Challenges in medical additive manufacturing
Maureen Antonia Johanna Maria van Eijnatten
The studies presented in this thesis were conducted at the Department of Oral and Maxillofacial Surgery / Oral Pathology and the 3D Innovation Lab of the VU University Medical Center, Amsterdam, The Netherlands.
The printing of this thesis has been financially supported by the department of Oral and Maxillofacial Surgery / Oral Pathology of the VU University Medical Center and the VU University Amsterdam.
Cover, layout & printing: Off Page, Amsterdam
ISBN: 978-94-6182-846-0
© Maureen van Eijnatten, 2017
VRIJE UNIVERSITEIT
Challenges in medical additive manufacturing
ACADEMISCH PROEFSCHRIFT
ter verkrijging van de graad Doctor aan
de Vrije Universiteit Amsterdam,
op gezag van de rector magnificus
prof.dr. V. Subramaniam,
in het openbaar te verdedigen
ten overstaan van de promotiecommissie
van de Faculteit der Geneeskunde
op vrijdag 8 december 2017 om 9.45 uur
in de aula van de universiteit,
De Boelelaan 1105
door
Maureen Antonia Johanna Maria van Eijnatten
geboren te Zwolle
promotor: prof.dr. T. Forouzanfar
copromotor: dr. J.E.H. Wolff
TABLE OF CONTENTS
Chapter 1 General introduction 7
Chapter 2 A Novel Method of Orbital Floor Reconstruction Using 17 Virtual Planning, 3-Dimensional Printing, and Autologous Bone
Chapter 3 Influence of CT parameters on STL model accuracy 27
Chapter 4 Impact of prone, supine and oblique patient positioning on CBCT 43 image quality, contrast to noise ratio and figure of merit value in the maxillofacial region
Chapter 5 CT image segmentation methods for bone used in medical 61 additive manufacturing: A Literature Review
Chapter 6 The impact of manual threshold selection in 87 medical additive manufacturing
Chapter 7 Accuracy of MDCT and CBCT in three-dimensional 103 evaluation of the upper airway morphology
Chapter 8 The accuracy of ultrashort echo time MRI sequences for 117 medical additive manufacturing
Chapter 9 General discussion 133
Chapter 10 English summary & Dutch summary 141
Chapter 11 Acknowledgements 153
1 GENERAL INTRODUCTION
GE
NE
RA
L INTR
OD
UC
TION
9
1In the early 1980s, a revolutionary layer-by-layer manufacturing technique was developed
to create industrial prototype parts from three-dimensional (3D) computational data. This
novel approach was termed “rapid prototyping” (RP) and by the general public at large
as “3D printing”. The first commercial photopolymer 3D printer was patented by Charles
W. Hull in 1986 [1]. Since then, a plethora of different 3D printing technologies have
been developed that include selective laser sintering (SLS), fused deposition modelling
(FDM) and inkjet-based technologies [2]. Overtime, the term “rapid prototyping” has
been replaced by “additive manufacturing” (AM), a term that currently encompasses
a wide range of different technologies and applications [3]. AM technologies are currently
being used in many fields, including the automotive, aerospace and consumer goods
(customisation) industries and dental and medical science. According to the Wohlers
Report 2017, the AM industry grew by 17.4% in 2016 with a global market share of $6.1
billion. Furthermore, total sales volume is expected to reach $25.6 billion in 2021 [4].
The major advantages of AM technologies over traditional subtractive manufacturing
technologies are the possibilities to customise and create almost any complex geometry
in a completely automated manner [5]. Another advantage is the speed with which
a complex geometrical prototype or functional part can be manufactured when compared
to conventional subtractive manufacturing technologies.
Over the past decades, advances in image processing and graphical processing power
have extended the role of medical imaging far beyond traditional, two-dimensional
visualisation. Novel image processing techniques can convert a stack of two-dimensional
(2D) computed tomography (CT) or magnetic resonance (MR) images into a virtual 3D
model. Such 3D models can be used as a blueprint to manufacture medical constructs
using AM. Expiring patents and price cuts in medical 3D imaging and AM technologies
have sparked an increase in the implementation of medical AM constructs in clinical
settings [6].
Currently, there is a paradigm shift in medicine away from mass production towards
iterations and from one-size-fits-all to personalised treatments. In this context, AM constructs
have proven to be particularly valuable in the field of oral and maxillofacial surgery due
to the plethora of complex bony geometries found in the skull area. Moreover, only minor
dislocations in the facial area can have a profound impact on the function and aesthetic
appearance of patients. In such cases, patient-specific AM anatomical models have been
shown to improve communication between clinicians and patients [7]. In addition, AM
models are being used for the preoperative planning and simulation of complex surgery in
order to significantly improve the efficiency and outcome of the surgery [8]. Maxillofacial
surgeons are also routinely using patient-specific AM drill guides [9] and saw guides [10]
to translate a surgical plan into the operating room. To date, AM has been sporadically
used to create patient-specific implants of the mandible [11], orbita [12]–[14], zygoma [15],
cranium [16], tibia [17], and, although still in its infancy, scaffolds for tissue engineering
[18]. Lastly, AM models of (rare) anatomical abnormalities can, in certain circumstances,
GE
NE
RA
L INTR
OD
UC
TION
10
1replace cadaveric materials in medical education and training [19], [20], thereby avoiding
ethical controversies and reducing costs [21].
The current medical AM process comprises three different steps (Figure 1) [22]. The first
step is 3D image acquisition using either CT or MRI technologies. In clinical settings,
however, CT images are currently most often used due to their superior hard tissue
contrast, spatial resolution and lower cost. The resulting imaging data is subsequently
reconstructed in two-dimensional (2D) grey scale images that are saved in Digital Imaging
and Communications in Medicine (DICOM) file formats. The second step is image
processing, in which the tissue of interest (e.g., “bone”) is segmented and converted into
a virtual 3D surface model by means of triangulation [23]. The resulting 3D surface model
is saved as a Standard Tessellation Language (STL) model, which can then be used to
design patient-specific constructs using dedicated computer-aided design (CAD) software.
The third and final step in the medical AM process is the conversion of the designated STL
model into a G-code that subsequently controls the path of the printing nozzle during AM.
One major challenge faced in the current medical AM workflow (Figure 1) is the lack
of cohesive convergence between the medical imaging, computing and manufacturing
technologies. This lack of convergence is mainly due to the fact that no single company
provides all the necessary devices and services required for the medical AM process.
The large variety of different CT/MRI scanners, software packages, 3D printing techniques
and materials available on the market makes it difficult to select appropriate devices and
software for medical AM. Moreover, such devices and software packages are often not
certified for medical AM. As a result, concerns have subsequently been raised amongst
policymakers and regulators responsible for product safety. Manufacturers of imaging
devices, 3D printers and reverse engineering software companies should therefore
agree on technical norms for the medical AM process. This is particularly important in
the light of the new European Union medical device regulation (MDR) [24] that will require
standardisation and CE certification for the whole medical AM process from 2020 onwards.
With regard to the first step in the AM workflow (Figure 1, step 1), there is general
agreement amongst physicians and medical technicians that fan-beam CT technologies,
Figure 1. The three different steps required to create a patient-specific AM construct.
GE
NE
RA
L INTR
OD
UC
TION
11
1such as multi-detector row computed tomography (MDCT), offer the best images for
medical AM [25]. However, it still remains unclear which CT technology offers the most
accurate images for medical AM purposes. Moreover, to the best of my knowledge,
no standardised scan protocols have been specifically developed for medical AM. [26]
Currently, manufacturers are exploring new ways to improve CT image quality by using
multiple photon energies (dual-energy CT) to enhance tissue characterisation. Furthermore,
as opposed to conventional CT scanners, cone-beam computed tomography (CBCT)
scanners are currently gaining popularity, especially in small dental and maxillofacial
clinics. This surge in popularity is mainly due to the markedley lower puchsing costs,
acceptable image quality and low radiation dose when compared with conventional CT
scanners [27]. Recent studies have shown, however, that the partial CBCT gantry rotation
produces a non-uniform dose distribution [28] that can in certain cases affect CBCT image
quality. Therefore, CBCT scanning technology should not be routinely used to fabricate
patient-specific implants. Interestingly, new research has shown that CBCT gantry and
head positioning may have a positive impact on image quality in the maxillofacial region,
which in turn may improve the overall accuracy of CBCT-derived AM constructs.
In the second step of the AM workflow (Figure 1, step 2), the major challenge is
the conversion of grey scale voxel images (DICOM) into a triangulated 3D surface model
(STL). This step is necessary since all CAD software packages approved by the United
States Food and Drug Administration (FDA) require STL file formats for anatomical
modelling and reverse engineering. The DICOM to STL conversion still requires multiple
manual processing steps that are dependent on the spatial abilities and knowledge of
anatomy of the technicians or physicians. The most important step in the DICOM to STL
conversion process is image segmentation, which refers to the partitioning of images into
regions of interest (ROIs) that correspond to specific anatomical structures. A wide range
of different image segmentation methods have been developed over the past decades
that include global thresholding, edge detection, region growing, statistical shape model
(SSM)- and atlas-based methods, morphological snakes, active contouring, random forests
and artificial neural networks [29]. Of these methods, the most commonly used image
segmentation method to date in medical AM is global thresholding. It should be noted
that most software packages only offer a single, default threshold value for compact bone,
soft tissue and cartilage. In addition, these default values are often not optimized for all
types of MDCT, DECT and CBCT images and do not take into account the variations in
grey values between the different scanners. Therefore, in most cases, manual threshold
selection is necessary to acquire an optimal STL model. Threshold selection still remains,
however, a subjective task that impedes the reproducibility and reliability of medical AM
constructs. Moreover, global thresholding often causes data loss in the STL model. Such
data loss can cause voids in the STL models that can subsequently lead to inaccuracies in
3D printed implants and lead to complications during surgery [30].
The aforementioned image segmentation process has been identified as one of
the major causes of inaccuracies in medical AM constructs [31]. However, adequate
GE
NE
RA
L INTR
OD
UC
TION
12
1methods to assess the accuracy and reliability of medical AM constructs are still lacking.
More specifically, novel methods are still being sought to validate the medical imaging
and image processing steps in the medical AM workflow (Figure 1, step 1 and 2). One
method of validating imaging devices and software packages is to use phantoms that
provide ground truth measurements in clinical settings. Such phantoms are commonly
composed of materials with realistic tissue (radio)densities. However, a major drawback
with such phantoms is that they are manufactured in generic forms and sizes that do not
resemble real patients. Thus, it is difficult to extrapolate the performance of an imaging
system in phantoms to humans. Fortunately, 3D printing offers new possibilities to
fabricate anthropomorphic phantoms that can be used to assess the accuracy of different
CT scanners and software packages.
Currently, the number of patients treated with medical AM constructs is rapidly increasing.
Accordingly, the number of CT and CBCT scans performed worldwide have subsequently
increased. In this context, it should be noted that all X-ray based imaging modalities
induce harmful ionizing radiation to the patient [32]. According to the ALARA (as low as
reasonably achievable) principle, the cumulative lifetime radiation dose of patients should
be minimised by reducing the overall number of examinations and by lowering the dose
resulting from each individual examination [33]. Therefore, every CT scan that is used for
medical AM needs foremost to be suitable for diagnostic purposes. Furthermore, novel
ionizing radiation-free MRI modalities are being developed to image the musculoskeletal
system [34]. MRI provides a noninvasive visualization of water microenvironments and
other sources of protons in the human body, which provides excellent imaging possibilities
for soft tissues. Cortical bone, however, has very short relaxation times [35], leading to fast
decay of the MRI signal. As a result, cortical bone commonly appears as a signal void on
MRI images obtained with routine clinical pulse sequences such as spin echo and gradient
echo. With ultra-short echo time (UTE) sequences, an MR signal can be detected directly
from bone [36], [37]. Therefore, UTE MRI sequences may be suitable to generate images
for medical AM purposes. However, the accuracy of UTE-derived STL models still needs to
be thoroughly evaluated. In addition, MRI remains an expensive imaging modality when
compared to CT.
In conclusion, AM offers possibilities for fabricating different types of patient-specific
constructs such as anatomical models, surgical guides and implants. However, despite
recent advances and promising case studies involving medical AM, notable scientific,
technological and regulatory challenges remain.
GENERAL AIM AND OUTLINE OF THE THESISThe aim of this thesis is to provide an insight into the current scientific and technological
challenges faced in medical additive manufacturing with respect to imaging and image
processing. More specifically, the different parameters that influence the accuracy of
patient-specific AM constructs will be identified.
GE
NE
RA
L INTR
OD
UC
TION
13
1REFERENCES1. Hull C W (1986) Apparatus for production
of three-dimensional objects by stereolithography. Google Patents, 1986
2. Campbell I, Bourell D, and Gibson I (2012) Additive manufacturing: rapid prototyping comes of age. Rapid Prototyp. J. 18:255–258
3. Gibson I, Rosen D, and Stucker B (2014) Additive manufacturing technologies - 3D Printing, Rapid Prototyping, and Direct Digital Manufacturing. Springer, New York
4. Wohlers Associates (2017) 3D Printing and Additive Manufacturing State of the Industry, Annual Worldwide Progress Report. Fort Collins, CO
5. Yan Y, Li S, Zhang R, Lin F, Wu R, Lu Q, Xiong Z, and Wang X (2009) Rapid Prototyping and Manufacturing Technology: Principle, Representative Technics, Applications, and Development Trends. Tsinghua Sci. Technol. 14, Supple:1–12
6. Ventola C L (2014) Medical Applications for 3D Printing : Current and Projected Uses. Pharm. Ther. 39:704–711
7. D’Urso P S, Barker T M, Earwaker W J, Bruce L J, Atkinson R L, Lanigan M W, Arvier J F, and Effeney D J (1999) Stereolithographic biomodelling in cranio-maxillofacial surgery: a prospective trial. J. Cranio-Maxillofacial Surg. 27:30–37
8. Paiva W S, Amorim R, Bezerra D A F, and Masini M (2007) Aplication of the stereolithography technique in complex spine surgery. Arq. Neuropsiquiatr. 65:443–445
9. Dawood A, Marti B M, Sauret-Jackson V, and Darwood A (2015) 3D printing in dentistry. Br Dent J 219:521–529
10. Abou-El Fetouh A, Barakat A, and Abdel-Ghany K (2011) Computer-guided rapid-prototyped templates for segmental
mandibular osteotomies: a preliminary report. Int. J. Med. Robot. 7:187–192
11. Cohen A, Laviv A, Berman P, Nashef R, and Abu-Tair J (2009) Mandibular reconstruction using stereolithographic 3-dimensional printing modeling technology. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 108:661–6
12. Fan X, Zhou H, Lin M, Fu Y, and Li J (2007) Late reconstruction of the complex orbital fractures with computer-aided design and computer-aided manufacturing technique. J. Craniofac. Surg. 18:665–673
13. Wolff J, Sándor G K B, Pyysalo M, Miettinen A, Koivumäki A-V, and Kainulainen V T (2013) Late reconstruction of orbital and naso-orbital deformities. Oral Maxillofac. Surg. Clin. North Am. 25:683–95
14. Stoor P, Suomalainen A, Lindqvist C, Mesimäki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific guiding implants for orbital wall defects. J. Cranio-Maxillofacial Surg. 42:1644–1649
15. Rotaru H, Schumacher R, Kim S-G, and Dinu C (2015) Selective laser melted titanium implants: a new technique for the reconstruction of extensive zygomatic complex defects. Maxillofac. Plast. Reconstr. Surg. 37:1
16. Jardini A L, Larosa M A, Maciel Filho R, Zavaglia C A D C, Bernardes L F, Lambert C S, Calderoni D R, and Kharmandayan P (2014) Cranial reconstruction: 3D biomodel and custom-built implant created using additive manufacturing. J. Cranio-Maxillofacial Surg. 42:1877–1884
17. Albano T, Grabowsky M B, Rodriguez L, Lavelle W, Uhl R, Ledet E, and Sanders G (2011) Designing patient-specific orthopaedic mesh implants to treat high-energy tibial fractures. Bioengineering Conference (NEBEC), 2011 IEEE 37th Annual Northeast
GE
NE
RA
L INTR
OD
UC
TION
14
118. Visscher D O, Farré-Guasch E, Helder M
N, Gibbs S, Forouzanfar T, van Zuijlen P P, and Wolff J (2016) Advances in Bioprinting Technologies for Craniofacial Reconstruction. Trends Biotechnol. 34:700-710
19. Bakhos D, Velut S, Robier A, Al Zahrani M, and Lescanne E (2010) Three-Dimensional Modeling of the Temporal Bone for Surgical Training. Otol. Neurotol. 31:328–334
20. Bruyère F, Leroux C, Brunereau L, and Lermusiaux P (2008) Rapid prototyping model for percutaneous nephrolithotomy training. J. Endourol. 22:91–6
21. McMenamin P G, Quayle M R, McHenry C R, and Adams J W (2014) The production of anatomical teaching resources using three-dimensional (3D) printing technology. Anat. Sci. Educ. 7:479–486
22. Bibb R, Eggbeer D, and Paterson A (2015) Medical modelling: The application of advanced design and rapid prototyping techniques in medicine. Woodhead Publishing, Cambridge, UK
23. Lorensen W E and Cline H E (1987) Marching cubes: A high resolution 3D surface construction algorithm. ACM siggraph computer graphics. 21:163–169
24. EU (2017) Revisions of medical device directives - European Commission (2017). Available: https://ec.europa.eu/growth/sectors/medical-devices/regulatory- f ramework/revis ion_en. [Accessed: 30-May-2017]
25. Liang X, Lambrichts I, Sun Y, Denis K, Hassan B, Li L, Pauwels R, and Jacobs R (2010) A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT). Part II: On 3D model accuracy. Eur. J. Radiol. 75:270–274
26. Mallepree T and Bergers D (2009) Accuracy of medical RP models. Rapid Prototyp. J. 15:325–332
27. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs
R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
28. Fahrig R, Dixon R, Payne T, Morin R L, Ganguly A, and Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med. Phys. 33:4541–4550
29. Sharma N and Aggarwal L M (2010) Automated medical image segmentation techniques. J. Med. Phys. 35:3–14
30. Stoor P, Suomalainen A, Lindqvist C, Mesimaki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific implants for reconstruction of orbital wall defects. J. Craniomaxillofac. Surg. 42:1644–1649
31. Smith E J, Anstey J A, Venne G, and Ellis R E (2013) Using additive manufacturing in accuracy evaluation of reconstructions from computed tomography. Proc Inst Mech Eng H 227:551–559
32. Koivisto J (2015) The use of MOSFET dosimeters and anthropomorphic phantoms in low dose dental CBCT applications. Helsinki, Finland: University of Helsinki
33. Implications of Commission Recommendations that Doses be kept as Low as Readily Achievable (1973) ICRP publication 22, Oxford, UK
34. Delso G, Carl M, Wiesinger F, Sacolick L, Porto M, Hüllner M, Boss A, and Veit-Haibach P (2014) Anatomic Evaluation of 3-Dimensional Ultrashort-Echo-Time Bone Maps for PET/MR Attenuation Correction. J. Nucl. Med. 55:780–785
35. Horch R A, Nyman J S, Gochberg D F, Dortch R D, and Does M D (2010) Characterization of 1H NMR signal in human cortical bone for magnetic resonance imaging. Magn. Reson. Med. 64:680–687
36. Keereman V, Fierens Y, Broux T, De Deene Y, Lonneux M, and Vandenberghe S (2010) MRI-based attenuation correction
GE
NE
RA
L INTR
OD
UC
TION
15
1for PET/MRI using ultrashort echo time sequences. J. Nucl. Med. 51:812–8
37. Aitken A P, Giese D, Tsoumpas C, Schleyer P, Kozerke S, Prieto C, and
Schaeffter T (2014) Improved UTE-based
attenuation correction for cranial
PET-MR using dynamic magnetic field
monitoring. Med. Phys. 41:12302
2 A NOVEL METHOD OF ORBITAL FLOOR RECONSTRUCTION
USING VIRTUAL PLANNING, 3-DIMENSIONAL PRINTING, AND AUTOLOGOUS BONE
Maarten Vehmeijer, Maureen van Eijnatten,
Niels Liberton, Jan Wolff
Journal of Oral and Maxillofacial Surgery (2016). 74 (8): 1608-1612doi: 10.1016/j.joms.2016.03.044
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
18
2
ABSTRACTFractures of the orbital floor are often a result of traffic accidents or interpersonal violence.
To date, numerous materials and methods have been used to reconstruct the orbital
floor. However, simple and cost-effective 3-dimensional (3D) printing technologies for
the treatment of orbital floor fractures are still sought. This study describes a simple,
precise, cost-effective method of treating orbital fractures using 3D printing technologies
in combination with autologous bone. Enophthalmos and diplopia developed in a 64-
year-old female patient with an orbital floor fracture. A virtual 3D model of the fracture
site was generated from computed tomography images of the patient. The fracture was
virtually closed using spline interpolation. Furthermore, a virtual individualized mould of
the defect site was created, which was manufactured using an inkjet printer. The tangible
mould was subsequently used during surgery to sculpture an individualized autologous
orbital floor implant. Virtual reconstruction of the orbital floor and the resulting mould
enhanced the overall accuracy and efficiency of the surgical procedure. The sculptured
autologous orbital floor implant showed an excellent fit in vivo. The combination of virtual
planning and 3D printing offers an accurate and cost-effective treatment method for orbital
floor fractures.
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
19
2
INTRODUCTIONBlow-out fractures are often a result of motor vehicle accidents, interpersonal violence, or
sports injuries. Such events can induce a blunt impact to the orbital area, which increases
the intraorbital pressure and subsequently leads to a fracture of one or more orbital walls.
This often results in a prolapse of the orbital content and, hence, displacement of the eye
globe into the maxillary sinus cavity, which can lead to a enophthalmos (infra-position of
the globe) and diplopia (double vision) of the eye.
To date, a wide range of different reconstruction materials have been used for
the treatment of orbital floor fractures. Nevertheless, over the past few decades, autogenous
bone grafts have been considered the gold standard [1]. However, a major drawback of
autologous bone compared with prefabricated titanium meshes is that autologous bone
is cumbersome to sculpture and can easily break if bent beyond its capacity. Furthermore,
bone grafts can in some cases resorb unpredictably. However, orbital graft resorption can
be minimized by proper fixation of the graft to the orbital floor [2].
Recent advancements in head imaging and 3-dimensional (3D) printing have opened
a wide range of new treatment possibilities [3]. Such technologies allow the creation of
patient-specific medical constructs. Currently, medical 3D printing is most commonly
used to manufacture medical models, drill guides, saw guides and individualized titanium
reconstruction plates [4], [5].
To date, biocompatible 3D printable materials are still being sought. Therefore, we
suggest combining currently available 3D printing technologies with autologous bone
grafts for the reconstruction of the orbital floor. The aim of this study was to develop
a simple and cost-effective method to produce individualized orbital implants using
computed tomography (CT) imaging and virtual planning, 3D printed moulds, and
autologous bone grafts.
MATERIALS AND METHODSA swollen and painful left eye developed in a 64-year-old female patient after a bicycle
accident. Clinical examination showed a left-sided periorbital hematoma, subconjuntival
ecchymosis, mild enophtalmus, and diplopia. Furthermore, the patient’s left eye movement
was restricted in the vertical plane. CT images showed a blowout fracture of the left
orbit (Figure 1).
The CT images were acquired using a GE Discovery CT750 HD Multi-Slice CT
scanner (GE Healthcare, Waukesha, WI). The following imaging parameters were used:
tube voltage, 120 kV; tube current, 300 mA; and reconstruction kernel, soft. The slices
were 0.6 mm thick, with 0.312-mm gaps between them. The resulting Digital Imaging
and Communications in Medicine (DICOM) files were imported into OsiriX MD imaging
software (Pixmeo, Geneva, Switzerland), and a virtual 3D model of the patient’s skull was
created using a threshold value of 120 Hounsfield units.
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
20
2
The virtual 3D model was used to manufacture a 3D printed mould of the virtually
reconstructed orbital floor. The virtual reconstruction was performed as follows:
The boundary of the orbital floor defect was delineated in the virtual 3D model using
3-matic Medical 9.0 software (Materialise, Leuven, Belgium). In a second step, 7 spline
interpolation curves were virtually drawn through the defect site to reconstruct the missing
orbital floor bone (Figure 2). In a third step, a rectangular box (60 x 60 x 30 mm) was
created around the reconstructed virtual 3D model of the orbital floor. The rectangular
box was necessary to 3D print a stable tangible mould of the orbital floor (Figure 3),
allowing the surgeon to apply force whilst sculpturing the autologous bone graft.
The aforementioned virtual reconstruction of the orbital floor took approximately 45
minutes. After successful virtual reconstruction, one tangible mould was printed using
a ZCorp inkjet printer (3D Systems, Rock Hill, SC). The total printing time required to
fabricate the tangible mould was 84 minutes.
Figure 1. Computed tomography image (coronal view) showing a blowout fracture of the left orbit with protruding soft tissue into the maxillary sinus (arrow).
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
21
2
During surgery, bone was harvested from the anterior iliac crest. The resulting bone
graft was thinned down to 1 mm in the outer area and to 2 to 3 mm in the central area.
The previously printed mould of the orbital floor was wrapped in a sterile plastic bag and
used as a template to mould and sculpture the graft into an individualized autologous
orbital floor implant (Figure 3).
The individualized implant was meticulously placed into the orbital floor and
fixated with two 3-hole miniplates (1.0 mm) and four 3-mm centre drive screws (KLS
Martin, Tuttlingen, Germany) to the inferior orbital rim. The surgical procedure took 40
minutes. Postoperative CT images revealed a harmonious reconstruction of the orbital
floor (Figure 4).
Figure 2. Virtual 3-dimensional model of the patient’s skull (grey), including the virtual mould of the orbital floor (A), the virtually reconstructed bone defect (B), and the spline interpolation curves (C).
Figure 3. Autologous bone graft (ie, implant) sculptured and moulded on a printed mould (A) with a reconstructed orbital floor (B).
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
22
2
RESULTS AND DISCUSSIONA wide variety of materials have been described for the treatment of orbital floor fractures,
ranging from polymers to metals [6]. In the past decade, individualized, 3D printed
titanium orbital implants have been used in maxillofacial surgery. However, such implants
are difficult to fabricate, are expensive, and sometimes lack dimensional precision [7].
Furthermore, they are difficult to adapt or sculpture during surgery because of their
mechanical properties. Medical titanium alloys typically have an elastic modulus (E-
modulus or Young’s modulus) of 120 GPa [8], whilst the E-modulus of orbital bone ranges
between 1.26 and 4.55 GPa [9]. Such differences can cause irritation and resorption of
the bone underneath the titanium implant [10]. The major advantage of the moulded
autologous bone implant used in this study is its biocompatibility, which subsequently
causes less irritation to surrounding tissues.
Historically, mirroring of the contralateral healthy orbit has been commonly used in
virtual planning and subsequent printing of orbital implants. However, this technique
has certain drawbacks. The major drawback is that the human face is often asymmetric
and therefore geometric variability between the two orbits is possible. Therefore,
Figure 4. Postoperative computed tomography image of the autologous orbital floor implant in situ.
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
23
2
we recommend using spline interpolation that offers the possibility of closing orbital
defects using the surrounding healthy bony structures as a reference. However, creating
an accurate spline-based mould requires technical and computational knowledge that
may be challenging for most clinicians. This problem could be solved using large image
databases combined with machine learning and pattern recognition algorithms that would
automatically reconstruct complex anatomic geometries in trauma patients.
The orbital floor mould used in this study was printed in 84 minutes using a high-
resolution 3D printer. During fabrication, the printer deposits a 0.1-mm-thick layer of powder
over which the printing head moves and deposits a binder that solidifies the powder.
The thin-layer increments subsequently offer very precise 3D printed constructs.
However, designing the very precise virtual model required for the fabrication of a mould
is slightly more challenging since it is dictated by the correctness of the conversion of CT
images into virtual 3D models (Figure 2). The accuracy of a virtual 3D model derived from
CT images currently ranges between 0.4 and 0.7 mm [11]. Nevertheless, the autologous
orbital floor implant used in this case report demonstrated an excellent fit in-situ and
subsequently reduced the operation time by approximately 40% (Figure 4).
The total manufacturing costs of the orbital floor mould were low (€85) compared with
other treatment options. To date, 3D printed titanium orbital implants have cost between
€3,000 and €5,000 euros. The costs of prefabricated titanium orbital plates or meshes
range between €250 euros (KLS Martin) and €600 and €800 (Synthes, West Chester, PA).
The treatment of bony defects with autologous bone still remains a challenge since
the amount of donor bone available is anatomically limited. A major drawback of
autogenous bone is, of course, the need for harvesting, which is accompanied by donor-
site morbidity. In addition, the sculpturing of autologous bone to fit the recipient site can
be time-consuming and manually cumbersome and subsequently result in inconsistent
cosmetic outcomes. However, this problem can be minimized using the treatment method
described in this study.
In conclusion, virtual preoperative planning combined with 3D printing offers
a geometrically predictable and low-cost method of creating autologous individualized
orbital floor implants.
A N
OV
EL M
ETH
OD
OF O
RB
ITAL FLO
OR
RE
CO
NSTR
UC
TION
24
2
REFERENCES1. Kontio R K, Laine P, Salo A, Paukku P,
Lindqvist C, and Suuronen R (2006) Reconstruction of internal orbital wall fracture with iliac crest free bone graft: clinical, computed tomography, and magnetic resonance imaging follow-up study. Plast. Reconstr. Surg. 118:1365–1374
2. Glassman R D, Manson P N, Vanderkolk C A, Iliff N T, Yaremchuk M J, Petty P, Defresne C R, and Markowitz B L (1990) Rigid fixation of internal orbital fractures. Plast. Reconstr. Surg. 86:1101–1103
3. Wolff J, Sándor G K B, Pyysalo M, Miettinen A, Koivumäki A-V, and Kainulainen V T (2013) Late reconstruction of orbital and naso-orbital deformities. Oral Maxillofac. Surg. Clin. North Am. 25:683–95
4. Kang S-H, Lee J-W, Lim S-H, Kim Y-H, and Kim M-K (2014) Verification of the usability of a navigation method in dental implant surgery: in vitro comparison with the stereolithographic surgical guide template method. J. Craniomaxillofac. Surg. 42:1530–5
5. Mazzoni S, Bianchi A, Schiariti G, Badiali G, and Marchetti C (2015) Computer-aided design and computer-aided manufacturing cutting guides and customized titanium plates are useful in upper maxilla waferless repositioning. J. Oral Maxillofac. Surg. 73:701–707
6. Avashia Y J, Sastry A, Fan K L, Mir H S, and Thaller S R (2012) Materials used for
reconstruction after orbital floor fracture. J. Craniofac. Surg. 23:1991–1997
7. Stoor P, Suomalainen A, Lindqvist C, Mesimaki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific implants for reconstruction of orbital wall defects. J. Craniomaxillofac. Surg. 42:1644–1649
8. Schuh A, Bigoney J, Hönle W, Zeiler G, Holzwarth U, and Forst R (2007) Second generation (low modulus) titanium alloys in total hip arthroplasty. Materwiss. Werksttech. 38:1003–1007
9. Ong H (2012) Biomechanical properties of the human orbital floor for selection of reconstruction materials using a mathematical engineering model [master’s thesis]. Groningen, The Netherlands, University of Goningen, 2012. Available at: http://irs.ub.rug.nl/dbi/53f59f60d8234
10. Al-Sukhun J, Penttila H, and Ashammakhi N (2012) Orbital stress analysis, Part IV: Use of a “stiffness-graded” biodegradable implants to repair orbital blow-out fracture. J. Craniofac. Surg. 23:126–130
11. van Eijnatten M, Rijkhorst E-J, Hofman M, Forouzanfar T, and Wolff J (2016) The accuracy of ultrashort echo time MRI sequences for medical additive manufacturing. Dentomaxillofacial Radiol. 45:20150424
3 INFLUENCE OF CT PARAMETERS ON STL MODEL ACCURACY
Maureen van Eijnatten, Ferco Berger, Pim de Graaf,
Juha Koivisto, Tymour Forouzanfar, Jan Wolff
Rapid Prototyping Journal (2017). 24 (4): 679-685doi: 10.1108/RPJ-07-2015-0092
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
28
3
ABSTRACTPurpose Additive manufactured (AM) skull models are increasingly used to plan complex
surgical cases and design custom implants. The accuracy of such constructs depends
on the standard tessellation language (STL) model, which is commonly obtained from
computed tomography (CT) data. The aims of this study were to assess the image quality
and the accuracy of STL models acquired using different CT scanners and acquisition
parameters.
Design/methodology/approach Images of three dry human skulls were acquired using two multi-detector row computed
tomography (MDCT) scanners, a dual energy computed tomography (DECT) scanner and
one cone beam computed tomography (CBCT) scanner. Different scanning protocols
were used on each scanner. All images were ranked according to their image quality and
converted into STL models. The STL models were compared to gold standard models.
Findings Image quality differed between the MDCT, DECT and CBCT scanners. Images acquired
using low-dose MDCT protocols were preferred over images acquired using routine
protocols. All CT-based STL models demonstrated non-uniform geometrical deviations of
up to +0.9 mm. The largest deviations were observed in CBCT-derived STL models.
Practical implications While patient-specific AM constructs can be fabricated with great accuracy using AM
technologies, their design is more challenging because it is dictated by the correctness
of the STL model. Inaccurate STL models can lead to ill-fitting implants that can cause
complications after surgery.
Originality/value This paper suggests that CT imaging technologies and their acquisition parameters affect
the accuracy of medical AM constructs.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
29
3
INTRODUCTIONAdditive manufacturing (AM), often referred to as three-dimensional (3D) printing, is
becoming increasingly popular in the field of medicine. Recent developments in high-
resolution imaging techniques and more affordable 3D printers have led to an increased
use of different AM technologies in clinical settings [1]. AM medical models are becoming
indispensable in medical education and communication [2], [3], treatment planning, and
surgical guidance [4], and have been shown to improve surgical outcomes and reduce
operating times [5]. Furthermore, AM drill and saw guides in combination with patient-
specific constructs such as implants or reconstruction plates are being more readily used
in maxillofacial surgery [6].
Medical AM differs from industrial AM in that it requires patient-specific 3D medical
images. To date, such images are commonly acquired using computed tomography (CT)
technology. The resulting CT datasets are archived as Digital Imaging and Communications
in Medicine (DICOM) files that are subsequently converted into standard tessellation
language (STL) models. Such models are necessary to design medical AM constructs using
commercially available computer-aided design (CAD) software packages.
A wide range of different CT technologies are currently available on the market for
diagnostic purposes. To date, multi-detector row computed tomography (MDCT) scanners
are most commonly used in hospital environments. These scanners have multiple rows
of detectors, hence short scanning times, and allow high-resolution imaging of bone
[7]. In addition, dual energy computed tomography (DECT) scanners have recently been
developed and offer enhanced tissue characterization by using two X-ray beams with
different photon energy spectra [8]. However, the scanning protocols used with these
advanced imaging technologies exhibit a wide variability in terms of image acquisition
and exposure parameters. These parameters include the scan mode (e.g. axial or helical),
tube potential (kV), exposure (mAs), rotation time, X-ray beam filter type and size, beam
collimation, detector configuration, helical pitch (or table feed per gantry rotation), and
field of view (FOV).
Cone beam computed tomography (CBCT) scanners are commonly used in private
clinics due to their small size and low costs. CBCT technology uses a cone shape X-ray
beam and a single flat panel detector to obtain 3D information of the head of a patient
by data acquisition through a single 200°-360° gantry rotation [9]. Cone beam technology
induces lower radiation doses compared with MDCT [10], [11] and has recently become
more affordable. This has led to an increase in popularity amongst maxillofacial radiologists
and surgeons [12]. Comparable to MDCT, modern CBCT scanners offer a wide range of
image acquisition parameters that include scan time, tube potential (kV), exposure (mAs),
rotation time, gantry rotation, FOV, and various metal artifact reduction algorithms.
The aforementioned CT technologies and resulting datasets are currently used for
diagnostic purposes in clinical settings. Such diagnostic datasets are also pro tempore
used to design medical constructs. However, it must be noted that imaging protocols
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
30
3
required for diagnostic purposes may differ markedly from those required by technicians
to design patient-specific AM constructs [13]. Nevertheless, all clinicians are bound to
the ALARA (as low as reasonably achievable) radiation safety principle. Therefore, all CT
scans required for patient-specific constructs need to be foremost diagnostable. The aims
of this study were to assess the image quality and the accuracy of STL models acquired
using different MDCT and CBCT scanners and acquisition parameters.
MATERIALS AND METHODSThree formalin-fixed human cadaver skulls of succumbed patients with intact bony structures
were obtained from the Department of Anatomy of the VU University Medical Center,
Amsterdam. All soft tissues were meticulously removed. The skulls were subsequently
scanned using a GOM ATOS™ optical 3D scanner (GOM, Braunschweig, Germany). This
device uses non-contact, blue-light technology to record a series of images of an object
from different angles, and combines these images into a triangular mesh representing
the surface of the object. The GOM ATOS™ optical 3D scanner offers a point accuracy
of 0.05 mm and was therefore used to obtain a “gold standard” STL model of each
skull (Figure 1).
Figure 1. Outline of the study.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
31
3
CT image acquisitionAll three dry human skulls were subsequently scanned using two different MDCT scanners
and a CBCT scanner (Figure 1): a Siemens Somatom Sensation 64-slice MDCT scanner
(Siemens Medical Solutions, Erlangen, Germany), a GE Discovery CT750 HD 64-slice
MDCT scanner (GE Healthcare, Little Chalfont, Buckinghamshire, UK), and a Vatech PaX
Zenith 3D CBCT scanner (Vatech, Fort Lee, USA). In addition, the GE scanner offered
a DECT image acquisition possibility using rapid switching X-ray beams with 140 kVp and
80 kVp voltage. All acquisition parameters are presented in Table 1.
Different scanning protocols were used on the aforementioned MDCT and CBCT
scanners. Initially, routine scanning protocols recommended by the device manufacturers
were used. In a second step, minimum and maximum tube potential (kV) and tube current
(mA) were used (Table 2) in order to assess the influence of these parameters on image
quality and accuracy of the STL models. Furthermore, the radiation output was calculated
using the dose-length product (DLP) for all MDCT protocols, and the dose-area product
(DAP) for all CBCT protocols.
All MDCT and DECT images were reconstructed using sharp reconstruction kernels
(Siemens: H70h, GE: BONEPLUS for routine mode and DETAIL for dual energy mode).
These kernels were selected after assessing all the different kernels available on the Siemens
and GE scanners.
Image quality analysisFollowing image acquisition and reconstruction, two radiologists with specific expertise
in maxillofacial bone imaging visually assessed and ranked all 57 DICOM datasets hence
Table 1. MDCT, DECT and CBCT exposure parameters.
Exposure parameters
Siemens Somatom Sensation 64-slice MDCT
GE Discovery CT750 HD 64-slice MDCT/DECT
Vatech PaX Zenith CBCT
Tube voltage (kVp) 80-140 80-140 50-120Tube current (mA) 150-650 150-650 5-10Rotation time (s) 1 1 15-24Q (mAs) 150-650 150-650 75-240Slice thickness (mm) 0.6 0.625 n.a.Slice increment (mm) 0.391 0.312 n.a.Pitch (mm) 0.55-1 0.5-1 n.aCTDIvol (mGy) 5.94-113 17.7-227 n.a.Voxel (mm) 0.391 x 0.391 x 0.6 0.312 x 0.312 x 0.625 0.2 x 0.2 x 0.2Voxel number 512 x 512 x 512 512 x 512 x 512 1056 x 1056 x 1056Scan angle 360º 360º 360ºPulsed No Only for dual energy YesScan FOV (cm) 20 20 Free (~24x19)
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
32
3
images (19 per skull). The radiologists were blinded from the acquisition parameters and
from the results of one another. A constant Window/Level setting was used to evaluate all
images equally (W6000; L100). The following anatomical structures were visually assessed:
1) Laminae papyracea; 2) Infra-orbital canal; 3) Sphenoid sinus septum; and 4) Orbital apex
(Figure 2). Each anatomical structure was given scores from 1 to 4 in ascending order; 1 =
anatomic structure not identifiable due to poor image quality; 2 = structure identified but
no details assessable and insufficient image quality; 3 = anatomic structure fully assessable
in all parts and acceptable image quality; 4 = very good delineation of structure and
excellent image quality.
Because of the small variations in the image quality of the different scanning protocols
(Table 3), a second more specific ranking system was necessary. In this ranking system,
all images were given scores from best to worst according to relative image quality
(Table 4). All image assessments and hence rankings were repeated after a 6-month
interval to minimize the radiologists’ learning effect.
Statistical analysis of the image qualityThe mean values and standard deviations (SD) of both ranking systems (Table 3 and 4) were
calculated using SPSS Statistics software (version 23.0 for Windows; SPSS Inc, Chicago,
USA). Furthermore, all rankings were compared using a nonparametric Mann-Whitney U
test with a 2-tailed p value. Statistical significance was established at p ≤ 0.05.
Image processingAfter image qualtiy analysis, All MDCT, DECT, and CBCT DICOM datasets were
subsequently converted into STL models using OsiriX MD software (Pixmeo, Geneva,
Switzerland). Automatic thresholding was used with a uniform intensity value of -500
Hounsfield Units (HU).
Table 2. MDCT, DECT and CBCT scanning protocols: tube potentials (kV) and exposures (mAs).
Scanning protocolMDCT (Siemens)kV / mAs
MDCT (GE)kV / mAs
DECT (GE)kV / mAs
CBCT (Vatech)kV / mAs
1. Routine 120 / 380 120 / 220 80 & 140 / 375 115 / 1442. Routine kV / max. mA 120 / 650 120 / 650 80 & 140 / 630 115 / 2403. Min. kV / max. mA 80 / 650 80 / 650 n.a.4. Max. kV / max. mA 140 / 550 140 / 510 120 / 2405. Min. kV / min. mA 80 / 150 80 / 150 50 / 756. Max. kV / Routine mA 120 / 1447. Metal artifact reduction 115 / 1448. Mediate kV / routine mA 80 / 144
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
33
3
Standard tessellation language deviation analysisAll STL files were imported into GOM Inspect software (GOM GmbH, Braunschweig,
Germany), and were cleared of outliers and aligned with the gold standard STL model
using a “Local Best-Fit” function. Geometrical differences between the STL models and
the gold standard were calculated using the “Surface Comparison on Actual” function and
visualized using colour maps. A detailed assessment of the orbital area was performed
using the “Select on Surface” function. All distance values were exported to Excel
software (Microsoft Office 2013, Redmond, WA, United States) and the mean and SD were
calculated. Finally, histograms and kurtosis were calculated using an excel data analysis
software plugin.
RESULTSThe image quality of the following four anatomical structures was visually assessed:
laminae papyracea; infra-orbital canal; sphenoid sinus septum and orbital apex (Figure 2).
Furthermore, the DLP of each MDCT examination and the DAP of each CBCT examination
was calculated. The results are summarized in Table 3. Scanner-specific rankings of
the different scanning protocols are presented in Table 4.
All visually examined images were subsequently converted into STL models.
The resulting STL models were compared to a gold standard skull STL model acquired
using an optical scanner. When compared with the gold standard STL models, all STL
models obtained from MDCT, DECT, and CBCT datasets demonstrated large non-
uniform deviations in the orbital wall, maxillary sinus, arcus zygomaticus, and nasal cavity.
Figure 2. Evaluated anatomical structures, 1) Laminae papyracea; 2) Infra-orbital canal; 3) Spenoid sinus septum; and 4) Orbital apex. The images were acquired using a Siemens MDCT scanner, 80 kV, 150 mA, window/level: 6000/100.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
34
3
Tab
le 3
. Im
age
qua
lity
sco
res
(sca
le 1
– 4
) and
do
se-l
eng
th p
rod
uct
(DLP
) / d
ose
-are
a p
rod
uct
(DA
P) o
f all
MD
CT,
DEC
T, a
nd C
BC
T d
atas
ets a
.
Scan
ning
pro
toco
l
MD
CT
(Sie
men
s)M
DC
T (G
E)
DE
CT
(GE
)C
BC
T (V
atec
h)
Mea
n ±
SD
pD
LP
(mG
y-cm
)M
ean
±
SDp
DLP
(m
Gy-
cm)
Mea
n ±
SD
pD
LP
(mG
y-cm
)M
ean
±
SDD
AP
(mG
y-cm
2 )
1. R
out
ine
3 ±
0.3
– 11
323
± 0
.5–
1632
2 ±
0.4
–14
942
± 0
.6–
25.0
8
2. R
out
ine
kV /
max
. mA
3 ±
0.3
118
653
± 0
.6.7
8546
762
± 0
.41
1614
1 ±
0.5
.029
*41
.51
3. M
in. k
V /
max
. mA
2 ±
0.6
.061
544
3 ±
0.8
116
764.
Max
. kV
/ m
ax. m
A3
± 0
.0.3
1723
903
± 0
.8.4
4051
111
± 0
.3.0
01*
43.3
65.
Min
. kV
/ m
in. m
A3
± 0
.31
131
3 ±
0.7
.797
398
1 ±
0.4
.004
*1.
646.
Max
. kV
/ R
out
ine
mA
1 ±
0.3
.001
*25
.20
7. M
etal
art
ifact
red
uctio
n2
± 0
.8.7
3225
.08
8. M
edia
te k
V /
ro
utin
e m
A2
± 0
.8.2
1111
.82
a p v
alue
s in
dic
ate
a co
mp
aris
on
with
the
ro
utin
e p
roto
col (
* =
p <
.05)
. All
p v
alue
s fo
r M
DC
T an
d D
EC
T sc
anni
ng p
roto
cols
are
no
nsig
nific
ant.
Tab
le 4
. Sc
anne
r-sp
ecifi
c ra
nkin
gs
of a
ll M
DC
T, D
ECT,
and
CB
CT
dat
aset
s a .
Scan
ning
pro
toco
l
MD
CT
(Sie
men
s)M
DC
T (G
E)
DE
CT
(GE
)C
BC
T (V
atec
h)
Mea
n ±
SD
pM
ean
± S
Dp
Mea
n ±
SD
pM
ean
± S
Dp
1. R
out
ine
3.2
± 1
.3–
2.9
± 1
.1–
1.9
± 0
.2–
6.2
± 0
.9–
2. R
out
ine
kV /
max
. mA
3.1
± 1
.1.2
842.
5 ±
1.3
.023
*1.
1 ±
0.2
.000
*3.
1 ±
1.3
.000
*3.
Min
. kV
/ m
ax. m
A1.
6 ±
1.0
.000
*3.
2 ±
1.2
.154
4. M
ax. k
V /
max
. mA
3.5
± 1
.2.1
791.
9 ±
1.3
.000
*1.
3 ±
0.6
.000
*5.
Min
. kV
/ m
in. m
A3.
7 ±
1.5
.074
4.3
± 1
.1.0
00*
3.3
± 1
.1.0
00*
6. M
ax. k
V /
Ro
utin
e m
A3.
3 ±
1.7
.000
*7.
Met
al a
rtifa
ct r
educ
tion
6.1
± 1
.0.7
798.
Med
iate
kV
/ r
out
ine
mA
4.8
± 1
.1.0
00*
a p v
alue
s in
dic
ate
a co
mp
aris
on
with
the
ro
utin
e p
roto
col (
* =
p <
.05)
.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
35
3
The calculated deviations of one skull (no. 2) are presented in Figure 3. The other two
skulls (no. 1, 3) presented similar deviations.
The STL models of skull 1, 2, and 3 acquired using the Siemens MDCT scanner showed
deviations of 0.58 mm, 0.49 mm, and 0.77 mm and SDs of 0.91, 0.81, 0.99, respectively
(Figure 3A). The GE MDCT STL deviations ranged from 0.56 mm, 0.45 mm, and 0.74 mm
with SDs of 0.88, 0.76, and 0.96, respectively (Figure 3B). DECT STL models resulted
in deviations of 0.44 mm, 0.45 mm, and 0.62 mm with SDs of 0.73, 0.71, and 0.84,
respectively (Figure 3C). Furthermore, DECT STL models showed less detail in the orbital
and maxillary sinus area when compared with the MDCT models. Finally, the CBCT scanner
presented deviations of 0.63 mm, 0.63 mm, and 0.80 mm with SDs of 0.88, 0.88, and 0.95,
respectively (Figure 3D).
The deviations between the gold standard STL models and STL models acquired from
all three skulls using four different CT scanners generally ranged between -0.9 mm and
Figure 3. Geometric deviations mapped on the surface of the STL models acquired using a routine scanning protocol on four different CT scanners. (a) MDCT (Siemens) - 120 kV/380mAs; (b) MDCT (GE) - 120 kV/380 mAs; (c) DECT (GE) - 80&140 kV/375 mAs; (d) CBCT (Vatech) - 115 kV/144 mAs
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
36
3
Figure 4. Frequency of the geometric deviations (in mm) of the STL models of all three skulls when compared with the gold standard.
+0.8 mm and are summarized in a histogram (Figure 4). The MDCT STL models showed
the smallest deviations, but the models were on average larger than the gold standard
STL model. The kurtosis – the steepness of a curve compared to a normal curve – of
the deviations in the MDCT STL models was 1.7 (Siemens) and 2.7 (GE), 0.08 in the DECT
STL models and -0.9 in the CBCT STL models.
Detailed assessments of the orbital area in two Siemens MDCT STL models of skull
2 are presented in Figure 5. When compared with the routine scanning protocol (A),
the min kV/min mA protocol (B) resulted in better anatomical details. The mean deviation
in the orbits was 0.25 ± 0.41 mm (routine protocol) and 0.29 ± 0.44 mm (min mA/min kV
protocol). In all MDCT STL models, the orbits were larger than the gold standard in 80 per
cent of the surface area.
Figure 5. An example of the deviations in the STL models attained using standard and min kV/min mA scanning protocols on the Siemens MDCT scanner.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
37
3
DISCUSSIONIn the present study, the image quality and the accuracy of skull-derived STL models
acquired using different MDCT and CBCT scanners and acquisition parameters
were assessed.
The image quality differed markedly between the MDCT, DECT and CBCT scanners
(Table 3). The MDCT scanners offered the best image quality. Furthermore, the images
acquired using min kV/min mA and hence low-dose MDCT scanning protocols were
preferred over the images acquired using routine protocols (Table 4). Moreover, compared
to the routine protocols, the low-dose protocols showed a significant reduction in the DLP
of 88 and 76 per cent on the Siemens and GE scanner, respectively (Table 3). These findings
have large clinical implications since the use of low-dose scanning protocols can markedly
reduce the effective dose in patients who require patient-specific AM constructs without
jeopardizing the overall diagnostic image quality.
After image quality assessment, the acquired DICOM datasets were converted into STL
models and geometrically compared to the corresponding gold standard skull models.
In this context, the MDCT scanners offered the most accurate STL models (Figure 3) with
absolute mean deviations ranging between 0.45 mm and 0.77 mm when compared with
the optical scan “gold standard” STL model. These results are in good agreement with
the findings of Thorhauer et al. (2010), who found mean deviations of 0.512 mm between
MDCT STL models and optical scan models of the knee [14]. Other studies reported mean
deviations of 0.13 mm [15] and 0.45 mm [16] on long cadaver bones and 0.137 mm on dry
human mandibles [17]. However, these studies only included large bony structures.
The STL models acquired using DECT scanning protocols demonstrated greater loss of
bony details when compared to the MDCT protocols, especially in the thin bony structures
of the orbit (Figure 3). Deviations ranging between 0.44 mm and 0.62 mm were found in
the DECT STL models. These findings could be due to the rapid switching between photon
energy spectra required for dual energy imaging on the GE scanner. Another explanation
could be the lack of a bone reconstruction kernel for dual energy image reconstruction.
The largest geometrical deviations were observed in the CBCT STL models (Figure 3),
with mean deviations ranging between 0.63 mm and 0.80 mm. Liang et al. (2010) reported
mean deviations of between 0.165 and 0.386 mm for CBCT STL models [17], though
deviations of up to 2-5 mm have also been reported [18]. The overall low accuracy of
the CBCT STL models found in this study could be caused by the relatively high amount of
noise, artefacts [19], and inhomogeneities in the CBCT X-ray beam [20].
The geometrical deviations that were observed in the orbital area of all STL models
could have been caused by the thin (less than 1 mm) and unique anatomy of the orbits.
Currently, most CT scanners have an in-plane spatial resolution of approximately 0.3
mm and a slice thickness of 0.6 mm and can therefore not detect such bony structures
(smaller than 0.6 mm). Moreover, the complex, scaffold-like structure of the orbital area
also comprises thick supportive bony structures. These structures attenuate the lower part
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
38
3
Figure 6. Beam hardening and photon starvation in the orbital area can cause an enlargement of the STL model when compared to the actual bone.
of the radiation energy spectrum and lead to beam hardening and photon starvation [21]
(Figure 6). Hardened energy photons are less prone to attenuate in thin bony structures,
causing a loss of contrast in these structures [22]. Interestingly, when compared with
routine MDCT scanning protocols, min kV/min mA and hence low-dose MDCT protocols
resulted in similar geometric deviations and an increase of detail in the orbital area (Figure
5). These results are in good agreement with a previous study by Oka et al. (2009), who
found minor geometrical differences between bone models constructed using normal and
low-dose CT parameters.
Clinical implicationsWhile patient-specific medical constructs can be fabricated with great accuracy (< 0.1 mm)
using AM technologies, their design is more challenging since it is dictated by the correctness
of the STL model. A recent study by Stoor et al. (2014) reported that 24 per cent of
their additively manufactured orbital implants did not fit due to a poor representation of
the orbital walls in the corresponding CT-based STL models. The consequence of ill-fitting
implants is that increased movements can jeopardize wound healing and tissue repair
and subsequently lead to complications after surgery [24]. Furthermore, geometrically
inaccurate implants can cause nerve impingements in complex anatomical areas such as
the orbit, with devastating outcomes. This opens new questions concerning the liability
of AM medical constructs. Therefore, technicians working in the medical field need to
understand the inaccuracies that can be introduced during the CT imaging and STL
conversion process required for the fabrication of medical AM constructs.
Limitations of this studyThe major limitation of this study was that skulls without soft tissues were used.
The photon energy spectrum of an X-ray beam changes as it passes through water [25],
which results in differences in absorption and scattering during clinical CT scanning.
Furthermore, the intensity values of soft tissues are closer to the values of bone, hence
the presence of soft tissues will make bone segmentation more difficult. Therefore,
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
39
3
the reported accuracies found in this study are not simply generalizable to clinical
conditions. Nevertheless, dry skulls combined with optical scanning technology offered
an highly accurate and detailed 3D geometric “gold standard” to evaluate the accuracy
of CT-based STL models. The acquisition of a gold standard is not straightforward for
complex anatomical structures, and even impossible to acquire in an vivo study.
CONCLUSIONCT imaging technologies and their acquisition parameters have an effect on the accuracy
of medical AM constructs. MDCT, DECT and CBCT scanners differ in image quality and
STL model accuracy in the maxillofacial area. Overall, MDCT scanners offered the best
image quality and STL models. The use of low-kV and low-mA instead of routine MDCT
protocols can markedly reduce the effective dose in patients who require patient-specific
AM constructs.
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
40
3
REFERENCES1. Ventola C L (2014) Medical Applications
for 3D Printing : Current and Projected Uses. Pharm. Ther. 39:704–711
2. Santler G, Karcher H, and Ruda C (1998) Indications and limitations of three-dimensional models in cranio-maxillofacial surgery. J. Cranio-Maxillofacial Surg. 26:11–16
3. D’Urso P S, Barker T M, Earwaker W J, Bruce L J, Atkinson R L, Lanigan M W, Arvier J F, and Effeney D J (1999) Stereolithographic biomodelling in cranio-maxillofacial surgery: a prospective trial. J. Cranio-Maxillofacial Surg. 27:30–37
4. Giannatsis J and Dedoussis V (2009) Additive fabrication technologies applied to medicine and health care: a review. Int. J. Adv. Manuf. Technol. 40:116–127
5. Guarino J, Tennyson S, McCain G, Bond L, Shea K, and King H (2007) Rapid Prototyping Technology for Surgeries of the Pediatric Spine and Pelvis Benefits Analysis. J. Pediatr. Orthop. 27:955–960
6. Parthasarathy J (2014) 3D modeling, custom implants and its future perspectives in craniofacial surgery. Ann. Maxillofac. Surg. 4:9–18
7. Suetens P (2009) Fundamentals of Medical Imaging. Cambridge University Press, New York, NY
8. Tawfik A M, Kerl J M, Razek A A, Bauer R W, Nour-Eldin N E, Vogl T J, and Mack M G (2011) Image quality and radiation dose of dual-energy CT of the head and neck compared with a standard 120-kVp acquisition. Am. J. Neuroradiol. 32:1994–1999
9. Lofthag-Hansen S (2010) Cone Beam Computed Tomography: Radiation dose and image quality assessments, University of Gothenburg, Gothenburg
10. Ludlow J B and Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and
maxillofacial radiology. Oral Surgery, Oral Med. Oral Pathol. Oral Radiol. Endodontology 106:106–114
11. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
12. Dawood A, Patel S, and Brown J (2009) Cone beam CT in dental practice. Br. Dent. J. 207:23–28
13. Mallepree T and Bergers D (2009) Accuracy of medical RP models. Rapid Prototyp. J. 15:325–332
14. Thorhauer E, Miyawaki M, Illingworth K, Holmes J A, and Anderst W (2010) Accuracy of bone and cartilage models obtained from CT and MRI. Poster presented at the 34th Annual Meeting of the American Society of Biomechanics, 18–21
15. Rathnayaka K, Momot K I, Noser H, Volp A, Schuetz M A, Sahama T, and Schmutz B (2014) Quantification of the accuracy of MRI generated 3D models of long bones compared to CT generated 3D models. Med. Eng. Phys. 34:357–363
16. Oka K, Murase T, Moritomo H, Goto A, Sugamoto K, and Yoshikawa H (2009) Accuracy analysis of three-dimensional bone surface models of the forearm constructed from multidetector computed tomography data. Int. J. Med. Robot. Comput. Assist. Surg. 5:452–457
17. Liang X, Lambrichts I, Sun Y, Denis K, Hassan B, Li L, Pauwels R, and Jacobs R (2010) A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT). Part II: On 3D model accuracy. Eur. J. Radiol. 75:270–274
18. Shahbazian M, Jacobs R, Wyatt J, Willems G, Pattijn V, Dhoore E, Van
INFLU
EN
CE
OF C
T PAR
AM
ETE
RS O
N STL M
OD
EL A
CC
UR
AC
Y
41
3
Lierde C, and Vinckier F (2010) Accuracy and surgical feasibility of a CBCT-based stereolithographic surgical guide aiding autotransplantation of teeth: In vitro validation. J. Oral Rehabil. 37:854–859
19. Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A, Bosmans H, Bogaerts R, Jacobs R, and Horner K (2013) Variability of dental cone beam CT grey values for density estimations. Br. J. Radiol. 86:20120135
20. Feldkamp L A, Davis L C, and Kress J W (1984) Practical cone-beam algorithm. J. Opt. Soc. Am. 1:612
21. Zbijewski W, Stayman J, and Muhit A (2012) CT Reconstruction Using Spectral and Morphological Prior Knowledge: Application to Imaging the Prosthetic Knee. The second international conference on image formation in X-ray computed tomography, Salt Lake City, UT, 434–438
22. Koivisto J, Wolff J, Järnstedt J, Dastidar P, and Kortesniemi M (2014) Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 118:355–362
23. Stoor P, Suomalainen A, Lindqvist C, Mesimäki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific guiding implants for orbital wall defects. J. Cranio-Maxillofacial Surg. 42:1644–1649
24. Lienau J, Schell H, Duda G N, Seebeck P, Muchow S, and Bail H J (2005) Initial vascularization and tissue differentiation are influenced by fixation stability. J. Orthop. Res. 23:639–645
25. Barrett J F and Keat N (2004) Artifacts in CT: Recognition and Avoidance. RadioGraphics ` 24:1679–1691
4 IMPACT OF PRONE, SUPINE AND OBLIQUE PATIENT POSITIONING
ON CBCT IMAGE QUALITY, CONTRAST TO NOISE RATIO
AND FIGURE OF MERIT VALUE IN THE MAXILLOFACIAL REGION
Juha Koivisto, Maureen van Eijnatten, Jorma Järnstedt,
Kisri Holli-Helenius, Prasun Dastidar, Jan Wolff
Dentomaxillofacial Radiology (2017). 46 (6): 20160418doi: 10.1259/dmfr.20160418
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
44
4
ABSTRACTObjectiveTo assess the impact of supine, prone and oblique patient imaging positions on the image
quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region using
a cone beam computed tomography scanner. Furthermore, the CBCT supine images were
compared to supine MSCT images.
MethodsOne fresh frozen cadaver head was scanned in prone, supine and oblique imaging
positions using a mobile CBCT scanner. MSCT images of the head were acquired in
a supine position. Two radiologists graded the CBCT and MSCT images at ten different
anatomical sites according to their image quality using a six-point scale. CNR and FOM
values were calculated at two different anatomical sites on the CBCT and MSCT images.
ResultsThe best image quality was achieved in prone imaging position for sinus, mandible
and maxilla, followed by supine and oblique imaging positions. 12-mA prone images
presented high delineation scores for all anatomical landmarks, except for the ear region
(carotid canal), which presented adequate to poor delineation scores for all studied head
positions and exposure parameters. The MSCT scanner offered similar image qualities to
the 7.5-mA supine images acquired using the mobile CBCT scanner. The prone imaging
position offered the best CNR and FOM values on the mobile CBCT scanner.
ConclusionsHead positioning has an impact on CBCT image quality. The best CBCT image quality can
be achieved using the prone and supine imaging positions. The oblique imaging position
offers inadequate image quality except in the sinus region.
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
45
4
INTRODUCTION Mobile cone beam computed tomography (CBCT) scanners were first introduced in the field
of maxillofacial surgery in 2005 [1]. Since then, mobile CBCT scanners [2]–[8] have become
increasingly popular as they offer comparable images to those acquired on traditional
multislice computed tomography (MSCT) scanners [9]. The major advantages of CBCT
scanners are their lower cost, size, weight and effective dose [10]–[12]. The lower effective
dose is the result of the default exposure parameters and the partial gantry rotation used
in CBCT scanners when compared to MSCT scanners [13], [14]. It has, however, been
reported that the partial CBCT gantry rotation produces a non-uniform dose distribution
[15] that can in some cases affect image quality. Nevertheless, the partial CBCT gantry
rotation offers acceptable diagnostic image quality at a lower effective dose than MSCT
scanners [16].
A recent study by Koivisto et al. [17] reported the positive effect of head positioning
on the effective dose in CBCT scanners. Furthermore, a study by Yan et al. [18] reported
a close relationship between image quality and effective dose when using low dose CBCT
protocols. Taking the results of the previous studies into account, it can be hypothesized
that CBCT images could be optimised by choosing the correct patient imaging position.
Currently, MSCT and CBCT image quality assessments are based on the visual
grading of predefined anatomical landmarks [19]–[21], the quantitative analysis of
physical properties such as modulation transfer function (MTF) [22] and contrast-to-noise
ratio (CNR) [23]. The image contrast is affected by the x-ray mean photon energy that
subsequently has an effect on the image quality [24]. Moreover, the relationship between
the image quality and patient dose is commonly quantified using the traditional figure of
merit (FOM) value that is defined as the (CNR) [2] divided by the incident patient exposure
[24]–[27]. The figure of merit value can provide useful information on the benefits and
drawbacks of different imaging positions in CBCT and MSCT scanners. It also provides
a means of comparing the ratio between the image quality and the resulting effective
dose. Furthermore the FOM offers the possibility of calculating a numerical value that
can be used to compare two inherently different imaging modalities. Moreover, the FOM
value can be used to select an appropriate protocol depending on the diagnostic task.
The combination of CNR and FOM calculations with image quality assessments offer an
adequate method for comparing and grading MSCT and CBCT images.
The objectives of this study were to assess the impact of three different patient imaging
positions (supine, prone, oblique) on CBCT image quality, contrast-to-noise ratio and
the figure of merit value in the maxillofacial region. The second objective was to compare
the CBCT supine images with MSCT images acquired in the same position.
MATERIALS AND METHODSThis study was performed according to the Ethical Principles for Medical Research Involving
Human Subjects as defined by the World Medical Association (Helsinki Declaration) [28].
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
46
4
One fresh frozen human cadaver head was provided by the Department of Anatomy,
Tampere University Hospital for image quality assessment.
ScannersCadaver images were obtained using a Planmed Verity mobile CBCT scanner (Planmed
Oy, Helsinki, Finland) and a Philips Brilliance-64 (64 slice) MSCT scanner (Philips Medical
Systems, Eindhoven, Netherlands). The mobile CBCT scanner uses a partial 210º gantry
scan and 3.0 mm Al and 0.5 mm Cu filters. The Philips Brilliance 64 MSCT scanner
uses a 360º rotation scan and combined filtration consisting of 2.5 mm aluminum and
1.2 mm titanium.
The mobile scanner was initially developed for extremity imaging and has recently been
approved by the American Food and Drug Administration (FDA) for maxillofacial imaging.
The scanner is equipped with a head positioning tray and a detachable forehead support.
During oblique imaging, the cadaver head was placed on the default tray and attached to
the forehead support. In order to acquire prone and supine images, the forehead support
was removed and the gantry was tilted accordingly.
Imaging positions A total of nine CBCT images were acquired in supine (Figure 1B), prone (Figure 1C) and
oblique (Figure 1D) imaging positions. Furthermore, one MSCT image was acquired using
the default supine imaging position (Figure 1A).
Exposure protocolsCBCT images were obtained using a constant 96 kV tube voltage. Three different tube
currents (1): 4.8 mA, (2): 7.5 mA and (3): 12 mA were used to encompass the full dynamic
Figure 1. A schematic presentation of the MSCT (A) and mobile CBCT (B, C, D) scanner used in this study with different head positions: supine (A, B), prone (C), and oblique (D).
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
47
4
range of the mobile CBCT scanner. All CBCT exposures were performed using the smallest
possible voxel size (0.2 mm). The MSCT exposures were performed using a default standard
high-resolution imaging protocol (120 kV, 93 mA).
All effective doses required to calculate the figure of merit (FOM) value were scaled
proportionally using the exposure values (mAs) attained in a previous study by Koivisto
et al. [17] using an anthropomorphic phantom. All exposure protocols, parameters and
calculated effective doses are presented in Table 1.
Standard axial, coronal and sagittal reconstructions were performed on a Philips
Brilliance Extended Workstation V 4.5.2.40007 (Philips Healthcare, the Netherlands).
Table 1. Exposure parameters of the CBCT and MSCT exposures
Planmed Verity Philips Brilliance
CBCT 64 MSCT
Imaging position Supine/ Supine/ Supine/ SupineProne/ Prone/ Prone/ -
Oblique Oblique Oblique -
Imaging protocol Protocol Protocol Protocol Protocol 4.8 mA 7.5 mA 12 mA High Resolution
Tube voltage (kVp) 96 96 96 120Tube current (mA) 4.8 7.5 12 93
Exposure time (s) 6 6 6 0.912
Q (mAs) 28.8 45 72 85
Slice thickness (mm) 0.2 0.2 0.2 0.67
Slice increment (mm) 0.2 0.2 0.2 0.33
Pitch (mm) NA NA NA 0.55
CTDIvol (mGy) - - - 11
Voxel size (mm) 0.2 x 0.2 x 0.2 0.2 x 0.2 x 0.2 0.2 x 0.2 x 0.2 0.33x0.352x0.352
Scan angle 210º 210º 210º 360º
Frame number 300 300 300 -
Radiation source Pulsed Pulsed Pulsed Continuous
Field of view (cm) 13 x 16 13 x 16 13 x 16 16 x 18
Eff.dose, Supine (µSv)* 125 195 312 781Eff.dose, Prone (µSv)* 97 151 242 -
Eff.dose, Oblique (µSv)* 29 45 72 -
NA, not applicable; CTDI, computed tomography dose index. *Calculated from the effective doses reported by Koivisto et al. 2014
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
48
4
The reconstructed image data consisting of 474 MSCT and 652 CBCT DICOM (Digital
Imaging and Communications in Medicine) files were subsequently exported to an Agfa
Impax PACS system (Agfa Healthcare, Mortsel, Belgium) and randomly visually graded
by two experienced radiologists using the same viewing conditions. All evaluations were
repeated after four weeks to assess the intra-observer reliability.
Image quality assessment All image quality assessments were performed on ten different anatomical landmarks in
four different anatomical regions: sinus (blue), ear (yellow), maxilla (green) and mandible
(red) (Figure 2) The image quality was graded by two radiologists using the following
scale: Level 5; excellent delineation of structures and excellent image quality, Level 4;
clear delineation of structures and excellent image quality, Level 3; anatomic structures
still fully assessable in all parts and good image quality, Level 2: structures identified and
results in adequate image quality, Level 1; anatomic structures not identifiable due to poor
image quality, Level 0; no diagnostic value.
Statistical analysis of the image quality The inter-rater agreement was evaluated by calculating the percentage of absolute
agreement and two-way random effects intra-class correlation coefficients (ICC) with
absolute agreement corresponding to 95% confidence interval (CI) [29]. The interclass
Figure 2. Anatomical landmarks assessed on the reconstructed MSCT and CBCT images
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
49
4
correlation coefficient (ICC) (Two-Way Mixed) model (2,2) was used to determine inter-
rater reliability [30].
The intra-rater reliability was calculated for four selected protocols (oblique position
(4.8 mA), supine position (12 mA) and prone position (7.5 mA)) using ICC (Two-Way Mixed)
model (3,1) [30]. ICCs range from 0 with no agreement to +1 with perfect agreement in
the interpretation of the ICC. Landis et al. [31] has characterized the values of reliability
coefficients as follows: less than 0.4, poor; 0.4 to 0.75, fair to good; and greater than 0.75,
excellent. Although arbitrary, these divisions provide useful benchmarks [29].
Contrast-to-noise ratioThe contrast-to-noise ratio (CNR) evaluation was performed in the fossa olfactorius
(anatomical landmark no. 6) and the apical aspect of tooth 43 (D43, anatomical landmark
no. 2). These anatomical sites were chosen since they represent commonly diagnosed
regions in the head area. The locations of ROIs were verified by superimposing CBCT and
MSCT images. During superimposition, all CBCT and corresponding MSCT images were
manually aligned into a coordinate system using control points chosen by the experienced
radiologist [32]–[34].
The CNR was calculated using the following equation:
(1)
where, SROI is the mean signal at the anatomical landmark in grey value unit, SBGR is
the mean signal in the background (HU) and σmean is the average standard deviation in
the anatomical landmark and background ROIs (Figure 3). Furthermore, the correlation
between CNR and the subjective image quality was calculated to assess the association
between the two image quality indicators.
The intra-rater reliability was calculated for four selected protocols (oblique position (4.8 mA), supine
position (12 mA) and prone position (7.5 mA)) using ICC (Two-Way Mixed) model (3,1) [30]. ICCs
range from 0 with no agreement to +1 with perfect agreement in the interpretation of the ICC. Landis et
al. [31] has characterized the values of reliability coefficients as follows: less than 0.4, poor; 0.4 to 0.75,
fair to good; and greater than 0.75, excellent. Although arbitrary, these divisions provide useful
benchmarks [29].
Contrast-to-noise ratio
The contrast-to-noise ratio (CNR) evaluation was performed in the fossa olfactorius (anatomical
landmark no. 6) and the apical aspect of tooth 43 (D43, anatomical landmark no. 2). These anatomical
sites were chosen since they represent commonly diagnosed regions in the head area. The locations of
ROIs were verified by superimposing CBCT and MSCT images. During superimposition, all CBCT and
corresponding MSCT images were manually aligned into a coordinate system using control points
chosen by the experienced radiologist [32]–[34].
The CNR was calculated using the following equation:
(1)
where, SROI is the mean signal at the anatomical landmark in grey value unit, SBGR is the mean signal in
the background (HU) and σmean is the average standard deviation in the anatomical landmark and
background ROIs (Figure 3). Furthermore, the correlation between CNR and the subjective image
quality was calculated to assess the association between the two image quality indicators.
Figure 3. Axial view of the tooth 43 (apex, D43) and background locations used for CNR calculation for MSCT
supine position (A) and for CBCT using supine (B), prone (C) and oblique (D) positions.
CNR SROI SBGR
mean
Figure 3. Axial view of the tooth 43 (apex, D43) and background locations used for CNR calculation for MSCT supine position (A) and for CBCT using supine (B), prone (C) and oblique (D) positions.
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
50
4
Figure of MeritIn this study, the figure of merit (FOM) value was used to assess the diagnostic efficacy of
the image quality versus the effective dose obtained using different imaging protocols.
The FOM value was calculated using the following equation described by Ogden et al. [26]
(2)
where E is the effective dose. In the present study, the FOM value was calculated for two
anatomical landmarks: the fossa olfactorius and the apex of tooth 43.
RESULTSImage quality assessmentThe average image quality results of nine different CBCT imaging protocols acquired in
different imaging positions and one MSCT protocol acquired in one position are presented
in Figure 4.
Statistical analysis of image qualityStatistical analysis was performed on all CBCT and MSCT image quality results.
The absolute agreement percentages, inter-rater intraclass correlation coefficients (ICC)
and 95% confidence intervals (CI) of all examination protocols are presented in Table 2.
The intra-rater reliability results of the two observers are presented in Table 3.
Figure of Merit
In this study, the figure of merit (FOM) value was used to assess the diagnostic efficacy of the image
quality versus the effective dose obtained using different imaging protocols. The FOM value was
calculated using the following equation described by Ogden et al. [26]
(2)
where E is the effective dose. In the present study, the FOM value was calculated for two anatomical
landmarks: the fossa olfactorius and the apex of tooth 43.
Results
Image quality assessment
The average image quality results of nine different CBCT imaging protocols acquired in different
imaging positions and one MSCT protocol acquired in one position are presented in Figure 4.
Figure 4. Results of subjective image quality assessments undertaken by two radiologists using different CBCT
and MSCT imaging protocols.
FOM CNR2
E
Figure 4. Results of subjective image quality assessments undertaken by two radiologists using different CBCT and MSCT imaging protocols.
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
51
4
Contrast-to-noise ratioThe mean CBCT contrast-to-noise values calculated in different imaging positions were as
follows: 3.6 in the prone imaging position, 1.8 in the supine imaging position and 1.5 in
the oblique imaging position. The Philips Brilliance 64 MSCT scanner mean CNR was 2.6.
Furthermore, the correlation between CNR and subjective image quality were 0.8217 for
tooth 43 and 0.3581 for the fossa olfactorius. The CNR values and the subjective image
qualities of the apex of tooth 43 and the fossa olfactorius are presented in Figure 5A and
Figure 5B, respectively.
Figure of MeritThe mean FOM values of the Planmed Verity CBCT scanner in different imaging positions
were as follows: 96.1 in the prone imaging position, 24.3 in the supine imaging position
and 22.3 in the oblique imaging position. The mean FOM value of the Philips Brilliance 64
MSCT scanner was 10.6. The FOM values of the apex of tooth 43 and the fossa olfactorius
obtained using one MSCT modality and multiple CBCT imaging modalities are presented
in Figure 6A and Figure 6B, respectively.
Table 2. Inter-rater absolute agreement (%), intra-class correlation (ICC) and confidence interval at 95% confidence level
Protocol Absolute agreement (%) ICC 95% Confidence Interval (CI)
MSCT (supine) 90 0.942 0.836, 0.984Oblique (12 mA) 90 0.967 0.908, 0.991Oblique (7.5 mA) 70 0.955 0.868, 0.988Oblique (4.8 mA) 90 0.964 0.897, 0.991Supine (12 mA) 80 0.893 0.696, 0.971Supine (7.5 mA) 70 0.880 0.659, 0.967Supine (4.8 mA) 70 0.887 0.675, 0.969Prone (12 mA) 70 0.889 0.687, 0.970Prone (7.5 mA) 70 0.898 0.710, 0.972Prone (4.8 mA) 80 0.940 0.827, 0.984
Table 3. Intra-rater reliability of selected protocols
Philips Brilliance 64 MSCTSupine 93 mA
Planmed Verity CBCT
Oblique 4.8 mA Supine 12 mA Prone 7.5 mA
Observer 1 0.780 0.873 0.571 0.856
Observer 2 1.000 0.877 0.667 0.816
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
52
4
DISCUSSIONIn this study, the impact of supine, prone and oblique imaging positioning on image quality
in four different skull regions was assessed. Furthermore, the contrast-to-noise ratio (CNR)
and the figure of merit value were calculated and compared. Finally, the image quality of
CBCT images acquired in the supine imaging position were compared to MSCT images
acquired in the same imaging position.
The combined mobile extremity/maxillofacial CBCT scanner used in this study is unique in
that it uses an adjustable gantry instead of a fixed gantry [35]. The adjustable gantry which
moves in an enclosed housing, offers the possibility of acquiring patient images in seated
and standing positions. Furthermore, supine, prone and oblique imaging positions are
possible. The position of the head with respect to the X-ray beam and reciprocal sensor in
the mobile CBCT scanner is almost identical to other dental CBCT scanners on the market
in the supine (Figure 2B) and prone (Figure 2C) imaging positions. Therefore, there are no
Figure 5. The image quality and CNR values of tooth 43 (A) and the fossa olfactorius (B) using different imaging positions and protocols.
Figure 6. The figure of merit values of tooth 43 (A) and fossa olfactorius (B) using different imaging positions and protocols.
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
53
4
major image quality differences between mobile CBCT scanners and other dental CBCT
scanners. However, the results found in this study concerning the oblique imaging position
are unique to the mobile CBCT scanner and therefore do not apply to other dental
CBCT scanners.
CBCT image qualitySinus: In the sinus region, the differences between the image qualities acquired using
the supine and oblique imaging positions were negligible. However, the best CBCT mean
image quality was observed in the prone imaging position. When compared to the prone
and oblique imaging positions, the supine imaging position offered only poor image
quality in the fossa olfactorius using 4.8 mA tube current. Furthermore, inadequate image
quality was observed in the fossa olfactorius using the prone and supine imaging positions
and 7.5 mA tube current. In the sinus region, only minor variations in the image quality
were observed when using 4.8 mA and 7.5 mA. However, when using the highest 12 mA
tube current, a significantly better image quality was observed in the sinus area. This
phenomenon could be caused by the better signal-to-noise ratio attained using 12 mA
tube current. Moreover, the better image quality observed in the prone imaging position
could be due to the closer proximity of the cadaver head to the scanner sensor.
The low image quality achieved in the supine imaging position can be explained
by the stronger attenuation and scattering resulting from the thick and complex bony
structures in the anterior part of the skull [32], [33]. Furthermore, the low image quality
observed in the oblique imaging position is likely to be caused by the longer radiation path
length and subsequent attenuation caused by the parietal bones at the top of the cranium
[22], [34]. These findings suggest that it could be advantageous to use the oblique imaging
position instead of the supine imaging position in the sinus region. This hypothesis is in
good agreement with a previous study by Heiland et al. who reported on the advantage
of using the oblique imaging position to image the sinuses on a Siremobil Iso-C3D CBCT
scanner [1]. However it must be noted that an oblique imaging position is not attainable
using a standard upright CBCT scanner.
Ear: In the ear region, the oblique imaging position resulted in poor image quality
of the carotid canal. Furthermore, the supine and prone imaging positions offered only
adequate image quality of the carotid canal. The low image quality observed in the oblique
position could have been caused by the attenuation of the parietal bones situated on
the top of the cranium.
The difficulty of acquiring adequate image quality in the carotid canal regardless of
the tube current (4.8 mA to 12 mA) could be due to the attenuation caused by the dense
bony structures of the temporal bone. These findings are in good agreement with a previous
study by Pereira et al. who reported on the difficulty of obtaining adequate images in
the TMJ region due to the dense bony structures in the petrous part of the temporal bone
[36]. These findings are further supported by the results of Kwong et al. [37] and Sohaib
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
54
4
et al. [38] who concluded that no significant image quality difference occurred when tube
currents were changed.
Maxilla: Using the mobile CBCT scanner, the prone and supine imaging positions
offered better image quality in the maxilla than the oblique imaging position. The prone
imaging position offered the best image quality of tooth 16 and the foramen infraorbitale
using 12 mA. The impact of different tube currents on image quality in all imaging positions
was moderate.
Mandible: In the mandible, the prone and supine imaging positions resulted in good
image quality for all three investigated tube currents. Furthermore, the oblique imaging
position demonstrated an adequate image quality of the apex of tooth 43 and poor image
quality of the foramen mentale using 7.5 mA tube current. However, only poor image
qualities were attainable in the mandible using the oblique imaging position and 4.8 mA
tube current. The impact of different tube currents on the image quality in the prone
imaging position was negligible and moderate in the oblique imaging position. However,
the tube current did not have any effect on image quality in the supine imaging position.
In general, the low image quality observed in this study using the oblique imaging
position is likely due to the anatomy of the head. Moreover, in the oblique position, the x-ray
beam enters the head on the back side of the cranium, and therefore the radiation path
length through the head becomes longer than in the prone position. This phenomenon
subsequently causes stronger attenuation, which has a negative effect on the overall
image quality [17].
Comparison between MSCT and CBCT images acquired in the supine positionThe MSCT, CBCT inter-rater and intra-rater reliabilities were assessed according to
the characterization method decribed by Landis et al. [31] The MSCT intra-rater reliability
was excellent. On the other hand, the CBCT intra-rater reliability was excellent in
the oblique and prone imaging positions and only good in the supine position.
In the supine position, the mobile CBCT scanner offered similar image quality to that
of the MSCT scanner when using 7.5 mA tube current. However, the mobile CBCT scanner
resulted in lower image quality and higher local variations in image quality when a 4.8 mA
tube current was used. Furthermore, when using 12 mA tube current, the mobile CBCT
scanner offered a good delineation of most anatomical structures and better image quality
than those acquired on the MSCT scanner.
Contrast-to-noise ratioIn this study, the best CNR value was observed in the apex of tooth 43 in the prone
imaging position. The lower CNR values observed in the oblique imaging position could
have been caused by the longer radiation path length and the parietal bones on the top
of the cranium. Interestingly, in the oblique imaging position, the delineation of the fossa
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
55
4
olfactorius resulted in a higher CNR value than in the supine imaging position. The lower
CNR value observed in the fossa olfactorius using the supine imaging position can be
explained by the stronger attenuation and scattering from the thick and complex bony
structures in the anterior nasofrontal area of the skull. [39] Furthermore, the correlation
between the CNR value and subjective image quality results was excellent in the apex of
tooth 43. However, only a fair correlation between the CNR value and image quality was
achieved in the fossa olfactorius. These results are in good agreement with a recent study
by Choi et al. [22] who described a strong association between subjective image quality
and CNR values.
Figure of MeritThe prone imaging position offered the best figure of merit value in the fossa olfactorius
and the apex of tooth 43. The differences between the oblique and supine imaging
positions were negligible. However, in the fossa olfactorius, the differences in the FOM
values attained in the oblique and supine imaging positions were significant. This could
have been caused by the highest CNR value attained at a moderate effective dose.
The MSCT scanner resulted in a higher FOM value than the mobile CBCT scanner in
the supine imaging position. However, the figure of merit value obtained using the MSCT
was lower than those acquired using the mobile CBCT scanner in prone and oblique
imaging positions.
The lower figure of merit value obtained on the MSCT scanner can be explained by
the manufacturer-recommended high-resolution imaging parameters that subsequently
resulted in a higher effective dose than in the mobile CBCT scanner. Nevertheless,
Table 4. A comparison between the MSCT and CBCT image qualities acquired in supine imaging positions.
RegionAnatomical landmark
Planmed Verity CBCT
Philips Brilliance 64 MSCT
4.8 mA 7.5 mA 12 mA High Resolution
Sinus Fossa olfact. poor poor good AdequateLamina.papyr. adequate adequate adequate AdequateHiatus max. adequate adequate good AdequateMiddle turb. adequate adequate good Adequate
Ostium adequate adequate good Adequate
Ear Carotid canal poor adequate adequate Adequate
Maxilla Foramen inf. adequate good good Good D 16 apex adequate good good Adequate
Mandible D 43 apex good good good Adequate Foramen.ment. good good good Good
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
56
4
the effective dose can be significantly reduced by using a lower mAs value that subsequently
does not markedly affect the diagnostic image quality in the sinus region [38].
One weakness of this study is the use of a frozen cadaver head for image quality
assessment During freezing residual water in the cadavar expands and subequently causes
anatomical alterations especially in the fine bony structures of the sino-nasal region.
Furthermore, frozen fatty acids in the brain [40] decrease in volume and consequently
increase the attenuation resulting in an increase of the HU- or the grey-values [40], [41].
The aforementioned factors subsequently reduce the image quality and contrast resolution
[42]. Moreover, the figure of merit value was assessed based on the CNR value from
a cadaver head while the effective dose used for calculating the FOM value was obtained
using an anthropomorphic phantom. Another drawback of this study is that the image
quality was assessed for only one anatomical landmark in the ear region while there is
a growing interest for diagnostic imaging of temporal bones due to the cochlear implants
[43]. Furthermore, an interpretation with the results obtained using the MSCT scanner
should be carried out with care since the images were acquired using standard exposure
parameters without image quality optimisation.
CONCLUSIONSHead imaging positioning has an impact on the overall CBCT image quality. The best
mean CBCT image quality, CNR value and FOM value were observed using the prone
imaging position for all diagnostic regions. The oblique imaging position offers inadequate
image quality except in the sinus region. The MSCT scanner offered similar mean image
quality in the supine position to that acquired on the mobile CBCT scanner using 7.5 mA
tube current.
CONFLICTS OF INTEREST Juha Koivisto is an employee of Planmeca Oy.
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
57
4
REFERENCES1. Heiland M, Schulze D, Blake F, and
Schmelzle R (2005) Intraoperative imaging of zygomaticomaxillary complex fractures using a 3D C-arm system. Int. J. Oral Maxillofac. Surg. 34:369–375
2. Koskinen S K, Haapamaki V V, Salo J, Lindfors N C, Kortesniemi M, Seppala L, and Mattila K T (2013) CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT). Skeletal Radiol. 42:649–657
3. Tuominen E K J, Kankare J, Koskinen S K, and Mattila K T (2013) Weight-bearing CT imaging of the lower extremity. Am. J. Roentgenol. 200:146–148
4. Zbijewski W and Stayman J W (2009) Volumetric soft tissue brain imaging on xCAT, a mobile flat-panel x-ray CT system. Proc. SPIE Med. Imaging. 7258:72582K–72582K–5
5. Zhang J, Weir V, Fajardo L, Lin J, Hsiung H, and Ritenour E R (2009) Dosimetric characterization of a cone-beam O-arm imaging system. J. Xray. Sci. Technol. 17:305–317
6. Siewerdsen J H, Daly M J, Bachar G, Moseley D J, Bootsma G, Brock K K, Ansell S, Wilson G A, Chhabra S, Jaffray D A, and Irish J C (2007) Multimode C-arm fluoroscopy, tomosynthesis, and cone-beam CT for image-guided interventions: from proof of principle to patient protocols. Proc. SPIE. 6510: 65101A–65101A–11
7. Siewerdsen J H, Daly M J, Chan H, Nithiananthan S, Hamming N, Brock K K, and Irish J C (2009) High-performance intraoperative cone-beam CT on a mobile C-arm: an integrated system for guidance of head and neck surgery. Proc. SPIE. 7261:72610J–72610J–8
8. Siewerdsen J H (2011) Cone-Beam CT with a Flat-Panel Detector: From Image Science to Image-Guided Surgery. Nucl. Instrum. Methods Phys. Res. A. 648:S241–S250
9. Schafer S, Nithiananthan S, Mirota D J, Uneri A, Stayman J W, Zbijewski W, Schmidgunst C, Kleinszig G, Khanna A J, and Siewerdsena J H (2011) Mobile C-arm cone-beam CT for guidance of spine surgery: image quality, radiation dose, and integration with interventional guidance. Med. Phys. 38:4563–4574
10. Siewerdsen J H, Chan Y, Rafferty M A, Moseley D J, Jaffray D A, and Irish J C (2005) Cone-beam CT with a flat-panel detector on a mobile C-arm: preclinical investigation in image-guided surgery of the head and neck. Proc. SPIE. 5744:789–797
11. Loubele M, Bogaerts R, Van Dijck E, Pauwels R, Vanheusden S, Suetens P, Marchal G, Sanderink G, and Jacobs R (2009) Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications. Eur. J. Radiol. 71:461–468
12. Ludlow J B and Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surgery, Oral Med. Oral Pathol. Oral Radiol. Endodontology. 106:106–114
13. Nemtoi A, Czink C, Haba D, and Gahleitner A (2013) Cone beam CT: a current overview of devices. Dentomaxillofac. Radiol. 42:20120443
14. Kim S, Yoo S, Yin F-F, Samei E, and Yoshizumi T (2010) Kilovoltage cone-beam CT: comparative dose and image quality evaluations in partial and full-angle scan protocols. Med. Phys. 37:3648–3659
15. Fahrig R, Dixon R, Payne T, Morin R L, Ganguly A, and Strobel N (2006) Dose and image quality for a cone-beam C-arm CT system. Med. Phys. 33:4541–4550
16. Lofthag-Hansen S, Thilander-Klang A, and Grondahl K (2011) Evaluation of subjective image quality in relation to diagnostic task for cone beam computed
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
58
4
tomography with different fields of view. Eur. J. Radiol. 80:483–488
17. Koivisto J, Wolff J, Järnstedt J, Dastidar P, and Kortesniemi M (2014) Assessment of the effective dose in supine, prone, and oblique positions in the maxillofacial region using a novel combined extremity and maxillofacial cone beam computed tomography scanner. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 118:355–362
18. Yan H, Cervino L, Jia X, and Jiang S B (2012) A comprehensive study on the relationship between the image quality and imaging dose in low-dose cone beam CT. Phys. Med. Biol. 57:2063–2080
19. Heiland M, Schulze D, Adam G, and Schmelzle R (2003) 3D-imaging of the facial skeleton with an isocentric mobile C-arm system (Siremobil Iso-C3D). Dentomaxillofac. Radiol. 32:21–25
20. Schulze D, Heiland M, Blake F, Rother U, and Schmelzle R (2005) Evaluation of quality of reformatted images from two cone-beam computed tomographic systems. J. cranio-maxillo-facial Surg. 33:19–23
21. Kamburoglu K, Murat S, Kolsuz E, Kurt H, Yuksel S, and Paksoy C (2011) Comparative assessment of subjective image quality of cross-sectional cone-beam computed tomography scans. J. Oral Sci. 53:501–508
22. Choi J-W, Lee S-S, Choi S-C, Heo M-S, Huh K-H, Yi W-J, Kang S-R, Han D-H, and Kim E-K (2015) Relationship between physical factors and subjective image quality of cone-beam computed tomography images according to diagnostic task. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 119:357–365
23. Yu L, Vrieze T J, Bruesewitz M R, Kofler J M, DeLone D R, Pallanch J F, Lindell E P, and McCollough C H (2010) Dose and image quality evaluation of a dedicated cone-beam CT system for high-contrast neurologic applications. AJR. Am. J. Roentgenol. 194:W193-201
24. Martin C J (2007) The importance of radiation quality for optimisation in radiology. Biomedical Imaging and Intervention Journal, 3:e38
25. Zamenhof R G (1982) The optimization of signal detectability in digital fluoroscopy. Med. Phys. 9:688–694
26. Ogden K M, Huda W, Khorasani M R, and Scalzetti E M (2008) A comparison of three CT voltage optimization strategies. Proc. SPIE. 691310–691351
27. Boone J M (1992) Parametrized x-ray absorption in diagnostic radiology from Monte Carlo calculations: implications for x-ray detector design. Med. Phys. 19:1467–1473
28. World Medical Association (2013) World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA 310:2191–2194
29. Prather H, Hunt D, Steger-May K, Hayes M H, Knaus E, and Clohisy J (2009) Inter-rater reliability of three musculoskeletal physical examination techniques used to assess motion in three planes while standing. PM R 1:629–635
30. Shrout P E and Fleiss J L (1979) Intraclass correlations: uses in assessing rater reliability. Psychol. Bull. 86:420–428
31. Landis J R and Kock G G (1977) The measurement of observer agreement for categorical data. Biometriscs 33:159–174
32. Gupta N and Garg N K (2014) Review on different techniques of image registration. Int. J. Eng. Comput. Sci. ISSN. 2319–7242
33. Wang C-W and Chen H-C (2013) Improved image alignment method in application to X-ray images and biological images. Bioinformatics 29:1879–1887
34. Korsager A S, Carl J, and Ostergaard L R (2016) Comparison of manual and automatic MR-CT registration for radiotherapy of prostate cancer. J. Appl. Clin. Med. Phys. 17:294–303
IMPA
CT O
F PRO
NE
, SUPIN
E A
ND
OB
LIQU
E PA
TIEN
T POSITIO
NIN
G O
N C
BC
T IMA
GE
QU
ALITY
59
4
35. Pauwels R, Araki K, Siewerdsen J H, and Thongvigitmanee S S (2015) Technical aspects of dental CBCT: state of the art. Dentomaxillofacial Radiol. 44:20140224
36. Pereira A M, Batigália F, Santos F S, Martins A T, and Lyra A O (2012) Clinical application of Cone-beam tomography in disorders of the head of mandible. J. Morphol 29:151–153
37. Kwong J C, Palomo J M, Landers M A, Figueroa A, and Hans M G (2008) Image quality produced by different cone-beam computed tomography settings. Am. J. Orthod. Dentofac. Orthop. 133:317–327
38. Sohaib S A, Peppercorn P D, Horrocks J A, Keene M H, Kenyon G S, and Reznek R H (2001) The effect of decreasing mAs on image quality and patient dose in sinus CT. Br. J. Radiol. 74:157–161
39. Timperley D G, Banks C, Robinson D, Roth J, Sacks R, and Harvey R J (2011) Lateral frontal sinus access in endoscopic
skull-base surgery. Int. Forum Allergy Rhinol. 1:290–295
40. Chang C-Y, Ke D-S, and Chen J-Y (2009) Essential fatty acids and human brain. Acta Neurol. Taiwan. 18:231–241
41. Pech P, Bergstrom K, Rauschning W, and Haughton V M (1987) Attenuation values, volume changes and artifacts in tissue due to freezing. Acta Radiol. 28:779–782
42. Hemmingsson A, Johansson A, and Rauschning W (1982) Attenuation in human muscle and fat tissue in vivo and in vitro. Acta Radiol. Diagn. 23:149–151
43. Zou J, Koivisto J, Lähelmä J, Aarnisalo A, Wolff J, and Pyykkö I (2015) Imaging Optimization of Temporal Bones With Cochlear Implant Using a High-resolution Cone Beam CT and the Corresponding Effective Dose. Ann. Otol. Rhinol. Laryngol. 124:466–473
5 CT IMAGE SEGMENTATION METHODS FOR BONE USED IN MEDICAL ADDITIVE MANUFACTURING:
A LITERATURE REVIEW
Maureen van Eijnatten, Roelof van Dijk, Johannes Dobbe,
Geert Streekstra, Juha Koivisto, Jan Wolff
Medical Engineering & Physics [accepted Sept 2017]
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
62
5
ABSTRACTAim of the studyThe accuracy of additive manufactured medical constructs is limited by errors introduced
during image segmentation. The aim of this study was to review the existing literature on
different image segmentation methods used in medical additive manufacturing.
MethodsThirty-two publications that reported on the accuracy of bone segmentation based on
computed tomography images were identified using PubMed, ScienceDirect, Scopus,
and Google Scholar. The advantages and disadvantages of the different segmentation
methods used in these studies were evaluated and reported accuracies were compared.
ResultsThe spread between the reported accuracies was large (0.04 mm to 1.9 mm). Global
thresholding was the most commonly used segmentation method with accuracies under
0.6 mm. The disadvantage of this method is the extensive manual post-processing
required. Advanced thresholding methods could improve the accuracy to under 0.38
mm. However, such methods are currently not included in commercial software packages.
Statistical shape model methods resulted in accuracies from 0.25 mm to 1.9 mm but are
only suitable for anatomical structures with moderate anatomical variations.
ConclusionsThresholding remains the most widely used segmentation method in medical additive
manufacturing. To improve the accuracy and reduce the costs of patient-specific additive
manufactured constructs, more advanced segmentation methods are required.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
63
5
INTRODUCTIONAdditive manufacturing (AM), also referred to as three-dimensional (3D) printing, is
becoming increasingly popular in medicine [1] since it offers the possibility to personalize
patient care. [2] The use of AM anatomical models results in more precise treatment planning,
better communication, [3,4] and improved training and education. [5,6] Furthermore, AM
can be used for the fabrication of drill guides, [7] saw guides, [8] and patient-specific
implants. [9] To date, medical AM is most commonly used in branches of surgery involving
the musculoskeletal system, such as oral and maxillofacial surgery, traumatology, and
orthopaedic surgery. However, it must be noted that the overall accuracy and repeatability
of medical AM constructs used for bone reconstruction still need to be improved. [10]
In this context, a recent systematic review by Martelli et al. (2016) identified 34 studies
(21.5%) that reported on the unsatisfactory accuracy of medical AM constructs. [11]
The current medical AM process used for the reconstruction of the musculoskeletal
system can be divided into four basic steps: imaging (1); image processing (2), optionally
followed by computer-aided design (3); and additive manufacturing (4). [12] Each of these
steps can introduce geometric deviations that can cause distortions in the resulting medical
AM constructs. [13] Recent studies, however, suggest that the majority of the inaccuracies
are introduced during imaging (Figure 1: step 1) and image processing (Figure 1:
step 2), rather than during the manufacturing, i.e., the 3D printing process, which is
generally considered to be precise. [11,14,15]
Step 1: ImagingCT scanners are best suited for imaging bony structures due to their superior hard tissue
contrast and spatial resolution. [16] Today, a plethora of different CT technologies are
available, ranging from single, helical CT to 128-slice dual-source CT configurations.
Cone-beam computed tomography (CBCT) scanners are becoming increasingly popular in
orthopaedic [17] and maxillofacial surgery [18] due to their lower radiation dose and costs.
Raw CT data acquired during image acquisition is commonly reconstructed as a Digital
Imaging and Communications in Medicine (DICOM) file.
One major challenge faced in medical AM is the large variety of different CT image
acquisition and reconstruction parameters currently available (see Figure 1; step 1 A&B).
To date, to the best of our knowledge, there are no standardized protocols available for
medical AM. Image slice thickness and slice interval have been identified as the primary
limiting factors for the overall accuracy of medical AM constructs, [19] especially when
reconstructing thin bony structures from axial plane images, such as the orbital floor, [20]
or where the imaging plane is nearly parallel to the bone surface to be reconstructed, such
as in the tibial plateau. Moreover, imaging noise, beam hardening, patient motion, and
metal artifacts can introduce inhomogeneities in the gray values of CT images.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
64
5
Step 2: Image processingThe medical AM process always requires image processing: the conversion of CT
images into 3D surface models (Figure 1, step 2). Such 3D surface models can be saved
in a wide range of different file formats. [21] Currently, the most commonly used file
format in medical AM is standard tessellation language (STL). Although rarely used, it is
theoretically possible to convert CT images into slice formats that can subsequently be
used for AM, thereby skipping the STL conversion process. However, it must be noted that
the computer-aided design (CAD) software packages currently available on the market for
medical AM still require STL files (Figure 1, step 3).
The conversion of DICOM datasets into 3D surface models has been identified as
a major source of inaccuracies in medical AM. [14,22,23] The most important step in this
conversion process is image segmentation (Figure 1, step 2A). [24] Segmentation refers
to the partitioning of images into regions of interest (ROIs) that correspond to anatomical
structures. At present, there are a plethora of different image segmentation methods
available for bony structures. [25] It remains unclear, however, which segmentation method
offers the most accurate 3D surface models. Therefore, the first aim of this study was to
review the existing literature on the different CT image segmentation methods currently
being used for bone segmentation in medical AM. The second aim was to evaluate
the impact of the different image segmentation methods on the geometric accuracy of
3D surface models.
MATERIALS AND METHODSExisting literature on the CT image segmentation of bony structures for medical AM
applications was reviewed using the PubMed Medline literature database, ScienceDirect,
Scopus, and Google Scholar. An initial database of 17,700 publications was generated
using the search terms “medical AND X” + “medicine AND X”, with X representing
the interchangeably used terms “additive manufacturing”, “rapid prototyping”, and “3D
printing”. The acquired database was subsequently filtered using the search terms “bone”
(7,630), “segmentation” (281), and “accuracy” (138). In order to acquire comparable
results, the following inclusion and exclusion criteria were used to isolate the publications
of interest to this study:
Inclusion criteria:
• Evaluates the geometric accuracy of 3D surface models of bone derived from CT
images
• Reports accuracy using a mean spatial distance-based metric in millimeters
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
65
5
Exclusion criteria:
• Uses non-biological materials or artificial phantoms
• Reports accuracy using non-spatial distance-based metrics, such as relative distance or
volume difference (%), maximum distance, 95%-value of all distances (95th percentile),
overlap-based metrics, and contour-based metrics.
• Uses a non-CT-based imaging technology, e.g., magnetic resonance imaging (MRI)
• Only reports on the accuracy of the final AM model and not on the preceding 3D
surface model
Figure 1. Overview of the parameters that can influence the accuracy of medical AM constructs. * This review focuses on the different CT image segmentation methods used in medical AM.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
66
5
Using the aforementioned inclusion and exclusion criteria, 10 publications were
identified that contained information on the geometric accuracy of 3D surface models
of bone acquired from CT images that were subsequently used for medical AM. From
a parallel search, 22 additional publications were identified that reported on the accuracy
of 3D surface models but did not focus on AM applications. Additional information was
collected from four review articles focusing on medical AM. [19,26–28]
All relevant details of the publications included in this study are summarized in
Table 1. The publications were subsequently sorted based on the type of image
segmentation method used in the study. The reported accuracies and standard deviations
(if available) are summarized in Table 2. In cases where multiple image segmentation
methods, software packages, anatomical structures, or CT modalities were evaluated, their
recorded accuracies are presented separately in Table 2. When multiple measurements
were performed using the same image segmentation method or anatomical structure,
the mean value and standard deviation of the reported accuracies are presented as one
entry in Table 2.
RESULTSA total of 32 publications were included in this study (Table 1). These publications
were published between 2002 and 2017. To the best of our knowledge, no papers on
the accuracy of medical AM constructs were published before 2002. [29] However, it must
be noted that the first publications on medical AM date back as far as the early 1990s. [30]
CT image segmentation methods used in medical AMThe reviewed publications reported that the following three semi-automated CT image
segmentation methods are often used in medical AM: global thresholding, [31] edge
detection, [32] and region growing. [33]
Of the image segmentation methods mentioned above, global thresholding is the most
commonly used bone segmentation method in medical AM. In global thresholding,
a single threshold value t for bone is manually [34] or automatically [31,35] selected. All
voxels with a gray value equal or greater than t are included in a segmented volume using
a binary mask Mx,y (Equation 1):
(1)
where Ix,y denotes the grey value at coordinates x and y in a CT image slice.
Global thresholding has certain drawbacks, namely, voxels reside on tissue boundaries
that contain more than one tissue type and induce a blurring of gray values across
boundaries. This phenomenon is referred to as the partial volume effect (PVE). As
8
Results A total of 32 publications were included in this study (Table 1). These publications were published
between 2002 and 2017. To the best of our knowledge, no papers on the accuracy of medical AM
constructs were published before 2002. [29] However, it must be noted that the first publications
on medical AM date back as far as the early 1990s. [30]
CT image segmentation methods used in medical AM
The reviewed publications reported that the following three semi-automated CT image
segmentation methods are often used in medical AM: global thresholding, [31] edge detection, [32]
and region growing. [33]
Of the image segmentation methods mentioned above, global thresholding is the most commonly
used bone segmentation method in medical AM. In global thresholding, a single threshold value t
for bone is manually [34] or automatically [31,35] selected. All voxels with a gray value equal or
greater than t are included in a segmented volume using a binary mask Mx,y (Equation 1):
M�,� � �1I�,� � �0I�,� � �,�1�
where Ix,y denotes the grey value at coordinates x and y in a CT image slice.
Global thresholding has certain drawbacks, namely, voxels reside on tissue boundaries that contain
more than one tissue type and induce a blurring of gray values across boundaries. This phenomenon
is referred to as the partial volume effect (PVE). As a consequence of the PVE, precise delineation
of tissue boundaries using only a single threshold value remains difficult and can result in an over-
or underestimation of the region of interest. Furthermore, a single threshold value does not take
artifacts, image noise, photon starvation, or variations in gray values between different CT scanners
and protocols into account, which can lead to inconsistent segmentation results. These drawbacks
have led to the development of more sophisticated thresholding methods such as local thresholding.
Local thresholding, also referred to as multi-level thresholding, divides an image into multiple
ROIs for which an individual thresholding bandwidth tk to tk+1 can be selected. [36] All voxels with
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
67
5
a consequence of the PVE, precise delineation of tissue boundaries using only a single
threshold value remains difficult and can result in an over- or underestimation of the region
of interest. Furthermore, a single threshold value does not take artifacts, image noise,
photon starvation, or variations in gray values between different CT scanners and protocols
into account, which can lead to inconsistent segmentation results. These drawbacks
have led to the development of more sophisticated thresholding methods such as
local thresholding.
Local thresholding, also referred to as multi-level thresholding, divides an image into
multiple ROIs for which an individual thresholding bandwidth tk to tk+1 can be selected. [36]
All voxels with a grey value between tk and tk+1 are included in k segmented volumes using
a binary mask Mx,y (Equation 2):
(2)
where k denotes the kth grey value band, of which tk is the lower limit and tk+1 the upper
limit. Furthermore, 3D adaptive thresholding algorithms have recently been developed
that offer the possibility to iteratively update the voxel classification in an additional
second step after global thresholding. [37]
Edge detection, on the other hand, identifies local edges on CT images by calculating
gray value gradients (derivatives). The gradients with a magnitude that is higher than
a chosen threshold value are defined as edges. A range of edge detection operators are
currently available that includes Sobel, Laplacian (2nd derivative) and Canny. [25] Canny edge
detection remains one of the most commonly used, fastest and most accurate operators.
[36] Edge detection is well suited for the segmentation of structures with different contrast
in different regions, such as long bones. However, local gray value changes induced by
noise and (metal) artifacts are often mistakenly identified as edges. It should also be noted
that, without further processing, edge detection methods do not necessarily segment all
bone voxels in the image and thus need to be combined with other methods, such as
region growing.
In region growing, a specific voxel is manually selected as a seed point for a specific
tissue type. Subsequently, the gray values of the neighboring voxels are compared to
the gray value of the seed point. Voxels that meet the predefined homogeneity criteria are
labeled and grouped together, thereby creating a region Ri of connected voxels in the CT
image I:
9
a grey value between tk and tk+1 are included in k segmented volumes using a binary mask Mx,y
(Equation 2):
M�,� �
��������,� � ����� � ��,� � ����� � ��,� � ����� � ��,� � �������,� � ����
,���
where k denotes the kth grey value band, of which tk is the lower limit and tk+1 the upper limit.
Furthermore, 3D adaptive thresholding algorithms have recently been developed that offer the
possibility to iteratively update the voxel classification in an additional second step after global
thresholding. [37]
Edge detection, on the other hand, identifies local edges on CT images by calculating gray value
gradients (derivatives). The gradients with a magnitude that is higher than a chosen threshold value
are defined as edges. A range of edge detection operators are currently available that includes
Sobel, Laplacian (2nd derivative) and Canny. [25] Canny edge detection remains one of the most
commonly used, fastest and most accurate operators. [36] Edge detection is well suited for the
segmentation of structures with different contrast in different regions, such as long bones.
However, local gray value changes induced by noise and (metal) artifacts are often mistakenly
identified as edges. It should also be noted that, without further processing, edge detection methods
do not necessarily segment all bone voxels in the image and thus need to be combined with other
methods, such as region growing.
In region growing, a specific voxel is manually selected as a seed point for a specific tissue type.
Subsequently, the gray values of the neighboring voxels are compared to the gray value of the seed
point. Voxels that meet the predefined homogeneity criteria are labeled and grouped together,
thereby creating a region Ri of connected voxels in the CT image I:
�� � �� � �� � �� �� � ��3�
In practice, region growing is seldom the only segmentation method used, but it is commonly
combined with other methods such as (global) thresholding. An advantage of region growing is
that it discards voxels that are not connected to the anatomical structure of interest, resulting in a
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
68
5
(3)
In practice, region growing is seldom the only segmentation method used, but it is
commonly combined with other methods such as (global) thresholding. An advantage of
region growing is that it discards voxels that are not connected to the anatomical structure
of interest, resulting in a shorter 3D printing time and material savings. A disadvantage
of region growing is that each separate bony structure requires an individual, manually
placed seed point. Furthermore, noise and the partial volume effect can cause voids or
erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been
developed. [38] The most established and validated advanced image segmentation
method to date is based on statistical shape models (SSM). SSM-based methods generate
a statistical shape model of an anatomical ROI from a training dataset. This training
dataset comprises a large number of fully segmented CT images, which are averaged to
form a mean anatomical shape model, along with common modes of variation and their
probabilities. The model given by:
(4)
describes every possible shape
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
as a linear combination of k eigenmodes
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
with
weights
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
, where
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
denotes the mean shape model. This model can be fitted to a new,
unsegmented image dataset, and the segmentation in terms of weights
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
then forms an
approximation of the segmented shape.[39]
The major challenge faced in SSM-based segmentation is the large number of
uniform training datasets required. Another drawback with SSM-based methods is that
they are prone to failure if the datasets differ strongly from the training shapes. In clinical
settings, such differences can occur, for example, in fractured bony structures that no
longer correspond to the initial anatomical shape. Another example is that an SSM trained
on a European population may not account for some of the morphological differences
present in an African population.
Accuracy of CT image segmentation methodsIn all the reviewed publications, the accuracy was assessed by comparing the CT-derived
3D surface models to a “ground truth” (Table 1). The ground truth was obtained either by
performing manual segmentation by an expert, by obtaining a laser surface scan or optical
3D surface scan from the dry or dissected bone, or by performing linear measurements on
the dry or dissected bone using calipers or a coordinate measurement machine (CMM).
In all publications that focused on SSM-based segmentation methods, the accuracy was
9
a grey value between tk and tk+1 are included in k segmented volumes using a binary mask Mx,y
(Equation 2):
M�,� �
��������,� � ����� � ��,� � ����� � ��,� � ����� � ��,� � �������,� � ����
,���
where k denotes the kth grey value band, of which tk is the lower limit and tk+1 the upper limit.
Furthermore, 3D adaptive thresholding algorithms have recently been developed that offer the
possibility to iteratively update the voxel classification in an additional second step after global
thresholding. [37]
Edge detection, on the other hand, identifies local edges on CT images by calculating gray value
gradients (derivatives). The gradients with a magnitude that is higher than a chosen threshold value
are defined as edges. A range of edge detection operators are currently available that includes
Sobel, Laplacian (2nd derivative) and Canny. [25] Canny edge detection remains one of the most
commonly used, fastest and most accurate operators. [36] Edge detection is well suited for the
segmentation of structures with different contrast in different regions, such as long bones.
However, local gray value changes induced by noise and (metal) artifacts are often mistakenly
identified as edges. It should also be noted that, without further processing, edge detection methods
do not necessarily segment all bone voxels in the image and thus need to be combined with other
methods, such as region growing.
In region growing, a specific voxel is manually selected as a seed point for a specific tissue type.
Subsequently, the gray values of the neighboring voxels are compared to the gray value of the seed
point. Voxels that meet the predefined homogeneity criteria are labeled and grouped together,
thereby creating a region Ri of connected voxels in the CT image I:
�� � �� � �� � �� �� � ��3�
In practice, region growing is seldom the only segmentation method used, but it is commonly
combined with other methods such as (global) thresholding. An advantage of region growing is
that it discards voxels that are not connected to the anatomical structure of interest, resulting in a
10
shorter 3D printing time and material savings. A disadvantage of region growing is that each
separate bony structure requires an individual, manually placed seed point. Furthermore, noise and
the partial volume effect can cause voids or erroneously connected structures in a segmented image.
In recent years, a number of advanced CT image segmentation methods have been developed. [38]
The most established and validated advanced image segmentation method to date is based on
statistical shape models (SSM). SSM-based methods generate a statistical shape model of an
anatomical ROI from a training dataset. This training dataset comprises a large number of fully
segmented CT images, which are averaged to form a mean anatomical shape model, along with
common modes of variation and their probabilities. The model given by:
��������� � �� � ∑ ��� ��, (4)
describes every possible shape �� as a linear combination of k eigenmodes �� with weights ���, where
�� denotes the mean shape model. This model can be fitted to a new, unsegmented image dataset,
and the segmentation in terms of weights ��� then forms an approximation of the segmented
shape.[39]
The major challenge faced in SSM-based segmentation is the large number of uniform training
datasets required. Another drawback with SSM-based methods is that they are prone to failure if
the datasets differ strongly from the training shapes. In clinical settings, such differences can occur,
for example, in fractured bony structures that no longer correspond to the initial anatomical shape.
Another example is that an SSM trained on a European population may not account for some of
the morphological differences present in an African population.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
69
5
evaluated using leave-one-out cross validation, in which the distances were measured
from the omitted training shape to the closest match of the “ground-truth” shape model
trained on all other available shapes.
In total, 22 of the reviewed publications (67%) evaluated the accuracy by calculating
the mean absolute distance between a CT-derived 3D surface model and a geometrically
aligned ground truth 3D surface model. Let us define the distance d(p,q) as the distance
between a point p belonging to the surface of a 3D surface model P and its closest point q
on the ground truth model Q. The mean surface distance (MSD) between the two surfaces
P and Q can then be defined as follows:
(5)
where |P| denotes the surface area of P. [40] Note that some variations in the definition
of d(p,q) were found in the reviewed publications. For example, Smith et al. (2013) used
perpendicular projections of the distance of each point p along the triangle normal vector,
[14] whilst Gassman et al. (2008) created a distance map based on the Euclidean distance,
or the shortest distance from each point p to the target surface Q. [41] Finally, Zhang et
al. [42] and Rajamani et al. [43] used the Hausdorff distance (HD), which can be defined
as follows:
(6)
Two publications [44,45] used the root-mean-square distance (RMSD) that can be derived
from Equation 5:
(7)
In four publications, [29,46–48] the mean absolute distance (MAD) was calculated by
performing a limited number of linear measurements n, which can be written as follows:
(8)
11
Accuracy of CT image segmentation methods In all the reviewed publications, the accuracy was assessed by comparing the CT-derived 3D
surface models to a “ground truth” (Table 1). The ground truth was obtained either by performing
manual segmentation by an expert, by obtaining a laser surface scan or optical 3D surface scan
from the dry or dissected bone, or by performing linear measurements on the dry or dissected bone
using calipers or a coordinate measurement machine (CMM). In all publications that focused on
SSM-based segmentation methods, the accuracy was evaluated using leave-one-out cross
validation, in which the distances were measured from the omitted training shape to the closest
match of the “ground-truth” shape model trained on all other available shapes.
In total, 22 of the reviewed publications (67%) evaluated the accuracy by calculating the mean
absolute distance between a CT-derived 3D surface model and a geometrically aligned ground truth
3D surface model. Let us define the distance d(p,q) as the distance between a point p belonging to
the surface of a 3D surface model P and its closest point q on the ground truth model Q. The mean
surface distance (MSD) between the two surfaces P and Q can then be defined as follows:
��� � 1|�|� ���, �������
,�5�
where |P| denotes the surface area of P. [40] Note that some variations in the definition of d(p,q)
were found in the reviewed publications. For example, Smith et al. (2013) used perpendicular
projections of the distance of each point p along the triangle normal vector, [14] whilst Gassman
et al. (2008) created a distance map based on the Euclidean distance, or the shortest distance from
each point p to the target surface Q. [41] Finally, Zhang et al. [42] and Rajamani et al. [43] used
the Hausdorff distance (HD), which can be defined as follows:
����, �� � ������ ������� ����, �������
Two publications [44,45] used the root-mean-square distance (RMSD) that can be derived from
Equation 5:
���� � � 1|�|� ���, �����
����7�
11
Accuracy of CT image segmentation methods In all the reviewed publications, the accuracy was assessed by comparing the CT-derived 3D
surface models to a “ground truth” (Table 1). The ground truth was obtained either by performing
manual segmentation by an expert, by obtaining a laser surface scan or optical 3D surface scan
from the dry or dissected bone, or by performing linear measurements on the dry or dissected bone
using calipers or a coordinate measurement machine (CMM). In all publications that focused on
SSM-based segmentation methods, the accuracy was evaluated using leave-one-out cross
validation, in which the distances were measured from the omitted training shape to the closest
match of the “ground-truth” shape model trained on all other available shapes.
In total, 22 of the reviewed publications (67%) evaluated the accuracy by calculating the mean
absolute distance between a CT-derived 3D surface model and a geometrically aligned ground truth
3D surface model. Let us define the distance d(p,q) as the distance between a point p belonging to
the surface of a 3D surface model P and its closest point q on the ground truth model Q. The mean
surface distance (MSD) between the two surfaces P and Q can then be defined as follows:
��� � 1|�|� ���, �������
,�5�
where |P| denotes the surface area of P. [40] Note that some variations in the definition of d(p,q)
were found in the reviewed publications. For example, Smith et al. (2013) used perpendicular
projections of the distance of each point p along the triangle normal vector, [14] whilst Gassman
et al. (2008) created a distance map based on the Euclidean distance, or the shortest distance from
each point p to the target surface Q. [41] Finally, Zhang et al. [42] and Rajamani et al. [43] used
the Hausdorff distance (HD), which can be defined as follows:
����, �� � ������ ������� ����, �������
Two publications [44,45] used the root-mean-square distance (RMSD) that can be derived from
Equation 5:
���� � � 1|�|� ���, �����
����7�
11
Accuracy of CT image segmentation methods In all the reviewed publications, the accuracy was assessed by comparing the CT-derived 3D
surface models to a “ground truth” (Table 1). The ground truth was obtained either by performing
manual segmentation by an expert, by obtaining a laser surface scan or optical 3D surface scan
from the dry or dissected bone, or by performing linear measurements on the dry or dissected bone
using calipers or a coordinate measurement machine (CMM). In all publications that focused on
SSM-based segmentation methods, the accuracy was evaluated using leave-one-out cross
validation, in which the distances were measured from the omitted training shape to the closest
match of the “ground-truth” shape model trained on all other available shapes.
In total, 22 of the reviewed publications (67%) evaluated the accuracy by calculating the mean
absolute distance between a CT-derived 3D surface model and a geometrically aligned ground truth
3D surface model. Let us define the distance d(p,q) as the distance between a point p belonging to
the surface of a 3D surface model P and its closest point q on the ground truth model Q. The mean
surface distance (MSD) between the two surfaces P and Q can then be defined as follows:
��� � 1|�|� ���, �������
,�5�
where |P| denotes the surface area of P. [40] Note that some variations in the definition of d(p,q)
were found in the reviewed publications. For example, Smith et al. (2013) used perpendicular
projections of the distance of each point p along the triangle normal vector, [14] whilst Gassman
et al. (2008) created a distance map based on the Euclidean distance, or the shortest distance from
each point p to the target surface Q. [41] Finally, Zhang et al. [42] and Rajamani et al. [43] used
the Hausdorff distance (HD), which can be defined as follows:
����, �� � ������ ������� ����, �������
Two publications [44,45] used the root-mean-square distance (RMSD) that can be derived from
Equation 5:
���� � � 1|�|� ���, �����
����7�
12
In four publications, [29,46–48] the mean absolute distance (MAD) was calculated by performing
a limited number of linear measurements n, which can be written as follows:
��� � 1��|�� � ��|
�,���
where D is a linear measurement performed on the ground truth, and x is a measurement performed
on the CT-derived 3D surface model. Additionally, in four publications, [49–52] the mean distance
(MD) was calculated as the signed mean of n measurements between different anatomical
landmarks, [49,52] reference points, [51] or photographs of a cryosectioned cadaver, [50] which
can be described as follows:
�� � 1����� � ���
����
Table 2 summarizes the geometric accuracies reported in the publications. All 3D surface model
accuracies generally varied between 0.04 mm and 1.9 mm. Manual segmentation resulted in an
accuracy of 0.20 mm ± 0.15 mm to 0.48 mm ± 0.51 mm. [53,54] In global thresholding, accuracies
varied between 0.04 mm ± 0.59 mm and 0.62 mm ± 0.76 mm. [14,29,36,46,48,49,52,55–59]
However, it must be noted that most authors reported that additional manual editing was performed.
Advanced thresholding methods resulted in accuracies below 0.38 mm ± 0.14 mm.
[36,37,42,45,46] SSM-based segmentation methods achieved accuracies ranging from 0.25 mm ±
0.20 mm to 1.9 mm ± 0.20 mm. [44,60–68] Finally, the publications on more experimental
segmentation methods (“other” in Table 2) reported geometric accuracies ranging between 0.28
mm ± 0.04 mm and 1.27 mm ± 0.92 mm. [41,47,50,51,59,69,70]
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
70
5
where D is a linear measurement performed on the ground truth, and x is a measurement
performed on the CT-derived 3D surface model. Additionally, in four publications,
[49–52] the mean distance (MD) was calculated as the signed mean of n measurements
between different anatomical landmarks, [49,52] reference points, [51] or photographs of
a cryosectioned cadaver, [50] which can be described as follows:
(9)
Table 2 summarizes the geometric accuracies reported in the publications. All 3D surface
model accuracies generally varied between 0.04 mm and 1.9 mm. Manual segmentation
resulted in an accuracy of 0.20 mm ± 0.15 mm to 0.48 mm ± 0.51 mm. [53,54] In global
thresholding, accuracies varied between 0.04 mm ± 0.59 mm and 0.62 mm ± 0.76 mm.
[14,29,36,46,48,49,52,55–59] However, it must be noted that most authors reported that
additional manual editing was performed. Advanced thresholding methods resulted in
accuracies below 0.38 mm ± 0.14 mm. [36,37,42,45,46] SSM-based segmentation methods
achieved accuracies ranging from 0.25 mm ± 0.20 mm to 1.9 mm ± 0.20 mm. [44,60–68]
Finally, the publications on more experimental segmentation methods (“other” in Table 2)
reported geometric accuracies ranging between 0.28 mm ± 0.04 mm and 1.27 mm ± 0.92
mm. [41,47,50,51,59,69,70]
DISCUSSIONTo date, the most common CT image segmentation method used for bone segmentation
in medical AM is global thresholding. All publications reviewed in this study that used
global thresholding reported geometric accuracies under 0.62 mm ± 0.76 mm [59]
(Table 2). However, the reported accuracies generally included additional extensive manual
post-processing, which is often very time consuming. Furthermore, the reproducibility of
manual post-processing remains a challenge, especially when suboptimal CT images with
thick slices, noise, metal artifacts or motion artifacts are used. In this context, Fasel et al.
[56] reported that a manual clean-up time of 10 hours was required to achieve an accuracy
of 0.25 mm ± 0.23 mm. Such extensive post-processing times are currently a major reason
for the high costs related to medical AM. [27]
In order to overcome this time-consuming manual task, more advanced thresholding
methods can be applied. In this context, an algorithm using multiple local thresholds,
automatically determined using edge detection, achieved an accuracy of 0.18 mm ± 0.02
mm without the need for any supplementary manual editing. [36] Furthermore, another
advanced thresholding method that iteratively reclassifies boundary voxels according to
their surroundings resulted in accuracies of 0.08 mm (calcaneus) and 0.20 mm (spine).
[37,42] However, to the best of our knowledge, such advanced thresholding methods
12
In four publications, [29,46–48] the mean absolute distance (MAD) was calculated by performing
a limited number of linear measurements n, which can be written as follows:
��� � 1��|�� � ��|
�,���
where D is a linear measurement performed on the ground truth, and x is a measurement performed
on the CT-derived 3D surface model. Additionally, in four publications, [49–52] the mean distance
(MD) was calculated as the signed mean of n measurements between different anatomical
landmarks, [49,52] reference points, [51] or photographs of a cryosectioned cadaver, [50] which
can be described as follows:
�� � 1����� � ���
����
Table 2 summarizes the geometric accuracies reported in the publications. All 3D surface model
accuracies generally varied between 0.04 mm and 1.9 mm. Manual segmentation resulted in an
accuracy of 0.20 mm ± 0.15 mm to 0.48 mm ± 0.51 mm. [53,54] In global thresholding, accuracies
varied between 0.04 mm ± 0.59 mm and 0.62 mm ± 0.76 mm. [14,29,36,46,48,49,52,55–59]
However, it must be noted that most authors reported that additional manual editing was performed.
Advanced thresholding methods resulted in accuracies below 0.38 mm ± 0.14 mm.
[36,37,42,45,46] SSM-based segmentation methods achieved accuracies ranging from 0.25 mm ±
0.20 mm to 1.9 mm ± 0.20 mm. [44,60–68] Finally, the publications on more experimental
segmentation methods (“other” in Table 2) reported geometric accuracies ranging between 0.28
mm ± 0.04 mm and 1.27 mm ± 0.92 mm. [41,47,50,51,59,69,70]
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
71
5
are currently not implemented in commercially available medical AM software packages.
In this context, Kang et al. [49] compared four different commercial software packages
and reported that global thresholding was the segmentation method of choice in all four
software packages.
Currently, only two commercial software packages for medical AM have been approved
by the United States Food and Drug Administration (FDA): MimicsTM (Materialise®,
Belgium), and Osirix® MD (Pixmeo, Switzerland). [71] Furthermore D2P™ (DICOM TO
PRINT) (3D Systems, Rock Hill, USA) has very recently (Jan 9, 2017) received 510k clearance
and is currently pending CE approval. These software packages are generally equipped
with global thresholding, region growing, dynamic region growing, local editing by a local
threshold, and multiple manual mask editing tools (e.g., delete, draw). More advanced
semi-automated image segmentation algorithms are currently available in Amira® 3D
analysis software for Life Sciences (FEI Visualization Sciences Group, Hillsboro, OR, USA)
that allow for interactive segmentation using local image histograms and edge detection
algorithms. However, this software is not presently cleared for medical use, although this
will be a requirement from 2020 onwards under the new European Union medical device
regulation (MDR) for patient-specific constructs. [72] In this context, standardization and
CE certification will be required for the whole medical AM process (Figure 1), including
the evaluation of the repeatability and accuracy of the image segmentation process.
The previously discussed drawbacks of global and advanced thresholding have
led to the development of novel SSM-based CT image segmentation approaches.
However, the SSM publications assessed in this study demonstrated a large variability in
segmentation accuracies (Table 2). Moreover, only two out of eleven publications (18%)
reported accuracies equal to or less than 0.5 mm. [60,61] The major challenge in SSM-based
segmentation methods is the quality and number of available training datasets, which
can affect the generalization ability and specificity of an SSM. [39] The generalization
ability quantifies to which extent the variability of the anatomical shapes in the patient
population are incorporated in the SSM. For example, when too few training datasets
are used, the SSM may fail to represent new shapes. The specificity of an SSM refers to
its ability to only generate segmented shapes that are similar to those incorporated in
the training datasets. Therefore, SSM-based segmentation methods are only applicable
for CT datasets with moderate anatomical variations. Note that 3D statistical deformable
models have also been used to segment bony structures, such as the knee joint, on
magnetic resonance images. [73]
Clinical implicationsThe majority of the accuracies that are depicted in Table 2 may be considered sufficient
for the creation of anatomical models, [2] of which hundreds are being successfully
manufactured and used in clinical settings every year. [74,75] However, the use of inaccurate
3D surface models to create patient-specific AM implants or guides can lead to ill-fitting
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
72
5
Table 1. Publications included in this review.
Ref Year Author Segmentation method Software Anatomy (#) MaterialCT scanner (manufacturer/# slices)
CT tube voltage / tube current
CT slice thickness / voxel size Ground truth
55 2016 Szymor Global thresholding Slicer Skull (1) Dry bone CBCT (Accuitomo) 90 kV / 5 mA 0.5 mm / 0.25 mm
Optical 3D scan
61 2015 Chu Random forest & multi-atlas & articulated SSM
Custom & Amira Femur (60) + acet. (60)
In-vivo MDCT 1.6 mm / 0.6 mm
Manual segmentation
48 2015 Poleti Global thresholding Dolphin (A)/ InVesalius (B)
Mandible (10) Dry bone CBCT (i-CAT) 120 kV / 18.5 mAs - / 0.3 mm Digital calipers
64 2014 Balestra Articulated SSM Custom & Amira Femur (26) + pelvis (26)
In-vivo MDCT Manual segmentation
56 2014 Fasel Global thresholding n.a. (3 institutes) Skull (3) Cadaver MDCT (Siemens Sensation/64) Optical 3D scan49 2014 Kang Global thresholding InVivoDental (A)
/ Mimics (B) / OnDemand3D (C) / OsiriX (D)
Skull (3) Dry bone MDCT (Siemens Sensation/64) 0.6 mm / 0.44 mm
Optical 3D scan
46 2014 Santolaria Global thresholding (A) / Advanced thresholding (B)
Mimics Mandible (1) Dry bone CBCT (Kodak 9500) - / 0.2 mm - 0.3 mm
CMM
57 2014 Van den Broeck Global thresholding Mimics Tibia (10) Cadaver MDCT (GE LightSpeed VCT) 120 kVp / 160 mAs
0.63 mm / 0.39 mm
Optical 3D scan
44 2014 Zhang Region-based SSM Custom & Stradwin Femur (41) In-vivo MDCT 1.6 mm / 1.0 mm
Manual segmentation
60 2013 Chang Base invariant wavelet active shape model
Custom Maxilla (19) In-vivo CBCT (Sirona) - / 0.29 mm Manual segmentation
47 2013 Engelbrecht “Commercial company” (A) / “doctor” (B)
Mimics & SimPlant Ortho Pro
Mandible (7) Cadaver CBCT (KaVo 3D exam) - / 0.3 mm Laser surface scan
14 2013 Smith Global thresholding Mimics Hip (6) + shoulder (6) joint
Cadaver MDCT (GE Lightspeed+XCR/16)
0.63 mm / - Laser surface scan
69 2013 Yokota SSM Custom Femur (200) + acet. (200)
In-vivo MDCT Manual segmentation
45 2013 Zhou 3D adaptive thresholding MATLAB Hip joint (70) In-vivo MDCT (GE Toshiba) 1.5 mm / 0.73 mm
Manual segmentation
69 2012 Oneill Semi-automatic morphological snakes
Custom Femur (6) In-vivo MDCT (GE Lightspeed Plus) 1.25 mm / 0.68 mm
Manual segmentation
58 2011 Anstey Global thresholding Mimics Femur (1) + acet. (1)
Cadaver MDCT (GE LightSpeed+ XCR/16)
0.63 / - Laser surface scan
36 2011 Rathnayaka Global thresholding (A) / multi-thresholding (B) / edge detection (C)
MATLAB & Amira Femur (5) Cadaver MDCT (Philips Brilliance/64) 140 kVp / - 0.5 mm / 0.39 mm
3D contact scanner
54 2010 Thorhauer Manual Mimics Knee (2) Cadaver MDCT (GE Lightspeed/64) 1 mm / 0.80 mm
Laser surface scan
63 2009 Kainmueller SSM Custom Femur (30) + acet. (50)
In-vivo MDCT 0.5 mm - 5mm / 0.5 mm - 0.9 mm
Manual segmentation
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
73
5
Table 1. Publications included in this review.
Ref Year Author Segmentation method Software Anatomy (#) MaterialCT scanner (manufacturer/# slices)
CT tube voltage / tube current
CT slice thickness / voxel size Ground truth
55 2016 Szymor Global thresholding Slicer Skull (1) Dry bone CBCT (Accuitomo) 90 kV / 5 mA 0.5 mm / 0.25 mm
Optical 3D scan
61 2015 Chu Random forest & multi-atlas & articulated SSM
Custom & Amira Femur (60) + acet. (60)
In-vivo MDCT 1.6 mm / 0.6 mm
Manual segmentation
48 2015 Poleti Global thresholding Dolphin (A)/ InVesalius (B)
Mandible (10) Dry bone CBCT (i-CAT) 120 kV / 18.5 mAs - / 0.3 mm Digital calipers
64 2014 Balestra Articulated SSM Custom & Amira Femur (26) + pelvis (26)
In-vivo MDCT Manual segmentation
56 2014 Fasel Global thresholding n.a. (3 institutes) Skull (3) Cadaver MDCT (Siemens Sensation/64) Optical 3D scan49 2014 Kang Global thresholding InVivoDental (A)
/ Mimics (B) / OnDemand3D (C) / OsiriX (D)
Skull (3) Dry bone MDCT (Siemens Sensation/64) 0.6 mm / 0.44 mm
Optical 3D scan
46 2014 Santolaria Global thresholding (A) / Advanced thresholding (B)
Mimics Mandible (1) Dry bone CBCT (Kodak 9500) - / 0.2 mm - 0.3 mm
CMM
57 2014 Van den Broeck Global thresholding Mimics Tibia (10) Cadaver MDCT (GE LightSpeed VCT) 120 kVp / 160 mAs
0.63 mm / 0.39 mm
Optical 3D scan
44 2014 Zhang Region-based SSM Custom & Stradwin Femur (41) In-vivo MDCT 1.6 mm / 1.0 mm
Manual segmentation
60 2013 Chang Base invariant wavelet active shape model
Custom Maxilla (19) In-vivo CBCT (Sirona) - / 0.29 mm Manual segmentation
47 2013 Engelbrecht “Commercial company” (A) / “doctor” (B)
Mimics & SimPlant Ortho Pro
Mandible (7) Cadaver CBCT (KaVo 3D exam) - / 0.3 mm Laser surface scan
14 2013 Smith Global thresholding Mimics Hip (6) + shoulder (6) joint
Cadaver MDCT (GE Lightspeed+XCR/16)
0.63 mm / - Laser surface scan
69 2013 Yokota SSM Custom Femur (200) + acet. (200)
In-vivo MDCT Manual segmentation
45 2013 Zhou 3D adaptive thresholding MATLAB Hip joint (70) In-vivo MDCT (GE Toshiba) 1.5 mm / 0.73 mm
Manual segmentation
69 2012 Oneill Semi-automatic morphological snakes
Custom Femur (6) In-vivo MDCT (GE Lightspeed Plus) 1.25 mm / 0.68 mm
Manual segmentation
58 2011 Anstey Global thresholding Mimics Femur (1) + acet. (1)
Cadaver MDCT (GE LightSpeed+ XCR/16)
0.63 / - Laser surface scan
36 2011 Rathnayaka Global thresholding (A) / multi-thresholding (B) / edge detection (C)
MATLAB & Amira Femur (5) Cadaver MDCT (Philips Brilliance/64) 140 kVp / - 0.5 mm / 0.39 mm
3D contact scanner
54 2010 Thorhauer Manual Mimics Knee (2) Cadaver MDCT (GE Lightspeed/64) 1 mm / 0.80 mm
Laser surface scan
63 2009 Kainmueller SSM Custom Femur (30) + acet. (50)
In-vivo MDCT 0.5 mm - 5mm / 0.5 mm - 0.9 mm
Manual segmentation
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
74
5
Table 1. (continued)
Ref Year Author Segmentation method Software Anatomy (#) MaterialCT scanner (manufacturer/# slices)
CT tube voltage / tube current
CT slice thickness / voxel size Ground truth
70 2009 Ramme Expectation maximization BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA
0.4 mm / 0.34 mm
Laser surface scan
42 2009 Zhang 3D adaptive thresholding Custom Calcaneus (1) + spine (2)
In-vivo MDCT (Toshiba HR / Siemens) 0.7 mm - 0.4 mm / 0.8 mm - 0.49 mm
Manual segmentation
53 2008 DeVries Manual BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA 0.4 mm / 0.34 mm
Laser surface scan
41 2008 Gassman Artificial neural network BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA 0.4 mm / 0.34 mm
Laser surface scan
59 2008 Gelaude Global thresholding (A) / self-developed contour-based algorithm (B)
Mimics Femur (15) Cadaver MDCT (Siemens Sensation) - / 1 mm Optical 3D scan
62 2008 Seim SSM Custom Pelvis (50) In-vivo MDCT 5 mm / 0.9 mm Manual segmentation
51 2006 Gelaude Adaptive contouring Mimics & MATLAB Pelvis (1) + humerus (1)
Cadaver MDCT 2 mm / - Manual segmentation
52 2006 Loubele Global thresholding Custom Skull (1) Skull phantom MDCT (Siemens Sensation/16) (A) / CBCT (I-CAT) (B)
Laser surface scan
66 2006 Rueda Active appearance model Custom Mandible (215) In-vivo MDCT Manual segmentation
60 2005 Schmidt Global thresholding / Self-developed border tracing algorithm *
Mimics / Custom Femur (1) + acet. (1)
Cryosectioned cadaver
MDCT 120 kV / 120 mA – 200 mA
1 mm / - Photographic images
67 2004 Lamecker SSM Custom Pelvis (23) In-vivo MDCT 5 mm / 1.4 mm Manual segmentation
43 2004 Rajamani SSM Custom Femur (14) In-vivo MDCT Manual segmentation
29 2002 Choi Global thresholding V-works Skull (1) Dry bone MDCT (Siemens Plus/4) 120 kVp / 200 mA 1 mm / - Anatomical landmarks
Acet. = acetabulum; CMM = coordinate measurement machine; MDCT = multi-detector row computed tomography; CBCT = cone-beam computed tomography. * Schmidt et al. (2005) did not disclose the identity of the segmentation methods used.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
75
5
Table 1. (continued)
Ref Year Author Segmentation method Software Anatomy (#) MaterialCT scanner (manufacturer/# slices)
CT tube voltage / tube current
CT slice thickness / voxel size Ground truth
70 2009 Ramme Expectation maximization BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA
0.4 mm / 0.34 mm
Laser surface scan
42 2009 Zhang 3D adaptive thresholding Custom Calcaneus (1) + spine (2)
In-vivo MDCT (Toshiba HR / Siemens) 0.7 mm - 0.4 mm / 0.8 mm - 0.49 mm
Manual segmentation
53 2008 DeVries Manual BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA 0.4 mm / 0.34 mm
Laser surface scan
41 2008 Gassman Artificial neural network BRAINS2 Phalanges (15) Cadaver MDCT (Siemens Sensation/64) 120 kVp / 105 mA 0.4 mm / 0.34 mm
Laser surface scan
59 2008 Gelaude Global thresholding (A) / self-developed contour-based algorithm (B)
Mimics Femur (15) Cadaver MDCT (Siemens Sensation) - / 1 mm Optical 3D scan
62 2008 Seim SSM Custom Pelvis (50) In-vivo MDCT 5 mm / 0.9 mm Manual segmentation
51 2006 Gelaude Adaptive contouring Mimics & MATLAB Pelvis (1) + humerus (1)
Cadaver MDCT 2 mm / - Manual segmentation
52 2006 Loubele Global thresholding Custom Skull (1) Skull phantom MDCT (Siemens Sensation/16) (A) / CBCT (I-CAT) (B)
Laser surface scan
66 2006 Rueda Active appearance model Custom Mandible (215) In-vivo MDCT Manual segmentation
60 2005 Schmidt Global thresholding / Self-developed border tracing algorithm *
Mimics / Custom Femur (1) + acet. (1)
Cryosectioned cadaver
MDCT 120 kV / 120 mA – 200 mA
1 mm / - Photographic images
67 2004 Lamecker SSM Custom Pelvis (23) In-vivo MDCT 5 mm / 1.4 mm Manual segmentation
43 2004 Rajamani SSM Custom Femur (14) In-vivo MDCT Manual segmentation
29 2002 Choi Global thresholding V-works Skull (1) Dry bone MDCT (Siemens Plus/4) 120 kVp / 200 mA 1 mm / - Anatomical landmarks
Acet. = acetabulum; CMM = coordinate measurement machine; MDCT = multi-detector row computed tomography; CBCT = cone-beam computed tomography. * Schmidt et al. (2005) did not disclose the identity of the segmentation methods used.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
76
5
Table 2. Mean and standard deviation of the reported accuracies of different CT image segmentation methods. Cad = cadaver, MSD = mean surface distance; RMSD = root-mean-square distance; MAD = mean absolute difference; MD = mean difference.
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
77
5
implants and complications during surgery. [23] Moreover, recent advances in computer-
assisted surgery, such as surgical navigation, [76] augmented reality, [77] and robot-guided
surgery, [78] will require highly accurate 3D surface models (< 1 mm). In this regard, it
should be noted that accuracy also depends on the resolution of the CT scanner. Currently,
routine diagnostic scan protocols typically have slice thicknesses ranging between 0.6
mm and 2 mm and voxel sizes ranging from 0.2 mm to 0.6 mm, which subsequently limits
the accuracy that can be achieved in clinical settings.
Suggestions for improvementThe limitations of the thresholding and SSM-based CT image segmentation methods
commonly used in medical AM are currently impeding the cost-effectiveness of patient-
specific AM constructs due to the overall time required for the segmentation and manual
post-processing. Therefore, more advanced methods are necessary that can accurately
segment different anatomical structures and their morphological variations. Therefore,
the authors of this study suggest that algorithms such as morphological snakes, [69]
artificial neural networks, [41] and expectation maximization [70] could markedly optimize
CT image segmentation in medical AM. In particular, deep convolutional neural networks
(CNNs) can be taught to obtain mid- and high-level abstractions from CT images. [79]
Based on the reviewed publications, the authors of this study suggest that future
research should focus on the clinical translation of advanced segmentation methods and
intelligent user interfaces. Such a novel interface would facilitate local segmentation using
multiple threshold values derived from local image histograms, edge detection algorithms,
and CNNs. Furthermore, an engineer should be able to interact with the segmentation
process in real time based on a preview of the 3D surface model. Such an approach will,
however, only be possible with advancements in graphical computing power.
Limitations of the reviewed studiesA major limitation of the reviewed publications was the incomplete specifications of
the CT data. A total of eleven publications (34%) did not mention the CT slice thickness
and/or the voxel size (see Table 1). Furthermore, in twenty-three publications (72%) the CT
tube voltage and tube current were not mentioned, and in eleven publications (34%)
the manufacturer of the CT scanner was not specified.
Another challenge faced when evaluating the publications was the diversity of metrics
used to assess the accuracies. A sizable body of work reported accuracies using relative
distance- based metrics such as volume percentages. [80] These publications were
not included in this review. Only publications that used mean distance-based accuracy
metrics in millimeters (mean surface distance, root-mean-square distance, mean absolute
difference and mean difference) were included. However, it should be noted that the root-
mean-square distance is frequently misinterpreted in practice, and that the mean absolute
difference and mean difference are commonly only based on a limited number of linear
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
78
5
measurements. We therefore recommend that the geometric accuracy of bone 3D surface
models be evaluated using the mean surface distance. Nevertheless, when only the mean
surface distance is reported, large deviations in local anatomical regions can either remain
unnoticed or can strongly influence the mean surface distance. [67] Therefore, the authors
of this study recommend that an additional surface comparison is provided using colors to
depict anatomy-specific deviations (a “heat map”). Such heat maps were used in eleven
publications (34%).
Another limitation of the reviewed studies was the lack of information on the amount
of manual editing that was required in order to achieve the reported accuracies. Manual
editing inherently introduces intra- and inter-operator variability. Therefore, we suggest
that the experience of the operator, the exact nature of manual editing, and the time spent
should always be specified in detail.
Finally, numerous studies included in this review acquired high-resolution CT scans
using dry bones (5 publications) or cadaver specimens (12 publications), which resulted
in mean accuracies of 0.30 mm ± 0.30 mm and 0.46 mm ± 0.35 mm, respectively. In-vivo
CT images (13 publications) resulted in a mean accuracy of 0.73 mm ± 0.36 mm. One
explanation for these differences is that the photon energy spectrum of an X-ray beam
changes as it passes through soft tissues, [81] which results in differences in absorption and
scattering during CT imaging. Moreover, the gray values of soft tissues are closer to those
of bone than air. Consequently, soft tissues make semi-automated bone segmentation
less accurate. [82] Therefore, the reported accuracies found in those studies using dry
bones are not translatable to clinical conditions. Furthermore, musculoskeletal bones
in humans differ markedly in morphology, which causes differences in the accuracy and
manual clean-up time. For example, segmentation of small scaffold-like orbital bones is
more challenging than that of large femoral bones.
Taking the above into account, we hypothesize that the best method to evaluate
the accuracy of medical AM constructs is to use multiple life-like cadaveric specimens
and to subsequently remove all soft tissues from the bone after CT imaging. This offers
the possibility to acquire an optical 3D surface scan of the bone that can be used as an
accurate ground truth (<0.05 mm). Note that boiling or chemical treatments should be
avoided since these treatments can introduce geometric deviations to the bone of up to
0.49 mm [59] and 0.43 mm, [57] respectively.
CONCLUSIONThis literature review revealed that a plethora of different CT image segmentation methods
are currently used for bone segmentation. The accuracy of these image segmentation
methods differed markedly. Global thresholding remains the most widely used CT
image segmentation method in medical AM but often requires extensive manual post-
processing. Advanced thresholding approaches could improve the accuracy of global
thresholding, but such methods are currently not implemented in commercially available
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
79
5
software packages. The development of fully automatic and adaptive image segmentation
algorithms could improve the accuracy and reduce the costs of patient-specific AM
constructs. More specifically, future research should focus on the development of image
segmentation methods using CNNs.
COMPETING INTERESTSNone declared
FUNDINGNone
ETHICAL APPROVALNot required
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
80
5
REFERENCES1. Wurm G, Lehner M, Tomancok B,
Kleiser R, and Nussbaumer K (2011) Cerebrovascular Biomodeling for Aneurysm Surgery: Simulation-Based Training by Means of Rapid Prototyping Technologies. Surg Innov 18:294–306
2. Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos A A, Cai T, Kumamaru K K, George E, Wake N, Caterson E J, Pomahac B, Ho V B, Grant G T, and Rybicki F J (2015) Medical 3D Printing for the Radiologist. Radiographics 35:1965–1988
3. D’Urso P S, Barker T M, Earwaker W J, Bruce L J, Atkinson R L, Lanigan M W, Arvier J F, and Effeney D J (1999) Stereolithographic biomodelling in cranio-maxillofacial surgery: a prospective trial. J. Cranio-Maxillofacial Surg. 27:30–37
4. Tam M D, Laycock S D, Bell D G, and Chojnowski A (2012) 3-D printout of a DICOM file to aid surgical planning in a 6 year old patient with a large scapular osteochondroma complicating congenital diaphyseal aclasia. J. Radiol. Case Rep. 6:31–37
5. McMenamin P G, Quayle M R, McHenry C R, and Adams J W (2014) The production of anatomical teaching resources using three-dimensional (3D) printing technology. Anat. Sci. Educ. 7:479–486
6. Bakhos D, Velut S, Robier A, Al Zahrani M, and Lescanne E (2010) Three-Dimensional Modeling of the Temporal Bone for Surgical Training. Otol. Neurotol. 31:328–334
7. Dawood A, Marti B M, Sauret-Jackson V, and Darwood A (2015) 3D printing in dentistry. Br Dent J 219:521–529
8. Ramasamy M, Giri, Raja R, Subramonian, Karthik, and Narendrakumar R (2013) Implant surgical guides: From the past to the present. J. Pharm. Bioallied Sci. 5:S98–S102
9. Jardini A L, Larosa M A, Maciel Filho R, Zavaglia C A D C, Bernardes L F, Lambert C S, Calderoni D R, and Kharmandayan P (2014) Cranial reconstruction: 3D biomodel and custom-built implant created using additive manufacturing. J. Cranio-Maxillofacial Surg. 42:1877–1884
10. Ibrahim D, Broilo T L, Heitz C, de Oliveira M G, de Oliveira H W, Nobre S M W, dos Santos Filho J H G, and Silva D N (2009) Dimensional error of selective laser sintering, three-dimensional printing and PolyJetTM models in the reproduction of mandibular anatomy. J. Cranio-Maxillofacial Surg. 37:167–173
11. Martelli N, Serrano C, van den Brink H, Pineau J, Prognon P, Borget I, and El Batti S (2016) Advantages and disadvantages of 3-dimensional printing in surgery: A systematic review. Surgery 159:1485–1500
12. Bibb R, Eggbeer D, and Paterson A (2015) Medical modelling: The application of advanced design and rapid prototyping techniques in medicine. 2nd ed. Woodhead Publishing
13. Ibrahim D, Broilo T L, Heitz C, de Oliveira M G, de Oliveira H W, Nobre S M W, Dos Santos Filho J H G, and Silva D N (2009) Dimensional error of selective laser sintering, three-dimensional printing and PolyJet models in the reproduction of mandibular anatomy. J. Craniomaxillofac. Surg. 37:167–73
14. Smith E J, Anstey J A, Venne G, and Ellis R E (2013) Using additive manufacturing in accuracy evaluation of reconstructions from computed tomography. Proc Inst Mech Eng H 227:551–559
15. Huotilainen E, Paloheimo M, Salmi M, Paloheimo K-S, Björkstrand R, Tuomi J, Markkola A, and Mäkitie A (2014) Imaging requirements for medical applications of additive manufacturing. Acta radiol. 55:78–85
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
81
5
16. Liang X, Lambrichts I, Sun Y, Denis K, Hassan B, Li L, Pauwels R, and Jacobs R (2010) A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT). Part II: On 3D model accuracy. Eur. J. Radiol. 75:270–274
17. Koivisto J, Kiljunen T, Kadesjö N, Shi X-Q, and Wolff J (2015) Effective radiation dose of a MSCT, two CBCT and one conventional radiography device in the ankle region. J. Foot Ankle Res. 8
18. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
19. Winder J and Bibb R (2005) Medical Rapid Prototyping Technologies: State of the Art and Current Limitations for Application in Oral and Maxillofacial Surgery. J. Oral Maxillofac. Surg. 63:1006–1015
20. Hughes CW, Page K, Bibb R, Taylor J, Revington P. The custom-made titanium orbital floor prosthesis in reconstruction for orbital floor fractures. British Journal of Oral and Maxillofacial Surgery 2003;41:50-53.
21. Hiller J D and Lipson H (2009) STL 2.0: a proposal for a universal multi-material Additive Manufacturing File format. In: Proceedings of the Solid Freeform Fabrication Symposium, 266–278
22. Huotilainen E, Jaanimets R, Valášek J, Marcián P, Salmi M, Tuomi J, Mäkitie A, and Wolff J (2014) Inaccuracies in additive manufactured medical skull models caused by the DICOM to STL conversion process. J. Cranio-Maxillofacial Surg. 42:e259–e265
23. Stoor P, Suomalainen A, Lindqvist C, Mesimaki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific implants for reconstruction of orbital wall defects. J. cranio-maxillo-facial Surg. 42:1644–1649
24. Fourie Z, Damstra J, Schepers R H, Gerrits P O, and Ren Y (2012) Segmentation process significantly influences the accuracy of 3D surface models derived from cone beam computed tomography. Eur. J. Radiol. 81:e524–e530
25. Sharma N and Aggarwal L M (2010) Automated medical image segmentation techniques. J. Med. Phys. 35:3–14
26. Hieu L C, Zlatov N, Sloten J Vander, Bohez E, Khanh L, Binh P H, Oris P, and Toshev Y (2005) Medical rapid prototyping applications and methods. Assem. Autom. 25:284–292
27. Rengier F, Mehndiratta A, von Tengg-Kobligk H, Zechmann C M, Unterhinninghofen R, Kauczor H-U, and Giesel F L (2010) 3D printing based on imaging data: review of medical applications. Int. J. Comput. Assist. Radiol. Surg. 5:335–341
28. Gibson I, Cheung L K, Chow S P, Cheung W L, Beh S L, Savalani M, and Lee S H (2006) The use of rapid prototyping to assist medical applications. Rapid Prototyp. J. 12:53–58
29. Choi J Y, Choi J H, Kim N K, Kim Y, Lee J K, Kim M K, Lee J H, and Kim M J (2002) Analysis of errors in medical rapid prototyping models. Int. J. Oral Maxillofac. Surg. 31:23–32
30. Klein H M, Schneider W, Alzen G, Voy E D, and Günther R W (1992) Pediatric craniofacial surgery: Comparison of milling and stereolithography for 3D model manufacturing. Pediatr. Radiol. 22:458–460
31. Otsu N (1975) A Threshold Selection Method from Gray-Level Histograms. Automatica 20:62–66
32. Canny J (1986) A Computational Approach to Edge Detection. IEEE Trans. Pattern Anal. Mach. Intell. PAMI-8:679–698
33. Adams R and Bischof L (1994) Seeded region growing. IEEE Trans. Pattern Anal. Mach. Intell. 16:641–647
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
82
5
34. van Eijnatten M, Koivisto J, Karhu K, Forouzanfar T, and Wolff J (2016) The impact of manual threshold selection in medical additive manufacturing. Int. J. Comput. Assist. Radiol. Surg. 12:607–615
35. Rathnayaka K, Momot K I, Noser H, Volp A, Schuetz M A, Sahama T, and Schmutz B (2014) Quantification of the accuracy of MRI generated 3D models of long bones compared to CT generated 3D models. Med. Eng. Phys. 34:357–363
36. Rathnayaka K, Sahama T, Schuetz M A, and Schmutz B (2011) Effects of CT image segmentation methods on the accuracy of long bone 3D reconstructions. Med. Eng. Phys. 33:226–33
37. Zhang J, Yan C H, Chui C K, and Ong S H (2010) Fast segmentation of bone in CT images using 3D adaptive thresholding. Comput. Biol. Med. 40:231–236
38. Withey D J and Koles Z J (2007) Medical Image Segmentation: Methods and Software. 2007 Jt. Meet. 6th Int. Symp. Noninvasive Funct. Source Imaging Brain Hear. Int. Conf. Funct. Biomed. Imaging (NFSI-ICFBI 2007): 140–143
39. Heimann T and Meinzer H P (2009) Statistical shape models for 3D medical image segmentation: A review. Med. Image Anal. 13:543–563
40. Aspert N, Santa-Cruz D, and Ebrahimi T (2002) MESH: measuring errors between surfaces using the Hausdorff distance. In: Proceedings. IEEE International Conference on Multimedia and Expo. 705–708 vol.1
41. Gassman E E, Powell S M, Kallemeyn N A, Devries N A, Shivanna K H, Magnotta V A, Ramme A J, Adams B D, and Grosland N M (2008) Automated bony region identification using artificial neural networks: reliability and validation measurements. Skeletal Radiol. 37:313–319
42. Zhang J (2009) Model-Based Segmentation and Registration of Multimodal Medical Images, National University of Singapore
43. Rajamani K, Nolte L, and Styner M (2004) A Novel Approach to Anatomical Structure Morphing for Intraoperative Visualization. In: Medical Image Computing and Computer-Assisted Intervention -- MICCAI 2004: 7th International Conference, Saint-Malo, France, September 26-29, 2004. Proceedings, Part II C. Barillot, D. R. Haynor, and P. Hellier, Eds. Berlin, Heidelberg, 478–485
44. Zhang J, Malcolm D, Hislop-Jambrich J, Thomas C D L, and Nielsen P M F (2014) An anatomical region-based statistical shape model of the human femur. Comput. Methods Biomech. Biomed. Eng. Imaging Vis. 2:176–185
45. Zhou S, Cheng Y, Wang Y, Dong K, Guo C, Bai J, and Tamura S (2013) Segmentation of the hip joint in CT volumes using adaptive thresholding classification and normal direction correction. J. Chinese Inst. Eng. 36:1059–1072
46. Santolaria J, Jiménez R, Rada M, and Loscos F (2014) Error compensation method for improving the accuracy of biomodels obtained from CBCT data. Med. Eng. Phys. 36:397–404
47. Engelbrecht W P, Fourie Z, Damstra J, Gerrits P O, and Ren Y (2013) The influence of the segmentation process on 3D measurements from cone beam computed tomography-derived surface models. Clin. Oral Investig. 17:1919–1927
48. Poleti M L, Fernandes T M F, Pagin O, Moretti M R, and Rubira-Bullen I R F (2016) Analysis of linear measurements on 3D surface models using CBCT data segmentation obtained by automatic standard pre-set thresholds in two segmentation software programs: an in vitro study. Clin. Oral Investig. 20:179–185
49. Kang S-H, Kim M-K, Kim H-J, Zhengguo P, and Lee S-H (2014) Accuracy Assessment of Image-Based Surface Meshing for Volumetric Computed Tomography
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
83
5
Images in the Craniofacial Region. J. Craniofac. Surg. 25:2051–2055
50. Schmidt J, Engh J, Viceconti M, and Ploeg H (2005) What is the accuracy of surface models created from visible human male computed tomography data? ASB 29th Annu. Meet., p. 63
51. Gelaude F, Sloten J Vander, and Lauwers B (2006) Semi-automated segmentation and visualisation of outer bone cortex from medical images. Comput. Methods Biomech. Biomed. Engin. 9:65–77
52. Loubele M, Maes F, Vandermeulen D, Denis K, Jacobs R, White S C, van Steenberghe D, Van Bael A, Loeckx D, Lambrichts I, and Suetens P (2006) Assessment of bone segmentation quality of CT scanners using laser scanning. Int J CARS 1:389–402
53. DeVries N a, Gassman E E, Kallemeyn N a, Shivanna K H, Magnotta V a, and Grosland N M (2008) Validation of phalanx bone three-dimensional surface segmentation from computed tomography images using laser scanning. Skeletal Radiol. 37:35–42
54. Thorhauer E, Miyawaki M, Illingworth K, Holmes J A, and Anderst W (2010) Accuracy of bone and cartilage models obtained from CT and MRI. In: Poster presented at the 34th Annual Meeting of the American Society of Biomechanics. 18–21
55. Szymor P, Kozakiewicz M, and Olszewski R (2016) Accuracy of open-source software segmentation and paper-based printed three-dimensional models. J. Cranio-Maxillofacial Surg. 44:202–209
56. Fasel J H D, Beinemann J, Schaller K, and Peitgen H-O (2014) Computer Science Tools for Manual Editing of Computed Tomographic Images: Impact on the Quality of 3D Printed Models. Surg. Sci. 5:439–443
57. Van den Broeck J, Vereecke E, Wirix-Speetjens R, and Vander Sloten J (2014)
Segmentation accuracy of long bones. Med. Eng. Phys. 36:949–953
58. Anstey J B, Smith E J, Rasquinha B, Rudan J F, and Ellis R E (2011) On the use of laser scans to validate reverse engineering of bony anatomy. Stud. Health Technol. Inform. 163:18–24
59. Gelaude F, Vander Sloten J, and Lauwers B (2008) Accuracy assessment of CT-based outer surface femur meshes. Comput. Aided Surg. 13:188–199
60. Chang Y-B, Xia J J, Yuan P, Kuo T-H, Xiong Z, Gateno J, and Zhou X (2013) 3D segmentation of maxilla in cone-beam computed tomography imaging using base invariant wavelet active shape model on customized two-manifold topology. J. Xray. Sci. Technol. 21:251–282
61. Chu C, Chen C, Liu L, and Zheng G (2015) FACTS: Fully Automatic CT Segmentation of a Hip Joint. Ann. Biomed. Eng. 43:1247–1259
62. Seim H, Kainmueller D, Heller M, Lamecker H, Zachow S, and Hege H-C (2008) Automatic Segmentation of the Pelvic Bones from CT Data Based on a Statistical Shape Model. In: Proceedings of the First Eurographics Conference on Visual Computing for Biomedicine, 93–100
63. Kainmueller D, Lamecker H, Zachow S, and Hege H-C (2009) An articulated statistical shape model for accurate hip joint segmentation. Conf. Proc. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc. 6345–6351
64. Balestra S, Schumann S, Heverhagen J, Nolte L, and Zheng G (2014) Articulated Statistical Shape Model-Based 2D-3D Reconstruction of a Hip Joint. In: Information Processing in Computer-Assisted Interventions: 5th International Conference, IPCAI 2014, Fukuoka, Japan, June 28, 2014. Proceedings D. Stoyanov, D. L. Collins, I. Sakuma, P. Abolmaesumi, and P. Jannin, Eds. Cham, 128–137
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
84
5
65. Rajamani K T, Joshi S C, and Styner M A (2004) Bone model morphing for enhanced surgical visualization. In Biomedical Imaging: Nano to Macro, 2004. IEEE International Symposium on. 1255–1258
66. Rueda S, Gil J A, Pichery R, and Alcañiz M (2006) Automatic Segmentation of Jaw Tissues in CT Using Active Appearance Models and Semi-automatic Landmarking. In: Medical Image Computing and Computer-Assisted Intervention -- MICCAI 2006: 9th International Conference, Copenhagen, Denmark, October 1-6, 2006. Proceedings, Part I R. Larsen, M. Nielsen, and J. Sporring, Eds. Berlin, Heidelberg, 167–174
67. Lamecker H, Seebass M, Hege H-C, and Deuflhard P (2004) A 3D statistical shape model of the pelvic bone for segmentation. Proc. SPIE 5370:1341–1351
68. Yokota F, Okada T, Takao M, Sugano N, Tada Y, Tomiyama N, and Sato Y (2013) Automated CT segmentation of diseased hip using hierarchical and conditional statistical shape models. In: International Conference on Medical Image Computing and Computer-Assisted Intervention. 190–197
69. O’Neill G T, Lee W-S, and Beaulé P (2011) Segmentation of cam-type femurs from CT scans. Vis. Comput. 28:205–218
70. Ramme A J, DeVries N, Kallemyn N a, Magnotta V a, and Grosland N M (2009) Semi-automated phalanx bone segmentation using the expectation maximization algorithm. J. Digit. Imaging 22:483–91
71. Pinto J M, Arrieta C, Andia M E, Uribe S, Ramos-Grez J, Vargas A, Irarrazaval P, and Tejos C (2015) Sensitivity analysis of geometric errors in additive manufacturing medical models. Med. Eng. Phys. 37:328–334
72. EU (2017) Revisions of medical device directives - European Commission. Available: https://ec.europa.eu/growth/sectors/medical-devices/regulatory-
framework/revision_en. [Accessed: 30-May-2017]
73. Zhu Z and Li G (2011) Construction of 3D Human Distal Femoral Surface Models Using a 3D Statistical Deformable Model. J. Biomech. 44:2362–2368
74. Vos W De, Casselman J, and Swennen G R J (2009) Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: A systematic review of the literature. Int. J. Oral Maxillofac. Surg. 38:609–625
75. Ventola C L (2014) Medical Applications for 3D Printing : Current and Projected Uses. Pharm. Ther. 39:704–711
76. He Y, Zhang Y, An J, Gong X, Feng Z, and Guo C (2013) Zygomatic surface marker-assisted surgical navigation: a new computer-assisted navigation method for accurate treatment of delayed zygomatic fractures. J. Oral Maxillofac. Surg. 71:2101–2114
77. Badiali G, Ferrari V, Cutolo F, Freschi C, Caramella D, Bianchi A, and Marchetti C (2014) Augmented reality as an aid in maxillofacial surgery: validation of a wearable system allowing maxillary repositioning. J. cranio-Maxillofacial Surg. 42:1970–1976
78. Majdani O, Rau T S, Baron S, Eilers H, Baier C, Heimann B, Ortmaier T, Bartling S, Lenarz T, and Leinung M (2009) A robot-guided minimally invasive approach for cochlear implant surgery: preliminary results of a temporal bone study. Int. J. Comput. Assist. Radiol. Surg. 4:475–486
79. Greenspan H, Ginneken B van, and Summers R M (2016) Guest Editorial Deep Learning in Medical Imaging: Overview and Future Promise of an Exciting New Technique. IEEE Trans. Med. Imaging 35:1153–1159
80. Udupa J K, Leblanc V R, Zhuge Y, Imielinska C, Schmidt H, Currie L M, Hirsch B E, and Woodburn J (2006) A framework for evaluating image segmentation algorithms. Comput. Med. Imaging Graph. 30:75–87
CT IM
AG
E SE
GM
EN
TATIO
N M
ETH
OD
S FOR
BO
NE
USE
D IN
ME
DIC
AL A
DD
ITIVE
MA
NU
FAC
TUR
ING
85
5
81. Barrett J F and Keat N (2004) Artifacts in CT: Recognition and Avoidance. RadioGraphics 24:1679–1691
82. van Eijnatten M, Berger F H, de Graaf P, Koivisto J, Forouzanfar T, and Wolff J (2017) Influence of CT parameters on STL model accuracy. Rapid Prototyp. J. 23:679–685
6 THE IMPACT OF MANUAL THRESHOLD SELECTION IN
MEDICAL ADDITIVE MANUFACTURING
Maureen van Eijnatten, Juha Koivisto, Kalle Karhu,
Tymour Forouzanfar, Jan Wolff
International Journal of Computer Assisted Radiology and Surgery (2016). 12(4): 607-615doi: 10.1007/s11548-016-1490-4
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
88
6
ABSTRACTPurposeMedical additive manufacturing requires standard tessellation language (STL) models.
Such models are commonly derived from computed tomography (CT) images using
thresholding. Threshold selection can be performed manually or automatically. The aim
of this study was to assess the impact of manual and default threshold selection on
the reliability and accuracy of skull STL models using different CT technologies.
MethodOne female and one male human cadaver head were imaged using multi-detector row
CT, dual-energy CT, and two cone-beam CT scanners. Four medical engineers manually
thresholded the bony structures on all CT images. The lowest and highest selected mean
threshold values and the default threshold value were used to generate skull STL models.
Geometric variations between all manually thresholded STL models were calculated.
Furthermore, in order to calculate the accuracy of the manually and default thresholded
STL models, all STL models were superimposed on an optical scan of the dry female and
male skulls (“gold standard”).
ResultsThe intra- and inter-observer variability of the manual threshold selection was good (intra-
class correlation coefficients >0.9). All engineers selected grey values closer to soft tissue
to compensate for bone voids. Geometric variations between the manually thresholded
STL models were 0.13 mm (multi-detector row CT), 0.59 mm (dual-energy CT), and 0.55
mm (cone-beam CT). All STL models demonstrated inaccuracies ranging from -0.8 mm
to +1.1 mm (multi-detector row CT), –0.7 mm to +2.0 mm (dual-energy CT), and -2.3 mm
to +4.8 mm (cone-beam CT).
ConclusionsThis study demonstrates that manual threshold selection results in better STL models
than default thresholding. The use of dual-energy CT and cone-beam CT technology in
its present form does not deliver reliable or accurate STL models for medical additive
manufacturing. New approaches are required that are based on pattern recognition and
machine learning algorithms.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
89
6
Figure 1. A schematic diagram of the three steps required to fabricate an AM medical construct.
INTRODUCTIONAdditive manufacturing (AM), also known as three-dimensional (3D) printing, refers to
a process where a series of successive layers are laid down to create a 3D construct. AM
combined with advanced medical imaging technologies such as computed tomography
(CT) and magnetic resonance imaging (MRI) has resulted in a paradigm shift in medicine
from traditional serial production to patient-specific constructs. This combination of
technologies offers new possibilities for the fabrication of implants, saw guides and drill
guides that are designed to meet the specific anatomical needs of patients [1].
The three-step medical AM process begins with image acquisition (Figure 1, Step 1),
which is commonly performed using a multi-detector row computed tomography (MDCT)
scanner. However, dual-energy computed tomography (DECT), which offers the possibility
of acquiring CT images using two different X-ray spectra, is becoming more common
in hospital environments [2]. Furthermore, cone-beam computed tomography (CBCT) is
being increasingly used in dentistry and maxillofacial surgery due to its low costs and
reduced radiation dose when compared with MDCT scanners [3].
Images acquired using CT technologies are commonly saved as Digital Imaging and
Communications in Medicine (DICOM) files. These files contain voxels with grey values that
are proportional to the attenuation coefficient in the corresponding volume of the patient.
In MDCT, these grey values are scaled according to Hounsfield units (HU): air (-1000 HU),
water (0 HU), and compact bone (+1000 HU). In CBCT technology, the degree of x-ray
attenuation is scaled using grey values, hence voxel values [4]. CBCT grey values are often
arbitrary and do not correspond to MDCT HU values [3], [5], [6]. Furthermore, a large
variability in the grey values has been reported between different CBCT scanners [7], [8].
At present, medical AM requires the conversion of DICOM images into virtual 3D
surface models that are commonly saved as standard tessellation language (STL) files
(Figure 1, Step 2). STL models are commonly used to design medical constructs using
computer-aided design (CAD) software. The DICOM-to-STL conversion process requires
the partitioning and hence the segmentation of voxels into different tissue types. The most
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
90
6
common segmentation method used to date is thresholding. During the thresholding
process, all voxels with a grey value that is equal or greater than a selected threshold value
t are included in a segmented volume [9] using a binary mask Mx,y (Equation 1):
(1)
where Ix,y denotes the grey value at coordinates x and y.
The medical image segmentation software packages available offer only a single,
default threshold value for compact bone, soft tissue, and cartilage. However, these
default values are often not optimized for all types of MDCT, DECT, and CBCT images
and do not take into account the variations in grey values between different scanners [10].
Therefore, in most cases, manual threshold selection is necessary to acquire an optimal
STL model. Threshold selection, however, still remains a subjective task [11], especially in
the head area due to the plethora of complex bony geometries (Figure 2). Furthermore,
minor dislocations in the facial area can have an impact on patient function and
aesthetic appearance.
At present, there is a paucity of literature on threshold selection in the head area for
medical purposes. Therefore, the aim of this study was to assess the impact of manual and
default threshold selection on the reliability and accuracy of skull STL models acquired
using different MDCT and CBCT technologies.
MATERIALS AND METHODSOne female and one male human cadaver head were anonymously provided by
the Department of Anatomy, VU University Medical Center Amsterdam, The Netherlands.
The two heads were embedded in a novel embalming liquid “Fix for Life” [12] that produces
near life-like cadavers. Ethical approval for this study was provided by the Medical Ethical
Committee of the VU University Medical Center (Ref. 2016.401).
M��� � � 0I��� � �1I��� � � , ( 1 )
Figure 2. The effect of threshold selection on skull STL models.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
91
6
The two “Fix for Life” cadaver heads were imaged using the following CT technologies:
GE Discovery CT750 HD 64-slice MDCT (GE Healthcare, Little Chalfont, Buckinghamshire,
UK), NewTom 5G CBCT (NewTom, Verona, Italy), and Vatech PaX Zenith 3D CBCT (Vatech,
Fort Lee, USA) (Figure 3, Step 1). The GE Discovery CT750 MDCT scanner was also
operated using a dual-energy imaging mode (DECT). All scanners and image acquisition
parameters are summarized in Table 1.
After CT image acquisition, all DICOM files were imported into Osirix® MD software
(Osirix Foundation, Geneva, Switzerland). This software is FDA-cleared, CE-labeled for
primary diagnostics, and is commonly used in medical AM. Osirix® MD software provides
options for both manual and default threshold selection.
Four medical engineers were subsequently requested to manually select the optimal
threshold value for bone in order to create an accurate STL model of the female and
male skull, hence facial bony structures (Figure 3, Step 2). All four engineers were blinded
for their own results and those of others. The manual threshold selection procedure was
repeated after a six-week interval in order to determine the intra-observer variability and
to calculate the mean threshold value. In addition, the inter-observer variability and intra-
class correlation coefficients (ICC) were calculated using SPSS® software (SPSS® version 22,
Chicago, Il, USA). ICC ranges between 0 and 1, with 0 corresponding to no agreement and
1 corresponding to complete agreement [13].
In order to graphically represent the distribution of grey values in the manually selected
and default threshold values, histograms were plotted for each of the four CT scanners
Figure 3. Outline of the study.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
92
6
Fig
ure
3. O
utlin
e o
f the
stu
dy.
GE
Dis
cove
ry C
T750
HD
64
-slic
e (M
DC
T)G
E D
isco
very
CT7
50 H
D
64-s
lice
(DE
CT)
New
Tom
5G
(CB
CT)
Vate
ch P
aX Z
enit
h 3D
(C
BC
T)
fem
ale
mal
efe
mal
eM
ale
fem
ale
mal
efe
mal
em
ale
Tub
e vo
ltag
e (k
V)
120
120
80,1
4080
,140
110
110
115
115
Tub
e cu
rren
t (m
A)
300
300
375
375
7 7
66
Exp
osu
re t
ime
(s)
0.63
8 0.
912
0.91
20.
699
1010
2424
Spac
ing
bet
wee
n sl
ices
(mm
)0.
312
0.31
20.
312
0.31
2n.
a.n.
a.n.
a.n.
a.Sl
ices
thi
ckne
ss (m
m)
0.62
50.
625
0.62
50.
625
0.30
00.
300
0.30
00.
300
Num
ber
of v
o x
els
512
x 51
2 x
767
512
x 51
2 x
919
512
x 51
2 x
767
512
x 51
2 x
919
610
x 61
0 x
539
610
x 61
0 x
541
800
x 80
0 x
632
800
x 80
0 x
632
Rec
ons
truc
tion
kern
elB
one
plu
sB
one
plu
sSo
ftSo
ftSt
and
ard
Stan
dar
dn.
a.n.
a.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
93
6
using MatLab® software (MatLab v.2012, MathWorks, Natick, Massachusetts, USA) (Figure
4). Only the highest and lowest mean selected threshold values presented on the eight
histograms were used to generate STL models (Figure 3, Step 3). The generated STL
models were subsequently geometrically compared to each other using GOM Inspect®
software (GOM Inspect v8, GOM mbH, Braunschweig, Germany) in order to calculate
the variations between the highest and lowest threshold STL models (Figure 3, Step 4).
In a final step, all soft tissues were manually removed from the cadaver heads using
standard dissection equipment (i.e., scrapers and scalpels) by a highly experienced
technician of the Department of Anatomy. Manual removal was opted for since this
procedure ensured minimal dimensional changes in the bony structures of the cadaver
skulls [14]. The resulting dry female and male skulls were subsequently scanned using
a GOM ATOS™ III optical 3D scanner (GOM GmbH, Braunschweig, Germany) with an
accuracy of < 0.05 mm to acquire a “gold standard” STL model of the skulls (Figure 3).
These “gold standard” STL models were subsequently superimposed on the STL models
generated using the highest and lowest manually selected and default threshold values in
order to calculate the accuracy of each thresholded STL model (Figure 3, Step 5).
RESULTSThe intra- and inter-observer reliability results of all manually selected threshold values
are presented in Table 2. All selected threshold values ranged from 113 HU to 303 HU for
the MDCT and DECT technologies and from 537 gv to 1281 gv for the CBCT technologies
(Figure 4 A - H). As shown in the histograms, all the selected threshold values differed
from the default threshold value provided by Osirix MD® software (500 HU). Furthermore,
the geometric variations between the highest and lowest thresholded STL models were
larger in the STL models derived from DECT and CBCT when compared with the MDCT-
derived STL models (Figure 5).
When compared to the “gold standard”, all manually and automatically thresholded
STL models demonstrated inaccuracies ranging from -0.8 mm to +1.1 mm, –0.7 mm to
+2.0 mm, and -2.3 mm to +4.8 mm for all STL models derived from MDCT, DECT, and
CBCT, respectively (Figure 6 A - K). The male skull presented comparable accuracies to
those observed on the female skull. The MDCT and DECT-derived STL models acquired
using the default threshold value demonstrated the highest loss of bone-HU values (Figure
6 C, F). The NewTom CBCT-derived STL model acquired using the default threshold value
(500 HU) provided by Osirix MD software resulted in an increase in artifacts and noise
(Figure 6 I). The Vatech CBCT DICOM images did not allow the creation of an STL model
using the 500-HU default threshold value since the grey values were not scaled to HU
values (Figure 4 D,H).
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
94
6
Table 2. Intra- and inter-observer variability of manual threshold selection by four medical engineers on CT images of a female and a male cadaver head.
Intra-observer variability Inter-observer variability between the engineers
Intra-class correlation coefficient (ICC)
ICCEngineer 2female / male
ICCEngineer 3female / male
ICCEngineer 4female / maleCadaver head female / male
Engineer 1 0.999 / 0.997 Engineer 1 0.994 / 0.988 0.980 / 0.987 0.970 / 0.954Engineer 2 0.995 / 0.995 Engineer 2 0.978 / 0.998 0.961 / 0.931Engineer 3 0.992 / 0.999 Engineer 3 0.914 / 0.917Engineer 4 0.969 / 0.989 Engineer 4
Figure 4 (A to H). The mean threshold values (HU) selected by four medical engineers and the pre-defined default threshold value (500 HU) are presented in histograms A to H. The y-axis of the histograms (frequencies) are set to a logarithmic scale.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
95
6
DISCUSSIONTo date, thresholding is the most commonly used segmentation method in medical AM.
However, accurate bone segmentation often requires manual threshold selection, which
still remains a subjective task. Moreover, recent studies suggest that the majority of
inaccuracies that occur during the medical AM process are introduced during the image
acquisition and image processing phases, rather than during the manufacturing, i.e., the 3D
printing process itself [15]–[17]. Such inaccuracies can markedly influence the resulting
STL model (see Figure 6) and subsequently lead to ill-fitting AM implants [18]. Therefore,
the aim of the present study was to assess the impact of manual and automatic default
threshold selection on the reliability and accuracy of skull STL models.
In the present study, all threshold values selected by the four engineers demonstrated
a good intra-observer reliability (ICC>0.9). Furthermore, the inter-observer reliability was
also good (ICC>0.9), as shown in Table 2. Interestingly, all engineers that were blinded
during the experiment selected threshold values for bone that were very close to the grey
values of soft tissues (Figure 4). This resulted in small disjointed structures in the STL
model (marked red in Figure 7) that represent the transition from bone into soft tissue grey
values. Such disjointed “soft-tissue” structures can be manually removed during STL post-
processing [19]. All engineers purposely selected the “soft tissue” threshold values during
bone segmentation in order to incorporate the maximum number of bone-specific grey
values. These grey values are allocated to voxels that represent different tissues during
the CT image reconstruction process. However, during this process, voxels on the bone-
to-soft tissue boundaries that are partially filled with soft tissue are commonly assigned
a lower grey value than bone. This phenomenon is coined the partial volume effect (PVE)
Figure 5. Geometric variations in mm between the highest and lowest thresholded STL models acquired using four different CT scanners (see also Figure 4).
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
96
6
Figure 6 (A - K). Accuracy of all STL models of the female skull acquired using the lowest (left) and highest (middle) mean threshold value selected by the four engineers and the default threshold value of 500 HU (right). The arrows indicate missing data (C, F) or excessive noise (I) in the default-threshold STL models.
[20]. As a consequence of the PVE, voxels may be erroneously allocated to “soft tissue”
instead of “bone”, resulting in data loss and hence bone voids in the STL model (Figure
6). Therefore, engineers should be aware of this phenomenon since data loss can lead to
large inaccuracies in individualized printed medical constructs [18], [20].
Another major finding in this study was the difference between the MDCT and CBCT
DICOM files that were used to construct STL models (Figure 4). One explanation for this
phenomenon is the inherent difference between these technologies. CBCT technology is
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
97
6
typically more heavily affected by image noise and distortions due to the “cone-beam”
geometry of the X-ray beam [21], [22]. In CBCT, the simultaneously irradiated area is
typically larger than in MDCT technology. This causes increased scatter levels and results
in cupping, reduced contrast, and other scatter-induced artifacts in the reconstructed
image. In addition, CBCT images are more subject to cone beam artifacts due to
the large cone beam angle and the imaging geometry comprising a single focal plane.
The cone beam artifacts result from violating Tuy’s sufficiency condition [23] that requires
that each plane intersecting a region of interest must intersect the focal trajectory, i.e.,
the path defining the radiation source position during the imaging. The embodiments
of cone beam artifacts are dependent on the reconstruction algorithm and the imaging
geometry. Typical cone beam artifacts include the elongation of structures in the axial
direction and negative undershoots at sharp edges in the transaxial planes [24]. In CBCT,
Figure 7. MDCT-derived low-threshold STL model of the female cadaver skull (grey) with disjointed “soft-tissue” structures (red).
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
98
6
the focal trajectory consists of a single planar circle or arc that results in a violation of
Tuy’s sufficiency condition in all regions outside the focal plane. The resulting cone beam
artifacts are more pronounced the further away the region of interest is from the focal
plane. In MDCT, the volume that satisfies Tuy’s sufficiency condition is notably larger due
to the helical nature of the focal trajectory.
The presence of artifacts makes the segmentation and hence the thresholding of bone-
specific grey values in CBCT images more cumbersome [25]. This subsequently leads to
a larger variation in manually selected threshold values for CBCT images (Figure 4) and to
the larger geometric variations of up to 0.55 mm in CBCT-derived STL models observed
in this study (Figure 5). DECT-derived STL models demonstrated geometric variations of
up to 0.59 mm (Figure 5). As a consequence of these geometric variations in STL models,
the use of DECT and CBCT technology in its present form does not deliver reproducible
STL models for medical AM. Therefore, the authors of this study suggest that only MDCT
technology should be used for AM applications because of the lower variability (0.13 mm,
see Figure 5) and higher accuracy (Figure 6) of the technology.
The present study demonstrates that the “human factor”, i.e., the medical engineer,
influences the outcome of the segmentation process. Moreover, no single bone threshold
value is applicable for all facial bones. The authors of this study therefore recommend
the use of individual threshold values for each anatomical buttress. Recently, attempts
have been made to develop novel segmentation algorithms using multi-thresholding
[26], adaptive thresholding [11], and semi-automatic region growing [27]. However, these
algorithms are still in an early stage of development [28] and do not take the inherent
differences between MDCT and CBCT technologies into account. Future research should
therefore focus on developing novel medical image segmentation software that is suitable
for different CT imaging modalities. Furthermore, new approaches should be developed
using pattern recognition and machine learning algorithms.
CONCLUSIONThis study shows that manual threshold selection results in better skull STL models than
default thresholding since all the medical engineers in our study selected grey values
closer to soft tissue to compensate for bone voids. Our study also showed that MDCT-
derived STL models offer the lowest variability and highest accuracy, whilst the use of
DECT and CBCT technology in its present form does not deliver reliable STL models
for medical AM. New approaches based on pattern recognition and machine learning
algorithms are required.
STATEMENTSConflicts of interestAuthor 2 Juha Koivisto and author 3 Kalle Karhu are currently employed by Planmeca
Ltd (Finland), a company that specializes in the manufacture of cone-beam computed
tomography scanners. The other authors declare that they have no conflict of interest.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
99
6
Ethical approvalThis article does not contain any studies with human participants or animals performed
by any of the authors. All human cadaveric materials that were used in the present study
(one female and one male head) were anonymously acquired through the body donor
program of the Department of Anatomy of the VU University Medical Center Amsterdam,
The Netherlands, in full accordance with Article 1 of the Dutch law on funeral services (http://
wetten.overheid.nl/BWBR0005009/2015-07-01) and European legislation. Furthermore,
ethical approval for this study was provided by the Medical Ethical Committee (METC) of
the VU University Medical Center (Ref. 2016.401).
Informed consentFor this type of study, no formal consent was required.
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
100
6
REFERENCES
1. Abou-El Fetouh A, Barakat A, and Abdel-Ghany K (2011) Computer-guided rapid-prototyped templates for segmental mandibular osteotomies: a preliminary report. Int. J. Med. Robot. 7:187–192
2. Johnson T R C (2012) Dual-energy CT: general principles. AJR. Am. J. Roentgenol. 199:3–8
3. Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A, Bosmans H, Bogaerts R, Jacobs R, and Horner K (2013) Variability of dental cone beam CT grey values for density estimations. Br. J. Radiol. 86:20120135
4. Razi T, Niknami M, and Alavi Ghazani F (2014) Relationship between Hounsfield Unit in CT Scan and Gray Scale in CBCT. J. Dent. Res. Dent. Clin. Dent. Prospects 8:107–110
5. Katsumata A, Hirukawa A, Okumura S, Naitoh M, Fujishita M, Ariji E, and Langlais R P (2007) Effects of image artifacts on gray-value density in limited-volume cone-beam computerized tomography. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 104:829–36
6. Silva I M de C C, Freitas D Q de, Ambrosano G M, Bóscolo F N, and Almeida S M (2012) Bone density: comparative evaluation of Hounsfield units in multislice and cone-beam computed tomography. Braz. Oral Res. 26:550–556
7. Nackaerts O, Maes F, Yan H, Couto Souza P, Pauwels R, and Jacobs R (2011) Analysis of intensity variability in multislice and cone beam computed tomography. Clin. Oral Implants Res. 22:873–9
8. Reeves T, Mah P, and McDavid W (2012) Deriving Hounsfield units using grey levels in cone beam CT: a clinical application. Dentomaxillofacial Radiol. 41:500–508
9. Sahoo PK, Soltani S, and Wong A K C (1988) A survey of thresholding techniques. Comput. Vision, Graph. Image Process. 41:233–260
10. Logan H, Wolfaardt J, Boulanger P, Hodgetts B, and Seikaly H (2013) Evaluation of the accuracy of Cone Beam Computerized Tomography (CBCT): medical imaging technology in head and neck reconstruction. J. Otolaryngol. - Head Neck Surg. 42:25
11. Nackaerts O, Depypere M, Zhang G, Vandenberghe B, Maes F, and Jacobs R (2015) Segmentation of Trabecular Jaw Bone on Cone Beam CT Datasets. Clin. Implant Dent. Relat. Res. 17:1082–1091
12. van Dam A, van Munsteren C, and de Ruiter M (2015) Fix for Life. The Development of a New Embalming Method to Preserve Life-like Morphology. FASEB J. 29:10
13. Bartko J J (1966) The intraclass correlation coefficient as a measure of reliability. Psychol. Rep. 19:3–11
14. Van den Broeck J, Vereecke E, Wirix-Speetjens R, and Van der Sloten J (2014) Segmentation accuracy of long bones. Med. Eng. Phys. 36:949–953
15. Huotilainen E, Paloheimo M, Salmi M, Paloheimo K-S, Björkstrand R, Tuomi J, Markkola A, and Mäkitie A (2014) Imaging requirements for medical applications of additive manufacturing. Acta radiol. 55:78–85
16. Gibson I, Rosen D, and Stucker B (2014) Additive manufacturing technologies - 3D Printing, Rapid Prototyping, and Direct Digital Manufacturing. Springer, New York
17. Smith E J, Anstey J A, Venne G, and Ellis R E (2013) Using additive manufacturing in accuracy evaluation of reconstructions from computed tomography. Proc Inst Mech Eng H 227:551–559
18. Stoor P, Suomalainen A, Lindqvist C, Mesimaki K, Danielsson D, Westermark A, and Kontio R K (2014) Rapid prototyped patient specific implants for reconstruction of orbital wall defects. J. cranio-maxillo-facial Surg. 42:1644–1649
THE
IMPA
CT O
F MA
NU
AL TH
RE
SHO
LD SE
LEC
TION
IN M
ED
ICA
L AD
DITIV
E M
AN
UFA
CTU
RIN
G
101
6
19. Mitsouras D, Liacouras P, Imanzadeh A, Giannopoulos A A, Cai T, Kumamaru K K, George E, Wake N, Caterson E J, Pomahac B, Ho V B, Grant G T, and Rybicki F J (2015) Medical 3D Printing for the Radiologist. Radiographics 35:1965–1988
20. Huotilainen E, Jaanimets R, Valášek J, Marcián P, Salmi M, Tuomi J, Mäkitie A, and Wolff J (2014) Inaccuracies in additive manufactured medical skull models caused by the DICOM to STL conversion process. J. Cranio-Maxillofacial Surg. 42:e259–e265
21. Feldkamp L A, Davis L C, and Kress J W (1984) Practical cone-beam algorithm. J. Opt. Soc. Am. 1:612
22. Endo M, Tsunoo T, Nakamori N, and Yoshida K (2001) Effect of scattered radiation on image noise in cone beam CT. Med. Phys. 28:469-474
23. Tuy H K (1983) An Inversion Formula for Cone-Beam Reconstruction. SIAM J. Appl. Math. 43:546–552
24. Clack R, Zeng G L, Weng Y, Christian P E, and Gullberg G T (1991) Cone beam single photon emission computed tomography using two orbits. In: Colchester
A C F, Hawkes D J (eds) Proceedings of the Information processing in medical imaging: 12th international conference, IPMI ’91 Wye, July 7–12, 1991. Springer, Berlin, pp 45–54
25. Maret D, Telmon N, Peters O A, Lepage B, Treil J, Inglèse J M, Peyre A, Kahn J L, and Sixou M (2012) Effect of voxel size on the accuracy of 3D reconstructions with cone beam CT. Dentomaxillofacial Radiol. 41:649–655
26. Rathnayaka K, Sahama T, Schuetz M A, and Schmutz B (2011) Effects of CT image segmentation methods on the accuracy of long bone 3D reconstructions. Med. Eng. Phys. 33:226–33
27. Xi T, Schreurs R, Heerink W J, Bergé S J, and Maal T J J (2014) A Novel Region-Growing Based Semi-Automatic Segmentation Protocol for Three-Dimensional Condylar Reconstruction Using Cone Beam Computed Tomography (CBCT). PLoS One 9:e111126
28. Pauwels R, Araki K, Siewerdsen J H, and Thongvigitmanee S S (2015) Technical aspects of dental CBCT: state of the art. Dentomaxillofacial Radiol. 44:20140224
7 ACCURACY OF MDCT AND CBCT IN THREE-DIMENSIONAL
EVALUATION OF THE UPPER AIRWAY MORPHOLOGY
Maureen van Eijnatten*, Hui Chen*,
Ghizlane Aarab, Tymour Forouzanfar, Jan de Lange,
Paul van der Stelt, Frank Lobbezoo, Jan Wolff
European Journal of Orthodontics (2017). cjx030 [epub ahead of print]doi: 10.1093/ejo/cjx030
* Both authors contributed equally to this work
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
104
7
ABSTRACTObjectivesTo assess the accuracy of five different computed tomography (CT) scanners for
the evaluation of the oropharynx morphology
MethodsAn existing cone-beam computed-tomography (CBCT) dataset was used to fabricate an
anthropomorphic phantom of the upper airway volume that extended from the uvula to
the epiglottis (oropharynx) with known dimensions (gold standard). This phantom was
scanned using two multi-detector row computed-tomography (MDCT) scanners (GE
Discovery CT750 HD, Siemens Somatom Sensation) and three CBCT scanners (NewTom
5G, 3D Accuitomo 170, Vatech PaX Zenith 3D). All CT images were segmented by two
observers and converted into standard tessellation language (STL) models. The volume
and the cross-sectional area of the oropharynx were measured on the acquired STL models.
Finally, all STL models were registered and compared with the gold standard.
ResultsThe intra- and inter-observer reliability of the oropharynx segmentation was fair to excellent.
The most accurate volume measurements were acquired using the Siemens MDCT (98.4%;
14.3 cm3) and Vatech CBCT (98.9%; 14.4 cm3) scanners. The GE MDCT, NewTom 5G CBCT
and Accuitomo CBCT scanners resulted in smaller volumes, viz., 92.1% (13.4 cm3), 91.5%
(13.3 cm3), and 94.6% (13.8 cm3), respectively. The most accurate cross-sectional area
measurements were acquired using the Siemens MDCT (94.6%; 282.4 mm2), Accuitomo
CBCT 95.1% (283.8 mm2), and Vatech CBCT 95.3% (284.5 mm2) scanners. The GE MDCT
and NewTom 5G CBCT scanners resulted in smaller areas, viz., 89.3% (266.5 mm2) and
89.8% (268.0 mm2), respectively.
LimitationsImages of the phantom were acquired using the vendor-supplied default airway scanning
protocol for each scanner.
ConclusionSignificant differences were observed in the volume and cross-sectional area measurements
of the oropharynx acquired using different MDCT and CBCT scanners. The Siemens MDCT
and the Vatech CBCT scanners were more accurate than the GE MDCT, NewTom 5G,
and Accuitomo CBCT scanners. In clinical settings, CBCT scanners offer an alternative to
MDCT scanners in the assessment of the oropharynx morphology.
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
105
7
INTRODUCTIONObstructive sleep apnea (OSA) is a sleep-related breathing disorder, often associated
with a compromised upper-airway space and an increase in upper-airway collapsibility
[1]. The most common complaints of OSA patients are excessive daytime sleepiness,
unrefreshing sleep, poor concentration, and fatigue [2]. OSA also has a range of
deleterious consequences that include increased cardiovascular morbidity, neurocognitive
impairment, and overall mortality [3]–[6]. An important role in the pathogenesis of OSA is
played by anatomical and functional abnormalities of the upper airway [7].
The three-dimensional (3D) morphology of the upper airway is currently assessed using
computed tomography (CT) technologies [8]. The two most common CT technologies
used to date for the assessment of the upper airway are multi-detector row computed
tomography (MDCT) [9] and cone-beam computed tomography (CBCT) [10]. The major
advantages of CBCT scanners are their lower radiation dose and costs [11], [12]. As
a result, CBCT scanners are being increasingly used for upper-airway imaging in OSA
patients [13]–[15]. CBCT scanners use a single, partial gantry rotation [16], which not only
accounts for lower radiation dose, but also produces acceptable diagnostic image quality
[17]. However, it remains unclear whether MDCT and CBCT scanners can provide accurate
3D images of the upper airway.
According to a systematic review by Alsufyani et al. [18], only one out of sixteen studies
focusing on the use of CBCT to automatically or semi-automatically model the upper airway
had a sufficiently sound methodology to test the accuracy of the upper airway dimensions.
The main challenge faced in the assessment of the upper airway accuracy using MDCT or
CBCT is the lack of a “gold standard” model with known dimensions. Recent studies have
used artificial models, hence phantoms of the upper airway, as a gold standard [19]–[21].
However, commercially available phantoms used in most of the aforementioned studies are
commonly manufactured in simple, generic forms and sizes and therefore do not resemble
the clinical situation. In the present study, a novel 3D printed anthropomorphic phantom
of the upper airway volume (oropharynx) with known dimensions was manufactured that
closely resembled a real patient in terms of size, shape, structure, and attenuation profiles.
The aim of this study was to assess the accuracy of two different MDCT scanners and
three different CBCT scanners using a novel 3D printed anthropomorphic phantom for
the evaluation of the oropharynx morphology.
MATERIALS AND METHODSA CBCT dataset of a 27-year-old female that had been previously acquired using
a NewTom 5G CBCT scanner (QR systems, Verona, Italy), was used to design and 3D print
an anthropomorphic phantom of the airway space (Figure 1 A and B). The aforementioned
CBCT dataset was converted into a virtual 3D surface, hence standard tessellation
language (STL) model of the upper airway volume that extended from the uvula to
the epiglottis: the oropharynx (Figure 3). This STL model served as the gold standard
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
106
7
Figure 1. A. Design of the oropharynx of the phantom (red = maxilla and mandible; green = cervical vertebrae; yellow = supports of the markers; black = markers at the level of the minimum cross-sectional area of the oropharynx; blue = upper airway; purple = base plane; grey and pale-yellow = mould of the skin). B. The 3D printed phantom. C. Sagittal image of the phantom using GE scanner. Arrow = a marker. D. Segmentation based on GE images (purple = oropharynx; green = base plane of the phantom).
in this study. The gold standard STL model of the oropharynx was subsequently used to
manufacture the phantom. All bony structures surrounding the oropharynx were 3D printed
using a High Performance Composite powder ZP151 (3D Systems, Rock Hill, USA). This
composite material was chosen due to its bone-like density that resembles the attenuation
profile of bone [22]. The soft tissue surrounding the oropharynx was fabricated using soft-
tissue-equivalent silicon (Dragon Skin 30, Smooth-On, Inc., Macungie, Pennsylvania, USA).
During the assembling of the phantom, three metal markers were positioned in a defined
plane to acquire a reproducible reference-point system (RPS) for the cross-sectional area
measurement of the oropharynx (Figure 1).
As Figure 2 shows, MDCT images of the oropharynx phantom were acquired using two
MDCT scanners: GE Discovery CT750 HD 64-slice MDCT (GE Healthcare, Little Chalfont,
Buckinghamshire, UK) and Siemens Somatom Sensation 64-slice MDCT (Siemens Medical
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
107
7
Solutions, Malvern PA, USA). Furthermore, the following three CBCT scanners were used
to acquire CBCT images of the phantom: NewTom 5G CBCT (QR systems, Verona, Italy),
3D Accuitomo 170 CBCT (J Morita, Kyoto, Japan), and Vatech PaX Zenith 3D CBCT
(Vatech, Fort Lee, USA). All MDCT and CBCT images were acquired using the vendor-
supplied default airway scanning protocol. All imaging parameters of the five CT scanners
are presented in Table 1.
The acquired CT datasets were saved as Digital Imaging and Communications in
Medicine (DICOM) files and were imported into Amira® software (v4.1, Visage Imaging
Inc., Carlsbad, CA, USA) (Figure 2C). Using thresholding, one maxillofacial radiologist and
one orthodontist then segmented all the acquired DICOM datasets of the oropharynx
Figure 2. Flowchart of this study.
Table 1. Image acquisition parameters for MDCT and CBCT scans.
GE MDCT
Siemens MDCT
NewTom 5G CBCT
Accuitomo CBCT
Vatech CBCT
Tube voltage (kV) 120 120 110 90 115Tube current (mA) 103 57 5.8 5 6Scan time (s) 0.5 0.5 3.6 17.5 24Slices thickness (mm) 0.625 0.600 0.300 1 0.200Number of voxels 512 x 512x
180512 x 512 x151
610 x 610 x 539
684 x 684 x 480
800 x 800 x 632
Reconstruction Soft B40f Standard N.A. N.A.DLP (mGy-cm)
DAP (mGy-cm2)
CTDIvol (mGy)
46.49
N.A.
N.A.
49
N.A.
N.A.
N.A.
12.232
N.A.
N.A.
N.A.
8.70
N.A.
17.67
N.A.
CBCT: cone beam computed tomography; CTDI: computed tomography dose index; DAP: dose-area product; DLP: dose-length product; Gy: Gray; MDCT: multi-detector row computed tomography; N.A.: not applicable.
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
108
7
(Table 2). Both observers were blinded for their own results and those of each other.
The segmentation procedure was performed five times for each CT scanner and was
subsequently repeated after a ten-day interval. This resulted in a total of 20 threshold
values per CT scanner (Table 2). These 20 threshold values per scanner were subsequently
used to segment the oropharynx. All segmented oropharynx volumes acquired from
the five MDCT and CBCT scanners (Figure 2D) were converted into STL models. These STL
models were subsequently imported into GOM Inspect v8® metrology software (GOM,
Braunschweig, Germany) for further airway analysis.
The region of interest (ROI) in this study was the volume of the oropharynx between two
parallel planes located 1.5 mm and 42 mm above the base plane of the phantom (Figure 3).
In order to obtain comparable oropharynx volume measurements, all STL models derived
from the five CT datasets were cropped accordingly. The volume of the oropharynx was
subsequently calculated using GOM Inspect software (Figure 4A). Furthermore, the cross-
sectional area of the oropharynx at the level of the metal markers in the phantom was
calculated (Figure 4B).
To determine the accuracy of the CT-derived STL models, all acquired STL models were
superimposed onto the gold standard STL model of the oropharynx using a verified surface
registration (local best-fit) algorithm in GOM Inspect software with an accuracy of 0.05 mm
[23]. All geometric deviations between the oropharynx STL models and the printed gold
standard phantom STL model are depicted in Figure 5.
Figure 3. The segmented volume of the oropharynx. (Red line = upper boundary of region of interest (ROI); yellow line = lower boundary of ROI; green line = the location of the markers)
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
109
7
Finally, statistical analysis was performed, using IBM Statistical Package for Social
Sciences for Windows (SPSS® version 21, Chicago, Il, USA). Statistical significance was
set at α=0.05. To determine the intra- and inter-observer reliability of the oropharynx
measurements, intraclass correlation coefficients (ICCs) were calculated. Reliability was
divided into three categories: poor (ICC<0.40); fair to good (0.40≤ICC≤0.75); excellent
(ICC>0.75) [24]. The accuracy of the scanners was calculated as the ratio of the phantom
measurements to the gold standard (%). One-way analysis of variance (ANOVA) was used
to test the differences in the volume and cross-sectional area measurements between
the five different CT scanners. Post-hoc analysis (Tukey’s honest significant difference test)
was run to establish which CT scanners produced significantly different results.
RESULTSAll threshold values used for the segmentation of the oropharynx are shown in table 2.
Intra- and inter-observer reliability hence ICCs of the threshold values ranged from 0.436
(fair to good) to 0.966 (excellent).
There were significant differences between the volume measurements of the oropharynx
STL models acquired using the five different CT scanners (F=84.21; P=0.00) (Figure 4A).
Tukey’s test showed that there were no significant differences in volume measurements
between the Siemens MDCT and Vatech CBCT scanners, and between the GE MDCT and
NewTom 5G CBCT scanners. The Siemens MDCT and Vatech CBCT scanners provided
the most accurate volume measurements of the oropharynx (Figure 4A). The NewTom
5G CBCT, Accuitomo CBCT, and the GE MDCT scanners resulted in smaller volume
measurements of the oropharynx (Figure 4A, Table 3).
There were also significant differences between the cross-sectional area measurements
of the oropharynx STL models acquired using the five different CT scanners (F=43.11;
P=0.00) (Figure 4B). The Siemens MDCT, Vatech CBCT and Accuitomo CBCT scanners
Table 2. Intraobserver and interobserver reliability of the threshold values (Hounsfield Units) chosen for five CT scanners estimated by intraclass correlation coefficients (ICCs), based on 20 measurements in total per scanner.
Scanner
Threshold values (HU) (Intra-observer reliability)
Threshold values (HU)(Inter-observer reliability)
Observer 1Mean±SD (ICC)
Observer 2Mean±SD (ICC) Mean±SD (ICC)
GE (MDCT) -339±47 (.966) -389±87 (.864) -364±73 (.818)Siemens (MDCT) -204±54 (.573) -250±77 (.748) -231±70 (.558)NewTom5G (CBCT) -114±35(.481) -102±40 (.720) -108±37 (.786)Accuitomo (CBCT) -289±40 (.661) -244±62 (.438) -267±56 (.507)Vatech (CBCT) -361±75 (.787) -330±63 (.436) -346±70 (.592)
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
110
7
A.
B.
Figure 4. A. Mean and standard deviation of the volume of the oropharynx derived from five CT scanners. GS = gold standard; NS = no significant difference. B. Mean and standard deviation of the cross-sectional area of the oropharynx derived from five CT scanners. GS = gold standard; NS = no significant difference. The accuracy (%) was calculated as the ratio between the phantom measurements and the gold standard values.
Table 3. Mean and standard deviation of the volume and the cross-sectional area of the upper airway derived from five CT scanners. GS = gold standard.
Variable GSGEMDCT
SiemensMDCT
NewTom 5GCBCT
AccuitomoCBCT
VatechCBCT
Volume of the upper airway (cm3)
14.5 13.4 ± 0.32 14.3 ± 0.31 13.3 ± 0.15 13.8 ± 0.21 14.4 ± 0.19
Area of the upper airway (mm2)
295.5 266.5 ± 6.1 282.4 ± 6.8 268.0 ± 3.8 283.8 ± 7.9 284.5 ± 5.2
provided the most accurate cross-sectional area measurements of the oropharynx (Figure
4B). The GE MDCT and NewTom 5G CBCT scanners resulted in smaller area measurements
of the oropharynx (Figure 4B, Table 3).
Figure 5 shows the oropharynx STL models acquired using five different CT scanners.
The largest geometric deviations of were observed in the vicinity of the uvula and
the epiglottis region (Figure 5).
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
111
7
DISCUSSION3D evaluation of the oropharynx offers new possibilities of assessing anatomical
abnormalities in OSA patients. In this study, significant differences (P < 0.001) were found
between the volume and cross-sectional area measurements of the oropharynx acquired
using different MDCT and CBCT scanners (Figure 4, Figure 5).
The most accurate volume measurements of the oropharynx were acquired using
the Siemens MDCT (98.4%; 14.3 cm3) and Vatech CBCT (98.9%; 14.4 cm3) scanners
Figure 5. The results of registration between all CT-derived oropharynx STL models (acquired using the mean threshold value) and the gold standard STL model. A = GE; B = Siemens; C = NewTom 5G; D = Accuitomo; E = Vatech; F = Gold standard STL model. Scale: Red = CT-derived STL model is larger than the gold standard; blue = CT-derived STL model is smaller than the gold standard; green = CT-derived STL model is close to the gold standard; black and white = outliers (> 0.8 mm).
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
112
7
(Figure 4 A). The GE MDCT, NewTom 5G CBCT and Accuitomo CBCT scanners resulted
in smaller volume measurements, viz., 92.1% (13.4 cm3), 91.5% (13.3 cm3), and 94.6%
(13.8 cm3), respectively. The most accurate cross-sectional area measurements of
the oropharynx were acquired using the Siemens MDCT (94.6%; 282.4 mm3), Accuitomo
CBCT (95.1%; 283.8 mm3) and Vatech CBCT (95.3%; 284.5 mm3) scanners (Figure 4 B).
The GE MDCT and NewTom 5G CBCT scanners resulted in smaller area measurements,
viz., 89.3% (266.5 mm3) and 89.8% (268.0 mm3), respectively. These results are in good
agreement with previous studies that reported an underestimation of the airway area in
both MDCT and CBCT images [25], [26]. However, it should be noted that the absolute
values of the aforementioned inaccuracies ranged between 13.3 cm3 and 14.4 cm3 (volume),
and between 266.5 mm2 and 284.5 mm2 (cross-sectional area) (Table 3). Consequently,
the authors of this study hypothesize that the reported inaccuracies should not affect
the radiological evaluation of OSA patients in clinical settings [27].
Even though the authors of this study used the recommended soft tissue imaging
protocols on all CT scanners, the STL models acquired using the GE and NewTom 5G
scanners were generally smaller in size than the gold standard STL model (Figure 4).
The smaller STL models caused a shift in the histograms towards the negative direction
(Figure 5 A, C). This phenomenon is probably due to the partial volume effect [28], in
which voxels in the vicinity of the air-to-soft tissue boundary are commonly allocated to
“soft tissue” instead of “air” during the segmentation process.
The higher accuracy of the Vatech STL models could be a result of the smaller spatial
resolution used in the default airway scanning protocol (Table 1) [29]. However, a very
recent study by Sang et al. [30] that investigated the influence of voxel size on the accuracy
of NewTom 5G and Vatech CBCT reported that increasing the voxel resolution from 0.30
to 0.15 mm does not always result in increased accuracy of 3D tooth reconstructions, while
different CBCT modalities (i.e. NewTom 5G vs. Vatech) can significantly affect the accuracy.
The largest geometric deviations were found in the uvula and epiglottis area (Figure 5).
Interestingly, the acquired oropharynx STL models were generally too large in the epiglottis
region and too small in the vicinity of the uvula. One explanation for this phenomenon
could be that the epiglottis has a concave-like geometry and the uvula is convex. These
findings are in good agreement with a previous study by Barone et al. [31] who observed
discrepancies between the segmentation of concave and convex shapes in teeth.
The results of the present study show that CBCT scanners offer an alternative to MDCT
scanners in the assessment of the oropharynx. This is in good agreement with a previous
study by Suomalainen et al. [32], who reported that CBCT scanners offer images similar to
those acquired using low-dose MDCT protocols. Therefore, taking the lower CBCT radiation
dose into consideration [11], [12], clinicians should preferably use CBCT modalities for
the analysis of the oropharynx. Moreover, all appropriate measures should be undertaken
to minimize the dose according to the International Commission on Radiological Protection
(ICRP) and the As Low As Reasonably Achievable (ALARA) principles [33].
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
113
7
Limitations of the current studyOne limitation of this study was that for each of the five CT scanners, only one image
dataset was acquired using a single image acquisition protocol. Since there are multiple
image acquisition protocols available for each MDCT and CBCT scanner, different protocols
should be considered in a future study. Another limitation was that the gold standard
values in the present study were obtained from the original STL model of the oropharynx
that was used to 3D print the phantom. 3D printing was performed using a Zprinter 250
inkjet powder printer (3D Systems, Rock Hill, USA), which has a layer thickness of 0.1 mm
and an in-plane resolution of approximately 0.05 mm [23]. Therefore, this process may
have introduced a manufacturing error, hence measurement uncertainty, of up to 0.2 mm
[34]. Nevertheless, this uncertainty can be considered clinically insignificant [35].
CONCLUSION Significant differences were observed in the volume and cross-sectional area measurements
of the oropharynx acquired using different MDCT and CBCT scanners. The Siemens MDCT
and the Vatech CBCT scanners were more accurate than the GE MDCT, NewTom 5G,
and Accuitomo CBCT scanners. In clinical settings, CBCT scanners offer an alternative to
MDCT scanners in the assessment of the oropharynx morphology.
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
114
7
REFERENCES
1. Ryan C M and Bradley T D (2005) Pathogenesis of obstructive sleep apnea. J. Appl. Physiol. 99:2440–2450
2. Aarab G, Lobbezoo F, Hamburger H L, and Naeije M (2011) Oral appliance therapy versus nasal continuous positive airway pressure in obstructive sleep apnea: a randomized, placebo-controlled trial. Respiration. 81:411–419
3. Marin J M, Carrizo S J, Vicente E, and Agusti A G N (2005) Long-term cardiovascular outcomes in men with obstructive sleep apnoea-hypopnoea with or without treatment with continuous positive airway pressure: an observational study. The Lancet 365:1046–1053
4. McNicholas W T, Strohl K P, White D, Levy P, Schmidt W, Wheatley J, Redline S, Carley D, Buysse D, Young T, Douglas N, and Fleetham J (1999) Sleep-Related Breathing Disorders in Adults: Recommendations for Syndrome Definition and Measurement Techniques in Clinical Research. Sleep 22:667–689
5. Parish J M and Somers V K (2004) Obstructive sleep apnea and cardiovascular disease. In Mayo Clinic Proceedings, 79:1036–1046
6. Villaneuva A T C, Buchanan P R, Yee B J, and Grunstein R R (2005) Ethnicity and obstructive sleep apnoea. Sleep Med. Rev. 9:419–436
7. Lee R W W, Sutherland K, and Cistulli P A (2010) Craniofacial morphology in obstructive sleep apnea: a review. Clin. Pulm. Med. 17:189–195
8. Slaats M A, Van Hoorenbeeck K, Van Eyck A, Vos W G, De Backer J W, Boudewyns A, De Backer W, and Verhulst S L (2015) Upper airway imaging in pediatric obstructive sleep apnea syndrome. Sleep Med. Rev. 21:59–71
9. Abramson Z, Susarla S, August M, Troulis M, and Kaban L (2010) Three-dimensional computed tomographic analysis of airway anatomy in patients
with obstructive sleep apnea. J. oral Maxillofac. Surg. 68:354–362
10. Ogawa T, Enciso R, Shintaku W H, and Clark G T (2007) Evaluation of Cross-section Airway Configuration of Obstructive Sleep Apnea. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod. 103:102–108
11. Ludlow J B and Ivanovic M (2008) Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surgery, Oral Med. Oral Pathol. Oral Radiol. Endodontology 106:106–114
12. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
13. Schendel S A, Broujerdi J A, and Jacobson R L (2014) Three-dimensional upper-airway changes with maxillomandibular advancement for obstructive sleep apnea treatment. Am. J. Orthod. Dentofac. Orthop. 146:385–393
14. Marcussen L, Henriksen J E, and Thygesen T (2017) Do Mandibular Advancement Devices Influence Patients’ Snoring and Obstructive Sleep Apnea? A Cone-Beam Computed Tomography Analysis of the Upper Airway Volume. J. Oral Maxillofac. Surg. 73:1816–1826
15. Cossellu G, Biagi R, Sarcina M, Mortellaro C, and Farronato G (2015) Three-dimensional evaluation of upper airway in patients with obstructive sleep apnea syndrome during oral appliance therapy. J. Craniofac. Surg. 26:745–748
16. Scarfe W C and Farman A G (2008) What is cone-beam CT and how does it work? Dent. Clin. North Am. 52:707–730
17. Lofthag-Hansen S, Thilander-Klang A, and Grondahl K (2011) Evaluation of subjective image quality in relation to
AC
CU
RA
CY O
F MD
CT A
ND
CB
CT
115
7
diagnostic task for cone beam computed tomography with different fields of view. Eur. J. Radiol. 80:483–488
18. Alsufyani N A, Flores-Mir C, and Major P W (2012) Three-dimensional segmentation of the upper airway using cone beam CT: a systematic review. Dentomaxillofacial Radiol. 41:276–284
19. Weissheimer A, de Menezes L M, Sameshima G T, Enciso R, Pham J, and Grauer D (2012) Imaging software accuracy for 3-dimensional analysis of the upper airway. Am. J. Orthod. Dentofac. Orthop. 142:801–813
20. Ghoneima A and Kula K (2013) Accuracy and reliability of cone-beam computed tomography for airway volume analysis. Eur. J. Orthod. 35:256–261
21. Schendel S A and Hatcher D (2010) Automated 3-dimensional airway analysis from cone-beam computed tomography data. J. oral Maxillofac. Surg. 68:696–701
22. Emadi N, Safi Y, Akbarzadeh Bagheban A, and Asgary S (2014) Comparison of CT-Number and Gray Scale Value of Different Dental Materials and Hard Tissues in CT and CBCT. Iran. Endod. J. 9:283–286
23. Mandal D and Syan C (eds) (2016) CAD/CAM, Robotics and factories of the future. Springer, New Delhi, India
24. Fleiss J L (1999) Design and analysis of clinical experiments. John Wiley & Sons, NY
25. Baumgaertel S, Palomo J M, Palomo L, and Hans M G (2009) Reliability and accuracy of cone-beam computed tomography dental measurements. Am. J. Orthod. Dentofac. Orthop. 136:19–25; discussion 25-28
26. Rodriguez A, Ranallo F N, Judy P F, Gierada D S, and Fain S B (2014) CT reconstruction techniques for improved accuracy of lung CT airway measurement. Med. Phys. 41:111911
27. Buchanan A, Cohen R, Looney S, Kalathingal S, and De Rossi S (2016) Cone-beam CT
analysis of patients with obstructive sleep apnea compared to normal controls. Imaging Sci. Dent. 46:9–16
28. Pham D L, Xu C, and Prince J L (2000) Current methods in medical image segmentation. Annu. Rev. Biomed. Eng. 2:315–337
29. Taft R M, Kondor S, and Grant G T (2011) Accuracy of rapid prototype models for head and neck reconstruction. J. Prosthet. Dent. 106:399–408
30. Sang Y-H, Hu H-C, Lu S-H, Wu Y-W, Li W-R, and Tang Z-H (2016) Accuracy Assessment of Three-dimensional Surface Reconstructions of In vivo Teeth from Cone-beam Computed Tomography. Chin. Med. J. (Engl). 129:1464–1470
31. Barone S, Paoli A, and Razionale A V (2016) CT segmentation of dental shapes by anatomy-driven reformation imaging and B-spline modelling. Int. J. Numer. Method. Biomed. Eng. 32:e02747
32. Suomalainen A, Kiljunen T, Kaser Y, Peltola J, and Kortesniemi M (2009) Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac. Radiol. 38:367–378
33. International Commission on Radiological Protection (2007) The 2007 Recommendations of the International Commission on Radiological Protection. Elsevier, Amsterdam
34. Lemu H G and Kurtovic S (2012) 3D Printing for Rapid Manufacturing: Study of Dimensional and Geometrical Accuracy. IFIP Adv. Inf. Commun. Technol. 384:470–479
35. Khalil W, EzEldeen M, Van De Casteele E, Shaheen E, Sun Y, Shahbazian M, Olszewski R, Politis C, and Jacobs R (2016) Validation of cone beam computed tomography-based tooth printing using different three-dimensional printing technologies. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 121:307–315
8 THE ACCURACY OF ULTRASHORT ECHO TIME MRI SEQUENCES FOR
MEDICAL ADDITIVE MANUFACTURING
Maureen van Eijnatten, Erik-Jan Rijkhorst, Mark Hofman,
Tymour Forouzanfar, Jan Wolff
Dentomaxillofacial Radiology (2016). 45(5): 20150424doi: 10.1259/dmfr.20150424
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
118
8
ABSTRACTObjectivesAdditively manufactured bone models, implants and drill guides are becoming
increasingly popular amongst maxillofacial surgeons and dentists. To date, such constructs
are commonly manufactured using computed tomography (CT) technology that induces
ionizing radiation. Recently, ultrashort echo time (UTE) magnetic resonance imaging (MRI)
sequences have been developed that allow radiation-free imaging of facial bones. The aim
of the present study was to assess the feasibility of UTE MRI sequences for medical additive
manufacturing (AM).
MethodsThree morphologically different dry human mandibles were scanned using a CT and MRI
scanner. Additionally, optical scans of all three mandibles were made to acquire a “gold
standard”. All CT and MRI scans were converted into Standard Tessellation Language (STL)
models and geometrically compared to the gold standard. To quantify the accuracy of
the AM process, the CT, MRI and gold standard STL models of one of the mandibles were
additively manufactured, optically scanned and compared to the original gold standard
STL model.
ResultsGeometric differences between all three CT-derived STL models and the gold standard
were less than 1.0 mm. All three MRI-derived STL models generally presented deviations
below 1.5 mm in the symphyseal and mandibular area. The AM process introduced minor
deviations of less than 0.5 mm.
ConclusionsThis study demonstrates that MR imaging using UTE sequences is a feasible alternative to
CT in generating STL models of the mandible, and would therefore be suitable for surgical
planning and AM. Further in-vivo studies are necessary to assess the usability of UTE MRI
sequences in clinical settings.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
119
8
INTRODUCTIONVirtual three-dimensional (3D) surgical planning and additive manufacturing (AM)
technologies are increasingly used in oral and maxillofacial surgery. To date, these
advanced technologies have proven to be invaluable in dental implant surgery [1], in
the restoration of mandibular fractures [2], in tumour resections [3] and in maxillofacial
reconstructions [4], [5]. Moreover, additively manufactured medical constructs such as
drill guides, [6] saw guides [7] or individualized implants [8] offer a predictive, functional
and aesthetic outcome, especially in patients with anatomical limitations, insufficient
mandibular or maxillary bone or poor bone density [9].
Medical AM comprises three basic steps. The first step is the acquisition of high-
resolution medical 3D images that are archived as a digital imaging and communications
in medicine (DICOM) file. The second step is to load the DICOM file into virtual planning
software that commonly converts the DICOM file automatically into a virtual 3D surface
model in the Standard Tessellation Language (STL) file format [10]. The resulting STL
model can then be used to design a medical construct using computer aided design (CAD)
software. The last step is the conversion of the STL model into a g-code that is required
to control the AM process.
Currently, 3D medical images intended for oral and maxillofacial AM are acquired
using multi-detector row computed tomography (MDCT) or cone beam computed
tomography (CBCT) technologies [11]. Both MDCT and CBCT provide high resolution
images of the maxillary and mandibular bone and the surrounding tissues [12]. However,
the aforementioned technologies have one major disadvantage: they expose the patient
to ionizing radiation. The dose levels range between 20 and 400 µSv for CBCT and around
1000 µSv for MDCT modalities, which is about 10 to 50 times more than a conventional
panoramic radiograph (about 20 µSv) [13], [14]. Even though device manufacturers
have been developing “ultra” low-dose CT scanning protocols [15]–[17], any X-ray-
based imaging technology inevitably results in some radiation exposure. Therefore, an
alternative, radiation-free imaging modality is still being sought, especially in applications
that require a large number of follow-up scans [18].
Over the past decade, MRI technology has been discussed as an alternative to CT
technology for implant planning [9], [19], [20]. To date, most conventional MRI sequences
only offer optimal soft tissue images and lack the ability to image bone/air interfaces in
the head area. This is because MR image acquisition is based on water microenvironments
and other sources of protons in the human body. Since cortical bone has a low proton
density and a very short T2 relaxation time of about 1.5 ms [21], conventional spin echo
and gradient echo MRI sequences are too slow to acquire a sufficient signal from bone.
As a result, new ultrashort echo time (UTE) sequences have been recently developed
and successfully used for bone imaging [22], [23]. The unique feature of these novel UTE
sequences is that they sample the MR signal with a minimum echo time due to radial
sampling, and rapidly switch to data recording after the radiofrequency (RF) pulse has
been delivered [23].
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
120
8
The MRI of bone using UTE sequences is being applied in several emerging clinical
technologies such as PET-MRI attenuation correction [22], high-intensity focused ultrasound
(HIFU) applications [24], MR-guided surgery [25] and MR-only treatment planning and
guidance in radiotherapy [26]. However, to the best of our knowledge no studies have
been performed on the feasibility of UTE MRI sequences in generating STL models for
AM purposes. In this study, we compare the geometric accuracy of UTE MRI-derived STL
models of three morphologically different human mandibles with CT-derived STL models.
Furthermore, additional geometric inaccuracies introduced during the AM process
are assessed.
MATERIALS AND METHODSThree morphologically different human cadaver mandibles of succumbed Dutch
patients with unknown clinical histories and intact bony structures were obtained from
the Department of Anatomy, VU University Medical Center, Amsterdam, The Netherlands.
The mandibles were boiled for 20 hours and all soft tissues were meticulously
removed manually.
The outline of this study is summarized in Figure 1. The three dry mandibles were
scanned using a CT and an MRI scanner (Step 1). In addition, an optical scan of all three
mandibles was acquired using an optical 3D scanner (Artec Spider™, Artec group, Moscow,
Russia) with a point accuracy of 0.05 mm. The optical scans of the three mandibles were
used as gold standard STL models. The CT and MRI scans were then converted to STL
files (Step 2) and aligned with the gold standard STL models. Subsequently, the geometric
deviations between the CT and MRI-derived STL models and the gold standard models
were calculated. Furthermore, the CT, MRI and gold standard STL model of one mandible
were additively manufactured (Step 3) and scanned again with the optical scanner (Step 4)
in order to map additional geometric deviations introduced during the AM process.
MRIThe mandibular cortical bone has a very short T2* of approximately 1.5 ms, and therefore
conventional spin or gradient echo sequences are too slow to acquire any bone signal. In
UTE imaging, the free induction decay signal is sampled directly with a sub-millisecond
echo time, resulting in a MR signal from the surrounding cortical bone [22], [23].
All three mandibles were scanned using an UTE sequence on a Philips Achieva 3T
MRI scanner (Philips Healthcare, Best, the Netherlands) with an eight-channel head
radiofrequency coil using a UTE single echo sequence. The following scan parameters
were used: repetition time = 4.8 ms, echo time1 = 0.14 ms and flip angle = 15°. A non-
selective hard radiofrequency pulse was used for excitation, followed by tuning the receive
coil. Immediately thereafter, free induction decay sampling was started during ramp-up of
the gradients. This procedure resulted in a time between excitation and readout of 0.14
ms. The 3D k-space was radially sampled and regridded to a Cartesian coordinate system
of isotropic 0.5 x 0.5 x 0.5 mm voxels.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
121
8
CT imagingCT imaging was performed with a Siemens Somatom Sensation 64-slice CT scanner
(Siemens Medical Solutions, Erlangen, Germany) using a low-dose scanning protocol.
The scan parameters were as follows: tube voltage = 80 kV, tube current exposure time
product = 150 mAs, pixel size = 0.3 mm x 0.3 mm, slice thickness = 0.6 mm. The images
were then reconstructed using a sharp bone convolution kernel (H60h).
Image processing and STL deviation analysisAll MRI and CT datasets were saved in DICOM file formats and subjected to the image
processing steps described in Figure 2. All CT datasets were imported into Insight
Segmentation and Registration Toolkit® software (Kitware Inc., Clifton Park, NY) and
manually thresholded using 0 Hounsfield Units (HU). The MRI datasets were thresholded
using 15 arbitrary units (AU) (mandible 1 and 2), and 12 AU (mandible 3) (Figure 2, Step 1).
These threshold values were chosen just above the MRI background noise signal level to
acquire an optimal model of the mandibular bone. Subsequently, all segmented datasets
were converted to STL models (Figure 2, Step 2) and imported into MeshLab software
(Visual Computing Lab, Pisa, Italy). The MRI and CT-derived STL models were aligned with
their corresponding gold standard STL models using an iterative closest point algorithm
(Figure 2, Step 3). The signed STL-to-STL distances, that is geometric deviations, between
all MRI and CT-derived STL models and the gold standard STL model were computed
Figure 1. Outline of the study. STL, Standard Tessellation Language; UTE, ultrashort echo time.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
122
8
using Visualization Toolkit® software (Kitware Inc.) (Figure 2, Step 4) and visualised using
colour maps and histograms (Figure 3).
The 95th percentiles of all STL-to-STL distances were calculated using MatLab software
v.2012, (MathWorks®, Natick, MA). To determine the geometric accuracy in the different
areas of interest for surgical planning, the 95th percentiles were calculated for the whole
mandible as well as for five different anatomical regions: 1) right ramus with condyle;
2) right body; 3) symphyseal area with alveolar ridge; 4) left body; 5) left ramus with
condyle (Table 1).
Additive manufacturingThe CT and MRI-derived STL models of mandible 1 were imported into GOM Inspect®
software (GOM GmbH, Braunschweig, Germany) and smoothed with a surface tolerance of
1.0 mm to correct stair-step artefacts. The smoothed CT-derived STL model, the smoothed
MRI-derived STL model and the gold standard STL model of mandible 1 obtained using
the optical scanner were all printed using a Zprinter 250 inkjet powder printer (3D Systems
Inc., Rock Hill, SC). The resulting additively manufactured models were once again scanned
with an optical scanner, aligned with the gold standard STL model, and subjected to STL
deviation analysis according to Steps 3 and 4 in Figure 2.
RESULTSAll geometric deviations between the optical scan “gold standard” STL models and
the CT-derived STL models were below 1.0 mm (Figure 3 A-C), with 95th percentiles < 0.5
mm (Table 1), with the exception of Mandible 1 (0.71 mm). The MRI-derived STL models
generally presented geometric deviations < 2.0 mm (Figure 3 D-F), with 95th percentiles <
1.5 mm (Table 1), again except for Mandible 1 (1.69 mm).
Figure 4 presents the mean and standard deviation (SD) of the signed geometric
deviations in the five different anatomical regions of all CT and MRI STL models. The mean
value indicates to what extent the STL model was smaller or larger than the gold standard,
and the SD indicates the spread in geometric deviations between the STL model and
Figure 2. Overview of image processing steps and software packages involved in the STL analysis. The Insight Segmentation and Registration Toolkit® software was obtained from Kitware Inc., Clifton Park, NY, the MeshLab software was obtained from Visual Computing Lab, Pisa, Italy, and the Visualization Toolkit® software from Kitware Inc., Clifton Park, NY. DICOM, digital imaging and communications in medicine.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
123
8
the gold standard. The deviations in all CT-derived STL models were uniformly distributed
over the different anatomical regions, whereas all MRI-derived STL models showed
a small spread of deviations in the symphyseal area and a larger spread of deviations in
the mandibular condyles (Figure 4).
Minor additional geometric deviations were reported in the additively manufactured
3D models of Mandible 1 (Figure 5). The additively manufactured gold standard 3D model
demonstrated geometric deviations of up to 0.5 mm (Table 2), especially in thin structures
such as the coronoid process and alveolar ridge. Geometric deviations in the additively
manufactured 3D model obtained using CT data (Figure 5 B) were mostly in the vicinity of
the coronoid process. In the 3D model obtained using MRI data (Figure 5 C), deviations
were observed in the coronoid process, mandibular condyles and alveolar ridge.
Figure 3. (A-F) Deviations between the CT-derived Standard Tessellation Language (STL) models and the gold-standard STL model (a–c), and the MRI-derived STL models and the gold-standard STL model (d–f), shown in colour maps and histograms.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
124
8
Dev
iatio
n(m
m)
1.Ram
us(R
)
2.Bod
y (R)
3.Alve
olar rid
ge
4.Bod
y (L)
5.Ram
us(L)
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5CT vs. gold std. MRI vs. gold std.
Dev
iatio
n(m
m)
1.Ram
us(R
)
2.Bod
y (R)
3.Alve
olar rid
ge
4.Bod
y (L)
5.Ram
us(L)
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5CT vs. gold std. MRI vs. gold std.
Dev
iatio
n(m
m)
1.Ram
us(R
)
2.Bod
y (R)
3.Alve
olar rid
ge
4.Bod
y (L)
5.Ram
us(L)
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5CT vs. gold std. MRI vs. gold std.
Mandible 1 Mandible 2 Mandible 3
Figure 4. Mean ± standard deviation of the geometric deviations of five anatomical regions of the mandible in CT- and MRI-derived Standard Tessellation Language models. Std., standard.
Table 1. The 95th percentiles of the geometric deviations of five anatomical regions of the mandible in CT and MRI-derived Standard Tessellation Language models.
CT MRI
Mandible 1 (mm)
Mandible 2 (mm)
Mandible 3 (mm)
Mandible 1 (mm)
Mandible 2 (mm)
Mandible 3 (mm)
Whole mandible 0.71 0.47 0.40 1.69 1.52 1.25 1. Right ramus with condyle 0.49 0.33 0.42 1.89 1.46 1.55 2. Right body 0.30 0.45 0.43 1.03 1.23 0.96 3. Symphyseal area 0.45 0.56 0.45 1.64 0.96 0.70 4. Left body 0.62 0.56 0.30 1.02 1.30 0.86 5. Left ramus with condyle 0.93 0.45 0.34 2.04 1.91 1.53
Table 1 The 95th percentiles of the geometric deviations of five anatomical regions of the mandible in CT and
MRI-derived Standard Tessellation Language models.
CT MRI
Mandible 1
(mm)
Mandible 2
(mm)
Mandible 3
(mm)
Mandible 1
(mm)
Mandible 2
(mm)
Mandible 3
(mm)
Whole mandible 0.71 0.47 0.40 1.69 1.52 1.25
1. Right ramus with condyle 0.49 0.33 0.42 1.89 1.46 1.55
2. Right body 0.30 0.45 0.43 1.03 1.23 0.96
3. Symphyseal area 0.45 0.56 0.45 1.64 0.96 0.70
4. Left body 0.62 0.56 0.30 1.02 1.30 0.86
5. Left ramus with condyle 0.93 0.45 0.34 2.04 1.91 1.53
Table 2. The 50th, 95th and 99th percentiles of the geometric deviations in additively manufactured three-dimensional models of Mandible 1.
Percentile Gold standard (mm) CT (mm) MRI (mm)
50th 0.11 0.21 0.48 95th 0.46 0.88 1.73 99th 1.12 1.32 2.63
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
125
8
Figure 5. (A-C) Deviations between the additively manufactured (AM) gold-standard three-dimensional (3D) model and the gold-standard Standard Tessellation Language (STL) model; the AM CT-derived 3D model and the gold-standard STL model; and the AM MRI-derived 3D model and the gold-standard STL model shown in colour maps and histograms.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
126
8
DISCUSSIONToday, there is an ever increasing need for high-resolution 3D images for use in surgical
planning [27], [28]. In this context, CT and CBCT imaging technologies are commonly used
and subsequently induce harmful radiation to the patient [29]. The cumulative lifetime
radiation dose of the population needs to be minimized by reducing the number of overall
examinations and the dose resulting from each individual exposure [30]. Hence, radiation-
free imaging modalities are still being sought for medical additive manufacturing.
MR and CT imagingThe novel UTE MRI sequence used in this study produced STL models that were
morphologically comparable to those acquired using CT technology (Figure 3). In
the MRI- derived STL models, the 95th percentile was generally found to be < 1.5 mm
(Table 1). The only exception was Mandible 1, which had a 95th percentile of 1.69 mm.
These larger geometric deviations were partially due to the submillimetre-thin alveolar
ridge of the mandible that could not be imaged due to the limited spatial resolution
of the MRI scanner (0.5 mm). Indeed, the 95th percentile of the geometric deviations in
the symphyseal area of mandible 1 was 1.64 mm, compared with 0.96 mm and 0.70 mm for
Mandible 2 and 3, respectively (Table 1). Another anatomical region that showed slightly
larger geometrical deviations in the MRI- derived STL models of all three mandibles were
the condyles (Table 1). This phenomenon was most likely caused by a drop in the UTE
MRI signal below the chosen threshold value at the edge of the condyles. This lower
signal could be due to the higher cortical bone density in the condyle regions. In clinical
settings, the presence of soft tissues would offer the possibility to use more advanced
segmentation methods [22], instead of the thresholding approach used in this study.
In MRI technology, the spatial resolution is dependent on the number of frequency
encoding steps and the size of the field of view. Since the size of the mandible determines
the minimal field of view, the resolution is dependent on the frequency encoding, that
is how often the free induction decay signal is sampled. Therefore, the resolution of
the current UTE MRI sequence could be further improved by increasing the duration of
the scan; however, this would be a disadvantage in a clinical setting.
In the CT-derived STL models, the 95th percentile of the geometric deviations
in the whole mandible was generally found to be < 0.5 mm (Table 1). All geometrical
deviations were < 1.0 mm (Figure 3). The capability of CT to capture small bone structures
is also dependent on the spatial resolution of the scanner (0.3 mm), which is directly linked
to the hardware properties such as the detector configuration of the scanner.
The aforementioned results are in agreement with a previous study by White et al. [31],
who obtained MRI and CT-derived additively manufactured 3D models of 10 ovine knees,
and compared caliper measurements of these models with caliper measurements taken
from the real bony anatomy. They reported a mean deviation of 2.15 ± 2.44 mm in their
MRI models and a mean deviation of 0.61 ± 0.41 mm in their CT models. Lee et al. [32]
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
127
8
reported a mean deviation of 0.7 ± 0.1 mm between their MRI and CT-derived STL models
of the femur. A more recent study by Van Den Broeck et al. [33] reported RMS errors of
0.55 mm in their CT-derived STL models of the tibia, and 0.56 mm in their MRI-derived STL
models. The results from our study differ from the ones reported by Lee et al. [32] and Van
Den Broeck et al. [33]. This may have been due to differences in the evaluated anatomical
structures and the measuring protocols. Moreover, the differences may also have been
caused by the different MRI sequences used. White et al. [31], Van Den Broeck et al. [33]
and Lee et al. [32] all used conventional spin or gradient echo sequences that commonly
require more manual segmentation of the bone, which in return can be subjective and very
time-consuming.
The UTE MRI protocol used in this study has certain clinical advantages when compared
to CT technologies. The major advantage of MR imaging is the lack of radiation exposure,
which in return allows for multiple imaging sessions to be undertaken. Furthermore,
the UTE sequence used in this study combined with other sequences available on the same
MRI device allows for hard and soft tissue visualisation, respectively. This can be especially
valuable when treating cancer, ameloblastoma patients and orbital fractures [34], [35].
Additive manufacturingIn order to quantify the accuracy of the AM process, the gold standard STL model of
Mandible 1 was first additively manufactured and then optically scanned and compared to
the original STL model (Figure 5 A). The 95th percentile of geometric deviations introduced
during the AM process using the gold standard STL model was 0.46 mm (Table 2). This
result suggests that only minor deviations were introduced during the AM process.
However, because an optical scanner was used to image the bone, the aforementioned
results are not obtainable in clinical settings.
The 95th percentiles of geometric deviations in the MRI and CT-derived additively
manufactured models were higher than those obtained with the optical scanner: 1.73 mm
and 0.88 mm, respectively (Table 2). These results suggest that both CT and MR imaging
have a greater influence on the accuracy of medical additive manufactured constructs than
the additive manufacturing process itself.
Limitations of this studyThe major limitation of the present study was that mandibles without soft tissues were
used. Therefore, the results of this study are not simply generalizable to clinical conditions.
In clinical settings, a second UTE MRI echo image is required to discriminate bone from air
and soft tissues [22]. Furthermore, motion artefacts due to the relatively long duration of
the UTE sequence (several minutes) and metal artefacts caused by orthodontic appliances
or fillings can affect the image quality in vivo. Further cadaver and patient studies are
recommended in order to assess the feasibility of UTE MRI sequences in clinical settings.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
128
8
CONCLUSIONThis study demonstrates that MR imaging of bone using an UTE sequence is a feasible
alternative to CT imaging in the generation of STL models of mandibles with different
morphologies, and would therefore be suitable for surgical planning and additive
manufacturing. The CT and MRI modalities generally introduced geometrical deviations
in the STL models of < 1.0 mm and < 1.5 mm, respectively. The additive manufacturing
process introduced minor deviations of < 0.5 mm. Further in vivo studies are necessary to
assess the feasibility of the UTE MRI sequence in clinical settings.
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
129
8
REFERENCES
1. Ramasamy M, Giri, Raja R, Subramonian, Karthik, and Narendrakumar R (2013) Implant surgical guides: From the past to the present. J. Pharm. Bioallied Sci. 5:S98–S102
2. Voss J O, Varjas V, Raguse J-D, Thieme N, Richards R G, and Kamer L (2016) Computed tomography-based virtual fracture reduction techniques in bimandibular fractures. J. Craniomaxillofac. Surg. 44:177–185
3. Yu H, Wang X, Zhang S, Zhang L, Xin P, and Shen S G (2013) Navigation-guided en bloc resection and defect reconstruction of craniomaxillary bony tumours. Int. J. Oral Maxillofac. Surg. 42:1409–1413
4. Schepers R H, Kraeima J, Vissink A, Lahoda L U, Roodenburg J L, Reintsema H, Raghoebar G M, and Witjes M J (2016) Accuracy of secondary maxillofacial reconstruction with prefabricated fibula grafts using 3D planning and guided reconstruction. J. Cranio-Maxillofacial Surg. 44:392-399
5. Li Y, Jiang Y, Ye B, Hu J, Chen Q, and Zhu S (2016) Treatment of Dentofacial Deformities Secondary to Osteochondroma of the Mandibular Condyle Using Virtual Surgical Planning and 3-Dimensional Printed Surgical Templates. J. Oral Maxillofac. Surg. 74:349–368
6. Dawood A, Marti B M, Sauret-Jackson V, and Darwood A (2015) 3D printing in dentistry. Br Dent J 219:521–529
7. Mazzoni S, Bianchi A, Schiariti G, Badiali G, and Marchetti C (2015) Computer-aided design and computer-aided manufacturing cutting guides and customized titanium plates are useful in upper maxilla waferless repositioning. J. Oral Maxillofac. Surg. 73:701–707
8. Kontio R (2014) Update on mandibular reconstruction: computer-aided design, imaging, stem cells and future
applications. Curr. Opin. Otolaryngol. Head Neck Surg. 22:307–315
9. Aguiar M F, Marques A P, Carvalho A C P, and Cavalcanti M G P (2008) Accuracy of magnetic resonance imaging compared with computed tomography for implant planning. Clin. Oral Implants Res. 19:362–365
10. Huotilainen E, Jaanimets R, Valášek J, Marcián P, Salmi M, Tuomi J, Mäkitie A, and Wolff J (2014) Inaccuracies in additive manufactured medical skull models caused by the DICOM to STL conversion process. J. Cranio-Maxillofacial Surg. 42:e259–e265
11. Huotilainen E, Paloheimo M, Salmi M, Paloheimo K-S, Björkstrand R, Tuomi J, Markkola A, and Mäkitie A (2014) Imaging requirements for medical applications of additive manufacturing. Acta radiol. 55:78–85
12. Rengier F, Mehndiratta A, von Tengg-Kobligk H, Zechmann C M, Unterhinninghofen R, Kauczor H-U, and Giesel F L (2010) 3D printing based on imaging data: review of medical applications. Int. J. Comput. Assist. Radiol. Surg. 5:335–341
13. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
14. Lorenzoni D C, Bolognese A M, Garib D G, Guedes F R, and and Eduardo Franzotti SantAnna (2012) Cone-Beam Computed Tomography and Radiographs in Dentistry: Aspects Related to Radiation Dose. Int. J. Dent. 2012:813768
15. Nauer C B, Rieke A, Zubler C, Candreia C, Arnold A, and Senn P (2011) Low-Dose Temporal Bone CT in Infants and Young
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
130
8
Children: Effective Dose and Image Quality. Am. J. Neuroradiol. 32:1375–1380
16. Ono K, Hiraoka T, Ono A, Komatsu E, Shigenaga T, Takaki H, Maeda T, Ogusu H, Yoshida S, Fukushima K, and Kai M (2013) Low-dose CT scan screening for lung cancer: comparison of images and radiation doses between low-dose CT and follow-up standard diagnostic CT. Springerplus 2:393
17. Lutz J, Jager V, and Hempel M J (2007) Delineation of temporal bone anatomy: feasibility of low-dose 64-row CT in regard to image quality. Eur. J. Radiol. 17:2638–2645
18. Singh S, Kalra M K, and Moore M A (2009) Dose reduction and compliance with pediatric CT protocols adapted to patient size, clinical indication, and number of prior studies. Radiology 252:200–208
19. Imamura H, Sato H, Matsuura T, Ishikawa M, and Zeze R (2004) A comparative study of computed tomography and magnetic resonance imaging for the detection of mandibular canals and cross-sectional areas in diagnosis prior to dental implant treatment. Clin. Implant Dent. Relat. Res. 6:75–81
20. Kawai T, Asaumi R, Kagawa T, Inadomi D, Yuasa K, and Yosue T (2014) The observation of the incisive branches of mandibular canal – the comparison between MRI vs. CBCT. Clin. Oral Implants Res. 25:2014
21. Horch R A, Nyman J S, Gochberg D F, Dortch R D, and Does M D (2010) Characterization of 1H NMR signal in human cortical bone for magnetic resonance imaging. Magn. Reson. Med. 64:680–687
22. Keereman V, Fierens Y, Broux T, De Deene Y, Lonneux M, and Vandenberghe S (2010) MRI-based attenuation correction for PET/MRI using ultrashort echo time sequences. J. Nucl. Med. 51:812–8
23. Aitken A P, Giese D, Tsoumpas C, Schleyer P, Kozerke S, Prieto C, and
Schaeffter T (2014) Improved UTE-based attenuation correction for cranial PET-MR using dynamic magnetic field monitoring. Med. Phys. 41:12302
24. Huisman M, Lam M K, Bartels L W, Nijenhuis R J, Moonen C T, Knuttel F M, Verkooijen H M, van Vulpen M, and van den Bosch M A (2014) Feasibility of volumetric MRI-guided high intensity focused ultrasound (MR-HIFU) for painful bone metastases. J. Ther. Ultrasound 2:16
25. Liao H, Inomata T, Sakuma I, and Dohi T (2010) 3-D augmented reality for MRI-guided surgery using integral videography autostereoscopic image overlay. IEEE Trans. Biomed. Eng. 57:1476–1486
26. Metcalfe P, Liney G P, Holloway L, Walker A, Barton M, Delaney G P, Vinod S, and Tomé W (2013) The Potential for an Enhanced Role for MRI in Radiation-Therapy Treatment Planning. Technol. Cancer Res. Treat. 12:429–446
27. Schendel S A and Jacobson R (2009) Three-Dimensional Imaging and Computer Simulation For Office-Based Surgery. J. Oral Maxillofac. Surg. 67:2107–2114
28. Giannatsis J and Dedoussis V (2009) Additive fabrication technologies applied to medicine and health care: a review. Int. J. Adv. Manuf. Technol. 40:116–127
29. Koivisto J (2015) The use of MOSFET dosimeters and anthropomorphic phantoms in low dose dental CBCT applications. Helsinki, Finland: University of Helsinki
30. Implications of commission recommendations that doses be kept as low as readily achievable (1973) ICRP publication 22. Oxford, UK: Pergamon Press
31. White D, Chelule K L, and Seedhom B B (2008) Accuracy of MRI vs CT imaging with particular reference to patient specific templates for total knee replacement surgery. Int. J. Med. Robot. Comput. Assist. Surg. 4:224–231
THE
AC
CU
RA
CY O
F ULTR
ASH
OR
T EC
HO
TIME
MR
I SEQ
UE
NC
ES
131
8
32. Lee Y S, Seon J K, Shin V I, Kim G-H, and Jeon M (2008) Anatomical evaluation of CT-MRI combined femoral model. Biomed. Eng. Online 7:6
33. Van den Broeck J, Vereecke E, Wirix-Speetjens R, and Vander Sloten J (2014) Segmentation accuracy of long bones. Med. Eng. Phys. 36:949–953
34. Apajalahti S, Kelppe J, Kontio R, and Hagström J (2015) Imaging characteristics of ameloblastomas and diagnostic value of computed
tomography and magnetic resonance imaging in a series of 26 patients. Oral Surg. Oral Med. Oral Pathol. Oral Radiol. 120:e118–e130
35. Schmutz B, Rahmel B, McNamara Z, Coulthard A, Schuetz M, and Lynham A (2014) Magnetic resonance imaging: An accurate, radiation-free, alternative to computed tomography for the primary imaging and three-dimensional reconstruction of the bony orbit. J. Oral Maxillofac. Surg. 72:611–618
9 GENERAL DISCUSSION
GE
NE
RA
L DISC
USSIO
N
135
9
Even though surgical techniques have improved over time, surgery still remains a manual
task. Current surgical techniques in the field of oral and maxillofacial surgery are still heavily
dependent on the dexterity and experience of the surgeon. In this context, medical AM
opens new avenues to improve the dexterity of the surgeon during surgical procedures.
The resulting decrease in operation time and improved accuracy can markedly reduce
treatment costs. For example, AM can be combined with virtual planning to create
accurate and low-cost individualized orbital floor implants. This novel treatment method
has shown to reduce operation time by approximately 40% [1]. However, creating accurate
medical AM constructs still requires technical and computational knowledge that can be
a challenge for clinicians. Therefore, AM constructs are currently most often designed
by engineers with in-depth knowledge of the field of anatomical modelling and reverse
engineering.
While the aforementioned AM constructs can be fabricated with great accuracy
(< 0.1 mm) [2] using current printing technologies, their design is currently limited by
the accuracy of the medical images used during the fabrication process. It must be noted
that image quality can differ markedly between MDCT, DECT and CBCT scanners. [3]
Moreover, scanner-specific acquisition parameters can also influence the image quality
[4]. Of all the scanners on the market, MDCT scanners offer the best image quality
and accuracy for medical AM [3]. However, radiation safety still needs to be taken into
consideration when acquiring images for AM, given the current concerns of radiation-
induced secondary malignancies [5]. Therefore, if possible, low-dose MDCT scanning
protocols should be used to minimise the effective dose in patients who require AM
constructs [3]. Even though current CBCT devices can attain lower dose values than MDCT
devices in the maxillofacial area [6], their image quality remains problematic for medical
AM purposes. The CBCT-derived STL models evaluated in this PhD thesis demonstrated
large non-uniform inaccuracies of up to 0.9 mm in the maxillofacial area (Chapter 3), which
was probably due to the relatively high amount of noise and number of artefacts [7] and
inhomogeneities in the CBCT X-ray beam [8]. Bearing the above-mentioned concerns in
mind, more emphasis should be put into developing UTE MRI sequences that are specially
designed for AM purposes [9].
One possible method of improving the overall image quality of CBCT scanners in
the maxillofacial region is to change the head position in the gantry. According to our
studies, the best CBCT image quality can be achieved on a mobile gantry device using prone
or supine imaging positions [10]. More specifically, the prone imaging position resulted in
the best image quality for the sinus, mandible and maxilla when compared with the supine
and oblique imaging positions [11]. The oblique imaging position offered inadequate
image quality except in the sinus region. These novel insights regarding the image quality
of CBCT scanners may help to improve the accuracy of CBCT images for medical AM
purposes. A more advanced method of improving the accuracy of CBCT images would
be to develop adaptive, flexible scan protocols that would allow physicians to zoom into
anatomical areas of interest during image acquisition. For example, an initial “coarse”
GE
NE
RA
L DISC
USSIO
N
136
9
CBCT scan of a patient’s head could be acquired using a large field of view, followed
by a more precise local scan of the region of interest, e.g., a tumour or fracture site. This
process would, however, require real-time reconstruction of CBCT images using powerful
cluster computers. Another promising new technology that could revolutionise the overall
image quality required in the AM process is spectral imaging [12]. Spectral imaging refers
to a technique that uses photon-counting detectors to separate photons coming from
a polychromatic X-ray source into a series of energy bins, for which the energy threshold
levels can be configured using dedicated software. Each tissue type in the human body
has a different response curve to photon energies. Therefore, it is possible to discriminate
tissues based on their response to different photon energies. This could greatly improve
tissue discrimination and delineation in CT and CBCT images.
In addition to the aforementioned CT imaging process, the conversion of the resulting
images into STL models remains a challenge in medical AM. The most challenging step in
the conversion process is image segmentation. Our literature review on bone segmentation
methods used in medical AM (Chapter 6) revealed a large spread in accuracies amongst
the current methods used ranging from 0.04 mm to 1.9 mm. Image segmentation is
often compromised by factors such as the partial volume effect, scattering artefacts and
noise [13].
To date, global thresholding remains the most commonly used segmentation method
in medical AM. However, the selection of a suitable threshold value for bone still remains
a manual task that can introduce geometric variations amongst engineers [14]. Generally,
global thresholding achieves accuracies under 0.6 mm [15]. Advanced thresholding
methods could improve the accuracy of CT-derived STL models to under 0.38 mm [16].
However, such methods are, to the best of our knowledge, not included in the FDA-
approved software packages for medical AM. Finally, it should be noted that the triangle-
based STL file format was not initially developed to describe complex biological structures
that encompass many curved surfaces and different tissue densities. Consequently, many
small triangles are needed to accurately describe a curved surface. A solution could be to
employ novel file formats such as STL2 [17] or the additive manufacturing file format (AMF)
[18]. However, these novel file formats are not yet supported by the software packages
and printers currently used in medical AM.
Current CT and CBCT scanners deliver voxel-based datasets in which anatomical
structures are represented by a continuous spectrum of grey values that are related
to attenuation coefficients. Medical AM, on the other hand, requires binary datasets
in the form of STL models. Therefore, the current medical AM process always requires
a conversion from continuous data into binary data, i.e., image segmentation. A major
drawback with the image segmentation process is that a large amount of the valuable
biological information that was initially present in the CT image in the form of grey values
is discarded. This can be considered as a waste of photons. One solution to this problem
would be to develop an algorithm that could directly compute mesh volumes (STL
models) from raw CT projection data. This could be achieved by restricting the number
GE
NE
RA
L DISC
USSIO
N
137
9
of grey values in a reconstructed CT image to a small, discrete number of values known
in advance: one for each tissue type, i.e., “air”, “soft tissue”, “cancellous bone” and
“cortical bone”. Consequently, since each voxel is already assigned to a specific tissue
type during the reconstruction phase, image segmentation as a whole becomes redundant.
This technique is known as discrete tomography [19]. The major advantage of discrete
tomography over conventional CT is that it maximises the amount of information obtained
from a CT scan for a given dose level: “making the most out of each photon”. Therefore,
I hypothesize that discrete tomography could markedly reduce the number of projections
and radiation dose needed to acquire an accurate STL model.
As mentioned above, the creation of accurate medical AM constructs still requires
a large amount of human intervention by experienced clinicians and engineers, particularly
in the imaging and image processing steps. This problem could be tackled by automating
certain processes in the medical AM workflow using novel artificial intelligence (AI)
approaches [20]. The basic mathematical concept of AI dates back to the 1940s when
researchers proposed the development of an electrical network that mimics the function of
neurons in the human brain [21]. This was the onset for further developments in the field of
AI and led to a variety of machine learning (ML) algorithms that are today used in the fields
of computer vision and medical imaging, among others. In recent years, the development
of graphic processing units (GPUs) has enabled the use of multi-layered ML algorithms
such as convolutional neural networks (CNNs) [22]. Currently, CNNs are becoming
the methodology of choice for analysing medical images [23] and show great potential
in performing accurate and automated image segmentation tasks [24]–[26]. Furthermore,
a CNN can be taught to reconstruct complex 3D shapes from CT data [27] and could
therefore theoretically be used to auto-complete STL models of disfigured skulls. A CNN
can be taught in a supervised or unsupervised manner. To date, the majority of CNNs
in the field of medical imaging use supervised learning, in which a CNN “learns from
experience” based on a large amount of labelled training data (e.g., medical images).
It should be pointed out that the performance of this class of CNN depends heavily on
the availability and the quality of the images used for training. As opposed to supervised
learning, in unsupervised learning a CNN is taught to independently discover patterns
in data. Taking the above into account, one can postulate that recent developments in
AI combined with medical imaging (“big”) data will revolutionise healthcare and will
subsequently offer new possibilities in the field of medical AM, virtual/augmented reality
and robotics. These developments will most likely lead to more automation, which, in
turn, will have a fundamental socio-economic impact on the global labour market in
the 21st century.
GE
NE
RA
L DISC
USSIO
N
138
9
REFERENCES
1. Vehmeijer M, van Eijnatten M, Liberton N, and Wolff J (2016) A Novel Method of Orbital Floor Reconstruction Using Virtual Planning, 3-Dimensional Printing, and Autologous Bone. J. Oral Maxillofac. Surg. 74:1608-1612
2. Pinto J M, Arrieta C, Andia M E, Uribe S, Ramos-Grez J, Vargas A, Irarrazaval P, and Tejos C (2015) Sensitivity analysis of geometric errors in additive manufacturing medical models. Med. Eng. Phys. 37:328–334
3. van Eijnatten M, Berger F H, de Graaf P, Koivisto J, Forouzanfar T, and Wolff J (2017) Influence of CT parameters on STL model accuracy. Rapid Prototyp. J. 23:679–685
4. Huotilainen E, Paloheimo M, Salmi M, Paloheimo K-S, Björkstrand R, Tuomi J, Markkola A, and Mäkitie A (2014) Imaging requirements for medical applications of additive manufacturing. Acta radiol. 55:78–85
5. Mathews J D, Forsythe A V, Brady Z, Butler M W, Goergen S K, Byrnes G B, Giles G G, Wallace A B, Anderson P R, Guiver T A, McGale P, Cain T M, Dowty J G, Bickerstaffe A C, and Darby S C (2013) Cancer risk in 680 000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346:f2360
6. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J, Walker A, Cockmartin L, Bosmans H, Jacobs R, Bogaerts R, and Horner K (2012) Effective dose range for dental cone beam computed tomography scanners. Eur. J. Radiol. 81:267–271
7. Pauwels R, Nackaerts O, Bellaiche N, Stamatakis H, Tsiklakis K, Walker A, Bosmans H, Bogaerts R, Jacobs R, and Horner K (2013) Variability of dental cone beam CT grey values for density estimations. Br. J. Radiol. 86:20120135
8. Feldkamp L A, Davis L C, and Kress J W (1984) Practical cone-beam algorithm. J. Opt. Soc. Am. 1:612
9. van Eijnatten M, Rijkhorst E-J, Hofman M, Forouzanfar T, and Wolff J (2016) The accuracy of ultrashort echo time MRI sequences for medical additive manufacturing. Dentomaxillofacial Radiol. 45:20150424
10. Koivisto J, van Eijnatten M, Jarnstedt J, Holli-Helenius K, Dastidar P, and Wolff J (2017) Impact of prone, supine and oblique patient positioning on CBCT image quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region. Dentomaxillofac. Radiol. 46:20160418
11. Heiland M, Schulze D, Blake F, and Schmelzle R (2005) Intraoperative imaging of zygomaticomaxillary complex fractures using a 3D C-arm system. Int. J. Oral Maxillofac. Surg. 34:369–375
12. Cormode D P, Si-Mohamed S, Bar-Ness D, Sigovan M, Naha P C, Balegamire J, Lavenne F, Coulon P, Roessl E, Bartels M, Rokni M, Blevis I, Boussel L, and Douek P (2017) Multicolor spectral photon-counting computed tomography: in vivo dual contrast imaging with a high count rate scanner. Scientific Reports, 7:4784
13. Sharma N and Aggarwal L M (2010) Automated medical image segmentation techniques. J. Med. Phys. 35:3–14
14. van Eijnatten M, Koivisto J, Karhu K, Forouzanfar T, and Wolff J (2016) The impact of manual threshold selection in medical additive manufacturing. Int. J. Comput. Assist. Radiol. Surg. 12:607–615
15. Gelaude F, Van der Sloten J, and Lauwers B (2008) Accuracy assessment of CT-based outer surface femur meshes. Comput. Aided Surg. 13:188–199
16. Rathnayaka K, Sahama T, Schuetz M A, and Schmutz B (2011) Effects of CT image segmentation methods on the accuracy
GE
NE
RA
L DISC
USSIO
N
139
9
of long bone 3D reconstructions. Med. Eng. Phys. 33:226–33
17. Hiller J D and Lipson H (2009) STL 2.0: a proposal for a universal multi-material Additive Manufacturing File format. In: Proceedings of the Solid Freeform Fabrication Symposium, 266–278
18. ASTM (2016) Standard Specification for Additive Manufacturing File Format (AMF). Available: http://standards.iso.org/iso/52915
19. Batenburg K J and Sijbers J (2006) Discrete tomography from micro-CT data: application to the mouse trabecular bone structure. In Proc. of SPIE 6142: 614240–614241
20. Lugar G and Stubblefield W (1993) Artificial Intelligence: structures and strategies for complex problem solving. Pearson education
21. McCulloch W S and Pitts W (1943) A logical calculus of the ideas immanent in nervous activity. Bull. Math. Biophys. 5:115–133
22. Hinton G E, Osindero S, and Teh Y-W (2006) A fast learning algorithm for deep belief nets. Neural Comput. 18:1527–1554
23. Litjens G, Kooi T, Bejnordi B E, Setio A A A, Ciompi F, Ghafoorian M, van der Laak J A W M, van Ginneken B, and Sánchez C I (2017) A survey on deep learning in medical image analysis. arXiv Prepr. arXiv1702.05747
24. Krizhevsky A, Sutskever I, and Geoffrey E. H (2012) ImageNet Classification with Deep Convolutional Neural Networks. Adv. Neural Inf. Process. Syst. 25:1–9
25. Prasoon A, Petersen K, Igel C, Lauze F, Dam E, and Nielsen M (2013) Deep feature learning for knee cartilage segmentation using a triplanar convolutional neural network. In: Lecture Notes in Computer Science, 8150 LNCS: 246–253
26. De Brébisson A and Montana G (2015) Deep neural networks for anatomical brain segmentation. In: IEEE Computer Society Conference on Computer Vision and Pattern Recognition Workshops. 20–28
27. Wu Z, Song S, Khosla A, Yu F, Zhang L, Tang X, and Xiao J (2015) 3d shapenets: A deep representation for volumetric shapes. In: Proceedings of the IEEE Conference on Computer Vision and Pattern Recognition. 1912–1920
10 ENGLISH SUMMARY & DUTCH SUMMARY
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
143
10
SUMMARYAM offers possibilities to fabricate different types of patient-specific constructs such as
anatomical models, surgical guides and implants. Such constructs can improve the surgeon’s
dexterity, decrease operation time and improve accuracy of surgical procedures. As
demonstrated in Chapter 2, AM can be combined with virtual planning to create accurate
and low-cost individualised orbital floor implants. A virtual 3D model of a patient with
an orbital floor fracture was generated from CT images of the patient. The fracture was
virtually closed using spline interpolation. The resulting 3D model was used to design
and manufacture a mould of the defect site using an inkjet printer. This tangible mould
was subsequently used during surgery to sculpture an individualised autologous orbital
floor implant. This novel treatment method enhanced the overall accuracy and efficiency
of the surgical procedure. The sculptured autologous orbital floor implant showed an
excellent fit in vivo and reduced the operation time by approximately 40%.
Despite recent advances and promising case studies involving medical AM, notable
scientific, technological and regulatory challenges remain. Each step in the medical AM
process, i.e., imaging (step 1), image processing (step 2) and manufacturing (step 3) can
introduce inaccuracies in the resulting construct. However, since the aforementioned
AM constructs can be fabricated with great accuracy (< 0.1 mm) using current printing
technologies, their design is currently limited by the accuracy of the imaging and image
processing steps. Therefore, the general aim of this thesis was to provide an insight into
the current scientific and technological challenges faced in medical additive manufacturing
with respect to imaging and image processing. More specifically, the different parameters
that influence the accuracy of patient-specific AM constructs were identified.
In Chapter 3, the image quality and the accuracy of STL models acquired using different
CT scanners and acquisition parameters is assessed. Images of three dry human skulls
were acquired using different scanning protocols on two multi-detector row computed
tomography (MDCT) scanners, a dual energy computed tomography (DECT) scanner and
one cone beam computed tomography (CBCT) scanner. All images were ranked according
to their image quality and converted into STL models. These STL models were subsequently
compared with corresponding gold standard models acquired using an optical 3D surface
scanner. It was found that the image quality differed between the MDCT, DECT and CBCT
scanners. Images acquired using low-dose MDCT protocols were preferred over images
acquired using routine protocols. All CT-based STL models demonstrated non-uniform
geometrical deviations of up to +0.9 mm. The largest deviations were observed in CBCT-
derived STL models. Therefore, it can be concluded that CT imaging technologies and
their acquisition parameters can markedly affect the accuracy of medical AM constructs.
In Chapter 4, the impact of head positioning on CBCT image quality is evaluated.
The impact of supine, prone and oblique patient imaging positions on the image quality,
contrast-to-noise ratio and figure of merit value in the maxillofacial region of a fresh frozen
cadaver head was assessed using a CBCT scanner. In addition, the CBCT supine images
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
144
10
were compared with supine MSCT images. The best CBCT image quality was achieved
in the prone imaging position for sinus, mandible and maxilla, followed by the supine
and oblique imaging positions. The MSCT scanner offered similar image qualities to
the 7.5-mA supine images acquired using the mobile CBCT scanner. The prone imaging
position offered the best CNR and FOM values on the mobile CBCT scanner. These
findings can help to improve CBCT image quality in the maxillofacial region, which, in
turn, may improve the overall accuracy of CBCT-derived AM constructs.
In addition to the CT imaging step, the accuracy of medical AM constructs is affected
by errors introduced during image processing, particularly during the image segmentation
process. Therefore, in Chapter 5, thirty-two publications that reported on the accuracy of
different CT image segmentation methods for bone used in medical AM are reviewed.
The advantages and disadvantages of the different image segmentation methods used
in these studies were evaluated and the reported accuracies were compared. The spread
between the reported accuracies was large (0.04 mm to 1.9 mm). Global thresholding
was the most commonly used segmentation method with accuracies under 0.6 mm.
The disadvantage of this method is the extensive manual post-processing required.
Advanced thresholding methods could improve the accuracy to under 0.38 mm. However,
such methods are currently not included in commercial software packages. Statistical
shape model methods resulted in accuracies from 0.25 mm to 1.9 mm but they are
only suitable for anatomical structures with moderate anatomical variations. To improve
the accuracy and reduce the cost of patient-specific AM constructs, more advanced image
segmentation methods are required.
Chapter 6 provides more insight into the most commonly used image segmentation
method in medical AM, namely global thresholding. The impact of manual and default
threshold selection on the reliability and accuracy of skull STL models acquired using
different CT technologies was evaluated. One female and one male human cadaver head
were imaged using MDCT, DECT and two CBCT scanners. Four medical engineers manually
thresholded the bony structures on all CT images. The lowest and highest selected mean
threshold values and the default threshold value were used to generate skull STL models.
Geometric variations between all manually thresholded STL models were calculated.
Additionally, in order to calculate the accuracy of the manually and default thresholded STL
models, all STL models were superimposed on an optical scan of the dry female and male
skulls (“gold standard”). The intra- and inter-observer variability of the manual threshold
selection was good (intra-class correlation coefficients >0.9). All engineers selected grey
values closer to soft tissue to compensate for bone voids. Geometric variations between
the manually thresholded STL models were 0.13 mm (MDCT), 0.59 mm (DECT) and 0.55
mm (CBCT). All STL models demonstrated inaccuracies ranging from −0.8 mm to +1.1 mm
(MDCT), −0.7 mm to +2.0 mm (DECT) and −2.3 mm to +4.8 mm (CBCT). The findings of
this study demonstrate that manual threshold selection results in better STL models than
default thresholding.
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
145
10
In Chapter 7, a novel method was developed to assess the accuracy and reliability of CT
scanners and software packages. To this end, an anthropomorphic phantom that provided
ground truth measurements for the evaluation of the oropharynx morphology was 3D
printed. This phantom was used to assess the accuracy of two multi-detector row computed
tomography (MDCT) scanners (GE Discovery CT750 HD, Siemens Somatom Sensation)
and three CBCT scanners (NewTom 5G, 3D Accuitomo 170, Vatech PaX Zenith 3D). All
CT images were segmented by two observers and converted into standard tessellation
language (STL) models. The volume and the cross-sectional area of the oropharynx
were measured on the acquired STL models. Finally, all STL models were registered and
compared with the gold standard. Significant differences were observed in the volume
and cross-sectional area measurements of the oropharynx acquired using the different
MDCT and CBCT scanners. The Siemens MDCT and the Vatech CBCT scanners were more
accurate than the GE MDCT, NewTom 5G, and Accuitomo CBCT scanners.
AM medical models, implants and drill guides are becoming increasingly popular
amongst maxillofacial surgeons and dentists. This rise in popularity has subsequently
increased the number of CT and CBCT scans performed worldwide. However, all X-ray
based imaging modalities induce harmful ionizing radiation to the patient. Therefore,
Chapter 8 assesses the feasibility of using radiation-free UTE MRI sequences for medical
AM. Three morphologically different dry human mandibles were scanned using a CT and
MRI scanner. All CT and MRI scans were converted into STL models and geometrically
compared with a corresponding gold standard STL model acquired using an optical 3D
scanner. To quantify the accuracy of the AM process, the CT, MRI and gold-standard STL
models of one of the mandibles were additively manufactured, optically scanned and
compared with the original gold standard STL model. Geometric differences between all
three CT-derived STL models and the gold standard were < 1.0 mm. All three MRI-derived
STL models generally presented deviations < 1.5 mm in the symphyseal and mandibular
area. It should be noted that the AM process only introduced minor deviations of < 0.5
mm. Hence, UTE MRI sequences offer an alternative to CT in generating STL models of
the mandible and could be suitable for surgical planning and AM.
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
147
10
SAMENVATTINGAdditive manufacturing (AM), oftewel driedimensionaal (3D-) printen, maakt het mogelijk
om patiëntspecifieke producten te fabriceren zoals anatomische modellen, chirurgische
boormallen en implantaten. Dergelijke toepassingen verkorten de operatieduur
en kunnen de nauwkeurigheid van de operatie verbeteren. In Hoofdstuk 2 wordt een
nieuwe behandelmethode gepresenteerd waarbij 3D-printen gecombineerd wordt met
virtuele chirurgische planning om nauwkeurige en goedkope patiëntspecifieke implantaten
te fabriceren voor de reconstructie van de bodem van de oogkas (orbitabodem). De eerste
stap van deze nieuwe behandelmethode is om computertomografie (CT-) beelden van
een patiënt met een orbitabodemfractuur te converteren naar een virtueel 3D-model.
Vervolgens wordt het defect in de orbitabodem virtueel gesloten. Op basis van het
resulterende 3D-model wordt vervolgens een mal ontworpen die geprint wordt met
een inkjet 3D-printer. De chirurg kan een dergelijke fysieke mal tijdens de operatie als
hulpmiddel gebruiken om van autoloog bot een patiëntspecifiek orbitabodemimplantaat
optimaal vorm te geven. Deze nieuwe behandelmethode leidt tot een hogere chirurgische
precisie en efficiëntie en kan de operatietijd significant verkorten.
Ondanks de recente ontwikkelingen en veelbelovende medische casuïstiek op het
gebied van 3D-printen zijn er nog vele wetenschappelijke, technologische en reglementaire
uitdagingen te overwinnen. Het 3D-printen van een medisch product vereist drie stappen:
1) medische beeldvorming; 2) beeldverwerking; en 3) 3D-printen. Elk van deze stappen
kan de nauwkeurigheid van medische 3D-geprinte producten beïnvloeden. Echter, omdat
moderne 3D-printtechnieken zeer nauwkeurig zijn (< 0,1 mm), is de nauwkeurigheid met
name afhankelijk van de medische beeldvorming en de daaropvolgende beeldverwerking.
Het doel van deze dissertatie is inzicht te verkrijgen in de huidige wetenschappelijke
en technologische uitdagingen op het gebied van medisch 3D-printen, met name
omtrent medische beeldvorming en beeldverwerking. Ons onderzoek heeft geleid tot
de identificatie van verschillende parameters die de nauwkeurigheid van medische
3D-geprinte producten beïnvloeden.
In Hoofdstuk 3 wordt de invloed van verschillende CT-scanners en scanparameters
onderzocht op de beeldkwaliteit en de nauwkeurigheid van de resulterende
virtuele 3D-modellen. Drie droge menselijke schedels werden gescand met twee
multidetectorcomputertomografie (MDCT) scanners, een dual-energy computertomografie
(DECT) scanner, en een cone-beam computertomografie (CBCT) scanner.
Op elk van deze CT-scanners werden verschillende scanprotocollen getest. Twee
ervaren radiologen beoordeelden de beeldkwaliteit van de resulterende CT-scans. Hierna
werden de scans geconverteerd naar virtuele 3D-modellen en werd de nauwkeurigheid
van deze 3D-modellen bepaald door ze te vergelijken met een “gouden standaard”
(optische 3D-scan). De radiologen namen significante verschillen in beeldkwaliteit waar
tussen de MDCT-, DECT- en CBCT-scanners. Bovendien prefereerden de radiologen het
lage-dosis-scanprotocol boven het standaard scanprotocol op de MDCT-scanner. Alle
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
148
10
virtuele 3D-modellen vertoonden niet-uniforme onnauwkeurigheden tot +0,9 mm, waarvan
de modellen verkregen met CBCT de grootste afwijkingen vertoonden. De belangrijkste
conclusie van Hoofdstuk 3 is dat het type CT-scanner en het gebruikte scanprotocol een
grote invloed kunnen hebben op de nauwkeurigheid van medische 3D-geprinte producten.
In Hoofdstuk 4 wordt de invloed van de hoofdpositie in de CBCT-scanner op
de beeldkwaliteit, contrast-to-noise ratio (CNR) en figure-of-merit (FOM) onderzocht.
Mobiele CBCT-scanners kunnen het hoofd van een patiënt in rugligging, in buikligging, of
zittend met het hoofd in een schuine positie scannen. In elk van deze drie posities werd een
CBCT-scan gemaakt van een vers ingevroren kadaverhoofd. Ook werd er ter vergelijking
een MDCT scan gemaakt van het kadaverhoofd. De beste beeldkwaliteit voor de
sinussen, mandibula en maxilla op de CBCT-scans werd verkregen in buikligging, gevolgd
door de rugligging. De schuine positie gaf de laagste beeldkwaliteit. De beeldkwaliteit
van de CBCT-scan gemaakt van het kadaverhoofd in rugligging met een buisstroom van
7,5 mA was vergelijkbaar met de MDCT-scan. De rugligging resulteerde in de beste CNR
en FOM voor de CBCT-scanner. Deze nieuwe inzichten over de kwaliteit van CBCT-scans
kunnen de algehele nauwkeurigheid van 3D-geprinte producten verkregen uit CBCT-scans
ten goede komen.
Het is niet alleen de medische beeldvorming die de nauwkeurigheid van medische
3D-printen beïnvloedt maar ook de beeldverwerking. De belangrijkste stap van
beeldverwerking is beeldsegmentatie. Hoofdstuk 5 geeft daarom een overzicht van
tweeëndertig wetenschappelijke artikelen die rapporteren over de nauwkeurigheid van
verschillende CT-beeldsegmentatiemethodes voor botstructuren. De voor- en nadelen
van deze methodes worden samengevat en de nauwkeurigheden worden met elkaar
vergeleken in een tabel. Deze tabel laat zien dat de nauwkeurigheden in de literatuur
sterk uiteenlopen: 0,04 mm tot 1,9 mm. De meestgebruikte segmentatiemethode
in medisch 3D-printen is de zogenaamde drempelmethode (“thresholding”). Deze
methode resulteert in nauwkeurigheden onder 0,6 mm. Een belangrijk nadeel
van de drempelmethode is dat men vaak handmatig ruis of artefacten van het virtuele
3D-model moet verwijderen. Geavanceerde drempelmethodes kunnen de nauwkeurigheid
van virtuele 3D-modellen verbeteren tot onder 0,38 mm. Echter, dergelijke methodes
zijn tot op heden niet beschikbaar in de commerciële softwarepakketten voor medisch
3D-printen. Ten slotte resulteren de segmentatiemethodes gebaseerd op statistische
vormmodellen in nauwkeurigheden tussen 0,25 mm en 1,9 mm, al zijn dergelijke methodes
alleen geschikt voor het segmenteren van botstructuren met beperkte anatomische
variaties. Kortom, betere beeldsegmentatiemethodes zouden de nauwkeurigheid en
de kosten van medisch 3D-printen ten goede komen.
Hoofdstuk 6 gaat dieper in op de drempelmethode. Een medisch ingenieur
selecteert meestal handmatig de juiste grijswaarde als drempelwaarde om botstructuren
te segmenteren in CT-beelden, al bevatten de meeste softwarepakketten ook een
automatische drempelwaarde. Daarom wordt in Hoofdstuk 6 de invloed onderzocht
van handmatige en automatische drempelwaardeselectie op de betrouwbaarheid en
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
149
10
de nauwkeurigheid van virtuele 3D modellen verkregen met verschillende
CT-scanners. Een mannelijk en een vrouwelijk gebalsemd kadaverhoofd werden gescand
met een MDCT-, een DECT- en twee CBCT-scanners. Vier medisch ingenieurs selecteerden
vervolgens in iedere CT-scan een drempelwaarde om bot te segmenteren. De laagste
en de hoogste geselecteerde drempelwaarde werden vervolgens gebruikt om virtuele
3D-modellen te genereren. Dit werd ook gedaan met de automatische drempelwaarde.
De geometrische variaties tussen alle handmatig gegenereerde 3D-modellen werden
berekend. Ook werd de nauwkeurigheid van alle 3D-modellen bepaald door deze
te vergelijken met een gouden standaard (een optische 3D-scan van de schedels
waarvan het zachte weefsel verwijderd was). De intra- en inter-observer variabiliteit van
de handmatige drempelwaardeselectie was goed (intra-class correlatiecoëfficiënt > 0,9).
Alle medisch ingenieurs selecteerden drempelwaardes die leken op de grijswaarden
van de zachte, omliggende weefsels om zoveel mogelijk bot te segmenteren. Echter, er
werden geometrische variaties tussen de resulterende 3D-modellen waargenomen van
0,13 mm (MDCT), 0,59 mm (DECT), en 0,55 mm (CBCT). De nauwkeurigheden liepen
uiteen van -0,8 mm tot +1,1 mm (MDCT), −0,7 mm tot +2,0 mm (DECT), en van −2,3
mm tot +4,8 mm (CBCT). Een andere interessante bevinding was dat de handmatig
geselecteerde drempelwaardes resulteerden in betere 3D-modellen dan die verkregen
met de automatische drempelwaarde.
In Hoofdstuk 7 wordt een nieuwe methode gepresenteerd om de betrouwbaarheid en
de nauwkeurigheid van CT-scanners en softwarepakketten te onderzoeken. Deze methode
bestaat uit het 3D-printen van een antropomorfisch fantoom waarvan de afmetingen
bekend zijn en die dus dienst kan doen als gouden standaard. Een dergelijk fantoom is
in Hoofdstuk 7 gebruikt om de betrouwbaarheid en de nauwkeurigheid van verschillende
CT-scanners te bepalen in het afbeelden van de bovenste luchtwegen (oropharynx). Het
fantoom werd gescand met twee MDCT-scanners (GE Discovery CT750 HD, Siemens
Somatom Sensation) en drie CBCT-scanners (NewTom 5G, 3D Accuitomo 170, Vatech
PaX Zenith 3D). De resulterende CT-scans werden door twee tandartsen gesegmenteerd
en vervolgens geconverteerd naar virtuele 3D-modellen. Van alle 3D-modellen werd het
volume en de doorsnede berekend en vergeleken met de gouden standaard. Er werden
significante verschillen in volume en doorsnede waargenomen tussen de verschillende
MDCT- en CBCT- scanners. De Siemens MDCT- en de Vatech-CBCT scanners waren
nauwkeuriger dan de GE MDCT-, NewTom 5G, en Accuitomo CBCT-scanners.
Medisch 3D-printen wint snel in populariteit onder mond-, kaak- en aangezichtschirurgen
en tandartsen. Dit heeft echter ook tot gevolg dat het aantal gemaakte CT- en CBCT-scans
wereldwijd toeneemt. Alle beeldvormende modaliteiten waarbij röntgenstraling gebruikt
wordt, stellen de patiënt bloot aan schadelijke straling. Daarom wordt in Hoofdstuk 8 een
nieuwe, stralingsvrije methode onderzocht om bot in beeld te brengen: een ultrashort
echo time (UTE) sequentie op een magnetic resonance imaging (MRI-) scanner. Drie
droge mandibulae werden gescand met een UTE MRI-sequentie en een CT-scanner.
De resulterende scans werden geconverteerd naar virtuele 3D-modellen en vergeleken
EN
GLISH
SUM
MA
RY & D
UTC
H SU
MM
ARY
150
10
met een gouden standaard (een optische 3D-scan). Ook werd de nauwkeurigheid van het
3D-printproces onderzocht door alle virtuele 3D-modellen te printen op de inkjet printer
van het VU medisch centrum en deze opnieuw te vergelijken met de gouden standaard.
Alle virtuele 3D-modellen verkregen met de CT-scanner resulteerden in nauwkeurigheden
< 1,0 mm. Ter vergelijking, alle virtuele 3D-modellen verkregen met de UTE MRI-sequentie
waren < 1,5 mm nauwkeurig in de symphysis mandibulae en de corpus mandibulae. Een
andere interessante bevinding was dat het 3D printproces slechts minimale afwijkingen
introduceerde (< 0,5 mm). UTE MRI-sequenties kunnen dus een alternatief bieden voor
CT voor het genereren van virtuele 3D-modellen van de mandibula. Dergelijke modellen
kunnen gebruikt worden voor virtuele chirurgische planning en 3D-printen.
11 ACKNOWLEDGEMENTS
AC
KN
OW
LED
GE
ME
NTS
155
11
ACKNOWLEDGEMENTSThe 3D Innovation Lab of the VU University Medical Center (VUmc) constantly explores and
invests in 3D imaging and printing technologies and a wide array of medical virtual planning
software solutions required for the fabrication of individualized medical constructs hence
human spare parts. This unique collaboration and education space formed the basis of
my PhD research. Therefore, I would like to express my gratitude to all people involved in
the 3D Innovation Lab: Niels Liberton, Sjoerd te Slaa, Frank Verver, Micha Paalman, Dafydd
Visscher, dr. Jack van Loon, Angela Ridwan-Pramana, dr. Sophie Kommers, Omar Hertgers,
Ivar Kommers, Rianne van Loenen, Professor Ruud Verdaasdonk and our colleagues from
the Academic Medical Center (AMC), Professor Theo Smit, dr. Geert Streekstra en dr.
Iwan Dobbe. Also thanks to our Bachelor and Master students that contributed to our
research over the past years: Roelof van Dijk, David Spies, Fedde Kuilman, Niki Tokkari,
Daan Brons, Karen Kiers, Guus van Baar, Janne Ploeger, Jordi Minnema. Meghan Simons,
Sevde Yen, Claire Olthoff, and Ezra Esmeijer. In this context, I would like to thank Theo
Faes for coordinating our Master students.
I would like to thank the department of Oral & Maxillofacial Surgery/Pathology and my
department head and promotor Professor Tymour Forouzanfar for facilitating my PhD. Also
thanks to all of my colleagues from the department. I am particularly grateful to Annelies
van der Geest and dr. Martijn van Steenbergen for their consistent support and excellent
organization skills.
Dr. Eric-Jan Rijkhorst and dr. Dennis Heijtel, I would like to thank you for the interesting
physics-related discussions and your contributions to our research. Eliane Kaaij and
Jasmine Rubirayoxall, thank you for your outstanding anatomical skills and passion. And
Professor Paul van der Stelt and dr. Hui Chen, thank you for the productive collaboration.
A special thanks to Planmeca Oy for providing financial support for my PhD project. Timo
Müller, Vesa Varjonen, Jaakko Lähelmä, dr. Kalle Karhu, dr. Ari Hietanen, and dr. Juha
Koivisto, I hope we can continue our excellent collaboration in the future. I would also
like to offer my thanks to DSM Biomedical B.V., and particularly Jac Koenen, for their
financial support. Finally, I would like to thank Anke Kleinveld and her colleagues from
the Institute for Education and Training for allowing me to become a tutor at the VUmc
School of Medical Sciences and thereby supporting my research at the VUmc. It was
an amazing experience for me to guide Bachelor students through their first years of
medical education.
Thanks so much to my family and friends for their unlimited encouragement and support. In
this context, I would like to thank my paranymphs Elise van Eijnatten and Mike van Rijssel.
Floor Maarse, thank you for being such a great swim teacher. Furthermore, I am especially
AC
KN
OW
LED
GE
ME
NTS
156
11
grateful to Daniël Doorman, for always taking the time to listen to me and standing by my
side. A special thanks to Peter Heath for revising the manuscript and to my dad for revising
the Dutch summary. I am also grateful to all of the co-authors of the related publications.
Finally, I am extremely grateful to my mentor and co-promotor Jan Wolff for believing in
me from the beginning of my research career and for his consistent support.