PresentationNeuroSpeed® PTA Balloon Catheter &Credo® Stent for intracranial stenosis

AC_P

rese

ntati

on_N

euro

Spee

d_Cr

edo_

1506

05

Confidential: Internal Use only22

Concept

NeuroSpeed® PTA Balloon Catheter in combination with Credo® Stent

Flexible• PTA alone with NeuroSpeed® PTA Balloon Catheter• PTA with NeuroSpeed® PTA Balloon Catheter + Stenting with Credo® Stent

through NeuroSpeed PTA Catheter

Gentle• Gentle compliance concept of the balloon catheter• Precise and controllable inflation behavior

• Small delivery system diameter• No exchange maneuver or additional

crossing of the lesion needed

• Stent for stabilisation and additional support

Optimized stenosis treamtent

reduced snow plow effectless procedurel stepsreduced procedural risks

Safe and accurate inflation

Confidential: Internal Use only33

Procedure:

• Delivery of NeuroSpeed® PTA catheter

• Inflation of NeuroSpeed® PTA catheter

• Deflation of NeuroSpeed® PTA catheter

• Delivery of Credo® Stent through low profile NeuroSpeed PTA Catheter if

necessary

Optimized Procedure

Confidential: Internal Use only44

The NeuroSpeed® PTA Balloon Catheter is indicated for dilatation of a stenosis in the cerebral arteries in order to improve perfusion.

NeuroSpeed® PTA Balloon CatheterProduct Overview

NeuroSpeed® PTA Balloon Catheter

Balloon diameter 2.0 / 2.5 / 3.0 / 3.5 and 4.0 mm

Balloon working length 8 mm

Inner diameter 0,0165‘‘

Outer diameter (dist./ prox.) 2.7 F / 3.7 F

Usable length 150 cm

Tip length 10 mm

Tip markers / Balloon markers 1 / 2

CE pendingfor 1.5 mm

Confidential: Internal Use only55

Product Overview – Radiopaque MarkerNeuroSpeed® PTA Balloon Catheter

Two radiopaque markers indicating the nominal length of the balloon

One radiopaque marker indicating the distal tip of the catheter ( For positioning control)

Longer soft tip for improved trackability

Confidential: Internal Use only66

Product Overview NeuroSpeed® PTA Balloon Catheter

NeuroSpeed® (4.0 x 8 mm) Diameter

PressureDogbone

(max.) nominal

4 bar 3.40 3.72

6 bar 3.60 4.00

8 bar 3.80 4.23

10 bar 3.90 4.37

• Adjustable diameter over pressure range but NO Dogbone Effect

• Less sizes needed for comparable performanceInflationPressure

Balloon Diameter(mm)

bar psi 2.00 2.50 3.00 3.50 4.00

1.0 14.5 1.66 1.93 2.18 2.97 2.92

2.0 29.0 1.72 2.09 2.42 3.06 3.26

3.0 43.5 1.78 2.22 2.62 3.17 3.50

4.0 58.0 1.84 2.33 2.78 3.25 3.72

5.0 72.5 1.92 2.42 2.89 3.39 3.87

6.0 87.0 2.00 2.50 3.00 3.50 4.00

7.0 101.5 2.10 2.59 3.14 3.62 4.14

8.0 116.0 2.16 2.65 3.22 3.69 4.23

9.0 130.5 2.22 2.70 3.29 3.76 4.30

10.0 145.0 2.27 2.75 3.38 3.83 4.37

11.0 159.5 2.33 2.82 3.46 3.90 4.45

12.0 174.0 2.39 2.87 3.54 3.97* 4.53*

13.0 188.5 2.46 2.92 3.64 4.04 4.62

14.0 203.0 2.52* 2.98* 3.73* 4.11 4.70

NeuroSpeed®

Compliance Chart

Nominal Pressure * Rated Burst Pressure

Confidential: Internal Use only77

Credo® Stent

Used in combination with the NeuroSpeed® PTA balloon catheter, the Credo® stent is indicated to increase the vessel diameter of intracranial arteries (for example the ICA, MCA, VA/C4-BA, V4) that are suitable for and accessible to the system (stent with PTA balloon catheter) in patients specific attributes.

Product Overview

Credo® Stent

Stent diameter 3.5 and 4.5 mm

Stent length 15 and 20 mm

Vessel diameter 2.0 – 4.0 mm

Delivery catheter NeuroSpeed® PTA Balloon Catheter

CE pending

Confidential: Internal Use only88

Credo® StentProduct Overview

Concept: Gentle stent performance for stabilization of the lesion but no trauma to surronding pathologic vessel tissue

Potential clinical benefits of a Nitinol self expandable StentThe higher the compression grade the higher the force it exerts

With higher compression grade, the forces needed to further compress the stent are higher When further compressed (higher stenosis grade) the stent affects the surrounding tissue

with higher intensity When less compressed (non stenotic section) the stent affects the surrounding tissue with

lower intensity

Confidential: Internal Use only99

Credo® StentProduct Overview – Radiopaque marker concept

Only stent marker are visible (not the whole stent) advantage for follow-up • Barely any

artefacts• Non shadowing

in angiography

3 Distal X-Ray markers Credo® Stentindicating the opening of the device

Distal transport wire marker Credo® Stentindicating the tip of the transport wire for correct placement within the catheter

Central transport wire markerCredo® Stentindicating the middel of the Stent

Distal radiopaque marker of the NeuroSpeed® PTA balloon catheterfor controlled Stent deployment

Distal and proximal radiopaque balloon marker of the NeuroSpeed® PTA balloon catheter indicating the nominal working length of the balloon

Proximal transport wire markerCredo® Stentindicating the point up to which the stent can be repositioned

Confidential: Internal Use only1010

Procedure

Delivery of low profile NeuroSpeed® and inflation of ballon

i. e. nominal pressure = 6bar

Pathology: symptomatic stenotic lesion

Confidential: Internal Use only1111

Procedure

If necessary CREDO® stent deployment after PTA without any exchange maneuver

Confidential: Internal Use only1212

Procedure

Delivery of low profile NeuroSpeed® and passage of lesion

Pathology: symptomatic stenotic lesion

Confidential: Internal Use only1313

Procedure

Gentle, contrallable Inflation of the NeuroSpeed® PTA catheter

i. e. nominal pressure = 6bar

Confidential: Internal Use only1414

Procedure

If necessary CREDO® stent deployment after PTA without any exchange maneuver

Confidential: Internal Use only1515

Procedure

If necessary CREDO® stent deployment after deflation without any exchange maneuver

Confidential: Internal Use only1616

Treatment Concept

Live presentation of the procedure …