Development of Sampling System for Investigations on Pollutant Transport
-
Upload
marton-balczo -
Category
Documents
-
view
212 -
download
0
Transcript of Development of Sampling System for Investigations on Pollutant Transport
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
1/6
DEVELOPMENT OF SAMPLING SYSTEM FOR
INVESTIGATIONS ON POLLUTANT TRANSPORT
Mrton BalczBudapest University of Technology and Economics, Department. of Fluid Mechanics
Zoltn SzucsnBudapest University of Technology and Economics, Department. of Fluid Mechanics
Gbor KalmrBudapest University of Technology and Economics, Department. of Fluid Mechanics
Istvn GoricsnBudapest University of Technology and Economics, Department. of Fluid Mechanics
ABSTRACT
In case of wind tunnel modelling of pollutant transport in the atmospheric boundary
layer tracer gas is used as pollutant. Observance of modelling and similarity laws provide
that the tracer gas concentration obtained from wind tunnel modelling is proportional to the
real pollution concentration. Simultaneous sampling in the different points of the modelled
area and thereafter the storage of samples are necessary, while the samples pass the gasdetector serially and the concentration of each sample is determined. The sampling is called
representative on condition the local wind speed and the sampling velocity are equal
(isokinetic sampling). However the local wind speed is varied from point to point thereby the
sampling speed has to be varied simultaneously. At the Department of Fluid Mechanics a
multi-channel automatic sampling system has been developed which is appropriate for
simultaneous and isokinetic sampling.
This paper presents the key components, operation and the results of preliminary tests of the
system.
Gas samples are stored in pneumatic cylinders actuated by linear stepper motors. The way of
gas samples is controlled by valves. The concentration of samples is measured by gas
chromatograph equipped with a flame ionization detector (FID). The whole measurement
system is controlled by a Labview program running on PC.
KEYWORDS
Pollutant transport modelling, PC controlled sampling system, isokinetic sampling
1 INTRODUCTIONThe micro scale dispersion investigations have growing importance especially in
cities. The widely used mathematical modelling of pollutant transport requires validation.
Beside the field tests where some unpredictable factors - for example uncontrollable and
fluctuating wind and background concentration during the measurements, etc. - causeuncertainties, the physical modelling is the next possibility. In case of wind tunnel
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
2/6
investigations controllable parameters can be used. The approaching wind, the background
concentration, the tracer release can be kept constant or can be measured. Naturally the
reliability of results is based on the quality of modelling. This latter is ensured by
conformance of modelling laws. A VDI Guideline on the wind tunnel modelling of
atmospheric flows [1] collects these requirements.
2 EXPERIMENTAL SETUP
Figure 1 shows a typical measurement arrangement. The tracer gas, usually methane
(CH4) or ethane (C2H6) which substitutes the real pollutant, is released by digital mass flow
controllers. After eventual mixing with air, the tracer is led to the sources. The distribution of
tracer gas over the modelled area has to be similar to the pollution distribution over the real
terrain, so special point, line or area sources are built in the model.
Figure 1. Experimental setup
Sampling of polluted air happens by suction through plastic tubes from the measurement
points and has to fulfil the following requirements:
a) Simultaneous samplingb) Isokinetic sampling
The first requirement means that samples from all measurement points have to be taken at the
same time. The second requirement states that the sampling velocity has to be equal to the
local wind velocity. Due to various local wind speeds the suction must be set to different flow
rates. The new mchannel sampler unit developed at the Department of Fluid Mechanics fulfils
both requirements. After sampling, the unit puts the samples serially into the detector unit.
Details of these units are presented in the next paragraphs.
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
3/6
3 THE SAMPLING AND MEASURING SYSTEM
The multi channel sampler equipment has been built of pneumatic elements. Figure 2
shows the sketch of the system. Each sample channel consists of a 100ml pneumatic cylinder
actuated by a computer controlled linear stepper motor. The sampling velocity is determined
by the velocity of the piston. Since each cylinder possesses an own step motor and every step
motor is individually controlled, the velocity of the piston thereby the sampling velocity canbe varied. Two electromagnetic valves have been coupled up the cylinder. All channels are
connected to a mechanical multiplexer, an earlier 48-port pressure scanning valve. Depending
on the valve positions each piston can:
a) Suck sample from the wind tunnel model into the cylinder
b) Put sample from the cylinder off the unit (exhaust)
c) Put sample from the cylinder through the mechanical multiplexer towards the
detector unit
Figure 2. The sampling system
The sampling process is the following (* - position does not matter):
Channel Action Purpose Position
valve 1 valve 2 multiplexer
Cleaning steps
all pull out piston from the
cylinder (each channel withthe calculated isokinetic
velocity)
move false sample from the
connection tube between modell andsampling unit into the cylinder and
suck correct sample into the tubes
left * *
all push piston in remove false sample from thecylinder and vent
right left *
Sampling step
all pull out piston from thecylinder (each channel with
the calculated isokineticvelocity)
fill cylinder with sample left * *
Measurement steps
1 push piston in (constant
velocity)
push sample of channel 1 through the
multiplexer towards the detector
right right 1
2 push piston in (constant
velocity)
push sample of channel 1 through the
multiplexer towards the detector
right right 2
... Repeat measurement steps until samples from all channels have been forwarded to the detector
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
4/6
The detector unit contains a gas chromatograph equipped with a flame ionization
detector (FID) and a 6-port, two-position sampling valve (see Figure 3). A FID can detect
hydrocarbons in a hydrogen flame. The carbon ions in the flame are exposed to an electric
field between two electrodes and cause an electric current between the electrodes that is
proportional to the amount of carbon in the flame. Detailed description of the FID is to find in
the literature e.g. in [2].The sampling valves main function is
a) to collect a known volume of sample gas ( 1 st position)
b) to put that known volume of sample into the FID. ( 2nd position)
In the 1st (standby) position the sample comes through and fills a 50l sample loop and
is exhausted. Through the other ports, air (also called carrier gas), goes to the FID detector.
After switching to 2nd (operate) position the carrier gas input is connected to the sample loop
thereby the known volume of sample determined by the volume of sample loop is pushed by
the carrier gas into the FID detector. At this valve position the sample coming from the
sampler unit is exhausted directly into the environment. After a certain time the valve is
switched again into standby position and the cycle is finished.
(a)
(b)
Figure 3. The FID and the sampling valve in standby (a) and operate (b) position.
After the known volume tracer gas and air mixture has been injected to the FID
detector, the output voltage of the electrometer connected to the FID detector fast increases.
The area under this peak is proportional to the concentration of tracer gas and air mixture (see
Figure 4). The mean concentration has been determined as an average of at least three or
optimal five peaks. This means that for the accurate results the sample loop has to fill and
inject three or five times from the same sample (cylinder). This requirements and the dead
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
5/6
volume of the valves and tubing has been determined the required volume of the sample
cylinder (100ml).
(a) (b)Figure 4. Concentration peaks on the signal checking monitor (a) calibration principle (b)
Before each wind tunnel measurement, a calibration of the FID is necessary. For this
purpose, the FID unit is connected to a calibration gas source with known concentration and
this known concentration is correlated to the area of the measured electrometer voltage peak.
A second point is given by the zero concentration of the carrier gas and the zero peak area
belonging to it, so the linear calibration equation can be determined.
4 CONTROL SYSTEM
The whole measurement process can be controlled by computer (Figure 5). Digital
mass flow controllers are connected via RS-485 cables and HART protocol (RS-232 can onlyhandle 1 unit, RS-485 up to 32 units).
Figure 5. Control and data acquisition system
The multi channel sampler is controlled by an USB digital I/O device, of which digitallines can set the valves and the multiplexer. Each linear stepper has an own microcontroller
-
7/29/2019 Development of Sampling System for Investigations on Pollutant Transport
6/6
chip which can start and stop, change direction and speed of the motor, and also manages the
two limit switches. Both valves and steppers have amplifier circuits to amplify TTL digital
I/O signals to the voltage and current required by the specific device. A separate USB device
reads the analogue output voltage of the FID and can also steer the electromagnetic actuator
of the 6-port sampling valve.
All devices can be controlled from a program written in National Instrument Labview.The program is planned to run a measurement fully automatically.
5 PROJECT STATUS
After the discussion of several sampling methods, the present sampler configuration
was accepted. In mid 2005 a 2-channel test system was built. A similar system (without
isokinetic sampling) in Karlsruhe, Germany has been visited [3]. Repeatability and linearity
of the FID unit were checked using calibration gas of different concentrations and were found
good. After first tests of the sampler, reduction of dead volumes (thinner tubing and valve
connectors) was proposed. Instead of joining the tubes coming from the channels towards the
FID in an aluminium block, the use of a mechanical multiplexer was recommended. At the
moment further tests are carried out concerning measurement accuracy of the whole system.
6 CONCLUSIONS
As a result of development work at Department of Fluid Mechanics an automatic, PC
controlled sampling and measuring system was born. The system can contain up to 48
individually controlled channels. Thereby in case of pollutant transport investigation in wind
tunnel simultaneously 48 sampling points can be appointed. Using this new device the
accuracy, the repeatability and the reliability of sampling-based concentration measurements
can be increased.
REFERENCES1. VDI 3783 Part 12. Environmental meteorology, Physical modelling of flow and
dispersion processes in the atmospheric boundary layer, Application of wind tunnels,
Beuth Verlag GmbH, 2000
2. H.R.Trnkler, E. Obermeier: Sensortechnik. Handbuch fr Praxis und Wissenschaft
Springer, 1998
3. Christof Gromke, Ph.D. student, oral communication, Laboratory of Building- and
Environmental Aerodynamics, Institute for Hydromechanics, University of Karlsruhe,
Germany, 11.24.2005
AFFILIATIONS
Corresponding author: Mrton Balcz, Ph.D. student, Department of Fluid Mechanics,Budapest University of Technology and Economics, H-1111 Budapest, Bertalan L. u. 4-6.,
tel./fax: +36 1 4631560, +36 1 4633464,[email protected]
Authors:
Zoltn Szucsn, research engineer, Department of Fluid Mechanics, Budapest University of
Technology and Economics, [email protected]
Gbor Kalmr, research engineer, Department of Fluid Mechanics, Budapest University of
Technology and Economics, [email protected]
Istvn Goricsn, assistant professor, Department of Fluid Mechanics, Budapest University of
Technology and Economics, [email protected]
Acknowledgements:
Authors acknowledge the support of Hungarian projects OTKA T037730 and NKFP
3A/088/2004.