1 / 17 Cheops Calibraons UGE Calibration team : Bruno Chazelas, Mirsad Sarajlic, Michael Sordet, Adrien Deline, Francois Wildi, Thibault Kuntzer

Transcript of Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source...

Page 1: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

1 / 17

Cheops Calibrations

UGE Calibration team : Bruno Chazelas, Mirsad Sarajlic, Michael Sordet, Adrien Deline,

Francois Wildi, Thibault Kuntzer

Page 2: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

2 / 17

Cheops :CHaracterizing ExOPlanet Satellite

● ESA-S , mission● Precsion photometer :

20 ppm photometric precision

● Can be pointed to a large portion of the sky

● Science goal :

– Mass / Radius relation for dense exoplanets (Li Zeng et al, 2017)

Page 3: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

3 / 17

Cheops instrument

● Defocalized telescope● Photometry of a single star● CCD e2v 47-20 : 1024x1024 pixels● Sun-Synchronous ● Field is rotating● Pointing ~ 2-3 arcsec● Data : imagette 200x200

Page 4: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

4 / 17

Calibration Test benchCalibration Source Module :

Super stable sourceNeutral densityFilter wheelMonochromator

CHEOPS TV chamber

Page 5: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

5 / 17

Super Stable source

LDLSLight source

Gamma Scientific photometer

Stability 5 ppm over 24hComputer

Mobile knifeIntegrating sphere

Page 6: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

6 / 17

Calibration Test bench

Page 7: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

7 / 17

Detector chain Calibration

Characteristics Value Comment

RON Nominal 230 kHz

7.2 ADU (14.1 e-)

RON Nominal 100 kHz

3.5 ADU (6.8 e-)

Gain Nominal 230 kHz

0.511 +/- 0.003

Dark 0.056 e-/s

Warm pixels 7 pixelsWith dark above 5 e-/s

Gain temperature sensitivity

-900 ppm/K

Non linearity amplitude

1.2 %

Full well 114-121 ke-

Page 8: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

8 / 17

Flat-fields

● Overall target precision : 0.1 %

● Direct measurement with integrating sphere

– Have to be corrected from sphere non uniformity

– Have to be corrected from distortion

● Spectral reconstruction :

– Monochromator + UVBRI Filter measurement

Page 9: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

9 / 17

Flat-fields : Distortion

● There is more flux on the border of the measured FF

● It does not affect point sources

● If not corrected huge photometric errors

● Recommended to redo measurement in space

Page 10: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

10 / 17

Flat-fields : Distortion measurement

Page 11: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

11 / 17

Flat-field : spectral synthesis

With the calibration measurement one can reproduce with high precision the “real” flat of a source with Smooth spectrum

Page 12: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

12 / 17

Flat-field : spectral synthesis

Page 13: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

13 / 17

Flat-field : precision

● Photometric precision ~ 0.06 %● Error on the correction of the non uniformity of the

integrating sphere ~ 0.04 %● Error due to the distortion correction ~ 0.034 %

● Total error ~ 0.08 %

Page 14: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

14 / 17

PSF

In the center of FoVE90 = 12.8 ± 0.3 pixels

Page 15: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

15 / 17

Performances

● No end-to-end test (lack of time)● But test with integrating sphere

– Temperature stability < 1.5 mK <=> < 2 ppm– Bias voltage stability < 40 µV <=> < 2 ppm – Test bench is the drifting element.– One can give minimum performance achieved

(we are in the right range!)

Page 16: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

16 / 17

Performances

Page 17: Cheops Calibrations - UNIGEobswildif/publications/2018_ICSO_Chazelas.pdf · Super Stable source LDLS Light source Gamma Scientific photometer Computer Stability 5 ppm over 24h Mobile

17 / 17

Conclusion

● All needed calibrations available● Instrument in good shape to make great science● Launch next year