Detectors and Preamplifiers

16
Detectors and Preamplifiers • Detector deliveries • Customer Acceptance Tests • Available detectors • AGATA triple detectors • Scanning program ter IKP University Cologne AGATA week 8-11 July 2008 U

description

Detectors and Preamplifiers. Detector deliveries Customer Acceptance Tests Available detectors AGATA triple detectors Scanning program. P. Reiter IKP University Cologne AGATA week 8-11 July 2008 Uppsala. Recent detector tests and future deliveries. Turkish detectors - PowerPoint PPT Presentation

Transcript of Detectors and Preamplifiers

Page 1: Detectors and Preamplifiers

Detectors and Preamplifiers

• Detector deliveries

• Customer Acceptance Tests

• Available detectors

• AGATA triple detectors

• Scanning program

P. Reiter IKP University Cologne AGATA week 8-11 July 2008 Uppsala

Page 2: Detectors and Preamplifiers

Detector Delivery Owner Status

A001* 11/2005 GANIL rejected – hole trapping

A001 06/2006 GANIL accepted

A002* 04/2006 TU Munich rejected – leakage currents

A002 06/2007 TU Munich accepted

A003* 11/2006 Liverpool rejected – leakage currents

A003 10/2007 Liverpool accepted

A004 09/2007 Ankara rejected – segments out of specification

A004* 07/2008 Ankara CAT at IKP, Cologne

A005 12/2007 Stockholm CAT at CEA, Saclay & IKP

B001* 11/2005 Padova rejected – segments out of specification

B001 06/2007 Padova rejected – leakage currents

B002 08/2006 GANIL accepted

B003 06/2007 Liverpool accepted

B004 12/2007 Ankara rejected – missing segment in capsule

B004* 06/2008 Ankara CAT at IKP, Cologne

B005 12/2007 Stockholm CAT at CEA, Saclay

C001 12/2005 Padova accepted

C002 07/2006 GANIL accepted

C003 03/2007 Liverpool accepted, broken, leakage currents

C004 09/2007 Ankara accepted

C005 12/2007 Stockholm CAT at CEA, Saclay

Page 3: Detectors and Preamplifiers

Detector Delivery Owner Status

A001 06/2006 GANIL Accepted, available

A002 06/2007 TU Munich Accepted, available

A003 10/2007 Liverpool Accepted, available

B002 08/2006 GANIL Accepted, available

B003 06/2007 Liverpool Accepted, available

C001 12/2005 Padova Accepted, available

C002 07/2006 GANIL Accepted, available

C004 09/2007 Ankara accepted

Page 4: Detectors and Preamplifiers

Recent detector tests and future deliveries

Turkish detectors

• B004 (repaired) delivered to IKP, Cologne, test started

• A004 (repaired) delivered July, 1st 2008

• shipment of A004, B004, C004 to Turkey autumn 2008

Swedish detectors

• A005, B005, C005 tests ongoing at CEA

- test results of A005 to be confirmed at IKP, Cologne

- test of B005 started

Liverpool detector C003: leakage current, out of warranty

Italian detectors

• A006 announced delivery July (week 29) 2008

• B006 announced delivery July (week 31) 2008

• B001 announced delivery July 2008

Page 5: Detectors and Preamplifiers

Summary available detector

• 8 detectors accepted (A1,A2,A3 B2,B3 C1,C2,C4) • 1 detector to be repaired at Canberra (B1)• 5 detectors ongoing test (A4, A5, B4, B5, C5)• 1 detector broken (C3) ----------------------------------------------------------------------------------------• 3 (Sweden A005, B005, C005) ongoing CAT at Saclay• 2 (Turkey A004, B004) ongoing CAT at Cologne

• 1 (Italy C001) scanning at Liverpool until July

• 1 (Italy B001) not accepted after CAT, not repaired• 1 (UK C003) is broken after CAT, needs repair

• 3 (Turkey A004, B004, C004) are shipped to Turkey fall 2008-----------------------------------------------------------------------------------------• At the moment 6 detectors are available for AGATA triple clusters

A001, A002, A003 B002, B003 C002

Page 6: Detectors and Preamplifiers

AGATA triple cluster detectors

First AGATA triple *A001 : B002 : C002 :

*Goes to LegnaroJuly 2008

First AGATA triple *A001 : B002 : C002 :

*Goes to LegnaroJuly 2008

Second AGATA tripleA003 : B003 : C004 : *

*Goes to Turkey in fall 2008,Replaced by C005

Second AGATA tripleA003 : B003 : C004 : *

*Goes to Turkey in fall 2008,Replaced by C005

Third AGATA tripleA002 : B005 or B001: * C001: **

*B005 after test or B001 after delivery & test** C001 after scanning

Third AGATA tripleA002 : B005 or B001: * C001: **

*B005 after test or B001 after delivery & test** C001 after scanning

Page 7: Detectors and Preamplifiers

Scanning program

LiverpoolC001 * A005

* Goes to IKP,Assembly of secondTriple cluster by CTTJuly 2008

LiverpoolC001 * A005

* Goes to IKP,Assembly of secondTriple cluster by CTTJuly 2008

OrsayS002 *B005 :

* Commissioning, confirmLiverpool results

OrsayS002 *B005 :

* Commissioning, confirmLiverpool results

See talks by A. Boston C. UnsworthA. KorichiF. CrespiThursday, 14:00-15:40

Page 8: Detectors and Preamplifiers

Future detector tests and triple assembly

Detectors for period fall and winter 2008/09• 13 detectors may be available A1, A2, A3, A5, A6 B1, B2, B3, B5, B6 C1, C2, C5 • A6, B1, B6 delivery July 2008

• redelivery of turkish detectors A4, B4, C4 ?

• January - June 2009 italian detectors C6, A7 ?

• detector repair of C3 ?

• summer 2008 order of german detectors A8, B7, C7 ?

Page 9: Detectors and Preamplifiers

Talks by • A. Wiens, IKP Cologne

• A. Goergen, CEA Saclay

• D. Quirion, Canberra

• B. Bruyneel, IKP Cologne

• G. Pascovici, IKP Cologne

Page 10: Detectors and Preamplifiers

1st Tripel Cluster Energies: Single vs Triple cryostat

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59,6 keV Single 59.6 keV Triple Specification low energy

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

A0

01

– F

WH

M [

keV

]B

00

2 –

FW

HM

[ke

V]

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59.6 keV Single 59.6 keV Triple Specification low energy

0

0.5

1

1.5

2

2.5

3

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

1332,5 keV Single 1332,5 keV Triple Specification high energy

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

Core A1A2A3A4A5A6 B1B2B3B4B5B6 C1C2C3C4C5C6 D1D2D3D4D5D6 E1E2E3E4E5E6 F1 F2 F3 F4 F5 F6

59,6 keV Single 59,6 keV Triple Specification low energy

C0

02

– F

WH

M [

keV

]

Resolution 60keV line Resolution 1.33MeV line

Page 11: Detectors and Preamplifiers

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

] F

WH

M[k

eV

59,6 keV Single 59,6 keV Triple Specification low energy

2nd Tripel Cluster Energies: Single vs Triple cryostatA

003

FW

HM

]ke

V[

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

] F

WH

M[k

eV

59,6 keV Single 59,6 keV Triple Specification low energy

0

0,2

0,4

0,6

0,8

1

1,2

1,4

1,6

C0

04

F

WH

M ]k

eV

[

59,6 keV Single 59,6 keV Triple Specification low energy

C00

4 F

WH

M [k

eV]

B

003

FW

HM

[keV

]

A00

3 F

WH

M [k

eV]

Resolution:

60keV line

Page 12: Detectors and Preamplifiers

Tripel Cluster Energy resolution: Single vs Triple cryostat

Excellent energy resolution in two AGATA triple

clusters

Mean values segments @ 60 keV :

especially at low energies (electronic noise):

- glued detectors less microphonics

- HEKO AGATA preamp

- electronic assembly (grounding, cabling)

First AGATA tripleA001 : 1011 +/- 53 eVB002 : 1039 +/- 70 eVC002 : 965 +/- 63 eV

First AGATA tripleA001 : 1011 +/- 53 eVB002 : 1039 +/- 70 eVC002 : 965 +/- 63 eV

Second AGATA tripleA003 : 1052 +/- 65 eVB003 : 995 +/- 73 eV C004 : 1148 +/- 104 eV *

Second AGATA tripleA003 : 1052 +/- 65 eVB003 : 995 +/- 73 eV C004 : 1148 +/- 104 eV *

* high frequency oscillation to be removed

Page 13: Detectors and Preamplifiers

For any 1406keV single event in the detector:

-3.0

-2.5

-2.0

-1.5

-1.0

-0.5

0.0

0 6 12 18 24 30 36

Hit segment number

Ba

selin

e S

hift

in A

1 (

keV

)Cross talk measurements

A2..A6

B1 F1

Co

re to

se

g X

talk

= u

nd

erst

ood

!

Segment labeling:

Sectors: A...F

Rings: 1...6

X

Bart Bruyneel, Andreas Wiens

Page 14: Detectors and Preamplifiers

B002 in Triple Cryostat @ 5000 V

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0 200 400 600 800 1000 1200

segment "i" + 36x segment "j"

Xta

lk a

mpl

itude

s fro

m s

eg "

i" to

seg

"j"

[%]

Measured Cross talk

Core to Seg Xtalk (theory)

Smallest capacities in ring 2

A2

B2 C2 D2 E2 F2

Page 15: Detectors and Preamplifiers

Cross talk in first AGATA Triple Cryostat

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

-0.25%

-0.20%

-0.15%

-0.10%

-0.05%

0.00%

0.05%

0 200 400 600 800 1000 1200 1400 1600

A001 B002 C002

Co

re B

Co

re C

Co

re A

Co

re C

Co

re A

Co

re B

Sector A Sector F

Sector A Sector F Sector A Sector F

A00

1B

002

C00

2

Page 16: Detectors and Preamplifiers

AGATA Triple Clusters: Summary & outlookTwo AGATA Triple Cluster Detectors operational:- first & second triple: cryostat, mechanics, vacuum, thermal properties of preamps o.k.- seg-preamps, core preamps & pulser with high dynamic range operational- very good energy resolution at low energies - no cross talk between detectors observed in 1st ATC between detectors- commissioning of 2nd ATC ongoing

Next steps:- transport of 1st ATC to Legnaro, first tests with digital electronics at Legnaro July 2008-- finish commissioning of 2nd ATC, - cross talk tests-- assembly of 3rd ATC constrained by -available detectors

ATC1

ATC2

ATC3