Page 1: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

DIGITAL TRANSFORMER RATIOMETER

E N G L I S H User Manual

8510

Page 2: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.

Serial #: ________________________________

Catalog #: 2136.50

Model #: 8510

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com

Page 3: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 1

Table of Contents

1. INTRODUCTION ............................................................................... 31.1 International Electrical Symbols ................................................41.2 DefinitionofMeasurementCategories .....................................41.3 ReceivingYourShipment ..........................................................51.4 OrderingInformation .................................................................5

1.4.1 AccessoriesandReplacementParts ............................5

2. PRODUCT FEATURES ...................................................................... 62.1 Description ................................................................................62.2 ControlFeatures .......................................................................82.3 CableIdentification ...................................................................9

3. SPECIFICATIONS........................................................................... 103.1 DTR®8510Specifications .......................................................103.2 BatteryChargerSpecifications ...............................................12

4. DISPLAY FUNCTIONS .................................................................... 134.1 ProgramFlow .........................................................................134.2 TopLevelMenu ......................................................................134.3 ConfigureInstrument ..............................................................14

4.3.1 SetClock ....................................................................164.3.2 SetupNameplate ........................................................174.3.3 SelectTestType ..........................................................204.3.4 Select Test Mode ........................................................204.3.5 SelectStorageMode ..................................................214.3.6 Select Filter .................................................................224.3.7 Erase Memory.............................................................234.3.8 SelectLanguage .........................................................244.3.9 RestoringFactoryDefaults .........................................25

4.4 Recall Data .............................................................................26

Page 4: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

2 Digital Transformer Ratiometer DTR® Model 8510

5. OPERATION .................................................................................. 285.1 PowerUp ................................................................................285.2 RunningaTest ........................................................................29

5.2.1 VT/PTTest ..................................................................295.2.2 StoringtheMeasurementRecord(Manual/Auto) .......31

5.2.2.1 ManualMode ........................................................315.2.2.2 AutoMode ............................................................32

5.2.3 VT/PTExcitationCurrent ............................................325.2.4 VT/PTContinuityTest .................................................335.2.5 CTTest ........................................................................33

5.3 TipsforMakingPreciseRatioMeasurements ........................355.4 Ratio Test - 1:1 .......................................................................36

6. CONNECTIONS .............................................................................. 376.1 ConnectionDiagrams .............................................................376.2 PolyphaseConnections ..........................................................38

7. DATAVIEW® SOFTWARE ................................................................ 397.1 InstallingDataView® ................................................................397.2 OpeningtheControlPanel .....................................................437.3 UsingtheControlPanel ..........................................................457.4 ConfiguringtheDTR® 8510 ....................................................477.5 RunningaTest ........................................................................497.6 DownloadingaTest ................................................................507.7 SavingtheMeasurementRecords .........................................517.8 ReportGeneration ..................................................................51

8. MAINTENANCE ............................................................................. 538.1 ChargingtheBatteries ............................................................538.2 Cleaning ..................................................................................56

APPENDIX A: DISPLAY MESSAGES ..................................................... 57RepairandCalibration ...........................................................................60TechnicalandSalesAssistance ............................................................60Limited Warranty ...................................................................................61WarrantyRepairs ...................................................................................61

Page 5: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 3

CHAPTER 1

INTRODUCTION

WARNING Thesesafetywarningsareprovidedtoensurethesafetyofpersonnelandproperoperationoftheinstrument.

This instrument is protected from accidental voltages of not•morethan50Vwithrespect toearth.Theguaranteed levelofprotectionofthisequipmentmaybecompromisedifusedinamannernotspecifiedbythemanufacturer.Readtheinstructionmanualcompletelyandfollowallsafety•informationbeforeattemptingtouseorservicethisinstrument.TheDigitalTransformerRatiometerDTR• ® Model 8510 is designedforuseonde-energized (“dead”) transformersonly.MakesurethetestsampleiscompletelydisconnectedfromACpowerandisfully discharged.OnlyqualifiedpersonnelshouldusetheDTR• ® 8510 .TheDTR• ®8510mustnotbeusedinamannerinwhichanyofitscomponents(includingtestcables)arereliedupontoprovideprotectionfromelectricshock.Nohighvoltageinsulation/protectionisprovidedbyanycomponentoftheDTR® 8510. Always make sure the circuit is fully discharged before attaching any test cables.Donottouch,adjust,orrepositiontestcableswhiletheDTR• ® is conductingatest.Usecautiononanyapparatus:potentiallyhighvoltagesand•currentsmaybepresentandposeashockhazard.Safetyistheresponsibilityoftheuser.•Onlyusethechargingunitsuppliedwiththeinstrumentto•rechargethebattery.NeveropentheinstrumentwhileitisconnectedtoACpower•orwhentestcablesareconnectedtotransformers,equipment,circuits,etc.

Page 6: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

4 Digital Transformer Ratiometer DTR® Model 8510

1.1 International Electrical Symbols

Signifies that the instrument is protected by double or reinforced insulation.

CAUTION - Risk of Danger! Indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument in all cases where this symbol is marked.

Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

Ground/Earth

Important instructions to read and understand completely.

Important information to acknowledge.

Battery

Fuse

USB socket

Compliance with the Low Voltage & Electromagnetic Compatibility European directives (73/23/CEE & 89/336/CEE)

In the European Union, this product is subject to a separate collection system for recycling electrical and electronic components In accordance with directive WEEE 2002/96/EC

1.2 Definition of Measurement CategoriesCAT I: FormeasurementsoncircuitsnotdirectlyconnectedtotheACsupply

walloutletsuchasprotectedsecondaries,signal level,and limitedenergycircuits.

CAT II: Formeasurementsperformedon circuits directly connected to theelectricaldistributionsystem.Examplesaremeasurementsonhouse-holdappliancesorportabletools.

CAT III: Formeasurementsperformedinthebuildinginstallationatthedistri-butionlevelsuchasonhardwiredequipmentinfixedinstallationandcircuitbreakers.

CAT IV: For measurements performed at the primary electrical supply(<1000V)suchasonprimaryovercurrentprotectiondevices,ripplecontrolunits,ormeters.

Page 7: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 5

1.3 Receiving Your ShipmentUponreceivingyourshipment,makesure that thecontentsareconsis-tentwith thepacking list.Notifyyourdistributorofanymissing items. Iftheequipmentappearstobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Savethedamagedpackingcontainertosubstantiateyourclaim.

1.4 Ordering InformationDigital Transformer Ratiometer Model 8510 (115V) ........Cat. #2136.50Includes NiMH batteries (installed), 115V power cord, set of 15 ft cables in a carrying bag, an external battery charger (90-264VAC 50/60Hz), DataView® software CD-ROM, warranty card and user manual.

1.4.1 Accessories and Replacement Parts

Fuse-setof5,0.5A250V(5x20mm,SlowBlow) ............... Cat. #2118.53

Fuse-setof5,4A125V(5x20mm,SlowBlow) .................. Cat. #2118.55

Lead-Setof2,30ft ............................................................Cat. #2136.76

Lead-Replacementsetof2,15ft ......................................Cat. #2136.77

Battery-Replacementsetof2,12VNiMHrechargeable....Cat. #2136.78

ReplacementBatteryCharger .............................................Cat. #2136.79

Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com for availability

Page 8: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

6 Digital Transformer Ratiometer DTR® Model 8510

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionTheDigital TransformerRatiometerDTR®Model 8510 is a lightweight,rugged,portableinstrumentdesignedforonsitetestingofpower,potentialandcurrenttransformers.

OperationoftheDTR®8510isfullyautomatic.Nousercalibration,rangeselection,handcrankingortediousbalancingisrequired.

During each test cycle, the DTR® 8510 automatically checks for:

H/Xleadreversal•

Continuityoftestedcircuits/windings(ifconfigured)•

Shortcircuitconditions(highcurrent)•

Upon completion of a test cycle, the DTR® 8510 displays:Turns Ratio:• Theratiooftheprimarytosecondaryvoltageatthetransformerterminalsduetotestexcitation.

Excitation Current: • TheRMSexcitationcurrentintheHwindingduetotestexcitationduringnegligibleloadingoftheassociatedXwinding.

Polarity:• Indicatesthepolarity(phase)ofXrelativetoH.

Deviation: • Indicatesthedeviationfromnameplateratioin%.

TurnsRatio,ExcitationCurrent,PolarityandDeviationareusefulparam-etersindiagnosingandpredictingavarietyoffaultsthatoccurinpower,potentialandcurrenttransformers.

Page 9: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 7

The DTR® 8510 will display the following messages:IncorrectLeadConnections•H/XReversal(accidentalstep-upmisconnection)•Short(excessexcitationcurrent)•OpenCircuits•CircuitContinuity•LowBattery•

Inaddition,theDTR®8510allowstheusertostorethedatainautomaticormanualmodeaftereachtesteliminatingtheneedtowritedownthetestresults.Eachmeasurement record is date and time stampedprovidingcompletetestinformation.

TheusercanalsosaveNameplatevoltagesandcomparetheresultsasthedataisbeinggathered.ThedatacanlaterbedownloadedtoaPCandanalyzedusing theDataView®softwarepackagewhich is includedwiththeproduct.

DataView®allowsfullcontroloftheinstrument.

The DTR® 8510 utilizes an advanced, low-voltage, step-down measure-ment technique in which the high voltage “H” windings are subjected to test excitation. This results in greater operator safety and the ability to test a much wider array of transformer types and sizes.

Page 10: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

8 Digital Transformer Ratiometer DTR® Model 8510

2.2 Control Features

50V CAT IV

OFF/CHARGE

ON

ENTER

MEM

TEST

CONTRAST

DIGITAL TRANSFORMER RATIOMETERDTR® MODEL 8510

USE SPECIFIEDCHARGER ONLY

USE ONLY ON DE-ENERGIZED TRANSFORMERS

H

X

52

1

3

4

6

7

8

10

11

9

1. Display: Displaysdata,statusandcontrolfeaturesoftheinstrument.

2. USB Connector: Allowsconnectiontoacomputerforinstrumentconfigurationset-upandstatuscheck,downloadingofthestoreddatausingtheDataView®softwareandrunningatest..

3. High-side “H” Cable Connector (primary):Connectionfortheprimarysideofthetransformer.

4. Low-side “X” Cable Connector (secondary):Connectionforthesecondarysideofthetransformer.

5. Battery Charge Input Connector:Allowsthesmartchargertochargethebatteries.

6. Power Switch: TurnstheinstrumentONorOFF(ifthechargerisnotconnected).Ifthechargerisconnected,thebatterieschargeinOFF/CHARGEposition.

7. Operation and Battery Low Indicator: GreenLEDindicatesthattheinstrumentisonandblinkswhenthebatteriesaregettinglow(<12V).Theinstrumentwillcompletelyshutdownwhenthebatteriesarebelow8.7V.

Page 11: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 9

8. Display Contrast Adjustment: Allowsadjustmentofthedisplaycontrast.

9. Function Keys:Allowsnavigationofthemenuandfunctionsoftheinstrument.

10. Backlight ( ) Control Button:TurnsthedisplaybacklightONorOFF.

11. Test Button: Runstheselectedtestwhenpressedandreleased.

2.3 Cable Identification

H CABLE SET

RED INLINE PLUG

BLACK INLINE PLUG

"H" IDENTIFIER

5 PINCONNECTOR

"H" IDENTIFIER

X CABLE SET

RED INLINE PLUG

BLACK INLINE PLUG

"X" IDENTIFIER

3 PINCONNECTOR

"X" IDENTIFIER

HH

XX

Primary (H) Cable:

Secondary (X) Cable:

Eachcableisclearlymarked.ThePrimary(H)cablehasa5-pinconnectorand the Secondary (X) cable has a 3-pin connector. They cannotinadvertently be connected incorrectly.

Page 12: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

10 Digital Transformer Ratiometer DTR® Model 8510

CHAPTER 3

SPECIFICATIONS

Reference Conditions: 23°C±5°C (30 to50%RH) range.Add25ppm/°C from-10° to18°Cand28°to50°Ctoallaccuracyspecifications.Noexternalelectricalormagneticfields.Outputcurrent≤150mAforVT/PTand≤50mAforCT.Calibrationcycleis1year.

3.1 DTR® 8510 Specifications

DTR® 8510

ELECTRICALRatio Range (VT/PT) Autoranging: 0.8000 to 8000:1

Accuracy (VT/PT) Ratio Range Accuracy (% of Reading)

0.8000 to 9.9999 ± 0.2%

10.000 to 999.99 ± 0.1%

1000.0 to 4999.9 ± 0.2%

5000.0 to 8000.0 ± 0.25%

Ratio Range (CT) Autoranging: 0.8000 to 1000.0

Accuracy (CT) Ratio Range Accuracy (% of Reading)

0.8000 to 1000.0 ± 0.5%

Excitation Signal VT/PT Mode: 32Vrms maxCT Mode: Auto Level 0 to 1A, 0.1 to 4.5Vrms

Excitation Current Display

Range: 0 to 1000mAAccuracy: ± (2% of Reading + 2mA)

Excitation Frequency 70Hz

Display Dual line alpha-numeric LCD, 16x2 characters with contrast adjustment and backlight control. Day/night visible.

Measurement Method In accordance with IEEE Std C57.12.90™-2006

Power Source Two 12V, 5x2, 1650mAH NiMH rechargeable battery packs

Battery Life Up to 10 hrs of continuous operation. Low battery indication.

Battery Charger Universal input (90 to 264Vrms input) smart recharger

Charging Time <4 Hrs

Data Storage 99 objects with 99 tests each

Date/ Time Battery-backed, Real-time clock

Page 13: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 11

Communication USB. 2.0 compliant, optically isolated, 115.2 KB

Software DataView® analysis software included

MECHANICALDimensions 10.70 x 9.76 x 5.12" (272 x 248 x 130 mm)

Weight 8.1 lbs (3.7kg)

Connection XLR connectors

Leads 15 ft (4.6m) H & X shielded with color-coded large industrial

alligator clips in carrying bag

Enclosure Heavy duty Polypropylene case, UL 94 V0

Vibration IEC 68-2-6 (1.5mm to 55Hz)

Shock IEC 68-2-27 (30G)

Drop IEC 68-2-32 (1m)

Index of Protection IP 40 (Instrument lid open) per EN 60529IP 53 (Instrument lid closed) per EN 60529

ENVIRONMENTALOperating Temperature 14° to 122°F (-10° to 50°C)

Storage Temperature -4° to 140°F (-20° to 60°C)

Relative Humidity 10 to 85% RH @ 35°C

Altitude Up to 2000 meters (6560 ft)

SAFETYSafety Rating EN 61010-1; 50V CAT IV;

Pollution Degree 2

Double Insulated Yes

CE Marking Pending

Specifications are subject to change without notice.

Page 14: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

12 Digital Transformer Ratiometer DTR® Model 8510

3.2 Battery Charger Specifications

DTR® 8510 BATTERY CHARGER

ELECTRICALNo-load Voltage 41V ±2V

Fast Charge Current 0.9A ±70mA

Top-off Charge 130mA ±40mA

Charge Termination Options -dV, dT/dt

Input Rating 90 to 264VAC /47 to 63Hz

Maximum Output Power 35W

Trickle Charge Current 50mA ±25mA

Leakage Current (from battery with mains off) <1mA

MECHANICALDimensions 4.22" x 2.64" x 1.44" (107 x 67 x 36.5mm)

Weight 0.55 lbs (250g)

Input Connection 2-pin IEC 320-C7

Output Connection NiMH: 3 PIN DIN

ENVIRONMENTALOperating Temperature -4 to 104°F (-20° to 40°C)

SAFETYInsulation Class II

Electrical Safety Approval EN 60601-1, EN 60950, EN 60335-2-29

EMC Standards EN 61000-6-3 (Emission), EN 61000-6-1 (Immunity)

Specifications are subject to change without notice.

Page 15: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 13

CHAPTER 4

DISPLAY FUNCTIONS

4.1 Program FlowInordertousetheinstrumentefficiently,itisimportanttounderstandtheprogramflow.

Pressingthe▼and▲keyssimultaneouslyallowsgoingupone•level(ifthereisalevel)intheprogramflow. (Verticalnavigationmovingupalevel).Pressingthe▼or▲keyallowsnavigationwithinthelevel. •(Horizontalnavigation).Pressingthe• ENTERkeyallowstheselectionanddisplayofchoicesonelevelbelow(ifthereisalevel)intheprogramflow. (Verticalnavigationmovingdownalevel).

4.2 Top Level MenuTheinstrumenthasatoplevelcontrolasshowninFig.4-1.Theusercannavigatearoundtheringbypressingthe▼or▲key.

Main Screen

Recall DataConfiguration

Figure 4-1

Oncetheinstrumentisturnedon,itinitializesandthendisplaystheMainScreen:

Figure 4-2

Page 16: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

14 Digital Transformer Ratiometer DTR® Model 8510

Main Screen: • DisplaystheTestType,TestMode,Filter[Fast(F),Normal(N)orSlow(S)],StorageMode[Auto(A)orManual(M)]andthePresentNameplatevalues(ifenabled).ConfigureInstrument:• Configurestheinstrument(see§4.3).Recall Data: • Storeddatacanbeselectedanddisplayed(see§4.4).

NOTE: The program flow is two or three levels deep at various points. The available choices for a level can be viewed by pressing the ▼or ▲ keys.

TheENTERkeyisusedtoselectthefollowingfunctions:Enterconfigurationmenu.•EnterRecallDatamenu.•Initiatestorageoftestresults,acceptlocationparametersandstore•theMeasurementRecord.

4.3 Configure InstrumentWith the display showing Configure Instrument, press theENTER key.Theconfigurationring(seeFigure4-4)canbenavigatedaroundusingthe▲or▼key.

Figure 4-3

Set Clock: • AllowsthesettingofTimeandDate.Setup Nameplate: • AllowsselectionandeditingofPresentNameplatevoltagesorratios.PredefinedratioscanonlybeeditedthroughDataView®.Select Test Type: • AllowstheselectionofTestType(VT/PTorCT).

Select Test Mode: • AllowsselectionoftheTestMode(RatioorContinuity/RatioinVT/PTMode).

Select Storage Mode: • Allowstheselectionofdatastoragemode(AutomaticorManual).

Page 17: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 15

Select Filter: • Allowsselectionoffilter(Fast,Normal,Slow). See§4.3.6.

Erase Memory: • Allowsdeletionofallthestoredmeasurementrecords

Select Language: • Allowsselectionoflanguageonthedisplay(English,French,German,Italian,Spanish,Portuguese)

NOTE: To move one level up and return to Configure Instrument, press the ▼▲ keys simultaneously.

SetClock

SelectStorage Mode

SetupNameplate

SelectLanguage

SelectTest TypeErase Memory

SelectTest Mode

SelectFilter

Figure 4-4

Page 18: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

16 Digital Transformer Ratiometer DTR® Model 8510

4.3.1 Set Clock

1. Within the top levelofConfigure Instrument,press the▼or▲keyuntilSet Clock appearsonthedisplay,thenpressENTER.

Figure 4-5

TheSet Clock configurationallowsyoutoformatthefollowing:

Date Format:Allowstheformattingofthedate. (MM/DD/YY,DD/MM/YY,YY/MM/DD).

Figure 4-6

Time Format:Allowstheformattingofthetime(12or24hours).

Figure 4-7

Set Date and Time:Allowsthesettingofthedateandtimeintheselected format.

Note: Timemustbesetinthe24-hourformat.ItwillbedisplayedwithAM/PMifthe12-hourformatischosen.

Figure 4-8

Example: Tosetthetimeto2:50PM,setthehourto14andtheminutesto 50.

Page 19: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 17

Tomaneuverbetween theSet Clockconfigurationsettings,performthefollowing:

1. Within the top levelofConfigure Instrument,press the▼or▲keyuntilSet Clockappearsonthedisplay,thenpressENTERtoconfigurethedate.

2. Use the▼or▲keys to view the choices.PressENTER tomakeaselection.Oncetheformatisselected,thedisplayreturnstoDate Format.

3. Usethe▼or▲keystonavigatetotheotherdateandtimeoptionsorpressboth▼▲keystogethertoreturntothetoplevel.

4.3.2 Setup Nameplate

Nameplatevaluesareprimaryandsecondaryvoltages(orratios)associatedwith a given transformer.This information is available on theNameplateof the transformer.Themeasured resultswillbecomparedwithselected(edited)NameplatevaluesandDeviationwillbedisplayedandstoredin%relativetothesevalues.

TheDTR® 8510allows the storageof 10Nameplate voltages or ratioswhichcanbeindividuallyselectedforagiventest.

NOTE:TheNameplatelistofupto10transformerscanbewrittentoand•savedonlyusingDataView®software.Thevaluesinthelistcanbeeditedforusebutcannotbesavedfromthefrontpanel.Thekeysonthefrontpanelcanselectanyoneofthe10•transformerNameplatevalues.Oncethetransformervaluesareselected,theycanbeeditedtoanewvalueusingthe▼,▲andENTERkeys.Whenthenewvaluesareselected(eveniftheyarenoteditedtonewvalues),itbecomesthePresentNameplateandwillbeusedforallsubsequentVT/PTandCTmeasurements.Deviationwillbereportedandstoredusingthesevalues.ThePresentNameplatevalueswillberetainedinmemoryeven•iftheinstrumentisturnedOFFandturnedON.However,itwillnotbesavedinthelistfromwhichitwasoriginallyselected.TheusercanselectanotherNameplatefromthelistinthe•futureandmayormaynoteditthevalues.AssoonasthenewNameplatevaluesareselected(and/oredited),itbecomesthenewPresentNameplate.

Page 20: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

18 Digital Transformer Ratiometer DTR® Model 8510

ThePresentNameplatevaluesaredisplayedintheMain•Screen.EachMeasurementRecordwillrecordthePresentNameplate•alongwithotherassociatedparametersandwillbeavailablewhenthedataisdownloaded.TheRatioofthenameplatevaluesmustbebetween1and•32767.TheNameplateneedstobeenabled.IftheNameplateis•notenabled,theDeviation%willnotbereportedafterthemeasurements.Subsequently,thestoredvalueswillnothaveNameplateandDeviation%intheMeasurementRecord.

NOTE: The Main Screen will display the Present Nameplate only if it is Enabled. If it is not Enabled, it will display "---" for Primary and Secondary.

ToconfigurethesettingsfortheNameplate,performthefollowing:

1. Within the top levelofConfigure Instrument,press the▼or▲keyuntilSetup Nameplate appearsonthedisplay,thenpressENTER.

Figure 4-9

2. ThedisplaywillshowEnablewithYes and Nochoices.Pressthe▼keytoselectYes(EnableNameplate).

Figure 4-10

Page 21: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 19

3. Thedisplaywillnowshowthefollowingscreen:

Figure 4-11

4. Select Listbypressingthe▼key.Thelistcanthenbescrolledfrom0to9Nameplatevaluesbyusingthe▼or▲key.SelectthedesiredNameplate by pressing ENTER. The display will return to Setup Nameplate.

Figure 4-12

5. PressENTER,thenYes(▼)onEnable.

6. Select Edit bypressingthe▼key.ThevaluesinthePresentName-platecanbeediteddigitbydigitbyselectingeachdigitwiththe▼or▲andthenpressingtheENTERkey.WhenthelastdigitontheSec-ondaryisselected,thedisplayreturnstoSetup Nameplate.

Figure 4-13NOTE:

Selectingonetransformerfromthelistwillreplacethevalues•inthePresentNameplateimmediately.ThevaluescanthenbechangedintheEditmodeifneeded.Selecting• EditwillallowtheusertomodifythevaluesinthePresentNameplate.The• Edit processchangesthevaluesinthePresentNameplateonly.Thevaluespreviouslystoredwouldremainthesame.Thestored List valuescanonlybechangedinDataView®.

Page 22: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

20 Digital Transformer Ratiometer DTR® Model 8510

InordertowritethePresentNameplatevalue,youmustpress•ENTERuntilSetup Nameplateisdisplayed.Otherwise,theratiowillnotbeentered.Iftheup/downarrowsarepressedaftersettingtheratio,thenitperformsacanceloperationandthechangeswillnotbesaved.

REMEMBER: In order to move one level up and return to Configure Instrument, press the ▼▲ keys simultaneously.

4.3.3 Select Test Type

1. FromthetoplevelofConfigure Instrument,pressthe▼or▲keyuntilSelect Test Typeappearsonthedisplay,thenpressENTER.

Figure 4-14

2. Select VT/PT or CTusing▼or▲andpressENTER.ThedisplaywillreturntoSelect Test Type.

NOTE: Once the Test Type is selected, it is retained until changed by the user. If the instrument is turned OFF and turned ON, the Test Type will be the same as last selected.

4.3.4 Select Test Mode

1. FromthetoplevelofConfigure Instrument,pressthe▼or▲keyuntilSelect Test Modeappearsonthedisplay,thenpressENTER.

Figure 4-15

2. Select Ratio or CONT/Ratiotestusing▼or▲andpressENTER.ThedisplaywillreturntoSelect Test Mode.

Page 23: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 21

CONT/Ratio:PerformsaContinuitytestfollowedbyRatiotestonlyinVT/PT.Continuitytestisnotavailable(N/A)inCTmode.

Ratio test:PerformsonlytheRatiotestintheVT/PTorCTmode.

NOTE: Test Mode can also be selected from the Main Screen by press-ing ENTER, which toggles the selection of CONT and Ratio.

4.3.5 Select Storage Mode

1. FromthetoplevelofConfigure Instrument,pressthe▼or▲keyuntilSelect Storage Modeappearsonthedisplay,thenpressENTER.

Figure 4-16

2. Select Auto or Manualusing▼or▲andpressENTER.ThedisplaywillreturntoConfigure Instrument > Select Storage Mode.

NOTE:In • Manual mode,theuserhasachoiceattheendofatestwhethertostoretheMeasurementRecordornot.TheuseralsohasachoiceofwheretostoretheMeasurementRecord.SoitispossiblefortheusertostoretheMeasurementRecordatvariouslocationsleavinggapsinthestoragememorymap.AllMeasurementRecordswillbedownloadednomatterwheretheyarestoredinthememory.In • Automode,theinstrumentwillstoretheMeasurementRecordattheendofateststartingwiththefirstavailablelocation.AllthesubsequentMeasurementRecordswillbestoredsequentially. IfthepreviouslystoredMeasurementRecordisencountered,theAutostoremodewillskipoverandstorethenewMeasurementRecordinthenextavailablelocation.

NOTE: If the DTR® 8510 is connected to a computer, Auto mode will be disabled. The Measurement Record will not be saved in the instrument but will be sent to the computer.

Page 24: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

22 Digital Transformer Ratiometer DTR® Model 8510

4.3.6 Select Filter

1. FromthetoplevelofConfigure Instrument,pressthe▼or▲keyuntilSelect Filterappearsonthedisplay,thenpressENTER.

Figure 4-17

2. Select Fast, Normal or Slowusing▼or▲andpressENTER.ThedisplaywillreturntoConfigure Instrument > Select Filter.

NOTE: The DTR® 8510 offers the user a choice of three digital filters for measurement, depending upon the external electrical noise. These are sophisticated digital filters in the design which offer maximum noise rejection for repeatable measurements.

Digital Filters:

ThedigitizationofsignalisdonewithSigma-Delta(Σ-Δ)modulator.Theoutput isahigh frequencybitstreamwhich is fed toadigitalfilter.Thefilteringofsignal isperformedwithsophisticatedsinc (Sinx/x)filter.Theprimaryfunctionofthefilteristoremovethequantizationnoiseintroducedduringthemodulationprocess.

Thefilterlimitsthebandoftheallowedfrequencytosignificantlylowerthanone-halfof the frequencyof themodulator itself.This reduces the1-bitoutputofacomparatorinaband-limited,verylownoiseoutput.Thecut-offfrequencyandthedecimateddatarateareappropriatelyprogrammedintheinternalregisteredandtheoutputisarithmeticallyaveragedtoprovideanaccuratefinalanswer.

Highdecimationresultsinlowernoisebutataslowerthroughputrate.Thecalculationofarithmeticmeanisaquickprocessathighclockspeedsofthemicroprocessors.

TheDTR® 8510 improves thequality ofmeasurements in the followingways:

(1) Lownoiseelectronics(2) Shieldedelectronicsandcables.(3) Differentialmeasurementstoreducecommon-modeeffects.(4) Ratiometriccalculationstoreducedynamicerrors.

Page 25: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 23

VT/PT

Fast Normal SlowApprox Measurement Time 6.5 S 9 S 19 SUses Electrically

Quiet Environment

In Most Applications

Electrically Very Noisy

Environment

CT

Fast Normal SlowApprox Measurement Time 14 S 16 S 43 SUses Electrically

Quiet Environment

In Most Applications

Electrically Very Noisy

Environment

4.3.7 Erase Memory

1. Press the▼or▲keyuntilErase Memoryappearson thedisplay,thenpressENTER.

Figure 4-18

2. Select Yes or Nousing▼or▲andpressENTER.ThedisplaywillreturntoErase Memory.

CAUTION: Once the memory is erased, the stored data cannot be recovered. Download the data, if desired, before the Erase Memory function is executed.

Page 26: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

24 Digital Transformer Ratiometer DTR® Model 8510

4.3.8 Select Language

1. Pressthe▼or▲keyuntilSelect Languageappearsonthedisplay,thenpressENTER.

Figure 4-19

2. Selectalanguageusing▼or▲andpressENTER.ThedisplaywillreturntoSelect Language.

Thefollowinglanguagesareavailable:-English -Italian-French -Spanish-German -Portuguese

Oncethelanguageischosen,allthemenuitems,errormessagesetc.willbedisplayedinthislanguage.TheconfigurationretainsthelanguageiftheinstrumentisturnedOFFandthenturnedON.

NOTE: The language can also be selected immediately at start up. Press and hold the TEST button while turning the unit ON. After initialization, the instrument will display the currently active language. Release the TEST button. The desired language can now be selected.

Page 27: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 25

4.3.9 Restoring Factory Defaults

Thefactorydefaultsettingsareshowninthefollowingtable:

Parameter Default ValueTest Mode RatioTest Type VT/PTFilter NormalStorage Mode ManualLanguage EnglishNameplate Disable

Nameplate Values

19920:720019920:480019920:240019920:216019920:1207200:24007200:21607200:1202400:1202160:120

Table 4-1

Tosettheinstrumentbacktothefactorydefaultconfiguration:

1. TurntheDTR® 8510 OFF.

2. Pressandholdthe▲button.

3. TurntheDTR® 8510 ON.

4. Releasethe▲buttonwhentheinstrumentbeeps.

NOTE: The previously stored Measurement Records will remain in the memory and the language will remain the same as before the default configuration was loaded.

Page 28: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

26 Digital Transformer Ratiometer DTR® Model 8510

4.4 Recall Data1. FromtheMainScreen,pressthe▼or▲keyuntilRecall Dataappears

onthedisplay,thenpressENTER.

Figure 4-20

TheMeasurement Record in OBJECT #01 and TEST #01 will be dis-played,aswellastheRatio.

NOTE: The V after the Ratio indicates the VT/PT test type. C would be displayed if the Ratio was from a CT test.

Figure 4-21

2. Pressthe TESTbuttontomovetosecondscreen,whichwilldisplayCurrentandDeviation%.

Figure 4-22

3. PresstheTESTbuttontomovetothethirdscreentodisplaytheDateand Time.

Page 29: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 27

Figure 4-23

4. Press theTEST button tomove to the fourth screen todisplay thePrimaryandSecondaryvoltagesofthePresentNameplateatthetimethemeasurementwasmade.

Figure 4-24

5. PresstheTESTbuttonagaintoreturntothefirstscreen.

6. SelectanyOBJECTandanyTESTusing the▼or▲keyand theENTERkeytoviewtheMeasurementRecordatanylocation.

NOTE:TheDTR• ®8510canstore99OBJECTs.EachOBJECTcanstoreupto99TESTs.In • Manualmode,theusercanselectthelocationtostoretheMeasurementRecord.TheMeasurementRecordsdonothavetonecessarilybesequentiallystored.So,duringaRecallDataoperation,itisimportantfortheusertoknowwherethedataisAppendixBonpage59providesaformfortheusertoidentify•theinformationstoredateachlocation.Theformcanbecopiedandusedasneeded.

Page 30: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

28 Digital Transformer Ratiometer DTR® Model 8510

CHAPTER 5

OPERATION

WARNING: The DTR® Model 8510 is designed for use on de-energized (“dead”) transformers only. Make sure the test sample is completely disconnected from AC power and is fully discharged.

5.1 Power UpTurnthepowerswitchtotheONposition.

Atpowerup,theinstrumentgoesthroughaninitializationprocess:Beepsonce•Displays“Initializing…PleaseWait”forashortduration•DisplaystheCompanyNameandFirmwareversionnumber•DisplaysDateandTimeforashortduration•TheinstrumentrestsattheMainScreenwhichdisplays•TestType,TestMode,Filter,StorageModeandthepresentNameplatevalues(Fig.5-1)TheGreenpowerLEDonthefrontpanelturnsonsolid•

Figure 5-1NOTE:

Ifthebatteriesareinsufficientlycharged:•-TheGreenLEDwillblinkifthereissomebatterycapacityleft.Theinstrumentwillbeoperationalunderthiscondition.

-ThedisplaywillshowLowBatteryandtheGreenLEDwillblinkifthebatterypackshavenocapacityleft.Theinstrumentcannotbeusedfortestingunderthesecircumstances.Thebatteriesmustberechargedbeforeuse(see§8.1).

Page 31: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 29

5.2 Running a Test

5.2.1 VT/PT Test

Inthissection,anexampleofrunningatestusingthefollowingconfigura-tion(see§4.3forhowtoconfiguretheinstrument)isshown:

SetDate/Time:July4,2010.6:00PM•SetvaluesinNameplate: •PrimaryVoltage:19920;SecondaryVoltage:7200SelectType:VT/PT•SelectTestMode:CONT/Ratio•StorageMode:Manual•Set Filter: Normal•SetLanguage:English•

PRIMARY SECONDARY

19920 : 7200CTH RED X RED

X BLACK

H BLACK

RATIO 2.767 : 1- Typical Connection

H BLACK

PRIMARY SECONDARY

19920 : 7200CT

H RED X RED

X BLACK

H RED

X RED

X BLACK

H BLACK

RATIO 1.383 : 1

- Alternate Connection

H BLACK

H RED

X BLACK

X RED

Figure 5-2

Page 32: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

30 Digital Transformer Ratiometer DTR® Model 8510

1. Connect thePrimary(H)andSecondary(X)cablesto theappropri-ate connectors on the DTR® 8510 and the transformer under test (seeFig.5-2).

2. PressTESTbutton.Thetestwillruncontinuityfirstandthenruntheratio test.

Figure 5-3

WARNING: It is important to make sure the Primary (H) and the Secondary (X) cables are connected properly BEFORE running the test. This is especially true for the high turn-ratio transformers since dangerous high voltage may appear on the DTR® 8510 terminals.

NOTE:Continuitychecksfortheintegrityofthecableconnectionsonly.•ItdoesnotcheckforH/Xreversepolarity.IfthecontinuityforthePrimaryconnectionsiswithinspecifiedlimits,thedisplayshows----orequivalenttoaClosedSwitch.IfthecontinuityforthePrimaryconnectionsisoutsidethespeci-fiedlimits,thedisplayshows--/--orequivalenttoanOpenSwitch.During theRatio test, the instrumentwill check forH/X reverse•polarity, short (excessive current) etc. before conducting Ratiotest.

Inthisexample,thedisplaywillappearasshownbelow.

Figure 5-4

Page 33: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 31

If theH/Xwere reversed, the instrumentwill display the followingerrormessageandterminatethetest.

Figure 5-5

NOTE:Iftherewereapolarityreversal,theRatiowouldbedisplayed•negativeanditwouldbeblinking.Thismeanseithertheconnectionsarereversedonagivensideorthetransformeriswoundbackwardsrelativetothemarking.Thetestmayberunasmanytimesasneededbypressingthe•TESTbutton.Anactivetestcanbecancelledatanytimebypressingthe•TEST button.MakesurethattheH• REDleadisalwaysconnectedsuchthatitdoesnotshorttotheXREDorXBLACKleaddirectlyorthroughearthground.See§6formoreconnectiondetails.

5.2.2 Storing the Measurement Record (Manual and Auto Mode)

Thestoragemodeyouareindependsonwhatyouselectedwhenconfig-uringtheinstrument(see§4.3)beforerunningatest.

5.2.2.1 Manual Mode

When Manual mode is selected, the Measurement Record storage isoptionaltotheuser.

1. TostoretheactivemeasurementusingManualmode,pressthe MEM (ENTER)button.Thedisplaylooksasshownbelow:

Figure 5-6

Page 34: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

32 Digital Transformer Ratiometer DTR® Model 8510

2. Select theOBJandTest locationusing the▼or▲keyandpressENTER.

3. PressENTER or TESTtostoretheMeasurementRecord.

NOTE:Afterthetestisrun,thenewlocationisautomaticallyincreased•sequentiallyfromthelaststoredlocation.Ifagivenlocationisnotavailableforstorage,theSavefieldwill•displayN/A.InordertoNOTSTOREtheMeasurementRecord,simplypress•theTESTbuttonafterthemeasurementiscompleted.

5.2.2.2 Auto Mode

WhenAutomodeisselected,theinstrumentrequiresnouserinput.

1. Attheendofthetest,theMeasurementRecordisstoredintheloca-tionnexttothefirstavailablelocationandthenwillstoresequentially.

NOTE: If the previously stored Measurement Record is encountered, the Auto store mode will skip over and store the new Measurement Record in the next available location.

5.2.3 VT/PT Excitation Current

Excitationcurrent,displayedinmilli-amperes(mA)istheRMScurrentinthePrimary(H)windingduetotestexcitation.TheloadingontheSecondary(X)windingduetotheDTR®8510isnegligible.

TheDTR• ®8510usesamaximumof32Vrmsexcitationvoltagefortesting.Thedisplayedexcitationcurrentwillalmostalwaysbeasmallfractionoftransformerno-loadexcitationatfull(rated)voltage.Mosttransformerswillrequirelessthan100mA.Ifyouarepresentlyusingaratiometerthatemployssecondary•(X)sideexcitation(e.g.commonhand-crankedinstruments)instep-upmode,thecurrentmaybemuchhigherthanwhattheDTR®8510displays.Thisisnormalandresultsdirectlyfromthefactthatinmost•step-downtransformers,thePrimary(H)sidehasamuchhigherimpedancethanitsassociatedSecondary(X)winding.

Page 35: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 33

5.2.4 VT/PT Continuity Test

TheContinuityTestfunctionoftheDTR®8510servesasausefulmeansofidentifyingopenprimaryandsecondarywindings,highresistancecon-nections,opencircuitbreakersandblownfuses.Whenselected,theDTR® 8510 checks for continuity between itsSecondary (X) andPrimary (H)cables.

TheDTR• ®8510willindicateOPENforcircuits>4KΩandCONTforcircuits<4KΩ.Checkthesevalues.Asmallfractionoftransformers(HighinductanceorhighwindingOhmicDCresistance)mayappearopen.TheinstrumentchecksforSecondary(X)connectionsfirst.Even•ifitisfoundopen,itcontinuestocheckPrimary(H)connections.Regardlessoftheoutcome,theinstrumentproceedstoperformpre-testchecks.Followingthepre-testchecks,ifeitherXorHorbothwerefoundtobeOPEN,theinstrumentwilldisplaytheappropriateerrormessage.PressingtheTESTbuttonagainwillstartanewContinuitytest(assumingtheuserfixedtheconnectionproblem)again.IfitpassestheContinuitytest,Ratiotestwillbeexecuted.Transformerswithwindingarrangementmultiple,parallelpaths•(e.g.,Deltatype)mayreadCONTevenifoneormorewindingsmaybeOPEN.

5.2.5 CT Test

Inthissection,anexampleofrunningatestusingthefollowingconfigura-tion(see§4.3forhowtoconfiguretheinstrument)isshown:

SetDate/Time:July4,2010.6:00PM•SetvaluesinNameplate: •PrimaryCurrent:19920;SecondaryCurrent:7200SelectType:CT•Select Test Mode: Ratio•StorageMode:Manual•Set Filter: Normal•SetLanguage:English•

Note: CTtestdoesnotallowaCONT/Ratiooption.ItonlyrunsaRatiotest.

WARNING: The DTR® Model 8510 is designed for use on de-energized (“dead”) transformers only. Make sure the test sample is completely disconnected from AC power and is fully discharged.

Page 36: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

34 Digital Transformer Ratiometer DTR® Model 8510

1. Connect the Primary (H) cable to the appropriate connectors ontheDTR®Model8510and the transformerunder test.Connect theSecondary(X)cableasaloopthroughthecurrenttransformerholeasshowninthefigure.

X BLACK

H RED

X RED

H BLACK

X2 X1

Figure 5-7

2. PresstheTESTbutton.DuringtheRatiotest,theinstrumentwillcheckforH/Xreversepolarity,short(excessivecurrent)etc.beforeconductingRatiotest.

NOTE:IftheH/Xisreversed,theinstrumentwilldisplaytheerror•message(seeFig.5-5)asinVT/PTtestandterminatethetest.Iftherewereapolarityreversal,theRatiowouldbeblinkingwith•anegativesign.Thismeanseithertheconnectionsarereversedonagivensideorthetransformeriswoundbackwardsrelativetothemarking.Thetestcanbecancelledatanytimebypressingthe• TEST buttonduringtheoperation.TheCTtestrampsupthevoltageinsteps.Attheendofthetest,•itrampsdownthevoltage.Therearecertainlowmass,lowratioCT’sforwhichtheDTR• ® 8510maynotbeabletoprovidecorrectratio.Theerrormessagewillbe“HighExcitationCurrent”.

Page 37: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 35

AsinVT/PT,theCTtestmayberunasmanytimesasneededbypressingtheTESTbutton.

Themeasurementrecordscanbesaved inManual or Auto mode as in VT/PTmode.

TheCTremainsactiveuntil theconfiguration ischanged.Powering theinstrumentdownandturningitbackonwillnotchangetheTestType.

5.3 Tips for Making Precise Ratio Measurements

TheDTR®Model8510isdesignedforstep-downtransformers.Itsourceslow voltage on the primary. It starts with a fraction of the final sourcevoltageandchecksthesecondaryvoltage.Ifitexceedstheprimarybyapre-determinedvalue,thesafetyfeaturestopsthemeasurement.TherecommendedprocedureistoreversetheleadsinthesecircumstancessothattheprimaryofthetransformerisconnectedtotheHcablesandthesecondaryofthetransformerisconnectedtotheXcables.

CAUTION: It is important to check that the H cables are connected to the primary side and the X cables are connected to the secondary side of the transformer before the test begins.

Alwaysverifytheintegrityofthecableconnectionsandreposition•theclipsasnecessaryforrobustandlow-resistanceconnections.Inspectthetransformerterminalsfordielectriccoatings,fungus,dirtor corrosion.

Runningacontinuitytestbeforetheratiotestwillprovideauseful•waytocheckwindingsandconnections.

Whentestingpolyphasetransformers,keepinmindthatinsome•cases,themeasuredratiosmustbemultipliedordividedby√3.Referto§6forpolyphaseconnectiondiagramsandassociatedratioequations.

Page 38: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

36 Digital Transformer Ratiometer DTR® Model 8510

5.4 Ratio Test - 1:1 AsimpletestcanbeconductedtotestfunctionalityoftheDTR® 8510.

1. ConnecttheHREDtoXREDcableandseparatelyconnecttheHBLACKtoXBLACK.Thisconnectionsimulatesa1:1transformer.

2. RunVT/PTtest.

Inthisconnectionmode,thetestresultsshouldyieldarationearlyequalto1.0000.Ifitdoesnot,theDTR®8510mayrequirerepairorre-calibration.

CONNECTIONSEQUIVALENT

H RED X RED

H BLACK X BLACK

1 : 1

Figure 5-8

Page 39: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 37

CHAPTER 6

CONNECTIONS

6.1 Connection DiagramsMakesurethattheHREDleadisalwaysconnectedsuchthatitdoesnotshorttotheXREDorXBLACKleaddirectlyorthroughearthground.

Inthediagramsbelow,thetopthreearegoodconnectionsbutthebottomtwoshouldbeavoided.

Measurements - OK

HR/HB

HB/HR

HR

HB

XR/XB

XR/XB

XB/XR

HR

HB

XR/XB

XR/XB

XB/XR

P

P

S

S

XB/XR

Measurements - NOT OK

HB

HR

XR/XB

HB

HR

XR/XB

XR/XBXB/XR

P

S

XB/XR

Page 40: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

38 Digital Transformer Ratiometer DTR® Model 8510

6.2 Polyphase Connections

A

H3

H2

B

CH1

H0

A

H3

H2

B

CH1

H0

A

H3

H2

B

CH1

H0

AH

3

H2

B C

H1

H2

H1

AH

3

H2

B C

H1

AH

3

H2

B C

H1

AH

3

H2

B C

H1

A

H3

H2

B

CH1

H0

a

X3

X2

b

c

X1

a

X3

X2

b

cX1

X0

aX

3

X2

b c

X1

X2

X1

a

X3

X2

b

c

X1

X0

a

c

X2

X1

b

X3

a

X3

X2

b

c

X1

X0

aX3

X2

b

c

X1

X0

a

c

X2

X1

b

X3

1

2

3

4

5

6

7

8

9

1 Ø

STD

∆ - ∆STD

∆ - ∆REV

∆ - Y

STD

∆ - Y

REV

Y - Y

STD

Y - Y

REV

Y - ∆STD

Y - ∆REV

REF

NO.

XFMR

TYPE

1 Ø H1 - H2 X1 - X2

VH

VX

PHASE

HIGH

VOLTAGE

WINDING

LOW

VOLTAGE

WINDING

TURNS

RATIO

- TRANSFORMER -

HIGH VOLTAGE

WINDING

LOW VOLTAGE

WINDING

H1 - H3(A)

H2 - H1(B)

H3 - H2(C)

X1 - X3(a)

X2 - X1(b)

X3 - X2(c)

A

B

C

H1 - H3(A)

H2 - H1(B)

H3 - H2(C)

X1 - X3(a)

X2 - X1(b)

X3 - X2(c)

A

B

C

H1 - H3(A)

H2 - H1(B)

H3 - H2(C)

X1 - X0(a)

X2 - X0(b)

X3 - X0(c)

A

B

C

H1 - H3(A)

H2 - H1(B)

H3 - H2(C)

X1 - X0 (a)

X2 - X0(b)

X3 - X0(c)

A

B

C

H1 - H0(A)

H2 - H0(B)

H3 - H0(C)

X1 - X0(a)

X2 - X0(b)

X3 - X0(c)

A

B

C

H1 - H0(A)

H2 - H0(B)

H3 - H0(C)

X1 - X0(a)

X2 - X0(b)

X3 - X0(c)

A

B

C

H1 - H0(A)

H2 - H0(B)

H3 - H0(C)

X1 - X2(a)

X2 - X3(b)

X3 - X1(c)

A

B

C

H1 - H0(A)

H2 - H0(B)

H3 - H0(C)

X1 - X2(a)

X2 - X3(b)

X3 - X1(c)

A

B

C

VH

VX

VH

VX

VH • √3

VX

VH • √3

VX

VH

VX

VH

VX

VH

VX • √3

VH

VX • √3

Page 41: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 39

CHAPTER 7

DATAVIEW® SOFTWARE

7.1 Installing DataView®

DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE SOFTWARE AND DRIVERS.

Minimum Computer Requirements:• WindowsXP/WindowsVista&Windows7(32/64bit)• 256MBofRAMforWindowsXP 1GBofRAMforWindowsVista&Windows7(32bit) 2GBorRAMforWindowsVista&Windows7(64bit)

• 80MBofharddiskspace(200MBrecommended)

• CD-ROMdrive

Windows is a registered trademark of Microsoft Corporation in the United States and other countries.

NOTE:When installing, theusermusthaveAdministrativeaccessrightsduringtheinstallation.Theusersaccessrightscanbechangedaftertheinstallationiscomplete.Besuretoselectthe"allusers"optionduringinstallationifdifferentusersneedtoaccessthesoftwareonthesamecomputer.

1. InserttheDataView®CDintoyourCD-ROMdrive. If auto-run isenabled, theSetupprogramwill start automatically. If

auto-runisnotenabled,selectRunfromtheStartmenuandtypeinD:\SETUP (ifyourCD-ROMdrive isdriveD. If this isnot thecase,substitutetheappropriatedriveletter).

NOTE: If installing onto a Vista/Windows 7 computer, the User Account Control dialog box may appear. Select the Allow option to proceed.

Page 42: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

40 Digital Transformer Ratiometer DTR® Model 8510

2. A Set-up window,similartotheonebelow,willappear.

Figure 7-1

Thereareseveraldifferentoptionstochoosefrom.Someoptions(*)requirean internet connection.

DataView, Version x.xx.xxxx• -InstallsDataView®ontothePC.*• Adobe Reader-LinkstotheAdobe®websitetodownloadthemostrecentversionofAdobe®Readertothecomputer.Adobe® Reader is required for viewing PDF documents supplied with DataView® that are accessible from the Help menu.*• DataView Updates-LinkstotheonlineAEMC®softwareupdatestocheckfornewsoftwareversionreleases.*• Firmware Upgrades-LinkstotheonlineAEMC®firmwareupdatestocheckfornewfirmwareversionreleases.Documents• -Showsalistofinstrumentrelateddocumentsthatyoucanview.Adobe®ReaderisrequiredforviewingPDFdocumentssuppliedwithDataView®.

3. DataView, Version x.xx.xxxxoptionshouldbeselectedbydefault.SelectthedesiredlanguageandthenclickonInstall.

4. TheInstallation Wizardwindowwillappear.ClickNext.

5. Toproceed,acceptthetermsofthelicenseagreementandclickNext.

6. IntheCustomer Informationwindow,enteraNameandCompany,thenclickNext.

Page 43: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 41

7. In theSetup Typewindow that appears, select the“Custom” radio buttonoption,thenclickNext.

NOTE: DataView® is common to several AEMC® Instruments products.

8. In theCustom Setup window that appears, select the instrumentcontrolpanelthatyouwanttoinstallbyclickingontheRedXiconandselectingtheoption"Willbeinstalledonlocalharddrive"fromthelist,thenclickNext.

NOTE: The PDF-XChange option must be selected to be able to generate PDF reports from within DataView®.

Figure 7-2

NOTE: Shortcuts for DataView® and each instrument control panel selected during the installation process will be added to your desktop. You can add another instrument at any time.

9. TheDestination Folderwindowappears.KeeptheinstalllocationatthedefaultofProgramFilesandclickNext.

Page 44: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

42 Digital Transformer Ratiometer DTR® Model 8510

10. IntheSelect Program Folderwindowthatappearsnext,selectthe(all users)radiobuttonifmorethanonepersonwillneedtoaccessthesoftwareonthesamecomputer.ClickNext.

11. IntheCompleting the Install Wizard for DataViewwindow,clickonNextandtheinstallationwillbegin.

12. If the instrumentselected for installation requires theuseofaUSBport,awarningboxwillappear,similartoFigure7-3.ClickOK.

Figure 7-3

NOTE: The installation of the drivers may take a few moments. Windows may even indicate that it is not responding, however it is running. Please wait for it to finish.

15. A Questiondialogboxappearsnext.ClickYestoreadtheprocedureforconnectingtheinstrumenttotheUSBportonthecomputer.

16.Next,theInstallation Wizard Completewindowwillappear.ClickonFinish.

NOTE: The Set-up window remains open. You may now select another option to download (e.g. Adobe® Reader), or close the window.

17. Restart your computer, then connect the instrument to theUSBorSerialportonthecomputer.

18.Onceconnected,theFound New Hardwaredialogboxwillappear.Windowswillcompletethedriverinstallationprocessautomatically.

NOTE: If you connected your instrument to the computer before install-ing the software and drivers, you may need to use the Add/Remove Hardware utility to remove the instrument driver before repeating the process.

Page 45: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 43

7.2 Opening the Control Panel

TheControlPanelcanbeopenedfromtheDTRiconorfromDataView®.

To open the Control Panel using the desktop icon:• Double-clickthe DTR Iconthatwascreatedduringinstallation,

locatedonthedesktop.• TheConnectionwindowwillappear.ClickOK.

Figure 7-4

To open the Control Panel from DataView®:

Double-clickontheDataView icononyourdesktop.TheDataView®splashscreen with a logo and relevant information will be displayed for a fewsecondsandthentheQuickStartdialogboxwillappear(seeFig.7-5).

Figure 7-5

Page 46: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

44 Digital Transformer Ratiometer DTR® Model 8510

Select Configure Instrument. The Select Instrument dialog box willappear.

Figure 7-6

Select DTRandpress OK.TheConnectiondialogboxwillappear(seeFig.7-4).

TheConnectionwindowspecifiestheconnected instrument inthecom-municationsdrop-downlist.

Ifmultipleunitsareattached,select thedesiredunitwith itsassociatedserialnumberfromthedrop-downlist.

Oncethedesiredinstrumenthasbeenselected,clickOK andtheDTR®

8510ControlPanelandConfigurationwindowwillopen(seeFig.7-8).

Page 47: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 45

7.3 Using the Control Panel

TheControl Panelisusedforinstrumentoperationandconfiguration.ToopentheControlPanelreferto§7.2.

Figure 7-7

Themainmenuatthetopofthescreenliststhefollowingcommands:

File

Open - OpensaDTR*.icpfile

Save - Savesthereportfileintheappropriatefolder

Save as - Permitstheusertosavethereportatthelocationofhischoice

Create DataView Report - Generatesreportfortheopentestresults

Print Test - Printstheresultofthetest

Print Preview -Providesapreviewoftheselectedtesttobeprinted

Print All Tests - PrintsallthetestsintheselectedObject

Print - Printstheselectedreal-timewindow

Print Preview All Tests -ProvidesapreviewofallteststobeprintedPrint Setup - DisplaysaPrintSetupwindowallowingyoutoselectadestinationprinterandchangevariousprintingoptionsExit - Closesthecontrolpanel

Page 48: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

46 Digital Transformer Ratiometer DTR® Model 8510

EditEdit - Editsthesessioninformationfortheselectedtestintheopened*.icpfileDelete - Deletesthecurrenttestfromthe*.icpfile

View

Toolbar - Allowsselectiontoviewthetoolbar

Status bar - Allowsselectiontoviewthestatusbar

Split - Allowssplittingthescreentoviewtheinformation

Instrument

Connect - Establishesaconnectionwiththeinstrument

Disconnect - Closestheconnectionwiththeinstrument

Configure-Displaystheinstrumentconfigurationwindow

Download - DownloadsrecordeddatastoredintheDTR® 8510

Run Test - Allowsrunningatestfromthecomputer

Help

Help Topics - DisplaystheControlPanelhelptopic

About DTR - DisplaysadialogboxprovidingthecurrentversionnumberoftheDTR®8510andDataView®software

Page 49: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 47

7.4 Configuring the DTR® 8510

ToConfiguretheDTR® 8510, perform the following steps:

1. OpentheControl Panelandconnecttotheinstrument.ToopentheControlPanelreferto§7.2

2. Next, selectConfigure from the Instrument menu or click on theConfigure icon .TheConfigure Instrumentdialogboxwillbedis-played.

Figure 7-8

Date Format: AllowsselectionoftheDateFormatasMM/DD/YYor DD/MM/YYorYY/MM/DD.

Time format: AllowsselectionofTimeformatas24-houror12-hourformat.

Default Test Type: AllowsselectionofVT/PTorCTtest.

Default Test Mode: AllowsselectionoftestforRatioonlyorCont/Ratio(ContinuityandRatio)mode.

Page 50: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

48 Digital Transformer Ratiometer DTR® Model 8510

Default Storage Mode: AllowsselectionofstorageofeachMeasurementRecordinManualorAutomaticmode.

Default Nameplate Values: Allowstheusertostorethemostfrequentlyusednameplatevalues(orratiovalues).

Present Nameplate Values: Allowstheselectionofthepresentnameplatevalues.TheDeviation(%)willbecalculatedbasedonthePresentNameplatevaluesforeachmeasurement.

Set Clock: Allowssettingtheclockontheinstrumentmanuallyorsynchronizedwiththeclockinthecomputer.

Figure 7-9

Erase Memory: Allowsthememorytobecompletelyerased.

Language: Allows the selection of English, French,German, Italian orPortuguese.

Write To Instrument: Allowsthepresentconfigurationvaluestobewrit-tenintheinstrument.

NOTE: None of the changes made on the configuration screen take effect until Write to Instrument command is executed.

Read from Instrument: Allowsthepresentconfigurationtoberead.

Cancel: Cancelstheprocesstomodifytheconfigurationandclosesthedialogbox.

NOTE: Parameters changed after Write to Instrument was executed will be lost.

Page 51: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 49

7.5 Running a Test1. OpentheControl Panelandconnecttotheinstrument.

2. Next,selectRun TestfromtheInstrumentmenuorclickontheRun Test icon onthesideortopmenu.

Figure 7-10

TheusercanselectTest Type, Test Mode, Test Filter, Present Name-plate valuesandwheretheMeasurement Recordwouldbestored.

3. PressStarttorunthetest.Theresult(MeasurementRecord)oftheTestwillbedisplayedinthecenter.

NOTE:IftheDTR• ®8510isinAutosavemode,theconnectiontothecomputerwilldisabletheautomaticsavingoftheMeasurementRecord.IftheTestneedstobecancelledwhiletestingisinprogress,•presstheTESTbuttonontheinstrumentorCANCELbuttononthescreen.IfnewTestType,TestMode,testFilterand/orPresent•Nameplatevaluesareselected,theConfigurationwillbechangedtoreflecttheminthisandthefollowingtests.

4. PressSavetosavetheMeasurementRecordintheselectedObjectandTestlocations,thenpressClosetoclosethedialogbox.

NOTE: Pressing Save in the dialog box only saves the test to the instru-ment, NOT the computer. To save the test on the computer, select Save from the File menu.

Page 52: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

50 Digital Transformer Ratiometer DTR® Model 8510

7.6 Downloading a TestThisselectionpermitsdownloadingofthestoredMeasurementRecordsfromtheinstrument.

To download a test:

1. ClickontheDownload icon inthetoolbar,orclickonRecorded Dataonthesidemenu.TheSelect Testsdialogboxwillappear:

Figure 7-11

This dialog box displays the content of thememory. NoMeasurementRecordsaredownloadedasyet.

2. Select the Measurement Records individually or press Select All. PressDownloadtogetthedatatothecomputer.

Figure 7-12

Page 53: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 51

7.7 Saving the Measurement Records1. Select Save or Save AsfromtheFilemenu.TheSaveAsdialogbox

willappear:

Figure 7-13

2. Savethefileinthedefaultlocationof"MyDocuments\DataView\Data-Files\DTR"orselectadifferentfoldertosaveto.

7.8 Report GenerationTogenerateareport,afileneedstobeopen.

1. Select OpenfromtheFilemenu.Ifnecessary,navigatetotheappro-priatefolderwherethe.icpfileislocated,selectitandpressOpen.

Figure 7-14

Page 54: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

52 Digital Transformer Ratiometer DTR® Model 8510

Oncethefileisopen,selectCreate DataView Report fromthe FilemenuorclickontheCreate DataView Reporticon.Areportwillappearonthescreen.

Figure 7-15

NOTE:Alldownloadedfilesandreportsaresavedinthedefaultlocation•of"MyDocuments"or"Documents"foreachuser.DataView• ®andControlPanel(CP)arefullyintegratedprogramsforinstrumentcontrolanddataanalysis/reportgeneration.CPcandownloadtheMeasurementRecordspreviouslystoredintheDTR®8510memory.TheseMeasurementRecordscanbestoredinthe*.icpformat.Reportscanbegeneratedbyasingleclickonthe“icon”orcreate DataView Report fromthe Filemenu.Whenareportiscreated,adatabasefile*.dvbisautomaticallycreated.DataView• ®canalsocalltheCPortheCPofotherinstruments.DataView• ®cannotopen*.icpfiles.WhenareportiscreatedfromtheCP,areportcanbestored.In•theprocessa*.dvwfileisgenerated.DataView• ®canalsocreateareportfromtheavailabledatabasefiles.A*.dvbfileisautomaticallycreatedwhentheCPcreatesareportfrom*.icpfile.Whenareportiscreated,a*.dvwfileiscreated.Thisreportcan•bedirectlyopenedfromDataView®.

NOTE: For more information about using DataView®, refer to the DataView® manual that is included on the installation CD. This manual can also be found online at www.aemc.com on our software download page for DataView®.

Page 55: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 53

CHAPTER 8

MAINTENANCE

Useonlyfactoryspecifiedreplacementparts.AEMC• ®willnotbeheldresponsibleforanyaccident,incident,ormalfunctionfollowingarepairdoneotherthanbyitsservicecenterorbyanapprovedrepaircenter.

WARNING: To avoid electrical shock, do not perform any service other than the operating instructions contained in this manual, unless you are qualified to do so.

Donotallowwaterorotherforeignsubstancesintothecase.•Disconnecttheunitfromallcircuitsandtestcablesbeforeopening•thecase.Usecautionwithmetallictoolsthatmayshortbatterypacks,power•supplies,etc.

8.1 Charging the BatteriesTheDTR®8510issuppliedwithanexternalsmartchargingunit.InordertochargetheNiMHbatterypacks,theinstrumentswitchmustbeintheOFF/CHARGEposition.

Connectthechargerunitoutputcabletothe3-pinDINconnectoronthefrontpanelof the instrumentandconnect theACpowercord to theACreceptacle.

Thechargingstartsassoonasthepowerisapplied,theprocessisauto-maticandwillterminateonitsownuponcompletion.

CAUTION:Donotuseanyothercharger.•DonotconnectDCpowersupply.Damagewilloccurtothe•instrument.

Page 56: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

54 Digital Transformer Ratiometer DTR® Model 8510

TheDTR• ®8510mustbechargedindoors.Disconnectthechargeruponcompletionofthecharging•process.NiMHbatterymanufacturersdonotallowtricklechargeforalongtimeonarecurringbasis.

Note: TheDTR• ®8510PowerSwitchmustbeintheOFF/CHARGEpositioninorderfortheunittocharge.TheDTR• ®8510cannotbeusedtotestwhileitischarging.Thereisnoindicationonthefrontpanelaboutthestatusof•chargingoritscompletion.TheStatusLEDisonthechargerunititself(seeFigure8-1).

Figure 8-1

Charge Cycle and Status LED Indication on the ChargerLED MODE

Orange Battery not connectedOrange Battery initialization & analysis

Red Fast ChargeGreen with Intermittent Orange Flash Top-off charge

Green Trickle chargeAlternating Red-Green Error

Table 8-1

Page 57: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 55

The smart charger has the following features:

Multiple charge detection methods -dV (typical) and dT/dt. •Typically,theindicationthatthebatteryisfullychargedisdetectedbyaslightdropinitsvoltage(-dV).However,ifthebatteriesareunusedforalongtime,theymaydisplaythesamedropinvoltageintheearlypartofthechargecycleeveniftheyarenotfullycharged.Thisisafalseindication.Inordertopreventtheterminationofchargecycleduetothisfalseindication,thereisanintegraltimerbuiltinthechargerwhichwilldisregardthe–dVforthefirstminutesofthechargecycle.

Thebatterypacksarebuiltwithathermistor.Thesmartchargerconstantlymonitorsthebatterypacktemperature.Abuilt-intemperaturechargecontrol(+dT/dt)securesoptimalcharge.

Safety timer termination. •Incasethebatteriesdonotexhibitsufficient–dV/dtforthechargertodetect,thereisaninternalsafetytimerwhichiscustom-programmedforthebatterypacksusedintheDTR® Model 8510.

Fast charging cycle. •Thischargecyclesuppliesamaximum900mAchargeuntil-dVis detected.

Integral start timer to prevent initial, false -dV detection. •Afewbatterycellsmayhaveavoltagedrop(-dV)inthefirstpartofthechargecycle.Thisisespeciallytrueforbatterycellswhichhavebeenidleforalongperiodoftime.Becauseofthis,astart-timerbuiltintothechargerprevents-dVdetectioninthefirstminutesofthechargercycle.

Reverse polarity/short circuit protection. •Thechargerisprotectedagainstreversepolaritybyanautomaticallyresettablepolyswitchfuseontheoutput.

LED status indicator • (seeTable8-1onpage54).

Universal input voltage. •Withtheuseofaproperpowercord,thechargercanbeconnectedtostandardACreceptacleanywhereintheworldwithouttheneedforanexternaltransformer.

Page 58: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

56 Digital Transformer Ratiometer DTR® Model 8510

Functionality:

ChargingbeginswhenthechargerisconnectedtotheDTR®8510andthepowerswitch is intheOFFposition.TheLEDwillbeorangebeforethefastchargestarts.TheLEDwillchangetoredwhenthefastchargecyclebegins.Whenthecellsarefullychargedandthevoltagedrops(becauseofthe-dVsignalfromthecells),thechargerwillgointoatop-offchargemodebeforeswitchingtotricklechargemode.

Duringthetop-offchargesession,theLEDwillbegreenwithshort,inter-mittentorangeflashes.Whenthetop-offchargeiscomplete,thechargerwill go into trickle chargemode, changing theLEDcolor togreen.Thechargecurrentisnowreducedtoasafelevel.However,tricklecharginginexcessof12hoursonroutinebasiscouldbedamagingforNiMHbat-teries.

If thesafety timerdisconnectsbefore-dV, the top-offchargewillnotbeengaged.ThechargerwillthengodirectlytotricklechargemodeandtheLEDwillturngreen.Ifthebatteryvoltageisfarbelownormal,thechargerwillstopthefastchargecurrentandswitchto tricklechargemode.TheLEDwillthenindicatean"error"byflickeringgreenandred.Thebatterypacksand/orthechargermayneedtobereplaced.

Ifthelinepoweristurnedoff,thechargerwillresetandstartanewchargecyclewhenitisturnedonagain.

8.2 Cleaning

Disconnect the instrument from any source of electricity.

Useasoftclothlightlymoistenedwithsoapywater.•

Wipewithamoistclothandthendrywithadrycloth.•

Neverusealcohol,solventsorhydrocarbons.•

Page 59: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 57

APPENDIX A

DISPLAY MESSAGES

Check for H<>X Cable ReversalTheDTR®Model8510mayhavedetectedapotentiallyhazardousstep-upconditionwhileattemptingtotestthetransformerasconnected.Verifyconnections.

High Excitation CurrentCheck test cable connections for inadvertent short circuits. Check forshortedtransformerwindings.

Low BatteryAsthebatteriesaredrainedduringuse,thegreenLEDwillbegintoflashwhenthebatteryvoltagereachesapre-determinedthreshold.Thetestscanberunasusualduringthistimebutthebatterieswillsoonneedtobecharged.Asthebatteriesreachapredeterminedthreshold,LOWBATTERYindica-tionwillappearonthedisplay.Allthebuttonswillbenon-operationalnow.Thebatteriesmustberechargedinordertocontinuetesting.

NOTE: The DTR® Model 8510 charges in <4 hours. The instrument can not be used to test when the batteries are charging.

X Circuit CONT or OPENThe------indicationisforCONT(Closedswitch).Itindicatesagoodconnection.The--/--indicationisforOPEN(Openswitch).Itindicatesapoororopenconnection.Themessage“CheckXConnection”willbedisplayed.

Page 60: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

58 Digital Transformer Ratiometer DTR® Model 8510

H Circuit CONT or OPENThe------indicationisforCONT(Closedswitch).The--/--indicationisforOPEN(Openswitch)Themessage“CheckHConnection”willbedisplayed.

Note: IfbothHandXconnectionsareopen,“CheckHXconnection”willbedisplayed.

Low X SignalCheckSecondary(X)cablesforinadvertentshort.ChecktomakesuretheSecondary(X)cableisconnectedproperlyattheinstrumentpanel.Secondary(X)windingmaybeshortedorthetransformercircuitbreakermaybeopenortheratiomaybehigherthantheinstrumentcanmeasure.

Low H SignalCheckPrimary(H)cablesforinadvertentshort.ChecktomakesurethePrimary(H)cableisconnectedproperlyattheinstrumentpanel.Primary(H)windingmaybeshorted.

Memory is Full! Data NOT storedTheinstrumentmemoryisfullandthereisnomorelocationleftforadditionalstorage.

Memory Error Service RequiredMeasurementRecordscannotbestoredorrecalledbecausethememorycannotbeaccessed.Factoryserviceisneeded.However,theDTR® 8510canbeusedtomakemeasurementsinanormalmanner.

Page 61: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 59

APPENDIX B

Transformer Serial Number Correlation to Object NumberObject #(1 - 99) Transformer Serial # # of Tests in Object

(1 - 99)

Page 62: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

60 Digital Transformer Ratiometer DTR® Model 8510

Repair and Calibration

Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.

For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15 Faraday Drive Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.

Technical and Sales Assistance

Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 200FoxboroughBoulevard Foxborough,MA02035USA Phone:(800)343-1391 (508)698-2115 Fax: (508)698-2118 E-mail: [email protected] www.aemc.com

NOTE: Do not ship instruments to our Foxborough, MA address.

Page 63: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

Digital Transformer Ratiometer DTR® Model 8510 61

Limited Warranty

TheDTR®Model8510iswarrantedtotheownerforaperiodofoneyearfromthedateoforiginalpurchaseagainstdefectsinmanufacture.Thislimitedwar-rantyisgivenbyAEMC®Instruments,notbythedistributorfromwhomitwaspurchased.Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.

For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.

What AEMC® Instruments will do:Ifamalfunctionoccurswithintheone-yearperiod,youmayreturntheinstrumenttousforrepair,providedwehaveyourwarrantyregistrationinformationonfileoraproofofpurchase.AEMC®Instrumentswill,atitsoption,repairorreplacethefaultymaterial.

REGISTER ONLINE AT:www.aemc.com

Warranty Repairs

What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments 15FaradayDrive•Dover,NH03820USA Phone:(800)945-2362(Ext.360) (603)749-6434(Ext.360) Fax: (603)742-2346or(603)749-6309 E-mail:[email protected]

Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.

NOTE: You must obtain a CSA# before returning any instrument.

Page 64: ENGLISH User Manual · 4 Digital Transformer Ratiometer DTR® Model 8510 1.1 International Electrical Symbols Signifies that the instrument is protected by double or reinforced insulation.

12/10

99-MAN 100353 v1

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346

www.aemc.com