ENHANCEMENTS TO THE ION MOBILITY … · Figure 5 Mobility separation of the fragment ions of the...
Transcript of ENHANCEMENTS TO THE ION MOBILITY … · Figure 5 Mobility separation of the fragment ions of the...
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS
Figure 1 A schematic diagram of the Synapt G2 HDMS instru-ment
INTRODUCTION One of the more recent developments in ion mobility (IM)
separation technology has been the travelling wave-based
approach used in a commercially available hybrid quadrupole/
IM/oa-ToF instrument (Synapt HDMS, Waters Corp.). Whilst
this technology has provided greater access to the potential of
ion mobility spectrometry in combination with mass
spectrometry for many analysts, the mobility resolution
afforded by the travelling wave separator is relatively low in
comparison with current, albeit significantly larger,
instruments in some research laboratories. What is presented
here is a novel approach towards increasing the mobility
resolving power of the travelling wave device together with a
description of an enhanced detection system for acquiring
mobility data.
Kevin Giles, Tony Gilbert, Martin Green and Garry Scott Waters MS Technologies Centre, Manchester, UK
The trap T-Wave accumulates ions whilst the previous mobility
separation occurs, then the ions are released in a packet into
the IMS T-Wave where mobility separation is performed. The
transfer T-Wave delivers the mobility separated ions to the oa-
ToF mass analyser. Ion arrival time (or drift time)
distributions are recorded by synchronising the oa-ToF mass
spectral acquisitions with the gated release of ions from the
trap T-Wave device. The T-Wave mobility separator uses a
repeating train of DC pulses to propel ions through the gas-
filled cell in a mobility dependent manner. To provide
increased mobility resolution, higher N2 pressures and higher
T-Wave pulse amplitudes were required compared to those on
a standard Synapt instrument. Increasing the N2 pressure
alone in the IMS cell is problematic since driving the ions into
this device then requires higher voltages which can lead to
unacceptable losses through scattering and/or fragmentation.
The solution to this was to employ a high pressure Helium-
filled cell on the front of the IMS cell to provide a low loss
interface to the high N2 pressure region, see Figure 2.
References (1) C Wu, W.F. Siems, J. Klasmeier and H.H. Hill, Anal. Chem. 72 (2000) 391
(2) AE Counterman , SJ Valentine, CA Srebalus, SC Henderson, CS Hoaglund, DE Clemmer, J.Am. Soc. Mass Spectrom., 9 (1998) 743
RESULTS AND DISCUSSION Inverse Sequence Peptides The inverse sequence peptides GRGDS and SDGRG (mw 490)
have been studied previously and the doubly charged
molecular ion species of each at m/z 246 were determined to
have a difference in collision cross section (Ω) of 5.2% in an N2
drift tube at atmospheric pressure1. In Figure 3, the mobility
separation attained on the standard IMS cell and the new high
pressure cell are shown for a mixture of the two peptides.
CONCLUSION
A new high resolution Travelling wave IMS cell operating at N2 pressures of ~2.5mb with a high pressure Helium-filled entry cell has been described
Mobility resolution increases of 3-4X are achieved over the standard IMS cell in a Synapt instrument
Near baseline mobility separation achieved on doubly charged isomeric peptide doublet with 5% difference in collision cross-section indicating Ω/∆Ω of ~45
Use of a new ADC-based detection system illustrating high dynamic range in mobility mode of acquisition
Glu-Fibrinopeptide B Fragmentation
The doubly charged precursor ion of glu-fibrinopeptide b at m/z
785.6 was selected using the quadrupole mass filter and
fragmented in the trap T-Wave collision cell. The fragments
were then separated by mobility in the IMS T-wave cell. Figure
5 shows the resulting total ion current arrival time distribution
and those of selected fragments from the standard mobility cell
and the new mobility cell. Again, highlighting the increased
mobility resolution.
ENHANCEMENTS TO THE ION MOBILITY PERFORMANCE OF A TRAVELLING WAVE SEPARATION DEVICE
OVERVIEW
Mobility resolution improvements on a travelling wave separator and an en-hanced detection system
Studies undertaken on a Synapt G2 HDMS instrument
Mobility resolution increase of 3-4x and increased dynamic range with an ADC-based detection system
Figure 2 A schematic diagram of the new T-Wave IMS cell with the Helium-filled entry cell
Helium Nitrogen
Helium Out
Nitrogen Out
Ions Out
Ions In
Helium Cell
T-Wave IMS Cell
Stacked Ring Ion Guide
Helium Nitrogen
Helium Out
Nitrogen Out
Ions Out
Ions In
Helium Cell
T-Wave IMS Cell
Stacked Ring Ion Guide
Using the helium cell, working pressure of up to 2.5mb of N2 in
the IMS were possible, compared to the 0.5mb typically used
on a standard Synapt IMS cell.
The increase in mobility resolution poses significant issues for
the TDC-based detection system used on a standard Synapt,
reducing the available dynamic range for both intensity and
mass accuracy. Whilst moving to an ADC-based detection
system was expected to alleviate these issues, no
commercially available systems were capable of dealing with
the demands of recording IMS arrival time data. Consequently
a completely new ADC system was designed and built in-
house.
The ADC system comprises an 8-bit ADC sampling at 3Ghz
feeding to a field programmable gate array for signal
processing and subsequently to a block of memory for
accumulating the 200 sequential mass spectra that form the
mobility arrival time spectrum. Summing of the mobility data
is performed on the ADC card and ultimately transferred to the
acquisition PC after each ‘scan’ period.
With the new, higher pressure mobility cell, the mobility reso-
lution (Ω/∆Ω) has increased to ~45 from the resolution of ~11
for the standard mobility cell.
Bradykinin Multimers The formation of multiply charged multimeric species of the
form (M+H)nn+ has been demonstrated for various peptides in-
cluding Bradykinin2. The first isotopes of these species have
the same m/z value but can be separated by their differences
in mobility. Figure 4 shows the mobility separation of the
nominal m/z 1061 ion species of Bradykinin using the standard
mobility cell and the new, high pressure cell.
With the new, higher pressure mobility cell, the mobility reso-
lution (Ω/∆Ω) has increased by approximately a factor of three
over the standard mobility cell. In addition, these data
illustrate that even at five times the pressure of the standard
cell, the new design with the helium cell is able to efficiently
transmit the relatively labile (M+H)33+ species.
Figure 4 m/z vs arrival time and m/z 1061 mobility chroma-togram (a) Synapt at 0.5mb N2 and a 300m/s 9V wave (b) Synapt G2 at 2.5mb N2 and a 600m/s 40V wave.
B K_10.raw : 1
K G_B K_01. raw : 1(a)
(b)
4
B K_10.raw : 1
K G_B K_01. raw : 1(a)
(b)
4
(M+H)33+
(M+H)22+
(M+H)11+
(M+H)33+ (M+H)2
2+
(M+H)11+
Figure 5 Mobility separation of the fragment ions of the dou-bly charged m/z 785.6 ion of glu-fibrinopeptide b (a) Synapt at 0.5mb N2 and a 300m/s 9V wave (b) Synapt G2 at 2.5mb N2 and a 800m/s 40V wave.
0 2 4 6 8 10 12
Arrival Time (ms)
Nor
mal
ised
Inte
nsity
(%)
0 2 4 6 8 10 12
Arrival Tim e (m s)
Nor
mal
ised
Inte
nsity
(%)
m/z 246 333 786 480 684 812 942 1056 1171 1285
0 2 4 6 8 10 12
Arrival Time (ms)
Nor
mal
ised
Inte
nsity
(%)
0 2 4 6 8 10 12
Arrival Tim e (m s)
Nor
mal
ised
Inte
nsity
(%)
m/z 246 333 786 480 684 812 942 1056 1171 1285
TIC
TIC
(a)
(b)
ADC-based Detection System
A solution of sodium formate was infused into the electrospray
source of the Synapt G2. Acquisitions were made in both non-
mobility and mobility mode. The mass spectra in Figure 6 are
summed mass spectra over 30 scans with a single point internal
lock mass correction using the m/z 566.8892 ion. The m/z
differences between the calculated and measured m/z values are
presented in Table 1. As can be seen there is excellent
agreement with the calculated values for both data sets. Figure
7 shows the mobility data for the sodium formate sample. The
mobility arrival time peaks for single m/z species are
approximately 3 ToF pushes (mass spectra) wide and the ion
concentration in these peaks is around 66X (200/3) that of the
non-mobility mode. From the acquired data, the m/z 430.9 ion
flux in non-mobility mode is calculated to be around 0.2 ions per
push (3000 counts/s), consequently, in mobility mode this
increases to 13 ions per push in the mobility peak.
METHODS Instrumentation The instrument used in these studies was a Synapt G2 HDMS
instrument (Waters Corp.), shown in Figure 1, which has a
hybrid quadrupole/IM/oa-ToF geometry. Ions are generated
using an electrospray ionisation source and pass through a
quadrupole mass filter to the IM section of the instrument which
contains three travelling wave (T-Wave) ion guides.
Experimental All samples were infused into the electrospray source at 5µL/
min. The source conditions were tuned for each sample to give
best signal. The trap and transfer T-Wave cells were operated
with Argon at a nominal pressure of 2x10-2mb. The helium cell
performance was optimised with a resultant gas flow of 200mL/
min (pressure not measured directly), approximately twice the
main IMS cell flow of N2. The standard Synapt was operated
with ~0.5mb of N2 in the IMS cell and the Synapt G2 with
~2.5mb N2. T-Wave settings are listed for each experiment.
Figure 3 The arrival time distributions of the inverse pep-tides at m/z 246. (a) Synapt at 0.5mb N2 and a 300m/s 6V wave (inset are the data for the peptides run individually) (b) Synapt G2 at 2.5mb N2 and a 1300m/s 40V wave.
0
20
40
60
80
100
1.5 2 2.5 3 3.5 4 4.5
Arrival Time (ms)
Rel
ativ
e In
tens
ity
(%
0
20
40
60
80
100
2 2.5 3 3.5 4 4.5 5
Arrival Time (ms)
Rel
ativ
e In
tens
ity (%
(GRGDS)2+
211.7 Ų(SDGRG)2+
222.7 Ų
(a)
(b)
1.5 2 2.5 3 3.5 4 4.5
0
20
40
60
80
100
1.5 2 2.5 3 3.5 4 4.5
Arrival Time (ms)
Rel
ativ
e In
tens
ity
(%
0
20
40
60
80
100
2 2.5 3 3.5 4 4.5 5
Arrival Time (ms)
Rel
ativ
e In
tens
ity (%
(GRGDS)2+
211.7 Ų(SDGRG)2+
222.7 Ų
(a)
(b)
1.5 2 2.5 3 3.5 4 4.5
FWHM ~0.09ms
m/z200 400 600 800 1000 1200
%
0
100
%
0
100KG_ADC_01 37 (0.806) TOF MS ES+
2.87e6430.9138226.9514
203.0322
362.9269
838.8392702.8635*566.8892
498.9016 634.8764
770.8513 906.8262 974.8135
1042.8010
1110.7886
1178.7756
KG_ADC_02 38 (0.689) TOF MS ES+ 1.89e6226.9522
203.0324
430.9147
362.9276
*566.8892
498.9020
702.8642
634.8767838.8397
906.8271 974.8143
1042.80201110.7887
1178.7767
Non-M obility
Mobility
m/z200 400 600 800 1000 1200
%
0
100
%
0
100KG_ADC_01 37 (0.806) TOF MS ES+
2.87e6430.9138226.9514
203.0322
362.9269
838.8392702.8635*566.8892
498.9016 634.8764
770.8513 906.8262 974.8135
1042.8010
1110.7886
1178.7756
KG_ADC_02 38 (0.689) TOF MS ES+ 1.89e6226.9522
203.0324
430.9147
362.9276
*566.8892
498.9020
702.8642
634.8767838.8397
906.8271 974.8143
1042.80201110.7887
1178.7767
Non-M obility
Mobility
Figure 6 Summed mass spectra for sodium formate (a) with-out mobility separation (b) with mobility separation
(a)
(b)
Figure 7 Mobility chromatogram and m/z vs arrival time for sodium formate
KG_ADC_02.raw : 1
m /z 8 3 8 .8
KG_ADC_02.raw : 1
m /z 8 3 8 .8
Table 1 Mass measurement accuracy for sodium formate: non-mobility data and mobility data
m/z 158.9646 226.9520 294.9395 362.9269 430.9143 498.9017 566.8892 634.8766ppm -1.37 -2.83 -0.22 0.03 -1.19 -0.28 0.07 -0.29ppm -0.11 0.7 -1.24 1.96 0.9 0.53 0.07 0.18
m/z 702.8640 770.8514 838.8389 906.8263 974.8137 1042.8011 1110.7886 1178.7760ppm -0.73 -0.17 0.41 -0.09 -0.21 -0.12 0.04 -0.32ppm 0.27 0.48 1 0.9 0.61 0.83 0.13 0.61