Chapter 20. Lung Isolation Devices
Transcript of Chapter 20. Lung Isolation Devices
-
8/11/2019 Chapter 20. Lung Isolation Devices
1/144
Chapter 20
Lung Isolation DevicesP.634
Indications for Lung Isolation
Tho rac i c P roc edu r es
A def la ted lung provides better operating conditions and reduced
trauma during thoracic procedures.
Con t r o l o f Con t am inat i o n o r Hem o r r hage
A lung isolation device can prevent infected material from one
lung from contaminating the other lung (1). W hen hemorrhage
occurs in one lung, an isolation device allows the unaffected lungto be ventilated (2,3,4,5,6,7,8,9,10,11).
Un i l a te r al Patho l og y
A bronchopleural or bronchocutaneous fistula may have such a
low resistance to gas flow that most of t he tidal volume passes
through it, making it impossible to adequately ventilate the other
lung (12,13,14). Large cysts or bullae may rupture under positive
pressure, making it mandatory that they be excluded from
ventilation. Another indication for lung separation is when lungs
have markedly different compliance or airway resistance such as
that which occurs following single-lung transplantation or
unilateral injury (15).
-
8/11/2019 Chapter 20. Lung Isolation Devices
2/144
The best method of producing lung isolation in a given situation
will depend on several factors, including the indication for lung
isolation, patient variables, available equipment, and the skill and
training of the anesthesia provider.
Anatomical Considerations
The right mainstem bronchus is shorter, straighter, and has a
larger diameter than the left. It takes off from the tr achea at an
angle of 25 degrees in adults. The left mainstem bronchusdiverges from the median plane at a 45-degree angle. These
angles are slightly larger in children (16). The right upper lobe
bronchus takeoff is very close to the origin of the right mainstem
bronchus. These anatomical features mean that it is usually
easier to intubate the right mainstem bronchus than the left, but it
is more difficult to place a tube in the right mainstem bronchus
without obstructing the upper lobe orifice.
Double-lumen Tubes
The double-lumen tube (DLT, DLET) is the device most
commonly used to provide lung isolation.
Desc r i p t i o n
A DLT is essentially two single- lumen tubes bonded together and
designated either as right- or left-sided, depending on which
mainstem bronchus the tube is designed to fit. The tracheal
lumen is designed to terminate above the carina. The distal
portion of the bronchial lumen is angled to fit into the appropriate
mainstem bronchus.
-
8/11/2019 Chapter 20. Lung Isolation Devices
3/144
The internal lumen of each tube is D-shaped with the straight
side of the D in the middle of the tube. The resistance of each
lumen of a 35 to 41 French (Fr) DLT has been found to be
comparable to the resistance of a size 6 to 7 mm internal
diameter single-lumen tracheal tube (17).
The bronchial cuff for right-sided tubes varies in shape,
depending on the manufacturer. On some tubes, the cuff has a
slot to allow ventilation of the right upper lobe. Some right-sided
DLTs have two bronchial cuffs with an opening for the right upper
lobe between them. The resting volume and compliance of
bronchial cuffs varies between different sizes and brands of DLTs
(18,19). Most manufacturers color the bronchial cuff blue. They
also use blue markings on the pilot balloon and/or the inflation
device for the bronchial cuff.
A few DLTs have a carinal hook to aid in proper placeme nt andminimize tube movement after placement. Potential problems
with carinal hooks include increased difficulty during intubation,
trauma to the airway, malposition of the tube because of the
hook, and interference with bronchial closure during
pneumonectomy. The hook can break off and become lost in the
bronchial tree.
Most manufacturers place a radiopaque marker at the bottom of
the tracheal cuff or at the end of the tracheal lumen. Other marks
may be placed above and/or below the bronchial cuff. Some have
a radiopaque line running the length of the tube.
Disposable DLTs are supplied in sterile packages, which include
a stylet, connectors, and suction catheter(s). A means to supply
-
8/11/2019 Chapter 20. Lung Isolation Devices
4/144
continuous positive airway pressure (CPAP) may be included with
the tube or can be purchased separately.
The connector (20) allows both lumens to attach to a breathing
system at the same time.
S iz i ng
Adult DLTs commonly come in sizes 35, 37, 39, and 41 Fr. The
French scale is the ex ternal diameter of the tracheal segment
times three. Some manufacturers also provide 26, 28, and 32 Frtubes for younger patients. Unfortunately, the French gauge
markings are of limited value in determining the most important
measurement-the diameter of the bronchial segment.
There are major variations between manufacturers in the
dimensions of the bronchial segment of DLTs of the same
nominal size and even among tubes of the
P.635
same size from the same manufacturer (21,22,23,24). An
International Technical Specification has recommended that
outside circumference of the bronchial segment be designated in
millimeters (25). The diameter of the bronchial segment with the
cuff inflated may not increase with tube size (26).
Marg i n o f Sa fet y
The margin of safety for a DLT is the length of the
tracheobronchial tree between the most distal and proximal
acceptable positions (27,28). The margin of safety will depend on
the length of the lumen into which the cuff is placed and the
-
8/11/2019 Chapter 20. Lung Isolation Devices
5/144
length of the cuff. If the cuff is short or the mainstem bronchus
long, the margin of safety will be greater.
Left-sided Tubes
The outermost acceptable position for a left-sided DLT is when
the bronchial cuff is just below the carina. If the tube were more
proximal, the bronchial cuff could obstruct the trachea and/or the
contralateral (right) mainstem bronchus. In this case, the seal
between the two lungs would be lost. The most acceptable distal
position is when the the bronchial segment tip is at the proximal
edge of the upper lobe bronchial orifice. More distal insertion
would result in obstruction of the upper lobe bronchus.
The average length of the left bronchus from the carina to the
takeoff of the upper lobe bronchus is 5.6 cm. This leaves a
relatively small margin for placement, as there could be up to 3.5
cm of movement with neck flexion and extension (29). There is
great variability in the length of the bronchial segment of left-
sided DLTs currently available (30).
Right-sided Tubes
The margin of safety is defined differently for right-sided tubes. A
right-sided DLT is acceptably positioned if the right upper lobe
ventilation opening or the space between the two cuffs is aligned
with the right upper lobe orifice. Thus, the margin of safety is the
length of the ventilation opening minus the diameter of the
orifice. The margin of safety for right-sided tubes is considerably
less than for left-sided tubes.
Spec i f i c Tub es
-
8/11/2019 Chapter 20. Lung Isolation Devices
6/144
Carlens Double-lumen Tube
The Carlens DLT (Figs. 20.1, 20.2) is intended to be inserted into
the left mainstem bronchus. It has a carinal hook (spur). This
tube may be especially useful with massive hemoptysis when
verification of tube placement is especially difficult (31).
View Figure
Figure 20.1Carlens double-lumen tube. A:Theconnector has ports for fiberscope insertion or suctioning
and areas where a clamp can be applied to occlude gas
flow. B:Note the carinal hook and the blue bronchialcuff.
P.636
-
8/11/2019 Chapter 20. Lung Isolation Devices
7/144
View Figure
Figure 20.2Carlens tube in place.
White Double-lumen TubeThe White DLT (Fig. 20.3) is designed to fit the right mainstem
bronchus. It has a carinal hook. The cuff for the right mainstem
bronchus is circumferential superior to the o pening to the upper
lobe bronchus and continues distally behind the opening.
Robertshaw Right Double-lumen Tube
The bronchial portion of the Robertshaw right DLT (Fig. 20.4) is
angled at 20 degrees (32). The bronchial cuff has a slotted
opening in its lateral aspect (Fig. 20.5). The bronchial cuff is
proximal to the slot on the lateral surface and extends
tangentially toward the medial surface.
Robertshaw Left Double-lumen Tube
-
8/11/2019 Chapter 20. Lung Isolation Devices
8/144
The Robertshaw left DLT is shown in Figures 20.6and 20.7. The
angle of the bronchial portion is 40 degrees (32). The average
length of the bronchial segment is 23 mm for sizes 35/37 and 25
mm for sizes 39/41 (30).
Broncho-Cath Right-sided Tube
The Broncho-Cath right-sided tube has a bronchial cuff that is
roughly the shape of an S, or slanted doughnut. The cuff edge
nearest the right upper lobe bronchus is closer to the trachea
than the part of the cuff touching the medial bronchial wall ( Figs.
20.8, 20.9). A s lot in the tube just beyond the cuff corresponds to
the opening of the upper lobe bronchus (27). The end of the
bronchial segment has no bevel (33).
The shape of t he right bronchial cuff allows the venti lation slot to
ride off the right upper lobe orifice, increasing the margin of
safety. One study found that when this tube was inserted blindly,
right upper lobe obstruction occurred in 89% of cases (34).
Placement using fiberoptic endoscopy resulted in much better
positioning (35).
-
8/11/2019 Chapter 20. Lung Isolation Devices
9/144
View Figure
Figure 20.3White double-lumen tube. (Picture courtesy
of Teleflex Medical.)
Broncho-Cath Left-sided TubeOn the Broncho-Cath left-sided DLT (Fig. 20.10), the bronchial
portion of the tube is at an angle of approximately 45 degrees
(33,36,37). The bronchial portion has a curved tip. The bevel was
removed in 1994 but reintroduced in 2001 (33,38). The average
length of the bronchial segment is 30 to 31 mm (30). The tube is
available with or without a carinal hook.
Sher-I-Bronch Right-sided Double-lumen
Tube
The Sher-I-Bronch right-sided DLT has two cuffs on the bronchial
segment, one proximal and one distal to the upper lobe
-
8/11/2019 Chapter 20. Lung Isolation Devices
10/144
ventilation slot. The slot is 13 to 14 mm long (Figs. 20.11, 20.12).
One study found that poor lung isolation was more common when
this tube was used (39). A case has been reported in which the
tube tip became trapped in the right upper lobe bronchus (40).
View Figure
Figure 20.4Robertshaw right double-lumen tube inplace.
P.637
-
8/11/2019 Chapter 20. Lung Isolation Devices
11/144
View Figure
Figure 20.5Robertshaw right double-lumen tube in
place.
Sher-I-Bronch Left-sided Double-lumen TubeThe bronchial segment of the Sher-I-Bronch left-sided DLT
diverges from the main tube at an angle of 34 degrees. The
average length of the bronchial segment is 35 mm (30). The
bronchial segment has a bevel (30). One study found that the
bronchial cuff on this tube required significantly higher pressures
to achieve one-lung isolation than cuffs on other DLTs (41).
Silbroncho Double-lumen Tube
The Silbroncho DLT (Fig. 20.13) is made of si l icone, which is
softer than polyvinylchloride (PVC). The tube does not contain
latex. The bronchial segment is wire-reinforced distal to the
tracheal cuff. This allows it to be flexible and prevents kinking.
-
8/11/2019 Chapter 20. Lung Isolation Devices
12/144
The wire reinforcement also makes tube position easy to
determine on x-ray. The bronchial cuff is small and near the end
of the tube. It is available in sizes 33, 35, 37, and 39 Fr.
View Figure
Figure 20.6Robertshaw left double-lumen tube.(Courtesy of Rusch, Inc.)
-
8/11/2019 Chapter 20. Lung Isolation Devices
13/144
View Figure
Figure 20.7Robertshaw left double-lumen tube.
Techn i ques
Tube Choice
R i gh t ve r su s Le f t
When surgery is performed on the right lung, a left-sided DLT
should be used (42,43,44). Because the margin of safety in
positioning a right-sided DLT is so small, some prefer to use aleft-sided DLT whenever possible for left lung surgery (28,45).
During left pneumonectomy, immediately before the left mainstem
bronchus is clamped, the DLT can be pulled from the bronchus
and used for ventilating the right lung. A disadvantage of this
technique is the risk of blood and secretions moving from the
-
8/11/2019 Chapter 20. Lung Isolation Devices
14/144
operative (left) bronchus to the nonoperative bronchus. Other
possible problems include the tube becoming dislodged or
sutured in place during surgery. A ball-valve obstruction may
occur as a result of secretions or mediastinal pressure pushing
the tracheal lumen against the tracheal wall ( 46). A left DLT ma y
not provide optimum conditions for ventilating the residual lung
after left upper lobectomy (47).
A r ight-sided DLT should be used when i t is important to avoid
manipulation/intubation of the left main bronchus (e.g., an
exophytic lesion), when the left main bronchus is narrowed or the
left mainstem bronchus is so cephalad that the bronchial lumen
will not enter the
P.638
left mainstem bronchus, left pneumonectomy, left lungtransplantation, left mainstem bronchus stent in place or when
there is tracheobronchial disruption on the left (43,48,49,50,51).
-
8/11/2019 Chapter 20. Lung Isolation Devices
15/144
View Figure
Figure 20.8Broncho-Cath right double-lumen tube.
Newer versions have no bevel at the end of the bronchialsegment. (Courtesy of Mallinckrodt Medical, Inc.)
Before placing a right DLT, the patient's chest x-ray or computedtomography scan can be closely examined to id entify a right
upper lobe bronchus takeoff, which would make it difficult to use
a right DLT. A left DLT should be placed in the right mainstem
bronchus in Kartagener's syndrome, which includes complete
situs inversus and a longer-than-normal right mainstem bronchus
(52).
-
8/11/2019 Chapter 20. Lung Isolation Devices
16/144
View Figure
Figure 20.9Broncho-Cath right double-lumen tube. The
bronchial cuff has the shape of an S or a slanteddoughnut, with the edge of the cuff nearest the rightupper lobe bronchus closer to the trachea than the part ofthe cuff touching the medial bronchial wall. A slot in thetube beyond the cuff corresponds to the opening of theright upper lobe bronchus. Newer versions have no bevelon the bronchial segment.
Size
Selecting an appropriately sized DLT for a given patient is critical
to minimize the frequency of complications (53).
A DLT that is too small may fai l to provide lung isolation or may
require bronchial cuff volumes and pressures that could produce
mucosal ischemia or bronchial rupture. Using too small a DLT
can result in the tube advancing too far into the bronchus, a
higher level of autoPEEP (positive end-expiratory pressure), or
barotrauma (54,55). An undersized tube may be more likely to be
displaced. Ventilation and suctioning are more difficult with a
small tube.
-
8/11/2019 Chapter 20. Lung Isolation Devices
17/144
Using a large DLT wil l result in less resistance to gas flow,
facilitate suctioning and passage of a fiberscope, and reduce the
risk of advancing the DLT too far into t he bronchus but may
result in trauma (56). Inability to insert a larger tube through the
larynx or past the carina or intrinsic or extrinsic obstruction in the
mainstem bronchus to be intubated may necessitate use of a
smaller tube.
A tube is oversized if the bronchial lumen wi l l not fi t into the
bronchus or there is no air leak with the bronchial cuff deflated
(22,38,57,58,59,60). It is too small if the bronchial cuff inflation
volume is greater than the resting cuff volume. Not more than 3
cc of air in the bronchial cuff should be required to create a seal.
When there is a high risk that f luids wil l seep past the bronchial
cuff, a smaller DLT should be used (61).
In adults, the dimensions of the cricoid ring best define those ofthe main bronchi (23). In children, age but
P.639
P.640
not weight is a predictor of bronchial diameter, and the right main
bronchial diameter is significantly larger than the left ( 62).
Suggested sizes for DLTs in children are shown in Table 20.1.
Tubes from different manufacturers vary in size and may not fit
this table.
-
8/11/2019 Chapter 20. Lung Isolation Devices
18/144
View Figure
Figure 20.10Broncho-Cath left double-lumen tube.
(Courtesy of Mallinckrodt Medical, Inc.)
-
8/11/2019 Chapter 20. Lung Isolation Devices
19/144
View Figure
Figure 20.11Sher-I-Bronch double-lumen tubes. A:
Left-sided tube (top). Right-sided tube (bottom). B:Close-up of right bronchial segment, showing opening tothe right upper lobe. (Courtesy of Sheridan, Inc.)
-
8/11/2019 Chapter 20. Lung Isolation Devices
20/144
-
8/11/2019 Chapter 20. Lung Isolation Devices
21/144
Age (years) Doubl -lumen Tube (French)
8 to 10 26
10 to 12 26 to 28
12 to 14 32
14 to 16 35
From Hammer GB, Fitzmaurice BG, Brodsky JB. Methods for single-lungventilation in pediatric patients. Anesth Analg 1999;89:14261429.
The size of the mainstem bronchus may be determined by
measuring the width of the patient's bronchus from a chest x-ray
or computed tomographic scan (22,54,60,63,64,65,66,67,68).
Unfortunately, it is not possible to accurately measure bronchial
width on many chest x- rays, and the correlation between the
tracheal and bronchial size may not be reliable enough to
determine the proper size DLT from the size of the trachea(56,59,63,69,70,71,72). Reliability may be increased by
measuring the tracheal diameter both anteropostally and
mediolaterally (73). The lung transplant patient may have a
significantly different sized bronchus than would have been
predicted from the size of the trachea (56).
-
8/11/2019 Chapter 20. Lung Isolation Devices
22/144
Preparing the Double-lumen Tube
The tracheal and bronchial cuffs should be inflated and checked
for leaks and symmetrical cuff inflation, making certain that each
inflation tube is associated with the proper cuff. The cuffs and
stylet should be lubricated with a water-soluble lubricant and the
stylet placed
P.641
inside the bronchial lumen, making certain that it does not extend
beyond the tip. The connector should be assembled so that it can
be quickly fitted to the tube and breathing system after
intubation.
View Figure
Figure 20.13A:The Silbroncho double-lumen tube. B:The bronchial segment is wire-reinforced distal to thetracheal cuff.
-
8/11/2019 Chapter 20. Lung Isolation Devices
23/144
Insertion
The DLT is advanced through the larynx with the angled tip
directed anteriorly. After the bronchial cuff has passed the cords,
the tube is turned 90 degrees so that the bronchial portion points
toward the appropriate bronchus. If the tube is to be placed in the
left mainstem bronchus, the head and neck should be rotated to
the right before rotating and advancing the tube (74). Leaving the
stylet in place for the entire intubation procedure rather than
removing it once the bronchial cuff has passed the vocal cords
may result in more rapid and accurate placement ( 75). However,
some recommend that the stylet be removed just after the tube
passes the vocal cords to prevent trauma (76).
A DLT is most accurately placed by inserting a fiberscope into
the bronchial lumen and directing it into the appropriate bronchus
under direct vision (77,78,79,80,81,82). Concurrent direct
laryngoscopy may be required to elevate the supraglottic tissues
to facilitate passing the DLT through the glottic opening after the
fiberscope is in the trachea (83). This ensures that the correct
bronchus is intubated on the first attempt and avoids inserting
the tube too deeply or the tube becoming kinked in the upper
lobe bronchus (40).
A DLT wi th a carinal hook should be inserted with the bronchial
segment concave anteriorly until the bronchial cuff passes the
cords. It should then be rotated 180 degrees so that the hook is
anterior and advanced until the hook passes the vocal cords. The
tube is then advanced until the hook engages the carina. The
-
8/11/2019 Chapter 20. Lung Isolation Devices
24/144
hook can be tied closely to the tube with a slip knot to facilitate
passage through the larynx and then untied.
Inserting the bronchial portion into the bronchus can be
performed blindly after insertion through the vocal cords. In some
cases (e.g., bronchorrhea, bleeding), blind placement may
succeed where the fiberoptic technique does not (79).
With blind insertion, the correct depth of insertion may be difficult
to determine. However, this method may be useful where rapid
lung isolation or collapse is necessary. In adults, there is a
correlation between the ideal depth of insertion of left DLTs and
patient height but not weight or age (24,84,85,86,87). The ideal
depth of insertion can be estimated from the chest x-ray
(84,85,86). Advancing the tube with the bronchial cuff partially
inflated unti l an increase in resistance is felt (or only one side of
the chest moves and compliance is reduced) may preventinserting the tube too deeply (88,89,90,91). The bronchial cuff is
then deflated and the tube advanced a distance equal to the
length of the bronchial cuff plus 1 to 1.5 cm to place it just
beyond the carina (92,93). Depth of insertion may be estimated
by monitoring bronchial cuff pressure (94).
Since the tube usually moves upward with positioning, some
clinicians recommend that the tube should initially be inserted
more deeply than would be the ideal position (36,38,95,96).
Others believe that the danger of trauma is increased if the tube
is intentionally placed too deeply (97).
If the patient is anatomically difficult to intubate, a single-lumen
tracheal tube may be placed by any of the means that facilitate
-
8/11/2019 Chapter 20. Lung Isolation Devices
25/144
difficult intubations (98). An exchange catheter can then be
inserted into the single-lumen tube and the DLT inserted over the
catheter after the single-lumen tube has been withdrawn (99).
The WuScope (Chapter 1 8) has been successfully employed to
place a DLT in a patient who was difficult to intubate (100). A
DLT can be placed by using a lighted stylet (Chapter 19) in the
bronchial lumen (101) or a retrograde intubation technique (102).
Awake fiberoptic bronchial intubation with a DLT has been
reported (103). A DLT ma y be introduced over a gum elastic
bougie (45,104,105,106). It may be helpful to pass the bougie
through the opening for the right upper lobe orifice on right DLTs
(105).
A DLT may be inserted through a tracheostomy (107,108).
Cuff Inflation
Once the tip is thought to be in a mainstem bronchus, the
tracheal cuff should be inflated in a manner similar to that of a
tracheal tube (109,110). It is more difficult to inflate the bronchial
cuff correctly. An overinflated bronchial cuff is more likely to
herniate into the trachea, cause the carina to be pushed toward
the opposite side, or result in narrowing of the bronchial segment
lumen. Inflating the bronchial cuff beyond its resting volume may
result in dangerously high pressure (18,111).
The bronchial cuff should be inflated with small incremental
volumes until an airtight seal is just attained (109,112). The total
volume should be less than 3 mL (41). One technique is to
immerse the proximal tracheal lumen in water during ventilation
-
8/11/2019 Chapter 20. Lung Isolation Devices
26/144
via the bronchial lumen. The bronchial cuff is inflated until no
bubbles are seen escaping during positive-pressure inspiration
(110,111) (Fig. 20.14). Variations of this are to connect a balloon
(113) or a capnograph (111) to t he tracheal lumen. Another
method is to apply suction to the tracheal lumen (110). Absence
of bronchial seal will cause the reservoir bag in a breathing
system that is connected to the bronchial lumen to collapse.
Inflating the bronchial cuff to an airtight seal may not prevent the
spread of blood or secretions (61). The bronchial cuff may also
be inflated with water (114).
Confirming Position
Confirming proper tube position is essential because the tube
may not perform properly if incorrectly
P.642
positioned. Position should be checked after insertion, after
repositioning the patient, and before beginning one-lung
ventilation, as these tubes often move during patient positioning
or surgical manipulations
(97,115,116,117,118,119,120,121,122,123). The most frequent
DLT movement is during lateral decubitus positioning. While
movement is usually outward, distal migration may also occur.
DLT position should be confirmed whenever there is evidence of
malfunction.
-
8/11/2019 Chapter 20. Lung Isolation Devices
27/144
View Figure
Figure 20.14Inflating the bronchial cuff. With the
tracheal cuff inflated, the bronchial cuff is slowly inflated.Right:A bronchial cuff leak is indicated by bubbles whenthe end of the tracheal lumen is placed under water. Left:With a seal, no bubbles appear.
Au s cu l t a t o r y Tech n i q u e s
Unfortunately, auscultation detects DLT malposition only part of
the time because breath sounds can be transmitted from one
region of the lung to adjacent areas (124,125,126,127). Studies
have shown that a significant percentage of DLTs thought to be
positioned satisfactorily by auscultation were found to be
inappropriately positioned on subsequent fiberoptic examination
(116,120,128). A DLT may function satisfactorily although not in
an ideal position. Another problem with auscultation is that once
the patient is prepped and draped, the chest is no longer
available for auscultation. One study found that auscultatory
-
8/11/2019 Chapter 20. Lung Isolation Devices
28/144
placement was not associated with an increased incidence of
complications during one-lung ventilation (39).
Clamping Method
Le f t -s i d ed Tubes
With only the tracheal cuff inflated and the tracheal lumen
connected to the breathing system, both lungs should be
auscultated in the axillary regions and upper lung fields to detect
differences. The bronchial cuff should then be inflated and bothlumens connected to th e breathing system. Auscultation should
then be repeated.
Next, the attachment between the breathing system and the
tracheal lumen should be occluded and the tracheal lumen
opened to atmosphere. Breath sounds should be heard only over
the left lung. If breath sounds are heard bilaterally, the tube is
too high in the trachea. Both cuffs should be deflated and the
tube advanced. If breath sounds are heard only over t he right
lung, the bronchial lumen is on the right side. If this is the case,
both cuffs should deflated, the tube withdrawn until its distal end
is above the carina, rotated, then reinserted. The steps outlined
should be repeated.
The attachment between the breathing system and the bronchial
lumen should then be clamped and the patient ventilated through
the tracheal lumen. The bronchial lumen should be opened to
atmosphere. Breath sounds should be heard only over the right
lung. If there is marked resistance to ventilation, the tube is
either too far into the left bronchus or is not deep enough. The
-
8/11/2019 Chapter 20. Lung Isolation Devices
29/144
position can be determined by deflating the bronchial cuff while
continuing to ventilate through the tracheal lumen with the
bronchial lumen clamped. If the tip is too deep in the left
bronchus, breath sounds will be heard only on the left side. If the
tube is not deep enough in the bronchus, breath sounds wil l be
present bilaterally. The tracheal cuff also should be deflated and,
depending on where breath sounds were heard, the tube pulled
back or advanced. Both cuffs should be reinflated and the
auscultatory sequence repeated.
R i gh t - s i d ed Tubes
Auscultation of a r ight-sided DLT is similar to that of a le f t-sided
tube. It is especially important to confirm ventilation of the right
upper lobe.
P.643
Single Connector Method
An alternative technique is the single connector technique in
which a single connector is used to ventilate each lung in turn by
simply transferring the distal end of the DLT from one lumen to
the other, eliminating the need for repeated clamping and
unclamping (126). This has the advantage of being simpler and
requiring fewer steps and may result in reduced risk of creating a
potentially harmful ball-valve effect in a partially obstructed lobe
or lung by detecting it earlier.
-
8/11/2019 Chapter 20. Lung Isolation Devices
30/144
After the DLT is advanced to the desired depth, a single
connector from the breathing system is attached to the bronchial
lumen. While performing manual ventilation, the bronchial cuff is
inflated to create an airtight seal. Auscultation is used to confirm
that the intended lung is being ventilated. The connector is then
transferred from the bronchial lumen to the tracheal lumen. The
tracheal cuff is inflated to create an airtight seal, and it is
confirmed that the proper lung is ventilated. The connector is
then transferred to the bronchial lumen and short ventilations
performed while listening over the apex of the lung for vesicular
breath sounds. If these are not heard, the single connector is
detached from the bronchial lumen and a double connector is
connected to the two lumens.
F le x i b l e Endos co p i c Techn i q ue s
Flexible endoscopy is the most accurate method for determining
DLT position (120,129,130,131,132). Many recommend that this
should be the standard of care (81,131,133,134,135,136). Others
believe that the fiberoptic scope is helpful but not essential
(95,124,137,138,139). There is general agreement that fiberoptic
endoscopy is always needed for right-sided DLTs. Fiberoptic
methods to confirm the position of the DLT may not work in the
presence of blood or secretions (140). W henever there is any
doubt, this method should be used to check the position. A
further advantage is that it can be used to remove b lood or
secretions. It can also be used after the patient is prepped and
draped if a question arises with regard to proper tube placement.
-
8/11/2019 Chapter 20. Lung Isolation Devices
31/144
Left-sided Tubes
A f iberscope is placed in the tracheal lumen through the openend of the tube or through a port in the connector that is specially
designed for this purpose. As the fiberscope is advanced, the
carina should come into view. The top surface of the blue
bronchial cuff should be seen below the carina in the left
mainstem bronchus. The bronchial cuff should not herniate over
the carina, neither should the carina be pushed t o the right. An
unobstructed view of the nonintubated right mainstem bronchus
should be obtained.
The fiberscope should then be advanced through the bronchial
lumen to check for narrowing of the lumen at the level of the cuff
and an unobstructed view of the distal bronchial tree.
Right-sided Tubes
Looking down the tracheal lumen, the bronchial cuff's upper
surface should be seen below the carina in the right mainstem
bronchus. The fiberscope is then placed in the bronchial lumen.
The right middle lower lobe bronchial carina should be seen
below the end of the tube. The endoscopist should be able to
look into the right upper lobe orifice by flexing the tip of the
fiberscope superiorly.
B r o n c h o s p i r ome t r y
Pressure-volume and flow-volume loops are discussed in Chapter
23. Changes in compliance or resistance may mean that the DLT
is not correctly placed
(141,142,143,144,145,146,147,148,149,150). If a DLT is
-
8/11/2019 Chapter 20. Lung Isolation Devices
32/144
advanced into the trachea with the bronchial cuff partially
inflated, compliance is reduced when the tube impacts the
bronchus (92).
Ches t X -ray
Chest x-ray may be useful to confirm tube position when a
fiberscope is not available or cannot be used. However, it is less
precise than the fiberoptic bronchoscopy and is costly, time
consuming, and awkward to perform.
Stabilizing the Tube
Once correct tube position is confirmed, the tube should be
secured in place. Special fixation methods have been
recommended (151). During positioning, the tube should be held
at the level of the incisors and the head immobilized in a neutral
or slightly flexed position to prevent the tube from migrating into
an incorrect position.
Intraoperative Care
The bronchial cuff should be kept deflated (unless the lung needs
to be isolated to prevent spread of blood or infection) unti l the
lung needs to be collapsed to minimize damage to the bronchial
mucosa (38). Lung collapse will be most rapid if lung separationis initiated at end-expiration. Suction is of limited utility because
the gas trapped in the lung is distal to collapsible airways. If
despite best efforts complete lung separation cannot be
accomplished and gas is introduced into the ipsilateral lung with
each breath, then continuous suction may be helpful to evacuate
-
8/11/2019 Chapter 20. Lung Isolation Devices
33/144
the gas as it enters the lung (38). Bronchial cuff pressure should
be monitored and adjusted to the minimum necessary to achieve
an airtight seal.
A useful technique when the bronchial lumen is in the surgically
operated lung is to pass a suction catheter through the bronchial
lumen when the lung is deflated and to leave it until ready for
reinflation (152). This may prevent the bronchial lumen from
becoming obstructed by blood or mucus.
P.644
Replacing a Double-lumen Tube with a
Single-lumen Tube
If mechanical ventilation needs to be continued at the conclusion
of a case in which a DLT was used, it is usually desirable to
replace the DLT with a standard tracheal tube. Personnel who
are caring for the patient in the postoperative period are not
usually familiar with a DLT. After pneumonectomy, when all of the
patient's ventilation is conducted through one lumen of the DLT,
the small diameter of the lumen may make if difficult for the
patient to breathe spontaneously. W ith high minute venti lation,
this can lead to auto-PEEP (17). Pressure-support ventilation
(Chapter 12) can be used to decrease the imposed work of
breathing during spontaneous ventilation if replacing the DLT
cannot be performed. Suctioning through a DLT is difficult.
-
8/11/2019 Chapter 20. Lung Isolation Devices
34/144
In most cases, the procedure is simply to remove the DLT and
insert a single-lumen tube in its place. If the patient was difficult
to intubate or circumstances make visualizing the larynx difficult,
other techniques should be considered. One technique would be
to insert an airway exchange catheter into the tracheal lumen
before the DLT is removed. Special extra-long exchange
catheters are needed for the exchange. After the DLT has been
removed, the single-lumen tube is then advanced over the
catheter (153,154). Oxygen insufflation via the catheter will
reduce the incidence of hypoxemia.
Another technique has been described (155). At the conclusion of
the case, both cuffs are deflated and the DLT withdrawn until the
bronchial lumen is above the carina. The bronchial cuff is then
inflated and the lungs ventilated through the bronchial lumen.
The tracheal lumen is clamped, and an opening is created in thewall of the tracheal lumen. A single-lumen tube is then slipped
over a fiberscope, and the fiberscope is advanced through the
hole in the tracheal lumen and into the trachea. The opening in
the DLT is extended, and the DLT is slowly removed. The
fiberscope is then removed.
Hazards A sso c i a t ed w i t h Dou b le -l um en
Tubes
Many of the hazards associated with single-lumen tracheal tubes
(Chapter 19) can also occur with DLTs.
Difficulty with Insertion and Positioning
-
8/11/2019 Chapter 20. Lung Isolation Devices
35/144
Inserting a DLT may be time consuming. W hen there is severe
hemorrhage, this can be a major problem. Multiple insertions and
repositionings increase the risk of trauma. While they are very
useful in adults and older children, they are of ten too large for
small children. They are often difficult to position and usually
must be replaced with a single-lumen tube at the end of surgery.
The rigidity and width of the DLT can make intubation
complicated. The tube does not allow preshaping before
intubation, and the large width makes it difficult to pass the DLT
through a tracheostomy stoma, small airway, or nasal passage.
Tube Malposition
Certain physical conditions may make it difficult or impossible for
a DLT to be correctly placed (156,157). Preoperative fiberoptic
endoscopy may detect many of these problems. Even if a correct
position is achieved initially, head movement, a change in body
positioning, or surgical manipulation may result in t ube
malposition. Displacement during positioning can be decreased
by using a neck brace (123,158). Malpositioning is increased
when anesthesia providers have limited experience in lung
isolation (159).
If DLT malposition is suspected, fiberoptic techniques are clearly
advantageous in defining the problem and afford a means of
visual correction. While some have reported that monitoring
carbon dioxide waveforms helps to detect DLT displacement
(160,161), capnography does not reliably indicate DLT
misplacement (142,148,149,162).
-
8/11/2019 Chapter 20. Lung Isolation Devices
36/144
Consequence s
Functional indications of misplacement include the following:
Unsatisfactory Lung Collapse
An obstruction in the unventi lated lumen can prevent the
unventilated lung from deflating. If the lung cannot be collapsed,
operating time will be increased, and the surgical result may be
compromised. Another cause may be a tumor fragment in the
airway (163).
Obstruction to Lung Inflation
If the bronchial cuff is not below the carina, it may obstruct the
trachea and right mainstem bronchus. With right-sided tubes,
misalignment of the port for the right upper lobe can result in
obstruction. If the bronchial cuff on a left-sided DLT is too deep,
it may obstruct the upper lobe bronchus.
Gas Trapping
Gas trapping or expiratory obstruction may be the result of a one-
way valve effect that allows inflation but not deflation. If
unrecognized, it can result in cardiorespiratory embarrassment
and/or lung parenchymal damage.
Failure of Lung SeparationIf the airway to a bronchopleural fistula cannot be isolated from
that to the normal lung, barotrauma may develop with positive-
pressure ventilation, or the air leak through the fistula may be so
large that ventilation of the normal lung is compromised. An
-
8/11/2019 Chapter 20. Lung Isolation Devices
37/144
incompletely protected dependent lung may be flooded with blood
or secretions.
Poss i b l e Mal p o s i t i o n s
Bronchial Lumen in the Wrong Mainstem
Bronchus
In some cases, the bronchial portion will enter the opposite
P.645
lung. This is usually easy to detect and correct by using
fiberoptic endoscopy (115). In some cases, it may be best to
leave the bronchial lumen in the operative bronchus and isolate
the operative lung by clamping the bronchial limb and using the
tracheal lumen for ventilation (118,122). This may be appropriate
for right lung surgery but not for surgery on the left lung, sincethe right upper lobe bronchus would almost certainly be
occluded. An alternative option is to withdraw the tube until it is
intratracheal and to use a bronchial blocker (see below) to block
the operative lung.
It may be possible for the surgeon to assist in correctly placing
the tube once the chest is open (164). If it is determined that the
tube is in the wrong bronchus, both cuffs are deflated, and the
tube is withdrawn into the trachea. The surgeon then compresses
the bronchus, and the anesthesia provider advances the tube into
the correct side with surgical guidance. The cuffs are then
reinflated.
-
8/11/2019 Chapter 20. Lung Isolation Devices
38/144
Bronchial Portion Inserted too Far into the
Appropriate BronchusIf a left-sided DLT is inserted too deeply, the problem may be
that the tube is too small (156,165,166,167). It will result in
obstruction of the upper lobe. In some patients, a left-sided DLT
placed so that the bronchial cuff is just distal to the carina still
may cause left upper lobe obstruction. A high peak airway
pressure during one-lung venti lation should suggest t hismalposition (135).
Tube too Proximal in the Airway
If the tube is not sufficiently advanced into the bronchus, the
bronchial cuff may protrude into the trachea. The need to inject
more than 3 mL of air into the bronchial cuff to achieve a seal
should alert the user that the tube may be malpositioned. Thebronchial segment may slip out of its bronchus, especially during
changes in the patient's position. In many cases, no untoward
sequelae will occur. However, there may be obstruction of gas
flow to the other lung and inability to isolate the surgical lung.
Tip of Bronchial Lumen above the Carina
The tip of the bronchial lumen may be above the carina becauseof a tracheal lesion that prevents the tube from being advanced
farther. With this malposition, there will be unsatisfactory lung
deflation and failure of lung separation.
Incorrect Placement with Respect to the Upper
Lobe Bronchus
-
8/11/2019 Chapter 20. Lung Isolation Devices
39/144
-
8/11/2019 Chapter 20. Lung Isolation Devices
40/144
oxygen source at ambient pressure to the opening to the
nonventi lated lung may h elp to prevent hypoxia
(175,176,177,178). This may also enhance collapse of the
nonventilated lung.
If hypoxemia is a problem despite proper tube position, CPAP
can be applied to the nondependent lung (179). Some DLT
manufacturers include a CPAP device with each DLT, or they
may be purchased separately (Fig. 20.15). Other measures to
improve oxygenation include dependent lung PEEP, occasional
ventilation of the nondependent lung (one breath every 5 to 10
minutes), insufflation of 2 to 3 L/minute of oxygen to the
nonventilated lung, and clamping of the pulmonary artery before
clamping the bronchus. Jet venti lation of the nondependent lobes
that are not being removed by using an airway exchange catheter
may be used to improve oxygenation (180).
Obstructed Ventilation
Many cases of obstruction are the result of a malpositioned tube.
In addition, inflating the bronchial cuff can cause narrowing of the
bronchial lumen (129,156) or may cause the carina to be
displaced laterally, producing obstruction of the other mainstem
bronchus (181). A defective tube or connector may c ause
obstruction (182,183).
Other causes of bronchial obstruction have been reported. In one
case, the bronchial cuff was left deflated until one-lung
ventilation was to begin, and necrotic tumor migrated into the
bronchus of the dependent lung, causing obstruction when one-
-
8/11/2019 Chapter 20. Lung Isolation Devices
41/144
lung ventilation was begun (184). The bronchial lumen can
become twisted (185). A carinal hook may obstruct the opening of
the tracheal lumen (186).
P.646
View Figure
Figure 20.15Device for applying continuous positiveairway pressure to a nonventilated lung. The adjustablevalve supplies pressures from 1 to 10 cm H2O.
A rela tive contraindication to using a DLT is a lesion (airwaynarrowing or endoluminal tumor) somewhere along the pathway
where the tube will reside. An aberrant tracheal bronchus may be
a contraindication for using a DLT (172,173).
Trauma
-
8/11/2019 Chapter 20. Lung Isolation Devices
42/144
Trauma to the r espiratory tract can occur whenever intubation
with a DLT is performed (187,188). Tears in the trachea and
mainstem bronchus have been reported
(76,187,188,189,190,191,192,193,194,195,196,197,198,199,200,
201,202,203).
Tube size is a factor. Large tubes have been involved more often
in injury than smaller ones (188). A tube that is too small and
requires excessive cuff inflation may cause ischemic injury. In
one reported case, an endobronchial polyp that developed at the
bronchial cuff site ended in a fatal hemorrhage (204).
Measures to reduce airway trauma include removing the stylet
after the tip of the tube has passed the vocal cords, avoiding cuff
overinflation, deflating the tracheal and bronchial cuffs when
repositioning the patient or the tube, and not advancing the tube
when resistance is encountered. Some bronchial cuffs canprovide one-lung isolation with significantly lower pressures than
others (41). It has been recommended that the bronchial cuff be
kept deflated until needed to minimize pressure on the bronchial
mucosa. This may not be prudent if there is a bronchial tumor, as
necrotic tumor may migrate into the other lung (184).
Tube ProblemsReported problems with DLTs include mislabeling, tracheal lumen
distortion that prevented a suction catheter from passing, a slit in
the septum, a defect that made the bronchial lumen kink on itself,
a split in the tubing to the bronchial cuff, a kink in the inflating
tube to the tracheal cuff, a protuberance in the wall of the tube
-
8/11/2019 Chapter 20. Lung Isolation Devices
43/144
with resultant tracheal lumen obstruction, and foreign bodies in
the DLT (205,206,207,208,209,210,211,212). The carinal hook
may bend backward, obstructing the opening of t he tracheal
lumen (213).
Tracheal or bronchial cuff rupture can occur. This most commonly
results from contact between the tube and the teeth or
laryngoscope during insertion. Proposed methods to avoid this
problem include the use of a retractable protective sheath (214),
a lubricated Penrose drain (215), and lubricated teeth guards
(216) and increasing the curve of the bronchial portion of the DLT
with the stylet (217). Another possible cause of cuff rupture is
movement during repositioning (218).
Surgical Complications
The bronchial cuff may be punctured by the surgeon ( 219,220). A
suture or staple may be placed through the DLT, or the surgical
procedure may result in a tight stenosis, which could entrap the
bronchial segment (221).
Failure to Seal
One of the reasons to use a DLT is to prevent material from
passing from one lung to the other during the surgical procedure.
Failure to prevent fluids from traversing the bronchial cuff could
result from malposition or from improper cuff inflation. Neither an
airtight bronchial seal nor a cuff pressure of 25 cm H 2O
guarantees protection against aspiration (61). Lubricating the cuff
with a gel will reduce the risk that fluid will leak past the cuff
(222,223).
-
8/11/2019 Chapter 20. Lung Isolation Devices
44/144
Difficult Extubation
Difficulty in removing a DLT may be due to anatomical
abnormalities, surgical fixation, or entanglement by other surgical
or anesthetic hardware (220,224).
P.647
Single-lumen Bronchial TubesAnother option for achieving lung separation is to use a single-
lumen tube to intubate a mainstem bronchus. Two bronchial
tubes may be used in situations where dual lung ventilation is
needed but circumstances bode against having one tube in the
trachea (225,226,227).
Single-lumen bronchial tubes are sometimes used in pediatricpatients whose airways are too small for DLTs or in whom a
bronchial blocker cannot be used
(1,165,226,228,229,230,231,232,233,234,235,236,237). In the
patient with massive hemoptysis, bronchial intubation with a
single-lumen tube is often the easiest and quickest method of
separating the lungs (5,10).
Equ i pmen t
Bronchial intubation is most often carried out with a conventional
tracheal tube. A cuffed tube will prevent re-expansion of the
collapsed lung. The distance from the tip of the tube to the
cephalad edge of the cuff must be shorter than the length of the
-
8/11/2019 Chapter 20. Lung Isolation Devices
45/144
mainstem bronchus to ensure that the cuff can lie entirely inside
the bronchus and the upper lobe bronchus is not obstructed
(238).
Special tubes with a single lumen, an angulated distal (bronchial)
tip, and cuffs at both the tracheal and bronchial positions are
available (239,240). They are longer than conventional single-
lumen tracheal tubes.
Techn i ques
Before insertion, the correct length and size of the tube should
be estimated from a chest x-ray or computed tomography (CT)
scan (238). If the nasotracheal route is used, most conventional
single-lumen tubes will not be long enough to provide a reliable
mainstem intubation. It is recommended that the tube should be
one half to one size smaller than the usual size selected for
tracheal intubation (237). For bronchial intubation in children, a
tracheal tube 0.5 to 2.0 mm smaller than recommended for the
particular patient should be used (229,231).
Right-sided intubation can usually be performed blindly, but the
tube is more reliably placed by using a bronchoscope. It may be
possible to align the Murphy eye with the right upper lobe
bronchus. It may be possible to rotate the tube so that the bevel
faces the upper lobe bronchial orifice.
Left mainstem intubation may be a chieved blindly by using a
stylet to curve the distal end of the tr acheal tube to the left (231)
but often requires bronchoscopic guidance. If blood or secretions
preclude fiberoptic visualization, fluoroscopy is another option.
-
8/11/2019 Chapter 20. Lung Isolation Devices
46/144
The chance of intubating the left bronchus will be increased if the
tube is rotated 180 degrees from its usual position before
advancing it beyond the vocal cords and the patient's head is
turned to the right (228,229,241,242,243).
A gum elastic bougie can be inserted into the chosen bronchus
by using a bronchoscope. The bronchial tube can then be
railroaded over the bougie into position (244).
Correct positioning can be confirmed by auscultation, x-rays,
and/or flexible bronchoscopy.
The tube should be withdrawn into the trachea when one-lung
ventilation is no longer needed.
Eva lua t i on
Advantages of using a single- lumen tube for lung separation
include simplicity and the rapidity with which lung separation can
often be achieved, particularly when the right lung must be
ventilated.
Disadvantages include frequent lack of ventilation of the right
upper lobe with right mainstem intubation (38,236). Left upper
lobe ventilation may also be excluded when the left mainstem
bronchus is relatively short (238). Neither suctioning nor
application of CPAP to the nonventilated lung is possible.
There may be failure to achieve an adequate seal, especially if
an uncuffed tracheal tube is used or if the cuff is not inside the
bronchus (226,236,238,245). Lung collapse will be incomplete,
and the healthy ventilated lung will not be protected from
contamination. Both lungs cannot be ventilated at the beginning
-
8/11/2019 Chapter 20. Lung Isolation Devices
47/144
of anesthesia, and the collapsed lung cannot be re-expanded and
ventilated until the tube is withdrawn into the trachea ( 236).
Bronchial-blocking Devices
Ind i ca t i ons and Use
Indications for bronchial blockers are similar to those for a DLT,
with the exception of independent lung ventilation (44,226,246).
They are often used in patients in whom using a DLT is not
possible or advisable (nasal intubation, small patient, difficult
intubation, patient with a tracheostomy, subglottic stenosis, thick
and excessive secretions, need for continued postoperative
intubation)
(240,247,248,249,250,251,252,253,254,255,256,257,258,259,260
,261). A blocker may be especially useful for providing lung
separation in a critically il l patient with a single-lumen trachealtube already in place. Another indication may be the patient on
anticoagulants, since placing a blocker is usually less traumatic
than inserting a DLT (8). A blocker may allow a larger fiberoptic
endoscope to be used and provide better suctioning than a DLT
(262). A modified bronchial blocker can be used for tracheal gas
insufflation to reduce carbon dioxide within the dead space by
delivering fresh gas near the end of the tracheal tube ( 263).
P.648
Another indication for a blocker is the need to block a segment of
a lung rather than the entire lung
-
8/11/2019 Chapter 20. Lung Isolation Devices
48/144
(9,14,258,259,264,265,266,267,268,269,270). This cannot be
done with a DLT. A blocker may be used to sequentially block
different parts of the lung (270).
A blocker may be used to achieve lung isolation in the patient
with an improperly positioned double-lumen or bronchial tube
(53,122,238,271,272,273). A blocker may be useful if both lungs
require sequential blockage (274). The blocker can be shifted to
the opposite lung when needed. If one blocker does not provide
complete one-lung isolation, a second blocker may be used
(173,275,276,277).
Finally, there is no need to change the tube at the end of the
operation if postoperative mechanical ventilation is needed if a
bronchial blocker is used.
Dev ices
Univent Bronchial-blocking Tube
Desc r i p t i o n
The Univent tube is a cuffed silicone tracheal tube with a small
additional internal lumen along its concave side
(7,11,226,236,278,279,280,281,282,283,284) (Figs. 20.16,
20.17). The s mall channel contains a movable tubular blockerthat has a blue high-pressure, low-volume cuff. The blocker can
be advanced sufficiently beyond the tip of the tube to block
airways smaller than a mainstem bronchus (246,260,264,265).
-
8/11/2019 Chapter 20. Lung Isolation Devices
49/144
View Figure
Figure 20.16Univent bronchial blocker. The cuffed
tracheal tube has a small lumen along its concave side,which contains a tubular cuffed bronchial blocker. The
blocker can be advanced into a mainstem bronchus orsmaller airway.
The blocker has external depth markings to help determine theblocker position in relation to the tube. There is a s light bend in
the blocker above the cuff. The blocker tip is radiopaque. A grip
allows the user to rotate the blocker. A locking clamp fixes its
depth below the tip of the tube.
As shown in Table 20.2, the Univent tube is available in several
sizes. It has a slightly larger-than-usual external diameter for its
internal diameter because of the space required by the blocker.
Univent tubes that are 5 mm and larger have a lumen that can be
used for suctioning, CPAP, or oxygen insufflation (285). Adult
versions of the Univent blocker are hollow.
The Univent blocker (Uniblocker) can be purchased separate
from the tracheal tube and used with other tracheal tubes
-
8/11/2019 Chapter 20. Lung Isolation Devices
50/144
coaxially or in parallel (265,286) (Fig. 20.18). It is supplied with a
swivel adaptor that fits onto the tracheal tube connector. This
adaptor allows connection to the breathing system and has a port
for a fiberoptic scope in addition to a port for the blocker.
Use
Before use, the bronchial blocker and tube cuffs should be
inflated and checked for leaks. Both the tube and the blocker
should be well lubricated. After the cuffs have been deflated, the
blocker should be pushed back and forth in the tube to ensure
free movement. The blocker should then be fully retracted into
the tube lumen and fixed in place by using the clamp.
If there is an unobstructed view of the larynx, the Univent tube is
inserted in the same way as a conventional tracheal tube. If the
patient has a difficult airway, it can be inserted over an airway
exchange catheter or other device (287). The blocker can be
extended and used as an introducer (288,289). After the blocker
has passed the vocal cords, the tube is t hreaded over it and into
place.
After the Univent tube is inserted, the tracheal tube cuff is
inflated, and the patient is ventilated. The blocker is visualized by
using a flexible fiberoptic endoscope through an airway adaptor
with a port for the scope and is maneuvered into the appropriate
bronchus. The blocker tip direction can be changed by twisting
the shaft (290). It may be useful to deflate the tracheal tube cuff
and rotate the tube so that the blocker is directed toward the side
to be occluded.
-
8/11/2019 Chapter 20. Lung Isolation Devices
51/144
A guide wire can be inserted through the blocker 's lumen and be
used to direct the blocker into place, especially when it is
necessary to block an airway smaller than a mainstem bronchus
(258).
Another method of placing the blocker in the r ight or le ft main
bronchus is to insert a fiberscope through the trachea tube into
the bronchus to be blocked and then to advance the tube into
that bronchus. The blocker is then advanced into the bronchus
and the tube withdrawn into the trachea, leaving the blocker in
the bronchus. This technique may result in trauma to the airway
(291).
P.649
-
8/11/2019 Chapter 20. Lung Isolation Devices
52/144
View Figure
Figure 20.17Univent bronchial-blocking tubes.
Top:The bronchial blocker is retracted. Bottom:The bronchial blocker is advanced, and the cuff isinflated. (Courtesy of Vitaid.)
TABLE 20.2 Univent Tubes
-
8/11/2019 Chapter 20. Lung Isolation Devices
53/144
I nternal Diameter
(millimeters)
Outer Diameter
(millimeters)a
Age (years)
3.5 (uncuffed) 7.5/8.0 6 to 10
4.5 8.5/9.0 10 to 14
6.0 9.7/11.0 14 to 16
6.5 10.2/12.0 16 to 18
7.0 11.6/12.5 Adult
7.5 11.2/13.0 Adult
8.0 11.7/13.5 Adult
8.0 12.2/14.0 Adult
9.0 12.7/14.5 Adult
a
Values are sagittal/transverse.From Hammer GB, Fitzmaurice BG, Brodsky JB.
Methods for single-lung ventilation in pediatricpatients. Anesth Analg 1999;89:14261429; Frolich
-
8/11/2019 Chapter 20. Lung Isolation Devices
54/144
MA. Postoperative atelectasis after one-lung ventilationwith a Univent tube in a child. J Clin Anesth2003;15:159163; Hammer GB, Brodsky JB, RedpathJH, et al. The Univent tube for single-lung ventilationin paediatric patients. Paed Anaesth 1998;8:5557;Tobias JD. Variations of one-lung ventilation. J ClinAnesth 2001;13:3539.
The bronchial blocker can be inserted blindly. The whole tube isturned so that its concavity faces the side to be blocked. The
blocker is advanced into the mainstem bronchus and the cuff
inflated. This method has not proved very successful and may be
associated with airway trauma (282,292).
The blocker position should be checked by using a fiberscope.
The cephalad tip of the bronchial cuff should be below the carina.
The blocker should then be fixed to the tracheal tube by using
the cap stopper and blocker grip.
When the bronchus needs to be blocked, the lung is deflated with
the blocker open to atmosphere. The bronchial blocker cuff
should be inflated by using the least amount of air that will
provide a seal. This can be achieved by attaching the sample l ine
-
8/11/2019 Chapter 20. Lung Isolation Devices
55/144
from a carbon dioxide analyzer to the proximal end of the blocker
and noting when the waveform disappears (293). Another method
is the bubble test, in which the end of the bronchial lumen is
placed into water in a beaker (294). When the bronchus is
sealed, no bubbles will be observed passing through the water.
The typical cuff inflation volume is 5 to 6 cc (112).
The blocker from a Univent tube can be removed from the tube
and inserted alongside a tracheal or tracheostomy tube
(265,295,296). The blocker is then
P.650
guided into place with a fiberscope inserted through the tube.
View Figure
Figure 20.18The Univent bronchial blocker can be
purchased separately from the tube. The connector has
ports for attachment to the tracheal tube and breathingsystem, for introducing the blocker, and for a fiberscope.
Note the cap that fits over the end of the blocker.
-
8/11/2019 Chapter 20. Lung Isolation Devices
56/144
When the blocker is no longer needed, the cuff is deflated and
withdrawn into the main tube. If the Univent is to be used for
postoperative ventilation, the blocker should be fully retracted
and disabled to avoid inappropriate use by caregivers who are
unfamiliar with the device (297).
Eva l ua t i o n
Most studies show that the Univent provides lung isolation
equivalent to that of a DLT (11,280,298,299,300,301).
The Univent may be easier to insert and position correctly than a
DLT and may be especially useful for the difficult-to-intubate
patient (287,288,289,302,303,304,305). It can be used in the
patient with a tracheostomy (254,253,296) and for nasal
intubation (247). It can be used for postoperative ventilation
without having to reintubate the patient.
It is possible to use suction, apply CPAP, or insufflate oxygen
through the blocker lumen (278,306). The blocker with the cuff
deflated has been used for jet ventilation during carinal resection
(307,308).
Reports of problems with the Univent include the cap becoming
dislodged from the tip of the blocker ( 309,310) and fragmentation
of the tube inner wall and connector (311,312).
The Univent's curved shape is fixed, and this may be a
disadvantage when sliding it over a bronchoscope. It will not
soften in a warm water bath (38). It may distort the neck anatomy
sufficiently to make internal jugular vein cannulation difficult
(313). Bronchial perforation by the blocker during blind insertion
-
8/11/2019 Chapter 20. Lung Isolation Devices
57/144
has been reported (292). A case has been reported of prolonged
postoperative atelectasis after using a Univent tube (262). It is
not recommended for use in children below the age of six years.
When comparing the Univent with a left DLT, there was a greater
incidence of malposition with the Univent (119), but comparison
of the Univent and a right DLT failed to demonstrate a clear
advantage (35). The Univent is more expensive than a DLT or
other blockers (79,119,310).
A disadvantage of the Univent tube is the large amount of cross-
sectional area occupied by the blocker channel, especially in the
smaller tubes (236). It is not available in a size that would fit an
infant or small child.
Another problem is that the small lumen is re la tively easily
blocked by blood or pus (285). As a result, blood or pus may
accumulate and contaminate the dependent lung when theblocker is deflated.
The blocker's low-volume, high pressure cuff may cause in
mucosal injury (314,315).
The larger external diameter may make it difficult to pass the
Univent between the vocal cords.
Arndt Bronchial BlockerDesc r i p t i o n
The Arndt bronchial blocker assembly (wire-guided bronchial
blocker, WEB, FWEB) is designed to be used for a patient with a
single-lumen tracheal tube already in place
(44,111,246,253,316,317,318,319,320,321,322,323). It consists
-
8/11/2019 Chapter 20. Lung Isolation Devices
58/144
of two parts: A blocking catheter and a special airway adaptor.
Either may be purchased and used separately.
Blocking Catheter
The Arndt blocking catheter (Fig. 20.19) has a low-pressure,
high-volume balloon that has either an elliptical or spherical
shape. The spherical balloon is relatively compliant unless
overinflated and takes an elliptical form when inflated in a small
bronchus. The 9 Fr catheter is available with an elliptical balloon
that provides a longer sealing profile. It is recommended that the
elliptical cuff be limited to left mainstem intubation (236,246).
P.651
-
8/11/2019 Chapter 20. Lung Isolation Devices
59/144
View Figure
Figure 20.19Arndt bronchial blocker with multiport
adaptor. The wire loop can be cinched around the tip ofthe fiberscope, or the fiberscope is passed through theloop. The adaptor has ports for attachment to the trachealtube and breathing system, for introducing the blocker,and for a fiberscope.
A f lexible nylon wire passes through the proximal end of thecatheter and extends to the distal end, then exits as a small loop
(Fig. 20.19). The size of the loop may be increased or decreased
by advancing or retracting the wire assembly.
There are three sizes: 9, 7, and 5 Fr. Characteristics of the
blockers are shown in Table 20.3. Near the distal end of the 9 Fr
catheter are side holes to facilitate lung deflation.
Airway Adaptor
The multiport adaptor (Fig. 20.19) allows simultaneous
introduction of a bronchoscope and the blocker while maintaining
mechanical ventilation. It has four ports:
-
8/11/2019 Chapter 20. Lung Isolation Devices
60/144
A 15-mm female connector that attaches to the tracheal
tube;
A side port wi th a 15-mm male connector that attaches to
the anesthesia breathing system;
A port angled approximately 30 degrees for the bronchial
blocker; and
A port for the f lexible endoscope.
The blocker port has a Tuohy-Borst type connector to maintain anairtight seal and lock the blocker in place (by tightening the
connector around the blocker) or to allow free movement of the
blocker (by loosening the connector). The bronchoscopy port has
a plastic sealing cap.
Use
Before use, the inside of the tracheal tube and the o utside of theblocker and the bronchoscope should be well lubricated with a
silicone spray. The loop should be adjusted so that it loosely
approximates the outside diameter of the bronchoscope. The
blocker balloon should be inflated to test for leaks and then fully
deflated.
For each size Web, there is a tracheal tube size for which coaxial
placement is limited by the diameters of the blocker and
fiberscope used during placement. When these limits are
exceeded, parallel placement may be required. Either t he web is
passed outside the tracheal tube with guide loop inoperable, or a
fiberscope is also placed in parallel with the tracheal tube before
or after intubation.
-
8/11/2019 Chapter 20. Lung Isolation Devices
61/144
TABLE 20.3 Arndt Endobronchial Blockers
Size(French)
Smallest Single-lumenTube In ternal Diameter
for Coaxial Use
(millimeters)
Length(centimeters)
Cuf f Shape Average Cuf fI nfl ation Volume
(cubic cent imeters)
9 7.5 78 and
65
Elliptical 6 to 12
Spherical 4 to 8
7 6.0 65 Spherical 2 to 6
5 4.5 65 and
50
Spherical 0.5 to 2.0
From Klafta JM. One-lung anesthesia; making it work (ASA Refresher Course#509). Park Ridge, IL: ASA, 2004.
P.652
The wire in the blocker lumen is used in either of two ways: (i)
when cinched tightly around the tip of the fiberscope, the
fiberscope carries the blocker into its desired location, and (i i)
when the fiberscope is passed through the loop, it provides a
track for blocker to pass through after the fiberscope is placed
into the bronchus.
-
8/11/2019 Chapter 20. Lung Isolation Devices
62/144
After the patient is intubated and venti lation has begun, the
multiport adaptor is connected between the breathing system and
the tracheal tube. The bronchoscope is advanced through the
guide loop. This allows the blocker to follow the bronchoscope.
Alternately , the wire loop may be placed over the end of the
blocker prior to attaching the airway adaptor to the tracheal tube
(324).
The bronchoscope is advanced into the airway to be blocked and
then the guide loop is slid over the end of the bronchoscope. The
bronchoscope is withdrawn slightly to visualize the blocker. The
blocker is then advanced or withdrawn into position (324,325). It
may be advisable to advance the blocker approximately 1 cm
beyond the optimal position when the patient is in the supine
position to avoid dislodging the blocker toward the trachea while
the patient's position is changed to the lateral decubitus position(246). The balloon is then inflated under direct vision. The
balloon should fil l the entire bronchial lumen and not herniate
into the trachea. Following placement, the balloon may be
deflated until one-lung ventilation is required.
Removing the wire loop will result in an open channel, which
allows CPAP application, oxygen insufflation, suctioning, or
intermittent inflation. Leaving the wire loop during the operation
might damage the airway and entails the risk that the loop may
accidentally be stapled into the bronchial closure (326). However,
leaving the loop in place allows the blocker to be repositioned
(327). It may be possible to reposition the blocker by using
fiberoptic endoscopy (326). Excessive force should not be used
-
8/11/2019 Chapter 20. Lung Isolation Devices
63/144
when removing the wire guide, because this may displace the
blocker.
Lung collapse can be expedited by attaching a syringe or
applying suction to the blocker channel (44,328). However, some
clinicians do not recommend using suction through this channel
because of the risk of developing negative pressure pulmonary
edema (326). When no longer needed, the blocker cuff should be
fully deflated and the blocker removed.
Eva l ua t i o n
The Arndt blocker system can be used in the patient who is
already intubated (316,317,329), in the patient with a
tracheostomy (318,324,325), or with a nasal intubation (253,318).
It has been used to provide single-lung ventilation in children as
young as 17 months (245). It may be especially useful for the
patient in whom a DLT would be difficult to use
(260,316,318,319,320,321,330). It has been used in a newborn
(322). It allows a larger internal cross-sectional area than a DLT
or Univent tube of similar outside diameter ( 331). It may require
fewer insertion attempts than a DLT (332). If the wire is removed,
the lumen can be used for suctioning or administering oxygen or
CPAP.
P r o b l ems
A disadvantage is that once the wi re loop is removed, it cannot
be reinserted through the channel to allow repositioning of the
blocker. Placement requires the availability of fiberoptic
equipment and someone able to use it. It takes longer to position
-
8/11/2019 Chapter 20. Lung Isolation Devices
64/144
and to achieve complete lung collapse compared with the Univent
tube or a DLT (328,332). The balloon may be sheared when it is
removed from the blocker port (333).
Cohen Tip-deflecting Bronchial Blocker
The Cohen tip-deflecting bronchial blocker (Fig. 20.20) has a 9 Fr
external diameter and a central lumen with a 1.6 mm diameter
(334). The high-volume, low-pressure blue balloon at the tip is
spherical in shape. The average inflation volume is 5 to 8 mL.
There are side holes between the tip and the balloon to evacuate
gas from the distal lung or to insufflate oxygen. A proximal
control wheel that can be operated with the thumb and forefinger
is used to adjust tip deflection (Fig. 20.21). The catheter has
depth markings and an indicator arrow that shows the direction in
which the tip deflects.
-
8/11/2019 Chapter 20. Lung Isolation Devices
65/144
View Figure
Figure 20.20Cohen tip-deflecting bronchial blocker. The
proximal control wheel is used to adjust tip deflection. Anarrow on the wheel indicates the direction to which the tipdeflects. (Courtesy of Cook Critical Care.)
P.653
-
8/11/2019 Chapter 20. Lung Isolation Devices
66/144
View Figure
Figure 20.21The tip of the Cohen bronchial blocker can
be manipulated to fit into either bronchus.
In most cases, the blocker and a fiberscope are inserted throughan appropriately sized tracheal tube. The blocker can also be
placed outside the tracheal tube and guided into place with a
bronchoscope placed through the tracheal tube.
A multiport airway adaptor li ke the Arndt is attached to the
breathing system end of the tracheal tube. Alternatively, a
standard swivel adaptor can be used for insertion.
An assistant with or without video f lexible fiberoptic scope may
be necessary because it may require two hands to manipulate the
blocker into position, one to deflect the tip and the other to rotate
and advance the catheter.
Embolectomy Catheter
-
8/11/2019 Chapter 20. Lung Isolation Devices
67/144
Desc r i p t i o n
An embolectomy (Fogarty) catheter can be used as a bronchial or
segmental blocker
(1,14,53,62,226,248,251,256,257,267,268,269,275,324,335,336,3
37,338,339,340,341,342,343,344). This catheter is readily
available in most operating suites where vascular surgery is
performed or can be purchased for this purpose. It comes with a
stylet in place so that it is possible to place a curvature in the
distal tip to facilitate guidance to the target bronchus. The
occlusion balloon has a high-pressure, low-volume cuff (246). It
comes in a variety of sizes. Adult bronchi can be blocked with 7
Fr catheters, whereas 2 Fr to 5 Fr catheters are suitable for
segmental or pediatric blockade.
Use
Prior to use, the blocker should be lubricated with jelly or silicone
spray to facilitate passage. The balloon should be tested for
leaks and then fully deflated. A catheter with a faulty or eccentric
balloon should be discarded.
The embolectomy catheter may be placed before or after
intubation with a single-lumen tracheal or tracheostomy tube and
can be passed either through or alongside the tube. Placing the
catheter alongside the tube allows the tube to splint the blocker
in position. An alternative method is to insert the catheter
through a hole made in the side of the tracheal tube (345).
If the patient is already intubated, the catheter containing a stylet
may be passed through a fiberoptic endoscope adaptor
-
8/11/2019 Chapter 20. Lung Isolation Devices
68/144
(335,341,342). This allows uninterrupted ventilation. The Arndt
multiport airway adaptor (previously mentioned) or similar device
(346) may also be used to prevent an air leak. An annular space
around the fiberoptic endoscope and embolectomy catheter
equivalent to a single-lumen tube with an inner diameter of 4 to 5
mm will be necessary to allow exhalation in a reasonable period
of time (347).
A f iberoptic endoscope is passed down the tracheal tube, and the
embolectomy catheter is guided into the appropriate bronchus
under direct vision. Twirl ing the catheter between the fingers at
the proximal end or rotating the tracheal tube will impart lateral
direction to the tip (98). Additional lateral direction may be
gained by rotating the tracheal tube to one side or the other.
After the tip is advanced into the proper position, the stylet is
removed, the catheter balloon is inflated under direct vision, andthen the fiberscope withdrawn.
A modification of this technique is to deliberately intubate a
mainstem bronchus with a tracheal tube one size smaller than
would be appropriate for the trachea, advance the blocker
through the tracheal tube, then withdraw the tracheal tube into
the trachea (348,349,350,351). It may be helpful to preshape the
tracheal tube by using a stylet (351). Once the catheter is in the
targeted bronchus, the stylet is removed. Removing the tracheal
tube followed by insertion of a larger tracheal tube beside the
Fogarty catheter will securely fix the catheter in place. The
catheter balloon is inflated slowly until no air enters the blocked
lung, as detected by using auscultation. The Fogarty catheter
-
8/11/2019 Chapter 20. Lung Isolation Devices
69/144
should be securely taped to the tracheal tube to prevent it from
being dislodged. A fiberoptic endoscope should be used to check
the position of the blocker. The catheter ballooon should be
deflated and the lung collapsed by using pressure on the chest
and/or suction through the tracheal tube (351). The catheter
balloon should then be reinflated to the same volume as
previously used.
Eva l ua t i o n
The use of an embolectomy catheter has many advantages. It
can be passed through a single-lumen tracheal tube in a n already
intubated patient, and there is no need for reintubation if
postoperative mechanical ventilation is needed. It may be useful
in
P.654
pediatric patients (62,236,267,268,324,337,338,340,345,352), the
patient with a tracheostomy (252,336,351,353), or for a nasal
intubation. The embolectomy catheter is less expensive than a
DLT, Univent tube, or Arndt blocker (326,339). Fogarty catheters
are relatively thinner for a given balloon volume when compared
with other bronchial blockers and thus will allow the use of larger
tracheal tubes, especially in pediatric patients when they are
placed side by side in the trachea (351).
P r o b l ems
A significant disadvantage is the lack of a hollow center.
Suctioning, oxygen insufflation, or applying CPAP to the blocked
-
8/11/2019 Chapter 20. Lung Isolation Devices
70/144
lung is not possible, and lung collapse takes longer and may not
be as complete as with a DLT or a blocker with a hollow lumen
(62). The obstructed lung segment cannot be re-expanded until
the blocker is removed. Another disadvantage is that it is made
of latex, so it cannot be used in the patient with potential latex
allergy (Chapter 15). Most of these devices have low-volume,
high-pressure cuffs and can damage the airway, although the
pressure may be less than that exerted by the cuff on a DLT
(109,236).
There are no reported complications with the embolectomy
catheter in adults (246). Bronchial rupture has been reported with
an overinflated balloon in a child (337).
Other
Balloon atrioseptostomy, Foley urinary, and Swan Ganz
pulmonary artery catheters have been used as bronchial
blockers. They have a central lumen for suction or administration
of oxygen to the blocked lung.
P rob l ems w i t h B r on ch i a l B l o c k er s
A blocker may not be suitable for the patient with a high r ight
upper lobe bronchus takeoff or a tracheal bronchus (173,354). It
may be necessary to either use a second blocker or a DLT under
these conditions. If the distance between the carina a nd the
tracheal bronchus is small, it may be possible to herniate the
balloon on the bronchial blocker so that it blocks the upper lobe
bronchus (355). It may be possible to use the tracheal cuff of the
Univent tube to block a tracheal bronchus (356).
-
8/11/2019 Chapter 20. Lung Isolation Devices
71/144
A blocker may be dislodged into the trachea (357). The inf lated
balloon may then block ventilation to both lungs, prevent collapse
of the operated lung, and/or ca use air trapping. Other
complications reported include fixation by surgical staples (358);
perforation of the blocker balloon by a surgical needle ( 359); and
accidental inflation of the blocker cuff when it was just below the
tip of the tube, resulting in obstruction to gas flow (360).
The narrow blocker lumen may result in ineffective removal of
secretions and pulmonary soiling after the cuff is deflated (361).
Lung deflation may not be as satisfactory or achieved as rapidly
with a bronchial blocker as with a DLT (332).
Development of severe hypoxemia has been reported after
continuous suctioning of the nondependent lung through a
bronchial blocker (362). If suction is used to facilitate lung
collapse, it should be applied for only a few seconds,intermittently and with low pressure (246).
References
1. Camci E, Tugrul M, Tugrul ST, et al. Techniques and
complications of one-lung ventilation in children with suppurative
lung disease: experience in 15 cases. J Cardiothorac Vasc
Anesth 2001;15:341345.
[Medline Link]
2. Morell RC, Prielipp RC, Foreman AS, et al. Intentional
occlusion of the right upper lobe bronchial orifice to tamponade
life-threatening hemoptysis. Anesthesiology 1995;82:1529
1531.
-
8/11/2019 Chapter 20. Lung Isolation Devices
72/144
[Fulltext Link]
[CrossRef]
[Medline Link]
3. Shivaram U, Finch P, Nowak P. Plastic endobronchial tubes in
the management of life-threatening hemoptysis. Chest
1987;92:11081110.
[CrossRef]
[Medline Link]
4. Cohen E. Nitric oxide, thoracoscopy and other new trends in
thoracic anesthesia (ASA Refresher Course #175). Dallas, TX:
ASA, 1999.
5. Mangar D. Treatment of pulmonary artery haemorrhage with a
single lumen tracheal tube. Can J Anaesth 1994;41:880.
[Medline Link]
6. Arya VK, Dutta A, Chari P, et al. Difficult retrogradeendotracheal intubation: the utility of a pharyngeal loop. Anesth
Analg 2002;94:470473.
[Fulltext Link]
[CrossRef]
[Medline Link]
7. Inoue H, Shohtsu A, Ogawa J, et al. Endotracheal tube with
moveable blocker to prevent aspiration of intratracheal bleeding.
Ann Thorac Surg 1984;37:497499.
[Medline Link]
8. Herenstein R, Russo JR, Mooka N, et al. Management of one-
lung anesthesia in an anticoagulated patient. Anesth Analg
1988;67:11201122.
-
8/11/2019 Chapter 20. Lung Isolation Devices
73/144
[Medline Link]
9. Kabon B, Waltl B, Leitgeb J, et al. First experience with
fiberoptically directed wire-guided endobronchial blockade in a
severe pulmonary bleeding in an emergency setting. Chest
2001;120:13991402.
[CrossRef]
[Medline Link]
10. Klafta JM, Olson JP. Emergent lung separation for
management of pulmonary artery rupture. Anesthesiology
1997;87:12481250.
[Fulltext Link]
[CrossRef]
[Medline Link]
11. Inoue H, Suzuki I, Mwasaki M, et al. Selective exclusion of
the injured lung. J Trauma 1993;34:496498.[CrossRef]
[Medline Link]
12. McGuire GP. Lung ventilation and bronchopleural fistula. Can
J Anaesth 1996;43:12751276.
[Medline Link]
13. Darwish RS, Gilbert TB, Fahy BG. Management of a
bronchopleural fistula using differential lung airway pressure
release ventilation. J Cardiothorac Vasc Anesth 2003;17:744
746.
[CrossRef]
[Medline Link]
-
8/11/2019 Chapter 20. Lung Isolation Devices
74/144
14. Otruba Z, Oxorn D. Lobar bronchial blockade in
bronchopleural fistula. Can J Anaesth