Transcript of Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer
![Page 1: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/1.jpg)
INSTRUMENTAL ANALYSIS
LAB.
Mr Haydar A.M.S
BSc, MSc
Faculty of Science
Chemistry Department
Email: [email protected]
![Page 2: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/2.jpg)
Exp. 3 Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer
ObjectivesIn this experiment, you will
• Prepare a set of Co+2 standard solutions of known concentrations,
• Measure the A or %T of the known solutions and construct a calibration curve, and
• Measure the A or %T for unknown solution and determine its concentration from the calibration curve.
![Page 3: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/3.jpg)
Ultraviolet/Visible Molecular Absorption Spectrometry
The instrument used to measure the amount of light absorbed by the sample is a
spectrophotometer.
![Page 4: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/4.jpg)
Components of a Spectrophotometer
4
![Page 5: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/5.jpg)
5
Wavelength Selectors
![Page 6: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/6.jpg)
Light source
Grating
Rotating the gratingchanges the wavelength going through the sample
slits
slits
Sample
filter
Phototube
Wavelength Selectors
When blank is the sample Io is determined
otherwise I is measured
Separates white lightinto various colors
detects light &measures intensity
- white light of constant intensity
6
![Page 7: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/7.jpg)
7
![Page 8: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/8.jpg)
8
Sample Containers
![Page 9: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/9.jpg)
Range
![Page 10: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/10.jpg)
![Page 11: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/11.jpg)
![Page 12: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/12.jpg)
Transmittance (T) is defined as the amount of light passing through the sample solution (I) divided by the amount of incident radiation (Io).
T = I/ Io
•Transmittance is often expressed as a percentage or
%T = I/Io × 100%
The Absorbance (A) of a medium is defined by the equation:
A = log Io/I = -log T
![Page 13: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/13.jpg)
As the radiation is absorbed in the sample, the total intensity of radiation is reduced as it travels through the sample. This results in a non-linear relationship between transmittance and concentration
Absorbance (A) is based on the amount of light absorbed by the solution and is defined as the log of the inverse of the transmittance.
A = log10
(1/T)
Within limits, the relationship between absorbance and concentration is linear.
![Page 14: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/14.jpg)
The blank
• The blank contains all substances except the analyte.
• Is used to set the absorbance to zero:
A blank = 0
• This removes any absorption of light due to these substances and the cell.
• All measured absorbance is due to analyte.
Analyte is a substance undergoing analysis.
![Page 15: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/15.jpg)
Beer’s lawBeer’s law states that the amount of light absorbed by a solution (colored) is directly
proportional to the path length (b) through the medium and the concentration (c) of the absorbing species,
A (absorbance) = log Io/I = a b c
• Where (a) is the proportionality constant called the absorptivity
• b is the path length of the sample i.e. the inside cross section of the sample cuvette (cm).
• When the concentration is expressed in (mol/L), the absorptivity is called the molar absorptivity and is given the symbol (є)
A = є b c (units of є = L.mole-1.cm-1)
![Page 16: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/16.jpg)
Determination the calibration curve of cobalt nitrate by spectrophotometer
![Page 17: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/17.jpg)
Procedure
1. Obtain TWO matched cuvettes, the first cuvette is for distilled water and the second one is
for the cobalt solution.
2. Prepare the following solutions
From stock solution (0.2M)Co(NO3)2.6H2O
(M)
Absorbance
(A)
0.02
0.04
0.06
0.08
0
Part A Calibration Curve
![Page 18: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/18.jpg)
Procedure Continued
3. Set the wavelength dial 505 nm
4. Measure the absorbance of the blank (water).
5. Measure the absorbance of the cobalt solutions and record the absorbance
6. Empty and rinse the cuvettes thoroughly with water.
7. plot the absorbance versus concentration
8. Measure LOD, LOQ and sensitivity
![Page 19: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/19.jpg)
Procedure Continued
1. Measure the absorbance of the unknown concentration
2. use a calibration curve that you plot it in part A and determine the concentration of the unknown solution.
Part B Determination the absorbance of the unknown
![Page 20: Determination the Calibration Curve of Cobalt Nitrate by Spectrophotometer](https://reader030.fdocuments.net/reader030/viewer/2022020218/55a41dfd1a28ab384f8b46f1/html5/thumbnails/20.jpg)
20
1.00
0.20
0.60
Absorbance
0.30 0.40 0.50
0.80
0.40
0.10 0.20 Concentration in mol / Litre
0 0