DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 1 of 25 Rev. C (January 2015) Beckman Coulter Professional Development DxC Integrated Workstation Daily Start Up For Training Purposes Only These job aids are shortened versions of procedures found in the references below. Information in the job aid is correct as of the date published. Verify you have the correct information. References: UniCel DxC Synchron Access Clinical System Integrated Workstations Instructions For Use PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1

Transcript of DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily...

Page 1: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 1 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

DxC Integrated Workstation Daily Start Up

For Training Purposes Only These job aids are shortened versions of procedures found in the references below. Information in the job aid is correct as of the date published. Verify you have the correct information.

References: UniCel DxC Synchron Access Clinical System Integrated Workstations Instructions For Use

PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1

Page 2: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 2 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

UniCel DxC Synchron Access Clinical System Integrated Workstations

WARNINGS AND PRECAUTIONS Read all product manuals and consult with Beckman Coulter-trained personnel before attempting to operate the instrument. Beckman Coulter, Inc. urges its customers and employees to comply with all national health and safety standards such as the use of barrier protection. This may include, but is not limited to, protective eyewear, gloves, suitable laboratory attire when operating or maintaining this or any other automated laboratory equipment.

INTENTION FOR USE This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU) or Help System. Information in the Instructions for Use Manual or Help System supersedes information in any other manual.

REVISION STATUS Rev. C (January 2015) Based on DxC software version 5.4 and DxI software version 5.1.

TRADEMARKS Beckman Coulter, the stylized logo, Access, DxI, Microtube, Synchron, and UniCel are trademarks of Beckman Coulter, Inc., and are registered with the USPTO. All other trademarks are the property of their respective owners.

Page 3: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 3 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Table of Contents

Component Flowchart Page

IW DxC Integrated Workstation Daily Start Up 4

IW View/Print Supply, Reagent and Calibration Status 5

DxI Check Backup 6

DxI Shake Solid Waste Container 7

DxI Run Clean Routine 8

DxI Add Reaction Vessels 9

DxI Change Liquid Waste Container 10

DxI Change Wash Buffer Container 11

DxI Change Substrate Bottle 12

DxI Change Solid Waste Container 13

DxI Load Reagent Pack 14

DxI Unload Reagent Pack 15

UCTA Change Wash Buffer Container 16

UCTA Change Auto-Gloss 17

DxC Load MC Reagents 18

DxC Load CC Reagents 19

DxC Unload CC Reagents 20

DxC Calibration 21

DxI Calibration 22

IW Auto Generation of Control (QC) 23

IW Manual Programming of Quality Control 24

IW Clear Racks 25

Page 4: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 4 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

DxC Integrated Workstation Daily Start Up

No

No

No

No

No

Yes

Yes

Yes

Yes

Print/View Integrated Workstation supply, reagent, and calibration status

Load DxI supplies

Perform the following DxI Daily Maintenance procedures (Utility Assay enabled):~ Check System Backup~ Shake Solid Waste Container~ Run DxI Clean Routine~ Initial and Date Electronic Maintenance Log

Load UCTA supplies

Load DxC reagents

Calibrate DxC chemistries

Process Quality Control

Calibrate DxI assays

UCTAsupplies required?

DxCcalibration required?

DxI calibration required?

DxCreagents required?

Yes

Clear racks

DxI supplies* required?

*Note: DxI supplies can be loaded while the Clean Routine is

processing, except when changing the

Solid Waste.

Page 5: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 5 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 6: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 6 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 7: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 7 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 8: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 8 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Select Main Menu or the Menu tab from the DxI console

Enter the rack ID and press Enter

Use the load list displayed on the screen to place the appropriate cleaning solutions in the

correct position on the rack

Load the rack on the DxI Sample Presentation Unit (SPU)

*If you run the B12 assay, perform the

Special Clean in place of the Daily Clean. See the

DxI Help System for information about

running the Special Clean

CAUTIONCitranox and Contrad 70 cleaning solutions may

cause eye or skin irritation. See the MSDS

for details

Select Sample Manager F1

Select New Request F3

Select Maintenance F3

At the Request Maintenance Screen

Select OK F1

Process Daily Clean Routine

Document completed maintenance in Electronic Maintenance Log

Select Daily Clean System*

Page 9: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 9 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 10: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 10 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Use supply status to determine which bulk liquid waste

container(s) needs to be changed

Rinse the container thoroughly and store it for future use

Remove the full container and replace it with an

empty container

Decontaminate* the contents of the full container and

dispose of the liquid waste

Pull out the liquid waste drawer

Connect the tubing to the empty container and

close the drawer

Press the disconnect button on the full container to release

the tubing

Change a Full Liquid Waste Container

*Decontaminate liquid waste based

on laboratory protocol

Page 11: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 11 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Use supply status to determine which container(s) of wash buffer needs to be replaced

Pull out the wash buffer supply drawer

Extend the neck of the new container and remove the cap and inner seal

Remove the perforated cardboard panels of the new container and gently invert the container 3-4

times.

Press the disconnect button on the empty container to release the tubing

Remove the empty container from the drawer and replace it with the new container

Remove the cap/draw tube assembly from the empty container and insert it into the new

container

Connect the tubing to the new container and close the drawer

Change an Empty DxI Wash Buffer Container

*Lot number tracking requirement defined by

laboratory policy

No

YesLot number tracking

required?*

Select Rgts/Calfrom the workstation console

Select the Supplies tab

Type the lot number into the DxI Wash Buffer field (Rear or Front) located in the DxI Supplies Status section of the screen

Procedure complete

Page 12: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 12 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 13: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 13 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 14: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 14 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Check the status of the reagent tray in-use light

Place pack(s) in the tray positions

Close the reagent load door

Gently mix the packs and open the reagent load door

Wait until the green light turns offIs the light on?

Yes

No

Load a DxI Reagent Pack

More reagent pack(s) to load?

Procedure complete

No

Yes

Page 15: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 15 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 16: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 16 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 17: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 17 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 18: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 18 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 19: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 19 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 20: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 20 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Select Rgts/Calfrom the workstation console

Page down (pages 3-8) to locate the position(s) of the reagent(s) to be unloaded from the

CC carousel.

Open the reagent carousel door when prompted

Grasp the cartridge in the load position (located at the blinking bar code reader) and move it in front of the bar code reader beam. Verify the reagent

information is removed from the screen. Remove the cartridge

Close the reagent carousel door

Additional reagents selected?

Yes

No

Select Load F1

Unload CC Reagent

Procedure complete

Select position(s) to unload reagent cartridges

Select the DxC tab

Page 21: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 21 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

No

Yes

Select Rgts/Cal from the workstation console

Review the calibrator load list. Verify the calibrator lot number in use matches the load listNote: For some calibrators, the lot number is N/A

Press the RUN button on the UCTA.

Locate the racks to be used for calibration. Pour the calibrator(s) into sample cups and place the

sample cups in either the assigned bar code labeled tube(s) or the assigned calibrator rack

number and position

Place the rack(s) in the UCTA load tray

Select Cal F4

Select List F5

When Calibration is complete view calibration status on Rgt/Cal screen

Calibration successful?

Consult references for troubleshooting

assistance

Calibrate DxC Chemistries

Procedure complete

Select each chemistry for calibration. Page down to view additional chemistries

Select the DxC tab

Page 22: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 22 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 23: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 23 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

No

Yes

Locate the rack containing the bar code labeled tube(s) for the control sample(s)*

Pour the control material into a sample cup

Nest the sample cup in the appropriate bar code labeled tube(s)

Load the rack(s) on the UCTA load tray

When Quality Control is complete, review for acceptability**

Quality Control acceptable?

Consult references for troubleshooting

assistance

**Quality Control acceptability defined by

laboratory policy

Press the RUN button on the UCTA

Process Quality Control using Auto Generation of Control

Procedure complete

*To ensure all tests for your QC are processed following a manual run, reset auto generation of control by clearing the

control IDs.

Page 24: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 24 of 25 Rev. C (January 2015) Beckman Coulter Professional Development

Page 25: DxC Integrated Workstation Daily Start Up - Beckman Coulter · DxC Integrated Workstation Daily Start Up ... PN A93829AC (March 2014) UniCel DxI Help System, DxI software v5.1 . DxC

DxC Integrated Workstation, Customer, Daily Start Up Job Aid Page 25 of 25 Rev. C (January 2015) Beckman Coulter Professional Development