Detroit Medical Center - Biomech2

download Detroit Medical Center - Biomech2

of 61

Transcript of Detroit Medical Center - Biomech2

  • 8/13/2019 Detroit Medical Center - Biomech2

    1/61

    Concepts in OrthopaedicBiomechanics

    Basic Science LecturesDepartment of Orthopaedic Surgery

    Detroit Medical Center

    Michele J. Grimm, Ph.D.Director of Orthopaedic Biomechanics

  • 8/13/2019 Detroit Medical Center - Biomech2

    2/61

    Mechanical Concepts -- Friction

    Friction -- Resistance generated between two objects when

    one slides over the other

    Friction results in a force parallel and opposite to the

    direction of motion of an object

    Direction of

    motionDirection of

    friction force

  • 8/13/2019 Detroit Medical Center - Biomech2

    3/61

    Mechanical Concepts -- Friction

    Dependent upon the normalforce between the objects

    and the coefficient of

    friction for the two objects

    Fk  = µ

    k *N

    Every pair of materials

    will have a different

    coefficient of friction

    Coefficient of friction isdependent on the surface

    roughness of the objects and

    the molecular interactions

     between the surfaces

    Fk 

    W = mass*gravity N = W Fk = µk*N F > Fk to move block 

    W = mass*gravity N = W*cosf Fk = µk*N F > Fk to move object  F 

    N f 

    Fk  

  • 8/13/2019 Detroit Medical Center - Biomech2

    4/61

    Question 21

    In the Figure, if the coefficient of friction between the 10

    kg box and the floor is 0.2, in order for the box to slide the

    horizontal force component FH must exceed

    (1) 2 kg

    (2) 10 kg

    (3) 50 kg

    (4) FV 

    (5) F

    F

    10 kg

  • 8/13/2019 Detroit Medical Center - Biomech2

    5/61

    Question 21

    In the Figure, if the coefficient of friction between the 10

    kg box and the floor is 0.2, in order for the box to slide the

    horizontal force component FH must exceed

    (1) 2 kg

    (2) 10 kg

    (3) 50 kg

    (4) FV 

    (5) F

  • 8/13/2019 Detroit Medical Center - Biomech2

    6/61

    Question 22

    A cervical traction device is applied to the skull of a

     patient as shown in the figure. Assume that the head

    weighs 100 N and the coefficient of friction between the

     back of the head and the bed is 0.2. What is the minimum

    weight required in order to provide force at the cervicalspine?

    (1) 10 N

    (2) 20 N

    (3) 55 N

    (4) 100 N

    (5) 200 N

  • 8/13/2019 Detroit Medical Center - Biomech2

    7/61

    Question 23

    A cervical traction device is applied to the skull of a

     patient as shown in the figure. Assume that the head

    weighs 100 N and the coefficient of friction between the

     back of the head and the bed is 0.2. What is the minimum

    weight required in order to provide force at the cervicalspine?

    (1) 10 N

    (2) 20 N 

    (3) 55 N

    (4) 100 N

    (5) 200 N

  • 8/13/2019 Detroit Medical Center - Biomech2

    8/61

    Mechanical Concepts -- Lubrication

    Lubrication -- use of a material of low friction coefficient

     between the two surfaces to reduce the frictional force

     between the two materials

    Often lubricants are a fluid (oil, water, synovial fluid), butmay also be a powder such as graphite

    Lubricant acts to form a layer between the surface

    irregularities of the materials, adheres to both surfaces, and

    aids in slip between the surfaces

  • 8/13/2019 Detroit Medical Center - Biomech2

    9/61

    Mechanical Concepts -- Lubrication

    Boundary layer lubrication fills the valleys and coats the

     peaks of the material surfaces, but still allows contact

     between the opposing materials

    Fluid film lubrication occurs when relative motion between

    the surfaces is at a high enough velocity for the surfaces no

    longer to be in contact

    Ex: Tires hydroplaning on wet streets

  • 8/13/2019 Detroit Medical Center - Biomech2

    10/61

    Lubrication of Joints

    Low coefficient of friction between cartilage surfaces is due tofluid-film lubrication

    Synovial fluid -- static loads (no motion)

    Cartilagenous exudate (water) from loading of cartilage in

    cyclical manner Osteoarthritis results in increased water content in the articular

    cartilage causing

    Increased cartilage permeability which results in poor fluid-

    film generation for lubricationDecreased tissue stiffness which alter the normal load-

    carriage mechanisms

    Both mechanisms result in accelerated wear and further break-

    down of the articular cartilage

  • 8/13/2019 Detroit Medical Center - Biomech2

    11/61

    Mechanical Concepts -- Wear

    Wear is defined as the removal of the surface material bymechanical action

    Adhesive wear occurs due to the sliding contact of two rough

    surfaces

    Examples includemetal on metal

     bone on bone

    Abrasive wear occurs due to the sliding contact of a rough

    surface on a softer material Examples include

    stainless steel on polyethylene (femoral head implant on

    acetabular component)

     bone on cartilage

  • 8/13/2019 Detroit Medical Center - Biomech2

    12/61

    Mechanical Concepts -- Wear

    Wear results in particulate debris which can cause an

    inflammatory response and may affect implant behavior

    The amount of wear that occurs depends on

    The coefficient of friction between two surfaces

    The hardness of the two surfaces The normal force between the surfaces

    The contact area between the surfaces

    Increased friction will result in increased wear due to the

    greater shear force at the contact surface

  • 8/13/2019 Detroit Medical Center - Biomech2

    13/61

    Mechanical Concepts -- Wear

    Greater contact area between the surfaces

    increases the area over which wear occurs

  • 8/13/2019 Detroit Medical Center - Biomech2

    14/61

    Question 23

    The wear rate of a polyethylene acetabular total hip

    component is most dependent on the

    (1) Femoral neck offset

    (2) Femoral head diameter

    (3) Component position

    (4) Acetabular component backing

    (5) Acetabular component fixation technique

  • 8/13/2019 Detroit Medical Center - Biomech2

    15/61

    Question 23

    The wear rate of a polyethylene acetabular total hip

    component is most dependent on the

    (1) Femoral neck offset

    (2) Femoral head diameter  

    (3) Component position

    (4) Acetabular component backing

    (5) Acetabular component fixation technique

  • 8/13/2019 Detroit Medical Center - Biomech2

    16/61

    Material vs. Structural Properties

    A structure is comprised of one or more materials Behavior of a structure depends not only on the mechanical

     properties of the material (aka material properties), but on

    Dimensions: thickness, width, length

    Manner of supporting load

    Type: uniaxial, biaxial, bending, torsion

    Configuration: column, eccentric column, end or center-

    loaded beam

    Orientation of load with respect to geometry

    Shape

    Structures develop different combinations of tensile,

    compressive, and shear stresses in various loading conditions

  • 8/13/2019 Detroit Medical Center - Biomech2

    17/61

    Structures

    Basic structural shapes include

    Beam - Two dimensions are smaller

    than other one

    Plate - One dimension is smaller thanother two

    Rod - Solid cylinder

    Tube - Hollow cylinder

    Cable - Supports tensile stress only Sphere

    Arch

  • 8/13/2019 Detroit Medical Center - Biomech2

    18/61

    Loading Conditions -- Column Loading

    Load axis lies along center of vertical rod or cylinder acting as a

    column or support

    Compression (or tension) distributed evenly across horizontal

    cross-section

    Shear stress develops on planes at 45 degree angles to axis of

    loading

  • 8/13/2019 Detroit Medical Center - Biomech2

    19/61

    Loading Conditions -- Torsion Caused by twisting force on two

    ends of rod or cylinder

    Develops shear stress along long

    and short axes of cylinder

    Tensile and compressive stresses

    develop at angles of 45 degrees tolong axis

    Bones undergo torsional loading

    during internal and external

    rotation Torsional failure may occur in

    incidents such as a twisting fall

    when skiing

  • 8/13/2019 Detroit Medical Center - Biomech2

    20/61

    Loading Conditions -- Bending Caused by uniaxial loading of a curved structure or eccentric

    loading of a straight beam

    As a beam bends under an applied load, one surface becomes

    convex and one becomes concave

    Tensile stress develops on convex surface

    Compressive stress develops on concave surface

    Stresses are maximum at surfaces and become zero at neutral

     plane

    Bones have slight curvature and thus experience bending under

    normal conditions

    Neutral Axis, Stress = 0 

    Region of Tensile Stress

    Region of Compressive Stress

  • 8/13/2019 Detroit Medical Center - Biomech2

    21/61

    Loading Conditions

    Eccentric Column Loading

    Common in skeleton

    Load is applied off center

    In addition to direct compression,

    additional compression and tension

    due to bending

    Can be seen as bending with an

    additional compressive force

    Loading of femoral neck an

    example of eccentric column

    loading

  • 8/13/2019 Detroit Medical Center - Biomech2

    22/61

    Unixaxial Loading

    Strength and stiffness are proportional to the cross-sectional area of

    the structure

    Examining the simple uniaxial loading condition allows us to see the

    effect of geometry when material properties are constant

    Case 1:

    Ligaments of equal length and different diameters, A and 2A,under the same force rupture at the same deformation (D)

    Stiffness of ligament 2A will be twice that of ligament A

    Force required to cause deformation (D) which results in

    rupture for ligament 2A will be twice that of ligament A

    2A

    A

    Load

    Deformation

  • 8/13/2019 Detroit Medical Center - Biomech2

    23/61

    Uniaxial Loading

    Case 2: Ligaments of equal diameters and different lengths, L and

    2L, under the same force will rupture at different

    deformations

    Stiffness of ligament 2L will be half that of ligament LDeformation at rupture for ligament 2L will be twice

    that of ligament L

    Force at rupture, and therefore strength, will be equal for

    the two ligaments

    Load

    Deformation

    2LL

  • 8/13/2019 Detroit Medical Center - Biomech2

    24/61

    Moments

    A Rotational Moment is produced by a force out of line

    with the axis of a structure

    The moment is equal to the product of the force and the

    distance from the point of force application to the

    center of rotation of the structure

    M = F*L

    L

    F

  • 8/13/2019 Detroit Medical Center - Biomech2

    25/61

    Equilibrium

    For an object to be at rest, or in equilibrium, the followingrules apply

    The sum of forces and moments must equal zero:

    Forces act through the center of mass of an object

    Moments tend to rotate an object about its center ofmass (or about a fixed rotational point, such as a pin or

     joint) and are caused by off axis forces

    Forces and moments must be treated separately in

    determining equilibrium conditionsA  B  A  B  A  B 

     A + B = R 

     Ac = Bd 

    Rc d

  • 8/13/2019 Detroit Medical Center - Biomech2

    26/61

    Bending

    Bending Rigidity: load required to produce a unit deflection of a beam

    Increases with increasing beam dimension which resists the

     bending [R a thickness3]

    Decreases with increasing "working length" of beam [R a 

    (1/length)3]

    Increases with redistribution of material away from neutral axis

    This distribution of material is defined by the cross-sectional

    moment of inertia (Ixx)

  • 8/13/2019 Detroit Medical Center - Biomech2

    27/61

    Bending

    Working Length: length of structure subjected to bending

    In internal fixation, it is the length between the two

    main points where bone may experience discontinuities

    in bending or torsional moments

    For example -- fracture surfaces

    L

  • 8/13/2019 Detroit Medical Center - Biomech2

    28/61

    Bending Bending Strength of a structure: maximum bending load

    that can be sustained The bending strength is proportional to the ratio c/Ixx 

    c is the distance from the neutral axis to the surface

    of the structure

    Ixx is the moment of inertia

    Limit to the increase in strength with moment of inertia

    reached when the material is too thin to support the

    load and begins to buckle

  • 8/13/2019 Detroit Medical Center - Biomech2

    29/61

    Question 24

    In the analysis of mechanical forces, a moment is caused

     by:

    (1) The product of a mass and its velocity

    (2) Two equal, parallel forces of opposite sense

    (3) Two equal, parallel forces of the same sense

    (4) A force acting at a distance from a point

    (5) A system of concurrent forces acting on a free body

    of equilibrium

  • 8/13/2019 Detroit Medical Center - Biomech2

    30/61

    Question 24

    In the analysis of mechanical forces, a moment is caused

     by:

    (1) The product of a mass and its velocity

    (2) Two equal, parallel forces of opposite sense

    (3) Two equal, parallel forces of the same sense

    (4) A force acting at a distance from a point 

    (5) A system of concurrent forces acting on a free body

    of equilibrium

  • 8/13/2019 Detroit Medical Center - Biomech2

    31/61

    Torsion

    Torsional Rigidity: torque required to produce a unit angular

    twist of a structure

    Decreases with increasing working length

    Increases with increasing structural radius (R a r 4) through

    the polar moment of inertia (J)

    For rectangular cross-sections, rigidity is related to the

    dimensions of the circle which can fit within the cross-

    section

  • 8/13/2019 Detroit Medical Center - Biomech2

    32/61

    Torsion

    A cylinder with a slot or cut on one side will have a greatly

    reduced torsional rigidity

    Reduced by a factor of 5 to 10

    Width of cut relatively unimportant

    Torsional Strength: maximum torque that can be sustained

    Most efficient shape for bending and torsion is a

    cylindrical tube because it provides an equivilent area

    moment about all axes

  • 8/13/2019 Detroit Medical Center - Biomech2

    33/61

    Question 25

    A set mass of material is most resistant to both torsion and

     bending when it is shaped like a:

    (1) A cylinder

    (2) A solid rod

    (3) A flat beam

    (4) An I-beam

    (5) An elliptical beam

  • 8/13/2019 Detroit Medical Center - Biomech2

    34/61

    Question 25

    A set mass of material is most resistant to both torsion and

     bending when it is shaped like a:

    (1) A cylinder  

    (2) A solid rod

    (3) A flat beam

    (4) An I-beam

    (5) An elliptical beam

  • 8/13/2019 Detroit Medical Center - Biomech2

    35/61

    Stress Concentrations

    A structure first begins to yield, exhibiting permanent

    deformation, and then fails at the point where the highest

    stress is experienced

    In almost all structures, except axially loaded columns or

    cubes, there is some variation in stress over the structure

    due to the geometry and mode of loading

    Thus for a beam in bending, the maximum stresses

    occur at the surfaces perpendicular to the direction of

    loading

  • 8/13/2019 Detroit Medical Center - Biomech2

    36/61

    Stress Concentrations

    Any discontinuities in a structure will produce stressconcentrations

    Examples: notches, holes, sharp angles, grooves, or other

    sharp transitions in structure

    Increases stress at location by 2 to 4 times and creates a highstress gradient

    In most cases, failure will initiate at the stress-concentrating

    defect. The exception is if the defect exists in a region of low

    stress in a structure under nonuniform stress. Stress defects in orthopaedic structures can be the result of

    surgical intervensions, such as bone graft removal or the

    insertion of bone screws, or the development of structural

    defects such as result from metastatic cancer

  • 8/13/2019 Detroit Medical Center - Biomech2

    37/61

    Stress Distribution

    In a composite structure, stresses

    and displacements will be

    distributed among the component

    materials dependent upon their

    material properties

    Structures in parallel

    Displacement will be constant for

    two components

    Stress in each component will

    depend on its elastic modulus:higher modulus --> higher stress

    Load distribution depends on

    stress and component area

    fraction

  • 8/13/2019 Detroit Medical Center - Biomech2

    38/61

    Stress Distribution Structures in series

    Stress will be constant for two components

    Displacement will depend on each

    component's elastic modulus

    Higher modulus --> lower strain

    Same principles apply for all modes of

    loading

    Failure of the composite will occur when

    stress in one component reaches its ultimate

    strength

    Failure in one area reduces the area

    supporting the load

    Stress increases in remaining regions -->

     potential failure

  • 8/13/2019 Detroit Medical Center - Biomech2

    39/61

    Loading of Bone

    Bone material is strongest in compression, intermediately

    strong in tension, and weakest in shear

    Strength also depends on orientation of trabeculae or osteons

    with respect to the load

    Pure Tensile Loading:

    Primarily due to muscle action

    Tensile fractures relatively uncommon, strict tension on a

     bone will typically result in dislocation of a joint

    Most tension fractures are avulsions of muscle origins or

    insertions or occur in sesmoid bones

  • 8/13/2019 Detroit Medical Center - Biomech2

    40/61

    Loading of Bone Pure Compressive Loading:

    Compression fractures occur primarily in trabecular bone

    Cause is ususally traumatic impact or excessive muscle forces

    or a severe reduction in mechanical properties, such as in

    osteoporosis

    Cracks propagate through lamina and trabecular structure is

    crushed

    Resistance of bone to tensile or compressive fracture is

    dependent on:

    Relative strength of cortical and trabecular components

    Fraction of cross-sectional area of each component

    If ultimate strength is exceeded in either component, failure

    will generally occur

  • 8/13/2019 Detroit Medical Center - Biomech2

    41/61

    Loading of Bone

    Bending

    Many bones loaded in eccentric column mode which

    induces bending

    Bending fractures may result from

     Normal modes of loading at increased forces

    Jumping from a height

    End loaded (cantilever) beam loading

    Stopping fall with hand resulting in radial/ulnar

    fractureThree-point bending

    Bumper injuries

    "Boot top" skiing fractures

  • 8/13/2019 Detroit Medical Center - Biomech2

    42/61

    Loading of Bone Bending

    Fracture typically initiates on the tension surface and

    then propagates medially

    As long bone cross-sections are asymmetrical, the

    rigidity and stress generated will not be the same for all

    directions of bending

    Torsion

    Produces high shear and tensile stresses resulting in

    spiral fractures

    Fracture will occur at the level of lowest polar moment

    of inertia

    Typically mid-region of diaphysis

  • 8/13/2019 Detroit Medical Center - Biomech2

    43/61

    Loading of Bone -- Stress Concentrations

    A hole in the cortex of a bone, with a size of to 20% of the

    diameter of the bone, has the following effects

    Reduces bending strength by 50 percent if located on

    tensile surface

    Reduces torsional strength by 50 percent and energy

    absorption capacity by 75 percent if located at junction

    of mid and distal tibia

    Whether hole is empty or filled with a screw appears to

    make little difference in effect on strength

    Remodeling of bone or early filling of holes with new

     bone tends to eliminate stress concentrations

  • 8/13/2019 Detroit Medical Center - Biomech2

    44/61

  • 8/13/2019 Detroit Medical Center - Biomech2

    45/61

    Question 26

    Cortical bone exhibits its greatest strength when subjected

    to:

    (1) Tension

    (2) Torsion

    (3) Bending

    (4) Pure shear

    (5) Compression

  • 8/13/2019 Detroit Medical Center - Biomech2

    46/61

    Question 26

    Cortical bone exhibits its greatest strength when subjected

    to:

    (1) Tension

    (2) Torsion

    (3) Bending

    (4) Pure shear

    (5) Compression 

  • 8/13/2019 Detroit Medical Center - Biomech2

    47/61

    Torsion and Bending of the Spine

    Torsion applied to the spine is transmitted between the

    vertebrae primarily through the facet joints, which are

    compressed on one side and pulled apart on the other

    If the torsional force is great enough, bony breakage

    occurs on the compression side and ligaments ruptureon the tensile side

    When the spine is flexed

    The interspinous ligaments, ligamenum flavum, and

     posteriour longitudinal ligaments on the posterior sideare in tension

    The disc and vertebral body are compressed on the

    anterior, concave side

  • 8/13/2019 Detroit Medical Center - Biomech2

    48/61

    Torsion and Bending of the Spine

    Possible protection mechanisms for an intervertebral discA. If the posterior ligaments are stiffer than the disc, the

    amount of deformation in forward flexion depends more on

    the geometry of the posterior elements than on the disc. High

    compressive stresses are prevented in the disc.

    B. Contraction of the most posterior aspect of the muscles

    tends to compress the posterior elements and stretch the

    anterior elements. Stresses on the spine due to bending are

    reduced.

    C. If a strong contraction of abdominal muscles exists, the

    abdominal cavity can support some of the weight of the

    upper body because lateral deformation of the abdomen is

     prevented. The load the spine must bear is decreased.

  • 8/13/2019 Detroit Medical Center - Biomech2

    49/61

    Internal Fixation Devices and

    Mechanical Considerations -- Screws

    To minimize stress concentrations at the bone-

    screw interface (when no plate is used) screws

    should be countersunk to reduce sharp angleinteractions

    To reduce bending of the screw, which can

    damage both the screw and the bone

    Screws should be inserted perpendicular tothe bone or plate surface

    Screw-plate junction should be rounded

    Internal Fi ation De ices and

  • 8/13/2019 Detroit Medical Center - Biomech2

    50/61

    Internal Fixation Devices and

    Mechanical Considerations -- Plates

    Plates are designed for several different applications Thick plates are designed to facilitate primary bone healing

    Thin plates are designed to allow physiological load bearing

    following healing

    A thick plate will have 5 to 10 times the rigidity of a thin one

    In design, the effect of screw holes, and the resultant stress

    concentrations, must be considered

    A uniform cross-section must be obtained, including at screw-

     plate interfaces, in order to eliminate stress-concentrationinduced weaknesses

    I t l Fi ti D i d M h i l

  • 8/13/2019 Detroit Medical Center - Biomech2

    51/61

    Internal Fixation Devices and Mechanical

    Considerations -- Bone-Plate Structures

    Bending RigidityA bone-plate structure is 100 times more rigid if the plate is

     placed on the side of the bone that experiences tension

    during bending rather than compression

    A fractured bone can withstand compressive loading, butnot tensile loading

    If the bending axis is at right angles to the plate, the rigidity

    is between the above two conditions, with the load falling

    much more on the screwsA two plate system, oriented at 90 degrees to each other, is

    insensitive to the plane or direction of bending and is more

    rigid than a single plate in its optimum orientation

  • 8/13/2019 Detroit Medical Center - Biomech2

    52/61

  • 8/13/2019 Detroit Medical Center - Biomech2

    53/61

    Question 27

    A 23-year old man who sustains a transverse tibial mid-shaft

    fracture undergoes internal fixation of the fracture with a

    stainless steel plate and screws. One screw hole of the six-hole

     plate is left open because it is located over the fracture site.

    Firm fixation is obtained when two screws are placed proximal

    to the fracture and three screws are placed distal to the fracture.Six weeks later, the plate fails at the open screw hole due to:

    (1) Failure to bone graft that fracture.

    (2) Delayed union due to stress shielding the bone.

    (3) Stress concentration at the open screw hole.

    (4) Use of a stainless plate rather than a titanium plate.

    (5) A difference in torsional stiffness between the proximal

    and distal halves of the plate.

  • 8/13/2019 Detroit Medical Center - Biomech2

    54/61

    Question 27

    A 23-year old man who sustains a transverse tibial mid-shaft

    fracture undergoes internal fixation of the fracture with a

    stainless steel plate and screws. One screw hole of the six-hole

     plate is left open because it is located over the fracture site.

    Firm fixation is obtained when two screws are placed proximal

    to the fracture and three screws are placed distal to the fracture.Six weeks later, the plate fails at the open screw hole due to:

    (1) Failure to bone graft that fracture.

    (2) Delayed union due to stress shielding the bone.

    (3) Stress concentration at the open screw hole.

    (4) Use of a stainless plate rather than a titanium plate.

    (5) A difference in torsional stiffness between the proximal

    and distal halves of the plate.

  • 8/13/2019 Detroit Medical Center - Biomech2

    55/61

    Response of Tissues to Load -- Bone

    Wolff's Law -- Living bone adapts dependent upon the

    stress or strain acting on the tissue

    There is an “ideal” loading situation for each bone

    which maintains normal bone mass

    When additional force is applied, bone is deposited to

    reduce the stress

    When force is removed, bone is resorbed as it no longer

    is needed to distribute the force

    These changes occur over time, not instantaneously

    When designing an implant or fixator, the ideal is to

    maintain the same loading conditions for the bone

  • 8/13/2019 Detroit Medical Center - Biomech2

    56/61

    Response of Tissues to Load -- Joints

    Mechanism of effect of immobilization (removal of load) onligaments and tendons is unknown

    Immobilization tends to make joints "stiffer" in flexion and

    extension

    Require greater force to move joint Possibly due to:

    Adhesions of articular tissue

    Pannus formation

    Capsular contractions - increased pressure

    Ligament shortenings

    Tensile properties of joints affected differently than "motion

     properties"

    Effect of Immobilization on Bone Ligament

  • 8/13/2019 Detroit Medical Center - Biomech2

    57/61

    Effect of Immobilization on Bone-Ligament-

    Bone Structure -- Tensile Properties

    Strength and energy absorbed to failure reduced

    significantly for joint

    Elastic modulus of ligaments reduced significantly

    Ligament modulus dominates deformation as it is much

    less than that of bone and components are in series

    Significant weakening of bone-ligament junction

    Failure ususally the result of an avulsion injury

    Joint may take up to 1 year to regain normal propertiesafter 8 weeks of immobilization

    Ligament properties seem to recover quickly

    Bone-ligament junction recovers more slowly

  • 8/13/2019 Detroit Medical Center - Biomech2

    58/61

    Question 28

    After six weeks of immobilization of the knee, the tensile load-deformation curves (structural properties) of the bone-ligament-

     bone complex and the stress-strain curves (mechanical

     properties) of the knee ligament substance will be:

    (1) Unchanged

    (2) Significantly softened with no change in load at failure

    and tensile strength

    (3) Significantly stiffened with no change in load at failure

    and tensile strength

    (4) Significantly softened with decreased load at failure and

    tensile strength

    (5) Significantly stiffened with decreased load at failure and

    tensile strength

  • 8/13/2019 Detroit Medical Center - Biomech2

    59/61

    Question 28

    After six weeks of immobilization of the knee, the tensile load-

    deformation curves (structural properties) of the bone-ligament-

     bone complex and the stress-strain curves (mechanical

     properties) of the knee ligament substance will be:

    (1) Unchanged

    (2) Significantly softened with no change in load at failure

    and tensile strength

    (3) Significantly stiffened with no change in load at failure

    and tensile strength

    (4) Significantly softened with decreased load at failure and

    tensile strength

    (5) Significantly stiffened with decreased load at failure and

    tensile strength

  • 8/13/2019 Detroit Medical Center - Biomech2

    60/61

    Stress Shielding in Bone

    Metallic implants with high elastic moduli typically used

    in bone repair and joint replacement

    Design of implant or fixation devices often creates parallel

    structures

    Femoral implant

    Bone plate

    Metal will bear larger portion of stress than bone, stress in

     bone is reduced below normal levels

    (1) Reduction in stress results in resorption of bone

    (2) Reduction in bone cross-sectional area causes

    further reduction in percent of stress borne by bone

    (3) Results in further resorption

  • 8/13/2019 Detroit Medical Center - Biomech2

    61/61

    Stress Shielding in Bone

    Therefore:

    Don't want stiffest bone plate or implant possible, even

    though these may be stronger as well

    Must consider methods to transfer more stress to the

     bone

    Porous coatings

    Press-fits

    Collared prostheses

    Less rigid prosthesis or plate to allow for greater

    stresses in bone