Charge collection measurement on slim edge sensors with the ALiBaVa system

25
Charge collection measurement on slim edge sensors with the ALiBaVa system Riccardo Mori*, Matteo Cartiglia, Hartmut Sadrozinski**, Mara Bruzzi*, Vitaliy Fadeyev**, *Dipartimento di Energetica, Università degli Studi di Firenze Istituto Nazionale di Fisica Nucleare, sezione di Firenze ** Santa Cruz Institute of Particle Physics, University of California Santa Cruz

description

Charge collection measurement on slim edge sensors with the ALiBaVa system. Riccardo Mori*, Matteo Cartiglia, Hartmut Sadrozinski**, Mara Bruzzi*, Vitaliy Fadeyev**, *Dipartimento di Energetica, Università degli Studi di Firenze Istituto Nazionale di Fisica Nucleare, sezione di Firenze - PowerPoint PPT Presentation

Transcript of Charge collection measurement on slim edge sensors with the ALiBaVa system

Page 1: Charge collection measurement on slim edge sensors with the ALiBaVa system

Charge collection measurement on

slim edge sensors with the

ALiBaVa systemRiccardo Mori*, Matteo Cartiglia, Hartmut Sadrozinski**, Mara Bruzzi*, Vitaliy Fadeyev**,

*Dipartimento di Energetica, Università degli Studi di Firenze

Istituto Nazionale di Fisica Nucleare, sezione di Firenze

** Santa Cruz Institute of Particle Physics, University of California Santa Cruz

Page 2: Charge collection measurement on slim edge sensors with the ALiBaVa system

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. Christophersen

RD50, 21-23 November 2011, CERN

Introduction

ALiBaVa[1] is a portable read-out system developed for test sensors in (s)LHC-like experiments. It provides a real-time GUI and data format for custom (ROOT) analysis.

There is the interest on the SNR and the efficiency of detectors. In particular it is interesting to know the performances of slim edges (cut) detectors, compared to the uncut ones.

It is interesting to study the relation between the occupancy and the efficiency of a detector by a threshold scan:

To vary the threshold mean to scale the occupancy by a factor proportional to the integral of the distribution[2].

Our objectives are to study the signal distributions and to characterize slim edge sensors:

– comparing cut and uncut detectors;

– comparing inner strips with edge-strips (outer strips).[1]Marco-Hernandez, IEEE TNS, 2009; [2] Spieler, Introduction to radiation detectors and electronics, 1998, ch. VIII.6.

Page 3: Charge collection measurement on slim edge sensors with the ALiBaVa system

Materials Sensors:

(CMS, p-on-n, unirradiated, 128 strips.) GLAST cut (Si oxide passivated, laser-scribed and cleaved), p-on-n,

unirradiated, 8 strips. (Thanks to Bernard Phlips and Marc Christophersen!) GLAST uncut, p-on-n, unirradiated, 8 strips.

96 um 345 umR. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. Christophersen

RD50, 21-23 November 2011, CERN

Page 4: Charge collection measurement on slim edge sensors with the ALiBaVa system

Experimental set-up

Set-up:

– 90Sr/90Y (disintegration energies 0.5/2.3 MeV, half-life 29y/64h).

– Scintillator + photomultiplier (PM) tube from UCSC.

– ALiBaVa system from INFN Florence. Settlings:

– Vbias=[1,10,20,50,100,150,200,500] V.

– Beetle settlings: high gain, “Kazu” configuration.

– Noise measurements: detector not in front of the source.

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 5: Charge collection measurement on slim edge sensors with the ALiBaVa system

Analysis

Noise measurement analysis: Fit of the distribution to get the sigma. Occupancy threshold scan and fits to cross-check.

Signal measurements analysis: Time cut. Clustering: amplitude cut for seed and neighbours. Fit of the distribution to get the MPV and a sigma of noise.

=> SNR=MPV/sigma (from noise distribution fit). Analysis vs. bias voltage. Analysis vs. channel index: look at the edges!

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

ADC

#

#

ADC

Page 6: Charge collection measurement on slim edge sensors with the ALiBaVa system

Experimental results

GLAST cut (inner channels): Noise: distribution

Sigma=4.46 ADCs

Vbias=200V

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 7: Charge collection measurement on slim edge sensors with the ALiBaVa system

GLAST cut (inner channels): Noise: occupancy threshold scan.

Sigma=6.34 ADCs

Experimental results

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 8: Charge collection measurement on slim edge sensors with the ALiBaVa system

GLAST cut (inner channels): Signal: in time and distributions

MPV=103.62*1.013 ADCs (from the CMS, expected 107.33!).Sigma(Gaussian)=18.62 ADCs.

Experimental results

SNR=103.26*1.013/4.46=23.64

Time cut: [25,45]nsCorrection factor: 1.013.

Without performing the time cut

Vbias=200V

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 9: Charge collection measurement on slim edge sensors with the ALiBaVa system

GLAST cut (inner channels): Results versus bias voltage

sigma versus bias voltage

Experimental results

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 10: Charge collection measurement on slim edge sensors with the ALiBaVa system

Comparison inner-outer channels: the anomaly of the outer channels

Experimental results

Noise distributions Signal distributions

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Outer noise larger because of the interstrip capacitance.

Outer signals have a tail!?!

Page 11: Charge collection measurement on slim edge sensors with the ALiBaVa system

Comparison inner-outer channels: the anomaly of the outer channels

Adding the neighbours...

Experimental results

An inner channel without adding the neighbours...

The highest neighbours distribution: the tail!!!

The outer channels has the guard ring as neighbors, which is not read out.

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 12: Charge collection measurement on slim edge sensors with the ALiBaVa system

Comparison inner-outer channels: reconstructing the outer channels

Experimental results

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 13: Charge collection measurement on slim edge sensors with the ALiBaVa system

Comparison cut-uncut: Noise:

Inner noise

Experimental results

About the same sigma!

Vbias=150V

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 14: Charge collection measurement on slim edge sensors with the ALiBaVa system

Comparison cut-uncut: Signal:

Experimental results

(cut side)

About the same pulse height!

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 15: Charge collection measurement on slim edge sensors with the ALiBaVa system

Discussion

Limits: No calibration. No separated pedestal run.

On the results: Width of the distribution:

Noise. Dispersion of the path length. Natural broadening of the Landau described by Meroli[4].

MPV decrease from 200V to 500V. The border strips clearly have a tail due to the charge sharing with

the bias ring, in general due to the contribution as a neighbours of channels unread-out.

[4]Meroli, http://meroli.web.cern.ch/meroli/Lectures_StragglingFunction.html

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 16: Charge collection measurement on slim edge sensors with the ALiBaVa system

Conclusions

Work done: Tested ALiBaVa. Tested detectors: SNR, efficiency. Compared cut-uncut.

Results: The cut sensor is as good as the uncut one: about the

same sigma and the same pulse height. Future works:

Calibration. Test of other sensors (GLAST wonder baby, p-type).

Acknowledgements In particular, to Nicola Pacifico.

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 17: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

I-V cut sensor. CMS sensor results. GLAST uncut sensor results. Results vs. channel index. Dispersion of the path length. Warnings: “Teppo's effect”. Set-up.

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 18: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

I-V of sensors:

0 100 200 300 400 500 600 700 800 900

-1.00E-05

0.00E+00

1.00E-05

2.00E-05

3.00E-05

4.00E-05

5.00E-05

6.00E-05

GLAST Skinny Cutoff IV Piece 04550

Column H

V bias

Cu

rren

t

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 19: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

I-V of individual channels:

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 20: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

CMS sensor results:

sigma=3.3 ADCs

SNR≈25

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 21: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

GLAST uncut sensor results:

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 22: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

Results vs. channel index (all channels):

Cut sensorVbias=150V

Uncut sensorVbias=150V

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 23: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

Dispersion of the path length:L m

in

L M

AX

L min=400umL MAX=460.42um

Assuming: MPV(L min)=100 ADCs=>MPV(L MAX)=115.10 ADCs=> Dispersion of 15.1 ADCs

=>ENOUGH TO EXPLAIN!

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 24: Charge collection measurement on slim edge sensors with the ALiBaVa system

Spares

Warnings: “Teppo's effect”[3]:

Be careful on the settling of the threshold: with a too high threshold, a Landau still appear and the MPV is biased (higher respect to the real).

[3] Maenpaa, Ph.D thesis, 2011.

Normal cutMPV=89.80 ADCs

Cutting moreMPV=92.28 ADCs

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN

Page 25: Charge collection measurement on slim edge sensors with the ALiBaVa system

Daughter board

Spares

Sr sensorScint. PM tube

Mother board

PSPS

PC(ROOT)

Set-up:

R. Mori, M. Cartiglia, H. Sadrozinski, M. Bruzzi, V. Fadeyev, B. Phlips, M. ChristophersenRD50, 21-23 November 2011, CERN