Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the...
Transcript of Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the...
![Page 1: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/1.jpg)
Electrocardiograph (ECG)
Demonstration Board
6/3/2013 Design Team 3 1
Team Members Justin Bohr
Xie He Yuan Mei
Chaoli Ang Nate Kesto Mike Mock
Sponsor Peter Semig
Texas Instruments – Precision Analog
Facilitator Professor Ramakrishna Mukkamala
![Page 2: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/2.jpg)
Project Background
Develop an ECG demonstration board for the Precision Analog group at Texas Instruments.
TI intends to showcase their integrated circuit (IC) performance using the demonstration board. Provides TI with customer interest in specific precision analog components.
6/3/2013 2 Design Team 3
![Page 3: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/3.jpg)
Definition of Success
6/3/2013 Design Team 3 3
Expectations Functionality
Professional Display
Process signals from CardioSim II (simulator)
Display ECG output signal with Stellaris EVB Oscilloscope
Include TI components in design
Project Objective Develop several iterations of analog front-end (AFE)
circuits to interface a cardio simulator with an oscilloscope
Simulate, build, and test to optimize a final design
![Page 4: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/4.jpg)
Development Strategy
Research ECG Systems
Software and Simulations
Breadboard Testing
Develop AFE Printed Circuit Board
Measure System Performance
Optimize Design
Develop Final Solution
6/3/2013 4 Design Team 3
![Page 5: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/5.jpg)
Research
Studied ECG systems to observe functionality
ECG signals
Noise Artifact Mitigation
Battery-Powered Applications
Observations
Bandwidth: 0.7 – 100 Hz
Amplitude: 1.0mVpp – 1.5mVpp
Active and passive filtering
Linear dropout regulator or Buck converter (9V to 5V)
6/3/2013 5 Design Team 3
![Page 6: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/6.jpg)
TI Components
Final Texas Instrument Circuit Components
• Instrumentation Amplifier – INA333
• Operational Amplifier – OPA378
• Linear Drop-Out Regulator (LDO) – TPS7A4201
• Buck Converter – TPS62120
OPA378 INA333 TPS7A4201
6/3/2013 6 Design Team 3
![Page 7: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/7.jpg)
ECG System Structure
Vre
f
VCC
VCC
Vref
VCC
VC
C
RG1 10k
RG2 10k
R3 10k
R5 1M
NS
NS1 2.5
R14 1
M
NS
NS2 2.5R7 1k
C4 47n
R4 1k
C2 47n
R8 63.4k R9 63.4k
C5 3
3p
C6 3
3p
R10 63.4k R11 63.4k
C7 3
3pC8 3
3p
C9 3
30p
+
-
+
U2 OPA333
+
-
+
U4 OPA333
R16 1
M
RG
+
-V-
RefOut
V+RG
U3 INA333
+Vin
C1 1
T
VF
VoA
Vin
L2 1T
R12 1
MC10 1u
+
-
+
U1 OPA333
NS
NS3 2.5
R15 1M
R13 1
M
V3 0
V4 0
R17 100k
R1 100k
R2 100k
ECG Signal
Input LP Filter
Instrumentation Amplifier
Output LP Filter
Stellaris μC Display
Right Leg Drive
Servo HP Filter
6/3/2013 7 Design Team 3
![Page 8: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/8.jpg)
Simulation Results
DC Servo loop attenuates low frequency wandering below 0.7 Hz
6/3/2013 8 Design Team 3
![Page 9: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/9.jpg)
Breadboard Testing
Using adapters for INA333 and OPA333
6/3/2013 9 Design Team 3
![Page 10: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/10.jpg)
Phase 1
Analog Front End Circuit (AFE)
Printed Circuit Board (PCB) Layout
Buck Converter Reference Design (TPS62120)
Test Points & Jumpers
6/3/2013 10 Design Team 3
![Page 11: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/11.jpg)
Phase 1 Results
PCB layout error
Successful operation
Buck converter measurements
Phase 1 AFE output using CardioSim II
40.00
45.00
50.00
55.00
60.00
65.00
0.09 0.9 9 90
Ga
in (
dB
)
Frequency (Hz)
Measured Bode Plot of System Bandwidth
Bandwidth: 0.7 Hz – 23 Hz Passband Gain: 64 dB
6/3/2013 11 Design Team 3
![Page 12: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/12.jpg)
Phase 1 Results
6/3/2013 12 Design Team 3
![Page 13: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/13.jpg)
Phase 2
Fixed PCB trace error
Battery Connector
Thumb Pad Test
LDO vs. Buck Converter Comparison
RLD switch
Increased System Bandwidth to 50 Hz
6/3/2013 13 Design Team 3
![Page 14: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/14.jpg)
Phase 2 Results
Successful implementation of thumb pad sensors
Power circuit comparison
Increase bandwidth to 50 Hz for more ECG spectral content
TPS7A4201 TPS62120
Input Voltage (V) 9.5 9.5
Input Current (mA) 0.74 0.43
Input Power (mW) 7.03 4.08
Output Voltage (V) 4.93 5.12
Output Current (mA) 0.67 0.71
Output Power (W) 3.30 3.63
Efficiency (%) 47.00 88.98
6/3/2013 14 Design Team 3
![Page 15: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/15.jpg)
Phase 2 Results
6/3/2013 15 Design Team 3
![Page 16: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/16.jpg)
Phase 3
LDO for Power solution
Thumb pads orientation
Organized layout
Selectable bandwidth 50 Hz or 100Hz
Created holes for board mounting
Battery on bottom of board
LED power indicator 6/3/2013 16 Design Team 3
![Page 17: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/17.jpg)
Phase 3 Results
Selectable output filters
Labeled sub-circuits for demonstration purposes
Battery life of 211 hrs
Optimized thumb pad position
Improved connections for inputs and outputs
6/3/2013 17 Design Team 3
![Page 18: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/18.jpg)
Phase 3 Results
6/3/2013 18 Design Team 3
![Page 19: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/19.jpg)
Acrylic Display
Mount AFE on standoffs
Display Stellaris at 45 degrees
SolidWorks 3D model
6/3/2013 19 Design Team 3
![Page 20: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/20.jpg)
Code Modifications
Larger timebase options including 100 ms/div and above
Smoother screen updating
Default settings changed for optimized ECG viewing
6/3/2013 20 Design Team 3
![Page 21: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/21.jpg)
Demonstration
6/3/2013 21 Design Team 3
![Page 22: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/22.jpg)
Future Design Improvements
Implement a FFT for digital filtering on the Stellaris
Calculate BPM from FFT and display on LCD screen
Design the analog system using higher-order filters
Integrating Stellaris display board and AFE board into one PCB
6/3/2013 22 Design Team 3
![Page 23: Electrocardiograph (ECG) Demonstration Board · Develop an ECG demonstration board for the Precision Analog group at Texas Instruments. TI intends to showcase their integrated circuit](https://reader034.fdocuments.net/reader034/viewer/2022042911/5f44202cfcace86ced452321/html5/thumbnails/23.jpg)
Questions?
6/3/2013 23 Design Team 3