributing to band broadening in the column, such as eddy diffusion, longitudinal diffusion,

    and resistance to mass transfer in the stationary phase and the mobile phase. Contribution of

    these factors to efficiency is represented in a Van Deemter equation, and a plot is used to

    determine the experimental conditions required to obtain the most efficient system [3].

    References:

    1- C. F. Poole, The Essence of Chromatography, 1st

    ed., Elsevier, Amsterdam, 2003.2- D. A. Skoog, J. J. Leary, Principles of Instrumental Analysis, 4th ed., Saunders College

    Publishing, Fort Worth, 1992.

    3- A. Weston,P. R. Brown,HPLC and CE: Principles and Practice, 1st ed., AcademicPress, San Diego,1997.

    4- D. S. Hage, Affinity Chromatography: A Review of Clinical Applications, ClinicalChemistry, 45, 5, 593-615, 1999.

    PRELIMINARY WORK

    Be prepared to answer the following questions prior to the laboratory session.

    - What is HPLC used for? What is the main difference between an HPLC and GC?- What is the basic principle of HPLC?- What are the types of liquid chromatography? What are the separation mechanisms in

    each type?

    - What are the most commonly used detectors in HPLC? Get familiarize with theirworking principles.

    - What is the principle of ion exclusion chromatography? (Please use instrumentalanalysis textbooks or other sources to learn the answer).

    - What are the basic working priciples of reciprocating piston pumps and six-portvalves?

    - Study the acid dissociation phenomena, look into the pK a values of the commonstrong and weak acids.

    http://www.google.com.tr/search?hl=tr&tbo=p&tbm=bks&q=inauthor:%22Andrea+Weston%22http://www.google.com.tr/search?hl=tr&tbo=p&tbm=bks&q=inauthor:%22Phyllis+R.+Brown%22http://www.clinchem.org/search?author1=David+S.+Hage&sortspec=date&submit=Submithttp://www.clinchem.org/search?author1=David+S.+Hage&sortspec=date&submit=Submithttp://www.google.com.tr/search?hl=tr&tbo=p&tbm=bks&q=inauthor:%22Phyllis+R.+Brown%22http://www.google.com.tr/search?hl=tr&tbo=p&tbm=bks&q=inauthor:%22Andrea+Weston%22
  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    10/13

    - Describe the following chromatographic parameters: column capacity factor,selectivity, resolution and number of theoretical plates. How would you manipulate

    these parameters in a chromatographic system?

    EXPERIMENTAL PROCEDURE

    1. Make sure that a baseline has been attained before starting to analysis.2. Take readings of the mobile phase flow rate, column temperature, minimum and

    maximum pressure limits of the pumps, elution profile, sample volume that is

    injected, composition of the mobile phase.

    3. Preparation of the mobile phase; (this will have been already prepared for therun, so you can skip this step; but you must know how it is prepared.)

    Prepare 8 mM of sulfuric acid solution with Millipore Q (high purity water). Put

    435 l of 96-98% (w/w) sulfuric acid (as received from Merck, 5 L bottles) into 1

    L of MilliQ. Filter it through 0.45 m cellulose acetate membrane filter. Degass the

    solution for 10-15 minutes at room temperature. Filtered and degassed mobile

    phase is now ready to use in HPLC. The mobile phase can be used at most 1 week;

    for the next week, the fresh solution should be prepared since microbial growth

    will occur in the solution and it will obviously absorb air with time.

    4. To obtain the calibration curves for the standards;The standards, oxalic acid, citric acid and malic acid will be prepared at a

    concentration of 500 ppm (stock solutions). For standard preparation, mobile

    phase solution should be used. Prepare different standards, each with one type of

    acid solution at a certain concentration less than 300 ppm. These standards will

    allow you to detect the retention times of each acid. Next, prepare three different

    standards, each with all three acids at a certain concentration again less than 300

    ppm, and different than the concentration of the first set of standards. Use a glass

    pipette (1ml) and sample vials for the preparation of these standards.

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    11/13

    5. Press single injection and upload the method file (recently uploaded in thesoftware). Also upload a data file which represents the name of the single sample

    that will be analyzed in the device. Fill the sample volume as 20 L and press OK.

    A small box is going to appear which specifies that the device waits for the

    manual injection.

    You will not be using an auto-sampler so you have to perform injections manually

    through the injection port and wait until each analysis will be completed. Before

    each injection make sure that you rinse your syringe (stainless steel syringe) with

    mobile phase to remove any contaminants. Slowly and completely fill the syringe

    with your sample. Before injection, make sure that the sample in the syringe does

    not contain any bubbles. Any air bubble can cause damage for the device. These

    syringes are non-dispensable and expensive. So please pay attention while using

    it.

    Before injection, the injection port should be in load mode (left). Put the syringe

    to the injection port gently. Make the injection. After you make the injection of 50

    l and remove the injector from the port, you will turn the port to the inject mode

    (right), at the same time you will see the running alert on the monitor.

    6. After you complete this procedure for all of your standards, repeat the sameprocedure for the samples with unknown compositions.

    7. Obtain your runs data report. Make sure you have all the data you need tocomplete the post-run calculations for your report.

    DATA SHEET

    Initial readings of the main menu

    Analysis Time:

    Flow rate of mobile phase:

    Column temperature:

    Pressure limits of the pump:

    UV absorbance wavelength:

    Injection volume:

    Composition of the mobile phase:

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    12/13

    STANDARDS CompoundConcentration

    (ppm)Retention

    timePeak Area

    1

    2

    3

    4All threeacids

    5All threeacids

    6All threeacids

    UNKNOWN

    Non-retained

    species,Rt

    Non-retained

    species,Area

    Rt, 1 Area1 Rt, 2 Area2 Rt, 3 Area3

    wh1 wh2 wh3

    CALCULATIONS

    1. Construct the calibration curve for each species. Use least squares procedures tocompute the slope, y-intercept, r2, and standard error of estimate.

    2. Determine the concentration of the components in the unknown solution.3. Choose one of your unknown samples and calculate:

    a. Capacity factor of the column for each species

  • 8/2/2019 EXP 35 - High Performance Liquid Chromatography

    13/13

    b. Selectivity of the column; i.e., selectivity for citric acid over oxalic acid, andfor malic acid over citric acid

    c. resolution of the two sets of neighboring peaksd. efficiency of the column for each species: i.e., the number of theoretical

    plates for each species

    Please discuss the results of your calculations. Your discussions should include your

    comments about the efficiency and reliability of this analysis, and if you have any suggestions

    to improve it.