- 1 -
CHATILLON® DFIS SeriesACCUFORCE® AF2 Series
Digital Force Gauges
Operating Instructions
Instruction Manual NC001562May 2001
Issue 3
CHATILLON DFIS Series Digital Force Gauge
AMETEK AF2 Series Shown with MHTCMechanical Test Stand
- 2 -
TABLE OF CONTENTS
INSTRUMENT PRECAUTIONS .............................................................................................. 4
1.0 INTRODUCTION AND PRODUCT OVERVIEW ......................................................... 5
1.1 PRODUCT FEATURES .................................................................................................. 5
1.2 ELECTRICAL SAFETY ................................................................................................... 5
1.3 RECHARGING BATTERIES ........................................................................................... 5
1.4 PHYSICAL DIMENSIONS ................................................................................................ 6
1.5 REVERSIBLE HOUSING ................................................................................................. 6
1.6 ABOUT THE KEYPAD ..................................................................................................... 7
1.7 ABOUT THE DISPLAY ..................................................................................................... 7
1.8 CONNECTION TO EXTERNAL DEVICES ................................................................... 8
1.9 INTERFACE CABLES ..................................................................................................... 9
1.10 SETTING UP FOR SAFE USE ...................................................................................... 10
2.0 FRONT PANEL OPERATION ....................................................................................... 11
2.1 NORMAL MODE .............................................................................................................. 11
2.1.1 Selecting Units of Measure .............................................................................. 11
2.1.2 Selecting Type of Measurement - Capturing Peaks ..................................... 11
2.1.3 Zeroing ............................................................................................................... 12
2.1.4 Transmitting Readings ..................................................................................... 12
2.2 CONFIGURATION MODE ............................................................................................... 13
2.2.1 Configuring Output Port- Analog Output .......................................................... 13
2.2.2 Configuring Output Port- Digital Outputs .......................................................... 13
2.2.2.1 Enabling the Mitutoyo Output ................................................................... 14
2.2.2.2 Enabling the RS-232 Port- Numerical Data Only .................................. 15
2.2.2.3 Enabling the RS-232 Port- Full Data ....................................................... 16
2.2.2.4 Modifying Communication Parameters .................................................. 17
2.3 DATA COLLECT MODE .................................................................................................. 18
2.3.1 About Data Collect Mode ................................................................................. 18
Page
- 3 -
3.0 CALIBRATING THE DFIS/AF2 SERIES .................................................................... 20
3.1 CALIBRATION PROCEDURE ........................................................................................ 20
4.0 DIAGNOSTICS ................................................................................................................ 23
4.1 SOFTWARE PART NUMBER ........................................................................................ 23
4.2 SOFTWARE REVISION LEVEL ..................................................................................... 23
4.3 CAPACITY AND UNITS .................................................................................................. 24
4.4 NO LOAD ZERO VALUE ................................................................................................. 24
4.5 RS232 LOOPBACK TEST .............................................................................................. 25
4.6 KEYPAD TEST ................................................................................................................. 25
5.0 TECHNICAL REFERENCE AND SPECIFICATIONS ............................................... 26
6.0 MODEL NUMBERS ........................................................................................................ 27
TABLE OF CONTENTSPage
PRODUCT WARRANTYThis instrument is warranted against defects in workmanship, material and design for one (1) year from thedate of delivery to the extent that AMETEK will, at its sole option, repair or replace the instrument or any partthereof which is defective provided, however, that the warranty shall not apply to instruments subjected totampering or abuse, or exposed to highly corrosive conditions.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED ANDAMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES, INCLUDING, WITHOUT LIMITATION, ANYWARRANTY OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY. AMETEK SHALLNOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES, INCLUDING, BUT NOTLIMITED TO, ANY ANTICIPATED OR LOST PROFITS.
This warranty is voidable if the purchaser fails to follow any and all instructions, warnings or cautions in theinstrument’s operation manual.
If a manufacturing defect is found, AMETEK will replace or repair the instrument or replace any defective partthereof without charge; however, AMETEK’s obligation hereunder does not include the cost of transportationwhich must be borne by the customer. AMETEK assumes no responsibility for damage in transit, and anyclaims for such damage should be presented to the carrier by the purchaser.
- 4 -
PRECAUTIONS
q Read the instruction manual completely before attempting to use the DFIS or ACCUFORCE2 (AF2) Series. By following the instructions contained in this manual, the optimum accuracy andperformance can be attained.
IMPORTANT NOTE: The DFIS and ACCUFORCE 2 (AF2) gauges are identical in function-ality. The procedures, specifications and operations in this manual apply to both productsunless noted.
q Turn off the DFIS/AF2 Series before connecting or disconnecting any cables to the instrument.
q Do not overload the DFIS/AF2 Series! The instrument can handle an overload of up to 150% of therated full scale. The display will read “- - - -” when the gauge exceeds approximately 120% of fullscale. When the force applied is removed and “- - - -” is no longer displayed, the gauge may not readabove 121% of full scale in the Tension Peak or Compression Peak modes. Therefore, you mustcontact your local distributor to ensure the load cell was not damaged, unless you are 100% sure theforce applied did not exceed 150% of full scale.
q The DFIS/AF2 Series is designed for axial loading only! Applying load on an angle or eccentricloading may cause an erroneous reading.
q Do not use tools to attach the shaft adapter and accessories to the gauge. Use “finger-tight”pressure only to prevent damage to the load cell.
IMPORTANT NOTE: Before using your new DFIS/AF2 Series, fully charge the rechargeableNicad battery by plugging the supplied battery charger into the battery charger jack at the topof the gauge and the proper AC outlet. Use only the battery charger supplied with the gauge.Using other chargers can overcharge and damage the battery.
- 5 -
1.0 INTRODUCTION AND PRODUCT OVERVIEWThe DFIS/AF2 Series family of digital force gauges manufactured by AMETEK Test and CalibrationInstruments Division. They have been designed to provide accurate compression and tension forcemeasurements via an internal load cell. These instruments have been precisely crafted to provide years ofreliable service. Each gauge comes equipped with a rechargeable battery pack, AC battery charger,Accessory Attachments, and a handy Carrying Case.
1.1 PRODUCT FEATURESThe key features of the DFIS/AF2 Series are:
q Load capacity up to 500 lbs (250 kg, 2.5 kN)
q Internal sampling rate of 1000 Hz
q Displays Tension (T) or Compression (C)
q Simple keypad operation
q Rugged aluminum housing offers long life. Top (Front) housing is reversible
q Mounting arrangements for attaching to Manual and Motorized Test Stands
q Readings can be scaled using several engineering units: LB, KG, N
q Field calibration
q Analog and Digital outputs through common connector
q Can be controlled remotely via RS-232 port
1.2 ELECTRICAL SAFETYThese instruments have been assessed against the essentail health and safety requirements of the LowVoltage and the EMC Directives listed below. Based on conformity with the listed directives, the DFIS andAF2 are compliant with the following:
Directive Test StandardEMC Directive (89/336/EEC) EN 50081-1 EMC Generic Emissions Standard
EN 50082-1 EMC Generic Immunity Standard
Low Voltage Directive (73/23/EEC) EN 61010-1 Safety Requirement for Electrical Equip
1.3 RECHARGING BATTERIESA rechargeable NiCad battery powers the DFIS/AF2Series. Since NiCad batteries continue to dischargewhen not in use, it may be necessary to recharge the unit before use. When the battery power begins to loseits charge, the Battery Indicator will appear on the display . To recharge the battery, plug the battery chargerincluded with the gauge into the correct voltage source and insert the charger jack into the receptacle on theend of the gauge . Charge the battery for 10-12 hours to yield approximately 8-10 hours of continuousoperation. The gauge can be operated continuously with the charger connected.
_ WARNING -
Before using the gauge, please fully charge the Nicad battery. Use only the battery charger supplied withthe gauge. Using other chargers can overcharge and damage the battery.
- 6 -
1.4 PHYSICAL DIMENSIONS
1.5 REVERSIBLE HOUSING
The DFIS/AF2 Series is shipped shaft-down for stand-mounted applications, and the display on the front ofthe gauge is configured accordingly. For performing operations in which the gauge will be hand-held, it maybe desirable to rotate the front of the casing so that the display and logo will be shown right side up. To doso, remove the two screws from the back of the casing, carefully rotate the front of the casing 180 degreesbeing careful not to pull, pinch, or twist the wires inside, and then reinstall the screws.
NOTE: When removing the screws to the back casing for any reason, please take the proper ESD (ElectroStatic Discharge) precautions. Touching the circuit board without taking the proper ESD precautions maycause damage to the circuit board.
- 7 -
1.7 ABOUT THE DISPLAYThe display is an LCD seven segment display that is easily read even at a distance. The display illustratesdifferent symbols such as: battery symbol (BATT) indicating the battery is running low; a “C” for compres-sion or a “T” for tension. The peak operational modes are also displayed (T PEAK, C PEAK). The rightportion of the display indicates the force measurement units (LB, KG, N). The display updates 4 times persecond.
1.6 ABOUT THE KEYPADThe keypad consists of four dedicated keys (PEAK, UNITS, ZERO, XMIT). The keypad is operational whenthe power switch, located on the side of the gauge, is in the ON position.
Figure 1.2Keypad Layout
PEAK key cycles through three different operatingmodes: capture of the Normal reading; PeakCompression reading; or Peak Tension reading.The selected mode is indicated in the upper left ofthe display screen.
UNITS key allows you to select the display readingin lbs (LB), kilograms (KG) or newtons (N).
ZERO key returns current display to zero reading.Depressing this key clears peak readings as well.
XMIT is used to initiate data transmission to anexternal device, e.g. personal computer. This keyis also used to display and modify the gauge’scommunications parameters.
UNITS
XMIT
PEAK
ZERO
T C PEAK BATT
LBKGN
TTension
CCompression
PEAKCombined with
“T” or “C” toindicate peak
BATTIndicates low
battery condition
LB, KG, NForce measurement
indicator
-WARNING -Force reading accuracy may be affected by a low battery condition. BEFORE using the force gauge forthe first time, fully charge the rechargeable NiCd battery. Plug the charger into the gauge’s charge jackand into an AC outlet.
Always charge the battery whenever the BATT indicator is highlited. A minimum charge of 8 hours isrecommended.
0 0 0 0. . ..1
- 8 -
The DFIS/AF2 Series is supplied with a 12-pin female connector to provide both digital and analog outputs.AMETEK offers a variety of cables to handle most applications (Refer to Section 1.9). Select the cable typerequired to connect the gauge to the peripheral device using the mating connector. The pinout of theconnector is shown above. Pin assignments are defined in the table below.
1 TXD O RS-232 Transmitted Data
2 RXD I RS-232 Received Data
3 GND - Ground Digital Ground
4 - Clock Mitutoyo Clock
5 I Ready Mitutoyo Ready
6 I Request Mitutoyo Request
7 O Data Mitutoyo Data
8 (Reserved)
9 (Reserved)
10 - Ground Digital Ground
11 - Analog GND Analog Ground
12 O Analog SIG Analog Output
Table 1.8 I/O Connector Pinout
PIN Symbol I/O Purpose Description
98
7
65
4
3
21
1012 11
I/O Connector(View from Solder End)
1.8 CONNECTION TO EXTERNAL DEVICESThe DFIS/AF2 Series can be configured to accept input from an external source, and/or to send output to anexternal data-recording device, such as a strip chart recorder, an RS-232 or MITUTOYO printer, or apersonal computer. It can be used in combination with CHATILLON Test Stands and application softwaresuch as NEXYGEN to meet a wide variety of application requirements. To connect the gauge to a device,attach the appropriate interface cable to the port located at the top of the gauge, and the other end to theexternal device.
Power Adapter
I/O Connector
- 9 -
Cable Description
NC000652 6 ft (182 cm), Connects to a personalcomputer with a 25-pin RS-232 connection.
NC000652-1 10 ft (305 cm), Connects gauge to a personalcomputer with a 25-pin RS-232 connection.
NC000652-2 15 ft (450 cm), Connectsto a personalcomputer with a 25-pin RS-232 connection.
NC000652-3 20 ft (610 cm), Connects to a personalcomputer with a 25-pin RS-232 connection.
NC000653 6 ft (182 cm) Connects gauge to X-Y recorder (double banana)
NC000654 5 ft ( 152 cm) Connects gauge to a Mitutoyo device, 10-pin
NC000697 10 ft (305 cm) Connects gauge to a Mitutoyo device,RS232
NC000698 10 ft (305 cm) Connects gauge to a Datamyte
NC000647 6 ft (182 cm) Connects to CHATILLON Model TCD
ENC0125 8 ft (243 cm) Connects to CHATILLON Model TCM
1.9 INTERFACE CABLES
- 10 -
1.10 SETTING UP FOR SAFE USEThe DFIS/AF2 Series should be properly setup before accurate and most of all, safe measurements can bemade.
Only when you have verified that you will be conducting a safe test, proceed as follows:
1. Determine what and how it will be tested and verify that the DFIS/AF2 Series can handle the test.
2. If desired, mount the gauge in the test fixture.
3. If you will be powering the instrument from the AC line, plug the charger into its jack on the instrument.Plug the charger into the correct AC voltage outlet (the required AC voltage and frequency are listed onthe charger).
4. Otherwise, if you will be powering the instrument from batteries, fully charge the batteries before using.
IMPORTANT NOTE: Before using your new DFIS/AF2 Series, fully charge the rechargeableNicad battery by plugging the supplied battery charger into the battery charger jack at the topof the gauge and the proper AC outlet. Use only the battery charger supplied with the gauge.Using other chargers can overcharge and damage the battery.
5. Attach all necessary attachments and plug in the cables. Route the cables so they do not interfere withthe test.
6. Zero the gauge by depressing the ZERO key.
7. Attach the specimen and begin to take measurements.
- 11 -
2.0 FRONT PANEL OPERATIONThe DFIS/AF2 Series can operate in one of the three modes: Normal, Data Collect and Configuration.
2.1 NORMAL MODEThe principal function of the Normal mode is to provide indications of load applied to the load cell.Instantaneous force, peak-tension and peak-compression readings can be displayed and scaled in avariety of units of measure.
2.1.1 Selecting Units of MeasureForce indications can be displayed in LB, KG, or N. The gauge displays the active units of measure on theLCD display. Pressing the UNITS key will cycle between the three available measuring units.
With gauge power ON, depress the UNITS key to cycle through available force measurement units (LB,KG, N). Release the key when the desired unit is displayed.
2.1.2 Selecting Type of Measurement – Capturing PeaksThe gauge continuously measures instantaneous tension or compression forces (normal readings). It alsocaptures peak-tension and peak-compression forces. The gauge displays which measurement is beingindicated. Pressing the PEAK key will cycle between the available force measurement modes.
With gauge power ON, depress the PEAK key to cycle through available force modes (C, T, T PEAK, CPEAK). Release the key when the desired unit is displayed. The display will indicate C until tension isapplied to the loadcell. This will cause the display to indicate tension (T).
UNITSLBKGN
C
T PEAK
C PEAKLB
LB
LB
PEAK
- 12 -
2.1.3 ZeroingZeroing allows you to exclude the weight of accessories and attachments from the indicated force readings.It also clears captured peak tension and compression readings. To zero the gauge, momentarily press theZERO key. Only the current measurement type (i.e., NORM, T-PK, C-PK) will be zeroed. To zero and clearall measurement types press and hold the PEAK key and then press the ZERO key. Zeroing the gaugetakes approximately 0.5 seconds.
2.1.4 Transmitting ReadingsPressing the XMIT key will cause the gauge to transmit a single reading (current measurement type andcurrent units) through the output port (MITUTOYO or RS-232). If the gauge is in the NORM mode, the datavalue (NORM, T-PK or C-PK) displayed on the screen will be transmitted. If the gauge is in the DATACOLLECT mode, where a series of measurements are taken, the gauge will transmit an averaged value ofthe outputs at the loadcell.
To TARE, affix the required adapter or fixture to the gauge. Orient the gauge so that it is positioned as itwill be during testing. Depress the ZERO key to tare the gauge.
To ZERO, after a test, depress the ZERO key to zero the force readings. Removing power to the gaugewill also zero the force readings.
C PEAK
XMIT4 1 2 7.
LB
To TRANSMIT, depress the XMIT key to transmit the instantaneous force reading from the gauge to apersonal computer or external device. The proper interface cable must be connected to the gauge and thedevice.
C PEAK
ZERO0 0 0 0.
LB
- 13 -
2.2 CONFIGURATION MODEThe Configuration Mode facilitates the communications setup for the gauge.
2.2.1 Configuring Output Port- Analog OutputThe analog output is always available at the output port without configuration. The analog output is0 to -1.50Vdc (min) in tension mode and 0 to 1.50Vdc (min) in compression mode.
2.2.2 Configuring Output Port- Digital OutputThe DFIS/AF2 gauge supports three types of digital output: Modified BCD (Mitutoyo), RS232 Numerical DataOnly and RS232 Full Data.
When the digital output is configured for RS232 Numerical Data Only, the format of the data being transmittedto the external device is:
- XXXXX^^^<CR><LF>
When the digital output is configured for RS232 Full Data, the format of the data being transmitted to theexternal device is:
- XXXXX^YY <CR><LF>
<LF> = Line Feed
<CR> = Carriage Return
^^^ = 3 spaces (Note: ^^^ are NOT shown)
X X X X X = 5-character Force Value with decimal point
SPACE = Compression Reading (C)- = Tension Reading (T)
<LF> = Line Feed
<CR> = Carriage Return
YY = Units of Measure (2 characters)
^ = 1 space (Note: ^ is NOT shown)
X X X X X = 5-character Force Value with decimal point
SPACE = Compression Reading (C)- = Tension Reading (T)
- 14 -
2.2.2.1 Enabling MITUTOYO Digital OutputTo enable the MITUTOYO digital output, turn the gauge power OFF.
- NOTICE -You must power down the gauge to configure the digital output and
communication specifications for the gauge.
To enable the MITUTOYO output, with gauge power OFF, DEPRESS and HOLD the XMIT key whilesimultaneously turning the gauge power switch to the ON position. Release the XMIT key after approxi-mately 2 seconds.
The gauge will display the current communication parameters for the gauge. The parameters will bedisplayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity.
- NOTICE -Your gauge is shipped with the following factory settings:
Baud Rate = 9600Units = 0 1 (Transmit with Units)
Word Length/Stop Bits = 7 - 2 (7 bit word/2 stop bitsParity = -P (No Parity)
To select the MITUTOYO output, change the BAUD RATE to 000H by depressing the UNITS key until 000His displayed. Once 000H is displayed (000H will flash), depress the XMIT key to accept and proceed to thenext communication parameter. If you do not need to select a new parameter, the display will first flash0000, then display the normal operating display to begin testing.
XMIT0 0 0 H
UNITS
- 15 -
2.2.2.2 Enabling RS232 Digital Output- Numerical Data OnlyTo enable the RS232 digital output with numerical data only, turn the gauge power OFF.
- NOTICE -You must power down the gauge to configure the digital output and
communication specifications for the gauge.
To enable the RS232 NUMERICAL DATA ONLY output, with gauge power OFF, DEPRESS and HOLD theXMIT key while simultaneously turning the gauge power switch to the ON position. Release the XMIT keyafter approximately 2 seconds.
The gauge will display the current communication parameters for the gauge. The parameters will bedisplayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity.
- NOTICE -Your gauge is shipped with the following factory settings:
Baud Rate = 9600Units = 0 1 (Transmit with Units)
Word Length/Stop Bits = 7 - 2 (7 bit word/2 stop bitsParity = -P (No Parity)
To select the RS232 NUMERICAL DATA ONLY output, change the UNITS to 0 2 by depressing the UNITSkey until 0 2 is displayed. Once 0 2 is displayed (0 2 will flash), depress the XMIT key to accept andproceed to the next communication parameter. If you do not need to select a new parameter, the displaywill first flash 0000, then display the normal operating display to begin testing.
XMIT0 2
UNITS
- 16 -
2.2.2.3 Enabling RS232 Digital Output- Full Data (Default)To enable the RS232 digital output with full data, turn the gauge power OFF. Full data means that UNITS willbe included with your force value during transmission to an external device.
- NOTICE -You must power down the gauge to configure the digital output and
communication specifications for the gauge.
To enable the RS232 FULL DATA output, with gauge power OFF, DEPRESS and HOLD the XMIT key whilesimultaneously turning the gauge power switch to the ON position. Release the XMIT key after approxi-mately 2 seconds.
The gauge will display the current communication parameters for the gauge. The parameters will bedisplayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity.
- NOTICE -Your gauge is shipped with the following factory settings:
Baud Rate = 9600Units = 0 1 (Transmit with Units)
Word Length/Stop Bits = 7 - 2 (7 bit word/2 stop bitsParity = -P (No Parity)
To select the RS232 FULL DATA output, change the UNITS to 0 1 by depressing the UNITS key until 01 is displayed. Once 0 1 is displayed (0 1 will flash), depress the XMIT key to accept and proceed to thenext communication parameter. If you do not need to select a new parameter, the display will first flash0000, then display the normal operating display to begin testing.
XMIT0 1
UNITS
- 17 -
2.2.2.4 Modifying Communication ParametersYou may change the communication parameters for the force gauge by sequencing through the communi-cation setup display.
- NOTICE -You must power down the gauge to configure the digital output and
communication specifications for the gauge.
To modify any communication parameter, with gauge power OFF, DEPRESS and HOLD the XMIT keywhile simultaneously turning the gauge power switch to the ON position. Release the XMIT key afterapproximately 2 seconds.
The gauge will display the current communication parameters for the gauge. The parameters will bedisplayed in the following order: Baud Rate, Units, Word Length/Stop Bits and Parity.
- NOTICE -Your gauge is shipped with the following factory settings:
Baud Rate = 9600Units = 0 1 (Transmit with Units)
Word Length/Stop Bits = 7 - 2 (7 bit word/2 stop bitsParity = -P (No Parity)
To change any communication parameter, gain access to a parameter by depressing the XMIT key. BAUDRATE will be the first parameter displayed, followed by UNITS, WORD LENGTH, STOP BITS and PARITYin this order respectively.
To change a parameter, once you have sequenced to the parameter, depress the UNITS key until youdisplay the option desired for that specific parameter. To enable the parameter, depress the XMIT key.The selected parameter will flash. You must then depress the XMIT key to go to the next parameter. Youmust sequence through each parameter regardless of whether or not you will be making a change. Afterthe PARITY parameter, the display will flash 0000 and then display the normal operating display.
XMIT9 6 0 0
UNITS
- 18 -
2.3 DATA COLLECT MODEThe DFIS/AF2Series can accept and execute commands through the RS-232 serial port. The command setis tailored to make it easy to configure and operate the instrument under computer control.
2.3.1 About Data Collect ModeData transmitted by the gauge in Data Collect mode is a filtered average of 45 instantaneous force readings.Digital filtering is employed to eliminate false readings resulting from noise and vibration.
The external device, which the gauge is communicating to, e.g. personal computer, must be configured totransmit and ASCII character “F” to the gauge. This signals the gauge to go into, or out of DATA COLLECTmode.
When the external device sends an ASCII character “?” or “X” to the gauge, the gauge will always respondeven if the instantaneous reading is a zero value. This permits real time data collection to begin at a forcereading or zero value. The ASCII character “?” or “X” is a request to the gauge to transmit data.
The gauge will return to Normal mode when it receives and ASCII character “F”. The table on the next pageshows the available ASCII characters and their functions.
The diagram below will illustrate how the gauge operates based on these ASCII characters.
C
D
D
T
N
D
NormalMode
TensionPeak Mode
CompressionPeak Mode
Data Collect Mode Data Collect Mode
Data Collect Mode
ASCII“P”
ASCII“P”
ASCII“P”
ASCII“P”
ASCII“P”
ASCII“P”
ASCII“F”
ASCII“F”
ASCII“F”
PC
ASCII“?” or “X”
Figure 2.3Data Collect Mode and ASCII Command
Relationship
- NOTICE -Each ASCII character transmission will advances the gauge one step, similar to a keypad entry.
- 19 -
The DFIS/AF2Series can accept and execute commands through the RS-232 serial port. The command setis tailored to make it easy to configure and operate the instrument under computer control. The stringcommands are sent as ASCII characters. The following string commands are recognized:
Command Response Description
Legend:^ is space (20 hex);9 is a number 0 through 9;± is a space or a minus;<CR> is Carriage Return (0D hex);<LF> is Line Feed (0A hex).
^^^^^^^^^^<CR><LF> On power-up, this line is sent.
A Sends currently selected units to externaldevice.
F Toggles between Data Collect and theNormal and Peak Modes
P Steps through Normal and Peak Modes:Normal, Tension Peak, Compression Peak
R Resets the gauge: zeroes all modes.
S Sends currently selected mode: Normal,Tension Peak, or Data Collect
U Steps through Units in the following order:lb, kg, N
X or ? Sends data on display with the position ofthe decimal place the same as on display.If in Data Collect Mode, sends the DataCollect filter instead. If “Transmit Units” isset to transmit without units, lb, kg, and Nare replaced by ^^.
Response during force overload.
Z Zeroes the currently selectedmode: Normal, Tension Peakor Compression Peak.
Unit^=^lb^<CR><LF>
Unit^=^kg^<CR><LF>
Unit^=^N^^<CR><LF>
^N-MODE^^^<CR><LF>
TP-MODE^^^<CR><LF>
CP-MODE^^^<CR><LF>
DC-MODE^^^<CR><LF>
±99.99^lb<CR><LF>
±99.99^kg<CR><LF>±999.9^N^<CR><LF>
ERROR^^^^^<CR><LF>
- 20 -
3.0 CALIBRATING THE DFIS/AF2 SERIESThe DFIS/AF2 Series may be field calibrated, however, for optimum performance, it is recommended thatyour gauge be factory calibrated at least every other year or more frequently depending on your frequencyof use. Field calibration is recommended every 6 to 12 months. A 5-point calibration is required where fivedifferent calibrated weights are applied to the load shaft at increments of 20%, 40%, 60%, 80% and 100% ofcapacity.
q You must have a way to mount the load cell vertically oriented before applying tension and compres-sion force.
q The DFIS/AF2 Series should be calibrated using deadweights representing 100% of the rated full scalewith 20% increments. For example, calibrating a 10 lb. gauge, the deadweights must be in 20%increments for a total of 10 lbs. at correct gravity.
q You must have a safe method of axially applying the deadweights to the load cell in both the tensionand compression directions.
LB
LB5 0
- WARNING -The DFIS/AF2 Series should always be calibrated to full capacity. A 500 lb capacity gauge should be
calibrated at a full scale reading of 500 lbs. If you have a 500 lb capacity gauge, yet you use thegauge between 0 and 100 lbs, YOU CANNOT calibrate the gauge as thought the full load is 100 lbs.
3.1 CALIBRATION PROCEDUREThis section leads you step-by-step through the calibration process for a DFIS/AF2Series:
q STEP 1 Gauge Pre-SetWith gauge power OFF, PRESS and HOLD the UNITS and PEAK key.Switch gauge power to ON. Select the units to be used during thecalibration. Depress the UNITS key to scroll through the availableunits (LB, KG, N). Press ZERO to select the desired unit.
q STEP 2 Select the desired capacity that will be calibrated (2, 10, 50, 100, 200or 500). Depress the UNITS key to scroll through the availablecapacities. Press ZERO to select the desired capacity.
- WARNING -You must depress ZERO at the completion of the calibration procedure to record the calibration data
into the instrument’s electronics. Data is NOT SAVED until the ZERO key is depressed.
If, at any time during the calibration procedure that incorrect data is being displayed, simultaneouslydepress PEAK and ZERO keys to restart the calibration procedure.
- 21 -
q STEP 3 Test Loadcell and A/D ConverterThe display indicates the loadcell reading as a percent of the A/Dspan. Display the % A/D Mode by depressing the UNITS key. The% A/D Mode is displayed and should be below 6.0. Press ZERO.
Note: If the % A/D Mode is greater than 6.0, the loadcell may bedamaged. The gauge should be returned to the factory or the loadcellshould be replaced.
- WARNING -Once the gauge has displayed “-00-” place the gauge flat onto its back side. DO NOT place the gaugein a vertical position and DO NOT mount the gauge to a test fixture. The gauge must be positioned on
its back to remove any effects of gravity prior to the calibration.
q STEP 4 The display should indicate -00-, which indicates zero offset of theloadcell and that the loadcell is ready to be measured. Make surethe gauge is laying flat on its back side. DO NOT place the gaugein a vertical position while zero offset is being measured.
Depress the ZERO key to prepare the gauge for calibration. Allow a5 minute warm up time before proceeding to Step 5.
q STEP 5 You may now position the gauge onto a test fixture (test stand) so thatit is in COMPRESSION mode. You do not need to zero or tare thegauge.
q STEP 6 Add FULL LOAD to the gauge and exercise the loadcell. Remove thefull load weight and repeat this process 2 additional times (Exercisethe loadcell with full load a total of 3 times).
After applying and removing full load for the third time, depress theZERO key.
The display will prompt you to add the full load weight.
Note: A blinking number represents an incorrect weight or a problemwith the load cell or gauge electronics. The gauge should be returnedfor factory service.
If “PEAK” is displayed, the weight is NOT steady. Wait for the weightto stabilize and for the “PEAK” indicator to disappear before going to thenext step. Once stable, depress the ZERO key.
1.6C
7.1C
- 0 0 -
0.00C
LB
100.00C
LB
80.00PEAK
LBC
- 22 -
q STEP 7 The gauge will display the next value of weight to be applied to theloadcell (80% capacity). Add 80% of full load to the gauge.
The display should indicate the same value of the calibrate loadweight. The display reading should not blink.Depress the ZERO key.
Note: A blinking number represents an incorrect weight or a problemwith the load cell or gauge electronics. The gauge should be returnedfor factory service.
q STEPS 8, 9, 10 The gauge will continue to display the next value of weight to beapplied to the loadcell. Repeat the process until all 5-points have beensuccessfully calibrated in COMPRESSION mode.
q STEP 11 Remove the gauge from the test fixture and from Compression modeand re-orient the gauge into TENSION mode.
Note: The display will indicate TENSION mode by a blinking “T”,blinking Units or blinking decimal point.
A blinking “C” indicates a COMPRESSION verify mode.
Repeat STEPS 6-10 with the gauge in TENSION mode.
q STEP 12 After the gauge has been calibrated in both Compression and Tensionmodes, depress the ZERO key to exit the calibration process andrecord calibration data. The gauge is now calibrated.
80.00LB
C
- WARNING -You must depress ZERO at the completion of the calibration procedure to record the calibration data
into the instrument’s electronics. Data is NOT SAVED until the ZERO key is depressed.
If, at any time during the calibration procedure that incorrect data is being displayed, simultaneouslydepress PEAK and ZERO keys to restart the calibration procedure.
- 23 -
4.0 DIAGNOSTICSIn the Self Test operation, the DFIS/AF2 Series performs a series of self-diagnostic tests. During selfdiagnostics, the gauge will display “ . . .” and then the following information:
- NOTICE -You must power down the gauge to perform self diagnostic tests.
To enable the SELF DIAGNOSTIC, with gauge power OFF, DEPRESS and HOLD the ZERO key whilesimultaneously turning the gauge power switch to the ON position.
ZERO. . .
To display the SOFTWARE PART NUMBER used with the gauge, depress the ZERO key. A four digitnumber will appear representing the software part number.
ZERO0877
ZERO-2.08
To display the SOFTWARE REVISION NUMBER used with the gauge, depress the ZERO key. The gaugewill display the revision number, e.g. Revision 2.08.
4.1 SOFTWARE PART NUMBER
4.2 SOFTWARE REVISION LEVEL
- 24 -
ZERO 10
To display the CAPACITY and UNITS used with the gauge, depress the ZERO key. The gauge will displaythe capacity and units used in the previous and most current calibration.
- WARNING -If no capacity or calibration data is found in the gauge’s EPROM, the display with blink “-EL”. This
indicates that the gauge must be recalibrated.
-EL
ZERO0.0
To display the NO LOAD ZERO, depress the ZERO key. The gauge will display the no load zero measuredduring the previous and most current calibration. The display will indicate the no load zero reading in % ofA/D range).
C
- WARNING -You must place the gauge on its back side in a horizontal position.
ZERO- 00 -
Place the gauge on its back side in a horizontal position, depress the ZERO key. The gauge will displaya blinking “-00-” then display the applied force in % of A/D range until ZERO is depressed again.
Press ZERO to display the current no load value.
4.3 CAPACITY AND UNITS
4.4 NO LOAD ZERO VALUE
LB
- 25 -
ZERO- 0 -
Depress the ZERO key and the gauge will compare the current and stored no load offsets (zeros). If thegauge displays “-0-”, the gauge is within specification. If another value is displayed, the current offset hasshifted too far from the stored offset and recalibration is required. The difference between the currentoffset and tolerance is displayed as a % of A/D range.
ZEROL00P
Depress the ZERO key and the gauge will transmit an ASCII character “U” (See Section X.X Data CollectMode) back to itself via an RS232 loopback connector (Not included). Depress ZERO again to exit theloopback test.
With a loopback connector (Txc jumpered to Rxd); depress XMIT to send a continuous “U”. The gauge willdisplay a count of the number of transmission errors. The display will blink. Depress XMIT and theloopback test will toggle ON (Blinking) and OFF (steady).
ZERO1
Depress the ZERO key and the gauge will perform a KEYPAD test. The gauge will display the number ofeach key that was depressed: UNITS = 1, PEAK = 2, XMIT = 3, and ZERO = 4.
Turn power OFF to exit diagnostics mode.
4.5 RS232 LOOPBACK TEST
4.6 KEYPAD TEST
- 26 -
5.0 TECHNICAL REFERENCE & SPECIFICATIONS
AccuracyDFIS Series: +0.15% of Full Scale +1 LSCAF2 Series: +0.2% of Full Scale +1 LSC
Tare Capacity10% in order to utilize Full Scale (in lb Mode); cantare more than 10%, however user may not haveuse of gauge’s Full Scale capacity during testing
Deflection0.007 +0.003 for loadcell
Safe OverloadGauge will display “- - - -” when the force appliedexceeds approximately 120% of the gauge’s ca-pacity
OverloadMaximum overload is 150%. Load Cell deform-ation may occur when overload exceeds 150%.Contact your local AMETEK Representative if youexperienced an overload beyond 150%.
Analog Output-1.50 (Min) to 1.50Vdc (Min) at Full Scale
Digital OutputRS-232 and MITUTOYO
Display Update8 updates per second in Normal or Peak Mode
Sampling Rate1,000 samples per second
Battery Life8-10 hours of continuous operation betweencharges
Temperature Range40 to 110oF (5 to 45oC)
Temperature Stability0.06% per oF
Warranty1 year
- 27 -
6.0 MODEL NUMBERS
Model Capacity
DFIS2 2 lb x 0.002 lb1 kg x 0.001 kg
10 N x 0.01 N
DFIS10 10 lb x 0.01 lb5 kg x 0.005 kg
50 N x 0.05 N
DFIS50 50 lb x 0.05 lb25 kg x 0.02 kg
250 N x 0.2 N
DFIS100 100 lb x 0.1 lb50 kg x 0.05 kg
500 N x 0.5 N
DFIS200 200 lb x 0.2 lb100 kg x 0.1 kg
1000 N x 1 N
DFIS500 500 lb x 0.5 lb250 kg x 0.2 kg
2500 N x 2 N
Model Capacity
AF2-002 2 lb x 0.002 lb1 kg x 0.001 kg
10 N x 0.01 N
AF2-010 10 lb x 0.01 lb5 kg x 0.005 kg
50 N x 0.05 N
AF2-050 50 lb x 0.05 lb25 kg x 0.02 kg
250 N x 0.2 N
AF2-100 100 lb x 0.1 lb50 kg x 0.05 kg
500 N x 0.5 N
AF2-200 200 lb x 0.2 lb100 kg x 0.1 kg
1000 N x 1 N
AF2-500 500 lb x 0.5 lb250 kg x 0.2 kg
2500 N x 2 N
CHATILLON® Brand AMETEK® Brand
- 28 -
Pub No. NC001562Issued 05/01 Printed in U.S.A.Copyright 2001, by AMETEK, Inc.
AMETEK and CHATILLON are registeredtrademarks of AMETEK, Inc.LLOYD INSTRUMENTS is an AMETEK Inc.trademark
Information within this document issubject to change without notice.
ISO 9001Manufacturer
Internet Addresses:
www.ametek.com
www.chatillon.com
www.lloyd-instruments.co.ukTEST AND CALIBRATION INSTRUMENTS
AMETEK Test and Calibration
Instruments Division
8600 Somerset Drive
Largo, Florida 33773
USA
Tel +1-727-536-7831
Tel +1-800-527-9999 (USA only)
Email: [email protected]
Fax +1-727-539-6882
AMETEK Lloyd Instruments
No. 7 Sherwood Place
Alexander Heights
6064 Perth
Australia
Tel +61-8-9343-5725
Fax +61-8-9343-5723
AMETEK Singapore Pvt. Ltd.
10 Ang Mo Kio Street 65
#05-12 TECHPOINT
Singapore
569059
Tel +65 484 2388
Fax +65 481 6588
AMETEK Lloyd Instruments
Forum House
12 Barnes Wallis Road
Segensworth East
Fareham
Hampshire PO 15 5TT
Tel +44 (0) 1489-486399
Fax +44 (0) 1489-8851118
AMETEK Lloyd Instruments SA
3 Avenue des Coudriers
Zone d’Activite de l’Observatoire
7810 Montigny-Le-Bretonneux
France
Tel +33-1-3057-4774
Fax +33-1-3057-5033
AMETEK Precision
Instruments Europe GmbH
Rudolf-Diesel-Strasse 16
D-40670, Meerbusch
Germany
Tel +49-2-159-9136 0
Fax +49-2-159-9136 39
Americas Europe Asia Pacific