Chromatographic Techniques WORD
Transcript of Chromatographic Techniques WORD
-
8/10/2019 Chromatographic Techniques WORD
1/56
Chromatographic Techniques
General principles
Distribution coefcients
The basis of all forms of chromatography is the partition or distribution coefficients (Kd),
which describes the way in which a compound distributes itself between immiscible
phases. For two such immiscible phases A and B, the value for coefficient is a constant
at a given temperature and is given by the epression
The term effective distribution coefficient is defined as the total amount! distinct from the
concentration, of substance present in "one phase divided total amount present in the
other phase. #t is in fact the distribution coeffient multiplied by the ratio of the volumes of
the two phases present.
Basically, all chromatographic systems consist
stationary phase, may be a solid, gel, li$uid or a solid%li$uid miture that is
immobili&ed,
mobile phase, which may be li$uid or gaseous and which flows over the
stationary phase.
Modes o chromatography
chromatographic separations may be achieved using three contrasting
modes:
Column chromatographyin which the stationary phase attached to a
suitable matrix (inert, insoluble support) is packed into a glass or metal
column and the mobile phase passed through the column either by gravity
feed or applied gas pressure.
Thin layer chromatography in which the stationary phase attached to a
suitable matrix is coated thinly onto a glass, plastic or metal foil plate. The mobile
liquid passes across the thin-layer plate, held either horiontally or vertically, by
capillary action. This mode of chromatography has the practical advantage over in
chromatography that a large number of samples can be studied simultaneously
Paper chromatography
-
8/10/2019 Chromatographic Techniques WORD
2/56
stationary liquid phase is supported by the cellulose !bres of a paper
sheet. "s in thin-layer chromatography, with which paper
chromatography has several similarities
the mobile phase passes along the paper sheet either by gravity feed
or by capillary action. This is one of the older forms of chromatographyand, although it is still used to demonstrate the principles of
chromatography,
Perormance o column chromatography
The principle of a column chromatographic separation may be depicted by
considering a column packed with a solid granular stationary phase to aheight of # cm, surrounded by the mobile liquid phase of which there is $ cm%
per cm of column, as shown.
&f %' g of a compound is added to the column in $ cm% of mobile phase,
then as this $ cm% moves on to the column to occupy position ", $ cm% of
mobile phase will leave the base of the column.
&f the compound added has an eective distribution coe*cient of $, it will
distribute itself equally between the solid and liquid phases (stage $).
&f a further $ cm% of mobile phase is introduced on to the column, the mobile
phase in section " will move down to +, taking $ g of the compound with it,
leaving $ g at " (stage ').
"t both " and + a redistribution of the compound will occur so that there is
g in the mobile phase and g in the solid phase.
-
8/10/2019 Chromatographic Techniques WORD
3/56
The addition of a further $ cm% of mobile phase to the column displaces the
mobile phase in " to + and that in + to giving the distribution of the
compound as shown in stage %(/$/).
"ddition of a further $ cm% of mobile phase leads to the distribution shown at
stage 0 (0/$'/$'/0), and a further $ cm% of mobile phase leads to thedistribution shown at stage # ('//$'/$'//').
&t is apparent that after a relatively small number of equilibrations the
compound distributes itself symmetrically within a band.
Column chromatographic components (gas or liquid mobile phase)
-
8/10/2019 Chromatographic Techniques WORD
4/56
1 " stationary phase2 hosen to be appropriate for the analytes to be
separated.
1 " column2 !lled with the matrix coated with the stationary phase, or
microbore type, which the stationary phase is coated directly on the inside
wall.
mobile phase and delivery system: hosen to complement the
stationary phase and to deliver a constant rate of 3ow into the column.
! n in"ector system2 To deliver test samples to the top of the column in a
reproducible manner.
! detector and chart recorder:To give a continuous record of the
presence of the analytes in the eluent as it emerges from the column.
4etection based on the measurement of parameter -visible or
ultraviolet absorption or 3uorescence. " peak on the chart recorder
represents each separated analyte.
! raction collector: 5or collecting the separated analytes for further
biochemical studies.
Column liquid chromatography can be subdivided according to
the bac# pressure generated $ithin the column during the
separation process.
-
8/10/2019 Chromatographic Techniques WORD
5/56
%o$ pressure liquid chromatography &%P%C' generates pressures of less
than # bar ($ bar 6 $0.# lbf 7n8' 6 9.$ :;a), since there is little resistance to
eluent 3ow owing to the physical nature of the stationary phase. ;= are often blurred and
their equipment and procedures are virtually identical.
+oth give excellent resolutions and hence the term high performance liquid
chromatography is preferred for both of them,
)election o stationary and mobile phases
)uccessul chromatographic separations depend upon the correct
choice stationary and mobile phases. This may be achieved by
setting up one o the ollo$ing:
-
8/10/2019 Chromatographic Techniques WORD
6/56
dsorption equilibrium:This is between a stationary solid phase and a
mobile liquid phase (adsorption chromatography? hydrophobic interaction
chromatography).
Partition equilibrium:This is between a stationary liquid phase and a
mobile liquid or gas phase (partition chromatography? perfusionchromatography? pair chromatography? chiral chromatography? gasliquid
chromatography).
*on+e,change equilibrium2 This is between a stationary, solid ion-
exchanger and mobile, liquid electrolyte phase (ion-exchange
chromatography? chromatofocusing? membrane chromatography).
-,clusion equilibrium:This is between a liquid phase trapped inside the of
a stationary porous structure and the same mobile liquid phase (molecular
exclusion or gel !ltration).
! inding equilibrium:This is between a stationary immobilised ligand and
a mobile liquid phase (a*nity chromatography? immunoa*nity
chromatography? lectin a*nity chromatography? metal chelate a*nity
chromatography? dye-ligand chromatography? covalent chromatography).
&n practice it is quite common for two or more of these equilibria to be
involved simultaneously in a particular chromatographic separation.
nalyte development and elution
-
8/10/2019 Chromatographic Techniques WORD
7/56
@eparation of the mixture of analytes applied to the stationary phase by the
mobile phase and their elution from the column.
olumn chromatographic techniques can be subdivided on the basis of thedevelopment and elution modes.
sample is dissolved in a suitable solvent and applied to the stationary phase
as a narrow? discrete band.
:obile phase, normally consisting of an organic solvent or a mixture of
solvents often incorporating a buered aqueous system, is then allowed to
3ow continuously over the stationary phase, resulting in the progressive
separation and elution of the sample analytes.
&f the composition of the mobile phase is constant, the process is said to be
isocratic mobile phase may be gradually changed, for example with respect
to salt concentration or polarity - referred to as gradient elution. The
composition of the mobile phase may be changed continuously or in a
stepwise manner.
@uccessful onal development results in the elution of pure samples of all the
analytes.
&n displacement or a*nity development, the analytes in the sample are
separated on the basis of their a*nity for the stationary phase,
The analytes are selectively eluted by using a mobile phase containing aspeci!c solute that has a higher a*nity for the stationary phase than have
the analytes in the sample.
Thus, as the mobile phase is added, this agent displaces the analytes from
the stationary phase in a competitive fashion, resulting in their repetitive
binding and displacement along the stationary phase and eventual elution
from the column in the order of their a*nity for the stationary phase,
-
8/10/2019 Chromatographic Techniques WORD
8/56
The one with the lowest a*nity being eluted !rst.
&n frontal development, the sample is continuously added to the stationary
phase, thereby forcing the analytes along the stationary phase in the order of
their a*nity for it. The analyte with the lowest a*nity accumulates at the
front of the moving sample band
&n practice, the technique is eectively restricted to the analysis of a single
trace impurity in an otherwise pure sample.
C(/0MT0G/P(*C P-/10/M2C- P/M-T-/)
/etention time and elution volume
The !rst is the time it takes the analyte molecules to pass through the free spaces
between the particles of the matrix coated with the stationary phase. This time is
referred to as dead time, (t:)
Aolume of the free space is referred to as the column void volume, (Ao)
The value of t: will be the same for all analytes and can be measured by
using an analyte that does not interact with the stationary phase but simply
spends all of the elution time in the mobile phase travelling through the void
volume.
The second component is the time the stationary phase retains the analyte
referred to as the ad7ustered retention time. This time is characteristic of the
analyte and is the dierence between the observed retention time and the
dead time2
Capacity actor '
Bne of the most important parameters in chromatography
"lso called retention factor and capcity ratio
-
8/10/2019 Chromatographic Techniques WORD
9/56
&t is simply the additional time that the analyte takes to elute from the
column relative to an unretained or excluded analyte that does not interact
with the stationary phase and which, by de!nition, has a k8 value of 9. Thus2
where :s is the mass of analyte in the stationary phase, :m is the mass of
analyte in the mobile phase, As is the volume of stationary phase, and Am is
the volume of mobile phase.
Plate (eight and /esolution
-
8/10/2019 Chromatographic Techniques WORD
10/56
Plate height
Chromatography columns + consist of a number of ad7acent ones in each
of which there is su*cient space for an analyte to completely equilibrate
between the two phases. Cach one is called a theoretical plate.
The length of the column containing one theoretical plate is referred to as the
plate height >. The numerical value of both D and > for a particular column is
expressed by reference to a particular analyte.
;late height is simply related to the width of the analyte peak
Pea# roadening
" number of processes oppose the formation of a narrow analyte peak
thereby increasing the plate height2
! pplication o the sample to the column: &t takes a !nite time to apply
the analyte mixture to the column, so that the part of the sample applied !rstwill already be moving along the column by the time the !nal part is applied.
The part of the sample applied !rst will elute at the front of the peak.
! %ongitudinal di4usion. 5ick8s law of diusion states that an analyte will
diuse from a region of high concentration to one of low concentration at a
rate determined by the concentration gradient between the two regions and
the diusion coe*cient of the analyte. Thus the analyte within a narrow band
-
8/10/2019 Chromatographic Techniques WORD
11/56
will tend to diuse outwards from the centre of the band, resulting in band
broadening.
! Multiple path$ays2 The random packing of the particles in the column
results in many routes between the particles for both mobile phase and
analytes. The smaller the particle sie the less serious is this problem.
! -quilibration time bet$een the t$o phases: &t takes a !nite time for
each analyte in the test sample to equilibrate between the stationary and
mobile phases as it passes down the column
symmetric Pea#
&n some chromatographic separations, ideal shaped peaks are not obtained.
&n cases where there is a gradual rise at the front of the peak and a sharp fall
after the peak, the phenomenon is known as fronting ( because ofoverloading the column) - reducing the amount of mixture applied to the
column resolves the problem.
&n cases where the rise in the peak is normal but the tail is protracted, the
phenomenon is known as tailing. (because -retention of analyte by a few
active sites on the stationary phase, commonly on the inert support matrix.
@uch sites strongly adsorb molecules of the analyte and only slowly release
them).
This problem can be overcome by chemically removing the sites, frequently
hydroxyl groups, by treating the matrix with a silanising reagent such ashexamethyldisilaine.This process is sometimes referred to as capping.
/esolution
-
8/10/2019 Chromatographic Techniques WORD
12/56
The success of a chromatographic separation is 7udged by the ability of the
system to resolve one analyte peak from another.
Eesolution (Es) is de!ned as the ratio of the dierence in retention timebetween the two peaks to the mean of their base widths.
)ample preparation
)olvent e,traction
;reliminary lean up is essential, particularly if the test analyte(s) is in a
complex matrix such as plasma, urine, cell homogenate or microbiological
culture medium (especially in >;=)
The most common extraction, This is based on the extraction of the analytes
from aqueous mixtures using a low boiling water-immiscible solvent such as
diethylether or dichloromethane.
This solvent extraction procedure tends lack selectivity and is often
unsatisfactory for the >;= analysis of compounds in the range of ng cm-% or
less.
-
8/10/2019 Chromatographic Techniques WORD
13/56
)olid+phase e,traction
&ts advantage over simple solvent extraction is that it exhibits greater
selectivity, mainly because it is a form of chromatography.
The test solution is passed through a small (few millimetres in length)
disposable column (cartridge) packed with relatively large particles of a
bonded silica similar to those used for >;=.
These selectively adsorb the analyte(s) under investigation and ideally allow
interfering compounds to pass through.
@everal commercial forms of this solid-phase extraction technique are
available that facilitate the simultaneous treatment of a large number of test
samples.
Column s$itching
@uited to the analysis of analytes in very low concentartions in complex
mixtures especially in >;=
The test solution is applied to a preliminary short column similar to the type
used in solid-phase extraction.
Bnce the test analyte has been adsorbed and impurities washed through the
column,
The analyte is eluted with a suitable organic solvent and the column eluate
transferred directly to an analytical >;= column.
)upercritical 5uid e,traction
@5C exploits the fact that gases such as carbon dioxide exist as a liquid under
certain critical conditions .
&n the case of carbon dioxide, these conditions are %$.$F and G.% :;a and
the resultant liquid carbon dioxide can be used as the extraction solvent,
behaving as a low polarity solvent comparable to hexane.
+y altering the physical conditions of the extract, the carbon dioxide can be
made to revert to a gas, thus simplifying the recovery of the extractedanalytes.
)ample Deriviti6ation
@ome functional groups, especially hydroxyl, present in a test analyte may
compromise the quality of its behavior in a chromatographic system.
-
8/10/2019 Chromatographic Techniques WORD
14/56
The technique of analyte pre- or post-column derivatisation may facilitate
better chromatographic separation and detection by masking these
functional groups.
Column Chromatography
Columns
The glass column used should have a means of supporting the stationary
phase as near to the base of the column as possible in order to minimise the
dead space below the column support in which post-column mixing of
separated analytes could occur.
ommercial columns possess either a porous glass plate fused on to the base
of the column or a suitable device for supporting a replaceable nylon net,which in turn supports the stationary phase.
" cheaper alternative is to use a small plug of glass wool together with a
minimal amount of quart sand or glass beads.
Matri, materials
-
8/10/2019 Chromatographic Techniques WORD
15/56
The matrix is the material used to support the stationary phase. The selection
of a matrix for a particular stationary phase is vital to the successful
chromatographic use of the phase.
Generally spea#ing7 a matri, needs to have:
>igh mechanical stability to encourage good 3ow rates and to
minimise pressure drop along the column?
igh capacity, i.e. density of functional groups to minimise bed
volume.
&t also needs to be available in a range of particle sies.
&n addition some forms of chromatography require a matrix with aporous structure, in which case the pores need to be of the correct sie
and shape.
5inally, the surface of a matrix needs to be inert to minimie the non-
selective adsorption of analytes.
Commonly used types o matrices
garose - a polysaccharide made up of 4-galactose and %,-anhydro- $
-galactose units. The unbranched polysaccharide chains are cross-linked with
agents such as ',%-dibromopropanol to give gels that are stable in the p>
range %-$0.)epharose and io+Gel .
Cellulose - a polysaccharide of +eta $-0-linked glucose units. 5or matrix use
it is cross-linked with epichlorohydrin, the extent of cross-linking dictating the
pore sie. &t is available in bead, microgranular and !brous forms, has good
p> stability and 3ow properties, and is highly hydrophilic.
De,tran - a polysaccharide consisting of alpha-$--linked glucose units. 5or
matrix use it is cross-linked with epichlorohydrin but is less stable to acid
hydrolysis than are cellulose matrices. &t is stable up to p> $' and is
hydrophilic. )ephade,.
Polyacrylamide - a polymer of acrylamide cross-linked with D,D8methylene-
bisacrylarnide &t is stable in the p> range '-$$.+io- ranges and are most
commonly used for exclusion and ion-exchange chromatography. They have
relatively low hydrophilicities.
-
8/10/2019 Chromatographic Techniques WORD
16/56
)ilica - " polymeric material produced from orthosilic acid. The numerous
silanol (@i-B>) groups make it hydrophilic. Hhen derivatised, excess silanol
groups can be removed by treatment with trichloromethylsilane. p> range %-
. &t is chemically inert but, like the silicas, tends to dissolve above p>.
)tationary phases
The chemical nature of the stationary phase depends upon the particular
form of chromatography to be carried out.
:ost stationary phases are available attached to the matrices in a range of
sies and shapes.
+oth properties are important because they in3uence the 3ow rate and
resolution characteristics.
The larger the particle, the faster the 3ow rate but, conversely, the smaller
the particle the larger the surface area-to-volume ratio and potentially the
greater their resolving power.
The best packing characteristics are given by spherical particles
;article sie is commonly expressed by a mesh sie, which is a measure of
the openings per inch in a sieve? hence the larger the mesh sie, the smaller
is the particle.
-
8/10/2019 Chromatographic Techniques WORD
17/56
$99-$'9 mesh is most common for routine use, whereas a '99-099 mesh is
used for higher resolution work.
Column pac#ing
:ost critical factors in achieving a successful separation by any form of
column chromatography.
;acking a column is normally carried out by gently pouring a slurry of the
stationary phase in the mobile phase into a column that has its outlet closed,
Cnsure that no air bubbles are trapped and that the packing settles evenly.
-
8/10/2019 Chromatographic Techniques WORD
18/56
;oor column packing gives rise to uneven 3ow (channeling) and reduced
resolution.
The slurry is added until the required height is obtained, Bnce the required
column height has been obtained, the 3ow of mobile phase through the
packed column is started by opening the outlet, and continued until thepacking has completely settled.
To prevent the surface of the packed material from being disturbed either by
the addition of mobile phase to the column or during the application of the
sample to the column, it is normal to place a suitable protection device, such
as a !lter paper disc or nylon gaue, on the surface of the column. @ome
commercial columns possess an adaptor and plunger, which serve the dual
purpose of protecting the surface of the column and providing an inlet.
Do part of it should be allowed to run dry? hence a layer of mobile phase
should always be maintained above the column surface.
&t is di*cult to generalie about the ideal column height-to-diameter ratio and
the total bed volume. 5or example, in exclusion chromatography, a height-to-
diameter ratio of $92 $ to '92 $ is normally suitable.
pplication o sample
)everal methods are available
" simple way is to remove most of the mobile phase from above the column
by suction and 7ust to drain the remainder into the column bed. The sample is
then carefully applied by pipette and it too is allowed 7ust to run into thecolumn. " small volume of mobile phase is then applied in a similar manner
to wash !nal traces of the sample into the bed. :ore mobile phase is then
carefully added to the column to a height of '-# cm. The column is then
connected to a suitable reservoir that contains more mobile phase so that the
height of the phase in the column can be maintained at '-# cm.
"n alternative procedure, which avoids the necessity to drain the column to
the surface of the bed, is to increase the density of the sample by addition of
sucrose to a concentration of about $I. Hhen this solution is layered on to
the liquid above the column bed, it will automatically @ink to the surface of
the column and hence be quickly passed into the column.This method of
sample application is satisfactory, provided that the presence of @ucrose in
no way interferes with the separation and subsequent analysis of the sample.
J
" third method involves the use of capillary tubing and/or syringe or
;eristaltic pump to pass the sample directly to the column surface. This latter
method is the most satisfactory of the three.
-
8/10/2019 Chromatographic Techniques WORD
19/56
&n all cases, care must be taken to avoid overloading the column with sample,
Btherwise irregular separation will occur. &t is also advantageous to apply the
sample in as small a volume of mobile phase.
Column development and sample elution
The components of the applied sample are separated by the continuous
passage of the mobile phase through the column. This is known as elution
development.
4uring the elution process it is essential that the 3ow of mobile phase is
maintained at a stable rate and this is most simply achieved by gravity feed.
"n alternative and Kmore eective method of maintaining stable 3ow rates is
to use a peristaltic pump.
olumn development using a single liquid as the mobile phase is known as an
isocratic elution. >owever, in many cases in order to increase the resolvingpower of the mobile phase, it is necessary continuously to change its p>,
ionic concentration or polarity. This is known as gradient elution.
Detectors and raction collection
"s the resolved analytes emerge in the eLuent from the column it is
necessary detect their presence.
5or coloured analytes this can be achieved simply by visual
observation but for colourless compounds alternatives are necessary.
4etection may be based on ultraviolet absorption, 3uorescence spectroscopy,
changes in refractive index of the eLuent, the presence of a radioactive
emission atom or the ease of oxidation or reduction of the analytes as
measured by an electrochemical detector.
Mltraviolet detectors are probably the most common form of detector for
biochemical analysis.
5luorescent detectors are also available.
*on+e,change chromatography
This form of chromatography relies on the attraction between oppositely
charged particles.
-
8/10/2019 Chromatographic Techniques WORD
20/56
:any biological materials, e.g. amino acids and proteins, have ionisable
groups and the fact that they may carry a net positive or negative charge can
be utilied in separating mixtures of such compounds.
The net charge exhibited by these compounds is dependent on their (pNa and
on the p> of the solution.
&on-exchange separations are carried out mainly in columns packed with an
ion exchanger.
There are two types of ion-exchanger, mainly cation and anion exchangers.
The cation exchangers possess negatively charged groups and these will
attract positively charged cations.
These exchangers are also called acidic ion exchangers because their
negative charges result from ioniation of acidic groups.
"nion exchangers have positively charged groups that will attract negatively
charged anions. The term basic ion-exchange materials is also used to
describe these exchangers, from the association of protons with basic groups.
The ion+e,change mechanism is thought to be composed o 8ve
distinct steps:
(i) 4iusion of the ion to the exchanger surface. This occurs very quickly in
homogeneous solutions.
(ii) 4iusion of the ion through the matrix structure of the exchanger tothe
exchange site. This is dependent upon the degree of cross-linkage of the
exchanger and the concentration of the solution. This process is thought to
be the feature that controls the rate of the whole ion- exchange process.
(iii) Cxchange of ions at the exchange site. This is thought to occur
instantaneously and to be an equilibrium process.The more highly charged
the molecule to be exchanged, the tighter it binds to the exchanger and the
less readily it is displaced by other ions
-
8/10/2019 Chromatographic Techniques WORD
21/56
(iv) 4iusion of the exchanged ion through the exchanger to the surface.
(v) @elective desorption by the eluent and diusion of the molecule into th
external eluent.
The selective desorption of the bound ion is achieved by changes in p>
and/or ionic concentration or by a*nity elution, in which case an ion that has
greater a*nity for the exchanger than the bound ion introduced into the
system.
Materials and applications
%o$ pressure ion+e,change chromatography can be carried out using
a variety o matrices and ionic groups.
:atrices used include polystyrene, cellulose and agarose.
5unctional ionic groups include sulphonate (-@B0) and quaternary ammonium(-DE%), both of which are strong exchangers because they are totally ionised
at all normal working p> values, and
arboxylate (-BB ) and diethylammonium (->D(>'>%)'), both of which
are termed weak exchangers because they are ionised over only a narrow
range of p> values.
"ll exchangers are characteried by a total exchange capacity, which is
de!ned as the number of milli equivalents of exchangeable ions available,
either gram of dried exchanger or per unit volume of hydrated resin.
ompounds that are stable over a wide range of p> may be separated by
either type of exchanger.
The choice between a strong and weak exchanger all depends on sample
stability and the eect of p> on sample charge.
Heak electrolytes requiring a very low or high p> for ionisation can be
separated only on strong exchangers, as only they operate over a wide p>
range.
&n contrast, for strong electrolytes, weak exchangers are advantageous for a
number of reasons including a reduced tendency to cause sampledenaturation, their inability to bind weakly charged impurities and their
enhanced elution characteristics.
The p> of the buer used should be at least one p> unit above or below the
isoionic point of the compounds being separated.
-
8/10/2019 Chromatographic Techniques WORD
22/56
&n general, cationic buers such as Tris, pyridine and alkylamine are used in
con7unction with anion exchanger and anionic buers such as acetate,
barbiturate and phosphate are used with cationic exchangers.
The precise initial buer p> and ionic strength should be such as 7ust to
allow the binding of the sample components to the exchanger.
Cqually, a buer of the lowest ionic strength that eects elution should
initially be used for the subsequent elution of the components.
This ensures that, initially, the minimum number of undesired substancesbind to the exchanger.
The amount of sample that can be applied to a column is dependent upon the
sie of the column and the capacity of the exchanger.
-
8/10/2019 Chromatographic Techniques WORD
23/56
&f, however gradient elution is to be used, the initial conditions chosen are
such that the entire sample is bound by the exchanger at the top of the
column.
&n this case the sample volume is not important and large volumes of dilute
solution can be applied, thereby eectively introducing a concentrationstage.
and ionic strength gradients may be employed.
+ut continuous gradients tend to give better resolution with less peak tailing.
gradient decreases and the ionic
strength increases,
Hhereas for cation exchangers both the p> and ionic gradients increase.
-,clusion &permeation' chromatography
The separation of molecules on the basis of their molecular sie and shape.
Mtilies the molecular sieve properties of a variety of porous materials.
;robably the most commonly used of such materials is a group of polymeric
organic compounds that possess a three-dimensional network of pores that
confers gel properties upon them.
The term gel !ltration is used to describe the separation of molecules of
varying molecular sie utiliing these gel materials.
;orous glass granules have also been used as molecular sieves and the term
controlled-pore glass chromatography introduced to describe this separation
technique.
The terms exclusion or permeation chromatography describe all molecular
separation processes using molecular sieves.
The general principle of exclusion chromatography is quite simple. =arge
molecules that are completely excluded from the pores will pass through the
interstitial spaces and will appear in the eLuent !rst.
@maller molecules will be distributed between the mobile phase inside and
outside the molecular sieve and will then pass through the column at a
slower rate, hence appearing last in the eLuent.
-
8/10/2019 Chromatographic Techniques WORD
24/56
5or a given type of gel, the distribution coe*cient Nd, of a particular analyte
between the inner and outer mobile phase is a function of its molecular sie.
&f the analyte is large and completely excluded from the mobile phase within
the gel, Nd 6 9,
whereas if the analyte is su*ciently small to gain complete access to the
inner mobile phase, Nd 6 $.
4ue to variation in pore sie between individual gel particles, there is some
inner mobile phase that will be available and some that will not be available
to analytes of intermediate sie?
hence Nd values vary between 9 and $.
&t is this complete variation of Nd between these two limits that makes
possible the separation of analytes within a narrow molecular @ie range on a
given gel.
-
8/10/2019 Chromatographic Techniques WORD
25/56
pplications
Puri8cation. ;uri!cation of biological macromolecules by facilitating theirseparation from $arger and smaller molecules. Airuses, proteins, enymes,
hormones, antibodies, nucleic acids and polysaccharides.
/elative molecular mass determination:The elution volumes of globular
protein are determined largely by their relative molecular mass (:r). &t has
been stated that over a considerable range of relative molecular masses, the
elution volume is approximately linear function of the logarithm of :r.
)olution concentration. @olutions of high :r substances can be
concentrated by the addition of dry @ephadex
-
8/10/2019 Chromatographic Techniques WORD
26/56
more e*cient than dialysis. "pplications include removal of phenol from
nucleic acid preparation, protein preparations and salts from samples eluted
from ion exchange chromatography.
fnity chromatography
;uri!cation by a*nity chromatography does not rely on dierences in the
physical properties of the molecules to be separated.
&nstead, it exploits the unique property of extremely speci!c biological
interactions to achieve separation and puri!cation.
"s a consequence, a*nity chromatography is theoretically capable of giving
absolute puri!cation, even from crude mixtures in a single process.
The technique was originally developed for the puri!cation of enymes, but it
has since been extended to nucleotides, nucleic acids, immunoglobulins,
membrane receptors and even to whole cells and cell fragments.
Mnder the correct experimental conditions, when a complex mixture
containing the speci!c compound to be puri!ed is added to the immobilised
ligand.
only that compound will bind to the ligand. "ll other compounds can therefore
be washed away and
-
8/10/2019 Chromatographic Techniques WORD
27/56
The compound subsequently recovered by displacement from the ligand.
The method requires a detailed prior knowledge of the structure and
biological speci!city of the compound to be puri!ed.
&n the case of an enyme, the ligand may be the substrate, a reversible
inhibitor or an allosteric activator.
The conditions chosen would normally be those that are optimal for enyme-
ligand binding.
Matri,
n ideal matri, or afnity chromatography must possess the ollo$ing
characteristics:
(i) &t must contain suitable and su*cient chemical groups to which the ligand
may be covalently coupled and it must be stable under the conditions of theattachment.
(ii) &t must be stable during binding of the macromolecule and its subsequent
elution.
(iii) &t must at the most interact only weakly with other macromolecules to
minimie nonspeci!c adsorption.
(iv) &t should exhibit good 3ow properties.
&n practice, particles that are uniform, spherical and rigid are used. The most
common ones are the cross-linked agarose (@epharose, +io
-
8/10/2019 Chromatographic Techniques WORD
28/56
Practical procedure
The ligand-treated matrix is packed into a column in the normal way for the
particular type of support.
The buer used must contain any cofactors, such as metal ions, necessary for
ligand macromolecule interaction.
Bnce the sample has been applied and the macromolecule bound, the
column eluted with more buer to remove non-speci!cally bound
contaminants.
The puri!ed compound is recovered from the ligand by either speci!c or non-
speci!c elution.
Don-speci!c elution may be achieved by a change in either p> or ionic
strength.
&f elution is achieved by a p> change, the p> of the collected fractions must
be read7usted to the optimum value to minimise the opportunity for protein
denaturation.
-
8/10/2019 Chromatographic Techniques WORD
29/56
"*nity elution involves the addition of a high concentration of substrate, or
reversible inhibitor of the macromolecule if it is an enyme or the addtion of
the ligand for which macromolecule has a higher a*nity than it has for the
immobilised ligand.
The puri!ed material is eventually recovered in a buered solution that maybe contaminated with speci!c eluting agents or high concentrations of salt
and these must be removed by such techniques as exclusion chromatography
before the isolation is complete.
pplications
" wide range of enymes O other proteins, including receptor proteins and
immunoglobulins, has been puri!ed by a*nity chromatography.
The principles have been extended to nucleic acids2 mED" for example, is
routinely isolated by selective hybridisation on poly(M)-@epharose 0+ by
exploiting its poly(") tail.
" valuable development of a*nity chromatography is its use for the
separation of a mixture of cells into homogeneous populations. The technique
relies on the antigenic properties of the cell surface or the chemical nature of
exposed carbohydrates residues on the cell surface or on a speci!c
membrane receptor-ligand interaction.
The immobilised ligands used include ;rotein ", which binds to the 5c region
of &g
-
8/10/2019 Chromatographic Techniques WORD
30/56
Clution of the bound protein quite often requires forceful conditions because
of the very tight binding with the antibody which may lead to protein
denaturation.
Cxamples of elution procedures include the use of high salt concentrations
with or without the use of a detergent
The use of urea, @4@, guanidine hydrochloride, all of which cause
denaturation.
The use of chaotropic agents such as thiocyanate, perchlorate and
tri3uoracetate or lowering the p> to about % may avoid denaturation.
Metal chelate chromatography
"lso called as immobilised metal a*nity chromatography
This is a special form of a*nity chromatography in which an ion such as
u'P, Qn'P, >g'P or d'P or a transition metal ion such as Di'P, or :n'P
is used to bind proteins selectively by reaction with groups of histidine
residues, thiol groups in cysteine residues and indole groups of tryptophan
residues.
The immobilisation of the protein involves the formation of coordinate bonds
that must be su*ciently stable to allow retention during the elution of non-
binding contaminating material.
The subsequent elution of the protein can be achieved either by simply
lowering the p> or by the use of complexing agents such C4T".
;roteins puri!ed by this technique include !brinogen, superoxide dismutase
and the histone nuclear proteins.
;ractical
"nalytical +iochemistry &(+T2 #%$,#%')
Terminology
:ole2 "mount of a substance present in a 3ask irrespective of volume. $
mole6:H in gms6 .9'%x$9'%6avogadronumber
-
8/10/2019 Chromatographic Techniques WORD
31/56
-
8/10/2019 Chromatographic Techniques WORD
32/56
;reparation of solution on the basis of density, wt/wtI i.e g of solute /$99g of
solution
4ensity6 wt / unit volume
@;B0(:H) and Da>';B0(:H) required to prepare
#99ml of 9.': solution.
'. alculate the volume of acetic acid required to prepare $99ml of 9.$:
solution. 4ensity6$.9#$g/ml, purity6RR.#I
%.alculate volume of concentrated >l required to prepare ': and 'D
solution.
9hat is color
-
8/10/2019 Chromatographic Techniques WORD
33/56
Clectromagnetic radiation between %9-G9 nm
olor is one aspect of appearance
olor 6 light source P ob7ect properties P eye P brain
The human eye is most sensitive at ### nm
Colorimetry
Two ob7ects may appear the same when viewed under one light source,
but dierent under another 6 metamerism
:etamerism is one of the ma7or industrial problems in color matching
olorimetry attempts to quantify the perception of color
)ources and illuminants
-
8/10/2019 Chromatographic Techniques WORD
34/56
@ource 6 physical entity that produces radiation
&lluminant 6 table of values of spectral power distribution
&lluminant 4# represents average daylight. 4#9 represents typical
indoor light
0b"ects
Bb7ects are characteried by the amount of light they emit and re3ect
or transmit at each wavelength of interest
Hhen light is incident on an ob7ect a part of it is absorbed, a part is
re3ected and a part may be trasmitted
The ob7ect may also emit light
"ll these characteristics contribute to the observed color
-
8/10/2019 Chromatographic Techniques WORD
35/56
Transmitted light
=ight is characteried by its frequency ( ), the number of times the
crest of the wave passes some point in space per second,
or by its wavelength (), the distance between two successive crests.
These two quantities are related by the speed of light, a fundamental
constant2 U6c6%V$9m/s.
;lanck related the frequency of light to its energy (C) according
to E6hU, where h is ;lanckWs constant, h6.'V$9X%0Y/s.
-
8/10/2019 Chromatographic Techniques WORD
36/56
" compound will absorb light when the radiation posesses the energy
needed to move an electron from its lowest energy (ground) state to
some excited state. The particular energies of radiation that a
substance absorbs dictate the colors that it exhibits. onversely the
color of a compound can help us to determine its electronic
con!guration.
Hhite light contains all wavelengths in this visible region. Hhen a
transparent sample (like most aqueous solutions) absorbs visible light,
the color we perceive is the sum of the remaining colors that are
transmitted by the ob7ect and strike our eyes.
-
8/10/2019 Chromatographic Techniques WORD
37/56
The incident light from a tungsten &visible light source) or
deuterium &;< light source) lamp is focused by a lens and passes
-
8/10/2019 Chromatographic Techniques WORD
38/56
-
8/10/2019 Chromatographic Techniques WORD
39/56
The eer+%ambert %a$
>ebc
Hhere " 6 absorbance (no units, sinceA = log10P0/ P)
e6 molar absorbtivity with units of = mol-$ cm-$6absorbanse if $:
solution
b6 path length of the sample - that is, the path length of the cuvette
in which the sample is contained. centimetres.
c >concentration of the compound in solution, expressed in mol =-$
" \ at low concentation
;ath length / cm 9 9.' 9.0 9. 9. $.9
IT $99 #9 '# $'.# .'# %.$'#
"bsorbance 9 9.% 9. 9.R $.' $.#
-
8/10/2019 Chromatographic Techniques WORD
40/56
The =aw is not obeyed at high concentrations
&nharent property of molecules2 &f molecule absorbs more light- low
conc can be detected.
"bsorption of photon \ concentration of absorbing molecules
\ Thickness of sample
4etermination of max of ;otassium dichromate
Aeri!cation of +eer8s =aw
'etermination of ma
repare *+mg%++ ml.repare+ ml of *+ -g%ml solution of potassium dichromate in
+.+* K/0 by diluting + times.
1ead absorbance from 23+nm to 3*+nm at a difference of *nm. 'raw the graph
between absorbance vs wavelength
'etermine the wavelength at which shows maimum absorbance ie ma
4erification of Beer!s law
repare different dilutions (+ml each) of potassium dichromate * -g%ml(+.ml),+
-g%ml(+.5ml),5+ -g%ml(+.3ml),2+ -g%ml(+.6ml),3+ -g%ml(+.7ml),6+ -g%ml(.5ml),7+-g%ml(.6ml),++ -g%ml(5ml).1ead absorbance at wavelength(nm) of maimum
absorbance.. 'raw the graph between absorbance vs concentration. 'raw the
interpretation.
8stimation of haemoglobin in blood%verification of Beer!s law
1eagent +.++9 Ammonium hydroide ( 3 ml ammonia%l)
-
8/10/2019 Chromatographic Techniques WORD
41/56
'ilute blood ml : ;;ml +.++9 ammoinum hydoide
'ilute blood 03/0 A*97nm
.+ml ;.+ml
5.+ 7.+
3.+ 6.+
6.+ 3.+
7.+ 5.+
+.+ +
8stimation of haemoglobin 0bgopper sulfate)
-
8/10/2019 Chromatographic Techniques WORD
42/56
(5) eutrali&e 03:with a/0 to get 02
(2) @team distillation of 02and trap in boric acid.
(3) Titrate with hydrochloric acid.
Calculation
ram nitrogen% gram of sample ?
(ml of sample C ml of blanD) (normality) of standard acid +.+3g%me$
weight of
sample
ml of hydrochloric acid re$uired to titrate sample solution.
'isadvantages not all itrogen is protein.
urine
yrimidine 'A, 1A, etc.
Erea
any plant tissues have G *+< nonC
protein itrogen.
< itrogen
6.5* ? < rotein
$iuret %est
-
8/10/2019 Chromatographic Techniques WORD
43/56
eptide >hains Biuret >omplees( purplecolor)
$iuret %est
1eproduciple
4ery few interfering agents
(ammonium salts being one such agent )
Fewer deviations than with the Howry or ultraviolet absorption methods
1e$uires large amounts protein (C5+mg)
How sensitivity
. 'ilute samples to an estimated to + mg%ml with buffer. Add ml to each assay tube.
'uplicate samples are recommended, and a range of dilutions should be used if the
actual concentration cannot be estimated.
5. repare a reference tube with ml buffer.
2. repare standards from + mg%ml bovine serum albumin, preferably calibrated using
absorbance at 57+ nm and the etinction coefficient. 1ange should be from to + mg
protein.
3. Add ; ml Biuret reagent to each tube, vorte immediately, and let stand 5+ min.
*. 1ead at **+ nm.
&olin"Ciocalteu ( 'ory ) Assay
-
8/10/2019 Chromatographic Techniques WORD
44/56
@ensitive over a wide range
>an be performed at room temperature
+C5+ times more sensitive than E4 detection.
>an be performed in a microplate format .
any substances interfere with the assay .
(strong acids, ammonium sulfate, buer, 8'TA, nonionic and cationic detergents,
carbohydrate, lipids and some salts. )
TaDes a considerable amount of time to perform
The assay is photosensitive, so illumination during the assay must be Dept consistent for
all samples
Amount of color varies with different proteins.
The incubation time is very critical for a reproducible assay.
The reaction is also dependent on p0 and a worDing range of p0 ; to+.* is essential.
eagents A) 5< sodium carbonate in +. a/0.
B) +.*< copper sulfate in 5opper reagent *+ml of A and 5ml of B (Freashly prepared
-
8/10/2019 Chromatographic Techniques WORD
45/56
Folins reagent commercial contains phophmolybdic and phosphotungstate in oC
phosphoric acid. 'ilute with 'I.
+. a/0 3g in H
@tandard B@A +. mg in +. a/0.
EnDnown *+ diluted serum
. Add standard B@A containing up to 5+,3+,6+,7+,++ ,3+,7+ Jg
( +.5,+.3,+.6,+.7,.+,.3,.7 ml )of protein. TaDe +. and +.5 ml of unDnown. For
blanD 5ml of +. a/0
5. Bring all tubes to 5 mH total volume with +. a/0.
2. repare the copper reagent and diluted FolinC>iocalteu reagent ( with 'I).
3. To each tube add * mH of copper reagent and thoroughly vorte.
*. #ncubate tubes at room temperature for + min.
6. Add +.* mH of diluted FolinC>iocalteu reagent. 4orte immediately.
9. #ncubate at room temperature for 2+ min.
7. 4orte the tubes, &ero the spectrophotometer with the blanD and measure
absorbance at *++C9*+ nm.
Ultraviolet Absorbance
#f you dont Dnow what the protein concentration of an unDnown sample is liDely to be,
the ultraviolet method might be a good starting point.
This is often used to estimate protein concentration prior to a more sensitive method
onitors the absorbance of aromatic amino acids, tyrosine and tryptophan
0igher orders of protein structure, many other cellular components, and particularly
nucleic acids, also may absorb E4 light
This method is the least sensitive of the methods
The real advantages of this method are that the sample is not destroyed and that it is
very rapid.
Any nonCprotein component of the solution that absorbs ultraviolet light will intefere with
the assay.
-
8/10/2019 Chromatographic Techniques WORD
46/56
recipitate with +A dissolve in phosphate buffer and used for protein estimation.
>ell and tissue fractionation samples often contain insoluble or colored components that
interfere.
The most common use for the absorbance assay is to monitor fractions from
chromatography columns, or any time a $uicD estimation is needed and error in protein
concentration is not a concern.
An absorbance assay is recommended for calibrating bovine serum albumin or other
pure protein solutions for use as standards.
roteins in solution absorb ultraviolet light with absorbance maima at 57+ and 5++ nm.
Amino acids with aromatic rings are the primary reason for the absorbance peaD at 57+
nm.
eptide bonds are primarily responsible for the peaD at 5++ nm.
@econdary, tertiary, and $uaternary structure all affect absorbance, therefore factors
such as p0, ionic strength, etc. can alter the absorbance spectrum
>oncentration (mg%ml) ? (.** A57+) C +.96 A56+)
>oncentration (mg%ml) ? Absorbance at 57+ nm divided by absorbance coefficient
Absorbance coefficients of some common protein standards
Bovine serum albumin (B@A) 62
Bovine, human, or rabbit #g 27
>hicDen ovalbumin 9+
Procedure*
Iarm up the E4 lamp (about * min.)
AdLust wavelength to 57+ nm
>alibrate to &ero absorbance with buffer solution only
easure absorbance of the protein solution ( 5* ug%ml, *+ ug%ml )
AdLust wavelength to 56+ nm
>alibrate to &ero absorbance with buffer solution only
-
8/10/2019 Chromatographic Techniques WORD
47/56
easure absorbance of the protein solution
>alculate the protein concentration by Absorbance coefficients and by formula.
$icinchoninic Acid ( $CA ) Assay
4ery sensitive and rapid if you use elevated temperatures
>ompatible with many detergents
IorDing reagent is stable
4ery little variation in response between different proteins
Broad linear worDing range
The reaction does not go to completion when performed at room temperature
. repare the re$uired amount of protein determination reagent by adding volume
copper sulfate solution to *+ volumes of bicinchoninic acid solution.
-
8/10/2019 Chromatographic Techniques WORD
48/56
5. @et up test tubes containing samples and Dnown amounts of bovine serum albumin in
the range of + to ++ micrograms. 8ach tube should contain +. mH total volume.
2. Add 5.+ mH of the protein determination reagent to each tube and vorte.
3. #ncubate the tubes at 6+o> for * min.
*. >ool the tubes to room temperature and determine the absorbance at *65 nm.
Dye"$inding ( $radford ) Assay
>BB primarily responds to arginine residues
(eight times as much as the other listed residues)
#f you have an arginine rich protein,
Mou may need to find a standard
that is arginine rich as well.
>BB binds to these residues in the anionic form
Absorbance maimum at *;* nm (blue)
The free dye in solution is in the cationic form,
Absorbance maimum at 39+ nm (red).
-
8/10/2019 Chromatographic Techniques WORD
49/56
Fast and inepensive
0ighly specific for protein
4ery sensitive NC5+ Jg (micro assay) 5+C5++ Jg (macro assay)O
>ompatible with a wide range of substances
8tinction coCefficient for the dyeCprotein comple is stable over + orders of magnitude
(assessed in albumin)
'ye reagent is comple is stable for approimately one hour
onClinear standard curve over wide ranges
1esponse to different proteins can vary widely, choice of standard is very important
Absorption spectra of anionic and cationic forms of the dye overlap.
@o the standard curve is nonClinear although all Dit providers of the Bradford assay insist
that the assay performs linearly.
The assay performs linearly over short concentration stretches.
-
8/10/2019 Chromatographic Techniques WORD
50/56
#f your sample is more than 5+ micrograms, a second order curve will fit much better than a
linear curve
. Iarm up the spectrophotometer for * min. before use
-
8/10/2019 Chromatographic Techniques WORD
51/56
5. 'ilute samples with buffer to an estimated concentration of to 5+ micrograms%milliliter
2. repare standards containing a range of to 5+ micrograms protein (albumin or gamma
globulin are recommended) to a volume of 5++ Jl (to a volume of ++ Jl if you are
adding a/0)
3. repare unDnowns to estimated amounts of to 5+ micrograms protein per tube to 5++
Jl (++ Jl if you are using a/0)
*. Add ++ Jl a/0 to each sample and vorte.
6. Add 7++ Jl dye reagent and incubate * min.
9. easure the absorbance at *;* nm.
Ultraviolet Absorbance
PuicD
@ample can be recovered
Eseful for estimation of protein before using a more accurate method
0ighly susceptible to contamination by buffers, biological materials and salts
rotein amino acid composition is etremely important, thus the choice of a standard is
very difficult, especially for purified proteins
Absorbance is heavily influence by p0 and ionic strength of the solution.
+stimation Procedure
. Qero spectrophotometer to water (or buffer)
5. TaDe the absorbance at 57+ nm in a $uart& cuvette
2. >hange wavelength to 56+ nm and &ero with water (or buffer)
3. TaDe absorption at 56+ nm in a $uart& cuvette
*. Ese the following e$uation to estimate the protein concentration
NroteinO (mg%mH) ? .**A57+R +.96A56+
$ioad DC Protein Assay
Based on Howry Assay with following improvements
-
8/10/2019 Chromatographic Techniques WORD
52/56
. 1eaches ;+< of its maimum color development within * minutes
5. The color changes not more than +< in 5 hours
repare * dilutions of samples and * dilutions of a protein standard containing from
+.5 mg%ml to about .* mg%ml protein. A standard curve should be prepared each time
the assay is performed. For best results, the standard should be prepared in the same
buffer as the sample.
B@A
mg%ml
+.53
(%6)
+.37
(5%6)
+.95
(2%6)
+.;6
(3%6)
.5+
(*%6)
.33
(6%6
f celletract
+ +.> +.5> +.2> +.*> >
ipet * ul of standards and samples into a microtiter plate
Add 5*ul of reagent A into each well
Add 5++ul reagent B into each well
Agitate the plate to mi the reagents
, - . / 0 1 2
A 3 34-/ 34/5 3461 ,4-3 ,4// empty
$ 3 34-/ 34/5 3461 ,4-3 ,4// empty
C 3 34-/ 34/5 3461 ,4-3 ,4// empty
D blan7 blan7 blan7 blan7 blan7 blan7 empty
+ 3 34,C 34-C 34.C 340C C empty
& 3 34,C 34-C 34.C 340C C empty
8 3 34,C 34-C 34.C 340C C empty
9 empty empty empty empty empty empty empty
-
8/10/2019 Chromatographic Techniques WORD
53/56
8mpty? Air
BlanD?1eagent A:B
+?1eagent A:B:*ul water
After * minutes, absorbance can be read at 9*+nm. The absorbance will be stable for
about hour
:tandard , - . / 0 1
A 3 343.2 343/6 34313 34306 34352
$ 3 343/- 3430, 34323 34316 34355
C 3 343.6 3430, 3432, 34350 34352
-
8/10/2019 Chromatographic Techniques WORD
54/56
Average 3 343.6. 34303. 343123 3432,3 34352.
Con4 3 34-/ 34/5 342- 3461 ,4//
lot >oncentration as M ais
Standard
y = 19.815x - 0.2861
R2= 0.9868
0
0.5
1
1.5
2
0 0.02 0.04 0.06 0.08 0.1
Absorbancd
Concentration
y ? ;.7* C +.576
NroteinO? ;.7*NAOC +.576
;s
lot >oncentration as = ais
Standard
y = 0.0498x + 0.015
R2= 0.9868
0
0.02
0.04
0.06
0.08
0.1
0 0.5 1 1.5 2
Concentration
Abs
orbance
y ? +.+3;7 : +.+*
NroteinO?(NAO C+.++*)%+.+3;7
>ould introduce errors into the calculation
NroteinO? ;.7*NAOC +.576
:tandard , - . / 0 1
+ 3 343-5 34303 3430. 34,-/ 34,1,
& 3 343-6 343/2 34321 34362 34,2.
-
8/10/2019 Chromatographic Techniques WORD
55/56
8 3 343,5 343/6 34312 34,3/ 34,05
Average 3 343-03 343/52 34310. 34,35. 34,1/3
Con4 3 34-, 3415 ,43, ,451 -465
&raction 3 34, 34- 34. 340 ,
protein concentration
y = .205x - 0.0644
R2= 0.9!41
-0.5
0
0.5
1
1.5
2
2.5
.5
0 0.2 0.4 0.6 0.8 1 1.2
"raction
co
ncentrati
rotein >oncentration? 2.66mg%ml
%ips
Ese clean glassware and supplies
aDe sure cuvettes are clean of all residues
rotein assays are strongly influenced by the composition of the proteins present in your
sample
Become familiar with spectrophotometry before proceeding
Always let a spectrophotometer warm up for *C5+ minutes before using
Know the limits of the spectrophotometer with which you are using
@tandard curves are not always linear
The protein used for your standard curve must maDe sense
aDe sure your standard curve covers the absorbance range of your unDnown with at
least two points on either side
aDe sure that your protein solution behaves in a reproducible manner to the assay
method by maDing a dilution curve
-
8/10/2019 Chromatographic Techniques WORD
56/56
Ese buffer and water blanDs to anchor down your standard curve
lace the protein concentration on the yCais of you standard curve plot so that you can
use the bestCfit e$uation directly for concentration determination
ALai ***