DUSTCHECK DPM-1 600/DCDPM0101K · The Dustcheck DPM1 ambient sampling for PM10 and PM2,5 particle...

2
14 AMS Analitica Quality management system certified | ITALY | | www.amsanalitica.com The Dustcheck DPM-1 is a compact sampling device with one lter in full compliance of EU EN- 12341 and EN-14907 PM10 and PM2,5 requirements. The Dustcheck DPM-1 is also available to US-EPA method CFR-40. The system designed for a lter holder support that allows the user to select up to ve days of programmed air sampling monitoring. Easy to use, the Dustcheck DPM-1 sequential sampler t with all sampling needs. Based with a powerful aspirating membrane pump (the EPA version has built in a single head diaphragm pump) that allows the system to perform sampling in a constant ow option to compensate any pressure drop coming from overload samples. The new microprocessor capabilities allow the user to select and store in the memory up to three dierent programs with sampling settings ready to start. The DustCheck DPM1 sampler is in EN and US-EPA compliance. External case is in aluminum with epoxy paint no corrosion. 600/DCDPM0101K DUSTCHECK DPM-1 PM10 e PM2,5 sampler Technical specifications High performances vacuum-proof pump from 6 m3/h Application range between 5 and 50 l/min Integrated volumetric meter for sample volume control, precision +/- 2% Mass-Flow device for sampling flow control, precision +/- 1% Barometric atmospheric pressure detection sensor and sensors for temperature detection at volumetric meter and suction point and for envi- ronmental temperature detection Sampling flow standard deviation setting and control in the event of high back pressure. Display control during all the sampling stages Real-time parameters adjustment during sampling stages Device management settings including leak test, sensor control and values measurement Possible two-sensor connection for speed and wind direction detection (sensors optional) RS232 output for data download in the SDC2000™ software Memory data storage and data download with removable USB drive. Mother board firmware updates with USB removable drive. Backlit high resolution plasma display Integrated serial printer on front panel Sampler remote management and PC programming of the sampling stages (control board optional) through GSM/GPRS modem (modem optional and SIM not available) Automatic sampling data back up in the event of power failure and automatic sampling restart from break point when power is restored. Breaking point are highlighted in the final summary. Optional filters cooling compartment compliant with EN-12341/14907 (optional accessory) Compliant with EN-12341 and EN-14907 European regulations as well as with US-EPA Cfr.40 (only if equipped with US-EPA impactors). Housed in a epoxy treated aluminum container capable of operation in adverse weather conditions. Unique position 47mm sampling filters Fully microprocessor based for sampling periods and data management Large backlight plasma display Main Features: • Single position 47 mm filter holder support • Outdoor shelter with secure lock key • Volume sampling for 1 single filter • Ambient temperature detection available with an optional support

Transcript of DUSTCHECK DPM-1 600/DCDPM0101K · The Dustcheck DPM1 ambient sampling for PM10 and PM2,5 particle...

Page 1: DUSTCHECK DPM-1 600/DCDPM0101K · The Dustcheck DPM1 ambient sampling for PM10 and PM2,5 particle sizing com-ply with the European EN-12341:2014 and the US-EPA Cfr.40, if equipped

14 AMS Analitica Quality management system certified | ITALY | | www.amsanalitica.com

The Dustcheck DPM-1 is a compact sampling device with one lter in full compliance of EU EN-12341 and EN-14907 PM10 and PM2,5 requirements. The Dustcheck DPM-1 is also available to US-EPA method CFR-40. The system designed for a lter holder support that allows the user to select up to ve days of programmed air sampling monitoring.Easy to use, the Dustcheck DPM-1 sequential sampler t with all sampling needs. Based with a powerful aspirating membrane pump (the EPA version has built in a single head diaphragm pump) that allows the system to perform sampling in a constant ow option to compensate any pressure drop coming from overload samples. The new microprocessor capabilities allow the user to select and store in the memory up to three dierent programs with sampling settings ready to start. The DustCheck DPM1 sampler is in EN and US-EPA compliance. External case is in aluminum with epoxy paint no corrosion.

600/DCDPM0101KDUSTCHECK DPM-1

PM10 e PM2,5 sampler

Technical specifications

High performances vacuum-proof pump from 6 m3/h

Application range between 5 and 50 l/min

Integrated volumetric meter for sample volume control, precision +/- 2%

Mass-Flow device for sampling flow control, precision +/- 1%

Barometric atmospheric pressure detection sensor and sensors for temperature detection at volumetric meter and suction point and for envi-ronmental temperature detection

Sampling flow standard deviation setting and control in the event of high back pressure. Display control during all the sampling stages

Real-time parameters adjustment during sampling stages

Device management settings including leak test, sensor control and values measurement

Possible two-sensor connection for speed and wind direction detection (sensors optional)

RS232 output for data download in the SDC2000™ software

Memory data storage and data download with removable USB drive. Mother board firmware updates with USB removable drive.

Backlit high resolution plasma display

Integrated serial printer on front panel

Sampler remote management and PC programming of the sampling stages (control board optional) through GSM/GPRS modem (modem optional and SIM not available)

Automatic sampling data back up in the event of power failure and automatic sampling restart from break point when power is restored.

Breaking point are highlighted in the final summary.

Optional filters cooling compartment compliant with EN-12341/14907 (optional accessory)

Compliant with EN-12341 and EN-14907 European regulations as well as with US-EPA Cfr.40 (only if equipped with US-EPA impactors). Housed in a epoxy treated aluminum container capable of operation in adverse weather conditions.

• Unique position 47mm sampling filters• Fully microprocessor based for sampling periods and data management• Large backlight plasma display

Main Features:

• Single position 47 mm filter holder support• Outdoor shelter with secure lock key• Volume sampling for 1 single filter• Ambient temperature detection available with an optional support

Page 2: DUSTCHECK DPM-1 600/DCDPM0101K · The Dustcheck DPM1 ambient sampling for PM10 and PM2,5 particle sizing com-ply with the European EN-12341:2014 and the US-EPA Cfr.40, if equipped

15AMS Analitica Quality management system certified | ITALY | | www.amsanalitica.com

600/DCDPM0101KDustCheck DPM-1

How to order

600/DCDPM0101K DustCheck DPM1 sampler (No inlets)

600/LVSPM10DC38K PM10 38 l/min-EN 12341 Sampling Inlet with Impactor

600/LVSPM25DC38K PM2,5 38 l/min-EN 14907 Sampling Inlet with Impactor

600/101.001.01 Impactor PM10 1m3/h

600/102.010.01 Impactor PM10 2,3m3/h

600/103.003.01 Impactor PM2,5 1m3/h

600/104.010.01 Impactor PM2,5 2,3m3/h

600/PM10-EPA01K LVS PM10/US-EPA 16,7 l/min.sampling head

600/WINS-2,5PM01K EPA Wins Impactor PM2,5 16,7 l/min

600/AFSC001K Field sampling support

600/PFC-47PM Replace filter cassette 47mm

FAN/MFQ047 Quartz micro fiber filters 47mm Box 25pcs

The Dustcheck DPM1 ambient sampling for PM10 and PM2,5 particle sizing com-ply with the European EN-12341:2014 and the US-EPA Cfr.40, if equipped with the EPA inlet at 16,7 liters per minute. The external shelter is protected by special paint type coating to environmental protection. All data stored in the memory are available for download with a dedicated SDC software.

The DustCheck DPM-1 supplied with:

1 filter holder with 2 filter cassettes

Built in serial printer

Outdoor ambient temperature with shelter

Instruction manual

AC Plug cable

RS 232 cable to connect DPM5 to PC and software

Calibration report