Competitors Product Analysis-1
-
Upload
hafizuddin-razali -
Category
Documents
-
view
215 -
download
0
Transcript of Competitors Product Analysis-1
-
7/30/2019 Competitors Product Analysis-1
1/10
FCompetitor Products Analysis
Table of Contents
-
7/30/2019 Competitors Product Analysis-1
2/10
Analysers from Some of our Main Competitors
ABB
AMETEK (Thermox)
Emerson (Rosemount)
Servomex
Yokogawa
Lamtec
The failures and problems mentioned on the following pages are failures andproblems which were reported from world market customers to ENOTEC
-
7/30/2019 Competitors Product Analysis-1
3/10
ABBThe measuring cell is pasted onthe flange plate with glass solder.
The thermocouple is at the same time the contact of electrodes at the reference air side.
1.The thermocouple fuses to the cell 2. High spot loading results in thedestruction of the platinum coating
Ingress of air at the joint, whichwill not be detected
The measured value is correct
with test gas, but too highwith process gas
Design of the measuring cell with dead-end
High deposit of dust on the ceramic filter disk
Long response time
Measuring cell temperature too low (700C) Incorrect measurement because of
cross sensitivity against combustiblesin the process gas (measured value toolow)
Bad process gas exchange
Process GasLow Pressure
Reference Gas
By vibration the cell breaks offafter few months.
No explosion approval for a complete probe.
-
7/30/2019 Competitors Product Analysis-1
4/10
AMETEK (Thermox)Close-Coupled extractivemeasurement:
Easily soiled in flue gas with dust
Corrosion problems, if thesampled gas is falling below thedew point
Leakage due to vibration through
pasted ZrO2 measuring cell
Inaccurate COe measurement through two PT100 means zero reading most of thetime
Measuring value is correct with test gas (positive pressure) but much to high withextracted flue gas (negative pressure) due to a leaking sampling probe
No explosion approval for a whole measuring device.
Blockages inside the sampling tubes
Leaking or blocked sampling tubes
O2 value too high
-
7/30/2019 Competitors Product Analysis-1
5/10
AMETEK (Thermox)
InSitu Measurement(WDG-INSITU)
Cell temperature 615C 650C,(higher temperatures are not possiblebecause of leakages arising from theceramic glue)
Incorrect measured valuesbecause of cross sensitivityagainst combustibles(measured values too low)
Ingress of air at the joint, which willnot be detected by calibration
Measured value is correct withtest gas, but much to high withprocess gas
Small probe filter easily blocked in coal and oil fired boilers
No explosion approval for a complete probe.
-
7/30/2019 Competitors Product Analysis-1
6/10
Emerson (Rosemount) OXYMITTER
Measuring cell is soldered, but has ameasuring cell design with dead-end
Bad process gas exchange (T90 time approx. 5 minwith flue gas)
Electronics at the terminal box Temperature problems and overheated electronics
By default no protection tube
Flow meter and pressure reducer forreference air and test gas have to bemounted separately.
High additional assembly effort, poor design, highmaintenance costs
No internal flow rate monitoring
Problems in case of measuring cell exchange through
increased corrosion
No automatic calibration for explosion approved probes
High wear of the probe body, abrasive flue gas willreduce the probe life time below 1 year
Flange connection of the cell is
located in the hottest area
Cell temperature too low (720C) Incorrect measured values because of cross sensitivity
against combustibles (measured values much too lowwith combustibles)
No explosion approval for a functional probe.
-
7/30/2019 Competitors Product Analysis-1
7/10
Emerson (Rosemount) OCX 4400
Easily soiled
Corrosion problems if the samplinggas is falling below the dew point
Inaccurate COe measurement through two PT100 means zero reading all the time
Measuring value is correct with test gas (positive pressure) but much to highwith extracted flue gas (negative pressure) due to a leaking sampling probe
Blockages inside the sampling tube
Leaking or blocked sampling tube
Close-CoupledExtractiveMeasurement:
-
7/30/2019 Competitors Product Analysis-1
8/10
Servomex
Great danger of soiling
Measuring gas tube blocks
O2 measuring cell temperature toolow
Incorrect measured values because ofcross sensitivity against combustibles(measured values much too low)
COe measurement by two PT100
Small critical nozzle
Extractive measurement Great danger of soiling
Only the control unit is ATEX explosion approved, not the probe.
High drift and bad resolution below500 ppm COe
-
7/30/2019 Competitors Product Analysis-1
9/10
Yokogawa
Low measuring cell temperature (700C),since the measuring cell is pasted
Incorrect measured values because ofcross sensitivity against combustibles(measured values much too low)
Design of the measuring cell with dead-end Bad gas exchange
Flange connection of the cell is located inthe hottest area and is directly exposed tothe flue gas, since the standard probe does
not have a filter.
Problems in case of measuring cellexchange through increased corrosion andsoiling of the cell
Ingress of air at the joint, which will not bedetected by calibration since leak airexpelled during calibration.
Measured value is correct with test gas, butincorrect with process gas
No long probes, because 2 m probes and above would shake themselves to pieces and parts
No explosion approved at all
-
7/30/2019 Competitors Product Analysis-1
10/10
Lamtec
Cell temperature (600C), highertemperatures are not possible becauseof arising leakages of the ceramic glue
Incorrect measured values because ofcross sensitivity against combustibles(measured values too low)
Only short probes just for clean flue gas probe approx. 30 mm long
O2 measuring cell from the motorcartechnology ( probe), measuring cell
pasted
Flue gas temperature only up to max.
300C. Cell leakage at higher flue gas
temperatures
Bypass principle which blocks easilywith dust
Measured value can be correct with testgas, but slow and inaccurate withprocess gas due to dust which settles onthe sensor
Lambda ProbeLS2
No explosion approved at all