Comparison of conventional lateral-flow assays and a new fluorescent immunoassay to detect influenza...
-
Upload
alejandro-r -
Category
Documents
-
view
215 -
download
2
Transcript of Comparison of conventional lateral-flow assays and a new fluorescent immunoassay to detect influenza...
G
V
S1
Ci
2
3
GQ14
D5
6
7
A8
R9
R10
A11
A
12
K13
L14
F15
E16
117
18
b19
l20
m21
m22
i23
l24
R25
o26
27
t28
p29
a30
o31
a32
p33
t34
t35
US
0h
ARTICLE IN PRESS Model
IRMET 12078 1–4
Journal of Virological Methods xxx (2013) xxx– xxx
Contents lists available at SciVerse ScienceDirect
Journal of Virological Methods
jou rn al h om epage: www.elsev ier .com/ locate / jv i romet
hort communication
omparison of conventional lateral-flow assays and a new fluorescentmmunoassay to detect influenza viruses
ary P. Leonardi ∗, Adele M. Wilson, Alejandro R. Zurettiepartment of Pathology & Laboratories, Nassau University Medical Center, East Meadow, NY, United States
rticle history:eceived 9 August 2012eceived in revised form 14 February 2013ccepted 18 February 2013vailable online xxx
eywords:ateral-flow immunoassayluorescence
a b s t r a c t
Sofia, a novel, fluorescent lateral-flow immunoassay was compared with two conventional colorimetricassays, Quickvue Influenza A+B and Directigen FLU A+B, to identify influenza viral antigen from patientnasopharyngeal specimens. A total of 118 frozen original influenza-positive specimens and 57 prospec-tive specimens were examined. Using rt-PCR as a referee assay, sensitivity values (%) for influenza A/B of80.0/74.8, 73.3/59.3 and 73.3/40.7 were obtained using the Sofia, Quickvue and Directigen assays, respec-tively. All assays demonstrated reduced sensitivity for influenza B as compared with influenza A virus.With respect to the Sofia assay, the sensitivity of influenza B for the Directigen assay was significantlydiminished.
nhanced influenza sensitivity False positive results were not observed in the Sofia and Directigen assays. The Quickvue assay produced3 false-positive results (2 influenza A and 1 influenza B) resulting in a specificity (%) of 96 and 98 forinfluenza A and B, respectively. Cross-reactivity to other respiratory viruses was not observed amongimmunoassays. A sensitivity rank (highest to low) of rt-PCR > culture > Sofia> Quickvue > Directigen wasestablished using dilutions of influenza A and B. Sofia provides enhanced sensitivity and objective result
ntion
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
interpretation over conve
. Introduction
Diagnostic immunoassays rely on specific antigen–antibodyinding for accurate results. Although variations in format exist,
ateral flow immunoassays generally use a hydrophobic reactionembrane strip containing bound, anti-target antibodies. Speci-en, treated with reagents to produce target antigen-conjugates,
s allowed to migrate along the membrane, producing a coloredine if the antigen-conjugate is captured by the bound antibodies.esult interpretation is subjective and dependent on the strengthf the color signal.
Advantages demonstrated by lateral flow assays have madehem a cornerstone for influenza virus detection. Rapid result timeermits point-of-care testing to occur in emergency departmentsnd physician office settings on a 24 h, 7-day-a-week basis. Thus,ptimal health care with respect to the administration of antiviralgents or the institution of infection control measures, includingatient isolation, can occur readily. Immunoassays are also rela-
Please cite this article in press as: Leonardi, G.P., et al., Comparison of condetect influenza viruses. J. Virol. Methods (2013), http://dx.doi.org/10.1016
ively inexpensive, require little or no equipment, little technicalraining, and produce results in about 15 min.
∗ Corresponding author at: Virology & Molecular Pathology Laboratory, Nassauniversity Medical Center, 2201 Hempstead Tpke, East Meadow, NY 11554, Unitedtates. Tel.: +1 516 572 8711; fax: +1 516 572 6409.
E-mail address: [email protected] (G.P. Leonardi).
54
55
56
166-0934/$ – see front matter © 2013 Published by Elsevier B.V.ttp://dx.doi.org/10.1016/j.jviromet.2013.02.008
al colorimetric immunoassays.© 2013 Published by Elsevier B.V.
Despite these advantages, the sensitivity of rapid influenza
immunoassays is lacking with respect to that obtainable using cul-ture or molecular assays. This was exemplified during the recent
influenza A/2009 H1N1 pandemic. Reports of low, unacceptable
sensitivity values, ranging from 18 to 70% depending on which ref-
eree assay was chosen for comparison, tarnished the perception
of immunoassay worth as diagnostic tools (CDC, 2009; Ginocchio
et al., 2009; Leonardi et al., 2010; Leonardi, 2011).
A novel lateral-flow immunoassay, Sofia (Quidel; San Diego,
CA, USA) has recently received CLIA-waived status for diagnos-
tic testing. Instead of subjective, colorimetric results, this assay
employs immunofluorescent technology to detect influenza nucle-
oprotein, producing objective data. Sofia was compared with two
other lateral-flow immunoassays, Quickvue Influenza A+B (Quidel)
and Directigen FLU A+B (Becton Dickinson, Sparks, MD, USA), to
identify influenza viruses from original nasopharyngeal specimens.
The Directigen and Quickvue immunoassays have been utilized
widely for diagnostic testing and research evaluations (Chartrand
et al., 2012). The latter assay has also been CLIA-waived for diag-
nostic use.
2. Materials and methods
ventional lateral-flow assays and a new fluorescent immunoassay to/j.jviromet.2013.02.008
2.1. Specimens 57
A total of 118 original patient nasopharyngeal swab speci- 58
mens, identified previously as influenza-positive by immunoassay, 59
ING Model
V
2 irologi
c60
i61
d62
t63
264
s65
f66
(67
a68
69
D70
w71
q72
r73
w74
o75
w76
s77
i78
w79
80
a81
m82
p83
u84
(85
a86
287
88
t89
i90
o91
u92
293
94
W95
m96
t97
n98
t99
s100
r101
p102
2103
104
v105
3106
3107
108
s109
f110
w111
b112
i113
B114
b115
116
117
118
119
120
121
122
123
124
125
126
127
128
129
130
131
132
133
134
135
136
137
138
139
140
141
142
143
144
145
146
147
148
149
150
151
152
153
154
155
156
157
158
159
160
161
162
163
164
165
166
167
168
169
170
171
172
173
174
175
ARTICLEIRMET 12078 1–4
G.P. Leonardi et al. / Journal of V
ulture and/or rt-PCR were used to evaluate immunoassays. Spec-mens were stored at either −20 or −70 ◦C and were obtaineduring the respiratory virus seasons from 2006 to 2011. In addi-ion, 57 nasopharyngeal specimens, obtained during the current011–2012 influenza season, were evaluated prospectively. Thesepecimens were stored at 2–8 ◦C prior to testing which was per-ormed within 72 h post-collection. Determination of patient agechildren 12 and under vs. adults) and influenza A subtype werelso made on study specimens.
Aliquots of consensus-positive influenza A and B virus (NYS-OH, proficiency testing event # 121, sample nos. 1312 & 1314)ere used to prepare sample dilutions which were tested subse-
uently by immunoassay, shell vial culture and rt-PCR methods. Aank order of sensitivity was established. Different dilution seriesere used for the influenza A and B samples because of the high titer
f the influenza A specimen. Dilutions were chosen and tabulatedhen an immunoassay failed to identify virus. At each dilution, a
ample was tested by each immunoassay once. When a negativemmunoassay results occurred at a given dilution, a second sample
as tested to confirm the result.Other common respiratory viruses (respiratory syncytial virus,
denovirus, parainfluenza types 1–3 and human metapneu-ovirus) were also tested by the immunoassays using original
atient specimens. Presence of these viruses was confirmedsing shell-vial culture and immunofluorescent antibody methodsD3Ultra DFA respiratory virus and D3 DFA Metapneumovirus IDssays; DHI, Athens, OH, USA).
.2. Immunoassays
All immunoassays were performed according to manufac-urer’s procedures. Specimens were tested simultaneously by allmmunoassays. Comparison of immunoassays with respect to easef use, result reporting, result time, and equipment needs was eval-ated.
.3. Resolution of discrepant immunoassay results by rt-PCR
A molecular rt-PCR assay (ProFLU+, GenProbe/Prodesse;aukesha, WI, USA) served as the “gold standard” confirmatoryethod to decipher discordant or negative results in specimens
ested among the immunoassays. Shell vial culture (R-mix; Diag-ostic Hybrids, Athens, OH, USA) was also inoculated (100 �L) ifhe remnant specimen volume allowed, so that comparative sen-itivity of the immunoassays and culture could be evaluated withespect to gold standard rt-PCR. Cultures were harvested at 48 host-inoculation and examined using fluorescent antibodies (DHI).
.4. Statistics
95% confidence intervals were computed for assay sensitivityalues using a modified method of Wald (Agresti and Coull, 1998).
. Results
.1. Specimens
In 118 influenza-positive specimens (92 type A and 26 type B), 3pecimens negative by all immunoassays and rt-PCR were excludedrom the study. Of the remaining 115 retrospective specimens, 20ere negative among all immunoassays (16 type A and 4 type B)
Please cite this article in press as: Leonardi, G.P., et al., Comparison of condetect influenza viruses. J. Virol. Methods (2013), http://dx.doi.org/10.1016
ut were confirmed as positive using rt-PCR. Virus was identifiedn 5 of the 57 prospective specimens (4 influenza A and 1 influenza). One prospective specimen was negative by all immunoassaysut confirmed as A/H3-like using rt-PCR.
PRESScal Methods xxx (2013) xxx– xxx
When retrospective and prospective specimens are separately
analyzed, similar sensitivity/specificity values (%) for influenza A
and B of approximately 80/100 in were obtained for the Sofia assay.
Quickvue and Directigen assay sensitivity for influenza A and B in
the retrospective samples were also essentially the same as when
all samples were combined. However, influenza A sensitivity val-
ues (%) in the Quickvue and Directigen assays were respectively
reduced in the prospective samples, producing values of 60 (3 of 5
positive) and 40 (2 of 5 positive samples). All assays produced 100%
sensitivity for influenza type B because all correctly identified the
one positive prospective sample.
Specimens from children, 12 and under accounted for 84 of
the total 175 (48%) samples studied. Of these, 58 of 84 (70.2%)
were influenza influenza-positive (51 type A, 7 type B). The pan-
demic influenza A/2009/H1N1subtype accounting for 38 of the
51 (74.5%) influenza A-positive specimens among children. Of thetotal 97 influenza A specimens studied, 48, 31 and 7 were identi-fied as pandemic A/2009/H1N1, seasonal A/H3N2-like and seasonalA/H1N1-like agents, respectively. Subtype identity was not deter-
mined for 11 of the 97 influenza A specimens.
3.2. Immunoassay performance
Discrepant results (10 influenza A and 12 influenza B specimens)
were produced among the immunoassays (Table 1). Using this data,
sensitivity values (%) for influenza A/B of 80.0/74.8, 73.3/59.3 and
73.3/40.7 were obtained with the Sofia, Quickvue and Directigen
assays, respectively. All immunoassays were more sensitive for
influenza A rather than influenza B viruses. No false positive results
were obtained for the Sofia and Directigen assays; the Quickvue
assay exhibited 3 false-positive results, producing specificity (%)
values of a 96.0/98.0 for influenza A/B, respectfully.
Among 33 cases of discrepant or negative immunoassay results,
enough specimen volume remained to permit inoculation of R-
mix shell-vial cultures. With respect to rt-PCR, culture identified
influenza in 27/33 samples (81.8% sensitivity). All immunoassays
produced negative results when tested with other common respi-
ratory viruses.
Dilutions of known positive-influenza A and B specimens,
obtained from remnant proficiency testing samples, compared
viral detection among immunoassay, culture and molecular assays
(Table 2). A rank order for sensitivity was established. From high
to low: rt-PCR > culture > Sofia > Quickvue > Directigen. The Directi-
gen and Quickvue assays performed poorly in identifying influenza
B samples, failing to identify antigen at dilutions of 1:30 and 1:75,
respectively. Sofia produced positive influenza B results at a dilu-
tion of 1:800. Although the use of a single sample to presently test
the various diagnostic assays at each dilution precludes the ability
to calculate statistical significance, the results clearly demonstrate a
trend toward increased sensitivity for the Sofia assay and warrants
further, more extensive investigation.
In comparing these immunoassays from a laboratory-use per-
spective, the Quickvue assay was simplest to perform, required
less procedural steps, offered the fastest result time and did not
require any operational equipment. However, result interpreta-
tion for Quickvue was cumbersome as the technician sometimes
labored to visualize a pink result line imposed on a pink mem-
brane background. Although the background of the Directigen
cassette membrane was less pink in color, visualization of faintly-
pink positive results were also sometimes difficult to interpret. The
Directigen assay did not require equipment in its operation and
produced results in approximately 15 min.
ventional lateral-flow assays and a new fluorescent immunoassay to/j.jviromet.2013.02.008
The Sofia assay required a fluorescent analyzer to perform the 176
assay. It also had the most procedural steps in assay operation. 177
Despite these limitations, Sofia demonstrated enhanced sensitiv- 178
ity over its competitors, especially with respect to influenza B 179
ARTICLE IN PRESSG Model
VIRMET 12078 1–4
G.P. Leonardi et al. / Journal of Virological Methods xxx (2013) xxx– xxx 3
Table 1Performance characteristics of 3 direct antigen detection assays using rt-PCRa as a referee assay to evaluate discrepant and negative immunoassay results.b
ITEM Influenza type A specimens Influenza type B specimens
Sofia Quickvue Directigen Sofia Quickvue Directigen
TP 72 66 66 22 16 11FN 18 24 24 5 11 16TN 51 49 51 51 50 51FP 0 2 0 0 1 0SENSITIVITY (%) 80.0 73.3 73.3 74.8 59.3 40.7CI (95%)c 70.5–87.0 63.3–81.4 63.3–81.4 62.8–91.0 40.7–75.5 24.5–59.3SPECIFICITY (%) 100 96 100 100 98 100CI (95%) 91.6–100 86–99.7 91.6–100 91.6–100 88.7–99.9 91.6–100PPV (%) 100 97 100 100 94 100NPV(%) 74 67 68 91 82 76.
TP: true positive; FN; false negative; TN: true negative; FP: false positive. Sensitivity was calculated as TP/TP + FN × 100%. Specificity was calculated as TN/TN + FP × 100%.PPV: positive predictive value, calculated as TP/TP + FP; NPV: negative predictive value, calculated as TN/TN + FN. CI: 95% confidence interval.
a ProFlu+ and ProFAST+ assays (GenProbe-Prodesse; Waukesha, WI).b In 33 cases of discrepant or negative results among the antigen detection assays, enough specimen remained for inoculation of R-mix shell vials. Viral antigen was
culture-confirmed in 27 of 33 cases (81.8% sensitivity).c Confidence Interval (95%). Statistical significance demonstrated between Sofia and Directigen influenza B sensitivity.
Table 2Ability of diagnostic assays to identify influenza A and B viruses using dilutions of positive proficiency testing samples (New York State Department of Health-CLEP Program).
Specimen Dilution Diagnostic assay resultsa
Antigen detection Cultureb rt-PCRc
Sofia Quickvue Directigen
Influenza A1:100 + + + + ND1:400 + − − + ND1:3000 + − − + ND1:5000 − ND ND + +1:10,000 − ND ND − +
Influenza B1:25 + + + + ND1:30 + + − + ND1:75 + + − + ND1:100 + − ND + ND1:800 + − ND + ND1:3200 − ND ND + +1: 12,800 − ND ND − +
ND: no data.a All assays were performed according to manufacturer’s recommended procedures.b 180 �
IoFlu+
v180
p181
s182
a183
w184
4185
186
i187
m188
p189
i190
i191
i192
i193
194
b195
m196
o197
p198
2199
200
201
202
203
204
205
206
207
208
209
210
211
212
213
214
215
216
217
Shell vial cultures (R-mix, DHI, Athens, OH) were inoculated with specimens (nc.).
c Nucleic acids extracted using an EasyMag (Biomerieux, Inc.) and tested using Pr
iruses. In addition, Sofia allowed objective results to be obtained,ermitted the potential integration into a laboratory informationystem, could store 500 results in the analyzer memory, provided
hard-copy result and could be operated in either a read now oralk-away testing mode.
. Discussion
Although the role of lateral flow immunoassays for rapidnfluenza testing was well established, the reliability of these
ethods came into question during the 2009 influenza A/H1N1andemic. This outbreak necessitated an accurate diagnosis of
nfluenza. Further, the ability to distinguish seasonal and pandemicnfluenza A/H1N1 subtypes was necessary because of the differ-ng antiviral drug sensitivity profiles of these viruses. Lateral-flowmmunoassays could not perform either function effectively.
In subsequent influenza seasons, enzyme immunoassays haveeen utilized in a limited capacity. Adequate assay specificity per-
Please cite this article in press as: Leonardi, G.P., et al., Comparison of condetect influenza viruses. J. Virol. Methods (2013), http://dx.doi.org/10.1016
its its use as a first screening tool to rule in disease. Meta-analysisf 159 studies comparing rapid influenza immunoassays reported aooled specificity (%) of 98.2, supporting this role (Chartrand et al.,012). However, highly heterogeneous sensitivity values (95% CI of
l), harvested after 24 h and stained with appropriate monoclonal antibodies (DHI,
and ProFAST+rt-PCR assays (Gen Probe/Prodesse; Waukesha, WI).
57.9% to 66.6%) clearly suggest that negative results cannot stand
alone and require confirmation by DFA, cell culture or molecularassay.
Direct fluorescent-antibody (DFA) immunoassays demonstrate
enhanced sensitivity over their enzymatic counterparts with
one investigation reporting similar performance characteristics
between DFA and PCR in identifying pandemic influenza (Pollack
et al., 2009). An influenza A/2009 H1N1-specific fluorescent
immunoassay was also developed, permitting laboratories lacking
molecular diagnostic capability the ability to definitively identify
this new pathogen (Leonardi, 2010). Unlike enzyme immunoassays,
DFA requires extensive technical training, equipment, and greater
assay turn-around-times. These pitfalls would thus preclude DFA
testing in any point-of-care or emergency room setting.
A decrease in sensitivity for influenza B as compared to influenza
type A was demonstrated among all immunoassays. This finding
parallels the meta-analysis study which reported pooled sensitiv-
ity values (%) of 64.6/52.2 for influenza A/B, respectively Chartrand
ventional lateral-flow assays and a new fluorescent immunoassay to/j.jviromet.2013.02.008
et al., 2012). A plausible explanation for this finding may rest with 218
the comparative “virulence” of the influenza types. Influenza A 219
leads to more severe disease and higher rates of hospitalization 220
and death than infection with type B (Fiore et al., 2011). Differences 221
ING Model
V
4 irologi
i222
w223
s224
s225
s226
p227
s228
s229
r230
231
o232
s233
b234
i235
d236
w237
238
s239
s240
s241
c242
t243
p244
A245
o246
o247
i248
249
w250
u251
f252
t253
f254
u255
t256
3257
s258
u259
260
S261
t262
p263
p264
w265
b266
g267
268
s269
f270
t271
d
272
273
274
275
276
277
278
279
280
281
282
283
284
285
286
287
288
289
290
291
292
293
294
295
296
297
298
299
300
301
302
303
304
305
306
307
308
309
310
311
312
313
314
315
316
317
318
319
Leonardi, G.P., 2011. Immunodiagnostic assays, influenza and the H1N1 pandemic. 320
ARTICLEIRMET 12078 1–4
G.P. Leonardi et al. / Journal of V
n virulence also exist among individual sub types of influenza A,here A/H3N2 demonstrates far greater disease severity than sea-
onal A/H1N1 infection (Chartrand et al., 2012). Greater diseaseeverity usually means greater viral load and thus, greater observedensitivity in studies that predominantly examine influenza A sam-les, particularly the A/H3N2-like subtype. Among influenza Apecimens evaluated presently, 7, 31 and 48 were identified aseasonal H1-like, seasonal H3-like and 2009 H1N1-like subtypes,espectively.
Variations in assay sensitivity can also be influenced by the agef patients submitting specimens for study. Influenza immunoas-ays reportedly perform better in children than adults, presumablyecause higher viral loads and longer duration of shedding occur
n the former (Fiore et al., 2011). Meta-analysis has reportedlyemonstrated an approximate 13% increase in assay sensitivityhen children were studied (Chartrand et al., 2012).
The Quickvue and Directigen assays produced modestly greaterensitivity values than obtained in the meta-analysis of pooledtudies examining these products. In addition, influenza A/Bensitivity for Sofia was even higher than observed among itsompetitors. A study bias based on the selection of specimensested may account for the increased sensitivity values observedresently. Factors including a high percentage of positive-influenza
to B specimens (ratio of approximately 4:1), the high percentagef children’s specimens studied (approximately 50%) and a paucityf seasonal H1N1 samples would all favor increased assay sensitiv-ty.
A second source of bias may exist with respect to samples whichere positive by all immunoassays but not confirmed presentlysing gold standard rt-PCR. Positive confirmation was not per-ormed because the amount of sample left following immunoassayesting was often insufficient to permit such analysis. Further, theact that these specimens were previously identified as positivesing methods considered more sensitive than immunoassay (cul-ure and/or rt-PCR) and were currently identified as positive by
immunoassays which detect differing antigens and or antigenicites seem to make the potential for false positive results highlynlikely.
In head-to-head, same-specimen, simultaneous evaluation, theofia outperformed its competitors providing greater accuracy overraditional colorimetric immunoassays. Sofia also produced com-arable sensitivity to culture, however, the frozen-retrospectiveatient specimens may have contained non-viable virus whichould enhance antigen immunoassay detection but negatively
ias culture results. In the dilution studies, culture demonstratedreater sensitivity over the Sofia assay.
Sofia may represent a next step in the evolution of immunoas-
Please cite this article in press as: Leonardi, G.P., et al., Comparison of condetect influenza viruses. J. Virol. Methods (2013), http://dx.doi.org/10.1016
ays, combining lateral flow and fluorescent antibody detectionormats. The choice of fluorescent rather than colorimetric detec-ion is at least partially responsible for the improved sensitivityemonstrated by Sofia because the former is known to increase
PRESScal Methods xxx (2013) xxx– xxx
sensitivity while also widening the dynamic assay detection range
over the latter (Gibbs, 2001). Even greater sensitivity in future
assays may be achieved with the development of luminescent
immunoassays, which presumably may amplify results similar to
that seen in PCR.
With respect to sensitivity, the Sofia assay demonstrated a
statistically significant increase in influenza B viruses over the
Directigen assay. Although a trend toward higher sensitivity for
influenza A was observed for Sofia, the results among assays were
equivalent statistically. A larger clinical study may be needed to
demonstrate any potential significance for influenza A viruses
among these assays.
Features offered by Sofia, including production of objective
results, operation in a run-now or walk-away mode, and having
hard copy and memory stored results make it valuable to the diag-
nostic laboratory. The enhanced sensitivity demonstrated by the
Sofia assay helps reestablish the value of these methods for rapiddiagnostic testing, allowing meaningful patient management deci-sions to be made at the point-of-care.
Acknowledgements
The authors are grateful to Dr. Paul Olivo, Quidel Corporation forhis assistance in the statistical analysis of the data.
Sofia and Quickvue assay kits used in this study were donated
by Quidel Corporation.
References
Agresti, A., Coull, B.A., 1998. Approximate is better than exact for interval estimationof binomial proportions. Am. Stat. 52, 119–126.
Centers for Disease Control and Prevention, 2009. Evaluation of rapid influenza diag-
nostic tests for detection of novel influenza A (H1N1) virus-United States, 2009.
Morb. Mortal. Wkly. Rep. 58, 826–829.
Chartrand, C., Leeflang, M.G., Minion, J., Brewer, T., Madhukar, P., 2012. Accuracy of
rapid influenza diagnostic tests. A meta-analysis. Ann. Int. Med. 156, 500–511.
Fiore, A.E., Fry, A., Shay, D., Gubareva, D., Breese, L., Uyeki, T.M., 2011. Centers for
Disease Control and Prevention (CDC). Antiviral agents for the treatment andchemoprophylaxis of influenza-recommendations of the Advisory Committee
on Immunization Practices (ACIP). Morb. Mortal. Wkly. Rep. 60, 1–24.
Gibbs, J. 2001. Selecting the Detection System – Colorimetric, Fluorescent, Lumines-
cent Methods. ELISA Technical Bulletin No 5. Corning Life Sciences, Corning, Inc.Corning, NY. Revised 1/2001.
Ginocchio, C., Zhang, F., Manji, R., Arora, S., Bornfreund, M., Falk, L., Lotikar, M.,
Kowaska, M., Becker, G., Korologos, G., de Gerinomo, M., Crawford, J.M., 2009.
Evaluation of multiple test methods for the detection of novel influenza A (H1N1)during the New York City outbreak. J. Clin. Virol. 45, 191–195.
Leonardi, G.P., 2010. Rapid identification of 2009 H1N1 influenza A virus using flu-
orescent antibody methods. Am. J. Clin. Pathol. 134, 910–915.
Leonardi, G.P., Mitrache, I., Pigal, A., Freedman, L., 2010. Public hospital-based expe-
rience during an outbreak of pandemic influenza A (H1N1) virus infections. J.
Clin. Microbiol. 48, 1189–1194.
ventional lateral-flow assays and a new fluorescent immunoassay to/j.jviromet.2013.02.008
Clin. Laboratory Int. 35 (April and May), 32–34. 321
Pollack, N.R., Doung, S., Cheng, A., Han, L., Smole, S., Kirby, J.E., 2009. Ruling out 322
novel H1N1 influenza virus infection with direct fluorescent antigen testing. 323
Clin. Infect. Dis. 49, 66–68. 324