DART 3+D Scanner Assembly Instructions and User Manual For SVP-3DS Rev 1 Document # 7.5.056 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906

Transcript of DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars...

Page 1: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

DART 3+D Scanner

Assembly Instructions and User Manual

For SVP-3DS

Rev 1

Document # 7.5.056

IonSense Inc.

999 Broadway

Suite 404

Saugus, MA 01906

Page 2: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

2

This manual details the steps necessary to install and use the 3+D Scanner SVP-

3DS for the DART-SVP.

Contents

HARDWARE ...................................................................................3

KEY OF 3+D SCANNER PARTS ................................................................ 3

3+D SCANNER ASSEMBLY ...................................................................... 4

SOFTWARE ................................................................................... 10

METHODS .............................................................................................. 11

3+D Imaging ...................................................................................... 13

Page 3: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

3

Hardware

Key of 3+D Scanner Parts

1. 300mm Linear Rail (X-Rail)

2. 100mm Linear Rail (Y-Rail)

3. X-Rail Cradle (DART-SVP Source Rev 1 ONLY)

4. 3+D Sampling Stage

5. Y-Rail Cradle

6. Allen Wrenches (1.5mm, 2.5mm, 3mm)

1

2

3

5

4

4

6

Page 4: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

4

3+D Scanner Assembly

DART-SVP Source Rev 1: (SN: SVPS00XXX)

Figure 2: Placing the X-Rail into the Cradle

Using the 3mm Allen wrench, remove the three screws that attach the X-cradle to the base of the DART

SVP. Take the X-Rail and slide the non-motor end (end opposite the cable) into the cradle, making sure

that the front of the stage faces the screw holes.

Reference the Hardware and Network Installation Manual v2 manual to determine the exact placement of

the rail in the cradle as this procedure depends on the version and revision of the rail.

Using the 1.5mm Allen wrench, lock the rail into position by tightening the two set screws on the front of

the cradle. Once the rail is placed back onto the DART these set screws will be inaccessible.

Figure 3: Tighten the set screws to lock the X-Rail into position.

Front of Stage Facing away

from MS/Towards Cradle Screws

Page 5: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

5

DART-SVP Source Rev 2: (SN: SVPS002XX)

The revision 2 DART-SVP source, pictured below, has an X-Rail Cradle but it is not removable from the

source baseplate. Simply place the linear rail directly into the cradle on the baseplate, and lock into place

using the two set screws.

As with the Rev 1 source, reference the Hardware and Network Installation Manual v2 to determine the

exact placement of the rail in the cradle.

Y-Rail Installation:

Place the X-Y mounting block on top of the X-Rail stage. Using four 2.5mm screws and the

2.5mm Allen wrench loosely secure the block onto the X-Rail stage, aligning the Y-Rail so that

the front piece of the Y-rail’s stage faces the DART-SVP cradle. Tighten the two 1.5mm set

screws on the side of the X-Y mounting block facing the DART-SVP for good alignment and

then finish tightening the four 2.5mm screws on top of the X-Y mounting block to finish the

attachment.

Page 6: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

6

Figure 4: Showing the alignment of the X- and Y-Rails before they are connected.

Figure 5: Secure the Y-Rail to the X-Rail using a 2.5mm Allen wrench.

To attach the Y-Rail to the X-Y mounting block slide the non-motor end of the rail into the

cradle, leaving it so the end of the rail is flush with the end of the mounting block. Using the

1.5mm Allen wrench, tighten the two set screws to lock the rail in place.

Page 7: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

7

Figure 6a: The Y-Rail with the X-Z mounting block; b: The Y-Rail in the X-Z mounting block;

c: Locking the Y-Rail in place.

Pull the DART SVP source housing all of the way back from the mass spectrometer inlet and

remove the ceramic tube from the VAPUR interface. Supporting the rail assembly carefully,

place the cradle back into the space between the MS interface and DART-SVP source, with the

three screws on the DART-SVP side of the set-up.

Figure 7: Securing the rail assembly to the DART-SVP base with the 3mm Allen wrench.

Page 8: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

8

Figure 8: The rail assembly attached to the DART-SVP source.

DART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART

source. Raise the DART source to its maximum height so that the 3+D sample tray can fit underneath it.

Loosen the knobs on the side of the DART and turn it downward three clicks so that it is now pointed

downward at a 45 degree angle. Tighten the knobs to lock it back into place.

Figure 9a and b: Rev 1 Source at maximum height and 45 degree angle

Page 9: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

9

DART-SVP Source Rev 2: Attach the SVP-45A angle module to the DART source housing. Use the

spinning knob to adjust to the proper height.

Figure 10: Rev 2 Source with SVP-45A angle module

Place the mounting block of the 3+D stage on the Y-Rail block and loosely secure the top of the mounting

block using four 2.5mm hex screws. Using a 1.5mm Allen wrench, tighten the two set screws on the back

side of the mounting block for proper alignment and then finish by completely tightening the four 2.5mm

hex screws on the top of the block.

Figure 11: Securing the 3+D stage to the Y-Rail.

Insert the longer ceramic tube into the VAPUR inlet, taking care to leave a 2mm gap between the end of

the ceramic tube and the API inlet. Refer to the specific Vapur Flange Manual for the mass spectrometer

that the DART-SVP is being installed on for more details.

Page 10: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

10

Software

Check the “About” page on the Settings tab of the SVP software to see the version of software currently

loaded. For this manual v3.x.x or higher should be uploaded.

The most up to date version of software can be downloaded from the IonSense website at this

address: http://www.ionsense.com/software

If the software needs to be updated then download the file from the IonSense website and

reference the DART Software Manual v3 for instructions on how to perform the update. A

printed version of this software manual comes with each DART system, or can be accessed here:

http://www.ionsense.com/manuals

Web Software: Page 19

iPod Software: Page 20

Page 11: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

11

Methods

Once the software is updated select the Methods tab at the bottom of the software screen. There

will be one method available for the 3+D Scanner called 3+D Imaging to add into the current list

of methods. To add in a method, click on the “Add” button in the upper right corner of the

Methods page and select the desired method.

NOTE: The methods may need to be added to the main list; for further details reference page 8

of the Software Manual v3.

3+D Imaging

This is the standard/generic method for the 3+D scanner. It performs a basic back and

forth scan of a 2D surface.

Before running the method, be sure the Linear Rail settings are correct. If they aren’t the

methods won’t run properly. To check, click on the Settings tab in the software; open both

Linear Rail #1 (X) and Linear Rail #2 (Y). A left handed setup should have left selected for both

Motor Location and Sample Direction and vice versa for a right handed setup. Below are the

correct settings for a left handed rail setup.

Page 12: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

12

Once the method is selected either hit Start to begin the method, or select Edit to change the

various parameters. The 3+D Imaging method parameters are described in more detail on the

following page.

Page 13: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

13

3+D Imaging

The following parameters can be adjusted:

Sample Count

Number of rows.

Ion Mode

Ions can be positive or negative.

Temperature

The temperature of the heater.

X Start Position

X location where the sample starts.

X Sample Speed

How fast the X rail moves over the sample.

X Length

Length of the sample.

Y Start Position

Y location where the sample starts.

Y Sample Speed

How fast the Y rail moves over the sample.

Y Spacing

Distance between rows.

Shutdown State

The system can be kept in run or standby mode for

repeated use or automatically shut down.

Standby Temperature

If the system is kept in run or standby mode, this

temperature will be maintained.

Page 14: DART 3+D Scanner - IonSenseDART-SVP Source Rev 1: Loosen the two knobs on the back of the pillars holding the blue DART source. Raise the DART source to its maximum height so that

14

Revision History

REV DCR # Description of Change Effective Date

1.0 43 Initial Release 11/11/2013