Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose...

40
Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy By Faiz Khan (Chapter 9 )

Transcript of Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose...

Page 1: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

Dose Distribution/Scatter

Analysis/Dose calculations

The Physics of Radiation Therapy

By Faiz Khan (Chapter 9 )

Page 2: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• It is not possible to measure dose

distribution directly in patients

during their treatment

• Data on dose distribution are almost

entirely derived from measurements

in phantoms

INTRODUCTION

Page 3: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Basic dose distribution data are usually measured in a water phantom, which closely approximates the radiation absorption and scattering properties of muscle and other soft tissue

• (Wikipedia.org) for infant water is 75% of body-weight, for adults it is about 60% in males and 55% in females.

• Another reason for the choice of water as a phantom material is that it is universally available with reproducible radiation properties.

PHANTOMS

Page 4: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Solid dry phantoms

– tissue or water equivalent, it must have the same

• effective atomic number

• number of electrons per gram

• mass density

– Since Compton effect is the main interaction for megavoltage photon beams used in hospital, the necessary condition for water equivalence is

• same electron density (number of electrons per cubic centimeter)

PHANTOMS

Page 5: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PHANTOMS

• The electron density ( e)

– NA is Avogadro's number and ai is the fraction by weight of the ith element of atomic number Z, and atomic weight Ai.

Page 6: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PHANTOMS

From Faiz Khan

Page 7: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PHANTOMS

• The most important radiation properties in

this regard are the

– mass attenuation coefficient

– the mass energy absorption coefficient

– electron mass stopping power

– angular scattering power ratios

Page 8: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PHANTOMS

• anthropomorphic

phantom

– Frequently used for

clinical dosimetry

– Incorporates materials

to simulate various

body tissues, muscle,

bone, lung, and air

cavities

Page 9: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

DEPTH DOSE DISTRIBUTION

• PDD (percentage depth dose) is the ratio of

dose at any depth to the dose at some

reference depth along central axis of the

beam and its variation depends on

– Beam energy

– Depth

– Field size

– Distance from source

– Beam collimation system

Page 10: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PERCENTAGE DEPTH DOSE

• For orthovoltage (up to

about 400 kVp) and

lower-energy x-rays, the

reference depth is usually

the surface (do = 0).

• For higher energies, the

reference depth is taken

at the position of the peak

absorbed dose (do = dm).

Page 11: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Beam Quality and Depth

• The PDD (beyond the dmax) increases with beam

energy.

Fig. 9.3 from Faiz Khan

Page 12: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Beam Quality and Depth

• the dose build-up region.

– the skin-sparing effect.

Page 13: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Beam Quality and Depth

• the dose build-up region.

Fig. 9.4 from Faiz Khan

Page 14: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

A comparison of relative surface dose (relative to dmax ) as a function of field

size for several linear

accelerator models.

Nominal Beam energy/ Field size

accelerator model/ 5x5 10x10 20x20

ioniz. ratio open tray open tray open tray

6-MV

Siemens KDS-2 8.25 8.44 13.27 14.53 22.73 27.02

IR=0.675

10-MV

Siemens KDS-2 5.28 5.78 10.04 11.95 19.82 25.27

IR=0.748

6-MV

Varian 2500 ----- 8.9 ----- 15.2 ----- 25.8

IR=0.677

10-MV

Philips Sl75-20 ----- 5.7 ----- 11 ----- 21.3

IR=0.735

Page 15: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Effect of Field Size and Shape

• Field size

– Geometrical

(Defined by mirror light at surface)

– Dosimetrical or physical

(Defined as the size at 50% of isodose level)

SAD

FS

Page 16: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Effect of Field Size and Shape

• PDD increases with increasing field size

– Because increase in scattered dose is greater at

larger depths than at the depth of Dm

– The field size dependence of percent depth dose

is less pronounced for the higher-energy than

for the lower-energy beams.

Page 17: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Effect of Field Size and Shape

• Equivalent square fields

– for central axis depth-dose distribution

Table 9.2 from Faiz Khan

Page 18: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Effect of Field Size and Shape

• Equivalent square fields

– for central axis depth-dose distribution

– A simple rule of thumb

• the A/P parameter (area to perimeter ratio), as such, does not apply to circular or irregularly shaped fields

Page 19: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Source-Surface Distance

• Dose rate in free space from a point source varies

inversely as the square of the distance. (IVSL)

– scattering material in the beam may cause deviation

from the inverse square law.

• PDD increases with SSD

– IVSL

dmd

d

dm

SSD

SSD’

Page 20: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Source-Surface Distance

• PDD increases with SSD

Page 21: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• PDD increases with SSD

– the Mayneord F Factor ( without considering changes in

scattering )

PDD - Dependence on Source-Surface Distance

Page 22: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Source-Surface Distance

• PDD increases with SSD

Page 23: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

PDD - Dependence on Source-Surface Distance

• PDD increases with SSD

– the Mayneord F Factor

• works reasonably well for small fields since the

scattering is minimal under these conditions.

• However, the method can give rise to significant

errors under extreme conditions such as lower

energy, large field, large depth, and large SSD

change.

Page 24: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

Source Distance

• Source to Skin Distance (SSD): the distance

from the source (tungsten target) to the surface

of the patient.

• Source to Axis Distance (SAD): the distance

from the source of photons to the isocenter (the

point around which the linac head can rotate in

360o).

Page 25: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

SSDSAD

SSD setup is mostly used for statictherapy

SAD setup is used for rotational/dynamic therapy

Page 26: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

TISSUE-AIR RATIO

• was first introduced by Johns et al. in 1953

• the SSD may vary depending on the shape

of the surface contour, the SAD remains

constant

• TAR has been defined to facilitate

calculations not only for rotation therapy but

also for stationary isocentric techniques as

well as irregular fields

Page 27: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

TISSUE-AIR RATIO

Page 28: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

TISSUE-AIR RATIO

• Effect of Distance

– Independent of the distance from the source.

• Variation with Energy, Depth, and Field Size

– Backscatter Factor

– As the beam energy is increased, the scatter is further

reduced and so is the backscatter factor

Page 29: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

TISSUE-AIR RATIO

• Effect of Distance

– Independent of the distance from the source.

• Variation with Energy, Depth, and Field Size

– Backscatter Factor

• increases with field size, its maximum value occurs

for beams having a half-value layer (term used to

specify the beam quality) between 0.6 and 0.8 mm Cu,

depending on field size.

• Thus, for the orthovoltage beams with usual filtration,

the backscatter factor can be as high as 1.5 for large

field sizes

Page 30: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

TISSUE-AIR RATIO

– Backscatter Factor

• For megavoltage beams (Co-60 and higher energies),

the backscatter factor is much smaller.

• BSF for a 10 x 10-cm field for Co-60 is 1.036. This

means that the Dmax, will be 3.6% higher than the dose

in free space at the same point

• Above about 8 MV, the scatter at the depth of Dmax,

becomes negligibly small and the backscatter factor

approaches its minimum value of unity

Page 31: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

Relationship between TAR and PDD

r is field size at depth d and f is SSD

Page 32: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

Relationship between TAR and PDD

• Conversion of Percent Depth Dose from One SSD to

Another-the TAR Method

Page 33: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Example

A patient is to be treated with an orthovoltage beam having a

half-value layer of 3 mm Cu. Supposing that the machine is

calibrated in terms of exposure rate in air, find the time

required to deliver 200 cGy (rad) at 5 cm depth, given the

following data: exposure rate = 100 R/min at 50 cm, field

size = 8 x 8 cm, SSD = 50 cm, percent depth dose = 64.8,

backscatter factor = 1.20, and rad/R = 0.95

Page 34: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Example

Page 35: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Example

A patient is to be treated with Co-60 radiation. Supposing that

the machine is calibrated in air in terms of dose rate free

space, find the treatment time to deliver 200 cGy (rad) at a

depth of 8 cm, given the following data: dose rate free space

= 150 cGy/min at 80.5 cm for a field size of 10 x 10 cm,

SSD = 80 cm, percent depth dose = 64.1, and backscatter

factor = 1.036

Page 36: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Example

Determine the time required to deliver 200 cGy (rad) with a Co-

60 ray beam at the isocenter which is placed at a 10 cm depth

in a patient, given the following data: SAD = 80 cm, field size =

6 x 12 cm (at the isocenter), dose rate in free space at the SAD

for this field = 120 cGy/min and TAR = 0.68 1

Page 37: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

SCATTER-AIR RATIO

• Useful in calculating scattered dose in the medium

• The computation of the primary and the scattered dose

separately is particularly useful in the dosimetry of irregular

fields

• Defined as the ratio of the scattered dose at a given point in

the phantom to the dose in free space at the same point

• depends on the beam energy, depth, and field size.

Here d is the depth and rd is field size at that depth

Page 38: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

SCATTER-AIR RATIO

• Dose Calculation in Irregular Fields-Clarkson's

Method

– Any field other than square, rectangular and

circular field is considered as irregular

– The scattered component of the depth dose,

which depends on the field size and shape, can

be calculated separately from the primary

component which is independent of the field

size and shape

Page 39: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

• Irregular Fields-Clarkson's Method

– This method is used to calculate the scatter contribution

from all parts of irregular field

– The irregular treatment field is divided into sectors (say

each of sector is 10o, then its contribution is 1/36)

– Each sector is characterized by its radius and can be

considered as a circular part of that radius.

– Scatter contribution from all these sectors is summed

using tables of SAR to get the average scatter in the

treatment field

Page 40: Dose Distribution/Scatter Analysis/Dose calculationsphy428-528.ahepl.org/Dose_PDD_TAR_1.pdf · Dose Distribution/Scatter Analysis/Dose calculations The Physics of Radiation Therapy

SCATTER-AIR RATIO

• Clarkson's Method

– Net (SAR)QC = (SAR)QC -

(SAR)QB + (SAR)QA

– TAR = TAR(0) + SAR

Blocked area

Treatment area