Dentall Unit

download Dentall Unit

of 40

description

Dentistry

Transcript of Dentall Unit

  • 0197

    IMPORTANTThis manual provides operating instructions for the CLESTA. The instructions contained in this booklet should be thoroughly read and

    understood before operating the unit. After the installation has been completed, keep this manual in a safe place and refer to it for future maintenance.

    If you have any questions about this Manual or this product, please contact us. If manual becomes unreadable or is lost, please request a new manual by contacting your dealer.

    Installation should be conducted by authorized personnel only. Follow instructions on installation manual.

    DENTAL UNIT

    INSTALLATION andOPERATING INSTRUCTIONS

  • Page

    SAFETYPRECAUTIONS------------------------------------------------- 1

    1.GENERALINTRODUCTIONS---------------------------------------- 5

    2.INSTALLATIONINSTRUCTIONS----------------------------------- 8 3.OPERATINGINSTRUCTIONS--------------------------------------- 19

    4.DAILYMAINTANANCE----------------------------------------------- 23

    5.MAINTANANCEANDINSPECTION5-1.GUIDEFORDAILYMAINTENANCEANDINSPECTION---- 27(Maintenanceandinspectionbyuser)5-2.DUIDEFORPERIODICALCHECK-UP------------------------ 29

    6.ELECTROMAGNETICCOMPATIBILITY(EMC)----------------- 30

    7.LISTOFCOMPATIBLEHANDPIECES----------------------------- 33

    TABLE OF CONTENTS

  • Intended Use of the ProductThisproductisanactivetherapeuticdeviceintendedfortheexclusiveusefordiagnoses,treatmentsandrelativeproceduresofdentistry.The product must be operated or handled by the qualified dentists or by dental staffs under

    thesupervisionofthedentist.Suchdentistsordentalstaffsshouldinstructand/orassistthepatientstoapproachtoandleavefromtheproduct.Patientsshouldnotbeallowedtooperateorhandletheproductunlesshe/sheissoinstructed.Theproductissuppliedtogetherwiththehandpieceslikeelectricmicromotor,airturbineand/ormotor,scalerandsoon.

    Compatibility of HandpiecesUsethecompatiblehandpiecesasshownontheattachedlistforthisunit.(Listofcompatiblehandpieces).

    Important Notes In case of the troubles, please contact Takara Belmont offices or your dealers.

    Donotdisassembleorattempttorepair. Disassembly, repair or modifications should only be done by a qualified repair technician.

    Attempts at disassembly, repair or modifications may lead to abnormal operation and accidents.

    Disposal of residue material Pleaserequestaspecialcontractorwhenyoudisposeamalgam.

    In case of disposal of equipment Whendisposingtheunit,appropriatelydisposecomplyingwithallcurrentapplicableregulationsandlocalcodes.InEUarea,EUdirective2002/96/EConwasteelectricalandelectronicequipment(WEEE)isappliedonthisproduct.Inthisdirective,environmentconsciousrecycling/abandonmentisobligated.

  • EC REP

    SN

    Protective earth (ground) ON (power) OFF (power)

    Chair last position Chair auto return Chair preset 1 Chair preset 2

    To raise the chair To recline the backrest To lower the chairTo raise the bacrest

    Chairmanual control

    Fiber optic handpiece light on/off

    Micro motor rotation speed

    Scaler power control

    Micro Motor Forward/Reverseselect

    Syringe Saliva ejector Bowl flush

    Cupfiller Service outlet (water) Water Air

    Water heater Serial number

    Authorizedrepresentative in the European community

    Manufacturer Date of manufacture

    Non-ionizingradiaion

    Separatecollection for electrical and electronicequipment

    Type Bapplied parts

    alternating current

    CautionIt means caution, warnings, or possibility to danger.

    Refer to instructionmanual/booklet

    Inthismanual,onthelabelsoronthecontrolpanelofCLESTA,followingsymbolsareused.Confirm the meaning of each symbol.

    SYMBOLS

  • Beforeuse,readtheSafetyprecautionscarefullytoensureproperuse.Thefollowinginformationisdesignedtoensuresafeuseofthisproductandtopreventinjuryanddamage to you and others. The precautions contained here are classified depending on the severity and degree of

    imminenceofpossibleinjuryordamageresultingfromimproperuse.Besuretofollowalltheinformation,whichisimportantforsafety.

    SAFETY PRECAUTIONS

    Thissymbolindicatesthatitisrecommendedtofollowtheseprecautionsforsafety.

    Classification of precautions Severityanddegreeofimminenceofpossibleinjuryordamage

    WARNING

    CAUTION

    NOTICE

    Thissymbolindicatesthatignoranceoftheseprecautionsmayleadtosevereinjuryorevendeathasaresultofimproperuse.Thissymbolindicatesthatignoranceoftheseprecautionsmayleadtomildormoderatephysicalinjuryordamagetopropertyasaresultofimproperuse.

    WARNING1.BesuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftimeBesuretoturnoffbreakersforequipmentintheclinicwhenthisproductwillnotbeusedforalongperiodoftime(followingthecompletionofwork,duringthesuspensionofbusiness,etc.).Insulationdegradationmaycause electrical fire.

    2.BesuretoturnoffthemainswitchuponcompletionofworkorduringworkbreaksBesure to turnoff themainswitchuponcompletionofworkorduringworkbreaks.Thisprevents incorrectoperationduetoaccidentalcontactandassociatedhazards.

    4.DonotplaceanundueloadonthearmDonotgetonorplaceanundueloadonthearmofthisunitordentalchairarmrest.Thiscouldcausetheunittotoppleorotheraccidents.5.BesuretoestablishagroundingconnectionBesure toestablishapropergroundingconnection.(Refer toavendorforgroundingconnection.)Failureorelectricleakagemayleadtoelectricshock.6.Neverdisassemble,repairormodifythisproductIndividuals other than certified repair technicians should not disassemble or attempt to repair and modify this product. This could lead to an accident, failure, electric shock or fire.

    3.DonotsitonotherthanseatWhenthebackrestisattheforwardposition.donotsitonorplaceanundueloadontheheadrestorlegreatofdentalchair.Thiscouldcausetheunittotoppleorcoulddamagetheunit.

    7.UsewithcautioninthepresenceofelectromagneticinterferencewavesDonotplace thisproductaroundequipmentgeneratingelectromagneticwaves(includingcommunicationsequipment,elevators,etc.)asincorrectoperationofthisproductmayoccurinthepresenceofelectromagneticinterferencewaves.Donotuseequipmentgeneratingelectromagneticwaves,suchasmobilephones,aroundthisproduct.

    -1-

  • SAFETY PRECAUTIONS

    WARNING8.BesuretoturnoffthemainswitchwhenelectrocauteryisinuseBesuretoturnoffthemainswitchwhenelectrocauteryisinuse,becausenoisemaycauseincorrectoperationofthisproduct.

    10.Donotplaceobjectsweighing3kgormoreontheDoctor'stable

    Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.Refertothesectionofmaintenance.

    11.BesuretousethemirrorcoverBesuretousethemirrorcoverofthedentallightwhenthelightisturnedon.Directcontactwithlampsmaycauseburns.SeetheInstructionManualofthedentallightforfurtherinformation.12.Besuretoturnoffthepowerwhenreplacinglamps

    9.Ensurethemaintenanceofthisproduct

    DonotplaceobjectsweighingthreekilogramsormoreontheDoctor'stable.ThiscouldcausedamagetotheDoctor'stable,defectivefunctionoraccidents.

    Besuretoturnoffthepowerwhenreplacingthedentallight.Thiscouldresultinelectricshock.Useonlydedicatedhalogenlamps.Immediatelyafterahalogenlamphasburntout,thelampandthelampholderarestillhot.Replacethelampaftertheycooldown.Donottouchhalogenlampswithbarehands.SeetheInstructionManualofthedentallightforfurtherinformation.13. Immediately wipe off any water spills or leakage on the floor

    Immediately wipe off any water spills or leakage on the floor. Decreased strength of the floor may lead tophysicalinjuryincludingfall,orpropertydamage.

    14.UsewithcautiononpatientswithacardiacpacemakerUsethisproductwithextremecautiononpatientswithacardiacpacemaker.Inthecaseofanyabnormalitiesinpatientsduringuse,immediatelyturnoffthisproductanddiscontinueuse.

    CAUTION1.OnlyexperiencedpersonnelshouldusethisproductOnlydentistsorotherdentalprofessionalsshouldusethisproduct.

    2. Confirm safety before use.

    Before use, confirm that the parts are correctly and safely operating and that there are no obstacles around this product.

    4.Discontinueuseifyoufeelthatsomethingiswrong

    Keepyoureyesonpatients(especially,children)sothatmischieforinadvertentoperationofequipmentwillnotleadtounexpectedaccidents.

    3.Payattentiontopatientsandchildren

    Alwaysbecarefultoinspectthisproductforlooseness,rattling,tilting,wobbling,sounds,temperature,odors,etc. Immediately discontinue use at the first feeling that something is wrong.

    -2-

  • SAFETY PRECAUTIONS

    CAUTION5.keepyoureyesonthepatientduringoperation.

    6.PayattentionduringmovementoftheDoctor'stablePayattentiontosurroundingswhenyoumovetheDoctor'stable.Injurybythetipsofhandpieces,etc.,mayoccur.BesuretomovetheDoctor'stablebyholdingthehandleoftheunit.Besuretomoveup&downtheDoctor'stablebyreleasingthebalancearmbrakeoftheunit.7.DonotplaceanythinghotontheDoctor'stable

    12.PrecautionsforcleaningtheresincoverForcleaning,donotusecleaningagentscontainingsolventorabrasives,thinnersoroil-basedalcohol(butanolandisopropylalcohol),whichmaycausecracks.

    DonotplaceanythinghotontheDoctor'stable.Thiscouldcausedeformationordiscoloration.

    10.DonotsmackorrubthisproductDonotsmackorrubthisproductforcefully.Thiscouldcausedamagetocoversordefectivefunction.

    8.Cautionswhenadjustingtheheightofthecart-typeinstrumenttable(Carttypeonly)Afteradjustingtheheight,besuretofastentheloosenedlockscrew.Failuretoheedthiswarningmaycausethetabletodrop,resultinginanaccident.9.Observethecarthose(Carttypeonly)Donottreadonthecarthose.Failuretoheedthiswarningmayresultindamagetothehose,aswellaspeopletrippingover.

    11.Precautionsforcleaningaspittoonbowl

    Confirm that the patient is seated in the proper position. Keep your eyes on the patient during the operation.Payspecialattentiontosurroundingsatautomaticoperationofthedentaltreatmenttable.Damagetothebackrest,stoolorDoctor'stablemayoccur.

    Thespittoonbowlismadeofglass.Handlewithcare.Donotwashitwithhotwater.Otherwise,itmaybreak.

    13.ImmediatelywipeoffdrugsolutionwhenitcomesintocontactwiththisunitShoulddrugsolutionorwatercomesintocontactwiththisunit,immediatelywipeitoffwithadrysofttowel,etc.Thiscouldresultindefectivefunctionorelectricleakageaswellasspottingorrusting.

    14.ClosethewatermainvalveandturnoffthemainswitchuponcompletionofworkBesuretoclosethewatermainvalveandthemainswitchattheendofeachworkdaytopreventwaterleakagefromoccurring.

    16.Payattentionduringtheheadrestoperation

    Besuretooperateswitcheswithyourhands,except thefootcontroller,whichisoperatedwithyourfoot.Operationwithbodypartsotherthanhandsmaycausedamageorincorrectoperation.

    17.BecarefulnottoheatanemptywaterheaterExercise caution as heating of an empty water heater may result in burning of the heater, leading to fire.

    Do not allow hands, fingers, or hair to become entangled in the moving parts of the headrest during operationg

    15.Besuretooperateswitcheswithyourhands

    -3-

  • SAFETY PRECAUTIONS

    CAUTION18.Precautionsforcleaningtheoperationpanel(membraneswitches)

    The water tank is intended only for use with purified water, distilled water and pure water. Do not use mouthwash orelectrolyzedwater,suchasConCoolorpovidoneiodine,astheymaycausecloggedtubingoraffectinternalvalvesandequipment.22. Precaautions for insert the bowlflush nozzle and the cupfiller nozzle to the spittoon bowl

    Penetrationofdropletsofsanitizingsprayintothebackoftheoperationpanelmaybeassociatedwithswitchfailure.Useapapertowelsoakedwithsanitizingsolutiontocleanthesurfaceoftheoperationpanel.

    20.Setthepressureofthewatertankat200kPaorlessAdjust theairsupplypressurefor thewater tankto200kPaorless.Anexcessivelyhighpressuremaycausedamagetothewatertank.

    19.Precautionsforcleaning

    21. Do not use water other than purified water, distilled water or pure water for the water tank

    NOTICE1.TroubleshootingandcontactinformationInthecaseofanyproblems,discontinueuse,turnoffthemainswitchandcontactthedealerorourcompany.

    2.CheckoperationofthecompressorWithnoairsupplied,thisproductdoesnotoperateevenafterturningonthemainswitch.Turnonthepowerofthecompressorbeforeoperatingthisproduct.

    4.Usetheturbinewithawatercheckvalve

    Thewaterunitisintendedforusewithtapwater.Cautionshouldbeexercisedastheuseofwaterotherthantapwater(waterthroughasterilizerofwatersystems,etc.)mayresultinfailureofequipment.

    5.HandlingofequipmentinthecaseofapowerfailurePutthehandpieceintheholderandturnoffthemainswitchifequipmentstopsworkingduringuseduetoapowerfailureorotherreasons.

    3.Precautionswhenusingwaterotherthantapwater

    Usetheturbinewithawatercheckvalve.Contactthedealerorourcompanywhenaturbinewithoutawatercheckvalvewillbeused.

    Neverusesandpaper,metalscrubbrushesandabrasivecleaningagentstocleantheunit.Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagentstoavoidcorrosionofmetals,etc.

    23.ReadthedocumentsaccompanyingthevariouspiecesofequipmentBeforeuse,besure tocarefullyreadthepackageinsertsandInstructionManualsaccompanyingthevariouspiecesofequipment(includingoptionalarticles)toensureproperuse.

    Aftercleaningofthespottoonsection,besuretoinsertthenozzlesstraightlytospittoonbowl.Ifnot,itmaybedamagetheo-ringanditmayleakthewater.

    -4-

  • -5-

    Bowl Flush Nozzle

    Cuspidor Top

    Assistant Holder Bar

    Saliva Ejector

    Assistant Holder

    Light Pole

    Cupfiller Nozzle

    Syringe

    High Volume Evacuator

    Foot Control

    Dental Light

    Horizontal Arm

    Light Post

    Cuspidor Unit

    Balance Arm

    Dental Size Film Viewer

    Doctor Table TrayControl Panel

    Handle

    Umbilical Hose

    Junction Box

    Syringe

    Waste Receptacle

    Handpiece Holder

    Handpiece Tubing

    Handpieces(Air Turbine, Air Motor, Scaler and etc.)Manufacturers recommend to use the handpieces with CE markings.

    [1] GENERAL INTRODUCTIONS(1)UNITTYPEVARIATIONS&MAJORPARTSIDENTIFICATIONS1.UNITMOUNTTYPE

    Fig.1-1Overview

  • -6-

    Dental Size Film Viewer

    Doctor Table Tray

    Caster

    Cart Base

    Cart Hose

    Junction Box

    Foot Control

    Handle

    Umbilical Hose

    2.SPLITCARTHOLDERTYPEWITHCUSPIDOR

    Fig.1-2Overview

  • 188

    171 287 364

    294493

    76

    529

    98

    98250

    610

    52

    860

    783 65

    0

    423

    521

    562

    905

    330

    14

    38

    180

    330

    9090

    (2)DIMENSIONS

    Fig.1-3Dimensions

    Ratedpowersupply-------------------------------------------------AC230V50/60HZ0.7AAirmainpressure---------------------------------------------------0.45to0.5MPaWatermainpressure------------------------------------------------0.1to0.2MPaNetweight-----------------------------------------------------------80.5kg(withoutdentallight)Dentallight----------------------------------------------------------AL-501Workingenvironment-----------------------------------------------Temperature:(Operating)+5~+40(Storage)-10~+50Humidity:10~80%Atmosphericpressure:600hPa~1060hPaServiceLife-----------------------------------------------------------10Years

    Classification of foot controller ------------------------------------IPX1(applicablestandardIEC60529)Protectionclassagainstelectricshock----------------------------ClassIequipmentAppliedparts---------------------------------------------------------TypeBappliedparts:Handpieceforunit(Listofcompatiblehandpieces)Equipment that is not suitable for use in air, flammable anesthetic gas, oxygen or nitrous oxide

    (3)SPECIFICATIONS

    -7-

    -mm-

  • -8-

    220

    130

    527

    85 121

    135

    11055 70

    70

    311

    243

    175

    CM/OP MIN.1000 SC MIN.1200

    MIN.800

    MIN

    .600

    CENTER OF CLESTA CHAIR

    VACUUMDRAIN

    VACUUMOPERATINGPOWER

    SOURCE

    AIR

    WATER

    [2] INSTALLATION INSTRUCTIONS(1)PREPARINGFORINSTALLATIONTherecommendedlocation,plumbingandlayoutofjunctionboxareshowninFug.2-1,2-2and2-3respectively.1.LOCATIONANDPLUMBNGLAYOUTOFJUNCTIONBOX

    Fig.2-1LocationandPlumbingLayout

    StandardType

  • -9-

    22013

    0

    527

    80 121

    140

    110

    7070

    311

    243

    175

    CM/OP MIN.1000 SC MIN.1200

    MIN.800M

    IN.6

    00

    VACUUM

    DRAIN

    VACUUMOPERATING

    POWER SOURCE

    AIR

    WATER

    CENTER OF CLESTA CHAIR

    BelmontType

    NOTE:CM/OP.....CHAIRMOUNTOVERTHEPATIENTTYPESC............SPLITCARTTYPE

    Fig.2-2LocationandPlumbingLayout

  • -10-

    40

    20 30

    80

    F

    A

    D EG

    B

    C

    VACUUM DRAIN POWER SOURCE &VACUUM OPERATING

    AIR & WATER

    2.PLUMBINGHEIGHTOFJUNCTUONBOX

    Fig.2-3PlumbingHeight

    A:NIPPLEB:SOCKETJOINTC:AIR

    WATER

    D:DRAIN

    E:POWERSOURCE

    SUPPLYCABLEEARTHWIREF:VACUUM

    G:VACUUMOPERATING

    ;PT1/2-1/2(Brass);UnplasticiziedPolyvinylChloridePipeFitting;UnplasticiziedPolyvinylChloridePipe(ShockResistingType)'HI13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipeforWaterWorksService'VP13'18mmO.D.-13mmI.D.;UnplasticiziedPolyvinylChloridePipe'VP40'48mmO.D.-40mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.;PolyvinylChlorideInsulatedandSheathedCable'VVF2.0-2C';PolyvinylChlorideInsulatedWire'VE1.6';UnplasticiziedPolyvinylChloridePipe'VP20'26mmO.D.-20mmI.D.;UnplasticiziedPolyvinylChlorideConduit'VE16'22mmO.D.-16mmI.D.

  • -11-

    1.CHECKTHEPACKINGCheckeachquantityofthepartsforinstallationaslistedbelow. *Standard Specification

    1)SHUTOFFVALVESFORAIR&WATER 1SET2)WOODSCREW 4PCS.3)FLEXIBLEPIPE 2PCS.4)DRAINELBOW 1PCE.5)VACUUMELBOW 1PCE.6)DRAINCAP 1PCE.7)BASKETSTRAINER 1PCE.8)CUPFILLERNOZZLE 1PCE.9)BOWLFLUSHNOZZLE 1PCE.10)SALIVAEJECTORTIP 1PCE.11)VACUUMTIPS(BENT&STRAIGHT) 1SET12)SILICONTIP 10PCS.13)LEVELADJUSTMENTSCREW 4PCS.14)WASTERECEPTACLE 1PCE.15)LIGHTPOLE 1SET16)SILICONMATFORDR.TABLE 1PCE.17)WIRECONNECTOR 3PCS.18)SEAL 4PCS.19)PATIENTCUP 1PCE.20)VIALSWITHSTAND 1SET21)COTTONCONTAINERWITHSTAND 1SET21)MOUNTINGBRACKETCOVER 1PCE.

    2.ATTACHUNITTOCHAIR2-1.MountingbracketAType(Fig.2-4)1)Raisetheupperstructureofdentalchairtothehighestposition.2)InstallthecuspidorunitonMountingBracketATypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.3)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.4)SticktheSealsovertheFlatHeadScrews.5)Removethestopperscrewfromthemountingbracket.6)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopper

    screw.

    [2] INSTALLATION PROCEDURES

    Fig.2-4AttachUnittoChair

    Flat Head Screw

    MountingBracket A

    Level Adjustment Screw

    Cuspidor Unit

    NutSpring Washer

    Mounting Bracket

    Seal

    StopperScrew

    Mounting BracketCover

    StopperScrew

  • -12-

    Dental LightPlastic Wahser

    Light Pole

    Set Screw

    Cover

    Cable from Dental Light

    Cable from Cuspidor Unit

    2-2.MountingBracketBType(Option)(Fig.2-5)1)Raisetheupperstructureofthedentalchairtothehighestposition.2)AttachtheAdaptortoMountingBracketBTypewith4pcs.oftheM10-40SocketHeadCapScrewsandSpringWashers.3)InstallthecuspidorunitonMountingBracketBTypewith4pcs.ofM10-50FlatHeadScrews,SpringWasherandNuts.4)Adjustthelevelofcuspidorunitwith4pcs.ofM10-25LevelAdjustmentScrews.5)SticktheSealsovertheFlatHeadScrews.6)Removethestopperscrewfromthemountingbracket.7)Attachthemountingbracketcoveronthe mounting bracket, and fix with the stopper

    screw.

    Fig.2-5AttachUnittoChair

    Fig.2-6InstallDentalLight

    3.INSTALLDENTALLIGHT(Fig.2-6)1)RemoveCoverfromthebottomoftheunit2)PutPlasticWasherontheLightPole.3)RuntheCablethroughLightPoleandattachtheDentalLighttotheLightPole.4)RuntheCablethroughLightPostandinstallthePole,thentighten2pcs.oftheSetScrews.5)ConnectthewiresfromtheDentalLightandCuspidorUnitinaccordancewiththe identification color.

    6)Re-attachtheCover.

    Flat Head Screw

    Socket Head ScrewSpringWasher

    Mounting Bracket BLevel Adjustment Screw

    Nut

    Adaptor

    Cuspidor Unit

    Spring Washer

    Seal

    Mounting Bracket

    Stopper

    Mounting BracketCover

    Stopper

  • 1 2 3 4 4 5 5 6 7 8

    1 2 3 4 4 5 5 6 7 8

    4 5

    4 5

    L

    C C

    L

    Umbilical Hose

    Umbilical Hose Clamp

    Brown 1/8Blue 5/16

    Yellow 1/4

    Light Green 1/4Wirings for Water Heater & Dental Light

    White Black

    White Black

    Brown 1/8Blue 1/16

    Yellow 1/4

    Light Green 1/4

    Wirings for Water Heater

    Wirings for Dental Light

    Blue Brown

    BlueBrown

    24V Line Voltage

    Small thin Screw Driver

    Electric Terminal

    Wiring

    -13-

    4.CONNECTUMBILICALHOSEWITHJUNCTIONBOX(Fig.2-7,2-8&2-9)ThissectionisonlyforSPLITCARTTYPE.Forothertypeofunit,proceedto5.1)OpenthecoverofJunctionBoxbylooseningtwoscrews.2)LoosenthescrewofUmbilicalHoseClamp.3)Insertallhoses,tubingsandwiringsofUmbilicalHosefromCuspidorunitintoJunctionBoxthroughtheUmbilicalHoseClamp.4)CorrectthetwistoftheUmbilicalHoseastheJunctionBoxisplacedattheplannedlocation.5)SettheendofUmbilicalHoseattheUmbilicalHoseClampthenre-tightenthescrewfortheUmbilicalHoseClamp.6)Connecttubingswithnylonsleevestothe fittings as shown in Fig.2-8 or 2-9.

    7)ConnectwiringstoElectricTerminalasshowninFig.2-8or2-9.

    *HowtoconnectwiringtotheElectricTerminal(Fig.2-10)DuringpressingtheslotoftheElectricTerminalbytheSmallThinScrewDriver,insertthewiringtotheElectricTerminal.

    Fig.2-7ConnectUmbilicalHose

    Fig.2-8 Standard Specification

    Fig.2-9 Low Voltage SpecificationFig.2-10HowtoConnectWiring

  • -14-

    Standard Type Belmont Type

    Center of Clesta Chair Center of Clesta Chair

    Water Shut Off Valve

    Air Shut Off Valve

    Water Shut Off Valve

    Air Shut Off Valve

    ElectricalReceptacle

    Shut Off Valve

    Nipple

    ElectricalReceptacle

    Wood ScrewVacuum Drain Hose

    Drain Hose

    Drain Elbow

    Vacuum Elbow

    Vacuum Drain HoseVacuum Elbow Drain HoseDrain Elbow

    Vacuum Pipe Drain Pipe

    Drain Bush (Option)

    Drain Joint (Option)

    Drain Bush (Option)

    5.CONNECTUTILITYSECTION(Fig.2-11&2-12)1)AttachShutOffValvestotheairandwatersupplylines,pointingtoproperdirectionsforfurtherconnections.2)AttachElectricalReceptacletoMainPowerSupplyCable.(TheElectricalReceptaclemustbepreparedlocally.)3)OpeneachShutOffValvetoremoveanychipsordirtintheairandwatersupplyline,thenclosetheValves.4)OpenthecoverofJunctionBoxbylooseningtwoscrews,andplaceJunction Box at the planned location, then fix it on

    the floor with 4pcs. of Wood Screws.

    5)CuttheDrainHoseandtheVacuumDrainHoseattheproperlength,andconnectandgluethemtoeachelbow.6)ConnecttheVacuumElbowtotheSuctionPipe.7)InserttheDrainElbowintotheDrainPipe.

    *AttachingDrainBushing(Option)totheDrainPipeisdesirable,andDrainElbow should be bonded firmly into the Drain Pipe.

    * For Air Vacuum specification, prepare Drain

    Joint(Option)asshowninFig.2-13,sothatthedrainlineandthevacuumdrainlinemaybeinsertedtogetherintotheDrainPipe.

    Fig.2-11ConnectSupplyLines

    Fig.2-12ConnectDrainLines(1)

    Fig.2-13ConnectDrainLines(2)

  • -15-

    Shut Off ValvePacking

    Filter

    Filter

    Standard Type Belmont Type

    Air Filter Air FilterWater Filter Water Filter

    1 2 3 4 4 5 5 6 7 8

    1 2 3 4 4 5 5 6 7 8

    Cable for Chair Control (9P)

    Cable for Chair Safety Device (4P)

    Cable from Chair

    Main Power Supply Cable

    VacuumOperatingCable (9P)

    Plug

    VacuumOperating Cable

    Cable for Chair Control (9P)

    Cable for Chair Safety Device (4P)

    Cable from Chair

    Main Power Supply Cable

    VacuumOperatingCable (9P)

    Plug

    VacuumOperating Cable

    SeeFig.2-14&2-15.8)ConnectWaterShutOffValveandWaterFilterwithStainlessFlexiblePipeandPacking.ConnectAirShutOffValveandAirFilterinthesameway.

    SeeFig.2-16or2-17.9)PreparetheElectricalPlug,andconnectittoMainPowerSupplyCable.(ElectricalPlugmustbepreparedlocally.)10)ConnectVacuumOperatingCabletotheElectricTerminalasshowninFig.2-16or2-17. (For Central Vacuum Specification only)

    11)Runthethreecablesfromchair(PowerSupplyCablewithplug,Cablewith9PconnectorandCablewith4Pconnector),throughtheholeoftheJunctionBoxwallabovetheUmbilicalHoseintotheJunctionBox.12)ConnectthePowerSupplyCableofthechairtothereceptacle,andconnectothertwocablestothecorrespondingconnectors.

    Fig.2-14ConnectFilters(1)

    Fig.2-15ConnectFilters(2)

    Fig.2-16ConnectElectricalcable(1)

    Fig.2-17ConnectElectricalcable(2)

  • -16-

    6.ATTACHFILMVIEWER(Fig.2-18&2-19)1)RemovetheDoctorTableTraybylooseningeightFlatHeadScrews.2)RunthecablefromFilmViewerintoDoctorTable.3)AttachFilmViewertothebracketforDoctorTablewithtwoPhilipsScrews.4)ConnectthecablefromFilmViewertotheElectricalTerminal.

    7)SETUPCUSIPDORTOP(Fig.2-20)1) Insert Cupfiller Nozzle and Bowl Flush

    Nozzle,andtighteneachnut.2)PutBasketStrainerandDrainCapintothecuspidorbowl.

    Flat Head Screw

    Philips Screw

    DoctorTableTary

    ElectricalTerminal

    FilmViewer

    Electrical Terminal

    Cupfiller NozzleBowl Flush Nozzle

    Drain Cap

    Basket Strainer

    Fig.2-18AttachFilmViewer

    Fig.2-19ConnectWiringsforFilmViewer

    Fig.2-20CuspidorTop

  • -17-

    (3)ADJUSTMENTBEFOREUSE1.AIR&WATERSUPPLYLINE(Fig.2-21&2-22)1)ChecktheAirPressureGaugeonthesideofJunctionBox.ThePressureshouldbeapproximately0.45Mpato0.5Mpaforair.Foradjustmentthepressure,pullupandturntheknobofAirRegulatorwhiletheair is flowing (use 3-way Syringe for example).

    Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise. After adjustment, push down the knob to fix it.

    2)CheckiftheWaterPressureGaugeindicatesthepressureataround0.1Mpato0.2Mpa.AdjustmentcanbedonebyWaterRegulator.Toadjustthepressure,liftandturnthewater regulator knob until the water starts flowing.

    (use3-waySyringeforexample).Toincreasethepressure,turnclockwise.Todecrease,turncounter-clockwise.Afteradjustment,lowertheknobbackinplace.

    3)Astheunitisreadytorun,turnontheMasterSwitchlocatedunderneaththedoctortable(seeFig.2-23.),thenrunthesalivaejectorfor a while to flow the water through the unit and

    makesurenoleakagetakesplaceattheutilitysection.SeeFig.2-23&26.4)TurnallCoolantWaterControlKnobscounter-clockwisefullyandpickupallhandpiecetubingfromtheholder,thendepressDrivePedalofthefootcontrol(SpraySwitch-ON)foramoment to flush out the water from the tubing into the

    cuspidorbowl.

    Pull Up

    Push DownKnob

    Increase decrease

    Air Regulator

    Water RegulatorIncrease

    decrease

    Knob

    Pull UpPush Down

    Water Air

    Air Pressure Gauge

    Water Pressure Gauge

    Master Switch

    ONOFF

    Coolant WaterControl Knob

    Increase decrease

    Fig.2-21InsideofJunctionBox

    Fig.2-22SideViewofJunctionBox

    Fig.2-23BottomViewofDoctorTable

  • -18-

    2.DriveAir&SprayforHandpieces(Fig.2-24,2-25&2-26)1)Connecthandpiecestoeachtubing,andmakeadjustmentonebyone.2)Toadjustthedriveairpressureforeachhandpiece,turntheappropriateAdjustScrewforDriveAironetheAuto-SelectValveinthedoctortable.Turnthescrewclockwisetoclosethelinefully.ThenduringdepressingDrivePedalofFootControlfully,graduallyincrease the sir flow by turning the screw counter-

    clockwiseuptoHandpiecePressureGaugereadsthefullrecommendedpressure.3)Toadjustsprayofhandpiece,useAdjustScrewforCoolantAirontheAuto-SelectValveandCoolantWaterControlKnob.

    Flat Head Screw

    DoctorTableTray

    Auto-SelectValve

    PressureGauge

    Adjust Screw for Drive Air

    Adjust Screw for Coolant Air

    Drive Pedal

    Fig.2-24InsideofDoctorTable

    Fig.2-25Auto-SelectValve

    Fig.2-26FootControl

  • -19-

    (1)DOCTORUNITSECTION1.MASTERSWITCH(Fig.3-1&3-2)1)TurnontheMasterSwitchandthePowerIndicatorislitingreen.2)Aftertheoperation,turnofftheMasterSwitch,thenthesuppliesofair,waterandelectricityareallshutdownattheutilitysection.

    2.TABLEPOSITION(Fig.3-1)Forstandardmodelofthechairmountoverpatienttype,theheightofthedoctortableisalways fixed by Air Brake. The height can

    beeasilychangedbypressingBrakeReleaseButton.

    3.CHAIRCONTROL(Fig.3-3)ThechairmovementcanbecontrolledbypressingChairControlonMembraneSwitchA.Pleaserefertooperatinginstructionsforchair.

    4.CUPLILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-3)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds at

    factory.For bowl flush flushing time is set for 6 seconds

    atfactory.

    5.3-WAYSYRINGE(Fig.3-2&3-4)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedunderneaththedoctortable.

    [3] OPERATION INSTRUCTIONSFilm ViewerPower Indicator

    Membrane Switch A

    Brake Release Button

    Membrane Switch B

    Handpiece Holder

    Master Switch ONOFFAir

    Water

    Flow Control Knobfor 3-Way Syringe

    Coolant Water Control Knob

    2

    LP0

    1

    Cupfiller Switch

    Chair Manual Control

    Chair Automatic Control

    Bowl Flush Switch

    w A

    Air Button

    Water Button

    Syringe Nozzle

    Fig.3-1DoctorUnit

    Fig.3-2BottomView

    Fig.3-3MembraneSwitchA

    Fig.3-43-WaySyringe

  • -20-

    6.OPERATINGHANDPIECE(Fig.3-5)1)Pickupahandpiecefromthepandpieceholder,anddepressDrivePedalofFootControl.2)Forthedry-cutting,turnoffSpraySwitchofFootControl.3)WhiledepressingChipPedal,onlythechipblowcomesoutfromthehandpiece.

    7.FILMVIEWER(Fig.3-6)FilmViewerSwitchislocatedontherightside of film viewer. Press the switch once to

    illuminatetheviewer,pressitagaintoturnitoff.

    8.FIBEROPTICHANDPIECE(OPTION)(Fig.3-7)FiberOpticHandpieceSwitchislocatedontheMembraneSwitchB.Pickupaoptichandpiecefromthehandpieceholderandthehandpiceislit.Returnthehandpiecetotheholderanditisputout.

    9.ELECTRICSCALER(OPTION)(Fig.3-7)PowerControlforElectricScalerislocatedundertheMembraneSwitchB.

    ON

    OFFCoolant Water Switch

    Drive Air Pedal

    Chip Blower Button

    Film Viewer Switch

    Fig.3-5FootControl

    Fig.3-6FilmViewer

    Fig.3-7MembraneSwitchB

    1

    0

    2 3

    4

    5

    Fiber Optic Handpiece Switch

    Power Control for Electric Scaler

    HandpiecesRefertohandpiecemanufacturersoperatinginstructions.

    CAUTION

  • -21-

    (2)ASSISTANTUNITSECTION1.CUPFILLER&BOWLFLUSHCONTROL(ByAIRTIMERSYSTEM)(Fig.3-8)Bypressingeachmembraneswitch,waterissuppliedforacertaintime.For cupfiller, filling time is set for 4 seconds a

    factory.For bowl flush. Flushing time is set for 5 seconds

    atfactory.

    2.3-WAYSYRINGE(Fig.3-4&3-9)1)Bypressingeitherorbothbuttons,Syringeoffersair,waterandspray.2)SyringeTipcanberotatedfreely.3)Toadjustair/water/spray,useFlowControlKnobslocatedinthecuspidorbody.

    3.MASTERVALVEBLOCK(Fig.3-9)Openthesidepanelofthecuspidorbody.Flow Control knobs for Cupfiller and

    BowlFlushareaccessible.

    4.CUSPIDORBODYROTATION&TELESCOPICASSISTANTHOLDERBAR(Fig.3-10)1)CuspidorBodycanrotateby90degreesforassistantsconvenience.2)Fortwo-handstreatment,AssistantHolderBarcanbeextendedby50cm.

    5.ASSISTANTINSTRUMENTHOLDER(Fig.3-10)Pickuptheinstrument(SalivaEjectororHighVolumeEvacuator)fromtheholder,anditstartsworking.

    Bowl Flush Switch

    Cupfiller Switch

    Cupfiller Flow Control Knob

    Bowl Flush Flow Control Knob

    Air Flow Control Knob for 3-Way Syringe

    Water Flow Control Knob for 3-Way Syringe

    Air Vacuum Flow Control Knob

    Increase decrease

    TelescopicAssistantHolder bar

    AssistantHolder

    Fig.3-8MembraneSwitchforAssistant

    Fig.3-9MasterValveBlock

    Fig.3-10CuspidorUnit

  • -22-

    Water Heater Switch

    Bowl Flush Switch

    Service Water OutletFlow Control Knob for Service Water Outlet

    - +Cup Base

    MANUAL SENSOR

    H

    M LH

    Touchless Switch

    6.WATERHEATERSWITCH(Fig.3-11)TurnontheWaterHeaterSwitchwhenthewarmwaterforthepatientcupisrequired.

    7.SERVICEWATEROUTLETWITHFLOWCONTROL(Fig.3-11)WaterOutletwithFlowControlislocatedonthepanelwhichadaptstothequickconnector.

    8.BOWLFLUSHSWITCH(OPTION)(Fig.3-11)TurnontheBowlFlushSwitchandthewaterisflushed to the bowl continuously.

    9.AUTOMATICCUPFILLER&BOWLFLUSH(OPTION)(Fig.3-12)1)Placethepatientcupweighting45to80gonCupBaseandwaterissuppliedfora certain quantity to the patient cup and flushed

    foracertaintimetothebowl.2)QuantityofwaterinpatientcupcanbeadjustedbyturningCupBase.

    (3)LIGHTSECTION(MODELIO5000)1.MODESELETIONSWITCH(Fig.3-13)TurntheswitchtoSENSORside,andON/OFFofthelightcanbemadebyapproachinghandofthesensor.(TouchlessSwitch).

    2.INTENSITYSWITCH(Fig.3-14)ThenintensityoflightcanbeselectedL(Low),M(Medium)orH(High).RefertothemanualofIO5000.

    Fig.3-11CuspidorControlPanel

    Fig.3-12 Automatic Cupfiller

    Fig.3-13ModeSelectionSwitch

    Fig.3-14IntensitySwitch

  • -23-

    Drain Cap

    BasketStrainer

    Solid Collector

    Solid Collector Filter

    (1)DRAIN&SUCTIONLINEOFUNIT(Fig.4-1)1.BasketStrainer.RemovetheBasketStrainerCap,andcleantheBasketStrainer.2.SolidCollectorFilterOpenthesidepanelofthecuspidorbodyandtakeoutSolidCollectorFilterandcleanit.PullingontheSolidCollectorCapwillbringtheSolidCollectorFilterwithit.

    (2)AIRSUPPLYLINE(Fig.4-2)1.AirFilterDrainValve 1) Release the water in the air filter by opening

    DrainValveofJunctionBox.2)Turnoffthemainswitchofthecompressor.Releasecompressedairanddrainaccordingtotheinstructionsofthecompressor. Confirm the air Pressure Gauge indicates

    0Mpa.

    [4] DAILY MAINTENANCE AND CARE

    Drain Valve

    Air Pressure Gauge

    Fig.4-1BasketStrainer&SolidCollector

    Fig.4-2JunctionBox

    Fig.4-3VacuumHndpiece&SalivaEjectorHandpiece

    Vacuum Tip

    Saliva EjectorHandpiece

    VacuumHandpiece

    Vacuum

    Vacuum Cap

    Saliva EjectorHose

    Body

    Body

    Hose

    Slide Knob Slide Knob

    SalivaEjectorCap

    Hose Connector

    Filter

    (3)HANDPIECE1-1.VacuumHandpieceandSalivaEjectorHandpieceForeffectivesterilization,washingforremovingcontaminationandimmersionbyacleaningagentarerequired.Then,rinsebywaterinordertoremoveresidualcleaningagentonmedicaldevice.Takefollowingproceduresfromcleaningtosterilization.*Usethedisposablesalivaejectortip.

    Disassembly (Fig.4-3)Disassemblethehandpieceforthepreparationof cleaning as the following figures show.

    Pullthehoseconnectortodisconnectthevacuumhose.

  • CAUTION

    Sterilization pouch

    Fig.4-7VacuumHandpieceandSalivaEjectorHandpiece

    -24-

    Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat405degrees.(Fig.4-4) Scrub the intubation or hole, slide groove and filter by a

    cleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat405degrees.(Fig.4-5)Wipeoffbyaclothfortheareawhichbrushisunabletoreach.(Fig.4-6)

    B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.

    Sterilization (Fig.4-7)VacuumTip/VacuumCap/VacuumHandpieceBody/SalivaEjectorHandpieceBodycanbeautoclave.Vacuumhandpiecebodyandsalivaejectorbodyhavetoassemblebeforeautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121 or 4 min. at 135.

    Sterilizationwithautoclaveispermittedupto250times.

    Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece forfollowing cases. Any waste material can not be removed by clogged hole. Contamination and solid material attached to handpiece can not be removed.

    StorageAftercleaningthevacuumtip,keepitinthecleanplace.

    Sterilization by class B cycles. Sterilization temperature is 135 degrees or less. The cap, filter and body are made of resin. They may become deteriorated if they are sterilized in an autoclave many times. After autoclave sterilization, the cap, filter, body and valve are subject to discoloration, which does not have a negative effect on performance. The slide knob can be autoclave 100 times and is expendable supplies. Skip the drying process if the temperature is to exceed 135. If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch.

    CAUTION

    Cloth

    Fitting cleaning brushor Tooth brush

    Slide groove

    Cloth

    Fig.4-4Surfacecleaning

    Fig.4-5Cleaningofholeandslidingpart

    Fig.4-6Cleaningthearea(brushisunabletoreach)

  • -25-

    2.VacuumHose/SalivaEjectorHose(Fig.4-9)Itispracticaltoremovethehosesinordertocleanthem.Note:Attheendoftheday,cleanthevacuumandsalivaejectorhandpiecesbysuckingintwocupsworthofwater.

    3.AIRMOTOR/TURBINE/3-WAYSYRINGESterilizethehandpieceaccordingtomanufacturersoperatingmanual.

    VacuumHose

    SalivaEjectorHose

    Fig.4-9VacuumHose&SalivaEjectorHose

    CAUTION

    4.Belmont77Syringe(Fig.4-10)DisassemblyRemovethenozzlefromsyringebyturningitindirectionA.

    Cleaning by hand A.Wipeoffthesurfacecontaminationbyaclothwhilerinsingthesurfacebyrunningcleanwarmwaterat405degrees. Scrubthetipandjointpartofnozzlebyacleaningbrushorbyatoothbrushwithrunningcleanwarmwaterat405degrees.(Fig.4-11)B.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.C.Checkwhethercontaminationisremovedornotaftercleaning.Continuethecleaningifcontaminationisremained.D.Immersedwithanalkalinedetergentfor5minutes.(WerecommendtouseID212madebyDURR)E.Rinsethoroughlybydistilledwateratordinarytemperatureorbycleanwaterformorethan1minute.

    Cleaning must be done within 1 hour after use.Throw out and do not take autoclave handpiece for following cases. Contamination and solid material attached to handpiece can not be removed.

    1-2.Cleaningvacuumandsalivaejectorlines(Fig.4-8)Thesuckingunitcomesintocontactwithsecretions,spitandbloodthatcontainbacteriaeveryday.Besuretocleanandsterilizeitattheendofeachworkday.Recommendedcleaner:OrotolPlusmadebyDURR.

    Donotusestronglyacidiccleaningagentsoralkalinepipecleaningagents,whichmaycausecorrosionofmetals,etc/

    CAUTION Fig.4-8CleaningVacuum/SalivaElectorLines

    Fig.4-11CleaningBelmont77SyringeNozzle

    A

    B

    w A

    Nozzle

    Nut

    Syringe body

    Fig.4-10Belmont77Syringe

    Cleaning brushor Tooth brush

  • (8)FILTERREPLACEMENT(Fig.4-13)The water filter in the junction box needs to bereplacedatleastonceayear.The air filter in the junction box needs to bereplacedatleastonceeverythreeyears.Contactyourlocalservicerepresentativeforreplacement.

    -26-

    Water Filter

    Air Filter

    Fig.4-13Filter

    (6)LUBRICATINGHANDPIECESRefertothemanualofthehandpieces.

    (4)HOSESANDTUBINGCleanhosesandtubingusingDURRFD333.

    (5)CLEANINGUNITAllsurfacescanbecleanedwithDURRFD333cleaner.Spraythecleaner(FD333)onclothandwipethesurfaceswiththecloth.

    CAUTIONDonotdrenchtheunit.Wipeallsurfacesdryaftercleaning.

    StorageAftercleaningthenozzle,keepitinthecleanplace.

    Sterilization by class B cycles. Sterilization temperature is 135 degrees or less. Skip the drying process if the temperature is to exceed 135. If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch. Before use, make sure that the nut is firmly tightened.

    Sterilization (Fig.4-12)Thenozzlecanbesterilizedwithautoclave.A.Insertthehandpieceinasterilizationpouchandsealit. B. Autoclave for 20 min. at 121 or 4 min. at 135.Sterilizationwithautoclaveispermittedupto250times.ApplyVaselinethinlyandevenlytothetwoO-ringsaftersterilization. Tighten the nut firmly in direction B to undo the nozzle.

    CAUTIONFig.4-12SterilizationBelmont77SyringeNozzle

    Sterilization pouch

    (7)MASTERSWITCHAfterthedailyoperationandmaintenance,turnofftheMasterSwitch.

  • [5] MAINTANANCE AND INSPECTION5-1. Guide for daily maintenance and inspection (Maintenance and inspection by user)Managementofmaintenanceandinspectionofmedicalequipmentshouldbeimplementedbytheuser(medicalinstitution).Incasetheuserdoesnotimplementsuchmanagement,itispermitted that such management is outsourced to a qualified entity such as a medical equipment repaircompany. For safe use of this product, it is necessary that inspection should be conducted in the specified frequencyonthetemsdescribedbelow.

    Air turbine revolution, water flow, air flow and so forth shall be free of abnormality.Micromotor revolution, water flow and so forth shall be free of abnormality. Scaler vibration, water flow and so forth shall be free of abnormality.

    No. Item Frequency Inspection method and diagnosis

    Influence if inspection not conducted

    Maintenance required in case of nonconformity

    Check of safety functions

    Before start

    Before start

    Before start

    Before start

    For each patient

    For each patient

    For each patient

    After closing

    Unexpected personal injury and troubles may arise due to motion of the chair during medical treatment and due to pinching between doctor section and chair.

    Contact your dealer or our office if any abnormality arises.

    Check for leakage of water and air

    Leakage of water and air shall not be observed around the product.

    The product will not normally work, and troubles may arise.

    Contact your dealer or our office if any abnormality arises.

    Cupfiller When a paper cup is placed on the cupfiller, the cup shall be detected and cupfilling shall be executed.* Malfunction may arise if the cup is of another material grade (such as stainless steel and plastics) or if the paper cup is of dark color or pattern.

    Cupfilling may not be executed.

    Execute re-inspection in accor-dance with "Method for operation" described in the instruction manual. Contact your dealer or our office if recovery is not achieved as a result of re-inspection.

    Check of motions of equipment

    Troubles such as injury in patient's oral cavity and equipment failure may arise.

    Control the water flow in accor-dance with "Control of compo-nents" described in the instruction manual. If any other abnormality arises, refer to the instruction manual attached to individual equipment. Contact your dealer or our office if recovery is not achieved.

    Check of air turbine bar

    Appropriate bar shall be positively mounted.Make sure to refer to the instruction manual attached to individual equipment.

    The bar will not normally work and troubles may arise.

    If abnormality such as flaw and deformation is found on the bar, replace the bar in accordance with the instruction manual attached to individual equipment.

    Check of scaler tip

    Appropriate tip shall be positively mounted and be correctly used.Make sure to refer to the instruction manual attached to the scaler.

    The tip will not normally work and troubles may arise.

    If the tip was worn or deformed, replace the tip in accordance with the instruction manual attached to the scaler.Contact your dealer or our office if any other trouble arises.

    Check of tightness of syringe nut

    The nut for fixing the nut of Type 77, 3-way syringe shall be positively tightened.

    Troubles may arise if the nut comes off.

    Turn and positively retighten the nut that fixes the nozzle.

    Matters attached to micromotor

    Excessive handpiece oil or the like shall not be attached to the motor section.

    The motor section will not work normally and troubles may arise.

    Execute care in accordance with the instruction manual attached to individual micromotor.

    Make sure the chair movement stops by any of the following actions.when foot controller pedal is depressed. During chair auto movement, depress of any operation switch. While setup is in progress with function switch on the doctor membrane switch panel.When the spittoon bowl is turned to patient side (Pedestal type)

    -27-

  • No. Item Frequency Inspection method and diagnosis

    Influence if inspection not conducted

    Maintenance required in case of nonconformity

    After closing

    CareVacuum and saliva ejector handpiece

    Flush the suction line, and then clean the filter of the vacuum or saliva ejector handpiece.

    Faulty suction may arise.

    Clean the suction line and filter in accordance with "Method for care" described in the instruction manual.

    9

    CareCuspidor section

    Clean the cuspidor and dust filter.

    After closing

    Faulty water drainage may arise.

    Execute dust removal and cleaning in accordance with "Method for care" described in the instruction manual.

    CareSolid collector

    Clean the filter of the solid collector.

    Vacuum suction will become weak.

    Clean the filter in accordance with "Method for care" described in the instruction manual.

    After closing

    CareExterior

    Chemical, filthy water and so forth shall not be found (attached or remaining) on the product exterior.

    Discoloration and deterioration to the exterior, and corrosion and rusting to metallic components may arise.

    Execute wiping in accordance with "Method for care" described in the instruction manual.

    Check of main switch and main valves

    The product main switch shall be off, and water/air main valves shall be closed.

    Product failure and troubles may arise.

    Contact your dealer or our office if the main switch cannot be turned off or if the main valve cannot be closed.

    Product's moving parts

    No abnormal noise or the like shall be produced from product's moving parts when the product is operated.

    The product will not normally work and troubles may arise.

    Contact your dealer or our office if any abnormality arises.

    CareDrain valve

    Drain the water from the air filter drain valve.

    Water may enter the air line, and equipment failure may arise.

    Drain the water from the air filter in accordance with "Method for care" described in the instruction manual.

    Check of water pressure and air pressure

    Check the water pressure and air pressure by reading pressure gauges in the U-Box section.Basic set pressure : Water: 0.1 to 0.2 MPaAir: 0.45 to 0.5 MPa

    The product will not normally work, and troubles may arise.

    Contact your dealer or our office if the set pressure is abnormally high or low.

    Check of conditions of table section

    The table shall be free of inclination, and water shall not flow on the table.

    Injury caused by falling of goods located on the table and other troubles may arise.

    Contact your dealer or our office if any abnormality arises.

    Care Oil mist separator

    The oil level in the oil mist separator shall be lower than the red line.

    Normal output will not be produced due to inferior handpiece exhaust.

    Discharge the oil in accordance with "Method for care" described in the instruction manual.

    After closing

    After closing

    10

    11

    12

    Once every week

    13

    14

    15

    16

    17

    18

    Once every week

    Once every month

    Once every month

    Once every month

    -28-

  • 5-2. Guide for Periodical Check-upSomepartsandcomponentsoftheproductsaredegradedordeteriorateddependingonthe

    frequencyofuse.Annualcheck-upandmaintenance,aswellasreplacementofconsumableparts,arerequired.Therequiredparts(includingconsumableparts)arelistedbelow.Itmaybedifferentfromthe

    followinglistdependingontheoptionoftheunit.Forcheck-upandrepair,callatechnicianofourauthorizeddealer.

    Parts and components that require periodical check-up

    Parts Description

    Vacuum handpiece body

    Saliva ejector handpiece body

    Foot controller

    Water supply hose

    Drain hose

    Air supply hose

    Electric wiring of moving part

    Standard LifetimeNo.

    3 years 1

    2 3 years

    3 5 years

    3 years

    3

    4

    4

    5 3 years

    Consumable parts

    No. Parts Description

    1

    2

    Parts Description Standard LifetimeNo.

    3 years 8

    9 3 years

    10 5 years

    5 years11

    12 3 years

    Regulator

    Valves

    Switches

    Film viewer body part

    Pressure gauge

    Arm section of moving part

    Control PCBs.

    6 3 years 13 7 years

    7 5 years 14 5 years

    Valve for vacum handpiece body

    Vacuum tip

    Handpiece tubings

    Vacuum hose

    Saliva ejector hose

    8

    No. Parts Description

    6

    7

    Filter for oil mist separator

    Filter (Air & Water)

    O-ring, Packing, Diaphragm

    5

    Executethemaintenanceinaccordancewiththisinstractionmanualandoperatingmanualattachedtoeachindividualequipment(Dentallight,Handpiece,etc..).Failuretomaintainthisproductmayleadtophysicalinjuryorpropertydamage.

    WARNING

    -29-

  • [6] ELECTROMAGNETIC COMPATIBILITY(EMC)MedicalelectricalequipmentneedsspecialprecautionsregardingEMCandneedstobeinstalledandputintoserviceaccordingtotheEMCinformationprovidedinthismanual.PortableandmobileRFcommunicationsequipmentcanaffectmedicalelectricalequipment.Theequipmentorsystemshouldnotbeusedadjacenttoorstackedwithotherequipment.Ifadjacentorstackeduseisnecessary, theequipmentorsystemshouldbeobservedtoverifynormaloperationintheconfiguration in which it will be used.

    Guidance and manufactures declaration electromagnetic emissionsThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.

    Emissions test Compliance Electromagnetic environment - guidanceRFemissionsCISPR11 Group1

    TheCLESTAUnitusesRFenergyonlyforitsinternalfunction.Therefore,itsRFemissionsareverylowandarenotlikelytocauseanyinterferenceinnearbyelectronicequipment.

    RFemissionsCISPR11 ClassB

    TheCLESTAUnitissuitableforuseinallestablishments,includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.

    HarmonicemissionsIEC61000-3-2 ClassAVoltage fluctuations/FlickeremissionsIEC61000-3-3

    Complies

    Guidance and manufactures declaration electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.

    Immunity test IEC 60601test level Compliance levelElectromagnetic environment-

    guidanceElectrostaticdischarge(ESD)IEC61000-4-2

    6kVcontact8kVair

    6kVcontact8kVair

    Floorsshouldbewood,concreteorceramic file. If floors are covered withsyntheticmaterial,therelativehumidityshouldbeatleast30%.

    Electricalfasttransient/burstIEC61000-4-4

    2kVforpowersupplylines1kVforinput/outputlines

    2kVforpowersupplylines1kVforinput/outputlines

    Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

    SurgeIEC61000-4-5

    1kVdifferentialmode2kVcommonmode

    1kVdifferentialmode2kVcommonmode

    Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.

    Voltagedips,shortinterruptionsandvoltagevariationsonpowersupplyinputlinesIEC61000-4-11

    95%dipinUT)for0.5cycle40%UT(60%dipinUT)for5cycle70%UT(30%dipinUT)for25cycle95%dipinUT)for5s

    95%dipinUT)for0.5cycle40%UT(60%dipinUT)for5cycle70%UT(30%dipinUT)for25cycle95%dipinUT)for5s

    Mainspowerqualityshouldbethatofatypicalcommercialorhospitalenvironment.IftheuseroftheCLESTAUnitrequirescontinuedoperationduringpowermainsinterruptions,itisrecommendedthattheCLESTAUnitbepoweredfromanuninterruptiblepowersupplyorabattery.

    Powerfrequency(50/60Hz)magnetic fieldIEC61000-4-8

    3A/m 3A/m Power frequency magnetic fields shouldbeatlevelscharacteristicofatypicallocationinatypicalcommercialorhospitalenvironment.

    NOTEUTisthea.c.mainsvoltagepriortoapplicationsofthetestlevel.

    -30-

  • Guidance and manufactures declaration electromagnetic immunityThe CLESTA Unit is intended for use in the electromagnetic environment specified below. The customer or the useroftheCLESTAUnitshouldassurethatitisusedinsuchanenvironment.Immunity test IEC 60601 test level Compliance level Electromagnetic environment -guidance

    PortableandmobileRFcommunicationsequipmentshouldbeusednoclosertoanypartoftheCLESTAUnit,includingcables,thantherecommendedseparationdistancecalculatedfromtheequationapplicationstotheFrequencyofthetransmitter.

    Recommended separation distance

    ConductedRFIEC61000-4-6

    3Vrms150kHzto80MHzoutsideISMbandsa

    3Vrms d = 1.2P

    RadiatedRFIEC61000-4-3

    3V/m80MHzto2.5GHz

    3V/m d = 1.2P 80MHzto800MHzd = 2.3P 800MHzto2.5GHz

    WherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufactureranddistherecommendedseparationdistanceinmetres(m).

    Field strengths from fixed RF transmitters, as determinedbyanelectromagneticsitesurvey,ashouldbelessthanthecompliancelevelineachfrequencyrange.b

    Interferencemayoccurinthevicinityofequipmentmarkedwiththefollowingsymbol:

    NOTE1At80MHzand800MHz,thehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedby adsorption and reflection from structures, objects and people.

    a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbepredictedtheoreticallywith accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the CLESTA Unit isusedexceedstheapplicableRFcompliancelevelabove,theCLESTAUnitshouldbeobservedtoverifynormaloperation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchasreorientingorrelocatingtheCLESTAUnit.

    b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

    Essential performance (purpose of IMMUNITY testing)Unless operated by the switches for chair control, the chair connected to the CLESTA does not make any movements, except for sounding a buzzer and switching on/off the indicator.

    -31-

  • Recommended separation distances betweenPortable and mobile RF communications equipment and the CLESTA Unit

    TheCLESTAUnitisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesarecontrolled.ThecustomerortheuseroftheCLESTAUnitcanhelppreventelectromagneticinterferencebymaintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheCLESTAUnitasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.

    Rated maximum output power of transmitter

    W

    Separation distance according to frequency of transmitterm

    150 kHz to 80 MHzd = 1.2P

    80 MHz to 800 MHzd = 1.2P

    800 MHz to 2.5 GHzd = 2.3P

    0.01 0.12 0.12 0.230.1 0.38 0.38 0.731 1.2 1.2 2.310 3.8 3.8 7.3100 12 12 23

    Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmetres(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthemaximumoutputpowerratingofthetransmitterinwatts(W)accordingtothetransmittermanufacturer.NOTE1At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyadsorption and reflection from structures, objects and people.

    -32-

  • [7] LIST OF COMPATIBLE HANDPIECES

    MODEL

    SYRINGELUZZANI(3-way) Minilight w/Light

    DCI(3-way)

    TURBINE BIEN AIR BORA S36L / UNIFIX with LIGHT

    NSK Ti-Max X

    AIR MOTOR BIEN AIR Aquilon 830 / UNIFIX with LIGHT/PM1132NSK EX-203/EX-6

    SCALER SATELEC SP4055 w/Light

    -33-

  • 2-1-1,Higashishinsaibashi,Chuo-ku,Osaka,JapanTEL : (06) 6213-5945 FAX : (06) 6212-3680

    TAKARA BELMONT CORPORATION

    NOTE

    Takara Belmont (UK) Ltd.Belmont HouseOne St.Andrews Way,Bow,London E3 3PA U.K.Tel: (44)20 7515 0333Fax:(44)20 7987 3596

    EC REP

    Book No.FEFA25H0Printed in JAPAN 2012-04