Chapter 20. Lung Isolation Devices

download Chapter 20. Lung Isolation Devices

of 144

Transcript of Chapter 20. Lung Isolation Devices

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    1/144

    Chapter 20

    Lung Isolation DevicesP.634

    Indications for Lung Isolation

    Tho rac i c P roc edu r es

    A def la ted lung provides better operating conditions and reduced

    trauma during thoracic procedures.

    Con t r o l o f Con t am inat i o n o r Hem o r r hage

    A lung isolation device can prevent infected material from one

    lung from contaminating the other lung (1). W hen hemorrhage

    occurs in one lung, an isolation device allows the unaffected lungto be ventilated (2,3,4,5,6,7,8,9,10,11).

    Un i l a te r al Patho l og y

    A bronchopleural or bronchocutaneous fistula may have such a

    low resistance to gas flow that most of t he tidal volume passes

    through it, making it impossible to adequately ventilate the other

    lung (12,13,14). Large cysts or bullae may rupture under positive

    pressure, making it mandatory that they be excluded from

    ventilation. Another indication for lung separation is when lungs

    have markedly different compliance or airway resistance such as

    that which occurs following single-lung transplantation or

    unilateral injury (15).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    2/144

    The best method of producing lung isolation in a given situation

    will depend on several factors, including the indication for lung

    isolation, patient variables, available equipment, and the skill and

    training of the anesthesia provider.

    Anatomical Considerations

    The right mainstem bronchus is shorter, straighter, and has a

    larger diameter than the left. It takes off from the tr achea at an

    angle of 25 degrees in adults. The left mainstem bronchusdiverges from the median plane at a 45-degree angle. These

    angles are slightly larger in children (16). The right upper lobe

    bronchus takeoff is very close to the origin of the right mainstem

    bronchus. These anatomical features mean that it is usually

    easier to intubate the right mainstem bronchus than the left, but it

    is more difficult to place a tube in the right mainstem bronchus

    without obstructing the upper lobe orifice.

    Double-lumen Tubes

    The double-lumen tube (DLT, DLET) is the device most

    commonly used to provide lung isolation.

    Desc r i p t i o n

    A DLT is essentially two single- lumen tubes bonded together and

    designated either as right- or left-sided, depending on which

    mainstem bronchus the tube is designed to fit. The tracheal

    lumen is designed to terminate above the carina. The distal

    portion of the bronchial lumen is angled to fit into the appropriate

    mainstem bronchus.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    3/144

    The internal lumen of each tube is D-shaped with the straight

    side of the D in the middle of the tube. The resistance of each

    lumen of a 35 to 41 French (Fr) DLT has been found to be

    comparable to the resistance of a size 6 to 7 mm internal

    diameter single-lumen tracheal tube (17).

    The bronchial cuff for right-sided tubes varies in shape,

    depending on the manufacturer. On some tubes, the cuff has a

    slot to allow ventilation of the right upper lobe. Some right-sided

    DLTs have two bronchial cuffs with an opening for the right upper

    lobe between them. The resting volume and compliance of

    bronchial cuffs varies between different sizes and brands of DLTs

    (18,19). Most manufacturers color the bronchial cuff blue. They

    also use blue markings on the pilot balloon and/or the inflation

    device for the bronchial cuff.

    A few DLTs have a carinal hook to aid in proper placeme nt andminimize tube movement after placement. Potential problems

    with carinal hooks include increased difficulty during intubation,

    trauma to the airway, malposition of the tube because of the

    hook, and interference with bronchial closure during

    pneumonectomy. The hook can break off and become lost in the

    bronchial tree.

    Most manufacturers place a radiopaque marker at the bottom of

    the tracheal cuff or at the end of the tracheal lumen. Other marks

    may be placed above and/or below the bronchial cuff. Some have

    a radiopaque line running the length of the tube.

    Disposable DLTs are supplied in sterile packages, which include

    a stylet, connectors, and suction catheter(s). A means to supply

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    4/144

    continuous positive airway pressure (CPAP) may be included with

    the tube or can be purchased separately.

    The connector (20) allows both lumens to attach to a breathing

    system at the same time.

    S iz i ng

    Adult DLTs commonly come in sizes 35, 37, 39, and 41 Fr. The

    French scale is the ex ternal diameter of the tracheal segment

    times three. Some manufacturers also provide 26, 28, and 32 Frtubes for younger patients. Unfortunately, the French gauge

    markings are of limited value in determining the most important

    measurement-the diameter of the bronchial segment.

    There are major variations between manufacturers in the

    dimensions of the bronchial segment of DLTs of the same

    nominal size and even among tubes of the

    P.635

    same size from the same manufacturer (21,22,23,24). An

    International Technical Specification has recommended that

    outside circumference of the bronchial segment be designated in

    millimeters (25). The diameter of the bronchial segment with the

    cuff inflated may not increase with tube size (26).

    Marg i n o f Sa fet y

    The margin of safety for a DLT is the length of the

    tracheobronchial tree between the most distal and proximal

    acceptable positions (27,28). The margin of safety will depend on

    the length of the lumen into which the cuff is placed and the

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    5/144

    length of the cuff. If the cuff is short or the mainstem bronchus

    long, the margin of safety will be greater.

    Left-sided Tubes

    The outermost acceptable position for a left-sided DLT is when

    the bronchial cuff is just below the carina. If the tube were more

    proximal, the bronchial cuff could obstruct the trachea and/or the

    contralateral (right) mainstem bronchus. In this case, the seal

    between the two lungs would be lost. The most acceptable distal

    position is when the the bronchial segment tip is at the proximal

    edge of the upper lobe bronchial orifice. More distal insertion

    would result in obstruction of the upper lobe bronchus.

    The average length of the left bronchus from the carina to the

    takeoff of the upper lobe bronchus is 5.6 cm. This leaves a

    relatively small margin for placement, as there could be up to 3.5

    cm of movement with neck flexion and extension (29). There is

    great variability in the length of the bronchial segment of left-

    sided DLTs currently available (30).

    Right-sided Tubes

    The margin of safety is defined differently for right-sided tubes. A

    right-sided DLT is acceptably positioned if the right upper lobe

    ventilation opening or the space between the two cuffs is aligned

    with the right upper lobe orifice. Thus, the margin of safety is the

    length of the ventilation opening minus the diameter of the

    orifice. The margin of safety for right-sided tubes is considerably

    less than for left-sided tubes.

    Spec i f i c Tub es

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    6/144

    Carlens Double-lumen Tube

    The Carlens DLT (Figs. 20.1, 20.2) is intended to be inserted into

    the left mainstem bronchus. It has a carinal hook (spur). This

    tube may be especially useful with massive hemoptysis when

    verification of tube placement is especially difficult (31).

    View Figure

    Figure 20.1Carlens double-lumen tube. A:Theconnector has ports for fiberscope insertion or suctioning

    and areas where a clamp can be applied to occlude gas

    flow. B:Note the carinal hook and the blue bronchialcuff.

    P.636

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    7/144

    View Figure

    Figure 20.2Carlens tube in place.

    White Double-lumen TubeThe White DLT (Fig. 20.3) is designed to fit the right mainstem

    bronchus. It has a carinal hook. The cuff for the right mainstem

    bronchus is circumferential superior to the o pening to the upper

    lobe bronchus and continues distally behind the opening.

    Robertshaw Right Double-lumen Tube

    The bronchial portion of the Robertshaw right DLT (Fig. 20.4) is

    angled at 20 degrees (32). The bronchial cuff has a slotted

    opening in its lateral aspect (Fig. 20.5). The bronchial cuff is

    proximal to the slot on the lateral surface and extends

    tangentially toward the medial surface.

    Robertshaw Left Double-lumen Tube

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    8/144

    The Robertshaw left DLT is shown in Figures 20.6and 20.7. The

    angle of the bronchial portion is 40 degrees (32). The average

    length of the bronchial segment is 23 mm for sizes 35/37 and 25

    mm for sizes 39/41 (30).

    Broncho-Cath Right-sided Tube

    The Broncho-Cath right-sided tube has a bronchial cuff that is

    roughly the shape of an S, or slanted doughnut. The cuff edge

    nearest the right upper lobe bronchus is closer to the trachea

    than the part of the cuff touching the medial bronchial wall ( Figs.

    20.8, 20.9). A s lot in the tube just beyond the cuff corresponds to

    the opening of the upper lobe bronchus (27). The end of the

    bronchial segment has no bevel (33).

    The shape of t he right bronchial cuff allows the venti lation slot to

    ride off the right upper lobe orifice, increasing the margin of

    safety. One study found that when this tube was inserted blindly,

    right upper lobe obstruction occurred in 89% of cases (34).

    Placement using fiberoptic endoscopy resulted in much better

    positioning (35).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    9/144

    View Figure

    Figure 20.3White double-lumen tube. (Picture courtesy

    of Teleflex Medical.)

    Broncho-Cath Left-sided TubeOn the Broncho-Cath left-sided DLT (Fig. 20.10), the bronchial

    portion of the tube is at an angle of approximately 45 degrees

    (33,36,37). The bronchial portion has a curved tip. The bevel was

    removed in 1994 but reintroduced in 2001 (33,38). The average

    length of the bronchial segment is 30 to 31 mm (30). The tube is

    available with or without a carinal hook.

    Sher-I-Bronch Right-sided Double-lumen

    Tube

    The Sher-I-Bronch right-sided DLT has two cuffs on the bronchial

    segment, one proximal and one distal to the upper lobe

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    10/144

    ventilation slot. The slot is 13 to 14 mm long (Figs. 20.11, 20.12).

    One study found that poor lung isolation was more common when

    this tube was used (39). A case has been reported in which the

    tube tip became trapped in the right upper lobe bronchus (40).

    View Figure

    Figure 20.4Robertshaw right double-lumen tube inplace.

    P.637

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    11/144

    View Figure

    Figure 20.5Robertshaw right double-lumen tube in

    place.

    Sher-I-Bronch Left-sided Double-lumen TubeThe bronchial segment of the Sher-I-Bronch left-sided DLT

    diverges from the main tube at an angle of 34 degrees. The

    average length of the bronchial segment is 35 mm (30). The

    bronchial segment has a bevel (30). One study found that the

    bronchial cuff on this tube required significantly higher pressures

    to achieve one-lung isolation than cuffs on other DLTs (41).

    Silbroncho Double-lumen Tube

    The Silbroncho DLT (Fig. 20.13) is made of si l icone, which is

    softer than polyvinylchloride (PVC). The tube does not contain

    latex. The bronchial segment is wire-reinforced distal to the

    tracheal cuff. This allows it to be flexible and prevents kinking.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    12/144

    The wire reinforcement also makes tube position easy to

    determine on x-ray. The bronchial cuff is small and near the end

    of the tube. It is available in sizes 33, 35, 37, and 39 Fr.

    View Figure

    Figure 20.6Robertshaw left double-lumen tube.(Courtesy of Rusch, Inc.)

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    13/144

    View Figure

    Figure 20.7Robertshaw left double-lumen tube.

    Techn i ques

    Tube Choice

    R i gh t ve r su s Le f t

    When surgery is performed on the right lung, a left-sided DLT

    should be used (42,43,44). Because the margin of safety in

    positioning a right-sided DLT is so small, some prefer to use aleft-sided DLT whenever possible for left lung surgery (28,45).

    During left pneumonectomy, immediately before the left mainstem

    bronchus is clamped, the DLT can be pulled from the bronchus

    and used for ventilating the right lung. A disadvantage of this

    technique is the risk of blood and secretions moving from the

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    14/144

    operative (left) bronchus to the nonoperative bronchus. Other

    possible problems include the tube becoming dislodged or

    sutured in place during surgery. A ball-valve obstruction may

    occur as a result of secretions or mediastinal pressure pushing

    the tracheal lumen against the tracheal wall ( 46). A left DLT ma y

    not provide optimum conditions for ventilating the residual lung

    after left upper lobectomy (47).

    A r ight-sided DLT should be used when i t is important to avoid

    manipulation/intubation of the left main bronchus (e.g., an

    exophytic lesion), when the left main bronchus is narrowed or the

    left mainstem bronchus is so cephalad that the bronchial lumen

    will not enter the

    P.638

    left mainstem bronchus, left pneumonectomy, left lungtransplantation, left mainstem bronchus stent in place or when

    there is tracheobronchial disruption on the left (43,48,49,50,51).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    15/144

    View Figure

    Figure 20.8Broncho-Cath right double-lumen tube.

    Newer versions have no bevel at the end of the bronchialsegment. (Courtesy of Mallinckrodt Medical, Inc.)

    Before placing a right DLT, the patient's chest x-ray or computedtomography scan can be closely examined to id entify a right

    upper lobe bronchus takeoff, which would make it difficult to use

    a right DLT. A left DLT should be placed in the right mainstem

    bronchus in Kartagener's syndrome, which includes complete

    situs inversus and a longer-than-normal right mainstem bronchus

    (52).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    16/144

    View Figure

    Figure 20.9Broncho-Cath right double-lumen tube. The

    bronchial cuff has the shape of an S or a slanteddoughnut, with the edge of the cuff nearest the rightupper lobe bronchus closer to the trachea than the part ofthe cuff touching the medial bronchial wall. A slot in thetube beyond the cuff corresponds to the opening of theright upper lobe bronchus. Newer versions have no bevelon the bronchial segment.

    Size

    Selecting an appropriately sized DLT for a given patient is critical

    to minimize the frequency of complications (53).

    A DLT that is too small may fai l to provide lung isolation or may

    require bronchial cuff volumes and pressures that could produce

    mucosal ischemia or bronchial rupture. Using too small a DLT

    can result in the tube advancing too far into the bronchus, a

    higher level of autoPEEP (positive end-expiratory pressure), or

    barotrauma (54,55). An undersized tube may be more likely to be

    displaced. Ventilation and suctioning are more difficult with a

    small tube.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    17/144

    Using a large DLT wil l result in less resistance to gas flow,

    facilitate suctioning and passage of a fiberscope, and reduce the

    risk of advancing the DLT too far into t he bronchus but may

    result in trauma (56). Inability to insert a larger tube through the

    larynx or past the carina or intrinsic or extrinsic obstruction in the

    mainstem bronchus to be intubated may necessitate use of a

    smaller tube.

    A tube is oversized if the bronchial lumen wi l l not fi t into the

    bronchus or there is no air leak with the bronchial cuff deflated

    (22,38,57,58,59,60). It is too small if the bronchial cuff inflation

    volume is greater than the resting cuff volume. Not more than 3

    cc of air in the bronchial cuff should be required to create a seal.

    When there is a high risk that f luids wil l seep past the bronchial

    cuff, a smaller DLT should be used (61).

    In adults, the dimensions of the cricoid ring best define those ofthe main bronchi (23). In children, age but

    P.639

    P.640

    not weight is a predictor of bronchial diameter, and the right main

    bronchial diameter is significantly larger than the left ( 62).

    Suggested sizes for DLTs in children are shown in Table 20.1.

    Tubes from different manufacturers vary in size and may not fit

    this table.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    18/144

    View Figure

    Figure 20.10Broncho-Cath left double-lumen tube.

    (Courtesy of Mallinckrodt Medical, Inc.)

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    19/144

    View Figure

    Figure 20.11Sher-I-Bronch double-lumen tubes. A:

    Left-sided tube (top). Right-sided tube (bottom). B:Close-up of right bronchial segment, showing opening tothe right upper lobe. (Courtesy of Sheridan, Inc.)

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    20/144

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    21/144

    Age (years) Doubl -lumen Tube (French)

    8 to 10 26

    10 to 12 26 to 28

    12 to 14 32

    14 to 16 35

    From Hammer GB, Fitzmaurice BG, Brodsky JB. Methods for single-lungventilation in pediatric patients. Anesth Analg 1999;89:14261429.

    The size of the mainstem bronchus may be determined by

    measuring the width of the patient's bronchus from a chest x-ray

    or computed tomographic scan (22,54,60,63,64,65,66,67,68).

    Unfortunately, it is not possible to accurately measure bronchial

    width on many chest x- rays, and the correlation between the

    tracheal and bronchial size may not be reliable enough to

    determine the proper size DLT from the size of the trachea(56,59,63,69,70,71,72). Reliability may be increased by

    measuring the tracheal diameter both anteropostally and

    mediolaterally (73). The lung transplant patient may have a

    significantly different sized bronchus than would have been

    predicted from the size of the trachea (56).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    22/144

    Preparing the Double-lumen Tube

    The tracheal and bronchial cuffs should be inflated and checked

    for leaks and symmetrical cuff inflation, making certain that each

    inflation tube is associated with the proper cuff. The cuffs and

    stylet should be lubricated with a water-soluble lubricant and the

    stylet placed

    P.641

    inside the bronchial lumen, making certain that it does not extend

    beyond the tip. The connector should be assembled so that it can

    be quickly fitted to the tube and breathing system after

    intubation.

    View Figure

    Figure 20.13A:The Silbroncho double-lumen tube. B:The bronchial segment is wire-reinforced distal to thetracheal cuff.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    23/144

    Insertion

    The DLT is advanced through the larynx with the angled tip

    directed anteriorly. After the bronchial cuff has passed the cords,

    the tube is turned 90 degrees so that the bronchial portion points

    toward the appropriate bronchus. If the tube is to be placed in the

    left mainstem bronchus, the head and neck should be rotated to

    the right before rotating and advancing the tube (74). Leaving the

    stylet in place for the entire intubation procedure rather than

    removing it once the bronchial cuff has passed the vocal cords

    may result in more rapid and accurate placement ( 75). However,

    some recommend that the stylet be removed just after the tube

    passes the vocal cords to prevent trauma (76).

    A DLT is most accurately placed by inserting a fiberscope into

    the bronchial lumen and directing it into the appropriate bronchus

    under direct vision (77,78,79,80,81,82). Concurrent direct

    laryngoscopy may be required to elevate the supraglottic tissues

    to facilitate passing the DLT through the glottic opening after the

    fiberscope is in the trachea (83). This ensures that the correct

    bronchus is intubated on the first attempt and avoids inserting

    the tube too deeply or the tube becoming kinked in the upper

    lobe bronchus (40).

    A DLT wi th a carinal hook should be inserted with the bronchial

    segment concave anteriorly until the bronchial cuff passes the

    cords. It should then be rotated 180 degrees so that the hook is

    anterior and advanced until the hook passes the vocal cords. The

    tube is then advanced until the hook engages the carina. The

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    24/144

    hook can be tied closely to the tube with a slip knot to facilitate

    passage through the larynx and then untied.

    Inserting the bronchial portion into the bronchus can be

    performed blindly after insertion through the vocal cords. In some

    cases (e.g., bronchorrhea, bleeding), blind placement may

    succeed where the fiberoptic technique does not (79).

    With blind insertion, the correct depth of insertion may be difficult

    to determine. However, this method may be useful where rapid

    lung isolation or collapse is necessary. In adults, there is a

    correlation between the ideal depth of insertion of left DLTs and

    patient height but not weight or age (24,84,85,86,87). The ideal

    depth of insertion can be estimated from the chest x-ray

    (84,85,86). Advancing the tube with the bronchial cuff partially

    inflated unti l an increase in resistance is felt (or only one side of

    the chest moves and compliance is reduced) may preventinserting the tube too deeply (88,89,90,91). The bronchial cuff is

    then deflated and the tube advanced a distance equal to the

    length of the bronchial cuff plus 1 to 1.5 cm to place it just

    beyond the carina (92,93). Depth of insertion may be estimated

    by monitoring bronchial cuff pressure (94).

    Since the tube usually moves upward with positioning, some

    clinicians recommend that the tube should initially be inserted

    more deeply than would be the ideal position (36,38,95,96).

    Others believe that the danger of trauma is increased if the tube

    is intentionally placed too deeply (97).

    If the patient is anatomically difficult to intubate, a single-lumen

    tracheal tube may be placed by any of the means that facilitate

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    25/144

    difficult intubations (98). An exchange catheter can then be

    inserted into the single-lumen tube and the DLT inserted over the

    catheter after the single-lumen tube has been withdrawn (99).

    The WuScope (Chapter 1 8) has been successfully employed to

    place a DLT in a patient who was difficult to intubate (100). A

    DLT can be placed by using a lighted stylet (Chapter 19) in the

    bronchial lumen (101) or a retrograde intubation technique (102).

    Awake fiberoptic bronchial intubation with a DLT has been

    reported (103). A DLT ma y be introduced over a gum elastic

    bougie (45,104,105,106). It may be helpful to pass the bougie

    through the opening for the right upper lobe orifice on right DLTs

    (105).

    A DLT may be inserted through a tracheostomy (107,108).

    Cuff Inflation

    Once the tip is thought to be in a mainstem bronchus, the

    tracheal cuff should be inflated in a manner similar to that of a

    tracheal tube (109,110). It is more difficult to inflate the bronchial

    cuff correctly. An overinflated bronchial cuff is more likely to

    herniate into the trachea, cause the carina to be pushed toward

    the opposite side, or result in narrowing of the bronchial segment

    lumen. Inflating the bronchial cuff beyond its resting volume may

    result in dangerously high pressure (18,111).

    The bronchial cuff should be inflated with small incremental

    volumes until an airtight seal is just attained (109,112). The total

    volume should be less than 3 mL (41). One technique is to

    immerse the proximal tracheal lumen in water during ventilation

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    26/144

    via the bronchial lumen. The bronchial cuff is inflated until no

    bubbles are seen escaping during positive-pressure inspiration

    (110,111) (Fig. 20.14). Variations of this are to connect a balloon

    (113) or a capnograph (111) to t he tracheal lumen. Another

    method is to apply suction to the tracheal lumen (110). Absence

    of bronchial seal will cause the reservoir bag in a breathing

    system that is connected to the bronchial lumen to collapse.

    Inflating the bronchial cuff to an airtight seal may not prevent the

    spread of blood or secretions (61). The bronchial cuff may also

    be inflated with water (114).

    Confirming Position

    Confirming proper tube position is essential because the tube

    may not perform properly if incorrectly

    P.642

    positioned. Position should be checked after insertion, after

    repositioning the patient, and before beginning one-lung

    ventilation, as these tubes often move during patient positioning

    or surgical manipulations

    (97,115,116,117,118,119,120,121,122,123). The most frequent

    DLT movement is during lateral decubitus positioning. While

    movement is usually outward, distal migration may also occur.

    DLT position should be confirmed whenever there is evidence of

    malfunction.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    27/144

    View Figure

    Figure 20.14Inflating the bronchial cuff. With the

    tracheal cuff inflated, the bronchial cuff is slowly inflated.Right:A bronchial cuff leak is indicated by bubbles whenthe end of the tracheal lumen is placed under water. Left:With a seal, no bubbles appear.

    Au s cu l t a t o r y Tech n i q u e s

    Unfortunately, auscultation detects DLT malposition only part of

    the time because breath sounds can be transmitted from one

    region of the lung to adjacent areas (124,125,126,127). Studies

    have shown that a significant percentage of DLTs thought to be

    positioned satisfactorily by auscultation were found to be

    inappropriately positioned on subsequent fiberoptic examination

    (116,120,128). A DLT may function satisfactorily although not in

    an ideal position. Another problem with auscultation is that once

    the patient is prepped and draped, the chest is no longer

    available for auscultation. One study found that auscultatory

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    28/144

    placement was not associated with an increased incidence of

    complications during one-lung ventilation (39).

    Clamping Method

    Le f t -s i d ed Tubes

    With only the tracheal cuff inflated and the tracheal lumen

    connected to the breathing system, both lungs should be

    auscultated in the axillary regions and upper lung fields to detect

    differences. The bronchial cuff should then be inflated and bothlumens connected to th e breathing system. Auscultation should

    then be repeated.

    Next, the attachment between the breathing system and the

    tracheal lumen should be occluded and the tracheal lumen

    opened to atmosphere. Breath sounds should be heard only over

    the left lung. If breath sounds are heard bilaterally, the tube is

    too high in the trachea. Both cuffs should be deflated and the

    tube advanced. If breath sounds are heard only over t he right

    lung, the bronchial lumen is on the right side. If this is the case,

    both cuffs should deflated, the tube withdrawn until its distal end

    is above the carina, rotated, then reinserted. The steps outlined

    should be repeated.

    The attachment between the breathing system and the bronchial

    lumen should then be clamped and the patient ventilated through

    the tracheal lumen. The bronchial lumen should be opened to

    atmosphere. Breath sounds should be heard only over the right

    lung. If there is marked resistance to ventilation, the tube is

    either too far into the left bronchus or is not deep enough. The

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    29/144

    position can be determined by deflating the bronchial cuff while

    continuing to ventilate through the tracheal lumen with the

    bronchial lumen clamped. If the tip is too deep in the left

    bronchus, breath sounds will be heard only on the left side. If the

    tube is not deep enough in the bronchus, breath sounds wil l be

    present bilaterally. The tracheal cuff also should be deflated and,

    depending on where breath sounds were heard, the tube pulled

    back or advanced. Both cuffs should be reinflated and the

    auscultatory sequence repeated.

    R i gh t - s i d ed Tubes

    Auscultation of a r ight-sided DLT is similar to that of a le f t-sided

    tube. It is especially important to confirm ventilation of the right

    upper lobe.

    P.643

    Single Connector Method

    An alternative technique is the single connector technique in

    which a single connector is used to ventilate each lung in turn by

    simply transferring the distal end of the DLT from one lumen to

    the other, eliminating the need for repeated clamping and

    unclamping (126). This has the advantage of being simpler and

    requiring fewer steps and may result in reduced risk of creating a

    potentially harmful ball-valve effect in a partially obstructed lobe

    or lung by detecting it earlier.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    30/144

    After the DLT is advanced to the desired depth, a single

    connector from the breathing system is attached to the bronchial

    lumen. While performing manual ventilation, the bronchial cuff is

    inflated to create an airtight seal. Auscultation is used to confirm

    that the intended lung is being ventilated. The connector is then

    transferred from the bronchial lumen to the tracheal lumen. The

    tracheal cuff is inflated to create an airtight seal, and it is

    confirmed that the proper lung is ventilated. The connector is

    then transferred to the bronchial lumen and short ventilations

    performed while listening over the apex of the lung for vesicular

    breath sounds. If these are not heard, the single connector is

    detached from the bronchial lumen and a double connector is

    connected to the two lumens.

    F le x i b l e Endos co p i c Techn i q ue s

    Flexible endoscopy is the most accurate method for determining

    DLT position (120,129,130,131,132). Many recommend that this

    should be the standard of care (81,131,133,134,135,136). Others

    believe that the fiberoptic scope is helpful but not essential

    (95,124,137,138,139). There is general agreement that fiberoptic

    endoscopy is always needed for right-sided DLTs. Fiberoptic

    methods to confirm the position of the DLT may not work in the

    presence of blood or secretions (140). W henever there is any

    doubt, this method should be used to check the position. A

    further advantage is that it can be used to remove b lood or

    secretions. It can also be used after the patient is prepped and

    draped if a question arises with regard to proper tube placement.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    31/144

    Left-sided Tubes

    A f iberscope is placed in the tracheal lumen through the openend of the tube or through a port in the connector that is specially

    designed for this purpose. As the fiberscope is advanced, the

    carina should come into view. The top surface of the blue

    bronchial cuff should be seen below the carina in the left

    mainstem bronchus. The bronchial cuff should not herniate over

    the carina, neither should the carina be pushed t o the right. An

    unobstructed view of the nonintubated right mainstem bronchus

    should be obtained.

    The fiberscope should then be advanced through the bronchial

    lumen to check for narrowing of the lumen at the level of the cuff

    and an unobstructed view of the distal bronchial tree.

    Right-sided Tubes

    Looking down the tracheal lumen, the bronchial cuff's upper

    surface should be seen below the carina in the right mainstem

    bronchus. The fiberscope is then placed in the bronchial lumen.

    The right middle lower lobe bronchial carina should be seen

    below the end of the tube. The endoscopist should be able to

    look into the right upper lobe orifice by flexing the tip of the

    fiberscope superiorly.

    B r o n c h o s p i r ome t r y

    Pressure-volume and flow-volume loops are discussed in Chapter

    23. Changes in compliance or resistance may mean that the DLT

    is not correctly placed

    (141,142,143,144,145,146,147,148,149,150). If a DLT is

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    32/144

    advanced into the trachea with the bronchial cuff partially

    inflated, compliance is reduced when the tube impacts the

    bronchus (92).

    Ches t X -ray

    Chest x-ray may be useful to confirm tube position when a

    fiberscope is not available or cannot be used. However, it is less

    precise than the fiberoptic bronchoscopy and is costly, time

    consuming, and awkward to perform.

    Stabilizing the Tube

    Once correct tube position is confirmed, the tube should be

    secured in place. Special fixation methods have been

    recommended (151). During positioning, the tube should be held

    at the level of the incisors and the head immobilized in a neutral

    or slightly flexed position to prevent the tube from migrating into

    an incorrect position.

    Intraoperative Care

    The bronchial cuff should be kept deflated (unless the lung needs

    to be isolated to prevent spread of blood or infection) unti l the

    lung needs to be collapsed to minimize damage to the bronchial

    mucosa (38). Lung collapse will be most rapid if lung separationis initiated at end-expiration. Suction is of limited utility because

    the gas trapped in the lung is distal to collapsible airways. If

    despite best efforts complete lung separation cannot be

    accomplished and gas is introduced into the ipsilateral lung with

    each breath, then continuous suction may be helpful to evacuate

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    33/144

    the gas as it enters the lung (38). Bronchial cuff pressure should

    be monitored and adjusted to the minimum necessary to achieve

    an airtight seal.

    A useful technique when the bronchial lumen is in the surgically

    operated lung is to pass a suction catheter through the bronchial

    lumen when the lung is deflated and to leave it until ready for

    reinflation (152). This may prevent the bronchial lumen from

    becoming obstructed by blood or mucus.

    P.644

    Replacing a Double-lumen Tube with a

    Single-lumen Tube

    If mechanical ventilation needs to be continued at the conclusion

    of a case in which a DLT was used, it is usually desirable to

    replace the DLT with a standard tracheal tube. Personnel who

    are caring for the patient in the postoperative period are not

    usually familiar with a DLT. After pneumonectomy, when all of the

    patient's ventilation is conducted through one lumen of the DLT,

    the small diameter of the lumen may make if difficult for the

    patient to breathe spontaneously. W ith high minute venti lation,

    this can lead to auto-PEEP (17). Pressure-support ventilation

    (Chapter 12) can be used to decrease the imposed work of

    breathing during spontaneous ventilation if replacing the DLT

    cannot be performed. Suctioning through a DLT is difficult.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    34/144

    In most cases, the procedure is simply to remove the DLT and

    insert a single-lumen tube in its place. If the patient was difficult

    to intubate or circumstances make visualizing the larynx difficult,

    other techniques should be considered. One technique would be

    to insert an airway exchange catheter into the tracheal lumen

    before the DLT is removed. Special extra-long exchange

    catheters are needed for the exchange. After the DLT has been

    removed, the single-lumen tube is then advanced over the

    catheter (153,154). Oxygen insufflation via the catheter will

    reduce the incidence of hypoxemia.

    Another technique has been described (155). At the conclusion of

    the case, both cuffs are deflated and the DLT withdrawn until the

    bronchial lumen is above the carina. The bronchial cuff is then

    inflated and the lungs ventilated through the bronchial lumen.

    The tracheal lumen is clamped, and an opening is created in thewall of the tracheal lumen. A single-lumen tube is then slipped

    over a fiberscope, and the fiberscope is advanced through the

    hole in the tracheal lumen and into the trachea. The opening in

    the DLT is extended, and the DLT is slowly removed. The

    fiberscope is then removed.

    Hazards A sso c i a t ed w i t h Dou b le -l um en

    Tubes

    Many of the hazards associated with single-lumen tracheal tubes

    (Chapter 19) can also occur with DLTs.

    Difficulty with Insertion and Positioning

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    35/144

    Inserting a DLT may be time consuming. W hen there is severe

    hemorrhage, this can be a major problem. Multiple insertions and

    repositionings increase the risk of trauma. While they are very

    useful in adults and older children, they are of ten too large for

    small children. They are often difficult to position and usually

    must be replaced with a single-lumen tube at the end of surgery.

    The rigidity and width of the DLT can make intubation

    complicated. The tube does not allow preshaping before

    intubation, and the large width makes it difficult to pass the DLT

    through a tracheostomy stoma, small airway, or nasal passage.

    Tube Malposition

    Certain physical conditions may make it difficult or impossible for

    a DLT to be correctly placed (156,157). Preoperative fiberoptic

    endoscopy may detect many of these problems. Even if a correct

    position is achieved initially, head movement, a change in body

    positioning, or surgical manipulation may result in t ube

    malposition. Displacement during positioning can be decreased

    by using a neck brace (123,158). Malpositioning is increased

    when anesthesia providers have limited experience in lung

    isolation (159).

    If DLT malposition is suspected, fiberoptic techniques are clearly

    advantageous in defining the problem and afford a means of

    visual correction. While some have reported that monitoring

    carbon dioxide waveforms helps to detect DLT displacement

    (160,161), capnography does not reliably indicate DLT

    misplacement (142,148,149,162).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    36/144

    Consequence s

    Functional indications of misplacement include the following:

    Unsatisfactory Lung Collapse

    An obstruction in the unventi lated lumen can prevent the

    unventilated lung from deflating. If the lung cannot be collapsed,

    operating time will be increased, and the surgical result may be

    compromised. Another cause may be a tumor fragment in the

    airway (163).

    Obstruction to Lung Inflation

    If the bronchial cuff is not below the carina, it may obstruct the

    trachea and right mainstem bronchus. With right-sided tubes,

    misalignment of the port for the right upper lobe can result in

    obstruction. If the bronchial cuff on a left-sided DLT is too deep,

    it may obstruct the upper lobe bronchus.

    Gas Trapping

    Gas trapping or expiratory obstruction may be the result of a one-

    way valve effect that allows inflation but not deflation. If

    unrecognized, it can result in cardiorespiratory embarrassment

    and/or lung parenchymal damage.

    Failure of Lung SeparationIf the airway to a bronchopleural fistula cannot be isolated from

    that to the normal lung, barotrauma may develop with positive-

    pressure ventilation, or the air leak through the fistula may be so

    large that ventilation of the normal lung is compromised. An

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    37/144

    incompletely protected dependent lung may be flooded with blood

    or secretions.

    Poss i b l e Mal p o s i t i o n s

    Bronchial Lumen in the Wrong Mainstem

    Bronchus

    In some cases, the bronchial portion will enter the opposite

    P.645

    lung. This is usually easy to detect and correct by using

    fiberoptic endoscopy (115). In some cases, it may be best to

    leave the bronchial lumen in the operative bronchus and isolate

    the operative lung by clamping the bronchial limb and using the

    tracheal lumen for ventilation (118,122). This may be appropriate

    for right lung surgery but not for surgery on the left lung, sincethe right upper lobe bronchus would almost certainly be

    occluded. An alternative option is to withdraw the tube until it is

    intratracheal and to use a bronchial blocker (see below) to block

    the operative lung.

    It may be possible for the surgeon to assist in correctly placing

    the tube once the chest is open (164). If it is determined that the

    tube is in the wrong bronchus, both cuffs are deflated, and the

    tube is withdrawn into the trachea. The surgeon then compresses

    the bronchus, and the anesthesia provider advances the tube into

    the correct side with surgical guidance. The cuffs are then

    reinflated.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    38/144

    Bronchial Portion Inserted too Far into the

    Appropriate BronchusIf a left-sided DLT is inserted too deeply, the problem may be

    that the tube is too small (156,165,166,167). It will result in

    obstruction of the upper lobe. In some patients, a left-sided DLT

    placed so that the bronchial cuff is just distal to the carina still

    may cause left upper lobe obstruction. A high peak airway

    pressure during one-lung venti lation should suggest t hismalposition (135).

    Tube too Proximal in the Airway

    If the tube is not sufficiently advanced into the bronchus, the

    bronchial cuff may protrude into the trachea. The need to inject

    more than 3 mL of air into the bronchial cuff to achieve a seal

    should alert the user that the tube may be malpositioned. Thebronchial segment may slip out of its bronchus, especially during

    changes in the patient's position. In many cases, no untoward

    sequelae will occur. However, there may be obstruction of gas

    flow to the other lung and inability to isolate the surgical lung.

    Tip of Bronchial Lumen above the Carina

    The tip of the bronchial lumen may be above the carina becauseof a tracheal lesion that prevents the tube from being advanced

    farther. With this malposition, there will be unsatisfactory lung

    deflation and failure of lung separation.

    Incorrect Placement with Respect to the Upper

    Lobe Bronchus

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    39/144

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    40/144

    oxygen source at ambient pressure to the opening to the

    nonventi lated lung may h elp to prevent hypoxia

    (175,176,177,178). This may also enhance collapse of the

    nonventilated lung.

    If hypoxemia is a problem despite proper tube position, CPAP

    can be applied to the nondependent lung (179). Some DLT

    manufacturers include a CPAP device with each DLT, or they

    may be purchased separately (Fig. 20.15). Other measures to

    improve oxygenation include dependent lung PEEP, occasional

    ventilation of the nondependent lung (one breath every 5 to 10

    minutes), insufflation of 2 to 3 L/minute of oxygen to the

    nonventilated lung, and clamping of the pulmonary artery before

    clamping the bronchus. Jet venti lation of the nondependent lobes

    that are not being removed by using an airway exchange catheter

    may be used to improve oxygenation (180).

    Obstructed Ventilation

    Many cases of obstruction are the result of a malpositioned tube.

    In addition, inflating the bronchial cuff can cause narrowing of the

    bronchial lumen (129,156) or may cause the carina to be

    displaced laterally, producing obstruction of the other mainstem

    bronchus (181). A defective tube or connector may c ause

    obstruction (182,183).

    Other causes of bronchial obstruction have been reported. In one

    case, the bronchial cuff was left deflated until one-lung

    ventilation was to begin, and necrotic tumor migrated into the

    bronchus of the dependent lung, causing obstruction when one-

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    41/144

    lung ventilation was begun (184). The bronchial lumen can

    become twisted (185). A carinal hook may obstruct the opening of

    the tracheal lumen (186).

    P.646

    View Figure

    Figure 20.15Device for applying continuous positiveairway pressure to a nonventilated lung. The adjustablevalve supplies pressures from 1 to 10 cm H2O.

    A rela tive contraindication to using a DLT is a lesion (airwaynarrowing or endoluminal tumor) somewhere along the pathway

    where the tube will reside. An aberrant tracheal bronchus may be

    a contraindication for using a DLT (172,173).

    Trauma

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    42/144

    Trauma to the r espiratory tract can occur whenever intubation

    with a DLT is performed (187,188). Tears in the trachea and

    mainstem bronchus have been reported

    (76,187,188,189,190,191,192,193,194,195,196,197,198,199,200,

    201,202,203).

    Tube size is a factor. Large tubes have been involved more often

    in injury than smaller ones (188). A tube that is too small and

    requires excessive cuff inflation may cause ischemic injury. In

    one reported case, an endobronchial polyp that developed at the

    bronchial cuff site ended in a fatal hemorrhage (204).

    Measures to reduce airway trauma include removing the stylet

    after the tip of the tube has passed the vocal cords, avoiding cuff

    overinflation, deflating the tracheal and bronchial cuffs when

    repositioning the patient or the tube, and not advancing the tube

    when resistance is encountered. Some bronchial cuffs canprovide one-lung isolation with significantly lower pressures than

    others (41). It has been recommended that the bronchial cuff be

    kept deflated until needed to minimize pressure on the bronchial

    mucosa. This may not be prudent if there is a bronchial tumor, as

    necrotic tumor may migrate into the other lung (184).

    Tube ProblemsReported problems with DLTs include mislabeling, tracheal lumen

    distortion that prevented a suction catheter from passing, a slit in

    the septum, a defect that made the bronchial lumen kink on itself,

    a split in the tubing to the bronchial cuff, a kink in the inflating

    tube to the tracheal cuff, a protuberance in the wall of the tube

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    43/144

    with resultant tracheal lumen obstruction, and foreign bodies in

    the DLT (205,206,207,208,209,210,211,212). The carinal hook

    may bend backward, obstructing the opening of t he tracheal

    lumen (213).

    Tracheal or bronchial cuff rupture can occur. This most commonly

    results from contact between the tube and the teeth or

    laryngoscope during insertion. Proposed methods to avoid this

    problem include the use of a retractable protective sheath (214),

    a lubricated Penrose drain (215), and lubricated teeth guards

    (216) and increasing the curve of the bronchial portion of the DLT

    with the stylet (217). Another possible cause of cuff rupture is

    movement during repositioning (218).

    Surgical Complications

    The bronchial cuff may be punctured by the surgeon ( 219,220). A

    suture or staple may be placed through the DLT, or the surgical

    procedure may result in a tight stenosis, which could entrap the

    bronchial segment (221).

    Failure to Seal

    One of the reasons to use a DLT is to prevent material from

    passing from one lung to the other during the surgical procedure.

    Failure to prevent fluids from traversing the bronchial cuff could

    result from malposition or from improper cuff inflation. Neither an

    airtight bronchial seal nor a cuff pressure of 25 cm H 2O

    guarantees protection against aspiration (61). Lubricating the cuff

    with a gel will reduce the risk that fluid will leak past the cuff

    (222,223).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    44/144

    Difficult Extubation

    Difficulty in removing a DLT may be due to anatomical

    abnormalities, surgical fixation, or entanglement by other surgical

    or anesthetic hardware (220,224).

    P.647

    Single-lumen Bronchial TubesAnother option for achieving lung separation is to use a single-

    lumen tube to intubate a mainstem bronchus. Two bronchial

    tubes may be used in situations where dual lung ventilation is

    needed but circumstances bode against having one tube in the

    trachea (225,226,227).

    Single-lumen bronchial tubes are sometimes used in pediatricpatients whose airways are too small for DLTs or in whom a

    bronchial blocker cannot be used

    (1,165,226,228,229,230,231,232,233,234,235,236,237). In the

    patient with massive hemoptysis, bronchial intubation with a

    single-lumen tube is often the easiest and quickest method of

    separating the lungs (5,10).

    Equ i pmen t

    Bronchial intubation is most often carried out with a conventional

    tracheal tube. A cuffed tube will prevent re-expansion of the

    collapsed lung. The distance from the tip of the tube to the

    cephalad edge of the cuff must be shorter than the length of the

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    45/144

    mainstem bronchus to ensure that the cuff can lie entirely inside

    the bronchus and the upper lobe bronchus is not obstructed

    (238).

    Special tubes with a single lumen, an angulated distal (bronchial)

    tip, and cuffs at both the tracheal and bronchial positions are

    available (239,240). They are longer than conventional single-

    lumen tracheal tubes.

    Techn i ques

    Before insertion, the correct length and size of the tube should

    be estimated from a chest x-ray or computed tomography (CT)

    scan (238). If the nasotracheal route is used, most conventional

    single-lumen tubes will not be long enough to provide a reliable

    mainstem intubation. It is recommended that the tube should be

    one half to one size smaller than the usual size selected for

    tracheal intubation (237). For bronchial intubation in children, a

    tracheal tube 0.5 to 2.0 mm smaller than recommended for the

    particular patient should be used (229,231).

    Right-sided intubation can usually be performed blindly, but the

    tube is more reliably placed by using a bronchoscope. It may be

    possible to align the Murphy eye with the right upper lobe

    bronchus. It may be possible to rotate the tube so that the bevel

    faces the upper lobe bronchial orifice.

    Left mainstem intubation may be a chieved blindly by using a

    stylet to curve the distal end of the tr acheal tube to the left (231)

    but often requires bronchoscopic guidance. If blood or secretions

    preclude fiberoptic visualization, fluoroscopy is another option.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    46/144

    The chance of intubating the left bronchus will be increased if the

    tube is rotated 180 degrees from its usual position before

    advancing it beyond the vocal cords and the patient's head is

    turned to the right (228,229,241,242,243).

    A gum elastic bougie can be inserted into the chosen bronchus

    by using a bronchoscope. The bronchial tube can then be

    railroaded over the bougie into position (244).

    Correct positioning can be confirmed by auscultation, x-rays,

    and/or flexible bronchoscopy.

    The tube should be withdrawn into the trachea when one-lung

    ventilation is no longer needed.

    Eva lua t i on

    Advantages of using a single- lumen tube for lung separation

    include simplicity and the rapidity with which lung separation can

    often be achieved, particularly when the right lung must be

    ventilated.

    Disadvantages include frequent lack of ventilation of the right

    upper lobe with right mainstem intubation (38,236). Left upper

    lobe ventilation may also be excluded when the left mainstem

    bronchus is relatively short (238). Neither suctioning nor

    application of CPAP to the nonventilated lung is possible.

    There may be failure to achieve an adequate seal, especially if

    an uncuffed tracheal tube is used or if the cuff is not inside the

    bronchus (226,236,238,245). Lung collapse will be incomplete,

    and the healthy ventilated lung will not be protected from

    contamination. Both lungs cannot be ventilated at the beginning

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    47/144

    of anesthesia, and the collapsed lung cannot be re-expanded and

    ventilated until the tube is withdrawn into the trachea ( 236).

    Bronchial-blocking Devices

    Ind i ca t i ons and Use

    Indications for bronchial blockers are similar to those for a DLT,

    with the exception of independent lung ventilation (44,226,246).

    They are often used in patients in whom using a DLT is not

    possible or advisable (nasal intubation, small patient, difficult

    intubation, patient with a tracheostomy, subglottic stenosis, thick

    and excessive secretions, need for continued postoperative

    intubation)

    (240,247,248,249,250,251,252,253,254,255,256,257,258,259,260

    ,261). A blocker may be especially useful for providing lung

    separation in a critically il l patient with a single-lumen trachealtube already in place. Another indication may be the patient on

    anticoagulants, since placing a blocker is usually less traumatic

    than inserting a DLT (8). A blocker may allow a larger fiberoptic

    endoscope to be used and provide better suctioning than a DLT

    (262). A modified bronchial blocker can be used for tracheal gas

    insufflation to reduce carbon dioxide within the dead space by

    delivering fresh gas near the end of the tracheal tube ( 263).

    P.648

    Another indication for a blocker is the need to block a segment of

    a lung rather than the entire lung

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    48/144

    (9,14,258,259,264,265,266,267,268,269,270). This cannot be

    done with a DLT. A blocker may be used to sequentially block

    different parts of the lung (270).

    A blocker may be used to achieve lung isolation in the patient

    with an improperly positioned double-lumen or bronchial tube

    (53,122,238,271,272,273). A blocker may be useful if both lungs

    require sequential blockage (274). The blocker can be shifted to

    the opposite lung when needed. If one blocker does not provide

    complete one-lung isolation, a second blocker may be used

    (173,275,276,277).

    Finally, there is no need to change the tube at the end of the

    operation if postoperative mechanical ventilation is needed if a

    bronchial blocker is used.

    Dev ices

    Univent Bronchial-blocking Tube

    Desc r i p t i o n

    The Univent tube is a cuffed silicone tracheal tube with a small

    additional internal lumen along its concave side

    (7,11,226,236,278,279,280,281,282,283,284) (Figs. 20.16,

    20.17). The s mall channel contains a movable tubular blockerthat has a blue high-pressure, low-volume cuff. The blocker can

    be advanced sufficiently beyond the tip of the tube to block

    airways smaller than a mainstem bronchus (246,260,264,265).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    49/144

    View Figure

    Figure 20.16Univent bronchial blocker. The cuffed

    tracheal tube has a small lumen along its concave side,which contains a tubular cuffed bronchial blocker. The

    blocker can be advanced into a mainstem bronchus orsmaller airway.

    The blocker has external depth markings to help determine theblocker position in relation to the tube. There is a s light bend in

    the blocker above the cuff. The blocker tip is radiopaque. A grip

    allows the user to rotate the blocker. A locking clamp fixes its

    depth below the tip of the tube.

    As shown in Table 20.2, the Univent tube is available in several

    sizes. It has a slightly larger-than-usual external diameter for its

    internal diameter because of the space required by the blocker.

    Univent tubes that are 5 mm and larger have a lumen that can be

    used for suctioning, CPAP, or oxygen insufflation (285). Adult

    versions of the Univent blocker are hollow.

    The Univent blocker (Uniblocker) can be purchased separate

    from the tracheal tube and used with other tracheal tubes

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    50/144

    coaxially or in parallel (265,286) (Fig. 20.18). It is supplied with a

    swivel adaptor that fits onto the tracheal tube connector. This

    adaptor allows connection to the breathing system and has a port

    for a fiberoptic scope in addition to a port for the blocker.

    Use

    Before use, the bronchial blocker and tube cuffs should be

    inflated and checked for leaks. Both the tube and the blocker

    should be well lubricated. After the cuffs have been deflated, the

    blocker should be pushed back and forth in the tube to ensure

    free movement. The blocker should then be fully retracted into

    the tube lumen and fixed in place by using the clamp.

    If there is an unobstructed view of the larynx, the Univent tube is

    inserted in the same way as a conventional tracheal tube. If the

    patient has a difficult airway, it can be inserted over an airway

    exchange catheter or other device (287). The blocker can be

    extended and used as an introducer (288,289). After the blocker

    has passed the vocal cords, the tube is t hreaded over it and into

    place.

    After the Univent tube is inserted, the tracheal tube cuff is

    inflated, and the patient is ventilated. The blocker is visualized by

    using a flexible fiberoptic endoscope through an airway adaptor

    with a port for the scope and is maneuvered into the appropriate

    bronchus. The blocker tip direction can be changed by twisting

    the shaft (290). It may be useful to deflate the tracheal tube cuff

    and rotate the tube so that the blocker is directed toward the side

    to be occluded.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    51/144

    A guide wire can be inserted through the blocker 's lumen and be

    used to direct the blocker into place, especially when it is

    necessary to block an airway smaller than a mainstem bronchus

    (258).

    Another method of placing the blocker in the r ight or le ft main

    bronchus is to insert a fiberscope through the trachea tube into

    the bronchus to be blocked and then to advance the tube into

    that bronchus. The blocker is then advanced into the bronchus

    and the tube withdrawn into the trachea, leaving the blocker in

    the bronchus. This technique may result in trauma to the airway

    (291).

    P.649

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    52/144

    View Figure

    Figure 20.17Univent bronchial-blocking tubes.

    Top:The bronchial blocker is retracted. Bottom:The bronchial blocker is advanced, and the cuff isinflated. (Courtesy of Vitaid.)

    TABLE 20.2 Univent Tubes

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    53/144

    I nternal Diameter

    (millimeters)

    Outer Diameter

    (millimeters)a

    Age (years)

    3.5 (uncuffed) 7.5/8.0 6 to 10

    4.5 8.5/9.0 10 to 14

    6.0 9.7/11.0 14 to 16

    6.5 10.2/12.0 16 to 18

    7.0 11.6/12.5 Adult

    7.5 11.2/13.0 Adult

    8.0 11.7/13.5 Adult

    8.0 12.2/14.0 Adult

    9.0 12.7/14.5 Adult

    a

    Values are sagittal/transverse.From Hammer GB, Fitzmaurice BG, Brodsky JB.

    Methods for single-lung ventilation in pediatricpatients. Anesth Analg 1999;89:14261429; Frolich

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    54/144

    MA. Postoperative atelectasis after one-lung ventilationwith a Univent tube in a child. J Clin Anesth2003;15:159163; Hammer GB, Brodsky JB, RedpathJH, et al. The Univent tube for single-lung ventilationin paediatric patients. Paed Anaesth 1998;8:5557;Tobias JD. Variations of one-lung ventilation. J ClinAnesth 2001;13:3539.

    The bronchial blocker can be inserted blindly. The whole tube isturned so that its concavity faces the side to be blocked. The

    blocker is advanced into the mainstem bronchus and the cuff

    inflated. This method has not proved very successful and may be

    associated with airway trauma (282,292).

    The blocker position should be checked by using a fiberscope.

    The cephalad tip of the bronchial cuff should be below the carina.

    The blocker should then be fixed to the tracheal tube by using

    the cap stopper and blocker grip.

    When the bronchus needs to be blocked, the lung is deflated with

    the blocker open to atmosphere. The bronchial blocker cuff

    should be inflated by using the least amount of air that will

    provide a seal. This can be achieved by attaching the sample l ine

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    55/144

    from a carbon dioxide analyzer to the proximal end of the blocker

    and noting when the waveform disappears (293). Another method

    is the bubble test, in which the end of the bronchial lumen is

    placed into water in a beaker (294). When the bronchus is

    sealed, no bubbles will be observed passing through the water.

    The typical cuff inflation volume is 5 to 6 cc (112).

    The blocker from a Univent tube can be removed from the tube

    and inserted alongside a tracheal or tracheostomy tube

    (265,295,296). The blocker is then

    P.650

    guided into place with a fiberscope inserted through the tube.

    View Figure

    Figure 20.18The Univent bronchial blocker can be

    purchased separately from the tube. The connector has

    ports for attachment to the tracheal tube and breathingsystem, for introducing the blocker, and for a fiberscope.

    Note the cap that fits over the end of the blocker.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    56/144

    When the blocker is no longer needed, the cuff is deflated and

    withdrawn into the main tube. If the Univent is to be used for

    postoperative ventilation, the blocker should be fully retracted

    and disabled to avoid inappropriate use by caregivers who are

    unfamiliar with the device (297).

    Eva l ua t i o n

    Most studies show that the Univent provides lung isolation

    equivalent to that of a DLT (11,280,298,299,300,301).

    The Univent may be easier to insert and position correctly than a

    DLT and may be especially useful for the difficult-to-intubate

    patient (287,288,289,302,303,304,305). It can be used in the

    patient with a tracheostomy (254,253,296) and for nasal

    intubation (247). It can be used for postoperative ventilation

    without having to reintubate the patient.

    It is possible to use suction, apply CPAP, or insufflate oxygen

    through the blocker lumen (278,306). The blocker with the cuff

    deflated has been used for jet ventilation during carinal resection

    (307,308).

    Reports of problems with the Univent include the cap becoming

    dislodged from the tip of the blocker ( 309,310) and fragmentation

    of the tube inner wall and connector (311,312).

    The Univent's curved shape is fixed, and this may be a

    disadvantage when sliding it over a bronchoscope. It will not

    soften in a warm water bath (38). It may distort the neck anatomy

    sufficiently to make internal jugular vein cannulation difficult

    (313). Bronchial perforation by the blocker during blind insertion

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    57/144

    has been reported (292). A case has been reported of prolonged

    postoperative atelectasis after using a Univent tube (262). It is

    not recommended for use in children below the age of six years.

    When comparing the Univent with a left DLT, there was a greater

    incidence of malposition with the Univent (119), but comparison

    of the Univent and a right DLT failed to demonstrate a clear

    advantage (35). The Univent is more expensive than a DLT or

    other blockers (79,119,310).

    A disadvantage of the Univent tube is the large amount of cross-

    sectional area occupied by the blocker channel, especially in the

    smaller tubes (236). It is not available in a size that would fit an

    infant or small child.

    Another problem is that the small lumen is re la tively easily

    blocked by blood or pus (285). As a result, blood or pus may

    accumulate and contaminate the dependent lung when theblocker is deflated.

    The blocker's low-volume, high pressure cuff may cause in

    mucosal injury (314,315).

    The larger external diameter may make it difficult to pass the

    Univent between the vocal cords.

    Arndt Bronchial BlockerDesc r i p t i o n

    The Arndt bronchial blocker assembly (wire-guided bronchial

    blocker, WEB, FWEB) is designed to be used for a patient with a

    single-lumen tracheal tube already in place

    (44,111,246,253,316,317,318,319,320,321,322,323). It consists

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    58/144

    of two parts: A blocking catheter and a special airway adaptor.

    Either may be purchased and used separately.

    Blocking Catheter

    The Arndt blocking catheter (Fig. 20.19) has a low-pressure,

    high-volume balloon that has either an elliptical or spherical

    shape. The spherical balloon is relatively compliant unless

    overinflated and takes an elliptical form when inflated in a small

    bronchus. The 9 Fr catheter is available with an elliptical balloon

    that provides a longer sealing profile. It is recommended that the

    elliptical cuff be limited to left mainstem intubation (236,246).

    P.651

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    59/144

    View Figure

    Figure 20.19Arndt bronchial blocker with multiport

    adaptor. The wire loop can be cinched around the tip ofthe fiberscope, or the fiberscope is passed through theloop. The adaptor has ports for attachment to the trachealtube and breathing system, for introducing the blocker,and for a fiberscope.

    A f lexible nylon wire passes through the proximal end of thecatheter and extends to the distal end, then exits as a small loop

    (Fig. 20.19). The size of the loop may be increased or decreased

    by advancing or retracting the wire assembly.

    There are three sizes: 9, 7, and 5 Fr. Characteristics of the

    blockers are shown in Table 20.3. Near the distal end of the 9 Fr

    catheter are side holes to facilitate lung deflation.

    Airway Adaptor

    The multiport adaptor (Fig. 20.19) allows simultaneous

    introduction of a bronchoscope and the blocker while maintaining

    mechanical ventilation. It has four ports:

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    60/144

    A 15-mm female connector that attaches to the tracheal

    tube;

    A side port wi th a 15-mm male connector that attaches to

    the anesthesia breathing system;

    A port angled approximately 30 degrees for the bronchial

    blocker; and

    A port for the f lexible endoscope.

    The blocker port has a Tuohy-Borst type connector to maintain anairtight seal and lock the blocker in place (by tightening the

    connector around the blocker) or to allow free movement of the

    blocker (by loosening the connector). The bronchoscopy port has

    a plastic sealing cap.

    Use

    Before use, the inside of the tracheal tube and the o utside of theblocker and the bronchoscope should be well lubricated with a

    silicone spray. The loop should be adjusted so that it loosely

    approximates the outside diameter of the bronchoscope. The

    blocker balloon should be inflated to test for leaks and then fully

    deflated.

    For each size Web, there is a tracheal tube size for which coaxial

    placement is limited by the diameters of the blocker and

    fiberscope used during placement. When these limits are

    exceeded, parallel placement may be required. Either t he web is

    passed outside the tracheal tube with guide loop inoperable, or a

    fiberscope is also placed in parallel with the tracheal tube before

    or after intubation.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    61/144

    TABLE 20.3 Arndt Endobronchial Blockers

    Size(French)

    Smallest Single-lumenTube In ternal Diameter

    for Coaxial Use

    (millimeters)

    Length(centimeters)

    Cuf f Shape Average Cuf fI nfl ation Volume

    (cubic cent imeters)

    9 7.5 78 and

    65

    Elliptical 6 to 12

    Spherical 4 to 8

    7 6.0 65 Spherical 2 to 6

    5 4.5 65 and

    50

    Spherical 0.5 to 2.0

    From Klafta JM. One-lung anesthesia; making it work (ASA Refresher Course#509). Park Ridge, IL: ASA, 2004.

    P.652

    The wire in the blocker lumen is used in either of two ways: (i)

    when cinched tightly around the tip of the fiberscope, the

    fiberscope carries the blocker into its desired location, and (i i)

    when the fiberscope is passed through the loop, it provides a

    track for blocker to pass through after the fiberscope is placed

    into the bronchus.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    62/144

    After the patient is intubated and venti lation has begun, the

    multiport adaptor is connected between the breathing system and

    the tracheal tube. The bronchoscope is advanced through the

    guide loop. This allows the blocker to follow the bronchoscope.

    Alternately , the wire loop may be placed over the end of the

    blocker prior to attaching the airway adaptor to the tracheal tube

    (324).

    The bronchoscope is advanced into the airway to be blocked and

    then the guide loop is slid over the end of the bronchoscope. The

    bronchoscope is withdrawn slightly to visualize the blocker. The

    blocker is then advanced or withdrawn into position (324,325). It

    may be advisable to advance the blocker approximately 1 cm

    beyond the optimal position when the patient is in the supine

    position to avoid dislodging the blocker toward the trachea while

    the patient's position is changed to the lateral decubitus position(246). The balloon is then inflated under direct vision. The

    balloon should fil l the entire bronchial lumen and not herniate

    into the trachea. Following placement, the balloon may be

    deflated until one-lung ventilation is required.

    Removing the wire loop will result in an open channel, which

    allows CPAP application, oxygen insufflation, suctioning, or

    intermittent inflation. Leaving the wire loop during the operation

    might damage the airway and entails the risk that the loop may

    accidentally be stapled into the bronchial closure (326). However,

    leaving the loop in place allows the blocker to be repositioned

    (327). It may be possible to reposition the blocker by using

    fiberoptic endoscopy (326). Excessive force should not be used

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    63/144

    when removing the wire guide, because this may displace the

    blocker.

    Lung collapse can be expedited by attaching a syringe or

    applying suction to the blocker channel (44,328). However, some

    clinicians do not recommend using suction through this channel

    because of the risk of developing negative pressure pulmonary

    edema (326). When no longer needed, the blocker cuff should be

    fully deflated and the blocker removed.

    Eva l ua t i o n

    The Arndt blocker system can be used in the patient who is

    already intubated (316,317,329), in the patient with a

    tracheostomy (318,324,325), or with a nasal intubation (253,318).

    It has been used to provide single-lung ventilation in children as

    young as 17 months (245). It may be especially useful for the

    patient in whom a DLT would be difficult to use

    (260,316,318,319,320,321,330). It has been used in a newborn

    (322). It allows a larger internal cross-sectional area than a DLT

    or Univent tube of similar outside diameter ( 331). It may require

    fewer insertion attempts than a DLT (332). If the wire is removed,

    the lumen can be used for suctioning or administering oxygen or

    CPAP.

    P r o b l ems

    A disadvantage is that once the wi re loop is removed, it cannot

    be reinserted through the channel to allow repositioning of the

    blocker. Placement requires the availability of fiberoptic

    equipment and someone able to use it. It takes longer to position

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    64/144

    and to achieve complete lung collapse compared with the Univent

    tube or a DLT (328,332). The balloon may be sheared when it is

    removed from the blocker port (333).

    Cohen Tip-deflecting Bronchial Blocker

    The Cohen tip-deflecting bronchial blocker (Fig. 20.20) has a 9 Fr

    external diameter and a central lumen with a 1.6 mm diameter

    (334). The high-volume, low-pressure blue balloon at the tip is

    spherical in shape. The average inflation volume is 5 to 8 mL.

    There are side holes between the tip and the balloon to evacuate

    gas from the distal lung or to insufflate oxygen. A proximal

    control wheel that can be operated with the thumb and forefinger

    is used to adjust tip deflection (Fig. 20.21). The catheter has

    depth markings and an indicator arrow that shows the direction in

    which the tip deflects.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    65/144

    View Figure

    Figure 20.20Cohen tip-deflecting bronchial blocker. The

    proximal control wheel is used to adjust tip deflection. Anarrow on the wheel indicates the direction to which the tipdeflects. (Courtesy of Cook Critical Care.)

    P.653

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    66/144

    View Figure

    Figure 20.21The tip of the Cohen bronchial blocker can

    be manipulated to fit into either bronchus.

    In most cases, the blocker and a fiberscope are inserted throughan appropriately sized tracheal tube. The blocker can also be

    placed outside the tracheal tube and guided into place with a

    bronchoscope placed through the tracheal tube.

    A multiport airway adaptor li ke the Arndt is attached to the

    breathing system end of the tracheal tube. Alternatively, a

    standard swivel adaptor can be used for insertion.

    An assistant with or without video f lexible fiberoptic scope may

    be necessary because it may require two hands to manipulate the

    blocker into position, one to deflect the tip and the other to rotate

    and advance the catheter.

    Embolectomy Catheter

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    67/144

    Desc r i p t i o n

    An embolectomy (Fogarty) catheter can be used as a bronchial or

    segmental blocker

    (1,14,53,62,226,248,251,256,257,267,268,269,275,324,335,336,3

    37,338,339,340,341,342,343,344). This catheter is readily

    available in most operating suites where vascular surgery is

    performed or can be purchased for this purpose. It comes with a

    stylet in place so that it is possible to place a curvature in the

    distal tip to facilitate guidance to the target bronchus. The

    occlusion balloon has a high-pressure, low-volume cuff (246). It

    comes in a variety of sizes. Adult bronchi can be blocked with 7

    Fr catheters, whereas 2 Fr to 5 Fr catheters are suitable for

    segmental or pediatric blockade.

    Use

    Prior to use, the blocker should be lubricated with jelly or silicone

    spray to facilitate passage. The balloon should be tested for

    leaks and then fully deflated. A catheter with a faulty or eccentric

    balloon should be discarded.

    The embolectomy catheter may be placed before or after

    intubation with a single-lumen tracheal or tracheostomy tube and

    can be passed either through or alongside the tube. Placing the

    catheter alongside the tube allows the tube to splint the blocker

    in position. An alternative method is to insert the catheter

    through a hole made in the side of the tracheal tube (345).

    If the patient is already intubated, the catheter containing a stylet

    may be passed through a fiberoptic endoscope adaptor

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    68/144

    (335,341,342). This allows uninterrupted ventilation. The Arndt

    multiport airway adaptor (previously mentioned) or similar device

    (346) may also be used to prevent an air leak. An annular space

    around the fiberoptic endoscope and embolectomy catheter

    equivalent to a single-lumen tube with an inner diameter of 4 to 5

    mm will be necessary to allow exhalation in a reasonable period

    of time (347).

    A f iberoptic endoscope is passed down the tracheal tube, and the

    embolectomy catheter is guided into the appropriate bronchus

    under direct vision. Twirl ing the catheter between the fingers at

    the proximal end or rotating the tracheal tube will impart lateral

    direction to the tip (98). Additional lateral direction may be

    gained by rotating the tracheal tube to one side or the other.

    After the tip is advanced into the proper position, the stylet is

    removed, the catheter balloon is inflated under direct vision, andthen the fiberscope withdrawn.

    A modification of this technique is to deliberately intubate a

    mainstem bronchus with a tracheal tube one size smaller than

    would be appropriate for the trachea, advance the blocker

    through the tracheal tube, then withdraw the tracheal tube into

    the trachea (348,349,350,351). It may be helpful to preshape the

    tracheal tube by using a stylet (351). Once the catheter is in the

    targeted bronchus, the stylet is removed. Removing the tracheal

    tube followed by insertion of a larger tracheal tube beside the

    Fogarty catheter will securely fix the catheter in place. The

    catheter balloon is inflated slowly until no air enters the blocked

    lung, as detected by using auscultation. The Fogarty catheter

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    69/144

    should be securely taped to the tracheal tube to prevent it from

    being dislodged. A fiberoptic endoscope should be used to check

    the position of the blocker. The catheter ballooon should be

    deflated and the lung collapsed by using pressure on the chest

    and/or suction through the tracheal tube (351). The catheter

    balloon should then be reinflated to the same volume as

    previously used.

    Eva l ua t i o n

    The use of an embolectomy catheter has many advantages. It

    can be passed through a single-lumen tracheal tube in a n already

    intubated patient, and there is no need for reintubation if

    postoperative mechanical ventilation is needed. It may be useful

    in

    P.654

    pediatric patients (62,236,267,268,324,337,338,340,345,352), the

    patient with a tracheostomy (252,336,351,353), or for a nasal

    intubation. The embolectomy catheter is less expensive than a

    DLT, Univent tube, or Arndt blocker (326,339). Fogarty catheters

    are relatively thinner for a given balloon volume when compared

    with other bronchial blockers and thus will allow the use of larger

    tracheal tubes, especially in pediatric patients when they are

    placed side by side in the trachea (351).

    P r o b l ems

    A significant disadvantage is the lack of a hollow center.

    Suctioning, oxygen insufflation, or applying CPAP to the blocked

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    70/144

    lung is not possible, and lung collapse takes longer and may not

    be as complete as with a DLT or a blocker with a hollow lumen

    (62). The obstructed lung segment cannot be re-expanded until

    the blocker is removed. Another disadvantage is that it is made

    of latex, so it cannot be used in the patient with potential latex

    allergy (Chapter 15). Most of these devices have low-volume,

    high-pressure cuffs and can damage the airway, although the

    pressure may be less than that exerted by the cuff on a DLT

    (109,236).

    There are no reported complications with the embolectomy

    catheter in adults (246). Bronchial rupture has been reported with

    an overinflated balloon in a child (337).

    Other

    Balloon atrioseptostomy, Foley urinary, and Swan Ganz

    pulmonary artery catheters have been used as bronchial

    blockers. They have a central lumen for suction or administration

    of oxygen to the blocked lung.

    P rob l ems w i t h B r on ch i a l B l o c k er s

    A blocker may not be suitable for the patient with a high r ight

    upper lobe bronchus takeoff or a tracheal bronchus (173,354). It

    may be necessary to either use a second blocker or a DLT under

    these conditions. If the distance between the carina a nd the

    tracheal bronchus is small, it may be possible to herniate the

    balloon on the bronchial blocker so that it blocks the upper lobe

    bronchus (355). It may be possible to use the tracheal cuff of the

    Univent tube to block a tracheal bronchus (356).

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    71/144

    A blocker may be dislodged into the trachea (357). The inf lated

    balloon may then block ventilation to both lungs, prevent collapse

    of the operated lung, and/or ca use air trapping. Other

    complications reported include fixation by surgical staples (358);

    perforation of the blocker balloon by a surgical needle ( 359); and

    accidental inflation of the blocker cuff when it was just below the

    tip of the tube, resulting in obstruction to gas flow (360).

    The narrow blocker lumen may result in ineffective removal of

    secretions and pulmonary soiling after the cuff is deflated (361).

    Lung deflation may not be as satisfactory or achieved as rapidly

    with a bronchial blocker as with a DLT (332).

    Development of severe hypoxemia has been reported after

    continuous suctioning of the nondependent lung through a

    bronchial blocker (362). If suction is used to facilitate lung

    collapse, it should be applied for only a few seconds,intermittently and with low pressure (246).

    References

    1. Camci E, Tugrul M, Tugrul ST, et al. Techniques and

    complications of one-lung ventilation in children with suppurative

    lung disease: experience in 15 cases. J Cardiothorac Vasc

    Anesth 2001;15:341345.

    [Medline Link]

    2. Morell RC, Prielipp RC, Foreman AS, et al. Intentional

    occlusion of the right upper lobe bronchial orifice to tamponade

    life-threatening hemoptysis. Anesthesiology 1995;82:1529

    1531.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    72/144

    [Fulltext Link]

    [CrossRef]

    [Medline Link]

    3. Shivaram U, Finch P, Nowak P. Plastic endobronchial tubes in

    the management of life-threatening hemoptysis. Chest

    1987;92:11081110.

    [CrossRef]

    [Medline Link]

    4. Cohen E. Nitric oxide, thoracoscopy and other new trends in

    thoracic anesthesia (ASA Refresher Course #175). Dallas, TX:

    ASA, 1999.

    5. Mangar D. Treatment of pulmonary artery haemorrhage with a

    single lumen tracheal tube. Can J Anaesth 1994;41:880.

    [Medline Link]

    6. Arya VK, Dutta A, Chari P, et al. Difficult retrogradeendotracheal intubation: the utility of a pharyngeal loop. Anesth

    Analg 2002;94:470473.

    [Fulltext Link]

    [CrossRef]

    [Medline Link]

    7. Inoue H, Shohtsu A, Ogawa J, et al. Endotracheal tube with

    moveable blocker to prevent aspiration of intratracheal bleeding.

    Ann Thorac Surg 1984;37:497499.

    [Medline Link]

    8. Herenstein R, Russo JR, Mooka N, et al. Management of one-

    lung anesthesia in an anticoagulated patient. Anesth Analg

    1988;67:11201122.

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    73/144

    [Medline Link]

    9. Kabon B, Waltl B, Leitgeb J, et al. First experience with

    fiberoptically directed wire-guided endobronchial blockade in a

    severe pulmonary bleeding in an emergency setting. Chest

    2001;120:13991402.

    [CrossRef]

    [Medline Link]

    10. Klafta JM, Olson JP. Emergent lung separation for

    management of pulmonary artery rupture. Anesthesiology

    1997;87:12481250.

    [Fulltext Link]

    [CrossRef]

    [Medline Link]

    11. Inoue H, Suzuki I, Mwasaki M, et al. Selective exclusion of

    the injured lung. J Trauma 1993;34:496498.[CrossRef]

    [Medline Link]

    12. McGuire GP. Lung ventilation and bronchopleural fistula. Can

    J Anaesth 1996;43:12751276.

    [Medline Link]

    13. Darwish RS, Gilbert TB, Fahy BG. Management of a

    bronchopleural fistula using differential lung airway pressure

    release ventilation. J Cardiothorac Vasc Anesth 2003;17:744

    746.

    [CrossRef]

    [Medline Link]

  • 8/11/2019 Chapter 20. Lung Isolation Devices

    74/144

    14. Otruba Z, Oxorn D. Lobar bronchial blockade in

    bronchopleural fistula. Can J Anaesth