Compact Data Logger with Color Monitor MobileCorder MV100 ...

11
Bulletin 7901-01E MV100 Compact Data Logger with Color Monitor Small and lightweight (approximately 4 kg) 5.5-inch TFT color LCD Three different removable storage options: 3.5-inch floppy disk, Zip disk, PCMCIA ATA flash memory card Ethernet capability (standard) Sophisticated software Highly reliable hardware http://www.yokogawa.co.jp/Measurement/English/ Yokogawa Electric WWW server:

Transcript of Compact Data Logger with Color Monitor MobileCorder MV100 ...

Page 1: Compact Data Logger with Color Monitor MobileCorder MV100 ...

Bulletin 7901-01E

MV100Compact Data Logger with Color Monitor

Small and lightweight (approximately 4 kg) 5.5-inch TFT color LCD Three different removable storage options: 3.5-inch floppy disk, Zip disk, PCMCIA ATA

flash memory card Ethernet capability (standard)

Sophisticated software Highly reliable hardware

http://www.yokogawa.co.jp/Measurement/English/Yokogawa Electric WWW server:

Page 2: Compact Data Logger with Color Monitor MobileCorder MV100 ...

Models:2-channel model: 125 ms measurement interval4-channel model: 125 ms measurement interval6-channel model: 1 second measurement interval*12-channel model: 1 second measurement interval*(*: Measurement interval is 2 seconds when theA/D integrating time is set to 100 ms.)

Removable storage medium: 3 options (3.5-inchfloppy disk, Zip disk, PCMCIA ATA flash memory card)

Inputs: DC voltages, thermocouples, resistancetemperature detectors, and digital inputs can be mixed.

Setting a new standard for data log-ger mobilityToday users want mobile capabilities in all types of de-vices that use information. Cellular phones and note-book computers are embodiments of this concept ofmobility. With the MobileCorder MV100, YOKOGAWAis now setting the mobility standard for data loggers.

Large-capacity recording memoryThe MV100's internal memory can save approximately 27 hours of continuous data, recordingin 1-second intervals (when using a 6-channel model). Data capacity can be increased toapproximately 5 months' worth of continuous under the same recording conditions by using aPCMCIA ATA flash memory card (160 MB) as a removable storage medium.

Advanced data communicationscapabilityThe MV100 is standard equipped with an Ethernet (10BASE-T) port for high-speed communi-cations. The Ethernet capability makes it possible to form a simple network of PCs and MV100units using a hub, or connect the MV100 to a LAN.

Application softwareThe standard application software includes data display functions and MV100 setting func-tions. Optional software (sold separately) is also available with more advanced networkingcapabilities (e.g., file transfers and data monitoring).

Compact Data Logger with Color MonitorMobileCorder MV100Compact Data Logger with Color MonitorMobileCorder MV100

Large-capacity recording memoryThe MV100's internal memory can save approximately 27 hours of continuous data, recordingin 1-second intervals (when using a 6-channel model). Data capacity can be increased toapproximately 5 months' worth of continuous under the same recording conditions by using aPCMCIA ATA flash memory card (160 MB) as a removable storage medium.

Advanced data communicationscapabilityThe MV100 is standard equipped with an Ethernet (10BASE-T) port for high-speed communi-cations. The Ethernet capability makes it possible to form a simple network of PCs and MV100units using a hub, or connect the MV100 to a LAN.

Application softwareThe standard application software includes data display functions and MV100 setting func-tions. Optional software (sold separately) is also available with more advanced networkingcapabilities (e.g., file transfers and data monitoring).

Page 3: Compact Data Logger with Color Monitor MobileCorder MV100 ...

4

1 5.5-inch wide-viewing-angle color LCD display2 MV100 status display area3 Trend display areaThe fastest trend display updating rate is 1 min/div (approximately 615mm/h in terms of display speed).

4 Digital display area The display updating interval is 1 second.

5 Power ON/OFF switch6 Removable storage drive

6

1

4

3

2

5

Full-scalephotograph

Page 4: Compact Data Logger with Color Monitor MobileCorder MV100 ...

5

Bar graph display(maximum 6-channel switching display)Vertical and horizontal bar graphs can be selected.

Digital display(maximum 6-channel switching display)Displays digital measurements, as well as channel/tag numbers, engineering units, and alarm statuses.

Historical trend displayAllows past data saved in memory to be playedback. In addition, historical and current trends canbe viewed at the same time.

Overview displayAllows digital readings and alarm statuses on allchannels (including calculation channels) to bemonitored.

Information displayDisplays an alarm summary, message summary,and report data.

Trend display(maximum 12 simultaneous channels)The trend display direction (vertical, horizontal) andbackground color (white, black) can be switched.

7 Ethernet (10BASE-T)8 RS-232, RS-422-A/485 (optional)9 Power inlet0 Alarm output, remote control, etc. (optional)A Input modulesDC voltage, TC, RTD, and digital inputs (can be mixed). Clampterminals (screw terminals are optional).

DC model (specify when placing order)The main unit operates on DC power only.

The power inlet ischanged.

Display type

The DC model comes with an ACadapter.

The MV100 has extendible legsfor supporting the MV100 at anangle.

System components

A

9

<

8

7

Rear view of MV100

(small and lightweight) System components

Page 5: Compact Data Logger with Color Monitor MobileCorder MV100 ...

6

Removable storagemedium

PC ordata server

PCMCIA ATA flash memory card(maximum 160 MB)

Transferdata

Simultaneous extended-period data storage and detailed analysis Measurement dataThe MV100 can save data in two formats (display data and event data).

Display data—for extended-period trend recordingThe display data format is used to save data displayed aswaveforms. Each time the waveform display is updated, twodata values (maximum and minimum values) measured sincethe previous update are saved.

Event data—for detailed analysisThe event data format is used to save all data in a specifieddata saving interval. Event data can be used in combinationwith the trigger functions to detect and analyze abnormal data.A pretrigger can also be set, making it possible to analyzedata before and after the trigger.

Measurement data

Reliable data storage in internal memoryThe MV100 saves measurement data in internal memory. Data in internal memory can also be transferred to PCs or data serverseither online or using a removable storage medium. The measurement data memory consists of nonvolatile flash memory (1.2 MB)that does not require a battery backup. This means data written to memory will not be lost due to events such as a power outage.

3.5-inch floppy disk(1.44MB : 2HD)

Zip disk(100MB)

Savedata

Internalmemory

Enlargement

Maximum value

Minimum value

1 minute (1 dot)

Conceptual illustration of displayed data file (with waveform refresh rate of 30 min/div)

These two data values are saved every minute.

1div 30 minutes (30 dots)

Data saved from pretrigger functions

Saved data

Trigger detection

Measurementdata

Page 6: Compact Data Logger with Color Monitor MobileCorder MV100 ...

7

Measurementchannels

2

4

6

12

125 msApproximately

4.1 hoursApproximately

4.1 hours

500 msApproximately

16.6 hoursApproximately

16.6 hours

1 secondApproximately

33.3 hoursApproximately

33.3 hoursApproximately

27.7 hoursApproximately

13.8 hours

10 secondsApproximately

13.8 daysApproximately

13.8 daysApproximately

11.5 daysApproximately

5.7 days

Maximuminternal memorydata saving times

Saving interval

Extended-period data saving1. Saving data to internal memoryThe tables below present examples of the maximum internal memory data saving times.Approximately one day's worth of event data can be saved, using 6 channels and a saving interval of one second.

2. Saving data to removable storage mediumMV100 data are saved as files to a removable storage medium.The tables below present examples of the maximum data saving times for a PCMCIA ATA flash memory card (160 MB).

Measurementchannels

2

4

6

12

2 secondsApproximately

2.3 daysApproximately

1.7 daysApproximately

1.1 daysApproximately

13.8 hours

4 secondsApproximately

4.6 daysApproximately

3.4 daysApproximately

2.3 daysApproximately

1.1 days

10 secondsApproximately

11.5 daysApproximately

8.6 daysApproximately

5.7 daysApproximately

2.8 days

1 minuteApproximately

69.4 daysApproximately

52 daysApproximately

34.7 daysApproximately

17.3 days

Maximuminternal memorydata saving times

Saving interval

Display updating interval (min/div)1 minute 2 minutes 5 minutes 30 minutes

Saving intervalMeasurementchannels

2

4

6

12

125 msApproximately

57 days Approximately

28 days

500 msApproximately

231 daysApproximately

115 days

1 secondApproximately

462 daysApproximately

231 daysApproximately

154 daysApproximately

77 days

10 secondsApproximately

12.6 yearsApproximately

6.3 yearsApproximately

4.2 yearsApproximately

2.1 years

Maximum data saving times for PCMCIA ATA flash memory card

Measurementchannels

2

4

6

12

2 secondsApproximately

462 daysApproximately

231 daysApproximately

154 daysApproximately

77 days

4 secondsApproximately

925 days Approximately

462 days Approximately

308 days Approximately

154 days

10 secondsApproximately

6.3 yearsApproximately

3.1 yearsApproximately

2.1 yearsApproximately

1 year

1 minuteApproximately

38 yearsApproximately

19 yearsApproximately

12.5 yearsApproximately

6.3 years

Maximum data saving times for PCMCIA ATA flash memory card

Saving interval

Display updating interval (min/div)1 minute 2 minutes 5 minutes 30 minutes

Display data file only (no calculation channel)

Display data file only (no calculation channel)Event data file only (no calculation channel)

Event data file only (no calculation channel)

File structureThe two data formats can be used in combinations such asthe following:1 Display data only2 Event data only3 Display data and event data in combination

Display data, event data, and a trigger function can be used incombination. With this approach, display data with a slowsample rate can be used for continuous extended-period re-cording, and event data with a faster sample rate can be usedto record short-term details.

Other dataIn addition to measurement data, the MV100 can also save the following types of data:

• Manual sampling data: Instantaneous values (the 50 most recent measurements) occurring ateach contact input or key input are saved in ASCII format.

• Time-series (TLOG) calculation data: Maximum value, minimum value, integrated (totalized) value,etc. during fixed interval (with calculation option)

• Report data: Hourly reports, daily reports, weekly reports, monthly reports (with calculation option)• Settings data: Settings for set mode and setup mode

Display data

Trigger occurs Trigger occurs

Waiting for trigger

Waiting for trigger

Measured phenomenon

Event Event

MemoryMemory (for storing five months' worth of continuous data)

Page 7: Compact Data Logger with Color Monitor MobileCorder MV100 ...

8

The FTP client function makes it possible to make periodic,automatic transfers to a file server of data saved in the MV100’sinternal memory. A maximum of two servers (primary andsecondary) are supported, so files are automatically transferredto the secondary server if the primary server fails.

MV100 functions for the Internet

FTP client

Service request

Ethernet

FTP server

Downloadable files

Display data filesEvent data filesReport files

The FTP server allows a client computer to download datafiles stored on the MV100ís storage medium. Files can bedownloaded using DAQEXPLORER (sold separately).

The MV100 can easily be connected directly to a PC, evenwithout using general communication protocols such as GP-IB and RS-232-C.

You can create a simple network of PCs and MV100 units con-nected through a hub, even if you do not have an existingnetwork.

PCs and MV100 units can communicate with each in an exist-ing LAN environment. This makes it possible to monitor test-ing in a laboratory building from a remote office area.

If you want to exchange data between remote LANs (such asbetween a main-office LAN and a laboratory LAN), you canconnect them through a POTS line or leased line to form aWAN. POTS: Acronym for Plain Old Telephone Service. Basicdial telephone connections to the public switched network,without any added features or functions.

PC direct connection Simple network

LAN network connection POTS network connection

FTP client function FTP server function FTP client

Ethernet

Automaticallytransferable files

Display data filesEvent data filesReport files

Primary SecondaryFTP servers

Office area

Router

Server

Router

LaboratoryBuilding

PC

PC

Office area

Router

Server

Router

Laboratory Building

PC

PC

Dial-up router

Dial-up router POTS line, leased line,

Network capabilities through Ethernet

HUB

Communications(Network) Communications

Page 8: Compact Data Logger with Color Monitor MobileCorder MV100 ...

9

Data viewerThe data viewer can be used to convert file formats and play backdata files saved on the MV100 (event data, display data, TLOG datafiles), and data files transferred to a file server using a protocolsuch as FTP (event data, display data, TLOG data files). The fileconversion function lets you convert MV100 data files to ASCII for-mat, as well as the formats of off-the-shelf spreadsheet programssuch as Lotus 1-2-3 and MS-Excel.

MV standard software (for Windows 95/98/NT4.0)

DAQEXPLORER (for Windows 95/98/NT4.0) (sold separately)

DesktopDesktop is an MV100 management program that provides a Win-dows desktop-like operating environment. The MV100 units in thenetwork are displayed as icons on the desktop. The user can up-load or download data files and configuration files through simpleGUI actions (e.g., clicking or dragging and dropping icons).*

* The only type of data file transfer allowed is downloading from the MV100.

Data MonitorThis program can be used to monitor measurement data on as manyas 16 MV100 units distributed in a network.

Configuration softwareThe configuration software can be used to enter various MV100configurations either online or using a removable medium.

This software package is provided as a standard part of the MV100.

Application softwareApplication software

Page 9: Compact Data Logger with Color Monitor MobileCorder MV100 ...

10

Function-specific specificationsM Input unit

MV100 inputs MV102: 2 channelsMV104: 4 channelsMV106: 6 channelsMV112: 12 channels

Common standard specificationsGeneral specifications

M StructureExternal dimensions: Approximately 152 (W)u225 (H)u240 (D) mmWeight: Approximately 4 kg

M Input unitInput types: Floating unbalanced input, inter-channel isolation (However, a

common terminal is used for b terminals of RTDs.)Measurement intervals: MV102, MV104: 125 ms

MV106, MV112: 1 second (Measurement interval is 2 secondswhen the A/D integrating time is set to 100 ms.)

Input ranges, measuring ranges, and measurement/display accuracy:(reference operating conditions: 23 ±2°C; 55 ±10% RH; supplyvoltage: 90 to 132, 180 to 250 VAC; supply frequency: 50/60 Hz±1%; warmup time: 30 minutes or longer; performance under con-ditions not affected by equipment operations such as vibrations)

Input

DCV

TC

RTD*5

DI

Range/Type

20mV

60mV

200mV

2V

6V

20V

R*1

S*1

B*1

K*1

E*1

J*1

T*1

N*1

W*2

L*3

U*3

Pt100*4

JPt100*4

Voltage input

Contact input

Measuring range

-20.00 to 20.00 mV

-60.00 to 60.00 mV

-200.00 to 200.00 mV

-2.000 to 2.000 V

-6.000 to 6.000 V

-20.00 to 20.00 V

-200.0 to 600.0°C

-200.0 to 550.0°C

OFF: Less than 2.4 V

ON: 2.4 V or greater

Contact ON/OFF

0.0 to 1760.0°C

0.0 to 1760.0°C

0.0 to 1820.0°C

-200.0 to 1370.0°C

-200.0 to 800.0°C

-200.0 to 1100.0°C

-200.0 to 400.0°C

0.0 to 1300.0°C

0.0 to 2315.0°C

-200.0 to 900.0°C

-200.0 to 400.0°C

32 to 3200°F32 to 3200°F32 to 3200°F

-328 to 2498°F

-328.0 to 1472.0°F-328.0 to 2012.0°F-328.0 to 752.0°F

32 to 2372°F-328.0 to 4199°F

-328.0 to 1652.0°F-328.0 to 752.0°F

*1 R, S, B, K, E, J, T, N: IEC584-1 (1995), DIN IEC584, JIS C 1602-1995*2 W: W-5%, Rd/W-26%, Rd (Hoskins Mfg. Co.) ASTM E988*3 L: Fe-CuNi, DIN43710, U: Cu-CuNi, DIN43710*4 Pt100: JIS C 1604-1997, IEC751-1995, DIN IEC751-1996,

JPt100: JIS C 1604-1989, JIS C 1606-1989*5 Measuring current: i = 1 mA

Measurement accuracy (digital

display)

±(0.1% of rdg + 2 digits)

±(0.15% of rdg + 1°C)

R, S: 0 to 100°C, ±3.7°C;

100 to 300°C, ±1.5°C B: 400 to

600°C, ±2°C; if less than 400°C,

accuracy is not guaranteed.

±(0.15% of rdg + 0.7°C)

If -200 to -100°C, then

±(0.15% of rdg + 1°C)

±(0.15% of rdg + 0.5°C)

±(0.15% of rdg + 0.5°C)

If -200 to -100°C, then

±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 1°C)

±(0.15% of rdg + 0.5°C)

If -200 to 100°C, then

±(0.15% of rdg + 0.7°C)

±(0.15% of rdg + 0.3°C)

10 µV

10 µV

100 µV

1 mV

1 mV

10 mV

0.1°C

Digital displaymaximumresolution

Reference junction compensation (RJC):INT (internal)/EXT (external) switching possible

RJC accuracy: Type R, S, B, W: ±1°CType K, J, E, T, N, L, U: ±0.5°C (when measured at 0°C or higher)

Maximum input voltage: 2 VDC or lower voltage range and thermocouple: ±10 VDC(continuous)6 V, 20 VDC voltage range: ±30 VDC (continuous)

Input resistance: 2 VDC or lower voltage range and thermocouple: 10 MΩ orgreater6 V, 20 VDC voltage range: Approximately 1 MΩ

Input external resistance: DC voltage, thermocouple input: 2 kΩ or lessRTD input: 10Ω or less per line (equal on all three lines)

Input bias current: 10 nA or lessMaximum common mode noise voltage:

250 VAC rms (50/60 Hz)Common mode rejection ratio: 120 dB (50/60 Hz ±0.1%; 500Ω unbalanced; negative terminal

to ground)Normal mode rejection ratio: 40 dB (50/60 Hz±0.1%)Thermocouple burnout: Sensor ON/OFF switching possible

Burnout upscale/downscale switching possibleCalculation: Difference calculation: Difference calculation between any channels

Difference calculation range: DCV, TC, RTDLinear scaling: Scaling range: DCV, TC, RTD

Scalable value: -30000 to 30000Square root scaling: Scaling range: DCV

Scalable value: -30000-30000

M Display unitDisplay: 5.5-inch TFT color LCD (320u240 dots)

* The LCD may contain some pixels that are either always on oralways off. Due to the characteristics of liquid crystals, varia-tions in brightness may occur. Please note that such variationsdo not mean the display is broken.

Display colors: Trend and bar graph displays: 12 colorsBackground: White or black

Trend display: Direction: Vertical or horizontalNumber of windows: Switching between 4 (4 groups)Thickness: 1, 2, or 3 dotsWaveform update rate: 1, 2, 5, 10, 20, or 30 minutes, or 1,

2, or 4 hours (per div)Bar graph display: Direction: Vertical or horizontal

Number of windows: Switching between 4 (4 groups)Scale: Can be set in range of 4 to 12.Horizontal bar graph reference position:

End or centerUpdate rate: 1 second

Digital display: Update rate: 1 secondOverview display: Measurement values and alarm statuses on all channelsInformation display: Alarm summary, message summary, memory information, media

information, etc.Other displayed information: Memory status, scale values (0, 100%, display ON/OFF switch-

ing capability)Grid (number of divisions can be set between 4 and 12), andhours : minutesTime (year / month / date, hours : minutes : seconds), Trip line(thickness: 1, 2, or 3 dots), Messages (maximum 16 charac-ters, up to 8 types), alarm marks

Data reference function: Data can be played back from internal memory or a removablestorage medium.Display types: Split screen (divided in 2) or whole screenTime axis operations: Zoom-in/-out display, scrolling

M Storage functionsRemovable storage drive: A drive for the following types of media can be selected when

you place your order:I 3.5-inch floppy disk (2HD)I Zip diskI PCMCIA ATA flash memory card

Data saving method: Manual saving or auto-savingManual saving: Saves data when a removable storage medium is inserted.Auto-saving: Saving display data: Saves data to a removable storage medium

periodically (every 10 minutes to 31 days); Saving event data:Saves data to a removable storage medium periodically (every3 minutes to 31 days) (when trigger is not yet specified). Orsaves data when sampling period ends (when trigger isspecified).

Data saving intervals: Display data files: Interval varies according to the waveformupdate rate.Event data files: Sampling interval is specified.

Event data file sampling intervals:MV102, MV104: 125, 250, 500 ms; 1, 2, 5, 10, 30, 60, or 120secondsMV106, MV112: 1, 2, 5, 10, 30, 60, or 120 seconds

Measurement data files: The following two types of files can be created:(1) Event data files (to save instantaneous values sampled

at specified sampling intervals)(2) Display data files (to save maximum and minimum val-

ues occurring in display update interval in measurementdata sampled at measurement interval)

The two file types can be combined as follows:(1) Event data file (trigger only) plus display data file(2) Display data file only(3) Event data file onlyData format: Yokogawa standard format (binary format)

Per-channel data: Display data: Measurement data: 4 bytes per dataCalculation data: 8 bytes per data

Event data: Measurement data: 2 bytes per dataCalculation data: 4 bytes per data

Sampling time: Example sampling times (MV106, 6 measurement channels, 0calculation channels)

Display updating (min/div)

Saving interval (seconds)

Sampling time

1 minute

2 seconds

Approximately 27 hours

5 minutes

10 seconds

Approximately 5 days

20 minutes

40 seconds

Approximately 23 days

30 minutes

60 seconds

Approximately 34 days

60 minutes

120 seconds

Approximately 69 days

240 minutes

480 seconds

Approximately 277 days

Display data file only

Saving interval

Sampling time

1 second

Approximately 27 hours

5 seconds

Approximately 5 days

10 seconds

Approximately 11 days

30 seconds

Approximately 34 days

60 seconds

Approximately 69 days

120 seconds

Approximately 138 days

Event data file only

Display updating (min/div)

Saving interval (seconds)

Sampling time

1 minute

2 seconds

Approximately 20 hours

5 minutes

10 seconds

Approximately 4 days

20 minutes

40 seconds

Approximately 17 days

30 minutes

60 seconds

Approximately 26 days

60 minutes

120 seconds

Approximately 52 days

240 minutes

480 seconds

Approximately 208 days

Display data file plus event data fileDisplay data file

Saving interval

Sampling time

1 second

Approximately 6.9 hours

5 seconds

Approximately 34 hours

10 seconds

Approximately 2 days

30 seconds

Approximately 8 days

60 seconds

Approximately 17 days

120 seconds

Approximately 34 days

Event data file

Manual sampling data: Storage trigger: Key input or contact inputData format: ASCII formatMaximum stored data: 50 data

Page 10: Compact Data Logger with Color Monitor MobileCorder MV100 ...

11

Microsoft, MS, and Windows are registered trademarks of Microsoft Corporation, USA.Lotus and 1-2-3 are registered trademarks of Lotus Development Corporation.MMX and Pentium are registered trademarks of Intel Corporation, USA.Ethernet is a registered trademarks of Xerox Corporation.Zip and logo marks thereof are registered trademarks or trademarks of Iomega Corporation, USA.Other company names and product names mentioned in this document are the registered trademarks ortrademarks of their respective companies.

Unit: mm

151.7 187.7 52

29.8

180

14.6144

TLOG data (with calculation option only):Time series integrated (totalized) value, maximum value, mini-mum value, average value, max-min valueStorage trigger: Data saved when TLOG time is up.

Report data (with calculation option only):Periodic average value, maximum value, minimum value, andintegrated (totalized) value.Types: Hourly reports, daily reports, daily + weekly reports, daily+ monthly reportsData format: ASCII

Screen copying function: Copying method: Key inputData format: PNGOutput to: Removable storage medium or online output

M Trigger functionsEvent data file: Select FREE, TRIG, or ROTATE mode.Display data + event data file: Select TRIG or ROTATE mode.Trigger source: Key input, remote control (optional), alarmPretrigger: Works with event data. 0, 5, 25, 50, 75, 95, or 100%

M Alarm functionsMaximum number: A maximum of four alarms can be set on each channel.Alarm types: High-low limits, High-low difference limits, rate-of-change in-

crease/decrease limitsRate-of-change alarm time interval:

Measurement intervalu1 to 15Display: Status (alarm type) and common alarm display in digital dis-

play area when alarm occursHold/no hold switching capability

Hysteresis: ON (0.5% of display span)/OFF switching (common to all chan-nels/levels)

Outputs: Number of points: 2, 4, 6Operation excitation/no excitation, hold/no hold switching ca-pability

Storage: Stored information: Alarm occurrence/clear time, alarm typeNumber of saved items: Maximum 120 (most recent)

M Communication functionsNetwork type: Ethernet (10BASE-T)Basic protocol: TCP/IPFile transfer function: Automatic transfer from MV100 (FTP client protocol)

File transfer in response to request from host computer (FTPserver protocol)

Real time monitor function: Real time online monitoring of MV100 measurement data (pro-prietary protocol)

Transferable files: Display data files, event data files, report data, and screenshotdata

FTP server functions: Directory operations on a removable storage medium, fileoutput, file deletion, and information on available memory spacein a storage medium

M Power supply unitI AC power supplyRated supply voltage: 100 to 240 VAC (automatic switching)Operating supply voltage range: 90 to 132, 180 to 250 VACRated supply frequency: 50/60 Hz (automatic switching)I DC power supplyRated supply voltage: 12 VDCOperating supply voltage range: 10 to 18 VDC

MV100 power consumptionI AC power supply

Supply voltage

100 VAC

240 VAC

With LCD saver on

Approximately 25 VA

Approximately 35 VA

Normal use

Approximately 30 VA

Approximately 40 VA

Maximum

Approximately 55 VA

Approximately 70 VA

I DC power supply: Approximately 30 VA

Normal operating conditionsSupply voltage: AC power supply: 90 to 132, 180 to 250 VAC

DC power supply: 10 to 18 VDCSupply frequency: 50 Hz ±2%, 60 Hz ±2%Ambient temperature: 5 to 40°CAmbient humidity: 20 to 80% RH (at 5 to 40°C)

Optional specificationsM Alarm output relay contacts (/A1, /A2, /A3)

Relay output from back side when alarm occurs.Number of outputs: 2, 4, or 6Relay contact capacitance: 250 VDC/0.1 A (resistance load), 250 VAC (50/60 Hz)/3 AOutput form: NO-C-NC (excitation/no excitation, AND/OR, hold/no hold

switching capability)M Serial communications (/C2, /C3)

Host computer remote control, setting, and data output to host capabilityInterface type: EIA RS-232 (/C2) or RS-422-A/485 (4-wire) (/C3) complianceProtocol: Proprietary protocolSynchronization method: Start-stopCommunication type (RS-422-A/485):

4-wire half-duplex multidrop connection (1:N (where N is 1 to 31))Transfer rate: 1200, 2400, 4800, 9600, 19200, 38400 bpsData length: 7 or 8 bitsStop bit: 1 bitParity: ODD, EVEN, NONEMaximum distance (RS-422-A/485): 500 metersCommunication modes: ASCII mode for control and settings I/O. ASCII or binary mode

for measurement data output.M Fail/memory end output (/F1)

Relay output from back side before start time specified for display data file overwriting orwhen system abnormality occurs (1, 2, 5, 10, 20, 50, or 100 hours can be specified).Relay contact capacitance: 250 VDC/0.1 A (resistance load), 250 VAC (50/60 Hz)/3 A

M Screw input terminals (/H3)The input terminals are screw tightened input terminals.

M Mathematical calculation functions (/M1)The MV100 is capable of the following calculations, as well as calculation channel trends/digital displaying and recording.Calculation channels: MV102, MV104: 4 channels

MV106, MV112: 12 channelsCalculation types: General calculations: Addition, subtraction, multiplication, divi-

sion, square root, absolute value, common logarithm, exponent,power, relationships (<, >, =, ≠), logical calculations (AND, OR,NOT, XOR)

Statistical calculations: Time series data average, maximum, minimum, and integrated(totalized) values

Constants: 12Communication digital input: Up to 12 (data) communication digital inputs are allowed. Can

be used for calculation equations other than statistics.Remote input: Up to eight remote inputs are allowed. Remote status (0/1) can

be used in calculation equations.Report functions: Report types: Hourly reports, daily reports, daily + weekly re-

ports, daily + monthly reportsCalculation types: Average, maximum, minimum, and integrated (totalized) valuesData format: ASCII

M Remote control (/R1)The following remote control operations are possible through contact input (up to eight canbe set).I Memory start/stop (level)I Event data file external trigger input (trigger, 250 ms or greater)I Time adjustment (adjusts time to reference time using contact; trigger, 250 ms or greater)I Calculation start/stop (level)I Calculation data reset (trigger, 250 ms or greater)I Manual sampling (trigger, 250 ms or greater)I Message writing (as many as 8 can be set; trigger, 250 ms or greater)I Load settings (as many as 3 can be set; trigger, 250 ms or greater)I Alarm ACK (trigger, 250 ms or greater)

Application softwareM MV standard software (part of standard MV100 package)

System requirementsOS: Microsoft Windows 95/98/NT4.0

Processor: MMX Pentium 166 MHz or higher (Pentium II 266 MHz or higherrecommended)

RAM: 32 MB or more (64 MB recommended)Disk drive: Floppy disk driveFree hard drive space: 10 MB or more (100 MB recommended)Display card: Display card capable of displaying 32,000 colors or more (64,000

or more recommended) and compatible with Windows 95/98/NT4.0

Printer: Printer and printer driver compatible with Windows 95/98/NT4.0Main functions (package): Hardware configurations (online or using a removable storage

medium)Data viewer (waveform playback)Printout of playback dataFile conversion (to ASCII, Lotus 1-2-3, and MS-Excel formats)

M DAQEXPLORER (sold separately)System requirementsOS: Microsoft Windows 95/98/NT4.0

Processor: MMX Pentium 166 MHz or higher (Pentium II 266 MHz or higherrecommended)

RAM: 64 MB or more (128 MB recommended)Disk drive: Floppy disk driveFree hard drive space: 10 MB or more (100 MB recommended)Display card: Display card capable of displaying 32,000 colors or more (64,000

or more recommended) and compatible with Windows 95/98/NT4.0

Printer: Printer and printer driver compatible with Windows 95/98/NT4.0Main functions (package): Desktop (file transfers, configurations, etc. using operations on

desktop)Data monitoringHardware configurations (online or using a removable storagemedium)Data viewerPrintout of playback dataFile conversion (to ASCII, Lotus 1-2-3, and MS-Excel formats)

External view

SpecificationsSpecifications

Page 11: Compact Data Logger with Color Monitor MobileCorder MV100 ...

MM-06E

Represented by :YOKOGAWA ELECTRIC CORPORATIONMeasurement Sales Dept.9-32, Nakacho 2-chome, Musashino-shi, Tokyo 180-8750, JAPANPhone: 81-422-52-6614, Fax: 81-422-52-6624YOKOGAWA CORPORATION OF AMERICA2 Dart Road, Newnan, Georgia 30265, U.S.A.Phone: 770-253-7000, Fax: 770-251-2088YOKOGAWA EUROPE B.V.Vanadiumweg 11, 3812 PX Amersfoort, THE NETHERLANDSPhone: 31-33-4-641611, Fax: 31-33-4-631202YOKOGAWA ENGINEERING ASIA PTE. LTD.11 Tampines Street 92, Singapore 1852, SINGAPOREPhone: 65-783-9537, Fax: 65-786-6650

YOKOGAWA ELECTRIC CORPORATIONTest & Measurement Business Division155 Takamuro-cho, Kofu-shi, Yamanashi-ken, 400-8558 JapanPhone: 81-552-43-0310, Fax: 81-552-43-0396

Subject to change without notice. [Ed : 01/b] Printed in Japan, 001(YG)All Rights Reserved, Copyright© 2000, Yokogawa Electric Corporation.

MV100

Model

MV102

MV104

MV106

MV112

Removable storage drive/slot

Display/softwarelanguage

Power supply

Power inlet, power cord

Options

Suffix Code

-1

-2

-3

-2

-1

-2

DFRS

Option

Code

/A1

/A2

/A3

/C2

/C3

/F1

/H3

/M1

/R1

Description

MobileCorder MV100 (2 channels)

MobileCorder MV100 (4 channels)

MobileCorder MV100 (6 channels)

MobileCorder MV100 (12 channels)

Floppy disk drive

Zip drive (one 100-MB Zip disk included)

PCMCIA ATA flash memory card slot (one 20-MB card included)

English ,German and French, deg F/summer & winter time(English MV standard software included)

100 or 240 VAC

12 VDC*1

3-pin power inlet with UL/CSA cable3-pin power inlet with VDE cable3-pin power inlet with SAA cable3-pin power inlet with BS cable

Alarm relay contact output : 2 points*2

Alarm relay contact output : 4 points*2

Alarm relay contact output : 6 points*2*3

RS-232 interface*4

RS-422-A/485 interface*4

Fail/memory end output relay*3

Screw input terminals

Mathematical function (including report function)

Remote control *1 An AC adapter is included as a standard accessory.*2 Only one from the /A1, /A2, and /A3 options can be specified.*3 The /A3 and /F1 options cannot be specified at the same time.*4 The /C2 and /C3 options cannot be specified at the same time.

• Compliance with safety standards/EMC standardsThe MV100 has satisfied the strict require-ments for international safety standards andElectromagnetic Compatibility (EMC) stan-dards. Compliance with these standards isproof of the MV100's high level of reliability.Of course, the MV100 also carries the CEmark, which is required in Europe.

Safety standards: Passed CSA22.2 No. 1010.1; compliance withEN61010-1Installation category (overvoltage category) II*1Contamination level 2*2

*1 The installation category (overvoltage category) is a number that defines the transientovervoltage, and includes the impulse withstand voltage specification. Category II appliesto electrical devices that are powered by fixed facilities such as distribution boards.

*2 The contamination level relates to the level of adherence of solids, liquids, and gases,which lowers the surface resistivity or withstand voltage. Contamination level 2 appliesonly to the ordinary indoor environment (non-conductive contamination).

EMC standard: EN61326-1 compliance

• Input isolation and channel isolationThe MV100's inputs andchannels are isolated forDC voltage and thermo-couple inputs.Isolation protects theMV100 from unexpectedhigh-voltage noise.

Highly reliable hardware

NOTICE Before operating the product, read the instruction manual thoroughly for

proper and safe operation. If this product is for use with a system requiring safeguards that directly

involve personnel safety, please contact the Yokogawa sales offices.

Model and Suffix CodesApplication Software

Model

MV10A

DXA200

Language

Suffix Code

-02

OS

Windows 95/98/NT

Windows 95/98/NT

Description

MV standard software (standard)

DAQEXPLORER (sold separately)

English version

Accessories (sold separately)

Model

790501

790502

705900

A1053MP

A1134UN

Description

Soft carrying case

Front cover

3.5-inch 2HD floppy disks (10 disks)

Zip disk (100 MB)

PCMCIA ATA flash memory card (20 MB)

Application Software

Accessories