Chapter 4(Equipment and Accessories)

download Chapter 4(Equipment and Accessories)

of 44

Transcript of Chapter 4(Equipment and Accessories)

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    1/44

    Equipment andAccessories

    (M. Anees Akhtar)

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    2/44

    Vacuum Boxes Vacuum Boxes are used for Bubble Leak testing in the

    field

    Vacuum boxes are available for rounded surfaces,

    corner seams and vertical seams. The box should be able to withstand an external pressure

    of 100 kPa.

    Flexible gaskets are provided to seal the enclosure to thetest surface when pressure is applied to the vacuum box.

    The vacuum box should be of convenient size such as 15cm by 75 cm (6 inches wide by 30 inches long).

    Each box contains a window in its top opposite the openbottom.

    Suitable connections, valves, lighting and gauges arethe accessories of the vacuum boxes.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    3/44

    Vacuum Boxes

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    4/44

    Vacuum Boxes

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    5/44

    Desirable features of Vacuum Boxes Desirable features for a vacuum box are. Ability to readily admit natural or artificial light Close proximity of viewing window to the surface of the

    weldment being inspected. Capability for easy initial seating when starting evacuation,

    and good sealing properties to hold the vacuum.

    Equipped with a highly portable, readily availableevacuation device with sufficiently high flow to be able toovercome a certain amount of seal leakage and rapidlyevacuate the box to the required pressure differential.

    Easily read dial gauge Quick acting valve for either shutting off the evacuation

    device or controlling the level of pressure differential. A vacuum box for bubble leak testing should be

    checked for workability before performing leak testing todetermine that the condition of the box components issuch that it is capable of attaining a higher pressuredifferential than is required for the test.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    6/44

    Application of Foaming Solutions following are the methods listed in order of

    decreasing preference

    Spray application using a pump-type garden spray

    can with a fine orifice.

    Spray application using a plastic squeeze bottle or anoil squirt can

    Brush application using a short handle painters brush,

    1 to 3 cm in width.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    7/44

    Gaskets The gasket system used with a vacuum box is

    critical to the ease with which the box can behandled and sealed to the test surface so as to be

    able to hold a vacuum. The gasket must be shaped so that the initial area

    of the gasket contact with the test surface is smallto make seating of the vacuum box easier.

    Neoprene gaskets of 20 to 40 durometer providegood flexibility.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    8/44

    Dial Gauges for Vacuum BoxA pressure or vacuum dial gauge must be readily

    visible to the operator controlling the pressure withinthe vacuum box during leak testing.

    Pressure indicating gauges used in testing shouldpreferably have dial graduations covering a rangeof about double the intended maximum pressure.

    In no case should the range be less than one and

    one half not more than four times that pressure.All gauges used shall be calibrated against a

    standard dead weight tester a calibrated mastergauge, or a mercury column, and recalibrated atleast once each 12 months.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    9/44

    Pirani Guages

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    10/44

    Pirani Guages Pirani gauges employ a wheatstone bridge circuit,which serve to heat a filament and to balance itsresistance against a standard resistor sealed off in highvacuum.

    A change of pressure causes a change of filamenttemperature and, consequently, of the filamentresistance, thus unbalancing the bridge.

    The pressure can then be measured in terms of theunbalanced voltage. Alternatively, the power required

    to maintain the filament temperature at a constant levelis a measure of pressure. The temperature in this case iskept constant by means of a feedback circuit.

    produce fairly accurate absolute pressure readings fromatmosphere to 0.1 or 0.01 Pa (1.0 or 0.1 mtorr).

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    11/44

    Hot Cathode Ionization Guage The high filament temperature (1500 C to 2200 C)causes electrons to escape and become free electrons.

    The free electrons create gas ions as they acceleratetowards positively charged grid.

    They are absorbed by the grid; forming electron currentcalled emission current.

    Gas ions are attracted to negatively charged ioncollector where they pick up electrons and becomeneutral molecule. This neutralizing electron current iscalled ion current.

    The magnitude of the ion current is an indicationof pressure.

    A conventional hot cathode ionization gauge measurespressures in the range of 10-3to 10-10mbar.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    12/44

    Hot Cathode Ionization

    Gauge(BayardAlpert Ionization

    gauge)

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    13/44

    Cold Cathode Ionization Gauge

    (Penning Gauge) Measuring Range

    10-210-7mbar The cold cathode ionization gauge is basically a direct current (dc) glow

    discharge tube . The tube has two metallic electrodes(an anode and a cathode), one at

    each end of the tube. A dc voltage is applied in series with a large current limiting resistor. The

    resistor ensures discharge stability and prevents excessive current flow. To strike a discharge, an initial supply of electrons is needed from the

    cathode. Field emission occurs when the cathode is held at a high voltage and

    electrons are freed. Once electrons are released from the cathode, they accelerate towards

    the anode creating positive gas ions by impact ionization. The positive

    ions accelerate towards the cathode. The energy from positive ionsimpacting the cathode releases an additional supply of secondaryelectrons sufficient in quantity to sustain the discharge process. Thequantity of electrons released by ion impacts on the cathode dependson the number of ions produced in the discharge.

    The number of ions produced depends on the gas pressure in the tube. Measuring the ion current at the cathode is an indication of pressure.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    14/44

    cold Cathode Ionization Gauge

    (Penning Gauge)

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    15/44

    Radioactive Ionization Gauge

    The ions are produced by radiations (usually alphaparticles) emitted from a radioactive source.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    16/44

    Evacuation system for Vacuum Box

    Leak TestingA small box mounted air ejector connected to a

    compressed air supply.

    A small portable vacuum pump.

    The vacuum box, when placed over theexamination area, should be evacuated at least35 kPa pressure differential with respect toatmospheric pressure.

    The 35 kPa pressure differential can be verified bythe dial gauge.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    17/44

    Lighting Good lighting is necessary for the safety and reliability of

    the test results.

    Two general sources of lighting:

    Artificial Lighting

    For continuous shift operation, artificial light is essential.

    For other types of operation, it must be relied upon from 20to 50% of the total working hours, excluding overtime work ornight work.

    General Lighting

    General lighting is the base or minimum amount of light

    required. The uniform distribution of light to produce equivalent seeing

    conditions throughout an interior.

    Localized general lighting sources are arranged 3 meter ormore above the work to provide sufficient light for generalsafety and for ordinary visual needs.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    18/44

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    19/44

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    20/44

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    21/44

    Absolute Pressure Dial Guages Aneroid Capsules Upto 350 kPa

    Bourdon Tube PressureGauge Above 350 kPa

    The ordinary dialpressure gauge hastypical accuracies inthe range of 0.25 to

    0.33% of full scaleindication whenrecently calibrated.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    22/44

    Barometers

    Can be used to measureabsolute pressure.

    The height of mercurycolumn indicatesatmospheric pressure.

    These pressuremeasurements are absolute,because the pressure insidethe sealed end of the tubeis always zero.

    There is only vacuum abovethe mercury column, sothere is no contribution dueto air pressure.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    23/44

    Instruments for Pressure

    Measurement Types of pressure gauges that provide visible

    indications during leak testing include;

    Absolute pressure dial gauges.

    Aneroid barometers.

    Ordinary dial gaugesindicating pressure relative to

    ambient atmospheric pressure as well as water.

    Manometer , UTube mercury manometers and

    mercury column barometer.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    24/44

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    25/44

    Instruments for Temperature

    MeasurementDew point sensor

    The dew pointis the temperature below whichthe water vapor in a volume of humid air at a

    constant barometric pressure will condense into liquidwater.

    Vapor pressure due to moisture adds to the totalpressure measured by the pressure measurementinstruments.

    It works on the principle of balancing the moisturecontent in the element with the moisture content ofthe surrounding atmosphere.

    Used to test volume subject to pressure changeleakage rate measurement tests.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    26/44

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    27/44

    Vacuum Pumps

    Vacuum pumps are classified into two

    groups Gas transfer pumps

    Capture pumps

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    28/44

    Low and Medium Vacuum

    Pumps Rotary Vane Pump

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    29/44

    High and Ultra High Vacuum

    PumpsDiffusion Pump

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    30/44

    High and Ultra High Vacuum

    Pumps Turbomolecular Pump

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    31/44

    Vacuum ValvesValves are required to isolate either part or all

    of the pumping group from the system which isbeing evacuated or are necessary to control

    the pumping sequence.Vacuum valves must;

    be free from leakage.

    Offer minimum flow resistance and

    Contain materials that do not outgas.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    32/44

    Vacuum Valves

    Diaphragm Valves Gate Valve

    Closed

    Open

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    33/44

    Vacuum Connector O Rings

    The O ring are made of an elastomer materialwhich is trapped and compressed between two

    surfaces to form the seal. Demountable joints for vacuum in the range

    from atmospheric pressure down to 10-7mbarnormally utilize O ring for the sealing method.

    Grooved flanges are used to issue the leak

    tight system and less compression suchgrooves should meet the followingrequirements.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    34/44

    Vacuum Connector O Rings

    The O ring are made of an elastomer material

    which is trapped and compressed between two

    surfaces to form the seal. Demountable joints for vacuum in the range

    from atmospheric pressure down to 10-7mbar

    normally utilize O ring for the sealing method.

    Grooved flanges are used to issue the leaktight system and less compression.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    35/44

    Vacuum Connector Groovs in the flanges should meet the following

    requirements. The cross-section of the groove should be of the order of 10

    percent greater than the cross section of the O ring, toallow for a metal to metal flange contact withoutovercompressing the O ring.

    To provide positive retention of the gasket, when the seal isnot made.

    To provide the minimum volume of trapped gas.

    To allow reuse of the O ring after the seal is broken. In thisrespect, overcompression or grooves with sharp corner oracute angles are of bad design as this leads to permanent

    deformation or cutting of the O ring. To permit easy flange machining end to wide machine

    tolerance.

    Orings generally provide better sealing than flat gaskets.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    36/44

    Vacuum Connector Hinged Clamps

    These clamps are

    used for quickrelease

    applications

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    37/44

    Mass Spectrometer Common functions of commercially available

    mass spectrometer leak detector instruments

    are: Pumping tracer gas samples from leaks in test

    objects into the vacuum of the instrument.

    The ionization of gas sample molecules by

    electron impact.

    The sorting and identification of positive ions

    according to their mass-to-charge ratios.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    38/44

    Mass Spectrometer Tube

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    39/44

    Sniffer It is a simple tracer gas

    collection device

    It is connected to leak

    detector with a pumping

    device. The pump provides the

    suction necessary to

    draw tracer gas samples

    to the sniffer inlet through

    the connecting hose andinto the sensing element

    of the leak detector.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    40/44

    Compound mastics for sealing to vacuum

    Apiezon Type

    (hydrocarbon)Silicone type Fomblin

    Good lubricant Yes No Yes

    Vapour pressure (mbar) Low Approx. 10-6 Low

    Temperature suitability 125C max. -40 to 200C -02 to 200C

    Cost Moderate Low High

    Chemical stability Moderate Good Very good

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    41/44

    Tuning and calibration Tuninga helium leak detector is a procedure

    of adjusting the electronic controls of theunit to achieve maximum signal strength,highest reading on the leak rate meter whileadmitting a known flow of helium to the leak

    detector.Calibrationof the instrument merely

    correlates the output signal with a knownflow of helium standard calibration leak.

    Calibration is usually accomplished byadjusting the gain, or amplifier feed back,until the leak rate meter reading agrees withthe value stamped on the standard leak.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    42/44

    Calibrated Leak Standards

    Reservoir Calibrated Leaks

    Reservoir leaks contain their own tracer gas

    supply.

    NonReservoir calibrated Leaks

    Non-reservoir leaks to which tracer gas is

    added during testing.

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    43/44

    Classification of Calibrated

    Leaks

    LEAKS

    Reservoir

    Permeation Capillary

    Fixed ValueVariable

    Value

    Porous Plug

    Non -Reservoir

    Capillary

    Fixed ValueVariable

    Value

    Porous Plug

  • 8/13/2019 Chapter 4(Equipment and Accessories)

    44/44

    Properties of calibrated LeaksAn ideal calibrated leak should have the

    following properties:

    The leakage rate should be constant, and should

    remain unaffected by ambient conditions. The calibration should be accurate.

    The physical size should be convenient.

    The calibrated leak should not be too delicate or

    fragile. The calibrated leak should have its own gas

    supply.