Druck - Pdcr 3500 Flyer
description
Transcript of Druck - Pdcr 3500 Flyer
-
Dual Output Pressure SensorPDCR 3500
!"#$%&
Total Error-65 to 250 F
-0.300
-0.200
-0.100
0.000
0.100
0.200
0.300
0 200 400 600 800 1000 1200 1400 1600 1800 2000
Pressure (psi)
Erro
r (%F
S)
SN 1667320
Pressure 1527.257 psi absSerial # 1667320Last cal date 09/21/2002Manufactured 04/10/2002
Digital Output Analog Output
'( )
* +, ) -./0
1 2#3
-
The PDCR 3500 offers a unique capability of digitally enhanced, highaccuracy output, coupled with high analog bandwidth. The transduceris a 1 inch diameter fluid isolated analog output device similar to the PMP 4000 Series. It achieves a total accuracy performance of up to0.1% FS over -65 to 250F with the support from internal digitalelectronics which correct for the repeatable errors without affectingthe analog bandwidth.
Due to internal microprocessor electronics, the PDCR 3500 can alsogive a digital pressure output in engineering units via RS485. TheRS485 is also used for re-calibration and housekeeping activities suchas serial number retrieval and diagnostics.
The PDCR 3500 is therefore two transducers in one. Customers canuse either the analog or digital output (or both) depending on therequirements of their application.
SPECIFICATION
Pressure RangeAny range between 5 psi to 1000 psi gauge or absolute, 1000 to 10,000 psi sealed gauge or absolute.Other units may be specified.
Overpressure2x pressure range with negligible change in calibration coefficients
Pressure Containment6x pressure range up to 3000 psi for ranges up to 1000 psi4x pressure range up to 20,000 psi for ranges above 1000 psi up to10,000 psi
Media CompatibilityFluids compatible with stainless steel 316L and Hastelloy C276
Excitation Voltage18 to 32 volts
Non-Linearity, Hysteresis and Repeatability Over CalibratedTemperature Range0.2% FS Pressure0.3% FS Pressure after 12 months0.5% FS Pressure after 36 months
Operating Temperature Range-65 to 250F
Calibrated Temperature Range-65 to 250F
Pressure Connection7/16 - 10 UNJF to MS33656-4Others available on request
Electrical Connection10 Pin bayonet plug to MIL-C-26482 Shell size 12
Analog OutputOutput 0 to 5 VDCResolution Infinite
Digital OuputHardware RS485 half duplex 5 Vpp signalProtocol ASCII based, Refer to TN476Software Device fitted with DK 235. V1
Default SettingsFunction ValueBaud Rate 9600
Continuing development sometimes necessitates specificationchanges without notice.
Druck is an ISO 9001 registered company.
4 , 5 1
,267 28
Druck Incorporated4 Dunham DriveNew Fairfield, CT 06812Tel: (203) 746-0400Fax: (203) 746-2494E-mail: [email protected]
INSTALLATION DRAWINGS: ,! 9