d n a s c i t s i r e t c a r eNod4-T ETHERNET a...

60
NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa – 294, Rue Georges Charpak – 74100 JUVIGNY - FRANCE Tél. : +33 (0)4 50 87 78 64 – www.scaime.com eNod4-T ETHERNET Digital Process Transmitter Characteristics and functioning

Transcript of d n a s c i t s i r e t c a r eNod4-T ETHERNET a...

Page 1: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa – 294, Rue Georges Charpak – 74100 JUVIGNY - FRANCE

Tél. : +33 (0)4 50 87 78 64 – www.scaime.com

eNod4-T ETHERNET Digital Process Transmitter

Cha

ract

eris

tics

and

func

tioni

ng

Page 2: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

2 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

1 SAFETY INSTRUCTIONS ............................................................................................ 4 1.1 Getting started ................................................................................................ 4 1.2 Protection ........................................................................................................ 5 1.3 Power supply .................................................................................................. 5 1.4 Grounding ....................................................................................................... 5 1.5 Installation ....................................................................................................... 5 1.6 Maintenance .................................................................................................. 5 1.7 Caution ............................................................................................................ 5

2 ENOD4 PRODUCT RANGE ...................................................................................... 6 2.1 General presentation ..................................................................................... 6 2.2 Versions ........................................................................................................... 6

2.2.1 Communication protocol versions ........................................................ 6 2.2.2 IO+ version ............................................................................................... 6 2.2.3 Bluetooth® version ................................................................................... 6

2.3 eNodView Software ....................................................................................... 7 2.4 eNodApp - Mobile Application .................................................................... 7

3 GENERAL CHARACTERISTICS ................................................................................. 8 3.1 Dimensions ...................................................................................................... 8 3.2 Electrical characteristics ............................................................................. 10 3.3 Legal for trade .............................................................................................. 11

4 CONNECTIONS ..................................................................................................... 13 4.1 Cabling basic rules ...................................................................................... 14

4.1.1 Prevent Electrostatic Discharges ......................................................... 14 4.1.2 Prevent Electromagnetic disturbances ............................................... 15

4.2 Power supply connection ............................................................................ 16 4.3 Load-cell wiring ............................................................................................ 16 4.4 IO+ version .................................................................................................... 16 4.5 Inputs / outputs connections ...................................................................... 18

4.5.1 Typical connections .............................................................................. 18 4.6 JB4 option ...................................................................................................... 19

4.6.1 Specifications ........................................................................................ 20 4.6.2 Excitation trimming procedure ............................................................ 20 4.6.3 Spring terminal connections ................................................................ 20

4.7 220V option ................................................................................................... 21 5 COMMUNICATION ............................................................................................... 22

5.1 Communication Interface connections ..................................................... 22 5.1.1 Process control communication .......................................................... 22 5.1.2 PC communication ............................................................................... 22 5.1.3 AUX Communication (for HMI) ............................................................ 22

5.2 Communication address selection ............................................................ 22 5.3 Communication rate selection ................................................................... 23 5.4 Communication protocols .......................................................................... 23 5.5 Simultaneous functioning of communications .......................................... 24 5.6 Bluetooth® Communication (Optional) ...................................................... 25

6 CALIBRATION AND SCALE ADJUSTMENT ............................................................ 26 6.1 Physical calibration ...................................................................................... 26 6.2 Theoretical calibration ................................................................................. 26 6.3 Scale adjustment coefficient ...................................................................... 26 6.4 Gravity correction ........................................................................................ 26

Page 3: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

3 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

6.5 Scale interval ................................................................................................ 26 7 FILTERS ................................................................................................................... 27

7.1 Filtering related to the A/D conversion rate .............................................. 27 7.2 Bessel low pass filter ..................................................................................... 27 7.3 Notch filter ..................................................................................................... 27 7.4 Self-adaptive filter ........................................................................................ 27

8 MEASUREMENT AND STATUS ................................................................................ 28 8.1 Gross measurement ..................................................................................... 28 8.2 Net measurement ........................................................................................ 28 8.3 Tare value ..................................................................................................... 28 8.4 Factory calibrated points ............................................................................ 28 8.5 Logical IN/OUT level .................................................................................... 28 8.6 Preset Tare value .......................................................................................... 28 8.7 Measurement status ..................................................................................... 28 8.8 Weighing diagnosis ...................................................................................... 28

8.8.1 Global weighing diagnosis .................................................................. 28 8.8.2 Sensor input control .............................................................................. 28

9 LOGICAL INPUTS ................................................................................................... 29 9.1 Inputs assignment ........................................................................................ 29 9.2 Inputs function description .......................................................................... 29

9.2.1 None ....................................................................................................... 29 9.2.2 Tare ......................................................................................................... 29 9.2.3 Cancel Tare ........................................................................................... 29 9.2.4 Zero ......................................................................................................... 29 9.2.5 Measure transmission (SCMBus/fast SCMBus only) ........................... 29 9.2.6 Continuous measure transmission (SCMBus/fast SCMBus only) ....... 30 9.2.7 Sensor input control .............................................................................. 30

10 LOGICAL OUTPUTS .............................................................................................. 31 10.1 Outputs assignment ................................................................................... 31 10.2 Description .................................................................................................. 31

10.2.1 None ..................................................................................................... 31 10.2.2 Set point ............................................................................................... 31 10.2.3 Motion: ................................................................................................. 31 10.2.4 Defective measurement: .................................................................... 31 10.2.5 Input image: ........................................................................................ 32 10.2.6 Level on request: ................................................................................. 32

10.3 Analog output (IO+ version) ...................................................................... 32 11 ENODTOUCH – USER INTERFACE (OPTIONAL) ................................................... 33

11.1 Online functioning ...................................................................................... 33 11.1.1 Startup screens .................................................................................... 33 11.1.2 Multi-eNod screens ............................................................................. 34 11.1.3 Main screen ......................................................................................... 38 11.1.4 Main setup menu ................................................................................ 41

Page 4: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

4 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

1 SAFETY INSTRUCTIONS Important information Be sure to read this manual thoroughly before starting the PRODUCT. This manual is part of the PRODUCT and comes with it. Rules for the prevention of risks and accidents as well as safety rules must be observed! The PRODUCT is manufactured in accordance with the current state of the art and in accordance with safety regulations laws of the Labour Code. The general safety requirements must be observed during all phases of operation and repair of equipment. Non-compliance of written warnings in this manual is contrary to the requirements of safe operation and normal operation of the PRODUCT. As a user of this product, it is necessary to strictly follow in your work environment, all warnings and requirements necessary to perform safely any operation on the PRODUCT.

This symbol means that there is a risk of electrisation that can lead to death or serious injuries instructions are not followed.

This symbol means that there is a risk of danger that can lead to injuries if safety instructions are not followed.

Earth protection connection

WEEE 2002/96/CE Waste of electrical and electronic equipment directive imposes to final users the management of electrical and electronic equipment (EEE). For the purpose, it is mandatory for them to selectively collect their EEE depositing them in dedicated places.

DANGER

Danger: dangerous situation that can lead immediately to death or serious injuries.

WARNING

Warning: situation potentially dangerous that could lead to death.

CAUTIOUS

Cautious: situation potentially dangerous that could lead to injuries.

1.1 Getting started

WARNING

Security provided by this product is provided for use for its intended purpose. Maintenance can only be performed by qualified staff.

Page 5: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

5 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

1.2 Protection

WARNING

The PRODUCT, with 220V option, must be linked to a protection device. This one must comply to the standards in force in the country (NFC 15-100 in France) and must be part of the building in which the PRODUCT has been installed. The circuit breaker must be 16 Amps and should have a response curve of type C.

WARNING

The mains plug of the PRODUCT serves as disconnecting device. Consequently, it must be connected near the device and easily accessible.

1.3 Power supply

WARNING

The PRODUCT, with 24VDC version, must be powered by a power supply in accordance to EN 60950-1. It must be located close to the PRODUCT.

1.4 Grounding

DANGER

To minimize any risk of electric shock, frame and all parts of the equipment must be connected to Earth.

1.5 Installation The PRODUCT must be fitted on a support by using the mounting brackets at the rear of the box.

1.6 Maintenance The PRDUCT can be cleaned using a damp slightly cloth.

DANGER

For your safety, before any technical intervention or cleaning, it is imperative that the equipment is turned off and not plugged in. It is forbidden to replace components with power cable connected.

1.7 Caution When installing, moving, or handling this product or peripherals connected to it, refer to the instructions below to connect and disconnect the various cords.

Connection: Disconnection: 1. Turn off the units. 2. Start by connecting all cords to the units. 3. Connect the interface cables to the connectors. 4. Connect the power cords into outlet. 5. Turn on the units.

1. Turn off the units. 2. Disconnect the power cords from the outlets. 3. Disconnect the interface cables from the

connectors. 4. Disconnect all cabled from the units.

Page 6: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

6 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

2 ENOD4 PRODUCT RANGE

2.1 General presentation eNod4 is a high speed digital process transmitter with programmable functions and powerful signal processing capabilities. eNod4 offers operating modes for advanced process control both static and dynamic. Quick and accurate:

Analog to digital conversion rate up to 1920 meas/s with maximum scaled resolution of ±500 000 points. Digital filtering and measurement scaling. Measurement transmission up to 1 000 meas/s.

Easy to integrate into automated system: USB, RS485 and CAN communication interfaces supporting ModBus RTU, CANopen® and PROFIBUS-DPV1

(depending on version) communication protocols. Digital Inputs/Outputs for process control. Setting of node number by rotary switches and communication baud rate by dip switches. Integrated selectable network termination resistors. Wiring by plug-in terminal blocs.

2.2 Versions

2.2.1 Communication protocol versions Strain gauges load-cell conditioner with CANopen® and ModBus RTU communication. Strain gauges load-cell conditioner with Profibus DP-V1 and ModBus RTU communication. Strain gauges load-cell conditioner with Modbus TCP and ModBus RTU communication. Strain gauges load-cell conditioner with EtherNet/IP and ModBus RTU communication. Strain gauges load-cell conditioner with Profinet IO and ModBus RTU communication. Strain gauges load-cell conditioner with EtherCAT and ModBus RTU communication.

EDS, GSD, ESI and GSDML configuration files for above protocols can be downloaded from our web site: http://www.scaime.com

2.2.2 IO+ version In conjunction with all communication protocol versions, eNod4 can supports an opto-insulated board fitted with:

2 additional digital inputs and 1 speed sensor dedicated input. 0-5V or 0-10V analog output voltage. 4-20mA, 0-24mA, 0-20mA or 4-20mA with alarm at 3.6mA analog output current.

2.2.3 Bluetooth® version In conjunction with all communication protocol versions and IO+ version, eNod4 can supports a communication board that allows to exchange information via Bluetooth®.

Radio information

Wireless communication Bluetooth® Low Energy 4.2

Max. RF output power -16.9 dBm

Frequency range 2.4 – 2.4835 GHz

Page 7: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

7 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

2.3 eNodView Software So as to configure eNod4, SCAIME provides eNodView software tool. eNodView is the software dedicated to eNod devices and digital load cell configuration from a PC. This simple graphical interface allows accessing the whole functionalities of eNod4 for a complete setting according to the application. eNodView features and functions:

eNod4 control from a PC Calibration system Modification/record of all parameters Measure acquisition with graphical display Numerical filters simulation Frequential analysis FFT Process control Network parameter

eNodView software is available in English and French version and can be downloaded from our web site: http://www.scaime.com or ordered to our sales department on a CD-ROM support.

2.4 eNodApp - Mobile Application So as to configure eNod4, SCAIME provides eNodApp application through the Google Play Store. eNodApp is the application dedicated to eNod devices (fitted with the Bluetooth® option) configuration from an Android device. This simple graphical interface allows accessing most of the functionalities of eNod4 for a quick setting according to the application. eNodApp features and functions:

eNod4 control from an Android device Calibration system Modification/record of all parameters

eNodApp application is available in English and French version and can be downloaded from the Google Play Store.

Page 8: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

8 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

3 GENERAL CHARACTERISTICS

3.1 Dimensions

Page 9: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

9 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Page 10: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

10 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

3.2 Electrical characteristics Power supply Unit

Supply voltage 10.....28 VDC 110.....240 VAC

Max supply power 4.4 W

Additional max supply power (IO+ version) 3 W

Temperature range

Storage temperature range -25...+85 °C

Working temperature range -10...+40 °C

Load cell sensor Minimum input resistance > 43 Ω

sensor connection 4 or 6 wires

Bridge excitation voltage 5 VDC

Communication RS 485 Half-duplex

Rate 9 600...115 200 bauds

Ethernet 2 x RJ45 10/100 Mbits/s

Logical inputs

Number 2 (+2 with IO+ version)

Type opto-insulated type 3

Low level voltage-current 0 / 5 VDC – 0 /1.5 mA

High level voltage-current 11 / 30 VDC – 2 / 9 mA 7 mA @ 24VDC

Logical outputs Number 4

Type solid state relay

Max. current @ 40°C 0,4 A

Max. voltage in open state 53 VDC or 37 VAC

Max resistor in close state 2 Ω

Metrological specifications on A3 connector input (load-cell type sensor) Input sensor range for a load cell sensor 7.8 mV/V

Thermal zero drift typical 1.5 ppm/°C

Thermal span drift typical 2 ppm/°C

Linearity deviation 0.003 % FS

Conversion rate 6.25 ... 1920 meas./s

Page 11: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

11 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Metrological specifications on analog output (IO+ version) Output voltage range 0-5 or 0-10 V

Output current range 4-20, 0-24 ou 0-20 mA

Max. load on current output 500 Ohm

Outputs resolution 16 bit

Max. linearity error 1 LSB

Total error +/- 0.07 %FSR

Thermal zero drift typical +/- 2 ppm/°C

Thermal span drift typical +/- 3 ppm/°C

Conversion rate A/D converter rate value Hz

Speed sensor power-supply (IO+ version)

Bridge excitation voltage (V+ …V-) 12.5 +/- 2 V

Bridge excitation current 30 mA

Isolation 1000 V

Speed sensor input (IO+ version)

IN5 HTL 0…2.5 / 5…30 VDC

IN5 TTL 0…0.5 / 2.4…5 VDC

Isolation 1000 V

3.3 Legal for trade

Characteristics Unit

Accuracy class III or IIII

Maximum number of weighing ranges 1

Maximum number of verification scale divisions 6000

Load cell(s) excitation voltage 5 V

Minimum voltage division per verification scale division (ΔUmin)

0.5 µV

Minimum voltage of the weighing range 0.5 * n * 10-3 mV

Maximum voltage of the weighing range 39 mV

Value of factor pi 0.5

Minimum impedance for the load cell 47 Ohm

Type of load cell(s) connection system 4-wire or 6-wire

Maximum impedance for the load cell 1500 Ohm

Maximal length/section measurement cable 166 m/mm2

eNod4* is an analog data processing unit evaluated as a part of a non-automatic weighing instrument (NAWI) with applicative software -T or an automatic weighing instrument (AWI) like an automatic gravimetric filling instrument with

Page 12: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

12 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

applicative software -D or a catchweigher with applicative software -C. This instrument is not intended for direct sales to the public. It is suitable for conditioning OIML R60 certified strain gauges load cell(s) with analog output. Legal for trade mode has to be activated internally in order to respect metrological requirements. All the functionalities will be conformed to the essential requirements for certified weighing instruments. Sealing will be material on the eNod4 housing and load cell connections. If a junction box is used it will have to be sealed too. eNod4 is fitted with a software sealing for metrological parameters and once activated, is composed of an event counter and a CRC value of specific and adjustment parameters. These values shall be marked on the terminal device connected after the last official verification when putting in service the instrument. Any mismatch between the displayed values on the terminal and those marked will signify a broken sealing. eNod4 is fitted with a data storage device (DSD) so called alibi memory. Any weighing result (or checkweigher result with applicative software -C, or dosing result with applicative software -D) is stored internally and can be recalled on demand. 130816 records can be stored permanently at maximum. Any record is identified by a unique 32-bit long number. This identifier is incremented each time a weighing result is stored and transmitted. The minimum time between two DSD recording operations is 50ms. * legal for trade use does not apply to software applicative for beltweigher (-B) and weigh feeder (-F)

Page 13: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

13 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4 CONNECTIONS

Repère Mark

Fonction Function

Repère Mark

Fonction Function

A1 alimentation power supply

1 +VDC

JP1 Câblage capteur 6 fils / 4 fils 6-wire / 4-wire loadcell wiring 2 GND

A3 connexion

capteur load cell

connection

1 Exc+ JP2 Résistance de terminaison connexion AUX

AUX connection termination resistor 2 Sens+

3 Exc- A8 connexion AUX AUX connection

1 RB/TB (B-) 4 Sens- 2 RA/TA (A+) 5 Sig+ 3 GND

Page 14: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

14 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.1 Cabling basic rules

4.1.1 Prevent Electrostatic Discharges

For ESD protection, cabling for the analog load cell must include a ferrite and a silicone protection sleeve, not included with the transmitter.

Install the ferrite by passing the analog cell connection through the ferrite and wrapping it around once, as shown on the following picture:

6 Sig-

CN2 Connexion IO+ IO+ connection

1 V+ 7 Shield 2 IN5-HTL

A4 Port USB 3 IN5-TTL

A5 entrées/sorties

IN / OUT

1 IN1+ 4 V- 2 IN1- 5 GNDA 3 IN2+ 6 IN4+ 4 IN2- 7 IN3+ 5 OUTCOM 8 Iout 6 OUT1 9 GNDA

7 OUT2 10 Vout

8 OUT3 SW1 Sélecteur Adresse haute (hex)

High Address selector (hex)Low 9 OUT4

SW2 Sélecteur Adresse basse (hex) Low Address selector (hex) A7

2 x RJ45 Ethernet

Ethernet ETH1 RJ45 (sortie) Ethernet ETH1 RJ45 (output)

SW3 Sélecteur Baud rate AUX/USB Aux/USB Baud rate selector

Ethernet ETH2 RJ45 (entrée) Ethernet ETH2 RJ45 (input)

SW4 bouton poussoir Reset reset push button NS / MS

LED NS: Network Status (Bus Fault Profinet) LED MS: Module Status (System Fault Profinet)

S1-S2-S3-S4 LED sorties logiques outputs LED PWR-USB LED alimentation & activité USB

power supply & USB activity LED

Page 15: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

15 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.1.2 Prevent Electromagnetic disturbances

Observe these basic rules to guard against Electromagnetic disturbances.

Rule 1: Large area grounding contact

When installing the devices, make sure that the surfaces of inactive metal parts are properly bonded to chassis ground (see following sections).

Bond all inactive metal parts to chassis ground, ensuring large area and low-impedance contact (large cross-sections).

When using screw connections on varnished or anodized metal parts, support contact with special contact washers or remove the protective insulating finish on the points of contact.

Wherever possible, avoid the use of aluminium parts for ground bonding. Aluminium oxidizes very easily and is therefore less suitable for ground bonding.

Provide a central connection between chassis ground and the ground/protective conductor system.

Rule 2: Proper cable routing

Organize your wiring system into cable groups (high-voltage/power supply/signal/measurement/data cables).

Always route high-voltage and data cables in separate ducts or in separate bundles.

Install the measurement cables as close as possible to grounded surfaces (e.g. supporting beans, metal rails, steel cabinet walls).

Rule 3: Fixing the cable shielding

Ensure proper fixation of the cable shielding.

Always use shielded data cables. Always connect both ends of the data cable shielding to ground on a large area.

Keep unshielded cable ends as short as possible.

Always use metal/metalized connector housings only for shielded data cables.

Shields of load cells must be connected to the pin ‘Shield’ of the load cell connector of the eNod4 or, in case of junction box, directly on the body of this one.

Rule 4: Special EMC measures

For cabinet or enclosure lighting in the immediate range of your controller, use incandescent lamps or interference suppressed fluorescent lamps.

Rule 5: Homogeneous reference potential

Create a homogeneous reference potential and ground all electrical equipment.

Use sufficiently dimensioned equipotential bonding conductors if potential differences exist or are expected between your system components.

Equipotential bonding is absolutely mandatory for applications in hazardous areas.

Page 16: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

16 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.2 Power supply connection

On the front panel a green light ‘PWR’, (D7) indicates if power is connected.

4.3 Load-cell wiring

4 wires load-cell: jumpers in place (by default at delivery). 6 wires load-cell: jumpers removed

4.4 IO+ version An analog board in current and voltage might be used with eNod4 to provide IO+ version. This has to be requested when ordering eNod4 product. The analog output is both current and voltage galvanically isolated at 1000V. Voltage output might be set either 0-5V or 0-10V, and the current output to 4-20mA, 0-24mA, 0-20mA or 4-20mA alarm 3.6mA. It is software setting and both output (current and voltage) might separately be enable.

The IO+ version is fitted with two additional inputs IN3 and IN4:

1

2

+VDC GND

A1

1 2 3 4 5 6 7

Exc+ Sens+ Exc- Sens-

Sig+ Sig- Shield

4/ 6 wire jumper

ON: 4 wires OFF: 6 wires

IOUT

N

Current output connection

CN2

Iout

GNDA

Vout

8

9

10 VOUT

N

Voltage output connection

CN2

Iout

GNDA

Vout

8

9

10

A3

Page 17: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

17 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

The IO+ version is fitted with a pulse input and a dedicated power supply for a speed sensor (belt weigh feeder, belt weigher). Two input voltage levels are proposed for the pulse input of the speed sensor: TTL logical level or high voltage 30 V maxi level.

Inputs connection

CN2

V+

IN5-TTL

IN5-HTL

V-

GNDA

IN4+

IN3+

Iout

GNDA

Vout

1

2

3

4

5

6

7

8

9

10

V supply

Speed sensor power supply

CN2

V+

IN5-TTL

IN5-HTL

V-

GNDA

IN4+

IN3+

Iout

GNDA

Vout

1

2

3

4

5

6

7

8

9

10

Speed sensor input connection

CN2

V+

IN5-TTL

IN5-HTL

V-

GNDA

IN4+

IN3+

Iout

GNDA

Vout

1

2

3

4

5

6

7

8

9

10

Page 18: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

18 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.5 Inputs / outputs connections

An indicator light in front panel is assigned to each Output.

4.5.1 Typical connections

Inputs : Connection to a detector Inputs : Connection to a push button

Outputs : Possible connections

D1 D2 D3 D4

PNP

+

-

S IN+

IN-

NPN

+

-

S

IN+

IN-

+

-

IN+

IN-

+

-

IN+

IN-

+ -

Load

OUT

OUTCOM

+

-

Load

OUT

OUTCOM

Load

OUT

OUTCOM

Page 19: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

19 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.6 JB4 option

Optional overvoltage protection

Corner balancing potentiometer

Additional resistor for special use* (not supplied)

* cut strip conductor before insertion

Page 20: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

20 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.6.1 Specifications Version Standard Version ATEX / IECEx

Cable diameter (output) Ø 6 … Ø 11.5 mm Ø 6 … Ø 10 mm

Cable diameter (inputs) Ø 2.5 … Ø 7 mm Ø 4 … Ø 8 mm

Wire cross section 0.14 mm2 … 1 mm2

Cable stripping length 7 …10 mm

4.6.2 Excitation trimming procedure

For load cells with matched outputs, shortcut the potentiometers by soldering bridges J1, J2, J3, J4. First adjustment with reference weight:

1. Turn the potentiometers fully clockwise for each load cell so that the potentiometer resistance is as low as possible.

2. Lay your reference weight at one corner of your weighing system and note the result of the weighing indicator. Repeat this operation for each corner.

3. Spot the corner with the lowest output. Load successively each corner with the same reference weight and adjust at each step the potentiometer to match the lowest output corner value.

4. The load cells outputs are now matched. Then proceed to normal zero and gain adjustment of the weighing indicator.

4.6.3 Spring terminal connections

Use a screwdriver with a maximum blade width of 3 mm. 1. Press the black button with the screwdriver. 2. Stick the wire end into the relevant opening. 3. Release the pressure on the screwdriver. 4. Pull slightly the cable to check it is held tight.

7-10mm

Page 21: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

21 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

4.7 220V option

Plug the supplied power cord into the AC outlet.

Page 22: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

22 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

5 COMMUNICATION

5.1 Communication Interface connections

5.1.1 Process control communication Version Type of communication Connector

eNod4 Modbus TCP Modbus TCP A7

eNod4 EtherNet/IP EtherNet/IP A7

eNod4 Profinet IO Profinet IO A7

eNod4 EtherCAT EtherCAT A7

5.1.2 PC communication eNod4 models can communicate with a PC using the protocols Modbus RTU or SCMBus through the USB connector accessible from the front panel.

USB Communication stops AUX communication when used.

The appropriate USB driver can be downloaded from our website: www.scaime.com. It is also available on CD to order from our sales department. Note: If eNodView software has been correctly installed, it is not mandatory to re-install the USB drivers when connecting another eNod4 on the same USB port (Windows only asks for the driver if the device is connected to another USB port).

5.1.3 AUX Communication (for HMI) GND (connector pin3) is connected with power supply GND. The common mode voltage admitted is 27VDC from GND power supply. When eNod4 is at the end of a bus line the 150 Ω integrated resistor can be used (connecting jumper). Warning: Do not add termination line jumper on both side if distances between 2 devices is short.

5.2 Communication address selection Rotary switches selection (SW1 and SW2) accessible from the front panel. The new address only is taken into account after a reset. Example: eNod4 address = 3AH = 58d

!

Page 23: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

23 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

5.3 Communication rate selection Dipswitch selection (SW3) is accessible from the front panel. The new baud rate only is taken into account after a reset.

Dipswitch Baud rate RS485 et USB

Bit rate CAN 1 2 3 4

ON ON ON X 9600 50 kbit/s OFF ON ON X 19200 50 kbit/s ON OFF ON X 38400 50 kbit/s OFF OFF ON X 57600 125 kbit/s ON ON OFF X 115200 250 kbit/s OFF ON OFF X 9600 500 kbit/s ON OFF OFF X 9600 1 Mbit/s

OFF OFF OFF X 9600 125 kbit/s

5.4 Communication protocols

Model Communication interface Protocols* Connector Front panel LED

eNod4 Ethernet

RS485 AUX ModBus RTU SCMBus A8 /

USB ModBus RTU SCMBus

Front panel USB A4

Led USB USB Led

Ethernet

Modbus TCP EtherNet/IP EtherCAT

A7

MS état du module MS Module Status NS état du réseau NS Network Status

Profinet

MS défaut système (vert) / défaut bus (rouge) MS system failure (green) / bus failure (red) NS état du réseau NS Network Status

Page 24: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

24 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

5.5 Simultaneous functioning of communications DIN version

Box version

Simultaneous communication Ethernet RS485 AUX

USB yes* No

Ethernet Yes**

(*) Simultaneous use of Ethernet with USB port can reduce performance of this interface. (**) In this configuration, we recommend low speed rate on AUX output.

PC Connection

Page 25: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

25 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

5.6 Bluetooth® Communication (Optional)

Information related to the radio emission are the followings:

Information Radio

Wireless communication Bluetooth® Low Energy 4.2

Max. RF output power -16.9 dBm

Frequencies range 2.4 – 2.4835 GHz

Activation or deactivation of Bluetooth® fonctionnality is done by dipswitch #4 on SW3.

Dipswitch Bluetooth®

1 2 3 4

X X X ON Active

X X X OFF Inactive

A blue light on front side indicates the status of the Bluetooth® communication.

Blue light Bluetooth® COM Status

Off Inactive

Fast blinking Initialization

Slow blinking Ready for pairing with mobile device

Steady On Active and paired

The communication through Bluetooth® allows, using the dedicated mobile app eNodApp, to visualize the different weighing values, the Inputs / Output states and the applicative cycle states from an Android device. Moreover, it allows to perform some actions such as:

- Perform a Zero or Tare, - Calibrate a sensor, - Parametrize Inputs/Outputs, - Set filtering parameters, - Read/Clear DSD records, - Set application parameters, - And more…

The theoretical range of the Bluetooth® communication is about 10 meters without obstacles. eNodApp application is available in English and French version and can be downloaded from the Google Play Store.

Page 26: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

26 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

6 CALIBRATION AND SCALE ADJUSTMENT eNod4 is factory calibrated to deliver 500 000 counts for 2mV/V with a load cell on the A3 input connector. Initial calibration can be modified for a better adjustment to the usage or because of characteristics of the sensor. To achieve these various types of adjustments the following options and procedures are available:

physical calibration theoretical calibration scale adjustment coefficient gravity correction

6.1 Physical calibration Physical calibration is done by applying to the sensor from 1 up to 3 known references.

6.2 Theoretical calibration The theoretical calibration allows defining eNod4 user span without using calibration reference. The information needed to achieve the procedure is the sensor sensitivity and its rated capacity. For example, a 15kg load cell with sensitivity equal to 1.870 mV/V at 15kg; put sensor maximum capacity 15 000 and sensor sensitivity 1,870.

6.3 Scale adjustment coefficient Initial calibration value can be modified with a scale adjustment coefficient. This coefficient has maximum and minimum values.

6.4 Gravity correction When eNod4 is used to condition a weighing sensor, it can be necessary to adjust measurement if the place of measurement is different from the place where eNod4 was calibrated. eNod4 automatically adapts its span by storing into its non-volatile memory these 2 parameters: ‘Calibration place g value’ and ‘Place of use g value’. Initial values for these coefficients are identical; they correspond to the g value of a calibration place located in ANNEMASSE FRANCE.

6.5 Scale interval The scale interval is the difference between 2 consecutives indications. Possible values are: 1, 2, 5, 10, 20, 50, and 100. Modification of scale interval is taking into account after a new calibration.

When using eNod4 for legal for trade purpose, it is imperatively required to activate the legal for trade switch

BEFORE any calibration procedure (cf § legal for trade switch). !

Page 27: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

27 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

7 FILTERS There are four available filtering levels which can be associated:

filtering related to the A/D conversion rate including rejection of the mains frequency (50 or 60 Hz) harmonics.

low-pass Bessel filter notch filter self-adaptive filter

7.1 Filtering related to the A/D conversion rate The signal resolution is related to the conversion rate. The conversion rate might be chosen as low as possible, particularly for static applications. For dynamic applications, a compromise must be found between the measurement rate and the low-pass filter cut-off frequency. The eNodView software can be used to determine appropriate filter values. Choose a measurement rate that rejects the mains frequency harmonics according to the place of use, 50 or 60Hz.

7.2 Bessel low pass filter A low-pass digital filter can be applied as an output of the A/D converter. The filter orders (available values are 2, 3 or 4) and cut-off frequency are adjustable. eNodView software can be used to determine appropriate filter values.

7.3 Notch filter A notch filter might be applied as an output of the low-pass filter (if used) or the A/D converter. It allows attenuating the frequencies within a band defined by high and low cut-off frequencies. The eNodView software can be used to determine appropriate filter values.

7.4 Self-adaptive filter This filter can be set in cascade after previous filters. It is particularly efficient for static measurements but avoid using it in dynamic or dosing processes. The aim of this filter is to eliminate erratic measurements and to average consistent measurements.

Page 28: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

28 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

8 MEASUREMENT AND STATUS The eNod4 transmits measure after signal and data processing through different protocols available. The accessible variables are:

8.1 Gross measurement The ‘gross measurement’ stands for the digital value after measurement scaling. It is affected by all the ‘zero’ functions (power-up zero, zero tracking and zero requests). Zero request value can be saved in non-volatile memory (option).

8.2 Net measurement The ‘net measurement’ stands for the digital value after measurement scaling and tare subtraction.

8.3 Tare value The ‘tare value’ stores the calibrated value that is subtracted from the ‘gross measurement’ so as to give the ‘net measurement’. This value can be saved in non-volatile memory (option).

8.4 Factory calibrated points The ‘factory calibrated points” contains the measurement value without the user calibration layer. It is directly linked to the analog input voltage.

8.5 Logical IN/OUT level The ‘logical IN/OUT level’ allows reading any time eNod4 logical inputs and outputs level.

8.6 Preset Tare value A previous calculated tare can be restored using this variable. This value can be saved in non-volatile memory (option).

8.7 Measurement status The measurement status contains information on eNod4 measurement parameters.

8.8 Weighing diagnosis

8.8.1 Global weighing diagnosis An internal alarm flag reflects the integrity of the whole measurement chain. It’s used to set logical output active or optional analog output in an error mode in order to warn about any defection on the measurement chain (defective measurement). This variable is set active when at least one of the followings conditions occurs:

- sensor input control result out of tolerances - sensor input control command in progress - sensor input control command failed (timeout) - sensor input reference command in progress - gross meas. < (- max capacity) - gross meas. > (max capacity) - analog signal out of the A/D converter input range - EEPROM failure

This internal alarm flag is featured with adjustable specific de-bounced time and minimal activation time.

8.8.2 Sensor input control eNod4 features a weighing diagnosis system allowing to check the integrity of analog sensor input by electrically simulating a load, resulting to a simulated weight value. This diagnostic system can be used together with the others defects detection systems in order to achieve overall integrity check of the measurement chain. This system involves two phases initiated by the user:

- The first, just after user calibration, allows taking a simulated reference weight value when the measuring chain integrity is OK.

- The second, when the user wants to check the integrity of the system, allows to make the difference between a new simulated weight value and the reference. Then this difference can be compared with a dedicated maximum tolerance value.

Page 29: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

29 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

9 LOGICAL INPUTS Each input can work individually in either positive or negative logic. A holding time (de-bounced time) attached to all inputs can be configured.

9.1 Inputs assignment Inputs can be assigned individually to one of the following functions:

none tare zero send measurement measurement window cancel tare sensor input control

Function Functioning mode

transmitter

None

Tare

Cancel tare

Zero

Measure transmission (SCMBus/Fast SCMBus only)

Continuous measure transmission (SCMBus/Fast SCMBus only)

Sensor input control

9.2 Inputs function description

9.2.1 None Inputs have no effect.

9.2.2 Tare Each input can be assigned to the tare function. Depending on the chosen logic (positive or negative) for the corresponding input, tare is triggered by a falling or rising edge.

9.2.3 Cancel Tare Depending on the chosen logic (positive or negative), the tare is cancelled by a falling or rising edge of the corresponding input.

9.2.4 Zero Each input can be assigned to the zero function. Depending on the chosen logic (positive or negative) for the corresponding input, zero acquisition is triggered by a falling or rising edge. A new zero is acquired only if its value is within a range 10% of maximum capacity or 2% if eNod4 is in legal for trade functioning mode. This new value is the useful zero value, a reset cancels it.

9.2.5 Measure transmission (SCMBus/fast SCMBus only) The request can apply to:

gross measurement net measurement factory calibrated points

A single measurement is transmitted per rising or falling edge (depending on the configured logic) on the input signal.

Page 30: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

30 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

9.2.6 Continuous measure transmission (SCMBus/fast SCMBus only)

The request can apply to: gross measurement net measurement factory calibrated points

While the input is kept at the right level, a series of measurements are transmitted at the period defined by the ‘sampling period’ setting. If the 2 inputs are assigned to this function, only input 2 is operating.

9.2.7 Sensor input control The assignment of logical input to sensor input control function allows performing special procedure to diagnose load cell sensor input. Beforehand, user must acquire reference value of the load cell input by sending 'Sensor input control reference' specific command (e.g. after the device is calibrated). Note: Load cell sensor input control must be realized if no process cycle is in progress.

Page 31: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

31 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

10 LOGICAL OUTPUTS Each output can be individually set to positive or negative logic.

10.1 Outputs assignment Outputs can be assigned individually to one of the following functions:

Function Functioning mode

Transmitter

None

Set point

Motion

Defective measurement

Input image

Level on request

10.2 Description

10.2.1 None The output has no function.

10.2.2 Set point The outputs can be assigned to configurable set points. Output 1 is assigned to set point 1, output 2 to set point 2 and so on. Set points can be assigned to gross measurement or net measurement and are defined by a high value, a low value and a functioning mode. Two functioning modes are possible:

Hysteresis Window

10.2.3 Motion:

The outputs can be assigned to measurements stability criteria.

10.2.4 Defective measurement: The output level is set when the internal alarm flag described in “Weighing diagnosis” § in the MEASUREMENT AND STATUS § is activated. This allows to warn about defection of the measurement chain.

Page 32: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

32 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

10.2.5 Input image: Outputs can be assigned to copying inputs state, either using the same logic or inverting the input state (negative logic). Outputs 1 and 3 can be assigned to input 1 and outputs 2 and 4 to input 2.

10.2.6 Level on request: Activation of outputs is triggered by master requests.

10.3 Analog output (IO+ version) An optional analog board in current and voltage might be used with eNod4 to provide IO+ version. This must be asked when ordering eNod4 product. Voltage output might be set either 0-5V or 0-10V, and the current output to 4-20mA, 0-24mA, 0-20mA or 4-20mA with alarm at 3.6mA. Both output (current and voltage) might separately be enable. Settings are effective after eNod4 reset. Analog output affectation function is common to both current and voltage output and might be assigned to followings:

None: analog outputs have no function. Gross measurement: analog outputs can be assigned to gross measurement copy. Maximal level value is

related to Maximum Capacity parameter and works in mono-quadrant functioning. Bi-quadrant option can only be applied to gross measurement copy. When this option is activated, the lowest value of current and voltage levels corresponds to -MC and the highest value to +MC.

Net measurement: analog outputs can be assigned to net measurement copy. Maximal value is related to Maximum Capacity parameter and works in mono-quadrant functioning only. The highest value of current and voltage levels corresponds to +MC in only one quadrant.

Level on request: analog outputs are driven by master requests through the external value to control analog output variable (in 0.01% of the full scale of current or voltage analog outputs).

When analog output is assigned to “Gross measurement” or “Net measurement” its value jumps to a special error value when the internal alarm flag described in “Weighing diagnosis” § in the MEASUREMENT AND STATUS § is activated. This allows to warn about defection of the measurement chain.

Page 33: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

33 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11 ENODTOUCH – USER INTERFACE (OPTIONAL)

11.1 Online functioning

11.1.1 Startup screens At first Power-On, eNodTouch is set to manage one eNod4 at the address 1, the connection screen is displayed

while eNodTouch try to communicate with connected eNod4 device. Then a redirection screen appears during 3 seconds for a redirection to a main screen if the communication is

established between eNodTouch and eNod4 communication or an error screen if eNodTouch does not manage to connect to the eNod.

If there is no eNod4 connected or if the display unit cannot communicate with eNod4 device, a communication error screen appears. In this case, check the wiring between eNodTouch and eNod4 (§ Electrical Powering) and check that the communication parameters of eNodTouch and eNod4 are correct (§Peripheral – Device PLC Settings).

Redirection screen

Connection screen Communication error screen

Page 34: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

34 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

When communication is established the main screen of eNod4 appears:

11.1.2 Multi-eNod screens

11.1.2.1 Increase eNod number To increase the eNod number and have an access to the multi-eNod screen, please follow the procedure:

- Touch the Fn button of the main screen, the main menu appears. - Enter in the eNodTouch menu. - Go to the Starting Parameters screen by touching the right arrow button.

Main Menu

eNodTouch menu Starting parameters

Page 35: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

35 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

By default the eNod number is 1, eNodTouch starts on the main screen of the first eNod. Increase eNod number if needed. The display module can start on the main screen of any eNod or on the multi-

eNod screen by disabling « Starting eNod » function. If starting eNod is disconnected or have communication problems, the multi-eNod screen appears. When settings are finished, touch the left arrow button, to restart eNod Touch.

According to the number of eNod settled, eNodTouch redirects to the following multi-eNod screens:

Channel name Weight value

Unit

Gross/Net Indicator

Multi eNod (2 eNod) Multi eNod (3 eNod)

Multi eNod (4 eNod) Multi eNod (5 eNod)

Page 36: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

36 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.2.2 Select an eNod

To display the main screen of an eNod, touch its display area (Example for eNod n°1):

To return to multi eNod screen, touch the channel red button:

eNod4_1

eNod4_3

eNod4_5

eNod4_2

eNod4_4

eNod4_6

Multi eNod (6 eNod)

Touch its display area Main screen of eNod

Touch of channel display Multi-eNod screen

Page 37: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

37 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.2.3 Communication error and disconnection When an eNod4 has communication problems or is disconnected, the data from other channels freezes for 5

seconds. During this time, a reconnection button appears, eNodTouch tries to recover the connection with eNod4.

After 5 seconds if eNodTouch can not recover the connection with eNod4, the button enters in the state « eNod

disconnected », the data display resumes.

Identify the communication problem in the following list: wiring problem (§1.4), communication parameters

misconfigured on eNodTouch or eNod4 (§2.2.3). When you found out the issue, touch the « eNod disconnected » button to re-establish the communication. The display module tries to recover the communication with eNod4 during 1 second, the button toggles to the

« Reconnection » state. After 1 second if data from the eNod4 appears, the communication is re-established, else the button toggles to « eNod disconnected » state.

Page 38: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

38 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.3 Main screen

11.1.3.1 Transmitter Mode (eNod4-T)

To switch from the main keyboard to the 2nd keyboard touch

To switch from the main screen to the multi-eNod screen touch

Bargraph

Gross/Net indicator

Stability indicator

Main keyboard

Inputs / Outputs status

Zero in ¼ of scale interval indicator

Weight display area

Channel selection button

Gross weight

2nd keyboard Keyboard masked

Page 39: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

39 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.3.2 Keyboard functions

eNod4 model Key Action

Main Keyboard and 2nd Keyboard

All

ZERO function: possible only under stability condition, according to stability criterion NOTE: The ZERO function allows to reset the gross measurement. It is only available if the measured value is included in +/- 10% of the full scale.

TARE function: possible only under stability condition, according to stability criterion NOTE: The TARE value is the value subtracted from the gross measurement to get the net measurement.

PRESET TARE function NOTE: A previous calculated TARE can be restored using this variable.

CANCEL TARE function NOTE: Cancel TARE allows to switch back to the gross measurement display.

Shift to 2nd keyboard

Enter to Setup menu

Access to thresholds adjustment screen (in Transmitter mode)

Enter in the Quick Access Menu that give access to sensor input diagnosis and allows to set applicative target weights for eNod4-C, D, F and B. Quick access can be configured to manage what is accessible in this menu for eNod4-D and F.

Weighing result acquisition (DSD) only in transmitter functioning mode.

Page 40: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

40 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

eNod4-C

DYNAMIC ZERO function: Zero calculation (measurement averaging) without stability condition NOTE: Dynamic ZERO function has the same result as the ZERO function except that the stability condition is not existing anymore (only measurement averaging).

eNod4-D and F

SUSPEND DOSING: Allows suspending the ongoing dosing batch cycle. D: only possible when the “Suspend/Resume cycle allowed”. F: only possible when batch mode is activated. To resume the cycle touch, START CYCLE.

MANUAL EMPTYING: D: Visible only in filling mode with Manual Emptying mode. Allows to activate the emptying logical output during the emptying phase. F: Start the extraction until the empty hopper level is reached.

eNod4-C, D, F and B

START CYCLE: C: Start Checkweigher cycle. D: Start the dosing cycle if starting conditions are satisfied. F: Start feeder cycle and the extraction. B: Start Belt dosing cycle. Invisible if process cycle is ongoing.

eNod4-D, F and B

STOP CYCLE: Stop the dosing cycle. The dosing counter is not incremented for eNod4-D.

eNod4-F

REFILLING: Allows starting manual refilling. A touch when refilling is ongoing stops the refilling. Visible only if hopper’s weight is under high refilling level. Any way refilling will stop automatically if hopper’s weight reaches high refilling level.

eNod4-B

DYNAMIC ZERO function: In belt mode and when the system is running, after receiving a “dynamic zero” command eNod4 calculates the average of integrated weight per length during the belt revolution time.

eNod4-B, F

STOP and RESTART PID at each cycle step.

eNod4-F and B

RESET TOTALIZATION: Reset the totalization value.

11.1.3.3 Diagnostic (available depending on version)

The Quick Access diagnostic screen can be called directly from the main screen with the touch

The Diagnostic screen can be called using the key

Page 41: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

41 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

NOTES:

- Sensor input reference must be performed before using the sensor input control functionality, see Sensor Input control Parameters menu in the Configuration Setup chapter.

- When you press the HR key, the High-Resolution net weight expressed in tenth of the user weight unit is displayed during 5 seconds. This may be used for weighing system metrological qualification.

11.1.4 Main setup menu eNodTouch setup menu allows to display and modify all the eNod4 parameters. For more information about

parameters, consult eNod4 software manuals. Setup menu access can be secured by password (see § 3.5.1). In that case, the following screen is displayed:

Logical inputs

Logical output

Net weight

Sensor input control result

Run the sensor input control command

Display of the High-Resolution net weight

(on legal for trade certified version)

Page 42: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

42 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

After entering a valid password (or OLEG if you’ve forgot your password), the following menu will be displayed:

eNod4-B: Dosing modes

eNod4-F : Dosing Modes eNod4-D : Dosing Modes

eNod4-C : Checkweigher Mode eNod4-T : Transmitter Mode

Page 43: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

43 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.4.1 eNodTouch setup

Parameter Possible Value Description

Language [Fr], [En] Change the language of eNodTouch software

Keyboard mask [Yes], [No] Enable/Disable keyboard display on main screen

Password [Yes], [No] Enable/Disable access protection to setup menu

4 Char By default, “OLEG” is a master password always valid

11.1.4.2 Starting parameters

Parameter Possible value Description

eNod number 1-6 Number of eNod4 supported

Starting eNod

[Yes], [No] Enable/Disable the function to start on main screen of a specific eNod

1 – eNod Number Number of starting eNod4.

If there is only one eNod4 supported this function is necessarily activate.

Page 44: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

44 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.4.3 Calibration setup

Parameter Possible Value Description Calibration Parameters

When using eNod4 for legal for trade purpose, it is imperatively required to activate the legal for trade switch

BEFORE any calibration procedure (cf § legal for trade switch). !

Page 45: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

45 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Range 0 – 10 000 000 Maximum weighing capacity

Division 1, 2, 5, 10, 20, 50 Difference between two consecutive indicated values

Decimals 0-7 Position of decimal point

Weight unit 4 visible char Weight measurement unit (g, kg, t, lb…) using for calibration

Time unit s, mn, h Time unit using for calibration (eNod4-F and B). Only s and h can be selected in eNod4-B.

Weight unit (display flow rate) g, kg, t Weight unit of flow rate which is displayed on main screen.

Time unit (display flow rate) s, mn, h Time unit of flow rate which is displayed on main screen.

Cut-off frequency (display flow rate) Cut-off frequency of flow rate which is displayed on main

screen.

Physical calibration Allows creating relation between the weighing sensor and the physical values.

Number of loads 1, 2, 3 Number of calibration loads used during the physical calibration.

Load 1/2/3 1 – 1 000 000 Weight values corresponding to each calibration segment.

Page 46: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

46 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

The steps of this physical calibration operation are the followings :

1. Define the number of points to realize the calibration and the load values for each,

2. Press ,

3. Emptying the scale and press ,

4. Wait for the end of data acquisition,

5. Place the Load 1 and press ,

6. Wait for the end of data acquisition,

7. Repeat steps 5 and 6 (if several points have been defined).

Remark : If you press ,

the operation will stop and the values will not be saved.

The latest valid calibration values will be restored.

Zero Calibration Allows acquiring stable measurement with no load on the scale to set the calibration Zero

The steps of this zero calibration operation are the followings :

1. Emptying the scale and press ,

2. Wait for the end of data acquisition.

Remark : If you press ,

the operation will stop and the values will not be saved.

The latest valid calibration values will be restored.

Page 47: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

47 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Single slope adjustment Allows acquiring stable measurement with a load on the scale to set the first segment. Performing physical calibration in this way do not allows multiple segments calibration. Use this calibration method only if you cannot acquire the segment calibration immediately after the Zero calibration (e.g.: big hopper). Zero calibration must have been done before.

The steps of this single slope adjustment operation are the followings :

1. Define the number of points to realize the calibration and the load values for each,

2. Press ,

3. Place the Load 1 and press ,

4. Wait for the end of data acquisition,

5. Repeat steps 3 and 4 (if several points have been defined).

Remark : If you press ,

the operation will stop and the values will not be saved.

The latest valid calibration values will be restored.

Theoretical calibration An automatic scaling to migrate from the factory calibration to the user calibration.

Page 48: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

48 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

S (mV/V) 0 – 6.00000 Load cell sensitivity in mV/V at maximum capacity

Span Calibration correction

Span. cor 0.9 – 1.100000 Allows to adjust initial calibration

Calibration g >0 (9.805470) Allows compensating the gravity difference between calibration place and using place. Use g >0 (9.805470)

Page 49: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

49 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.4.4 I/O setup

(Available depending on version) (IO+ version only)

Page 50: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

50 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Parameter Possible Value Description Input 1-4 Parameters

Function Function assigned to the related input: None, Zero, Tare, Clear, start cycle, stop cycle, Suspend cycle, Dynamic Zero, Manual emptying, Belt fault

Logic [Positive]-[Negative] defines the edge (or level) that triggers input function

Debounced time 0 – 6553 ms Minimum required time of the logical inputs before their activation

Output 1-4 Parameters

eNod4-F external totalizer (Page 2)

Page 51: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

51 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Function

Function assigned to the related output: None, Thresholds, Stability, Coarse feed, High feed, fine feed, Cycle ongoing, result available, Out of tolerance, Dosing fault, Input copy, Remote control, Out of tolerance +, Out of tolerance –

Logic [Positive], [Negative] Defines idle output state.

Mode [Windows], [Hysteresis] Thresholds functioning principle

Comparison [Gross], [Net], [result] Comparison value for thresholds control

Weight / pulse external totalizer 0 – 65535

For external totalization purpose, eNod4 sends a pulse on logical output when the totalization value reaches multiple of this parameter. When an overflow is occurred on pulses output an alarm is set.

Thresholds control Allows modifying the Threshold values of the concerned outputs.

Remote Control

Allows fixing the level [On], [Off], of the concerned outputs.

Page 52: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

52 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Outputs 1-4 timings

In Checkweigher, allows modifying timings on outputs with functions « Remote control » (activation time), « Result within tolerances » and « Result out of tolerances » (activation time and delay).

Parameters Analog output (IO+ version only)

Analog output remote (Page 2) with “remote control” function or “Flow Rate Ctrl” function

Page 53: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

53 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Fonction The allocation of the analog output is unique for both outputs (current and voltage)

Polarity Bipolar / Unipolar

Analog outputs can be assigned to gross measurement copy. Maximal level value is related to maximum capacity (MC) parameter and works in mono-quadrant functioning. Bipolar option can only be applied to gross measurement copy. When this option is activated, the lowest value of current and voltage levels corresponds to -MC and the highest value to +MC.

Voltage Output Voltage output might be set either 0-5V or 0-10V.

Current Output Current output to 4-20 mA, 0-24 mA, 0-20 mA or 4-20 mA alarm 3.6 mA.

Defective measurement outputs parameters (Available depending on version)

Changes Defective measurement outputs timing.

An internal alarm flag reflects the integrity of the whole measurement chain. It’s used to set logical output active or optional analog output in an error mode in order to warn about any defection on the measurement chain (defective measurement). This variable is set active when at least one of the followings conditions occurs:

- sensor input control results out of tolerances

- sensor input control command in progress - sensor input control command failed

(timeout) - sensor input reference command in

progress - gross meas. < (- max capacity) - gross meas. > (max capacity) - analog signal out of the A/D converter

input range - EEPROM failure

This internal alarm flag is featured with adjustable specific de-bounced time and minimal activation time:

Defective measurement debounced time

0 – 65535 ms The internal alarm flag which reflects the integrity of the whole measurement chain is set active only after error conditions have always been true during this de-bounced time.

Defective measurement alarm activation time

0 – 65535 ms

The internal alarm flag which reflects the integrity of the whole measurement chain remains active for this minimal “defective measurement alarm activation time” when it come to be active and whatever the error conditions are during activation.

Page 54: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

54 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

11.1.4.5 Configuration setup

Parameter Possible Value Description Metrology Parameters

Page 55: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

55 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Stability criterion None, 0.25 0.5, 1, 2d Defines a stability interval, related to scale interval.

Zero at Startup [Yes], [No] Enable the Automatic Zeroing at power-on

Zero Tracking [Yes], [No] Enable the Zero tracking function

Legal for trade Switch

(Available depending on version)

[Yes], [No]

Enable the legal for trade mode.

The prerequisite conditions for the activation of legal for trade mode are:

- The weight unit must be one of the following: mg, g, kg, t, ct, µg, oz. (possibility to put spaces after unit)

- The stability criterion must be 0.25 d. - Filter settling time must be less than 1 second in

transmitter mode. - The quotient (Maximum capacity / Scale interval) must

be less or equal to 6000. - On eNod4-C, zero tracking device must not be activated

together with dynamic zero tracking device. On eNod4-T, C and D this mode has an effect on user calibration. Thus, this mode must be activated before processing any calibration.

Save Tare in FRAM [Yes], [No] It allows saving tare in non-volatile memory (FRAM).

Save Zero in FRAM [Yes], [No] It allows saving zero in non-volatile memory (FRAM).

Audit trail (Available depending on version)

Page 56: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

56 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Legal for trade sealing [Yes], [No]

Once legal for trade sealing activated, it forbids the change of any parameters that can affect the eNod4 metrological specifications. Such parameters appear in blue and are disabled over the eNodTouch menus.

Counter (audit trail) This number is incremented on each legal for trade sealing switching.

Checksum Checksum calculated over set of sealed parameters when legal for trade sealing is being activated.

eNod program version Program version of eNod4 embedded software.

eNod metrological version Version of set of metrological functions embedded in eNod4.

eNodTouch version Program version of eNodTouch embedded software.

Data Storage Device records consultation access.

Data Storage Device (Available depending on version)

Record ID to read Record of the ID to be read.

Erase all records from the DSD memory. This is possible only

when the legal for trade sealing is disabled.

Page 57: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

57 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Ethernet Parameters (Ethernet versions only)

IP IP address

Mask Subnet mask

Gateway Default gateway (must match with IP address and Mask)

DHCP Assignment of IP address from a DHCP server on this subnet

Swap data Swap data (bytes) in Modbus TCP registers (16 bits words)

Name Profinet station name (up to 10 chars with Option « Use rotary switch in Name » else 15, lower case, figures, dots and dashes only)

Option « Use rotary switch in Name »

Allows on Save automatic appending of hexadecimal representation of rotary switches in the Name.

Byte order Byte order in Profinet exchanges ([Little Endian], [Big Endian])

Ethernet/IP Modbus TCP

Profinet

Page 58: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

58 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Digital Filtering parameters

A/D converter parameters

Rejection [50Hz]-[60Hz] The A/D converter have In-built adjustable low-pass filter with frequency rejection of 50 Hz or 60 Hz harmonics.

Frequency 6 – 1600 Hz Filtered measurement frequency is adjustable from 6.25 to 1600 Hz

Bessel low-pass filter parameters

Filter Order 0, 2, 3, 4 For the suppression of periodic vibrations, a Bessel low-pass digital filter can be applied as an output of the A/D converter.

Minimum cut-off frequency value depends on the A/D converter rate:

0.10 Hz for 6.25 meas/s, 4 Hz for 400 meas/s

Cut-Off Frequency 0.1 to 200Hz

Notch Filter parameters

High & Low Frequency * 0.1 to 200Hz Frequency band to attenuate with the digital Notch filter

* May not be implemented depending on eNod4 applicative version.

Page 59: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

59 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Reset, factory setting Allows resetting eNod4 or making eNod4 come back to factory settings

Channel name Name or rename the channel with a maximum of 4 characters

Parameter Possible Value Description Sensor Input Control Parameters (Available depending on version)

Page 60: d n a s c i t s i r e t c a r eNod4-T ETHERNET a helitontarti.com.tr/dokumanlar/nu-enod4t-eth-e-0319_216702-f.pdf · NU-eNod4T-ETH-E-0418_216702-F.docx SCAIME SAS – Technosite Altéa

60 SCAIME Manual : NU-eNod4T-ETH-E-0418_216702-F.docx

Reference

Reference value expressed in factory calibrated points for the sensor(s) input control test. The value is automatically determined and stored after executing the sensor input reference command.

Result

Result of sensor(s) input control test expressed in 1/10 of user weight unit. Its value is automatically determined and stored after executing the sensor input control command. This test result represents the weight difference between the reference value and the current test value.

Result max. tolerance 0.0 – 6553.5

The Sensor input control result variable is compared with the Sensor input control result max. Tolerance parameter which is expressed in 1/10 of user weight unit and has a default value of 30. If the sensor input control result value is greater than or equal to “Sensor input control result max. Tolerance” then an error is reported

Sensor input control command will cause eNod4 to handle special test on sensor input and to deliver a test result. This command must not be realized when any process cycle that use weight is in progress.

Sensor input reference command will cause eNod4 to handle special sequence to acquire sensor input control reference value of the load cell sensor input. This command must be performed before using the Sensor input control command and each time the weight measurement chain (sensor, connection, calibration) is modified. This command must not be realized when any process cycle that use weight is in progress.