QUICK START GUIDE Chlorine + ADVANCED PHOTOMETER SYSTEM DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to NSF/ANSI Standard 50

Transcript of DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to...

Page 1: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

QUICK START GUIDE

Chlorine+ADVANCED PHOTOMETER SYSTEM

DIGITAL ACCURACY FOR RELIABLE WATER CARE

Certified to NSF/ANSI Standard 50

Page 2: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

2

©2018, Industrial Test Systems, Inc. Printed in USA

CONTENTSWelcome. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Chlorine+ eXact® EZ Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Install “AAA” Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Fill, Dip, Read Total Bromine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5 Chlorine Dioxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7 Free Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9 Total Chlorine High . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-11 Hydrogen Peroxide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12-13 Ozone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14-15 Peracetic Acid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16-17 pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18-19 Total Chlorine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20-21Acrylic Calibration Key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22For Best Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 DPD Interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24About Built-in Sample Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Warranty (2 years) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Chlorine+ eXact EZ Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . 25 Method Verifi cation Ready Snap® 3. . . . . . . . . . . . . . . . . . . . . 26 Compliance Testing (USEPA) . . . . . . . . . . . . . . . . . . . . . . . . . 26MAHC AND NSF/ANSI 50 CERTIFICATION . . . . . . . . . . . . . . . 27 Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27eXact® Micro Tests & Reagents . . . . . . . . . . . . . . . . . . . . . . . . 28Patent Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

WELCOME.Thank you for your Chlorine+ eXact® EZ purchase! This guide will quickly walk you through the technical details of your new photometer. After initial set-up, test procedures, and tips, you will be on your way to digital water testing! Each test will require the use of one or more of the testing methods outlined in this manual.YOUR CHLORINE+ EXACT® EZ COMES WITH:• Cleaning Brush• Quick Start Guide (this booklet)• Acrylic Calibration Key (SEE PAGE 22)• 42 foil packet strips (6 of each): Free Chlorine, Total Chlorine, pH,

Total Chlorine High, Glycine, Peracetic Acid, and Peroxide

WHAT YOU WILL NEED TO GET STARTED:• Four (4) AAA batteries• #4 Phillips head screwdriverNote: This system has been calibrated for use with only our eXact® Micro reagents.

Note: The Chlorine+ eXact EZ Starter Kits include eXact Strip/Reagent bottles with 25 tests each (reagents vary with each kit) and not individual foil packets listed above.

Page 3: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

3

Sensafe.com

CHLORINE+ EXACT® EZ OVERVIEWYOUR NEW CHLORINE+ EXACT EZ PHOTOMETER IS IDEAL FOR TESTING AND MAINTAINING YOUR DRINKING WATER AND FOOD PROCESS WATER.

SAMPLE CELL - 4 MLBuilt-in plastic 4mL cell for collecting your water sample

LCD DISPLAYDisplays results and test name

LIGHT SOURCEUses 525nm wavelength and 11mm path-length for testing

READ BUTTONStarts test timer

MENU BUTTONCycles through available tests

BASEInstall/replace batteries here (IP67 rated waterproof)

RESULT UNITS

LOW BATTERY

LCD DISPLAY

ZERO/ON BUTTONTurns the meter on and creates a baseline for your water testing

CELL COVERCovers the cell for mixing and bright light situations

Protected against the effects of immersion in water to depth between 15cm and 1m

Part no. 486205

SCREWUnscrew to remove base

INSTALL “AAA” BATTERIES (NOT INCLUDED)1. Use a #4 Phillips head screwdriver to remove the screw

from the base of your Chlorine+ eXact® EZ .2. Remove the base.3. Install Four (4) new AAA batteries as illustrated inside

your photometer’s battery compartment. We recommend using high quality batteries.

4. Replace the base fi rmly with pressure while tightening the screw. The meter will turn on automatically.

5. Tighten the screw with #4 Phillips head screwdriver.Be sure not to over tighten.

Page 4: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

4

©2018, Industrial Test Systems, Inc. Printed in USA

TOTAL BROMINE

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows bR.

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.00PPM, indicating the meter is ready for testing.

4

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

Page 5: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

5

Sensafe.com

TOTAL BROMINE

REMOVE STRIP Remove one eXact® Strip Micro bR, Part #486644 and set in a dry, convenient place. Replace cap on bottle.5

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Total Bromine. This result is automatically stored in the bR menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

FILL, DIP, READ

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip. See page 23 for important tips.

6DPD will stain the CELL wall if allowed to remain in the cell. To remove staining, rinse cell thoroughly and fi ll with water then add two (2) drops of bleach (5-8%) and clean with brush until stain is removed. Caution: Avoid contact of bleach with eyes and clothing.

Page 6: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

6

©2018, Industrial Test Systems, Inc. Printed in USA

CHLORINE DIOXIDE

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows Cd.

REMOVE STRIPS Remove one eXact® Strip Micro Glycine,Part #484014 and one eXact® Strip Micro Cd, Part #486633 and set in a dry, convenient place. Replace caps on bottles.

4

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

Page 7: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

7

Sensafe.com

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip Glycine in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Remove and discard the strip.

5CHLORINE DIOXIDE

AUTO-ZEROThe meter will automatically ZERO and the display will read 0.00PPM. Get ready to dip eXact® Strip Cd.6

FILL, DIP, READ

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Chlorine Dioxide. This result is automatically stored in the Cd menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

8

DIP STRIPThe meter will automatically initiate a 20 second countdown. DIP the eXact® strip Cd in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip. The display will automatically start to count up for 100 seconds.

7DPD will stain the CELL wall if allowed to remain in the cell. To remove staining, rinse cell thoroughly and fi ll with water then add two (2) drops of bleach (5-8%) and clean with brush until stain is removed. Caution: Avoid contact of bleach with eyes and clothing.

Page 8: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

8

©2018, Industrial Test Systems, Inc. Printed in USA

FREE CHLORINEFILL, DIP, READ

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows CL.

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.00PPM, indicating the meter is ready for testing.

4

Page 9: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

9

Sensafe.com

FREE CHLORINE

REMOVE STRIP Remove one eXact® Strip Micro CL (DPD-1), Part #486637 and set in a dry, convenient place. Replace cap on bottle.5

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Free Chlorine. If result is 6.0 or greater, repeat steps 5-7 using same sample and a fresh strip. This result is automatically stored in the CL menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

FILL, DIP, READ

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip.

6DPD will stain the CELL wall if allowed to remain in the cell. To remove staining, rinse cell thoroughly and fi ll with water then add two (2) drops of bleach (5-8%) and clean with brush until stain is removed. Caution: Avoid contact of bleach with eyes and clothing.

Page 10: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

10

©2018, Industrial Test Systems, Inc. Printed in USA

TOTAL CHLORINE HIGH

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows CLH.

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.0PPM, indicating the meter is ready for testing.

4

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

Page 11: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

11

Sensafe.com

TOTAL CHLORINE HIGH

REMOVE STRIP Remove one eXact® Strip Micro CLH, Part #486672 and set in a dry, convenient place. Replace cap on bottle.

5FILL, DIP, READ

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Remove and discard the strip. The display will automatically start to count up for 120 seconds. See page 23 for important tips.

6

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Total Chlorine. This result is automatically stored in the CLH menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

Page 12: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

12

©2018, Industrial Test Systems, Inc. Printed in USA

HYDROGEN PEROXIDE

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows HP.

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.0PPM, indicating the meter is ready for testing.

4

Page 13: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

13

Sensafe.com

HYDROGEN PEROXIDE

REMOVE STRIP Remove one eXact® Strip Micro HP, Part #486648 and set in a dry, convenient place. Replace cap on bottle.5

FILL, DIP, READ

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Hydrogen Peroxide. This result is automatically stored in the HP menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Remove and discard the strip. The display will automatically start to count up for 100 seconds.

6If sample temperature is less than 42°F, allow sample to reach room temperature before testing. Another option is to ignore the result on the display and press READ again to start another 20 second countdown/100 second count-up. The extra reaction time is needed in cold samples.

Page 14: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

14

©2018, Industrial Test Systems, Inc. Printed in USA

OZONE

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows O3.

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.00PPM, indicating the meter is ready for testing.

4

Page 15: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

15

Sensafe.com

OZONE

REMOVE STRIP Remove one eXact® Strip Micro O3, Part #486634 and set in a dry, convenient place. Replace cap on bottle.5

FILL, DIP, READ

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Ozone. This result is automatically stored in the O3 menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip. See page 23 for important tips.

6DPD will stain the CELL wall if allowed to remain in the cell. To remove staining, rinse cell thoroughly and fi ll with water then add two (2) drops of bleach (5-8%) and clean with brush until stain is removed. Caution: Avoid contact of bleach with eyes and clothing.

Do not empty cell if Chlorine or Bromine may be present in the sample. Instead, press READ again and simultaneously DIP a Glycine strip into the reacted sample from step 7. Take this second result and subtract it from the fi rst result (obtained in step 7 above). This new value is your Ozone result.For a more detailed procedure, visit sensafe.com/exact-ez-chlorine-plus.

Page 16: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

16

©2018, Industrial Test Systems, Inc. Printed in USA

PERACETIC ACID

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows PA.

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0PPM, indicating the meter is ready for testing.

4

Page 17: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

17

Sensafe.com

PERACETIC ACID

REMOVE STRIP Remove one eXact® Strip Micro PA, Part #486675 and set in a dry, convenient place. Replace cap on bottle.5

FILL, DIP, READ

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Remove and discard the strip. The display will automatically start to count up for 120 seconds. See page 23 for important tips.

6

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Peracetic Acid. This result is automatically stored in the PA menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

Page 18: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

18

©2018, Industrial Test Systems, Inc. Printed in USA

PH

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows PH.

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.00pH, indicating the meter is ready for testing.

4

Requires water sample with a Total Alkalinity minimum of 20ppm.

Page 19: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

19

Sensafe.com

PH

REMOVE STRIP Remove one eXact® Strip Micro PH, Part #486639-II and set in a dry, convenient place. Replace cap on bottle.

5FILL, DIP, READ

DIP STRIPPress to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip.

6

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as pH. This result is automatically stored in the PH menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

Page 20: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

20

©2018, Industrial Test Systems, Inc. Printed in USA

TOTAL CHLORINE

1 POWER ON PHOTOMETERPress the button to power on the Chlorine+ eXact® EZ.

3 SELECT TEST Press and re-press the button until the display shows TC.

FILL, DIP, READ

FILL CELL Before testing, rinse CELL and clean with brush thoroughly. Finally rinse the cell 3 times with the water sample to be tested, then FILL cell to capacity to begin test.

2

CAP CELL AND ZERO METER Place the Cell Cover onto the CELL and press

and the photometer display reads 0.00PPM, indicating the meter is ready for testing.

4

Page 21: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

21

Sensafe.com

DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously DIP the eXact® strip in the sample, gently touching the bottom of the cell. Use a gentle constant back and forth motion (2 strokes per second) until the timer displays “1”. Be careful to not spill the sample from the CELL. Remove and discard the strip. The display will automatically start to count up for 120 seconds. NOTE: Standard Method (4500-Cl G, procedure for total chlorine) requires the reading to be made after 2 minutes from the time the KI is added.

6

DPD will stain the CELL wall if allowed to remain in the cell. To remove staining, rinse cell thoroughly and fi ll with water then add two (2) drops of bleach (5-8%) and clean with brush until stain is removed. Caution: Avoid contact of bleach with eyes and clothing.

TOTAL CHLORINE

REMOVE STRIP Remove one eXact® Strip Micro CL (DPD-4), Part #486670 and set in a dry, convenient place. Replace cap on bottle.5

FILL, DIP, READ

CAP CELL AND READ RESULTSPlace the Cell Cover onto the CELL and READ result displayed as Total Chlorine. If result is 6.0 or greater, repeat steps 5-7 using same sample and a fresh strip. This result is automatically stored in the TC menu. After testing is complete, rinse the sample cell immediately and clean with brush to remove reagents which coat the CELL wall.

7

Page 22: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

22

©2018, Industrial Test Systems, Inc. Printed in USA

ACRYLIC CALIBRATION KEY

READ RESULT & RECORDResult is displayed on the Chlorine+ eXact® EZ. Record value below. For additional tests, repeat steps 5-6.

Expect the result to be within 0.15 units from previous calibration. If variation is greater, please re-check your ZERO procedure (steps 3-4) and be sure to use clean water (deionized or distilled if necessary).

The Acrylic Calibration Key is a tool for verifying the calibration of your Chlorine+ eXact® EZ Photometer compared to its original factory settings. Run this test immediately upon receipt and record value for future reference. Follow steps 1-4 on page 8 (do not use cap during ZERO). Then, follow the remaining steps below.

DIP KEY AND PRESS READ Place the Acrylic Calibration Key into the water sample in the center of the CELL. Be sure the key is positioned upright (vertical) and to the bottom of the CELL. Press READ

5

6

Acrylic Calibration Key Records:

DateTested

RecordedValue

Comments

Page 23: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

23

Sensafe.com

FOR BEST ACCURACY

The Chlorine+ eXact® EZ photometer has a 5 minute auto-shutoff timer.

Each test menu can store 20 results. To retrieve the stored results, go to the desired test using the MENU key. When the desired test is displayed, press and hold down the MENU key. Continue holding down the MENU key to scroll the stored results for that test, starting with the most recent result. The meter will display, from memory, the last 20 readings in sequence beginning with -20, which is the latest result, followed by -19, which is the 2nd latest result, etc; and fi nally -01, which is the oldest result retained. Only the last 20 readings are stored in each menu. This meter is able to store 100 results in memory (20 in each menu).

Before testing, rinse the meter sample cell with the sample water 3 times.

Always fi ll the cell to capacity (4mL); be careful not to splash liquid over the side.

Test immediately after fi lling the cell with the water sample.

Due to the strip slitting process, you may fi nd one or two strips that are noticeably smaller or larger in width than the normal strips in the bottle. These should be discarded. Using these strips may give unreliable results.

When testing pH, it is recommended to run the pH test prior to running Chlorine. If you choose to run the pH after Chlorine, ensure you thoroughly clean the cell with water and the brush provided.

Meter is not compatible for use with powder pillows, tablets, or liquids from other manufacturers.

Dip strip for entire countdown.

Even if all pads are not immersed in water, DO NOT BEND THE STRIP. Make sure the strip touches the bottom of the CELL while dipping the strip for 20 seconds.

Each eXact® Strip Micro is valid for ONLY one test. Discard strip after use.

Dry the outside of the meter and inside the mixing cap before storage to prevent corrosion.

Remove batteries before storing for prolonged periods.

Store the meter and test materials out of direct sunlight and away from chemical storage areas.

Minimize exposure of meter and test reagents to heat above 80°F (27°C).

When installing batteries, verify the O-ring is still attached to the screw before tightening.

If water temperature is above 95°F/35°C, press READ when the timer displays “10” for Total Bromine, Free Chlorine, Total Chlorine, and pH tests.

Cleaning the cell with water and brush after each test is recommended for best accuracy and prior to storage of unit.

Page 24: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

24

©2018, Industrial Test Systems, Inc. Printed in USA

TROUBLESHOOTING/INTERFERENCESListed below are possible situations that may arise while testing. Please contact one of our knowledgeable customer service representatives if you require further assistance.

Subject Cause SolutionDim screen or no response from meter

Low battery Replace batteries

“LO” on LCD while zeroing

Low battery Replace batteriesDirty cell Clean cellCloudy sample Dilute sample or use fi lterBad LED Contact ITS

“HI” on LCD while reading

Result above detection level

Re-run test to verify result

“LO” on LCD while reading

Result below detection level

Re-run test to verify result

eXact® Strip Micro CL (DPD-1, DPD-3, DPD-4) Interferences

Interferant Interfering Levels & Treatments

Acidity If sample has acidity above 150mg/L CaCO3 test may not develop full color. Neutralize to pH 6.0 to 7.0 with 0.5N Sodium hydroxide.

Alkalinity If sample has alkalinity above 200mg/L CaCO3 test may not develop full color. Neutralize to pH 6.0 to 7.0 with 0.5N Sulfuric acid.

Bromine & Bromamines, Br2

Color similar to other oxidizer reaction at all levels.

Chlorine, Cl2 Color similar to other oxidizer reaction at all levels.

Chlorine Dioxide, ClO2 Color similar to other oxidizer reaction at all levels.

Copper, Cu+2 Color development is reduced above 10 ppm (mg/L).

Iodine, I2 Color similar to other oxidizer reaction at all levels.

Manganese (Mn+4, Mn+6);Chromium (Cr+6)

See AWWA procedure 4500-CL F, 1(d) for removal of interferences.

Monochloramines (NH2Cl)(applies to DPD-1 only)

Monochloramine interferences are known to occur in free chlorine DPD methods. This interference is dependent on temperature and monochloramine concentration.

Ozone, O3 Color similar to other oxidizer reaction at all levels.

Peroxides Interference is possible.

pH Typical pH samples of potable water with a pH of 6.0 to 9.0 are OK. If outside this range adjust to pH 6.0 to 7.0 using acid (0.5N Sulfuric acid) or base (0.5N Sodium hydroxide).

Page 25: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

25

Sensafe.com

BUILT IN SAMPLE CELLThe built-in sample cell is made of transparent plastic; the sturdy cell design will last for over 20,000 readings. Our studies have shown that scratches on the cell will not compromise the accuracy of your results because of the cell’s fi xed position.

WARRANTY (2 YEARS)Registration of your eXact® photometer must be received within 30 days from date of purchase to activate the warranty. Registration is available over the phone (+1-803-329-9712 Ext. 0) or online at sensafe.com/product-registration (Personal data is kept confi dential). The eXact® photometer is warranted to be free from defects in materials and workmanship for a period of two (2) years from the date of purchase by the customer. ITS will repair or replace, at its discretion, product which is deemed to be faulty due to manufacturing defect. Warranty does not cover product damage caused by abuse (such as crushing a tablet in the cell), battery corrosion damage, or improper use. If the meter is faulty or otherwise defective contact ITS by phone (+1-803-329-9712 Ext. 0) or email ([email protected]) to describe the problem and obtain a return authorization form before returning the photometer to ITS. Damage caused by improper packing of the photometer for return shipment to ITS will not be covered by the warranty. Customer is responsible for shipping charges to ITS. ITS pays postage when photometer is returned to customer. A maximum processing fee of $75 will be charged for repair or replacement of non-registered photometers and damages not covered by this warranty. The repair or replacement of the photometer will not extend or renew the period of guarantee. This warranty does not affect your statutory rights. The warranty is not transferable.

CHLORINE+ EXACT® EZ ACCURACYAll tests have been calibrated using certifi ed reference standards and analytical spectrophotometric methods. The Chlorine+ eXact® EZ has been factory calibrated and will stay valid because of its exceptional quality. We are so confi dent in the Chlorine+ eXact® EZ, we offer an industry leading 2-year warranty.

We built the Chlorine+ eXact® EZ to be easy, accurate and environmentally friendly. We have achieved this by utilizing our patented eXact® Strip Micro Technology, which uses 60% less water and chemistry than alternative methods. Instead of using a 10mL water sample, eXact® Strip Micro uses a 4mL water sample. The accuracy of the meter is maintained by designing the sample cell with an 11mm path-length.

ABOUT

Page 26: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

26

©2018, Industrial Test Systems, Inc. Printed in USA

COMPLIANCE TESTINGThis DPD test system for Chlorine and Chlorine Dioxide are accepted for reporting by most health departments because the tests are USEPA (DIN Standard 38 408 G4/G5, ISO 7393/2) accepted for testing requirements for Free Chlorine, Total Chlorine, and Chlorine Dioxide.

The compliance requirement is a photometer wavelength to measure between 490 and 530nm. The Chlorine+ eXact® EZ smart photometer uses a 525nm wavelength and 11 mm path-length. The eXact® Strip Micro CL/Cd (DPD-1) use the same reagents and proportions, and the resulting solution pH is maintained between 6.2 and 6.5 as specifi ed by AWWA method 4500-Cl G/ClO2-D. The USEPA does not “approve” commercial DPD delivery systems. The eXact® Strip Micro CL (DPD-1) for Free Chlorine, the eXact® Strip Micro CL (DPD-4) for Total Chlorine, and the eXact® Strip Micro Cd (DPD-1) for Chlorine Dioxide meet your reportable testing requirements because the eXact® Strip Micro strips deliver the same chemicals in identical proportions. Consult with your local health department for offi cial regulations.

COMPONENT (FREE CHLORINE) AWWA 4500-CL G CHLORINE+ EXACT® EZAnhydrous DPD sulfate 1.5% 1.5%

Anhydrous Na2HPO4 33.4% 33.4%

Anhydrous KH2PO4Na2 64.0% 64.0%

EDTA 1.1% 1.1%

ABOUT

Ready Snap® 3 is a calibration verifi cation solution with predetermined color to verify the accuracy of your Chlorine+ eXact® EZ photometer.

The easy 3 step procedure (snap, fi ll, and test) allows for quick verifi cation of test parameters. Each box contains 10 ampoules of 10mL solution with no dilution necessary.

For Ready Snap® 3 values, visit us online at sensafe.com/exact-ez-chlorine-plus/

METHOD VERIFICATION

Page 27: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

27

Sensafe.com

MODEL AQUATIC HEALTH CODEAND NSF/ANSI 50 CERTIFICATION

The Model Aquatic Health Code (MAHC) is a set of guidelines published by the Centers for Disease Control and Prevention (CDC). This document brings together the latest knowledge based on science and best practices to help state and local government offi cials develop and update pool codes. They may use the code in whole, choose to use parts, or modify to fi t their needs. Use of the MAHC is intended to save time and resources spent individually developing and updating codes across the country, while giving agencies the benefi t of the latest science and best practices to help keep pools fun, safe, and healthy.

The MAHC requires NSF/ANSI 50 certifi cation of water quality testing devices (WQTD) used in recreational facilities such as public swimming pools, interactive fountains, and Water parks. Third-party certifi cation to NSF/ANSI 50 allows manufacturers to make verifi ed claims regarding the performance, accuracy and operating range of their WQTD. The performance testing of a WQTD involves accuracy and repeatability testing on two different lots of new production. Unlike most NSF/ANSI 50 certifi cations, WQTDs require follow-up testing of the product at the end of the manufacturer’s specifi ed shelf life. Certifi ed products are given an accuracy rating to one of three levels: L1, L2 or L3, with L1 being the highest accuracy rating.

TEST RANGE ACCURACY RATING

PARAMETERS

Total Bromine 0 - 17 ppm L1

Free Chlorine 0 - 12 ppm L1

pH 6.4 - 8.4 pH L1

TECHNICAL SUPPORTPlease visit sensafe.com/exact-ez-chlorine-plus/for the latest technical information and how-to-videos. For additional technical support, call (803) 329-9712.

Industrial Test Systems, Inc.1875 Langston Street, Rock Hill, SC 29730800-861-9712 or 803-329-9712 [email protected] | sensafe.com

ITS Europe, Ltd.The UK Centre for Homeland SecurityBuilding 7, Chilmark, Salisbury, Wiltshire SP3 5DU UK +44 1722 717911 [email protected] | itseurope.co.uk

Industrial Test Systems, Inc.

th292988

Manufacturing quality products since 1989

Page 28: DIGITAL ACCURACY FOR RELIABLE WATER CARE · DIGITAL ACCURACY FOR RELIABLE WATER CARE Certified to ... Replace the base fi rmly with pressure while tightening the screw. The meter

† Value provided represents best possible accuracy under laboratory conditions, but may vary throughout the detection range. For a complete list of accuracies throughout all ranges, please visit sensafe.com/exact-ez-chlorine-plus/. 1 Requires the use of 2 strips if reading is above 6 ppm.

PATENT INFORMATION US Patent #7,333,194; Euro Pat No. 1 725 864 DE FR UK; South Africa Pat No 2007/0628 by Industrial Test Systems, Inc., 1875 Langston Street, Rock Hill, SC USA.EXACT® is a registered trademark of Industrial Test Systems, Inc. Rock Hill, SC USA.

EXACT® MICRO TESTS & REAGENTS

PARAMETER / TEST PART # RANGE ppm

% BEST ACCURACY

# OFTESTS

Bromine, Total (DPD-4) 486644 0.00 – 17.0 NSF-50 L1 100

Chlorine Dioxide (DPD-1) 486633 0.00 – 6.0 5 100

Chlorine, Free (DPD–1) 1 486637 0.00 – 12.0 NSF-50 L1 100

Chlorine, Total (DPD–4) 1 486670 0.00 – 12.0 5 100

Chlorine, Total High 486672 1.0 – 200 5 50

Hydrogen Peroxide 486648 0.5 – 130 5 50

Ozone (DPD-4) 486634 0.00 – 2.00 10 100

Peracetic Acid (PAA) 486675 2 – 590 5 100

pH 486639-II 6.4 – 8.4 pH NSF-50 L1 100

R080318-C+EZQG

Chlorine+ADVANCED PHOTOMETER SYSTEM

NSF Certifi ed Tests

Visit us online: sensafe.com/exact-ez-chlorine-plus/for up–to–date product information.

SENSAFE ITSSENSAFE

Industrial Test Systems, Inc.1875 Langston StreetRock Hill, SC 29730 1–800–861–9712—Inside the U.S.A.1–803–329–9712—Outside the U.S.A.Fax:1–803–329–9743Email: [email protected]

ITS EUROPE, LTDThe UK Centre for Homeland SecurityBuilding 7, ChilmarkSalisbury, Wiltshire, SP3 5DU, UKPhone: +44 (0)1722 717911Email: [email protected]

MANUFACTURED BY DISTRIBUTED BY