Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY
description
Transcript of Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY
![Page 1: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/1.jpg)
Kanellopoulos, MD www.brilliantvision.com
Laboratory evaluation of LASIk flap parameters created with 2
femtosecond lasers ASCRS, San Diego 2011
Laboratory evaluation of LASIk flap parameters created with 2
femtosecond lasers ASCRS, San Diego 2011
Chatzilaou, Georgios, MD¹A. John Kanellopoulos, MD²
Professor NYU Medical School, NYDirector, Laservision.gr Institute, Athens, Greece
Chatzilaou, Georgios, MD¹A. John Kanellopoulos, MD²
Professor NYU Medical School, NYDirector, Laservision.gr Institute, Athens, Greece
Financial Interest: None¹Travel Reimbursement from Alcon, USA, Wavelight, Bausch and Lomb²
![Page 2: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/2.jpg)
Kanellopoulos, MD www.brilliantvision.com
PurposeThe laboratory evaluation of LASIK flap of a novel femtosecond laser versus our standard.The laboratory evaluation of LASIK flap of a novel femtosecond laser versus our standard.
FS 200FS60
![Page 3: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/3.jpg)
FS60
Kanellopoulos, MD www.brilliantvision.com
![Page 4: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/4.jpg)
Wavelight FS200
Kanellopoulos, MD www.brilliantvision.com
![Page 5: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/5.jpg)
Kanellopoulos, MD www.brilliantvision.com
Methods20 cadaver corneas, underwent LASIK flap creation Group A(10 corneas)with the Wavelight, Ultraflap 150kHz and Group B (10 corneas) with the INTRALASE FS60Peri-operatively the flap diameter (FD), hinge length (HL) , total cornea and flap pachymetry(tP & fP), and flap creation interval were measured. Following tissue fixation, light microscopy sections were calculated for: peripheral, para-central and central thickness, intra-flap thickness variability (IFTV), diameter, hinge length and overall interface quality.
20 cadaver corneas, underwent LASIK flap creation Group A(10 corneas)with the Wavelight, Ultraflap 150kHz and Group B (10 corneas) with the INTRALASE FS60Peri-operatively the flap diameter (FD), hinge length (HL) , total cornea and flap pachymetry(tP & fP), and flap creation interval were measured. Following tissue fixation, light microscopy sections were calculated for: peripheral, para-central and central thickness, intra-flap thickness variability (IFTV), diameter, hinge length and overall interface quality.
![Page 6: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/6.jpg)
Kanellopoulos, MD www.brilliantvision.com
Results
Group A: In mean values: fP: 110 microns +/- 5μm, IFTV: (+/- 5)μm. FD: 8.2 mm, HL: 3.55 mm +/- 0.34, creation time 15 secondsGroup B: In mean values: fP: 115 +/-5μm(intended: 110), IFTV: (+/-5)μm, FD: 8.1 mm(intended: 8), HL: 3.25mm +/-0.24, creation time 17 seconds
Group A: In mean values: fP: 110 microns +/- 5μm, IFTV: (+/- 5)μm. FD: 8.2 mm, HL: 3.55 mm +/- 0.34, creation time 15 secondsGroup B: In mean values: fP: 115 +/-5μm(intended: 110), IFTV: (+/-5)μm, FD: 8.1 mm(intended: 8), HL: 3.25mm +/-0.24, creation time 17 seconds
![Page 7: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/7.jpg)
Conclusions
• This study suggests highly reproducible flap parameters with both platforms. Further clinical studies may validate these data.
• This study suggests highly reproducible flap parameters with both platforms. Further clinical studies may validate these data.
Kanellopoulos, MD www.brilliantvision.com
![Page 8: Chatzilaou, Georgios , MD¹ A. John Kanellopoulos, MD² Professor NYU Medical School, NY](https://reader035.fdocuments.net/reader035/viewer/2022070413/56814d40550346895dba64b9/html5/thumbnails/8.jpg)
Thank youThank you
• www.brilliantvision.com• www.brilliantvision.com
Kanellopoulos, MD www.brilliantvision.com