Non-Interpretive SkillsMRI Safety
![Page 2: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/2.jpg)
Before You Begin
This module is intended primarily for clinical medical students or interns intending to learn or review non-interpretive radiology skills.
Please note that while not integral, this module series assumes some familiarity with basic imaging techniques and interpretive skills. If you wish to learn or review these concepts, please see our “Interpretive Skills” module series.
If material is repeated from another module, it will be outlined as this text is so that you are aware
![Page 3: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/3.jpg)
Objectives
By the end of this module, you will be able to:
• State the four main categories of risk for patients undergoing MRI
• Discuss strategies for decreasing the risks of MRI
• Name the person(s) who is responsible for MRI safety
![Page 4: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/4.jpg)
Introduction
• Safety issues in medical imaging are often discussed “in general,” with
• For MRI safety in particular, this method doesn’t work nearly as well as actual examples…
![Page 5: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/5.jpg)
Introduction
• MRI offers an amazing level of diagnostic ability and involves no ionizing radiation
• For these and many other reasons, MRI is often viewed as “better” or “safer” than CT
• But, the mechanism for achieving MRI’s benefits creates several other safety issues
![Page 6: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/6.jpg)
Format
• We’ll go through four common cases that we may encounter in inpatient and outpatient MRI patients - first as unknown cases, and then with answers and descriptions
• Please note: this module does NOT address all elements of MRI safety, but focuses on the four that you are most likely to see day to day
![Page 7: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/7.jpg)
![Page 8: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/8.jpg)
Case 1
• A patient is being rushed to the MRI suite for imaging of a suspected cerebrovascular accident (CVA). He was intubated for respiratory distress and arrives on a ventilator. The ICU team wants bring him directly into the scanner room. Is there anything you would like to check beforehand?
![Page 9: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/9.jpg)
Case 1
![Page 10: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/10.jpg)
Case 2
• A patient arrives for an MRI for evaluation of “possible focal nodular hyperplasia vs. adenoma” of the liver. They have no information in the electronic record, only a CT from an outside institution without a report.
• Should this patient under go MRI?
![Page 11: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/11.jpg)
Coronal CT – IV + Oral Contrast
Case 2
![Page 12: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/12.jpg)
Case 3
• A patient arrives for an MRI for evaluation of headaches. She is a veteran, and has served two tours in Afghanistan. Her only imaging is a sinus-protocol CT done earlier that day (no report yet).
• Should this patient undergo MRI?
![Page 13: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/13.jpg)
Case 3
![Page 14: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/14.jpg)
Case 4
• Two patients arrive for their MRI evaluations (one for a shoulder rotator-cuff protocol, and the other for a brachial plexus protocol).
• Should these patients undergo MRI?
![Page 15: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/15.jpg)
Case 4
![Page 16: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/16.jpg)
Or, from the point-of-view of our patients…
![Page 17: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/17.jpg)
![Page 18: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/18.jpg)
• How safe does this feel now? ☺
• FYI, another patient issue we won’t talk about
is claustrophobia – which is easy to minimize
until you’ve been in the bore of a scanner
• See let’s get in there …
![Page 19: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/19.jpg)
THE “BIG MAGNET”
• You’ve heard about the “big” MRI magnet – but how big is “big?” First, some units.
• 1 Tesla (T) = 10,000 Gauss
• Earth’s magnetic field = 0.5 Gauss Okay, so...?
• 1.5 T = 30,000x Earth’s magnetic field
• 3 T = 60,000x Earth’s magnetic field
The “lower strength” in practice
Most MRIs in practice
![Page 20: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/20.jpg)
THE “BIG MAGNET”
• We put our patients – knowingly – into a magnetic field that is 60,000x stronger than Earth’s – WHY?
• We’ll skip most of the physics – but we actually needthat strength to be able to send and receive enough signal from the hydrogen nuclei that we “see”
• Thus, our patient needs to go into that magnetic field
• What does NOT need to go into that magnetic field..?
![Page 21: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/21.jpg)
Leaving these “outside” may seem
obvious
FERRO-MAGNETIC = NOT OKAY
Placed in 1.5T or 3T these “everyday objects” become
flying projectiles
![Page 22: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/22.jpg)
The results of taking a non-MRI compatible
O2 tank into the scanner can be DISASTROUS…
But these may not be obvious – especially
when you are dealing with an incredibly sick patient that needs a scan “RIGHT NOW”
Case 1
![Page 23: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/23.jpg)
![Page 24: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/24.jpg)
First category of MRI risk:
Projectile
![Page 25: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/25.jpg)
Coronal CT – IV + Oral Contrast
Anything problematic
here?
Case 2
![Page 26: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/26.jpg)
Coronal CT – IV + Oral Contrast
What are THESE?
(and what study can
show them more
clearly?)
Case 2
![Page 27: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/27.jpg)
Case 2
![Page 28: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/28.jpg)
• We are good at remembering the “big magnet”
• But we are NOT good at remembering that the magnet only gets the hydrogen nuclei organized so that the radiofrequency (RF) pulses can interact with them
• Those RF pulses involve a HUGE amount of electricity – and PACEMAKERS and other implanted electronic devices can malfunction due to the induction of current!
• What “other implanted devices?”
Case 2
![Page 29: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/29.jpg)
For example: ????????
![Page 30: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/30.jpg)
• In recent years, companies have begun making MRI-safe pacemakers. BUT…
• If no product info is available, or the pacemaker was placed long ago – assume it is UN-safe to go into the MRI!
COCHLEAR IMPLANT
DEEP BRAIN STIMULATOR
Case 2
![Page 31: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/31.jpg)
COCHLEAR IMPLANT
DEEP BRAIN STIMULATOR
Second category of MRI risk:
Induction of Current
Case 2
![Page 32: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/32.jpg)
Anything problematic
here?
Case 3
![Page 33: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/33.jpg)
On further questioning-
Case 3
The patient had a “blast injury” where she had a facial contusion and mild corneal
injury but no chronic symptoms…
… from this metallic debris in her orbit!
![Page 34: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/34.jpg)
Case 3
• So the big magnet – the big “STATIC” magnet – organizes the hydrogens
• The RF pulses create the signal
• But how do we LOCALIZE that signal in 3D space? Smaller “gradient” magnets that CHANGE during the scan
![Page 35: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/35.jpg)
Case 3
The torque on a ferromagnetic object caused by the STATIC and CHANGING
magnetic fields create TORSIONAL force
Those objects can rotate or translate, injuring adjacent structures – the most
vulnerable area for this being the ORBIT!
![Page 36: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/36.jpg)
ANY possible metal exposure to eyes / face? Get:
Pre-MR Screening Orbit Radiograph
Another example - A “possible” cerebral aneurysm clip/ unspecified brain surgery?
Head CT
Third category of MRI risk:
Torsional Force
Case 3
![Page 37: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/37.jpg)
Recap…
3. Induction of Current
1. Projectile
2. Torsional Force What’s left?
![Page 38: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/38.jpg)
Case 4
So what exactly is wrong with a little body art?
Nothing – unless you are in an MRI with a big static magnet, smaller changing/gradient magnets, and a
huge amount of electricity…
![Page 39: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/39.jpg)
Question – what makes black tattoo ink black?
Case 4
![Page 40: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/40.jpg)
Question – what makes black tattoo ink black?
Ferrous Oxide
= Millions of tiny magnets …
Case 4
![Page 41: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/41.jpg)
Resistive heating occurs in these magnets, proportional to the strength of the
magnet and time of the scan
Burns have been reported (usually in unconscious or sedated patients – like patients
being sedated due to claustrophobia ! )
FYI – the only way we avoid resistive heating with the “big magnet” – to achieve 1.5T or 3T field strength – is to cool it to the point of being a “superconductor” free of resistance
– with a whole ton of pressurized Helium and Nitrogen
Case 4
![Page 42: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/42.jpg)
• Tattoos are not an absolute contraindication
• But, the risk increases if they are at the level of interest for imaging (= where the greatest amount of energy is directed)
• What to do? Consent, frequent checks during the scan, and in some cases, packing with ice
Fourth category of MRI risk:
Resistive Heating
Case 4
![Page 43: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/43.jpg)
SUMMARY – MRI Safety
KEY = SCREENING:
• EQUIPMENT
• DEVICES
• Exposure to metal (anywhere)
• ANY possible craniofacial metal exposure? ORBIT RADIOGRAPH / CT
![Page 44: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/44.jpg)
SUMMARY – MRI Safety
•You’ve heard this before, but now more than ever:
•SEE SOMETHING = SAYSOMETHING
•MRI safety is
EVERYONE’sresponsibility!
![Page 45: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/45.jpg)
References
•www.MRI Safety.com (excellent free site with
information about MRI safety of a
variety of medical devices)
•AAPM “Physics of MR Safety:” (https://www.aapm.org/meetings/am
os2/pdf/59-17207-59975-979.pdf)
![Page 46: CT and US Anatomy - ACR](https://reader034.fdocuments.net/reader034/viewer/2022050403/62705406cb50aa3dcd6771b5/html5/thumbnails/46.jpg)
END