Dry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop

Transcript of Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic...

Page 1: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Dry Calibration of Ultrasonic

Flowmeters with Natural Gas

up to 285 bar

Jos van der Grinten - NMi EuroLoop

Page 2: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Dry Calibration of

Ultrasonic Flow Meters with

Natural Gas up to 285 bar

Jos van der Grinten

Chief Metrologist, NMi EuroLoop

Page 3: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Contents

• Introduction

• Metering accurately at 285 bar?

• Theory

• Experimental set-up

• Results

• Conclusions

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 3

Page 4: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Introduction

Pipelines

• BBL

• Northstream

• Southstream

Metering

• >> 100 bar

• USMs

• No calibration

facilities

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 4

Page 5: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Metering accurately at 285 bar?

Method

• Calibrate at highest available pressure (60 bar)

covering the meter’s flowrate range

• Check speed of sound (SoS)

– at zero flow conditions,

– at pressures up to the operational pressure ,

– using natural gas

• If criteria are met,

use 60 bar curve at 285 bar

expand uncertainty

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 5

Page 6: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Metering accurately at 285 bar?

Assumptions

• Meter curve weakly dependent on pressure

USM is real flow velocity meter

evidenced by type approval tests

• Change in path length << path length

or is compensated

• If speed of sound is correctly measured,

then flow velocity is measured correctly as well

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 6

Page 7: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Theory – USM principle

• If increased pressure causes no change of the

geometry, c will match theory and consequently

v will be measured correctly

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 7

L v α

revfw

fwrev

revfw

revfw

fwrev

revfw

rev

fw

TT

TTLv

TL

TLv

TT

TTLc

TL

TLc

TLvc

TLvc

)cos(2)cos(2

22

)cos(

)cos(

Page 8: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Theory – Speed of Sound (p, t)

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 8

-105

20

35

350.0

400.0

450.0

500.0

550.0

600.0

650.0

700.0

1

21

41

61

8110

11211

4116

118120

122124

126128

1301

Temperature [°C]

Spe

ed

of

Sou

nd

[m

/s]

Absolute Pressure [bar]

650.0-700.0

600.0-650.0

550.0-600.0

500.0-550.0

450.0-500.0

400.0-450.0

350.0-400.0

Natural Gas

Page 9: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Theory – Speed of Sound (p, t)

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 9

-105

20

35

300.0

350.0

400.0

450.0

500.0

550.0

600.0

650.0

1

21

41

61

8110

11211

4116

118120

122124

126128

1301

Temperature [°C]

Spe

ed

of

Sou

nd

[m

/s]

Absolute Pressure [bar]

600.0-650.0

550.0-600.0

500.0-550.0

450.0-500.0

400.0-450.0

350.0-400.0

300.0-350.0

Nitrogen

Page 10: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Experimental set-up

• 6 x USM 16” ANSI 2500# NorSok

• Gauge pressure 0 – 350 bar at Pr point

• Barometer in control room

• Pt100 temperature transmitters (HART)

– Inside temperature

– Body temperature

– Ambient temperature

• All instruments interfaced with our CMS that logs

all transmitters

• Meter log recorded at our diagnostics computer

• Gas composition averaged during filling process

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 10

Page 11: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Experimental set-up

• Test pressures: 31, 151, 285 bara

• Measurands: SoS, VoG, TUSM, Tbody, Tamb, P

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 11

T

10” 10”

P

T

T

GC

T

Piston compressor

Page 12: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Experimental set-up – preparation

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 12

Page 13: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Experimental set-up

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 13

Page 14: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Problem solving

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 14

Page 15: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Criteria for acceptance

• Max deviation between measured SoS and the

SoS calculated from the gas composition:

within ± 0.3% (originally ± 0.2%)

– Equation of State both AGA-10 and GERG2008

– Average SoS

• Dispersion of SoS values between chords,

high – low or 2·s(c1..n) ≤ 0.5 m/s

• Zero flow velocity: average within ± 0.006 m/s

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 15

Page 16: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Results – process

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 16

6.0

8.0

10.0

12.0

14.0

16.0

18.0

20.0

0

100

200

300

400

500

600

700

14:00 20:00 2:00 8:00 14:00 20:00 2:00 8:00 14:00 20:00 2:00 8:00 14:00 20:00 2:00 8:00

Tem

pra

ture

[°C

]

Pre

ssu

re [b

ara]

& S

pe

ed o

f So

un

d [

m/s

]

Time [hh:mm]

Pressure Test and Dry Calibration – USM 1

Meter pressure P [bara]

SoS

SoS GERG 2008 [m/s]

Ambient t [°C]

Meter body t [°C]

Gas inside meter t [°C]

Page 17: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Results – SoS and theory

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 17

-2.0

-1.0

0.0

1.0

0

50

100

150

200

250

300

350

20:00 22:00 0:00 2:00 4:00 6:00 8:00 10:00 12:00 14:00 16:00 18:00 20:00 22:00

SoS

rel d

iff

[%]

Pre

ssu

re [b

ara]

Time [hh:mm]

SoS and theory – Dry Calibration – USM 2

Meter pressure P [bara]

SoS rel diff GERG 2008[%]

SoS rel diff AGA-10 [%]

target ± 0.30%

Page 18: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Results – SoS path dispersion

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 18

-1.0

0.0

1.0

2.0

3.0

4.0

5.0

6.0

0

50

100

150

200

250

300

350

14:00 16:00 18:00 20:00 22:00 0:00 2:00 4:00 6:00 8:00 10:00

2·s

tde

v(So

S) [

m/s

]

Pre

ssu

re [b

ara]

Time [hh:mm]

SoS path dispersion – Dry Calibration – USM 3

Meter pressure P [bara]

2*stdev Ch_SoS [m/s]

target ± 0.10%

Page 19: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Results – VoG at zero flow

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 19

-0.035

-0.025

-0.015

-0.005

0.005

0.015

0.025

0.035

0

50

100

150

200

250

300

350

8:00 10:00 12:00 14:00 16:00 18:00 20:00 22:00 0:00 2:00 4:00

Vo

G [

m/s

]

Pre

ssu

re [b

ara]

Time [hh:mm]

Zero Flow Test – USM 6 Master & Slave

Meter pressure P [bara]

1 Ch Slave VoG [m/s]

4 Ch Avg M VoG [m/s]

target ± 0.006 m/s

Page 20: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Results – uncertainty

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 20

• GERG2008 uncertainty 0.10%

• AGA-10 uncertainty 0.1% - 0.3% - 0.5%

• pressure 0.08%, temperature 0.04%

Pres-

sure

[bar a]

Temper-

ature

[°C]

# GERG

2008

Average

[%]

AGA-10

average

[%]

stdev

[%]

UCMC,

GERG

[%]

Utot,

GERG

[%]

284.09 ±

0.25

14.59

± 0.10

2791 -0.27 -0.12 0.01 0.12 0.13

283.26 ±

0.46

13.29

± 0.18

13010 -0.27 -0.13 0.01 0.12 0.13

151.61 ±

0.06

12.79

± 0.05

1861 -0.15 -0.07 0.00 0.12 0.13

32.23

± 0.01

10.22

± 0.03

1831 -0.48 -0.45 0.01 0.12 0.13

Page 21: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

AGA-10 / GERG2008 – 1 [%]

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 21

-100

1020

3040

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0.10%

0.15%

0.20%

121

4161

81101

121141

161181

201221

241261

281301

Temperature [°C]

Spe

ed

of

Sou

nd

[m

/s]

Absolute Pressure [bar]

0.15%-0.20%

0.10%-0.15%

0.05%-0.10%

0.00%-0.05%

-0.05%-0.00%

-0.10%--0.05%

-0.15%--0.10%

-0.20%--0.15%

Page 22: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Conclusions

• Dry calibrations

– Nitrogen generally used

– EuroLoop can also use natural gas

• Advantage of nat. gas is that dry calibrations are

performed with substantially higher SoS values:

Nat. Gas 660 m/s versus Nitrogen 470 m/s

• At 151 bar and 5°C ~ 15°C SoS is hardly

depending on temperature

• SoS calculations between AGA-10 and

GERG2008 differ at 285 bar by 0.15% and

match within their uncertainties

3rd European Flow Measurement Workshop, Noordwijk, 17 - 19 March 2015 22

Page 23: Dry Calibration of Ultrasonic Flowmeters with Natural · PDF fileDry Calibration of Ultrasonic Flowmeters with Natural Gas up to 285 bar Jos van der Grinten - NMi EuroLoop . Dry Calibration

Thank you !