Chapter 16 Face Masks and Airways P.444 Face Masks The face mask allows gas administration to the patient from the breathing system without introducing any apparatus into the patient's mouth. The ability to hold the mask and to administer positive pressure ventilation through the mask is a basic skill that all anesthesia providers must master. In the past, the face mask was often been used to administer an entire anesthetic. More recently, the introduction of supraglottic airway devices ( Chapter 17 ) has led to a decrease in this practice. While some supraglottic airway devices have the word mask in their titles, they are very different from face masks and are dealt with separately in Chapter 17 . The face mask is also used to administer noninvasive positive pressure ventilation (NPPV) for treatment of respiratory failure ( 1 , 2 ). Description A face mask may be constructed of a number of substances, including black rubber, clear plastic, an elastomeric material, or a combination of these. The majority of anesthetics today employ a disposable plastic mask. These masks are designed to fit a wide variety of patients and are not as easy to use as many older masks that were sized and anatomically designed to fit a narrower range of patients. Body The body (shell, dome) constitutes the main part of the mask ( 3 ). A transparent body allows observation for vomitus, secretions, blood, lip color, and exhaled moisture. A transparent mask may be better accepted by a conscious patient ( 4 ). Seal The seal (rim, flap, edge) is the part of the mask that comes in contact with the face. Two general types are available. One is a pad (cushion) that is inflated with air or filled with a material that will conform to the face when pressure is applied. The second type is a flap that is a flexible extension of the body that conforms to the contour of the face. It is pressed onto the face to create a seal. Connector

Transcript of Chapter 16. Face Masks & Airways

Page 1: Chapter 16. Face Masks & Airways

Chapter 16 Face Masks and Airways P.444

Face Masks The face mask allows gas administration to the pat ien t from the breathing system

wi thout introducing any apparatus in to the pat ient's mouth. The abi l i ty to hold the

mask and to administer posi tive pressure venti lat ion through the mask is a basic

sk il l that a ll anesthesia prov iders mus t master. In the past, the face mask was often

been used to administe r an ent ire anesthetic . More recently, the introduction of

supraglo tt ic airway dev ices (Chapter 17) has led to a dec rease in this practice .

Whi le some supraglottic ai rway devices have the word mask in their ti tles , they are

very different f rom face masks and are dealt wi th separately in Chapter 17.

The face mask is also used to administer noninvasive posi tive pressure vent ilation

(NPPV) for treatment of respira tory failure (1,2).

Description A face mask may be constructed of a number of substances, inc luding black rubber,

c lear plast ic , an elastomeric material , o r a combinat ion of these. The majori ty of

anesthet ics today employ a disposable plast ic mask. These masks are designed to

f it a wide variety of pat ients and are not as easy to use as many older masks that

were s ized and anatomical ly designed to f i t a narrower range of pa tients.

Body The body (shell, dome) const i tutes the main part of the mask (3). A transparent

body allows observation fo r vomitus, secre tions , blood, l ip color, and exhaled

moisture . A transparent mask may be better accepted by a conscious pat ient (4).

Seal The seal (rim, f lap , edge) is the part of the mask tha t comes in contac t wi th the

face. Two general types are available. One is a pad (cushion) that is inf lated wi th

ai r or f i l led wi th a material that wil l conform to the face when pressure is applied.

The second type is a f lap that is a f lex ible extension of the body that conforms to

the contour of the face. I t is p ressed onto the face to create a seal .

Connector

Page 2: Chapter 16. Face Masks & Airways

The connector (orif ice, col la r, mount) is at the opposi te s ide f rom the seal. I t

consists of a th ickened f i tt ing wi th a 22-mm in ternal diameter. A ring wi th hooks

(Fig . 16.1) may be placed around the connector to allow a mask strap to be

attached.

Specific Masks Masks come in a variety of sizes and shapes (F igs . 16.1, 16.2). An assortment

should be kept readily available, because no one wi l l f i t every face we ll .

View Figure

Figure 16.1 Clear, disposable masks. (A, Picture courtesy of Kendall; B, Courtesy of Rusch, Inc.)

P.445

Page 3: Chapter 16. Face Masks & Airways

View Figure

Figure 16.2 Black rubber masks. (Courtesy of Sun Med.)

Rendell-Baker-Soucek Mask The Rendell -Baker-Soucek (RBS) mask (Fig. 16.3), which is designed for the

pedia tric pa tient, has a triangular body. I t has a low dead space (5,6). Some of

these masks are scented and may have a pacif ier (F ig. 16.3B). This mask has been

used for the patient with a tracheostomy (7,8,9) and a patient with acromegaly (10).

I t may also be useful when a mask to cover only the nose is needed (11).

Endoscopic Masks An endoscopic mask is designed to al low mask venti lat ion whi le an endoscope is

being used (12,13,14,15). One is shown in F igure 16.4. A port/diaphragm in the

mask body al lows a f iberscope to be inserted in to the nose or mouth. A l ighted

stylet (Chapter 19) may also be used (16). A tracheal tube previously loaded over

the f iberscope or l ighted style t can then be advanced, if desired.

Scented Masks Since face masks are often used for an inha la tion anesthesia induct ion or for

preoxygenation prior to induction , efforts have been made to make this experience

more acceptable by using scents to camouflage the odors of inha la tional agents

(17,18,19,20,21,22,23,24). The scent may be incorporated into the mask by the

manufac turer o r app lied to the mask by the anesthesia provider (Fig. 16 .5). Some

masks are color coded according to the scent. The e thyl alcohol in some appl ied

f rui t f lavors may affect the accuracy of some gas monitors (25 ,26). Pre-scented

masks from the manufacturer do not present this problem (25).

Page 4: Chapter 16. Face Masks & Airways

Techniques of Use The face mask shou ld form a tight seal on the patient 's face while f i t ting

comfortably in the user's hand. A proper seal is essential ly for preoxygenation

(27,28). A poor f it requires the anes thesia provider to main tain steady pressure.

This may lead to c ramped hands and ti red muscles and l imits the user's ab il i ty to

perform other tasks. With spontaneous respirat ion, a loose seal wi l l resul t in

P.446

ai r dilu tion. With assisted or controlled venti lation, adequate gas exchange may be

impossib le wi th a poor mask f it . An inadequate seal can be compensated for some

ex tent by increasing the fresh gas f low, bu t th is is wasteful and contaminates the

room with anes thetic gases and vapors.

View Figure

Figure 16.3 Rendell-Baker-Soucek masks. A: Clear plastic version. (Courtesy of Rusch, Inc.) B: With pacifier. (Courtesy of Ohio Medical Products, a division of Airco, Inc.) C: They are also available in black rubber.

Page 5: Chapter 16. Face Masks & Airways

View Figure

Figure 16.4 Endoscopic masks. (Pictures courtesy of VBM Medical, Inc.)

Correct mask use starts wi th selec tion of the appropriate s ize and shape. Th is may

require some tr ial and error. The smalles t mask that wi l l do the job is the most

desirable because i t wi l l cause the leas t increase in dead space, usua lly be easiest

to hold, and wi l l be less l ikely to resul t in pressure on the eyes. If a sea l is d if ficult

to establ ish, reshaping the mark 's malleable perimeter, altering the amount of air in

the seal or se lec ting a different mask may be helpful. If there is a leak , the mask

should be checked to make certain there is an adequate sea l between the breathing

system connector and the mask (29).

One-hand Method There are several methods of holding a mask to maintain an open ai rway and a

t igh t sea l. A commonly used method is shown in Figure 16.6. The thumb and index

f inger on the lef t hand are placed on the mask body on opposi te s ides of the

connec tor. These fingers push downward to hold the mask to the face and prevent

leaks . Addi tional downward pressure, if required, can be exerted by the anesthesia

provider's chin pushing down on the mask elbow. The remaining three f ingers are

placed on the mandible or the inferior part of the mask. I t is important that the

applied pressure does not decrease a irway patency. Care should be taken to

prevent pressure on the eyes. In some cases, i t may be necessary to extend the

f ingers to the right s ide of the mask to get a good seal. It may be benef ic ia l to

gather part of the le ft cheek around the lef t base of the mask wi th the palm of the

left hand.

Page 6: Chapter 16. Face Masks & Airways

View Figure

Figure 16.5 Adding scent to the face mask may make inhalational inductions more pleasant.

P.447

View Figure

Figure 16.6 Holding the mask with one hand.

Two-handed Method A second method can be used to open any but the most dif f icul t ai rway and obtain a

t igh t fi t (Fig. 16.7). It requires two hands, so a second person is necessary if

assisted or control led resp iration is needed. The thumbs are placed on ei ther side

of the mask body. The index f ingers are placed under the angles of the jaw. The

Page 7: Chapter 16. Face Masks & Airways

mandib le is li f ted and the head extended. If a leak is present, downward pressure

on the mask can be inc reased by pressing down on the mask wi th the anesthesia

provider's chin on the mask elbow (Fig.16.7B). If a second person is not availab le,

the anes thesia venti lator can be used to supply posi tive pressure wh ile the ai rway

is held open by both hands (30).

Two-handed Jaw Thrust Another method to open the airway is to have one person stand a t the head of the

patient and perform a jaw th rus t at the ang le of the left mandible whi le the right

hand compresses the reservoir bag (31). The second person s tands at the patient 's

shoulder, fac ing the f irs t person. This second person's right hand covers the lef t

hand of the fi rs t person, and the lef t hand ach ieves righ t-s ided jaw thrust and mask

seal.

Claw Hand Technique The c law hand technique is useful for children undergoing short-durat ion

ophthalmic procedures (32). The anesthesia prov ider stands a t the s ide o f the bed,

P.448

fac ing the chi ld . The face mask is appl ied to the face by using the right hand wi th

the palmer surface fac ing upward. The ring f inger goes under the angle of the jaw,

and the middle f inger is placed under the angle of the jaw on the lef t . The index

f inger and thumb encirc le the body of the mask. The anesthesia provider then

t igh tens the grip on the mask to achieve a good f it . The l i tt le f inger is kept free.

Page 8: Chapter 16. Face Masks & Airways

View Figure

Figure 16.7 A: Holding the mask with two hands. Also shown is the Esmarch-Heiberg maneuver, which involves dorsiflexion at the atlanto-occipital joint and protrusion of the mandible anteriorly by exerting a forward thrust on the rami. B: The anesthesiologist's chin on the mask elbow helps to create a better seal between the mask and the patient's face.

Bearded Patient Techniques Achieving a sat isfac tory mask f it is of ten d if f icu lt when the patien t has a beard. One

so lut ion is to shave the beard (33). This is usual ly not acceptab le to the patient.

Another solut ion is to use a supraglottic ai rway device (34).

In some cases, i t may be possible to place the round end of the mask between the

lower l ip and the a lveolar ridge (35). The beard may be covered wi th a c lear

adhesive drape, a def ibri llator pad wi th a ho le cu t in the middle, plastic c ling wrap,

or gel and gauze (36,37,38,39,40,41,42,43). A small mask can be placed over the

nose and the mouth held shut (44).

Mask Ventilation of the Tracheostomy Stoma A Rende ll -Baker-Soucek mask can be used over a tracheotomy stoma to achieve

control led or assis ted vent ilation (8,9). It is placed around the s toma wi th the nasal

port ion poin ting in a caudal di rection so that the mand ibular curve rests on the

tracheal reg ion and the apex on the suprasternal notch.

Difficult Face Mask Ventilation Diff icul t mask vent ilat ion is reported to occur in 5% to 6% of anesthetics (45,46,47).

A varie ty of fac ial charac teris tics (fa t, emaciated, and edentulous faces as wel l as

those wi th prominent nares, burns, f la t noses, receding jaws) or other p rob lems

(dra inage tubes in the nose) wi l l be encountered in c l inical p rac tice. Pred ictors of

diff icul t mask vent ilat ion inc lude male gender, a beard, lack of teeth, age over 55

Page 9: Chapter 16. Face Masks & Airways

years , macroglossia , high body mass index, a his tory of snoring , increased

Mal lampati score, and low thyromenta l dis tance (46,47).

A histo ry of s leep apnea does not seem to be a pred ic tive sign of d if f icul t mask

venti lation (48). The ai rway in these patients may be improved by plac ing the head

in the sn if f ing posi tion (49).

The edentulous patient p resents the mos t common prob lem. There is loss of bone

of the alveolar ridge, causing a loss of dis tance between the po in ts where the mask

rests on the mandible and the nose. Furthermore, the buccinator muscle loses i ts

tone in these patien ts . The cheeks sag, creat ing gaps between them and the mask.

Alveolar process resorption resul ts in shrink ing a t the corners o f the mouth . Means

to improve mask f i t inc lude inserting an oral a irway, leaving the patient 's dentures

in p lace, packing the cheeks wi th gauze sponges , and insert ing the inferior margin

of the mask between the g ingiva of the mandible and the lower l ip (50,51).

Pat ients with fac ial deformit ies are part icularly challenging . Mask appl ication wi th

the nasa l project ion poin ted inferiorly has been used for children with certain facial

deformities (3) and fo r patients wi th acromegaly (10).

I f the mask is too smal l wi th an oral airway in place, the oral ai rway should be

removed and a nasal ai rway used. If a mask is too long, the mouth can be

elongated by inserting an oral ai rway.

I f mask venti lat ion is necessary for the patient who has a nasogastric tube in place,

there is usually a leak around the tube as i t ex i ts the side of the mask . This leak

can be improved by adding denture adhesive around the tube at the point where i t

ex its the mask (52).

Dead Space The face mask and its adaptor normally const itute the major part of the mechanical

dead space. Dead space may be decreased by increasing the pressure on the

mask, changing the volume of the cushion, us ing a smaller mask, ex tending the

separation between the insp iratory and expira tory channels close to or into the

mask, or blowing a jet of f resh gas into the mask. The leak induced wi th a poorly

f it ting face mask wi l l reduce the dead space o f the breathing system during

spontaneous venti la tion (53).

Mask Straps A mask strap (mask holder, inhaler re tainer, head strap, head harness, mask

harness, mask retainer, headband, head-res training strap) is used to ho ld the mask

f irmly on the face. Part icular care needs to be pa id to main taining the airway when

Page 10: Chapter 16. Face Masks & Airways

us ing a mask s trap because obstruc tion is more l ikely to go unrecognized than

when the mask is being he ld by the anesthesia provider's hand.

A typical mask strap (Fig . 16.8) consists of thin s tr ips arranged in a c i rc le wi th two

or four p rojec tions . The head res ts in the c irc le, and the straps attach around the

mask connector. The straps a t the jaw may tend to pull the jaw pos teriorly.

Crossing the two lower s traps under the chin may resul t in a better f it and

counteract the pul l f rom the upper straps so that there is less tendency for the

mask to creep up above the bridge of the nose (54). The best s trap applica tion is a

matte r of indiv idual pre ference and may be the best resul t of a tr ial -and-error

process (55).

Care must be taken tha t the s traps are no tighter than necessary to achieve a seal

in o rder to avoid pressure damage f rom the mask or the straps. They should be

released periodically and the mask moved sl igh tly. Gauze

P.449

sponges p laced between the straps and the skin wil l help to pro tec t the face f rom

excess ive pressure. Another r isk of using a mask strap is that i t wi l l take longer to

remove the mask if vomit ing or regurgitation occurs.

View Figure

Figure 16.8 Mask straps.

Advantages Using a face mask is associated with a lower incidence of sore th roat and requires

less anesthet ic depth than using a supraglottic device or a tracheal tube. Muscle

Page 11: Chapter 16. Face Masks & Airways

relaxants do not need to be used to tolerate the mask. The face mask may be the

most cost-eff ic ient method to manage the airway for short cases (56).

Disadvantages With a face mask, one or more of the anes thesia provider's hands are in continuous

use, and higher f resh gas flows are of ten needed. During remote anes thesia

(magnetic resonance imaging and computerized tomography scanning) ai rway

access is diff icul t. Compared with pat ients who are managed wi th a supraglo tt ic

ai rway device , pat ients who are managed wi th a face mask have more episodes of

oxygen desaturation, require more intraoperat ive airway manipulat ions, and present

more dif f icult ies in maintaining an ai rway (57,58). In spontaneously breathing

patients, the work of breathing is higher wi th a face mask than wi th a supraglo ttic

ai rway device or a trachea l tube (59). Using an a irway and/or cont inuous posi t ive

pressure wi l l reduce the work of breathing.

Complications Skin Problems Dermatit is may occur if the pat ient is al le rgic to the material f rom which the mask is

made (60). The pattern of the dermati tis follows the area of contact between the

mask and skin. Chemical or gas steri l ization of reusable masks can leave a residue

that can cause a sk in reac tion (61,62). Pressure nec ros is under the face mask has

been reported following prolonged mask app licat ion in the presence of hypotension

(1).

Nerve Injury Pressure f rom a mask or mask strap may resu lt in pressure inju ry to underlying

nerves. Forward jaw displacement may cause a s tretching nerve inju ry. Fortunately,

the sensory and motor dysfunct ions reported have been transient

(63,64,65,66,67,68). If excessive pressure on the face or extreme forward jaw

displacement must be exerted, tracheal intubation or a suprag lo tt ic ai rway dev ice

should be considered. The mask should be removed f rom the face periodically and

readjusted to make certa in that cont inuous pressure is not appl ied to one area.

Foreign Body Aspiration The diaphragm of an endoscopic mask may rupture during tracheal tube insert ion,

and a piece may be pushed into the pat ien t's tracheobronchial tree (69,70,71,72).

Other parts of a mask or mask strap may be aspira ted (73,74).

Page 12: Chapter 16. Face Masks & Airways

Gastric Inflation When posi tive pressure venti lat ion is used wi th a face mask, gases are likely to

enter the stomach (75,76). I t is recommended that inspiratory pressure be kept

below 20 cm H2O (75).

Eye Injury A corneal abrasion may be caused by a face mask inadvertently placed on an open

eye (77). Chemicals that enter the mask cushion during c leaning and disinfection

can be expelled from cracks in the cushion and come in to contact wi th the eye

when the mask is app lied to the face (78,79,80,81). Pressure on the medial ang les

of the eyes and supraorb ital margins may resul t in eyel id edema, chemos is of the

conjunctiva, pressure on the supraorb ital or supratrochlear nerve, cornea l inju ry,

and temporary blindness f rom central ret inal artery occlusion (82).

Mask Defects A mask wi th a plast ic membrane tha t occluded the connector has been reported

(83). Another mask had a metal

P.450

wi re st icking out of it (84). Incorrect assembly of a face mask has been reported

(85).

Cervical Spine Movement Most but not al l s tudies show tha t mask vent i lat ion moves the cerv ical spine more

than commonly used methods of tracheal intubat ion (86 ,87,88,89). This may be of

s ignif icance in the patient with an uns table cerv ica l spine injury.

Latex Allergy I f rubber is a component of a face mask, a serious reac tion can occur in the pat ient

wi th latex al lergy (90,91). This is discussed more fully in Chapter 15. Because of

the seriousness of this problem, non-latex masks are recommended wherever

possible.

Lack of Correlation between Arterial and End-tidal Carbon

Dioxide The dif fe rence between arte rial and end-tidal carbon dioxide levels is higher wi th

face mask venti lat ion than wi th a trachea l tube or supraglo tt ic device, particularly

wi th small tidal volumes (92).

Page 13: Chapter 16. Face Masks & Airways

Environmental Pollution Studies show that using a face mask is associated with greater operating room

pollut ion wi th anesthet ic gases and vapors than when a trachea l tube or

supraglo tt ic airway dev ice is used (93 ,94). Po llut ion can be reduced by using a

c lose active scavenging dev ice (95,96).

User Fatigue Holding a mask securely onto the face and at the same time main taining the correc t

jaw posi tion can be dif f icult and may resul t in operator fatigue. Fai lure to main tain

the correc t jaw posi tion may resul t in loss of ai rway patency or gas tric dis tention.

Jaw Pain Pos toperative jaw pain is more common af te r mask anesthesia than when a

supraglo tt ic airway dev ice is used (97).

Airways

Purpose A fundamental responsib il i ty of the anesthesia provider is to maintain a patent

ai rway. Unl ike other maneuvers to maintain a patent a irway, such as chin l if t , jaw

thrust, and tracheal intubation, cerv ical spine movement does not occur when an

ai rway is inserted (98).

Figure 16.9A shows the normal unobs tructed ai rway in a supine patient. The airway

passage has a rigid posterior wall , supported by the cerv ical vertebrae, and a

co llaps ib le anterio r wal l, consis ting of the tongue and epiglo tt is . Figure 16.9B

shows the most common cause of ai rway obstruc tion. Under anes thesia, the

muscles in the f loor of the mouth and pharynx support ing the tongue relax, and the

tongue and epiglottis fal l back into the pos terior pharynx, occluding the airway. The

purpose of an ai rway is to l i f t the tongue and epiglottis away f rom the posterio r

pharyngea l wal l and prevent them from obstructing the space above the larynx .

Using maneuvers such as dors if lex ion at the atlanto-occipi ta l jo in t and protrusion of

the mandible anteriorly may s ti l l be necessary to ensure a patent airway (99). An

oral or nasa l ai rway decreases the work of b reathing during spontaneous breathing

us ing a face mask (59).

Page 14: Chapter 16. Face Masks & Airways

View Figure

Figure 16.9 A: The normal airway. The tongue and other soft tissues are forward, allowing an unobstructed air passage. B: The obstructed airway. The tongue and epiglottis fall back to the posterior pharyngeal wall, occluding the airway. (Courtesy of V. Robideaux, M.D.)

Types Oropharyngeal Airways Figure 16.10 shows an oropharyngeal (oral) airway in place. The bi te portion is

between the teeth and lips , and the flange is outside the l ips . The pharyngeal end

rests between the posterio r wal l of the pharynx and the

P.451

base of the tongue and, by exert ing pressure along the base of the tongue, also

pulls the epig lo tt is forward .

Page 15: Chapter 16. Face Masks & Airways

View Figure

Figure 16.10 Oropharyngeal airway in place. The airway follows the curvature of the tongue, pulling it and the epiglottis away from the posterior pharyngeal wall and providing a channel for air passage. (Courtesy of V. Robideaux, M.D.)

In addi tion to helping to maintain an open ai rway, an oropharyngea l ai rway may be

used to prevent a pat ient f rom bi ting and occluding an oral tracheal tube, p rotect

the tongue f rom bi ting, fac il i tate oropharyngeal suct ioning, obta in a better mask f it ,

or p rov ide a pa thway for insert ing dev ices in to the esophagus or pharynx (100).

Oral ai rways have not been associated with an increased inc idence of sore throat

or o ther symptoms (101,102) or bacteremia (103).

Description An oropharyngeal ai rway (Fig. 16.11) may be made of elastomeric material o r

plast ic . I t has a f lange at the buccal end to prevent i t f rom moving deeper into the

mouth. The f lange may a lso serve as a means to f ix the ai rway in place. The bite

port ion is straigh t and f its between the teeth or gums. I t mus t be f i rm enough that

the pat ient cannot c lose the lumen by b it ing. The curved portion ex tends backward

to correspond to the shape of the tongue and palate.

Page 16: Chapter 16. Face Masks & Airways

View Figure

Figure 16.11 Oropharyngeal airway. All oral airways have a flange to prevent overinsertion, a straight bite block portion, and a curved section.

The American national s tandard (104) specif ies that the size of ora l ai rways be

designated by a number that is the length in centimeters (Fig. 16.11).

Specific Airways Guedel Airway The Guedel airway (Fig . 16.12) has a large f lange, a reinforced b ite portion, and a

tubular channel. Modif icat ions to a id f lex ib le f iberoptic intubation have been

described (105,106,107,108).

Berman Airway The Berman ai rway (Figs . 16.10, 16.11) has a center support and open sides. The

center support may have openings . There is a f lange at the buccal end. The s ide

opening can be opened wider to engage or disengage a tracheal tube.

Patil-Syracuse Endoscopic Airway The Patil-Syracuse endoscopic a irway was designed to aid f iberoptic intubat ion

(109). I t has late ral channels and a central groove on the l ingual surface to al low a

f iberscope wi th a tracheal tube to pass. A s lit in the dista l end allows the f iberscope

to be manipulated in the anteroposterio r di rec tion but

P.452

l imi ts lateral movement (110). I t is made from aluminum. The ai rway must be

removed f rom the oropharynx before a tracheal tube can be advanced over the

f iberscope and into the g lott is .

Page 17: Chapter 16. Face Masks & Airways

View Figure

Figure 16.12 Guedel airways. The bite portions are color coded to provide easy identification of size. (Courtesy of Mercury Medical.)

View Figure

Figure 16.13 Williams airway intubators. (Courtesy of Mercury Medical.)

Williams Airway Intubator The Wi ll iams ai rway intubator was designed for b lind orotracheal intubations

(111,112,113). It can also be used to aid f iberopt ic intubat ions or as an oral airway

(114).

The airway, shown in Figure 16.13, is p lastic and avai lable in two s izes, #9 and

#10, which wi l l admit up to an 8.0 or 8.5 in ternal diameter (ID) t racheal tube,

respectively. The tracheal tube connector should be removed during intubation,

Page 18: Chapter 16. Face Masks & Airways

because i t wi l l not pass through the airway unless the a irway is modif ied (115). The

proximal half is cylindrical , whi le the distal half is open on its l ingual surface.

A comparison of the Wi ll iams airway intubator wi th the Ovassapian f iberoptic

intubat ing airway found that the Wil l iams airway intubator provided a better v iew of

the glott is in a s ignif icant number of patients (116).

Ovassapian Fiberoptic Intubating Airway The Ovas-sapian f iberoptic intubating a irway (F ig. 16.14) was designed to del iver a

f iberscope as c lose to the larynx as poss ib le (110). It has a f lat l ingual surface that

gradually widens at the distal end. At the bucca l end are two vert ical s idewalls .

Between the sidewalls is a pair of guide wal ls that curve toward each other. The

guide wa lls are f lex ible and permit the ai rway to be removed f rom around the

tracheal tube. The prox imal half is tubu lar so that i t can func tion as a bi te block .

The distal half is open pos teriorly. I t wi l l accommodate a tracheal tube up to 9 .0

mm ID. Placing a black l ine along the midd le of the ai rway helps to identify the

midline and fac il i tates advancing the f lex ib le endoscope (117,118,119).

View Figure

Figure 16.14 Ovassapian fiberoptic intubating airway. (Courtesy of A. Ovassapian, M.D.)

Page 19: Chapter 16. Face Masks & Airways

View Figure

Figure 16.15 Berman intubation pharyngeal airways.

A comparison of this ai rway with the Wi ll iams airway intubator and the Berman

intubat ing airway (see below) found tha t both other ai rways provided a good v iew o f

the glott is (116,120).

Berman Intubating/Pharyngeal Airway The Berman intubat ing/pharyngeal a irway (Berman II) (Fig. 16 .15) is tubular along

i ts en ti re length. I t is open on one s ide so that i t can be sp li t and removed from

around a tracheal tube. I t can be used as an oral airway or as an aid to f iberoptic

or b lind orotracheal intubation (121). When the f iberscope is in the a irway, the t ip

cannot be bent, l imi ting scope maneuverabil ity. Part ially wi thdrawing the ai rway wi l l

improve maneuverabil i ty (110).

One study found that the Berman ai rway offered an advantage over tongue

retraction for f iberoptic intubation (122). Comparison wi th the Ovassapian

intubat ing airway found that the Berman ai rway o ffers somewhat easier

v isualiza tion of the cords (120). However, if the tracheal tube impinges on the

ai rway, i t is more di ff icul t to complete the intubat ion wi th the Berman ai rway.

Manipu lat ing the Berman ai rway in place can lead to successful trachea l in tubat ion

(120).

Use Pharyngeal and laryngea l ref lexes should be depressed before an oral ai rway is

inserted to avoid cough ing or laryngospasm.

Page 20: Chapter 16. Face Masks & Airways

Selec ting the correc t size is important. Too smal l an a irway may cause the tongue

to k ink and fo rce part of

P.453

i t agains t the roof of the mouth, causing obs truction. Too large an ai rway may

cause obs truct ion by displacing the epiglottis posterio rly and may traumatize the

larynx. The correc t s ize can be estimated by holding the ai rway next to the patien t's

mouth. The tip should rest cephalad to the angle of the mandible. The best c ri te ria

for proper s ize and posit ion are unobstructed gas exchange. I f the airway

repeated ly comes out of the mouth , it shou ld be removed and a smaller s ize tr ied.

View Figure

Figure 16.16 Insertion of oral airway. The airway is inserted 180° from the final resting position.

Wetting or lubricating the ai rway may fac il i tate insertion. The jaw is opened wi th

the lef t hand. The tee th or gums are separated by pressing the thumb against the

lower teeth or gum and the index or thi rd f inger against the upper teeth or gum.

One method to insert an ai rway is shown in Figure 16.16. The a irway is inserted

wi th i ts concave s ide fac ing the upper l ip . When the junct ion of the b ite portion and

the curved sect ion is near the inc isors, the ai rway is rotated 180° and s lipped

behind the tongue into the f inal posi tion. If res istance is met during insertion , it can

usual ly be overcome by a jaw thrust.

An al te rnate method of insert ion is shown in F igure 16.17. A tongue blade is used

to push forward and depress the tongue. The ai rway is inserted wi th the concave

s ide toward the tongue. As i t is advanced, it is ro tated to s lide around behind the

tongue.

Page 21: Chapter 16. Face Masks & Airways

I f the a irway has been used to faci l i ta te f iberoptic intubat ion, i t may be bette r to

remove the ai rway after the f iberscope has entered the trachea because the airway

might prevent the tracheal tube f rom passing into the trachea (123).

Bite Block A bi te block (gag, mouth prop, bi te protector) is placed between the molar teeth or

gums but not the inc isors (124). It is intended to prevent the teeth from bit ing on a

tracheal tube, supraglo tt ic a irway dev ice, f iberscope, or other device. Not on ly wi l l

i t protect these devices, but i t may also avoid dental in jury (125). A bi te block is

also used during elec troconvulsive therapy and in unconscious indiv iduals to

protect the tongue and lips. An oral ai rway should not be used for these purposes

(124,126,127). It is ineffec tive and may be harmful to the patient in this role

because a ll the power of the bi te is concentrated on the inc isors, which are not

designed for this pressure and are l iable to break or loosen. Because a bite block

does not extend in to the pharynx, it is usually less i rri ta ting than an oral ai rway.

View Figure

Figure 16.17 Alternative method of inserting an oral airway. A tongue blade is used to displace the tongue forward.

A varie ty of bite blocks have been developed (Figs. 16.18,16.19,16.20). Some have

a channel for gas to pass. Many have an a ttached string that can be p inned to the

P.454

patient 's gown or taped to the sk in so that it can be easily retr ieved. The curved

port ion of an ora l ai rway can be removed, leaving the remaining port ion as a bi te

block (128,129). A gauze roll is soft and allows pressure to be distribu ted over

Page 22: Chapter 16. Face Masks & Airways

several teeth (127). A bi te b lock may be part of a device used to secure a tracheal

tube.

View Figure

Figure 16.18 Bite block. This is placed between the teeth or gums (preferably in the molar area) to prevent occlusion of a tracheal tube or damage to a fiberoptic endoscope or to keep the mouth open for suctioning.

View Figure

Figure 16.19 This bite block is designed to be placed between the molar teeth with the flat portion extending toward the side of the face. The flat portion is used to grip for insertion and removal. (Picture courtesy of Hudson RCI.)

A bi te block may become deformed so that it does not p revent bi t ing (130). A bi te

block may be aspira ted (130).

Nasopharyngeal Airways A nasopharyngeal airway (nasal ai rway, nasal trumpet) is shown in posi tion in

Figure 16.21. When fully inserted, the pharyngeal end should be below the base of

the tongue but above the epiglottis (131).

Page 23: Chapter 16. Face Masks & Airways

A nasal ai rway is be tter tole rated than an oral ai rway if the patient has intact ai rway

ref lexes. I t is p referable to use a nasal airway if the pat ien t's teeth are loose or in

poor condit ion or there is trauma or pathology of the oral cavi ty . It can be used

when the mouth cannot be opened.

View Figure

Figure 16.20 Oberto mouth prop, which is used for protecting the teeth during electroconvulsive therapy. (Courtesy of Rusch, Inc.)

View Figure

Figure 16.21 The nasopharyngeal airway in place. The airway passes through the nose and extends to just above the epiglottis. (Courtesy of V. Robideaux, M.D.)

Contraindicat ions to us ing a nasopharyngeal ai rway inc lude ant icoagula tion; a

basilar skul l f rac ture; pathology, seps is , or deformity of the nose or nasopharynx;

or a history of nosebleeds requ iring medical treatment. There is no ev idence that

nasal airways cause s ign if icant bacteremia (103,132).

Page 24: Chapter 16. Face Masks & Airways

Nasopharyngeal airways have been used during and after pharyngeal surgery

(133,134), in infan ts wi th Pierre Rob in syndrome (135), to app ly continuous posit ive

ai rway pressure (CPAP) (136), to fac i li tate suct ioning (137), as a guide for a

f iberscope (137), to treat s ingultus (hiccups), as a guide fo r a nasogastric tube

(138,139), to dilate the nasal passages in preparat ion for nasotrachea l intubation

(138) and as a means to main tain the airway and admin ister anes thesia during

denta l surgery (140).

A nasopharyngeal airway can be fi tted wi th a tracheal tube connector and used with

an anesthesia breathing sys tem (141,142,143,144,145,146,147,148). These dev ices

have been used to maintain vent i lation during oral fiberopt ic endoscopy (109) and

to admin is ter cont inuous posi tive pressure (136). Environmental gas contaminat ion

may be a problem with this techn ique (149).

Description A nasopharyngeal airway resembles a shortened tracheal tube wi th a f lange at the

outer end to prevent i t f rom completely pass ing into the naris . Some airways come

wi th a safety pin that can be inserted in to the flange or ai rway wa ll (150). The

f lange is movable on some models . The American standard (104) requ ires that the

s ize of a nasopharyngeal ai rway be designated by the inside diameter in

mi ll imeters.

P.455

Page 25: Chapter 16. Face Masks & Airways

View Figure

Figure 16.22 Nasopharyngeal airways. The one on the right does not contain latex.

Specific Airways A varie ty of nasopharyngeal airways are ava ilable. Some are shown in Figure

16.22 .

Linder Nasopharyngeal Airway The Linder nasopharyngea l (bubble-tip) ai rway is shown in F igure 16.23. It is

plast ic with a large f lange (151,152). The distal end has no bevel . The ai rway is

suppl ied wi th an introducer, which has a balloon on its tip. The bal loon can be

inf lated and def lated by attaching a syringe to the one-way v alve at the other end of

the in troducer.

Page 26: Chapter 16. Face Masks & Airways

View Figure

Figure 16.23 Linder nasopharyngeal airway. (Courtesy of Polamedco, Inc.)

Before insert ion, the introducer is inserted into the ai rway unt i l the t ip of the

balloon is just pas t the end. Air is in jected unti l the bal loon t ip is inf lated to

approximately the outs ide diameter of the tube. The complete assembly is

lubricated and then inserted through the nos tri l . Af te r it is in p lace, the balloon is

deflated and the introducer removed.

Cuffed Nasopharyngeal Airway The cuffed nasopharyngea l (pharyngeal) airway is s imilar to a short cuffed tracheal

tube (147,148,153). I t is inserted through the nose into the pharynx, the cuff is

inf lated, and then is pulled back unt i l res is tance is fel t.

Binasal Airway The binasal a irway (F ig. 16.24) consists of two nasal ai rways jo ined together by an

adaptor fo r attachment to the breath ing system

(154,155,156,157,158,159,160,161). I t can be used to administer anesthesia or to

provide CPAP to babies.

Insertion The diameter of the nasa l ai rway should be the same as needed to insert a trachea l

tube (0 .5 to 1 mm smal le r than for an ora l tracheal tube).

Page 27: Chapter 16. Face Masks & Airways

View Figure

Figure 16.24 Binasal airway. (Courtesy of Rusch, Inc.)

P.456

View Figure

Figure 16.25 Insertion of a nasal airway. A: Correct method. The airway is inserted perpendicularly, in line with the nasal passage. B: Incorrect method. The airway is being pushed away from the air passage and into the turbinates.

Before insert ion, the nasal ai rway should be lubricated thoroughly along i ts en ti re

length. Each side of the nose should be inspected for s ize, pa tency, and the

Page 28: Chapter 16. Face Masks & Airways

presence of polyps. A vasoconstric tor may be appl ied before insert ion to reduce

trauma.

The nasopharyngeal ai rway shou ld be inserted as shown in Figure 16.25A. The

ai rway is he ld wi th the bevel against the septum and gently advanced posterio rly

whi le being rotated back and fo rth. I f res is tance is encountered during insert ion ,

the other nostri l or a smal ler size ai rway should be used. F igure 16.25B shows an

incorrect method for insert ing the ai rway. The a irway is being pushed toward the

roof of the nose.

The airway may be adjus ted to fi t the pharynx by s liding i t in or out. If the tube is

inserted too deeply, laryngeal ref lexes may be stimu lated ; if too short, ai rway

obstruct ion wil l not be relieved. Al though the correct nasal airway length fo r a

patient correlates with s imple external measurements of the face and neck, s tudies

indicate a correlat ion only with the patient 's height (131).

Complications Airway Obstruction The t ip of an ai rway can press the epiglo tt is o r tongue against the pos terior

pharyngea l wal l and cover the laryngeal aperture (162,163). With a nasopharyngea l

ai rway, neck movement in rotat ion or anteropos teriorly may resul t in the lumen

becoming obstructed (164). The use of a fenestrated ai rway may overcome this

problem. The nasopharyngeal ai rway lumen may be compressed ins ide the nose

(131).

A foreign body can enter an airway and cause complete or partial obstruc tion

(165,166). In one case, the plast ic packaging became s tretched over the end of the

ai rway, causing obstruct ion (167).

I f a nasopharyngeal a irway perfo rates the re tropharyngeal space, the space may

expand and cause a irway obs truct ion (168).

Trauma Inju ry to the nose and posterior pharynx is a po tential compl ication of nasa l

ai rways. Epis taxis is usual ly self -l imit ing but can present a serious prob lem in some

patients. To control severe nasal and nasopharyngeal b leeding, an epis taxis a irway

(Fig . 16.26) can be used. Pharyngeal perforat ion and retropharyngeal abscess

formation can occur (168). The lip or tongue may be caught between the teeth and

an oral a irway.

Tissue Edema

Page 29: Chapter 16. Face Masks & Airways

Facial, neck , or tongue edema, either uni lateral or bilateral, can occur fol lowing

surgery, especially in the s i t ting pos ition, and can resul t in a irway obs truction

(169,170,171,172,173). Pressure f rom an oral ai rway may be a

P.457

contribut ing factor. To prevent th is compl ica tion, the oral airway should not be lef t

in p lace fo r an extended period . A bite block should be used ins tead. Excess ive

head or neck flexion should be avoided, and the head and neck should be checked

f requently during long cases for edema or ecchymoses. Uvular edema possibly

caused by the uvula becoming entrapped between the hard pa late and an

oropharyngea l ai rway has been reported (174). Another patient developed

temporary deafness secondary to edema of the uvula and soft pala te following

prolonged nasopharyngeal ai rway use (175). Transien t sal ivary gland swel l ing may

occur wi th oral airway use (176,177).

View Figure

Figure 16.26 Epistaxis airway. This is inserted into the nose and inflated to provide local pressure on the bleeding site. It is available in several sizes. (Courtesy of Rusch, Inc.)

Ulceration and Necrosis Ulcerat ion of the nose or tongue can occur if an ai rway remains in place for a long

period of t ime (178).

Central Nervous System Trauma The use of a nasal ai rway in a patient wi th a basi lar skull f racture can resu lt in i ts

entering the anterior cranial fossa (179,180).

Dental Damage

Page 30: Chapter 16. Face Masks & Airways

Teeth can be damaged or avulsed if the patient bites down hard on an oral a irway

(181,182,183). An oral ai rway should be avo ided if there is ev idence of periodontal

disease, teeth weakened by caries or res torat ions, c rowns, f ixed partial dentures,

pronounced proclination (the front tee th having a forward inc l inat ion and

overlapp ing the lower f ront teeth), or isolated tee th. In these cases , the use of a

nasopharyngea l ai rway and/or a bi te block between the back teeth may be

preferable .

Laryngospasm and Coughing Inserting an ai rway before adequate anesthet ic depth is es tablished may cause

coughing or laryngospasm, especial ly if the ai rway contacts the epig lo tt is or voca l

cords .

Retention, Aspiration, or Swallowing Part or a ll of an ai rway may become disp laced into the pharynx, tracheobronchial

tree , or esophagus (184,185,186,187,188,189,190,191). Placing a safety pin

through the nasal ai rway f lange may prevent i t f rom sl ipping into the nose (192).

Devices Caught in Airway In one case, a cuff became detached when an esophageal s tethoscope was

removed f rom a pat ient wi th an oral airway in place (193). It was postulated that the

cuff became caught in the side grooves of the a irway. Another case was reported

where a fiberscope inadvertent ly traversed a fenes trated ora l ai rway, making i t

impossib le to pass a tracheal tube (194).

Equipment Failure An oral ai rway may fracture at the connection between the b ite portion and the

curved sect ion (195,196,197). A defect in the Wi l liams intubat ing airway that could

tear the tracheal tube cuff has been reported (198).

Latex Allergy I f an ai rway contains la tex , a severe react ion may occur i f the patien t is sensi t ive to

latex (90). Non-latex oral and nasal airways are readi ly identif ied and available.

Chapter 15 provides more details on latex al lergy.

Gastric Distention A nasopharyngeal airway that is too long may enter the esophagus wi th resu ltant

gastric d is ten tion (131).

References

Page 31: Chapter 16. Face Masks & Airways

1. Smurthwai te GJ , Ford P. Skin necrosis following continuous pos it ive pressure

wi th a face mask . Anaesthesia 1993;48:147–148.

[Med line Link]

2. Turner R. NPPV: face versus interface. Resp ir Care 1997;42:389–393.

3. Diaz JH. A new transparent disposable plast ic face mask fo r children and adul ts .

Anesthesiology 1993;78:1195–1196.

[Fu ll text Link]

[CrossRef]

[Med line Link]

4. Carson KD, Moriari ty DC. Teach ing ai rway management ski l ls—use of a c lear

PVC face mask . Anaesthesia 1995;50:571.

[CrossRef]

[Med line Link]

5. Rendel l -Baker L, Soucek DH. New paedia tric face masks and anaesthetic

equipment. Br Med J 1962;1 :1690.

[Med line Link]

6. Palme C, Nystrom B, Tunel l R. An evaluation of the eff ic iency of face masks in

the resuscitat ion of newborn infants. Lancet 1985;207–210.

7. Aghdami A, El lis R, Rah KH. A pedia tric face mask can be a useful aid in lung

venti lation on postlaryngectomy pat ients. Anesthesiology 1985;63:335.

[Fu ll text Link]

[CrossRef]

[Med line Link]

8. Northwood D, Wade MJ. Novel use of the Rendell -Baker Soucek mask.

Anaesthesia 1991;46:319.

[CrossRef]

[Med line Link]

9. Kannan S, Birch JP. Control led venti la tion th rough a tracheostomy stoma.

Anaesth Intens Care 2001;29:557.

[Med line Link]

10. Edmondson WC, Rushton A. The upside down facemask again. Anaes thesia

1992;47:361.

[CrossRef]

[Med line Link]

11. Wil l iamson R. Another use fo r paed ia tric masks. Anaes thesia 1992;47:361–

362.

Page 32: Chapter 16. Face Masks & Airways

[CrossRef]

[Med line Link]

12. Rauch RY, Brener CE. Airway management for pediatric

esophagogastroduodenoscopy using an endoscopy mask. Anesth Analg

2003;96:303–304.

[Fu ll text Link]

[CrossRef]

[Med line Link]

13. Fre i FJ, Ummenhofer W. A special mask for teaching f iber-optic in tubat ion in

pedia tric pa tients. Anesth Analg 1993;76:458.

[Med line Link]

14. Fre i FJ, Wengen DF, Rutishauser M, et al . The a irway endoscopy mask : useful

device for f ibreopt ic evaluat ion and in tubation of the paediatric ai rway. Paed

Anaesth 1995;5:319–324.

15. Aoyama K, Yasunaga E, Takenaka I, et al . Posit ive pressure venti lation during

f ibreoptic in tubat ion : comparison of the laryngeal mask airway, in tubating laryngeal

mask and endoscopy mask techn iques. Br J Anaes th 2002;88:246–254.

[Fu ll text Link]

[CrossRef]

[Med line Link]

16. Uda R, Ohtsuka M, Minami T, et al. Use of a ligh ted stylet for trachea l

intubat ion through an intubating port of a mask. Anesthesiology 1999;91:1560–

1561.

[Fu ll text Link]

[CrossRef]

[Med line Link]

17. Carson K. Pat ients prefer scented face masks. Can J Anaesth 1995;42:482–

483.

18. Hinkle A. Pat ien t safety and scented pediatric anesthesia facemasks. J Cl in

Anesth 1997;9:521–522.

[CrossRef]

[Med line Link]

19. Hinkle AJ . Scented masks in pediatric anesthesia. Anesthes iology

1987;66:104–105.

[Fu ll text Link]

[CrossRef]

Page 33: Chapter 16. Face Masks & Airways

[Med line Link]

20. Mather C. “Smelly agents.” Anaes thesia 1993;48:540.

[CrossRef]

[Med line Link]

21. Mayhew JF. “Smelly agents .” Anaesthesia 1993;49:1021.

[Med line Link]

22. Yamashita M, Motokawa K. “Frui t-flavored” mask induc tion for children.

Anesthesiology 1986;64:837.

[Med line Link]

23. Mehta S , Stone DN, Whitehead HF. Use of essent ial oi l to p romote induc tion of

anaesthesia in chi ld ren . Anaesthesia 1998;53:720.

[Fu ll text Link]

[CrossRef]

[Med line Link]

24. Cornwal l D. Lip balm “scent” for paediatric anaesthesia. Anaesthesia

2005;60:827.

[Fu ll text Link]

[CrossRef]

[Med line Link]

25. Hinkle AJ . Pat ient sa fety and scented pedia tric anesthes ia facemasks . A reply.

J Cl in Anesth 1998;10:262–263.

[Med line Link]

26. Yash ita M, Tsuneto S. “Normac” falsely recognizes “f ruit extrac t” as an

anesthet ic agent. Anesthesiology 1987;66:97.

[Fu ll text Link]

[CrossRef]

[Med line Link]

27. Gagnon C, Fortier L-P, Donati F. When a leak is unavoidable, preoxygenation is

equally ineffect ive wi th v ital capac ity or t idal vo lume breathing. Can J Anesth

2006;53:86–91.

28. McGowan P, Skinner A. Preoxygenation—the importance of a good face mask

seal. Br J Anaesth 1995;75:777–778.

[Med line Link]

29. Longmate A. Failure of a resuscitation bag and mask. Anaesthesia

2005;60:942–943.

[Fu ll text Link]

Page 34: Chapter 16. Face Masks & Airways

[CrossRef]

[Med line Link]

30. Benyamin RM, Wafai Y, Salem MR, e t al. Two-handed mask vent ilation o f the

diff icul t ai rway by a single ind iv idual. Anes thesiology 1998;88:1134.

[Fu ll text Link]

[CrossRef]

[Med line Link]

31. Benumof JL. Defini tion and inc idence of the dif f icul t ai rway. In: Benumof JL, ed .

Airway Management. Princ iples and Prac tice. St. Louis: Mosby, 1996: 121–125.

32. Podder S, Dutta A, Yaddanapudi S, e t al . Challenges in paediatric mask

holding: the “c law hand” technique. Anaesthesia 2001;56:697–699.

[Fu ll text Link]

[CrossRef]

[Med line Link]

33. Moore EW, Cowan CM. A close shave. Anaes thesia 1999;54:305.

[Fu ll text Link]

[CrossRef]

[Med line Link]

34. Alexander R. Overcoming the beard . Anaesthesia 1999;54:100.

[Fu ll text Link]

[CrossRef]

[Med line Link]

P.458

35. Crooke J. The bearded ai rway. Anaesthes ia 1999;54:500.

[Fu ll text Link]

[CrossRef]

[Med line Link]

36. Hancock AC. The bearded ai rway. Anaesthesia 1999;54:408.

[Fu ll text Link]

[CrossRef]

[Med line Link]

37. Johnson JO, Bradway JA, Blood T. A hairy s i tuat ion . Anes thesiology

1999;91:595.

[Fu ll text Link]

Page 35: Chapter 16. Face Masks & Airways

[CrossRef]

[Med line Link]

38. Thomas DI. Overcoming the beard. Anaesthesia 1999;54:100.

[Fu ll text Link]

[CrossRef]

[Med line Link]

39. Dalgleish DJ. A hairy s i tuat ion. Anesthesiology 2000;92:1199.

[Fu ll text Link]

[CrossRef]

[Med line Link]

40. Vincent C, Ames WA. The bearded ai rway. Anaesthesia 1998;53:1034.

[Fu ll text Link]

[CrossRef]

[Med line Link]

41. Mort imer AJ. An a irtight seal between facemask and beard. Anaes thesia

1980;41:670–671.

[CrossRef]

42. Levy N. Gel pads for the dif ficul t ai rway trol ley. Anaesthesia 2005;60:102.

[Fu ll text Link]

[CrossRef]

[Med line Link]

43. Ames W. How to sk in a bearded cat. Anaesthesia 2000;55:619.

[Fu ll text Link]

[CrossRef]

[Med line Link]

44. Garewal DS. D if f icul t mask vent i la tion. Anesthes iology 2000;92:1199.

[Fu ll text Link]

[CrossRef]

[Med line Link]

45. Adnert F. D if f icul t mask venti la tion. An underest imated aspec t of the problem of

the diff icult airway? Anesthes io logy 2000;92:1217–1218.

46. Langeron O, Masso E, Huraux C, et al . Predict ion of diff icul t mask venti lat ion.

Anesthesiology 2000;92:1229–1236.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 36: Chapter 16. Face Masks & Airways

47. Masso E, Langeron O, Mottet P, e t al . Cri teria fo r predicting dif ficul t mask

venti lation . Br J Anaesth 1999;82:23.

48. Rocchicc ioli CC, Beloucif SS, Blumen MM, et a l. Predict ion of dif f icul t mask

venti lation in pat ients wi th obstruc tive sleep apnea syndrome and predicative signs

of dif ficu lt airway. Presented at the American Society of Anesthesiologists Annual

Meeting, New Orleans, 2001.

49. Isono S, Tanaka A, Ish ikawa T, et al. Sn if f ing posi tion improves pharyngeal

ai rway patency in anesthetized patien ts wi th obstruc tive sleep apnea.

Anesthesiology 2005;103:489–494.

[Fu ll text Link]

[CrossRef]

[Med line Link]

50. Lanier WL. Improving anesthesia mask f it in edentulous pat ients. Anesth Analg

1987;66:1053.

[Fu ll text Link]

[CrossRef]

[Med line Link]

51. Kubota Y, Toyoda Y, Kubota H. Face mask f i t ting fo r edentu lous pat ien ts .

Anesth Analg 1993;76:450.

[Med line Link]

52. Noguchi T, Shiga Y, Koga K, et al . A method to improve a gas leak on mask

venti lation in the patient with a nasogastric tube. Anesthes io logy 2001;94:545.

[Fu ll text Link]

[CrossRef]

[Med line Link]

53. Mil le r DM, Adams AP, Light D. Dead space and paediatric anaes thetic

equipment: a physical lung mode l s tudy. Anaesthesia 2004;59:600–606.

[Fu ll text Link]

[CrossRef]

[Med line Link]

54. Chander S. A new head strap. Anes th Analg 1980;59:457–458.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 37: Chapter 16. Face Masks & Airways

55. McGee JP, Vender JS. Nonin tubat ion management of the ai rway. In: Benumof

JL , ed. Cl inical Procedures in Anesthesia and Intensive Care. Philadelphia: JB

Lippincott Co., 1992: 89–114.

56. Macario A, Chang PC, S tempel DB, e t al . A cost analysis of the laryngea l mask

ai rway for elective surgery in adul t ou tpatients. Anesthesio logy 1995;83:250–257.

[Fu ll text Link]

[CrossRef]

[Med line Link]

57. Johnston DF, Wrigley SR, Robb PJ, et al. The la ryngeal mask ai rway in

paediatric anaes thesia. Anaesthesia 1990;45:924–947.

[CrossRef]

[Med line Link]

58. Smith I, Whi te PF. Use of the laryngeal mask airway as an al ternative to a face

mask during outpatient a rthroscopy. Anesthes io logy 1992;77:850–855.

[Fu ll text Link]

[CrossRef]

[Med line Link]

59. Keidan I , Fine GF, Kagawa T, et al. Work of breathing during spontaneous

venti lation in anesthetized children: a comparat ive s tudy among the face mask,

laryngea l mask ai rway and endotracheal tube. Anesth Analg 2000;91:1381–1388.

[Fu ll text Link]

[CrossRef]

[Med line Link]

60. Begenau VG. Al le rgic dermatit is due to rubber: report of case. Anes thesiology

1951;12:771–772.

[Fu ll text Link]

[CrossRef]

[Med line Link]

61. Anonymous . Physic ian and the law. Anesth Analg 1970;49:889.

62. Potgie ter SV, Mostert JW. A hazard assoc ia ted wi th the use of a face mask:

case report. S A Med J 1959;33:989–990.

63. Azar I , Lear E. Lower l ip numbness fol lowing anes thesia. Anesthesiology

1986;65:450–451.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 38: Chapter 16. Face Masks & Airways

64. Ananthanarayan C, Ro lb in SH, Hew E. Facial nerve para lysis fol lowing mask

anaesthesia . Can J Anaesth 1988;35:102–103.

[Med line Link]

65. Glauber DT. Facial paralys is after general anesthes ia . Anes thesiology

1986;65:516–517.

[Fu ll text Link]

[CrossRef]

[Med line Link]

66. Barron DW. Supra-orbi tal neurapraxia. Anaes thesia 1955;10:374.

[CrossRef]

[Med line Link]

67. Keats AS. Pos t-anaesthetic cepha lagia . Anaesthesia 1956;11:341–343.

[CrossRef]

[Med line Link]

68. Bhuiyan MS, Chapman M. Mental nerve in ju ry following facemask anaesthesia.

Anaesthesia 2006;61:516–517.

[Fu ll text Link]

[CrossRef]

[Med line Link]

69. Zornow MH, Mitchell MM. Foreign body aspirat ion during f iberoptic assis ted

intubat ion. Anesthesiology 1986;64:303.

[Fu ll text Link]

[CrossRef]

[Med line Link]

70. Wil l iams L, Teague PD, Nagia AH. Foreign body f rom a Pat il-Syracuse mask.

Anesth Analg 1991;73:359–360.

[CrossRef]

[Med line Link]

71. Wil l ians L , Grow M, Teague PD, e t al . Foreign body f rom a Pat i l-Syracuse

mask. Anesth Analg 1991;73:359–360.

[CrossRef]

[Med line Link]

72. Waring PH, Vinik HR. A potential complicat ion of the Pat il -Syracuse endoscopy

mask. Anesth Analg 1991;73:668–669.

[Fu ll text Link]

[CrossRef]

Page 39: Chapter 16. Face Masks & Airways

[Med line Link]

73. Gunter JB. A potential chok ing hazard. Anesthes iology 1996;85:228.

[Fu ll text Link]

[CrossRef]

[Med line Link]

74. Ricketts AL, Latto P. Potential foreign body. Br J Anaesth 1986;58:685.

[CrossRef]

[Med line Link]

75. Weiler N, Heinrichs W, Dick W. Assessment of pulmonary mechanics and

gastric inf la tion pressure during mask venti lat ion. Prehosp Disaster Med 1995;100:

101–105.

[Med line Link]

76. Ho-Tai L, Devi tt J , Noel A, et a l. Gas leak and gas tric insuff la tion during

control led vent ilation: face mask versus laryngea l mask a irway. Can J Anaes th

1998;45: 206–211.

[Med line Link]

77. Snow JC, Kripke BJ, Norton ML, e t al . Corneal injuries during general

anesthesia. Anesth Analg 1975;54:465–467.

[Med line Link]

78. Anonymous . Allegedly defective anesthetic mask blamed in eye injury su it .

Biomed Safe Stand 1985;15:109.

79. Durkan W, Fleming N. Potent ia l eye damage from reusable masks.

Anesthesiology 1987;67:444.

[Fu ll text Link]

[CrossRef]

[Med line Link]

80. Murray WJ. A case of eye injury from a reusable anesthet ic mask.

Anesthesiology 1988;68:302.

[Fu ll text Link]

[CrossRef]

[Med line Link]

81. Murray WJ, Ruddy MP. Toxic eye in jury during induct ion of anesthesia. South

Med J 1985;78:1012–1013.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 40: Chapter 16. Face Masks & Airways

82. Munn KA, Will iams RT, Shafto CM. Transient unilateral blindness following

general anaesthes ia: case report. Can J Anaesth 1978;25:433–435.

83. Cook WP, Gravenstein JS. Breathing c ircui t occlusion due to a defect ive

paediatric face mask. Can J Anaesth 1988;35:205–206.

[Med line Link]

84. Gordon HL, Tweedie IE. Facemask hazard . Anaesthesia 1989;44:84.

[CrossRef]

[Med line Link]

85. McDoua ll SF, Campbell RC, Tasker RC. Clinical ale rt : failure in face mask seal

as a consequence of incorrect face mask assembly. Br J Anaesth 2003;91:924–

925.

[Fu ll text Link]

[CrossRef]

[Med line Link]

86. Hauswald M, Sk lar DR, Tandberg D, et al . Cerv ica l spine movement during

ai rway management. Cinef luoroscopic appra isal in human cadavers. Am J Emerg

Med 1991;9:535–538.

[CrossRef]

[Med line Link]

87. Brimacombe J, Kel ler C, Kunzel KH, et a l. Cerv ica l spine motion during ai rway

management: as c inefluoroscopic study of the posteriorly destabi lized cerv ical

vertebrae in human cadavers. Anesth Analg 2000;91:1274–1278.

[Fu ll text Link]

[CrossRef]

[Med line Link]

88. Donaldson WF, Hei l BV, Donaldson VP, et al . The effect of ai rway maneuvers

on the unstable C1-C2 segment: a cadaver study. Spine 1997;22:1215–1218.

[Fu ll text Link]

[CrossRef]

[Med line Link]

89. Turkstra TP, Craen RA, Pelz DM, et al . Cerv ical spine motion: a f luorscopic

comparison during intubat ion with l ighted stylet, GlideScope and Macintosh

laryngoscope. Anesth Analg 2005;101:910–915.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 41: Chapter 16. Face Masks & Airways

90. Paris ian S. Latex allergies causing more anesthesia problems. APSF Newslett

1992;7:1 .

91. McKinstry LJ , Fenton WJ, Barrett P. Anaesthes ia and the pat ient wi th latex

al lergy. Can J Anaesth 1992;39:587–589.

[Med line Link]

92. Ivers D, Verbough C, Th i HP, et al. The qua li ty of breathing and capnography

during la ryngeal mask and face mask venti lat ion. Anaesthesia 1995;50:858–862.

[CrossRef]

[Med line Link]

93. Barnett R, Gal lant B, Fossey S , et al . Nitrous oxide environmental pol lut ion. A

comparison between face mask , la ryngeal mask and endotrachea l intubat ion. Can J

Anaesth 1992;39:A151.

94. Sarma VJ , Leman J . Laryngeal mask and anaesthet ic waste gas concentra tions.

Anaesthesia 1990;45:791–792.

[CrossRef]

[Med line Link]

95. Carlsson P, Ljungqvist B, Hallen B. The effec t of local scaveng ing on

occupationa l exposure to ni trous oxide. Ac ta Anaesth Scand 1983;27:470–475.

[Med line Link]

96. Sik MJ, Lewis RB, Evele igh DJ . Assessment of a scavenging dev ice for use in

paediatric anaes thesia. Br J Anaesth 1990;64:117–123.

[CrossRef]

[Med line Link]

97. Brimacombe J, Holyoake L, Ke ller C, e t al . Pharyngolaryngeal, neck and jaw

discomfort af ter anesthesia with the face mask and laryngea l mask a irway at high

and low cuff volumes in males and females. Anesthesiology 2000;93:26–31.

[Fu ll text Link]

[CrossRef]

[Med line Link]

98. Aprahamian C, Thompson B, Finger WA, et al. Experimental cerv ica l spine

injury model. Evaluat ion of ai rway management and spl in ting techn iques. Ann

Emerg Med 1984;13:584–587.

[CrossRef]

[Med line Link]

99. Marsh AM, Nunn JF, Taylor SJ , et al. Airway obs truction associated wi th the

use of the Guedel airway. Br J Anaesth 1991;67:517–523.

Page 42: Chapter 16. Face Masks & Airways

[CrossRef]

[Med line Link]

100. At las GM. A comparison of f iberoptic-compatible ora l ai rways. J Clin Anaesth

2004;16:66–73.

101. Browne B, Adams CN. Postopera tive sore th roat related to the use of a Guede l

ai rway. Anaesthesia 1988;43:590–591.

[CrossRef]

[Med line Link]

102. Monroe MC, Gravenste in N, Saga-Rumley S. Pos topera tive sore th roat. Ef fect

of o ropharyngeal ai rway in orotracheally in tubated pat ien ts . Anes th Ana lg

1990;70:512–516.

[Fu ll text Link]

[CrossRef]

[Med line Link]

103. Al i MT, Tremewen DR, Hay AJ, e t al . The occurrence of bacteremia associated

wi th the use of o ral and nasopharyngea l ai rways. Anaesthes ia 1992;47:153–155.

[CrossRef]

[Med line Link]

104. American Society fo r Testing and Materials . S tandard specif ication for

anesthet ic equipment—oropharyngeal and nasopharyngeal airways (ASTM F-1573-

95). West Conshohocken, PA: Author, 1995.

105. Wil ton NCT. Aids fo r fiberopt ically guided in tubation in chi ldren.

Anesthesiology 1991;75:549–550.

[Fu ll text Link]

[CrossRef]

[Med line Link]

106. Ayoub C, Itani M, Lteif A, et al . A modif ied Guedel airway for continuous

oxygenation and suctioning during f ibreopt ic bronchoscopy. Anaesthesia

2000;55:455–457.

[Fu ll text Link]

[CrossRef]

[Med line Link]

107. Itan i M. The modif ied Guedel airway for f iberoptic in tubat ion . Middle East J

Anaesthesiol 1999;15:207–210.

[Med line Link]

Page 43: Chapter 16. Face Masks & Airways

108. McGinley J, McAdoo J. Airway adjunc t to an unant ic ipated dif f icult airway.

Anesth Analg 1999;88:467–468.

[Fu ll text Link]

[CrossRef]

[Med line Link]

109. Pat i l V, Stehl ing LC, Zauder HL, et al . Mechanical aids for f iberoptic

endoscopy. Anesthesiology 1982;57:69–70.

[Fu ll text Link]

[CrossRef]

[Med line Link]

110. Ovassapian A. Fiberoptic tracheal intubation in adul ts . In : Fiberopt ic

Endoscopy and the Diff icult Airway. 2nd ed. Phi ladelphia: Lippincott-Raven, 1996:

71–103.

111. Palazzo MGA, Sol tice NJ. A new a id to f iberoptic bronchoscopy. Anaesth

Intens Care 1983;11:388–389.

[Med line Link]

112. Will iams RT. Comments from an experienced user of the airway intubator.

Anesthesiology 1984;61:108–109.

[Med line Link]

113. Will iams RT, Harrison RE. Prone tracheal in tubat ion s implif ied using an ai rway

intubator. Can Anaesth Soc J 1981;28:288–289.

[Med line Link]

114. Loken RG. Mechanical airway maintenance. Can J Anaes th 1989;36:362–

363.

[Med line Link]

115. Chung D. A modif ied Will iams ai rway in tubator to assist f ib reoptic in tubation.

Can J Anaesth 1998;45:95.

[Med line Link]

116. Greenland KB, Lam MC, Irwin MG. Comparison of the Wi ll iams ai rway

intubator and Ovassapian fibreopt ic intubating ai rway for f ibreopt ic orotrachea l

intubat ion. Anaesthes ia 2004;59:173–176.

[Fu ll text Link]

[CrossRef]

[Med line Link]

117. Aoyama K, Seto A, Takenaka I . Simple modif ication of the Ovassapian

f iberoptic in tubat ing ai rway. Anesthes io logy 1999;91:897.

Page 44: Chapter 16. Face Masks & Airways

[Fu ll text Link]

[CrossRef]

[Med line Link]

118. Rav indran RS. Another advantage o f mark ing Ovassapian fiber-optic

intubat ing airway. Anes thesiology 2000;92:1843.

[Fu ll text Link]

[CrossRef]

[Med line Link]

119. Aoyama K, Nagaoka E, Takenaka I, et al . Another advantage of marking

Ovassap ian f iber-op tic in tubating ai rway. Anesthes iology 2000;92:1843.

[Fu ll text Link]

[CrossRef]

[Med line Link]

120. Randel l T, Val l i H, Hakala P . Comparison between the Ovassap ian intubating

ai rway and the Berman intubating ai rway in f iberoptic intubation . Eur J Anaesth

1997;14:380–384.

[CrossRef]

[Med line Link]

121. Berman RA. A method fo r bl ind oral intubation of the trachea or esophagus .

Anesth Analg 1977;56:866–867.

[Fu ll text Link]

[CrossRef]

[Med line Link]

122. Smith JE, Mackenzie AA, Scott-Knight VCE. Comparison of two methods of

f iberscope-gu ided tracheal intubat ion. Br J Anaesth 1991;66:546–550.

[CrossRef]

[Med line Link]

123. Pandit JJ, Dravid RM, Iyer R, e t al . Oro tracheal f ib reoptic intubat ion for rapid

sequence induct ion of anaesthes ia . Anaesthesia 2002;57:123–127.

[Fu ll text Link]

[CrossRef]

[Med line Link]

124. King H-K, Lewis K. Guedel oropharyngeal ai rway does not prevent pat ien t

bi ting on the endotracheal tube. Anaesth Intens Care 1996;24:729–730.

[Med line Link]

Page 45: Chapter 16. Face Masks & Airways

125. Quinn JB, Schul theis LW, Schumacher GE. A too th broken after laryngoscopy:

unlikely to be caused by the force appl ied by the anes thesiologist. Anesth Analg

2005;100:594–596.

[Fu ll text Link]

[CrossRef]

[Med line Link]

126. Kwan KM, Kok P, Koay CK. Prevent ion of tube occlusion caused by bi ting oral

bi te block versus oropharyngeal ai rway. Anaesth Intens Care 2000;28:227.

[Med line Link]

127. Negus B. Gauze b ite block. Anaes th Intens Care 1997;25:589.

[Med line Link]

P.459

128. Chaffe A , Street MN. A modif ied airway. Anaesthesia 1988;43:611.

[CrossRef]

[Med line Link]

129. Schwartz AJ, Dougal RM, Lee KW. Modif ication of ora l ai rway as a b ite block.

Anesth Analg 1980;59:225.

[Fu ll text Link]

[CrossRef]

[Med line Link]

130. Thomas SP, Bamigbade TA. Fa ilure of a SafeBi te bi te protector. Anaesthesia

2005;60:1047–1048.

[Fu ll text Link]

[CrossRef]

[Med line Link]

131. Stoneham MD. The nasopharyngeal a irway. Assessment of posi tion by

f iberoptic la ryngoscopy. Anaesthesia 1993;48:575–580.

[CrossRef]

[Med line Link]

132. Rooney R, Crummy EJ, McShane AJ. Bacteraemia fol lowing nasopharyngeal

ai rway insertion. Anaes thesia 1992;47:1099.

[CrossRef]

[Med line Link]

Page 46: Chapter 16. Face Masks & Airways

133. Stall ings JO, Lines J. Use of nasopharyngeal tubes as aids to lateral port

construc tion and maintenance of the ai rway in pharyngeal f lap surgery. Plast

Reconstr Surg 1976;58:379–380.

[Fu ll text Link]

[CrossRef]

[Med line Link]

134. Donnel ly J, Thirlwel l J . Nasopharyngeal ai rway in infants after pa latal surgery.

Br J Anaesth 1992;67:227.

[CrossRef]

[Med line Link]

135. Heaf DP, Helms PJ, Dinwiddie R, et a l. Nasopharyngeal ai rways in Pierre

Rob in syndrome. J Pedia tr 1982;100:698–703.

[CrossRef]

[Med line Link]

136. Young ML, Hanson C. An al te rnative to tracheostomy fol lowing

transsphenoidal hypophysectomy in a patien t with acromegaly and s leep apnea.

Anesth Analg 1993;76:446–449.

[Med line Link]

137. Wanner A, Zighelbo im A, Sackner MA. Nasopharyngeal a irway: a fac il i tated

access to the trachea. For nasotracheal suction , beds ide bronchof iberscopy, and

se lec tive bronchography. Ann Int Med 1971;75:593–595.

[Med line Link]

138. Lewis JD. Faci li tation of nasogastric and nasotracheal intubation with a

nasopharyngea l ai rway. Am J Emerg Med 1986;4:426.

[CrossRef]

[Med line Link]

139. Shetty S, Henthorn RW, Ganta R. A method to reduce nasopharyngeal trauma

f rom nasogas tric tube placement. Anesth Analg 1994;78:410–411.

[Med line Link]

140. Bagshaw ONT, Southee R, Ruiz K. A comparison of the nasal mask and the

nasopharyngea l ai rway in paedia tric chair dental anaesthesia. Anaesthesia

1997;52:786–796.

[Fu ll text Link]

[CrossRef]

[Med line Link]

Page 47: Chapter 16. Face Masks & Airways

141. McMi llon CD. Another use for nasopharyngeal ai rway. Anesthesiology

2000;93: 298.

[Fu ll text Link]

[CrossRef]

[Med line Link]

142. Beatt ie C. The modif ied nasal trumpet maneuver. Anesth Analg 2002;94:467–

469.

[Fu ll text Link]

[CrossRef]

[Med line Link]

143. Metz S, Beatt ie C. A modif ied nasal trumpet to fac il i ta te f ibreoptic in tubation .

Br J Anaesth 2003;90:388–391.

[Fu ll text Link]

[CrossRef]

[Med line Link]

144. Maha jan R. Simple tracheal tube—an alternat ive to the modif ied nasal

trumpet. B r J Anaes th 2003;91:759–760.

[Fu ll text Link]

[CrossRef]

[Med line Link]

145. Walz R. The modif ied nasal trumpet maneuver. Anesth Analg 2002;95:1821.

[Fu ll text Link]

[CrossRef]

[Med line Link]

146. Feldman SA, Fauvel NJ, Ooi R. The cuffed pharyngea l ai rway. Eur J Anaesth

1991;8:291–295.

[Med line Link]

147. Ralston SJ, Charters P. Cuffed nasopharyngeal tube as “dedica ted ai rway” in

diff icul t intubat ion. Anaesthesia 1994;49:133–136.

[CrossRef]

[Med line Link]

148. Boheimer NO, Feldman SA, Soni N. A se lf -retaining nasopharyngeal a irway.

Anaesthesia 1990;45:72–73.

[CrossRef]

[Med line Link]

Page 48: Chapter 16. Face Masks & Airways

149. Dominguez E. Another use for nasopharyngeal ai rway. In reply.

Anesthesiology 2000;93:298–299.

[Fu ll text Link]

[CrossRef]

[Med line Link]

150. Mobbs AP. Retained nasopharyngeal ai rway. A reply. Anaes thesia

1990;44:447.

151. Gal lagher WJ, Pearce AC, Power SJ. Assessment of a new nasopharyngea l

ai rway. Br J Anaesth 1988;60:112–115.

[CrossRef]

[Med line Link]

152. Long TMW. A traumatic nasopharyngeal intubat ion for upper airway

obstruct ion. Anaesthesia 1988;43:510–511.

[CrossRef]

[Med line Link]

153. Feldman SA, Fauvel NJ, Ooi R. The cuffed pharyngea l ai rway. Eur J Anaes th

1991;8:291–295.

[Med line Link]

154. Elam JO, Ti tel JH, Feingold A , et al . Simpl if ied airway management during

anesthesia or resusci ta tion: a binasal pharyngeal system. Anesth Analg

1969;48:307–316.

[Fu ll text Link]

[CrossRef]

[Med line Link]

155. Feingold A, Runyan NJ . Experience with the latex binasal pharyngeal ai rway.

Anesth Analg 1973;52:263–266.

[Fu ll text Link]

[CrossRef]

[Med line Link]

156. Komesaroff D. An unusual indica tion fo r the double nasopharyngeal ai rway: a

case report. Anesth Analg 1971;50:240–243.

[Fu ll text Link]

[CrossRef]

[Med line Link]

157. Weisman H, Weis TW, Elam JO. Use of double nasopharyngeal airways in

anesthesia. Anesth Analg 1969;48:356–361.

Page 49: Chapter 16. Face Masks & Airways

[Fu ll text Link]

[CrossRef]

[Med line Link]

158. Weisman H, Bauer RO, Huddy RA, et al. An improved b inasopharyngeal ai rway

system for anesthesia. Anesth Analg 1972;51:11–13.

[CrossRef]

[Med line Link]

159. Katz H, Mach ida RC, Wootton DG, et al . A technic of genera l anesthesia fo r

blepharoplasty and rhytidectomy. Anesth Analg 1976;55:165–167.

[CrossRef]

[Med line Link]

160. Katz H, Richards JA. General anesthes ia wi thout intubat ion fo r cataract

surgery. Anesthesiology Rev 1976;29–31.

161. Weisman H, Bauer RO, Elam JO. Binasopharyngeal ai rways. A rediscovered

system for anesthesia. Anesth Prog 1969;16:101–104.

[Med line Link]

162. Marsh AM, Nunn JF, Taylor SJ, et al. Airway obs truct ion associated with the

use of the Guedel airway. Br J Anaesth 1991;67:517–523.

[CrossRef]

[Med line Link]

163. Brown TCK. The a irway in mucopolysaccharidoses . Anaesth Intens Care

1992;12:178.

[Med line Link]

164. Inkster S. The fenestra ted nasopharyngeal ai rway. Anaesthesia 1997;50:567.

[CrossRef]

[Med line Link]

165. McNicol LR. Unusual cause of obs tructed ai rway in a chi ld. Anaesthesia

1986;41:668–669.

[CrossRef]

[Med line Link]

166. Walker MB. An unusual destina tion fo r ora l premedicat ion. Anaes thesia

1994;49: 839.

[Med line Link]

167. McNicol LR. Unusual cause of obs tructed ai rway in a chi ld. Anaesthesia

1986;41:668–669.

[CrossRef]

Page 50: Chapter 16. Face Masks & Airways

[Med line Link]

168. Baumann RC, MacGregor DA. Dissection of the posterior pharynx resul ting in

acute ai rway obs truct ion. Anesthesiology 1995;82:1516–1518.

[Fu ll text Link]

[CrossRef]

[Med line Link]

169. Drummond JC. Anesthes ia for posterior fossa procedures (ASA Annua l

Ref resher Courses). Park Ridge, IL: Author, 1994.

170. Haynes GR, Gramling-Babb P, Baker JD II I. Fatal outcome of postoperative

macroglossia as a compl icat ion of a c ran io tomy. Am J Anesth 1995;22:41–44.

171. Lam AM, Vav ilala MS. Macroglossia : compartment syndrome of the tongue?

Anesthesiology 2000;92:1832–1833.

172. Moore JK, Chaudhri S, Moore AP, et al. Macroglossia and pos terio r fossa

disease. Anaes thesia 1988;43:382–385.

[CrossRef]

[Med line Link]

173. Bhadri N, Rao GSU. Unilate ral fac ia l and neck swel l ing after inf ratentoria l

surgery in the late ral posi tion. Anesth Analg 1999;89:1290–1291.

[Fu ll text Link]

[CrossRef]

[Med line Link]

174. Shulman MS. Uvular edema wi thout endotracheal intubat ion. Anesthesiology

1981;55:82–83.

[Fu ll text Link]

[CrossRef]

[Med line Link]

175. Jones GL, Redfern RM, Jerwood DC. Deafness—a compl icat ion of

nasopharyngea l ai rways. Anaesth In tens Care 2002;30:807–808.

[Med line Link]

176. Gupta R. Uni late ral transien t sialadenopathy: another complicat ion of

oropharyngea l ai rway. Anesthesiology 1998;88:551–552.

[Fu ll text Link]

[CrossRef]

[Med line Link]

177. Bhavani-Shankar K. Sa livary gland enlargement caused by chemical agents.

Anesth Analg 1999;89:804.

Page 51: Chapter 16. Face Masks & Airways

[Fu ll text Link]

[CrossRef]

[Med line Link]

178. Moore MW, Rauscher L. A compl ication of oropharyngeal a irway placement.

Anesthesiology 1977;47:526.

[Fu ll text Link]

[CrossRef]

[Med line Link]

179. Muzzi DA, Losasso TJ, Cucchiara RF. Compl icat ion from a nasopharyngeal

ai rway in a pat ient wi th a basilar skul l f racture. Anesthesio logy 1991;74:366–368.

[Fu ll text Link]

[CrossRef]

[Med line Link]

180. Schade K, Borzo tta A, Michaels A. Intrac ran ial malposi t ion of nasopharyngeal

ai rway. J Trauma 2000;49:967–968.

[Fu ll text Link]

[CrossRef]

[Med line Link]

181. Pol lard BJ , O'Leary J . Guedel ai rway and tooth damage. Anaesth Intens Care

1981;9:395.

[Med line Link]

182. Clokie C, Metcalf I, Holland A . Dental trauma in anaesthesia . Can J Anaesth

1989;36:675–680.

[Med line Link]

183. Burton JF, Baker AB. Dental damage during anaes thesia and surgery. Anaes th

Intens Care 1987;15:262–268.

[Med line Link]

184. Anonymous . Nasopharyngeal airway. Biomed Safe Stand 1984;14:111–112.

185. Daly SM, Weinberg B, Murphy RJC, et al . Unrecognized aspiration of an

oropharyngea l ai rway. Pediatr Rad iol 1983;13:227–228.

[CrossRef]

[Med line Link]

186. Howat DDC. Disposable nasopharyngeal a irways—a potential hazard.

Anaesthesia 1982;37:101.

[CrossRef]

[Med line Link]

Page 52: Chapter 16. Face Masks & Airways

187. Hayes JD, Lockrem JD. Asp irat ion of a nasal ai rway. A case report and

principles of management. Anesthes iology 1985;62:534–535.

[Fu ll text Link]

[CrossRef]

[Med line Link]

188. Ho G, Weinger MB. Unusual complica tion of nasotracheal suctioning .

Anesthesiology 1993;79:630.

[Fu ll text Link]

[CrossRef]

[Med line Link]

189. Milam MG, Mil ler KS. Aspirat ion of an art if ic ia l nasopharyngeal airway. Ches t

1988;93:223–224.

[Med line Link]

190. Zal tzman J, Ferman A. An acute li fe -threatening complicat ion caused by a

Guedel ai rway. Cri t Care Med 1987;15:1074.

[CrossRef]

[Med line Link]

191. Smith BL. Retained nasopharyngeal a irway. Anaes thesia 1989;44:447.

[CrossRef]

[Med line Link]

192. Stoneham MD. Nasopharyngeal ai rways and nasotrachea l suc tion.

Anesthesiology 1994;80:480.

[Fu ll text Link]

[CrossRef]

[Med line Link]

193. Gandhi S, Dhamee MS. Detachment of an esophageal s tethoscope cuff—

possible role of an oral a irway. Anesthesiology 1983;58:202.

[Fu ll text Link]

[CrossRef]

[Med line Link]

194. Topf AI. Compl ication associated wi th the use of an oral ai rway.

Anesthesiology 1996;84:485.

[Fu ll text Link]

[CrossRef]

[Med line Link]

195. Lloyd-Will iams R. Fractured airways . Anaesth Intens Care 1985;13:335–336.

Page 53: Chapter 16. Face Masks & Airways

[Med line Link]

196. Anonymous . Universal hospital suppl ies. S izes 2 and 3 Guedel ai rways: bi te

block may d is lodge. Heal th Dev ices Alerts 2004;28:5.

197. Nandalan SP, Hiremath A. Crit ical inc ident due to a Guede l ai rway.

Anaesthesia 2004;59:832.

[Fu ll text Link]

[CrossRef]

[Med line Link]

198. Michelsen LG, Valdes-Murau H. An in tubat ing airway wi th teeth.

Anesthesiology 2001;94:938.

[Fu ll text Link]

[CrossRef]

[Med line Link]

P.460

Questions For the fol lowing quest ions, answer

• i f A, B, and C are correct

• i f A and C are correct

• i f B and D are correct

• is D is correct

• i f A, B, C, and D are correct.

1. Complications associated with use of a face mask include

A. Movement of the cerv ical spine

B. Facial nerve injury

C. Chemical i rr itation o f the eyes

D. Sinus itis

View Answer2. Which maneuvers can be used to secure a patent airway?

A. Jaw thrust

B. Posterior protrusion of the mandible

C. Chin l if t

D. Posterior f lexion of the atlan to-occipital jo int

View Answer3. Airways developed to aid in fiberoptic intubation include

Page 54: Chapter 16. Face Masks & Airways

A. Patil-Syracuse

B. Wi ll iams

C. Ovassapian

D. Guedel a irway

View Answer4. Contraindications to nasal airways include A. Hemorrhagic disorders

B. Sepsis

C. Basilar skul l f racture

D. Se izure disorders

View Answer5. Which external measurements correlate with the proper length of a nasal airway? A. Tip of nose to 2 cm above the thyroid carti lage

B. The dis tance from the tip of the thumb to the t ip of the index f inger measured to

the hand and back

C. Tip of the earlobe to cricoid carti lage

D. The patien t's height

View Answer6. Complications associated with the use of oral airways include A. Swell ing of the tongue

B. Edema of the uvula

C. Ulcerat ion of the tongue

D. Swallowing of the ai rway

View Answer7. Predictors of difficult mask ventila tion include

A. High body mass index

B. Age over 65 years

C. History of snoring

D. History of s leep apnea

View Answer8. Techniques to improve mask fit in the edentulous patient include A. Packing the cheeks wi th gauze sponges

B. Not removing the dentures

C. Insert ing the infe rior margin of the mask between the gingiva of the mandib le

and the lower l ip

D. Us ing a nasal airway

View Answer