ENSC 220 Lab Tutorial
Transcript of ENSC 220 Lab Tutorial
![Page 1: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/1.jpg)
1
ENSC 220 Lab Tutorial
Fall 2008
Credits: Dr. Ash Parameswaran, notes
Dr. Atousa Hajshirmohammadi, layout
![Page 2: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/2.jpg)
2
ENSC LAB (9000 level)
![Page 3: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/3.jpg)
3
Component rack
![Page 4: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/4.jpg)
4
Lab bench
Digital multimeter
Function generator
Oscilloscope
Dual power supply
![Page 5: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/5.jpg)
5
Bread board
The red lines indicate holes that are electrically connected
![Page 6: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/6.jpg)
6
Neatly done wiring
![Page 7: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/7.jpg)
7
Poorly done wiring
If you wire your circuit this manner,please do not bring it to us
for troubleshooting.
![Page 8: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/8.jpg)
8
Dual power supply
Powerswitch
Voltageknob
Current limit
Tracking(or)
independent
Voltage or current display
![Page 9: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/9.jpg)
9
Digital multimeter
Power switch
Range selectMode select
AC/DC measurement
Voltage/resistancemeasurement
![Page 10: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/10.jpg)
10
Digital multimeter(current measurement)
Fuse is inside this socket
Current measurement
![Page 11: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/11.jpg)
11
DMM specs
Specification and accuracy
![Page 12: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/12.jpg)
12
Meter accuracy
DC Volts ± (0.1% of the reading + 1 digit)
DC Amps ± (0.3% of the reading + 1 digit)
Resistance ± (0.2% of the reading + 1 digit)
![Page 13: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/13.jpg)
13
What does the +1 digit mean?
If in a circuit, we read 0.812 Volts using the 2 V range setting:
The accuracy is ± (0.1% of the reading + 1 digit)
Hence, the accuracy is ± 0.000812 V However the meter can only display 3 digitsbeyond the decimal point (2V range)
Therefore, actual reading accuracy is ± 0.001 V
For this measurement, the value is [0.812 ± 0.001] V
DisplayRange
200 mV 00.02 V 0.00020 V 0.00200 V 00.0
0.813 V
0.811 V
![Page 14: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/14.jpg)
14
How to set-up power supply:single supply (source)
Connect wires toterminals and short
the free end
Make sure this switch
is off
Turn onthe switch
Flick this switch to measurecurrent This LED lights-up
Rotate these two knobsanticlockwise one turn
Rotate the voltage knobslightly clock wise
Rotate the current knobclockwise and set the desired
max current Remove this short
Rotate the voltage knobclockwise and set the desired
voltage
![Page 15: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/15.jpg)
15
How to set-up power supply:dual supply (source)
Set current limit for both sides as already described
+ -
GndSet this switch for tracking when you wish both supplies to have
equal magnitude
![Page 16: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/16.jpg)
16
Sample experiment
Circuit:
0.5 V 470 Ohm
I
What is the value of I?
Theoretical value:
0.5 V
470 Ohm= 1.063829 mA
![Page 17: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/17.jpg)
17
First level experiment/calculation
Measure the resistance using an Ohm-meter
2 kOhm range, therefore the value is 0.470 ± 0.001 kOhm
Max value = 471 Ohm, Min value = 469 Ohm
![Page 18: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/18.jpg)
18
First level experiment/calculation (continued)
Measure the voltage of the supply
2 V range, therefore the value is 0.500 ± 0.001 V
Max value = 0.501 V, Min value = 0.499 V
![Page 19: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/19.jpg)
19
Measure current
2 mA range, therefore the value is 0.877 ± 0.003 mAMax value = 0.880 mA, Min value = 0.874 mA
Our theoretical value is = 1.063829 mASomething strange is happening here!
![Page 20: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/20.jpg)
20
Internal resistance of the meter (shunt)
2 mA 100 Ohm20 mA 10 Ohm
200 mA 1 Ohm
0.500 ± 0.001 V 0.470 ± 0.001 kOhm
100 Ohm(meter shunt)
Realistic circuit model:
![Page 21: ENSC 220 Lab Tutorial](https://reader030.fdocuments.net/reader030/viewer/2022032709/623bd105efab8815da741eb3/html5/thumbnails/21.jpg)
21
Reconcile measurement results with expected values
Max value = 0.501 V, Min value = 0.499 V
Max value = 471 Ohm, Min value = 469 Ohm
Voltage measurement:
Resistance measurement:
Calculated current:
Max value =
Min value =
0.501 V569 Ohms
0.499 V571 Ohms
= 0.8804 mA
plus 100 Ohm
= 0.8739 mA
Max value = 0.880 mA
Min value = 0.874 mA
Measured current:
Excellent reconciliation when the calculatedand measured values match well within the meter’s accuracy