Chapter 1 Function overview

28
Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber ESPEC CORP. | 9 Chapter 1 Function overview This chapter gives an overview of functions and explains the specifications and operation methods for the instrumentation. N-instrumentation configuration The N-instrumentation is network instrumentation equipped with controller functions (constant operation and program operation), network functions and external memory functions. PC application for external memory functions PATTERN MANAGER Lite* * Can be downloaded from Espec Test Navi, the reliability testing information website. Bench-top type temperature (and humidity) chamber Web application N-instrumentation P-200 External memory Instrumentation internal data External memory port Ethernet port (LAN port)

Transcript of Chapter 1 Function overview

Page 1: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 9

Chapter 1 Function overview This chapter gives an overview of functions and explains the specifications and operation methods for the instrumentation.

N-instrumentation configuration

The N-instrumentation is network instrumentation equipped with controller functions (constant operation and program operation), network functions and external memory functions.

PC application for external memory functions

PATTERN MANAGER Lite*

* Can be downloaded from

Espec Test Navi, the reliability

testing information website.

Bench-top type temperature

(and humidity) chamber

Web application

N-instrumentation P-200

External memory

Instrumentation internal data

External

memory port

Ethernet port

(LAN port)

Page 2: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

10 | ESPEC CORP.

P-200 features Selectable operation style

・ You can select the test operation/management style according to the environment that the chamber will be used in. Perform test preparation, setup, monitor, and analysis management with ease at any location at any time.

See the Network and Option guides.

Easier operation

・ Make use of the shortcut keys on the chamber screen. You can call up your regular, favorite operation screens with ease. (Quick access) See "1.2.1 Common operations area" in this guide.

・ Synchronize with measuring instruments without a PC. More convenient with additional functions

・ Additional instrumentation functions can be downloaded over the network. (Add-ons) See "Chapter 7 Add-ons/system updating" in this guide and the Network guide.

Support for daily management

・ ESPEC online support is available.(*) It reduces chamber management work when an error occurs and decreases down time. * You can use this service through a separate contract (paid service limited to Japan only).

◆ Reference ◆

Font used in the controller

Ricoh TrueType font designed by Ricoh Company, Ltd. is used for this product's controller.

Page 3: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 11

1.1 Instrumentation organization

The organization of the instrumentation screen is shown in Figure 1.1.

Monitor Temp/Hum Temp/Hum

Details

Program Details

Counter monitor

Ext output

Ext output

Time Signals

Meas-sync

Trend Graph

Constant Setup Temp/Hum

(No. 1 to No. 3) Hum Control

Details setting Equipment

Time Signals

Alarm Setting

Options

Program Setup NEW PGM

EDIT

COPY

DELETE

VIEW

Chamber Setup Set Timer

Set Protection

Reset Time Meter

Alarm Report

ROM Information

Reminder

Set Back Trace

Accessory

Add-ons/Updates History

Configuration

Page 4: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

12 | ESPEC CORP.

Chamber Setup Configuration Set Communication

Operation Process

Control Attain Range

Display Setup

Set Sound

Date & Screen Saver

Register User Password

Sensor Offset

Set Chamber Detail

Set Meas-sync

Set Recorder

Set Options

Set Quick Access

Service

External Memory Set Sampling/Save Data

Read/Write Program Pattern

Write Back Trace Data

Add-ons/System Updates

Figure 1.1 Instrumentation overall configuration

Add-ons are functions that can be added to the chamber after purchase. For details, see "Chapter 7 Add-ons/system updating".

Page 5: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 13

1.2 Screen configuration

The instrumentation screen consists of the following three areas.

① Common operations area ② Standard display area ③ Tab area

Common operations area

Standard display area

Tab area

Page 6: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

14 | ESPEC CORP.

1.2.1 Common operations area

Display functions and key functions are indicated as shown in the table below. Icon Function name Display function Function when key is pressed

Operation display area

Indicates the chamber operation state.

Displays the display operation screen.*1

Set Language None Displays the screen for setting the

language.

Alarm Indicates the alarm generation condition. Blinks when an alarm occurs.

Displays the ALARM screen.*1

Information Indicates the information availability condition. Blinks when a notification is generated.

Displays the INFO screen.*1

• If there are no notifications, the accessory icon is displayed, so the key function of the accessory key is performed.

Accessory None Displays the ACCESSORY screen.*1

• This operation can also be performed from the Chamber Setup screen.

Timer Indicates the timer set

condition. Displays the TIMER screen.*1

Time display area

Shows the current time. Displays the time/screen setting screen.

Quick Access A red disabled sign is

displayed when the quick access menu is disabled.

Displays the quick access screen.

*1 Performing a Show Time key operation while one of these screens is displayed will cause the message "Not available from this screen" to appear. Close the currently displayed screen and then perform the desired operation.

Page 7: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 15

Operation display area

STOP

CONSTANT For details, see "Chapter 3 Constant operation."*

PROGRAM: RUN For details, see "Chapter 4 Program operation."*

PROGRAM: PAUSE

Hold Last Program

Remote Operation For details of remote operation, see "Option guide /interface."

Remote Pause

Hold Last Remote Operation

* For details on how to use the network function to monitor, manage, configure and operate the chamber, see the Network guide.

Set Language

Set the controller's display language. Select the language, and then press [Back].

Page 8: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

16 | ESPEC CORP.

Alarm

Displays the alarm output status of the equipment.

There are two types of notifications issued by the equipment.

Alarm: This is issued when the equipment detects an abnormality in the equipment and the equipment stops operation. However, if backup operation is enabled, the equipment continues to operate as much as possible.

Warning: This is issued when the equipment detects a minor abnormality in the equipment and the equipment continues to operate. There is one type of warning that will stop operation.

If an alarm occurs, the Alarm screen is displayed automatically, the ALARM icon blinks, and a buzzer sounds.

Press [Stop Beep] on the Alarm screen to stop the resonant beep. Press the alarm name in the Alarm column to display the Alarm Help screen.

See Basic guide "Chapter 6.5 List of alarms." For details on setting the beep, see "1.7 Setting beep."

[CLEAR] key on the ALARM HELP screen If you remove the cause of the alarm while the alarm is occurring and the equipment has stopped operating, the equipment returns to normal operation automatically. The event remains in the Alarm screen's history, so press the [CLEAR] key when you want to erase the history display.

Page 9: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 17

Accessory

Press this icon to display the screen for viewing the hour meter, turning off the screen, and setting drainage. You can display the ACCESSORY screen from the Chamber Setup tab.

ACCESSORY screen

Turning off the screen and reducing brightness can be set with the timer. See "1.8 Setting date and screen."

Hour Meter: Displays the total operation time. During accumulation, the square blinks at the bottom right of the frame.

Turn OFF the screen: Press this key to turn off the screen. To cancel turning off the screen, press the screen.

Set Drain (temperature and humidity type only; when the automatic humidifying tray water change function is installed):

Set whether to enable automatic draining or whether to start/stop manual draining from the humidifying tray and wick pan. For details, see Options guide "Automatic humidifying tray water change function." * For chambers that do not have this options, drain using a drain

hose. For details see the Operation manual / Basic guide.

Notice

If the chamber operates at a high temperature when there is water in the humidifying tray, the water may evaporate, possibly resulting in high temperature in the chamber. To perform high temperature operation, connect a drain hose and drain the water.

Page 10: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

18 | ESPEC CORP.

Information

If there is an awaiting equipment information notification, the ACCESSORY icon changes to the INFO icon. Press the information icon when it is blinking to display the Information screen. If there is no longer any equipment information to be notified, the screen disappears, the icon stops blinking, and the icon changes back to ACCESSORY.

Information screen

For each information item, its explanation and action to be taken are displayed on the screen. Take appropriate action in accordance with the method described on the screen.

Information Description and Required Action

Ref in Preparation The system is preparing for refrigerator startup.

Backup operation in

progress

A backup operation was triggered by a problem in the

humidifier system.

Door open A door open condition was detected.

Check the door open condition.

Auto humidifier water

switching in progress

(Optional)

Humidifier switching is in progress so temperature and

humidity are unstable.

Check Hum Tray It's time for periodic humidifying tray inspection. Check the reminder function.

Check Wet-bulb Wick It's time for periodic wet-bulb wick inspection. Check the reminder function.

Reminder No. 1 The specified time has elapsed. Check the reminder function.

Reminder No. 2 The specified time has elapsed. Check the reminder function.

Reminder No. 3 The specified time has elapsed. Check the reminder function.

Page 11: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 19

Timer

The timer functions (start timer, end timer, and quick timer) can be configured, set, and canceled. When the timer is set, the icon color changes to yellowish green.

Set Timer screen

For the setting method, see "5.2 Setting and canceling the timer."

Page 12: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

20 | ESPEC CORP.

Time display

Displays the present time. Press this button to display the Date & Screen Saver screen.

For the setting method, See "1.8 Setting date and screen."

Quick Access

Displays the quick access menu. This screen also has a Set Quick Access key.

For the setting method, see "Chamber Setup" or "Convenient functions."

Page 13: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 21

1.2.2 Tab area

Monitor, Constant Setup, Program Setup, Chamber Setup, and External Memory can be selected from the tabs at the bottom of the screen.

Monitor

The status of the equipment can be checked. The monitorable items differ depending on the operating status.

For details, see "Chapter 3 Constant operation" and "Chapter 4 Program operation."

Constant Setup

This is the function for performing constant operation.

For details, see "Chapter 3 Constant operation."

Program Setup

This is the function for performing program operation. For programs, up to 8 patterns with up to 99 steps each can be set.

For details, see "Chapter 4 Program operation."

Chamber Setup

Used to set or check the management functions required for operation: protection setup, alarm report, version information, maintenance, back trace setup, and accessory.

For details, see "Chapter 5 Chamber setup."

External Memory

This is the function for writing the data stored in the internal memory to an external memory, for reading the data stored in an external memory into internal memory, or for adding functions (add-ons). • Set Sampling/Save Data • Read/Write Program Pattern • Set Back Trace Data Writing • Add-ons/system updating

For details, see "6.1 External memory."

Page 14: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

22 | ESPEC CORP.

1.2.3 Instrumentation power switch area

USB memory lamp

Displays the status of the external memory

For details, see "6.1 External memory."

Communication lamp

Blinks during communication

Blinks during Ethernet communication; when a communication connection is established through the RS-485, RS-232C, or GPIB (option) function; and when Web function is in use

Timer lamp

Blinks when the timer is set

Instrumentation power switch

Primary power supply off: Unlit

Primary power supply on: Lit (orange) (During electrical power receipt)

Instrumentation panel power on: Lit (green)

Screen illumination off: Blinking (green)

Page 15: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 23

1.3 Turning the power ON/OFF

1.3.1 Turning on the power

Turn on the power by performing the following procedure.

<Procedure>

1) Turn on the circuit breaker of the instrumentation.

2) The startup screen is displayed.

3) Check that the lamp of the instrumentation power switch lights (in orange), and

then press the instrumentation power switch. After the opening screen appears, the monitor information (temperature and humidity) is displayed.

<Temperature, humidity>

4) Press a tab to move to the desired screen.

Page 16: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

24 | ESPEC CORP.

Notice

Operations during startup

This chamber is equipped with flash memory as recording media for saving the sampling data and e-mail settings. One of the characteristics of flash memory is that its data may be corrupted if the power is turned off while memory is being accessed. This chamber is equipped with a recovery function that checks the flash memory during startup and restores corrupted data to the factory default values. When the recovery function is operating, the startup time (the time that the Booting screen is displayed) becomes longer. If the startup state (when the Booting screen is displayed) is maintained for 15 minutes without the display changing, contact your distributor or ESPEC.

Maintenance of screen system The controller of this chamber has a mechanism of maintenance of screen system, and the indication part is to be restarted automatically for the purpose even during test. Note that this restarting action is only of the indication part, and the chamber operation continues when restarting.

Chamber/system protection

Do not repeatedly turn the power supply circuit breaker on and immediately turn it off. Also, do not turn the power supply circuit breaker off during startup (when the Booting screen is displayed). Before turning the power off, check that "Booting" and "ALARM" are not displayed on the controller screen for at least 20 seconds after the power supply circuit breaker is turned on.

Page 17: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 25

◆ Note ◆

Data protection If the circuit breaker is repeatedly turned ON and OFF in short intervals, the internal memory (recording media) may no longer be recognized or the recorded data may be corrupted. Please periodically write the recorded sampling data to external memory. For details on the writing procedure, refer to "6.2 Setting sampling/writing" in Chapter 6 of the Controller guide.

Alarm display when booting If one of the following messages is displayed on the controller screen when the chamber is booting and if the keys related to saving data and recording cannot be used after booting, check the alarm report.

Screen message or screen status Details and action

"Chamber stopped because of faulty display device connection"

⇒ Initial communication error. Please contact a service representative.

"The recording area for the recorded data has

been corrupted.

The recording area was initialized."

⇒ A format is being performed. Subsequently the booting screen is displayed. Do not turn OFF the circuit breaker until the monitor screen is displayed. The internal memory may deteriorate or there may be a problem with the recording area. After the chamber boots, check the alarm report from the Chamber Setup screen. If there is a record of the "Recorded data deletion alarm" occurring, you can operate the chamber, but a portion of the recorded data may have been deleted.

"The recording area for the recorded data has

been corrupted.

Recovering and checking the recorded data"

⇒ A file check is being performed. Subsequently the booting screen is displayed. Do not turn OFF the circuit breaker until the monitor screen is displayed. The internal memory may deteriorate or there may be a problem with the recording area. After the chamber boots, check the alarm report from the Chamber Setup screen. If there is a record of the "Recorded data deletion alarm" occurring, you can operate the chamber, but a portion of the recorded data may have been deleted.

Continued on the next page.

Page 18: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

26 | ESPEC CORP.

◆ Note ◆

Continued from the previous page.

Screen message or screen status Details and action <Controller screen operations> • Set/Write Sampling Data screen - The [Rec] key to save the data to internal memory cannot be pressed • Write Back Trace Data screen - Back trace data cannot be recorded - The manual trigger [Exec] key cannot be pressed - Cannot write to external memory • Add-ons/System Updates screen - The [Update] key cannot be pressed <Web screen operations> • Back trace data cannot be written to external memory: The [Save] key cannot be pressed • Back trace data cannot be downloaded

⇒ After the chamber boots, check the alarm report from the Chamber Setup screen. If there is a record of the "Recording media recognition alarm" occurring, you can operate the chamber, but there is a risk the save/write/download functions for recorded data cannot be used normally.

Reference

When the circuit breaker of the equipment is turned on, a startup screen will appear. Then, the operation lamps, other lamps, and LEDs on switches will turn on (for a few seconds) and off one by one from the bottom, and finally all LEDs will turn on.

If the quick access function is enabled, the quick access screen will appear after startup.

1.3.2 Turning off the power

<Procedure>

1) Press the instrumentation power switch to display the execution confirmation screen. Press [YES].

Notice

Note that pressing the instrumentation power switch during operation and selecting [YES] will suspend the test and turn off the power of the instrumentation.

Page 19: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 27

1.4 Touch panel keys

This section explains major keys displayed on the touch panel.

Reference

Handling the touch panel with a pointed or hard substance may damage the touch panel and cause malfunctions. Be sure to use your fingers to operate the touch panel.

Do not touch two or more places at the same time. Doing so may cause malfunctions.

1.4.1 Keys used for moving to the next process

[YES] Press this key to execute an operation. Be sure to check the message on the screen before pressing this key.

[NO] Press this key to cancel an operation. The program cancels the operation and goes back to the previous operation.

[SAVE] Press this key to update the data. The confirmation screen is displayed.

[QUIT] Press this key to cancel a data update. The confirmation screen is displayed.

[CLOSE] Press this key to close the screen that is open.

1.4.2 Keys to be selected from multiple items

Press the key that corresponds to the item you want to select from among multiple setup items.

Example: [ON] [OFF] The set item is highlighted in yellow.

Page 20: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

28 | ESPEC CORP.

1.4.3 Ten (numeric input) key

Ten-key pop-up window

CLR: Clears all characters in the character string window. DEL: Deletes the last character entered. ENT: Sets the entered characters and closes the ten-key window. QUIT: Cancels the entered characters and closes the ten-key window.

1.5 Setting the language Select the language to be displayed. Items are displayed in the language that you select. You can set the language to Japanese, English, Chinese (simplified), Chinese (traditional), or Korean.

Page 21: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 29

1.6 Setting screens

Switches the display method used when the Monitor or Program Setup tab and the process value display (temperature/humidity type) that is used when humidity control is OFF.

<Procedure>

1) Press [Configuration] on the Chamber Setup tab.

Page 22: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

30 | ESPEC CORP.

2) Press [Display Setup] on the Configuration screen.

Reference

For details on setting the quick access function, see "Chamber setup" or "Convenient functions."

Page 23: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 31

Tab function (display hold function)

This is a function for holding the last display selected on the Monitor screen. The default is ON.

Example of screen transition when [OFF] is selected

When the screen is changed from the Constant Setup screen to the Monitor screen, [Temp/Hum] is displayed.

Example of screen transition when [ON] is selected:

When the screen is changed from the Constant Setup screen to the Monitor screen, the last screen displayed on the Monitor screen is displayed. For example, if [Details] on the Monitor screen was displayed before the transition, the screen returns to [Details] on the Monitor screen.

Process value display when humidity control is OFF

Set the humidity process value display that is used when you operate the equipment with humidity control OFF. When [YES] is selected, the humidity process value is displayed regardless of the humidity control setup. When [NO] is selected, the humidity process value is displayed as "---." when humidity control is off.

Page 24: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

32 | ESPEC CORP.

1.7 Setting beep Set the beep volume in the event of an alarm or warning, as well as ON/OFF of key touch sound. Press [Set Sound] on the Configuration screen.

Table 1.1 Beep setting items

Alarm Beep Set ON/OFF of beep in the event of the occurrence of an alarm.

Warning Beep Set ON/OFF of beep in event of the occurrence of a warning.

Key Touch Set ON/OFF of key touch sound.

Volume Set the beep volume. * The default is the maximum value.

CAUTION

We strongly advise you to keep the beep sound on whenever possible to prevent any delay in the detection of an alarm or warning.

If the beep is turned off, alarms and warnings are not indicated with a sound and are indicated only with the red, blinking operation lamp and on the warning occurrence screen.

Set the beep volume in accordance to the ambient environment. Lowering the beep volume affects the key touch sound and the beep volume for indicating alarms and warnings.

Page 25: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 33

1.8 Setting date and screen Set the date and the screen saver and luminance timer. Press [Date & Screen Saver] on the Configuration screen. The Date & Screen Saver screen can also be displayed by pressing the Time Display section in the common operations area. If a password has been registered, enter the password.

Set the date. Set the year, month, day, hours, minutes, and seconds. After you finish setting the date, press [Apply] to enter the setting.

Table 1.2 Setting the date

Input range

Year 2010 to 2037

Month 1 to 12

Day 1 to 31

Hour 00 to 23

Minute 00 to 59

Second 00 to 59

Page 26: Chapter 1 Function overview

Bench-top type temperature (and humidity) chamber Controller guide Chapter 1 Function overview

34 | ESPEC CORP.

Reference

The time accuracy is within 60 seconds per one month period. Because the clock used here is built into the instrumentation, we recommend that you periodically adjust the time.

The input range differs depending on the months entered. Be sure to enter the month first, followed by the date. Example: If March 31 is set, and then the month is changed to February, the date is

changed automatically to March 3, because February does not have 31 days. The date and time cannot be changed in the following cases.

・ Program operation is being executed. (including program pause and last program holding)

・ The timer is set ・ Sampling recording ・ External memory recording ・ Operation Mode screen is displayed. ・ Alarm is displayed. ・ Accessory or information screen is displayed.

Use this procedure to set the clock time zone. Note that after changing this setting, you will need to turn the chamber circuit breaker off and then back on again. Select the time zone to change to from standard time (UTC) from the pull-down menu under "Time zone" and click [Apply]. Turn the chamber circuit breaker off and then back on again.

Page 27: Chapter 1 Function overview

Controller guide Chapter 1 Function overview Bench-top type temperature (and humidity) chamber

ESPEC CORP. | 35

Setting Screen Saver ON/OFF. If key operation is not performed for a certain period of time, the screen is turned off to reduce the power consumption of the back light. Setting time: 1 to 59 minutes Setting Luminance Timer ON/OFF. If key operation is not performed for a certain period of time, the brightness of the back light is lowered. Setting time: 1 to 59 minutes

Reference

To turn off the screen immediately, press the ACCESSORY icon and then press [Turn OFF the screen]. * If the INFO icon is blinking, press the INFO icon before pressing [ACCESSORY] or press

the Chamber Setup tab. If you want to temporarily turn the screen back on or temporarily cancel the brightness reduction, press the instrumentation power switch or touch the screen.