TG-F1056-0E FEB. 2011K TG-F1056-2E OCT. 2013K CDT3000 Series Multi-Digital Calorie Monitor OUTLINE The CDT3000 calorie monitor consists of a pitot tube type flow element with a built-in temperature sensor and a differential pres- sure instrument equipped with root extraction and calorie calculation functions. By inputting an external temperature signal and calculat- ing temperature difference, it indicates and outputs thermal flow and totalized calorie in addition to flow rate. Applicable pipe size ranges from 20 mm to 450 mm. FEATURES Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion of CDT3000 Simultaneous measurement of both flow and temperature saves your installation cost Negligible small pressure loss Selectable from wide variety of flow meter ranges suitable for your applications Applicable for both horizontal and vertical pipes By inputting temperature signal on the fluid feed line it allows thermal flow measurement and totalized calorie calculation to serve as a calorie monitor for your energy conservation Easy to read with a large LCD display STANDARD SPECIFICATIONS Measuring fluid : Water, cold water, hot water Max. pressure : 2 MPa Max. allowable differential pressure : 700 kPa (bias pressure) Not applicable for vacuum service Temperature and relative humidity Fluid temperature : -10 to 80°C *1 (Use the lining or plastics pipe within its allowable temperature) Ambient : -10 to 50°C < 85% RH Storage : -20 to 60°C < 85% RH Without freezing and condensation *1 When the instrument is used for the services exceeding 70°C, do not insulate nozzle and differential pressure instrument as the maximum allowable temperature of sensor is -10 to 70°C. Pipe size : 20 mm to 450 mm, 3/4" to 18" Applicable pipe : Carbon steel pipe for ordinary piping (JIS G 3452) for 1 MPa class Carbon steel pipe for pressure service (JIS G 3454 Sch 40) for 2 MPa class Note : If other pipes than above are used, the calibration is re- quired. Contact us for details. Accuracy of flow measurement : ±5% F.S. as standard (±10% F.S. for a flow rate of less than 30%) Pressure influence : ±0.5% F.S./MPa Flow measuring range : 10 to 100% F.S. Low cutoff : Less than 7% Built-in temperature sensor : RTD (Resistance Temperature Detector) Pt 100, 3 wire, JIS grade A Temperature measuring Accuracy : ±5% F.S.(F.S. =130°C) Range : -10 to 120°C Indication resolution : 0.1°C Measuring period : 1 s Materials of wet parts : See "MATERIAL AND OUTLINE DIMEN- SION" Painting : Melamine resin painting only for indicator housing Painting color for front : Wine red (Munsell 10RP3/8) for rear : Light gray (Munsell N7.5) Housing enclosure : IP65, equivalent to JIS C0920 Protected against powerful jets except for an air introduction port at the bottom of housing. Process connection : R3/4 Male Installation posture : Perpendicular to front face Power supply : 24 V DC±10% Power consumption : 2 VA or less Specifications of indication parts Measuring value indication part 3-1/2 digit LCD : Letter height 18 mm, indication range 0 to 1999 Note : Data outside indication range are expressed as FFF Bar graph indication with 11 segments Filter : Selectable from 0, 2, 4, 8, 16 ,32s Moving average Flow rate indication interval : 0.5s with sampling period 0.25s Contents of indication : Changeable from following (1) to (6) by a key operation (1) Flow rate m 3 /h (2) Thermal flow MJ/h, 10 MJ/h, 100 MJ/h, GJ/h (3) Temperature difference °C (4) Feed line temperature °C (5) Return line temperature °C (6) Differential pressure kPa Totalized value indication part *2 7-1/2 digit LCD : Letter height 5 mm, indication range 0 to 19999999 Counting frequency : 2 Hz or less Totalized value saving function : Saved automatically in EEPROM* every 1 hour or by a key operation * Use an external counter preferably because the rewritable times of EEPROM are 1000000, about 10 years. Use the totalized value on the indicator mainly as a monitor- ing purpose. Reset of totalized values : By a key operation Contents of indication : Changeable from following (1) or (2) by a key operation (1) Totalized volume :m 3 [ 0.001 to 1000] (2) Totalized calorie : MJ [ 0.001 to 1000] LCD back light : Operating 1 minute after a key opera- tion Required straight runs Note : D is inside diameter of pipe Downstream 5D or more 3D or more Pipe size 32 mm or less 40 mm or more Upstream 10D or more 5D or more Mass : Approx. 2 kg

Transcript of CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ......

Page 1: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

TG-F1056-0E FEB. 2011K

TG-F1056-2E OCT. 2013K

CDT3000 SeriesMulti-Digital Calorie Monitor

OUTLINEThe CDT3000 calorie monitor consists of a pitot tube type flowelement with a built-in temperature sensor and a differential pres-sure instrument equipped with root extraction and calorie calculationfunctions. By inputting an external temperature signal and calculat-ing temperature difference, it indicates and outputs thermal flow andtotalized calorie in addition to flow rate. Applicable pipe size rangesfrom 20 mm to 450 mm.

FEATURES Easy installation by welding a 20 mm, 3/4 inch socket on exist

ing piping for insertion of CDT3000 Simultaneous measurement of both flow and temperature saves

your installation cost Negligible small pressure loss Selectable from wide variety of flow meter ranges suitable for

your applications Applicable for both horizontal and vertical pipes By inputting temperature signal on the fluid feed line it allows

thermal flow measurement and totalized calorie calculation toserve as a calorie monitor for your energy conservation

Easy to read with a large LCD display

STANDARD SPECIFICATIONS Measuring fluid : Water, cold water, hot water Max. pressure : 2 MPa Max. allowable differential pressure

: 700 kPa (bias pressure)Not applicable for vacuum service

Temperature and relative humidityFluid temperature : -10 to 80°C *1

(Use the lining or plastics pipe within itsallowable temperature)

Ambient : -10 to 50°C < 85% RH Storage : -20 to 60°C < 85% RH

Without freezing and condensation *1 When the instrument is used for the services exceeding 70°C,

do not insulate nozzle and differential pressure instrument asthe maximum allowable temperature of sensor is -10 to 70°C.

Pipe size : 20 mm to 450 mm, 3/4" to 18" Applicable pipe : Carbon steel pipe for ordinary piping (JIS

G 3452) for 1 MPa classCarbon steel pipe for pressure service(JIS G 3454 Sch 40) for 2 MPa class

Note : If other pipes than above are used, the calibration is re-quired. Contact us for details.

Accuracy of flow measurement : ±5% F.S. as standard (±10%F.S. for a flow rate of less than30%)

Pressure influence : ±0.5% F.S./MPa Flow measuring range : 10 to 100% F.S. Low cutoff : Less than 7% Built-in temperature sensor : RTD (Resistance Temperature

Detector) Pt 100Ω, 3 wire, JISgrade A

Temperature measuringAccuracy : ±5% F.S.(F.S. =130°C)Range : -10 to 120°CIndication resolution : 0.1°CMeasuring period : 1 s

Materials of wet parts : See "MATERIAL AND OUTLINE DIMEN-SION"

Painting : Melamine resin painting only for indicatorhousing

Painting color for front : Wine red (Munsell 10RP3/8)for rear : Light gray (Munsell N7.5)

Housing enclosure : IP65, equivalent to JIS C0920 Protectedagainst powerful jets except for an airintroduction port at the bottom of housing.

Process connection : R3/4 Male

Installation posture : Perpendicular to front face Power supply : 24 V DC±10% Power consumption : 2 VA or less Specifications of indication parts

Measuring value indication part• 3-1/2 digit LCD : Letter height 18 mm, indication range 0 to

1999Note : Data outside indication range are expressed as FFF

• Bar graph indication with 11 segments• Filter : Selectable from 0, 2, 4, 8, 16 ,32s Moving

average• Flow rate indication interval : 0.5s with sampling period 0.25s• Contents of indication : Changeable from following

(1) to (6) by a key operation(1) Flow rate m3/h(2) Thermal flow MJ/h, 10 MJ/h, 100 MJ/h, GJ/h(3) Temperature difference °C(4) Feed line temperature °C(5) Return line temperature °C(6) Differential pressure kPa

Totalized value indication part *2• 7-1/2 digit LCD : Letter height 5 mm, indication range 0 to

19999999• Counting frequency : 2 Hz or less• Totalized value saving function : Saved automatically in

EEPROM* every 1 hour or by a key operation* Use an external counter preferably because the rewritable

times of EEPROM are 1000000, about 10 years.Use the totalized value on the indicator mainly as a monitor-ing purpose.

• Reset of totalized values : By a key operation• Contents of indication : Changeable from following (1) or (2)

by a key operation(1) Totalized volume : m3 [ 0.001 to 1000](2) Totalized calorie : MJ [ 0.001 to 1000]

LCD back light : Operating 1 minute after a key opera-tion

Required straight runs

Note : D is inside diameter of pipe

Downstream

5D or more

3D or more

Pipe size

32 mm or less

40 mm or more

Upstream

10D or more

5D or more

Mass : Approx. 2 kg

Page 2: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E2 TOKYO KEISO CO., LTD.

Note 1: The inside diameter of pipe size 20 to 32 mm in 1 MPaclass is based on three-way socket JIS B 2301.Please specify the three-way sockets by using optionalcodes if requested.

Standard full scale range (maximum flow rate)when using 1 MPa class, 40 to 450 mm SGP pipe

Note 2: The inside diameter of pipe size 40 to 450 mm in 1 MPaclass is based on Carbon steel pipe for ordinary piping SGPJIS G 3452.

Standard full scale range (maximum flow rate)when using 2 MPa class, 20 to 32 mm three-way socket

Note 3: The inside diameter of pipe size 20 to 32 mm in 2 MPaclass is based on three-way socket with sch40.The three-way sockets are supplied by customers.

Standard full scale range (maximum flow rate)when using 2 MPa class, 40 to 450 mm STPG pipe

Note 4: The inside diameter of pipe size 40 to 450 mm in 2 MPaclass is based on Carbon steel pipe for pressure serviceSTPG (JIS G 3454 Sch 40).

Standard full scale range (maximum flow rate)when using 1 MPa class, 32 to 150 mm PVC lining pipe

Specifications of outputs Current outputs : Selectable from “Not required” or “The

followings”• Contents : Select from “one of (1) to (6) in Measuring

value indication”• Signal : 4 to 20 mA DC• Allowable load resistance : 500 Ω• Accuracy of output : ±0.5% F.S. at 23°C Accuracy of

indication itself• Temperature influence : ±0.03% F.S./ °C for both zero and

span• Response : 1s or less when filter is set as 0

Pulse outputs*2 : Selectable from “Not required” or “Thefollowings”

• Contents : Select from “ (1) or (2) in Totalized valueindication”

• Output type : Photo MOS relay®• Load rating : 30 VDC, 100 mA Max.• Pulse frequency : 2 Hz or less, Synchronized with totalized

value indication• Pulse width : ON 200 to 300 ms

Alarm outputs : Alarm output becomes ON when one offollowings occurs. You can also set to “Notrequired.”

(1) Flow rate indication exceeds 1999.(2) Thermal flow indication exceeds 1999.(3) Temperature difference exceeds 120% of its setting value.(4) Feed line temperature is less than -10°C or more than 120°C.(5) Return line temperature is less than -10°C or more than 120°C.(6) Differential pressure is less than -10% F.S. or more than 110%

F.S.• Output type : Photo MOS relay®• Load rating : 30 VDC, 100 mA Max.• Response : 1s or less to the indication

*2 Select a proper flow range so that normal flow rate becomesbetween 50% and 100% of the full scale range especially at thetotalized flow and pulse output-emphasized measurement. Theaccuracy of flow rate measurement is ±5% of full scale. Accord-ingly in the totalized flow measurement, if the normal flow rate israther low compared to flow range, the accumulated error re-sults in higher.

Standard full scale range (maximum flow rate)when using 1 MPa class, 20 to 32 mm three-way socket

Note 5: Above full scale ranges are based on the flow rate when theflow runs through the inside diameter of the pipe lined withPVC. The pipe consists of a tee joint of the Eslo coat LXmade by Sekisui Chemical Co.Åfs Eslon(R). Its maximumoperating pressure is 1 MPa. If you use the tee joints withthe different inside diameter from above, the extra calibra-tion is required.If your size is different from the one in the standard dimen-sions of D and L of PVC lining pipe described on page 7,consult us.

Pipe size Full scale flow rate m3/h Note 1Inside diameter

mmmm inch Code

20

25

32

3/4

1

1-1/4

05

3

4.5

8.5

10

4

6.5

12

20

6

9

17

(26)

(34)

(43)

Pipe size Full scale flow rate m3/h Note 2Inside diameter

mmmm inch Code

40

50

65

80

100

125

150

200

250

300

350

400

450

1-1/2

2

2-1/2

3

4

5

6

8

10

12

14

16

18

05

9

15

25

35

60

90

130

240

350

500

650

850

1100

10

13

20

35

50

85

130

190

340

500

750

900

1200

1500

20

18

30

50

75

120

190

270

480

720

1000

1300

1700

41.6

52.9

67.9

80.7

105.3

130.8

155.2

204.7

254.2

304.7

339.8

390.6

441.4

Pipe size Full scale flow rate m3/h Note 3Inside

diameter mm

mm inch Code

20

25

32

3/4

1

1-1/4

10

3

4.5

9

20

4

6.5

13

50

6

9

16

21.4

27.2

35.5

Pipe size Full scale flow rate m3/h Note 4Inside

diameter mm

mm inch Code

40

50

65

80

100

125

150

200

250

300

350

400

450

1-1/2

2

2-1/2

3

4

5

6

8

10

12

14

16

18

05

9

15

24

35

55

90

130

230

350

500

600

800

1000

10

12.5

20

34

48

80

130

180

320

500

700

900

1150

1500

20

18

30

48

70

110

180

260

460

700

1000

1250

1660

41.2

52.7

65.9

78.1

102.3

126.6

151.0

199.9

248.8

297.9

333.4

381.0

428.6

Pipe size Full scale flow rate m3/h Note 5Inside

diameter mm

mm inch Code

32

40

50

65

80

100

125

150

1-1/4

1-1/2

2

2-1/2

3

4

5

6

05

5

6.5

11

20

30

50

85

120

10

7

9

16

28

40

75

120

170

20

10

13

22

40

60

100

170

240

29.5

34.7

46.2

59.7

70.9

95.2

119.7

142

Refer to the " STANDARD SETTING LIST(1), (2) " on pages 10 and 11 for the details of Standard full scale range (maxi-mum flow rate), Maximum thermal flow and Multiplier.

Page 3: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E TOKYO KEISO CO., LTD. 3

MATERIAL AND OUTLINE DIMENSION

Feed line temperature sensor inputPower supply and signal output

CDT32

CDT34

(23)

72(1

5.5)

21 24

93

(0.5)93

(184)

(84) (100) l

R3/4

35

1 3 8 3 2 4 5 6 7

3

(41) (47)

(55)

(199

)

(127

)72

(105.5)

(84)

(66.5)

R3/4

l

1

8

3 3

2

4

5

6

7

1

2

3

4

5

6

7

8

No. Part name MaterialIndicatorDiaphragmBodyO-ringDrain hole sealIsolation valveValve shaftShaft O-ringVent valveCap nutFittingsO-ringPitot tubeTemperature sensor

Lead wire

Aluminum alloySUS 316LSUS 316Fluorocarbon rubberAlumina ceramicsSCS 14SUS 316NBRSUS 316SCS 13SCS 14NBRSUS 316SUS 316

PTFE insulated and fluorocarbon rubber sheathed

Pipesizemm

Standard "I" length mm*3 *4

*3 For SGP and STPG pipes*4 For PVC lining pipe

20 25 32 40 50 65 80 100 125 150 200 250 300 350 400 450

30 32 42 63 69 79 88 102 117 133 98 113 128 135 151 166

65 70 76 88 98 116 131 148

Note : The above length "I" of pitot tube is our standard size. When your sizes of pipe and socket are different from ours, check your sizes with "L" and "D" in the drawing on page 7 and consult us if necessary.

Page 4: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E4 TOKYO KEISO CO., LTD.

INDICATION PART

INTERCONNECTION

Measuring value indication part

Bar graph meter indication

Indicated in proportion to set value

Operation monitor

SET

ADJ.

H/C

Blinking during set mode

Remains ON while zero adjustment is enabled

Cooling or heating monitorDuring heating : ONDuring cooling : OFF

MODE key DOWN key UP key ENTER key

Totalized value indication part

Measuring unit monitorAlarm output monitor

Pulse output monitor

m3/h

MJ/h

°C ('T)

°C (TS)

°C (TR)

m3

MJ

Flow rate

Thermal flow

Temperature difference

Feed line temperature

Return line temperature

Totalized volume

Totalized calorie

The measured values of thermal flow, totalized volume and calorie are indicated in the following units and multipliers as specified.

Thermal flowTotal volumeTotal calorie

Flow rate, temperature difference, feed and return line temperature are indicated in fixed units.

0.001 m3

0.001 MJ0.01 m3

0.01 MJ0.1 m3

0.1 MJ

MJ/hm3

MJ

10 MJ/h10 m3

10 MJ

100 MJ/h100 m3

100 MJ

GJ/h1000 m3

1000 MJ

Applicable multi-core cable with shield

Outside diameter : ø4 to 8 mmCore wire cross section : 0.3 to 0.5 mm2 stranded

Terminalblock

Power supply 124 V DC ±10%

Receivinginstrument

Power supply 230 V DC, 100 mA or less

Load

Load

Shielded wire Power supply 330 V DC, 100 mA or less

Feed line RTDPt 100 Ω, 3 wire, JIS grade A

Return line RTDBuilt-in and pre-wired in Pitot tube before shipment

Note : RTD stands for resistance temperature detector Since the pulse and alarm are output by photo MOS relay®, the external power sources are acceptable as shown above power sources 1, 2, 3.

+ - + - + -

a1

COM

a2

A

B

b

A

B

b

POWER24 V DC

ANA. OUT4 to 20 m A DC

NC

OUT1PulseOUT2Alarm

TS

Pt100

TR

Pt100

+ -

+ -

+ -

+ -

Page 5: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E TOKYO KEISO CO., LTD. 5

*5 Insert “/ ” between each optional code if more than one options are involved.You are requested to see the option item list on the next chapter to check if your requirements are available or not.

Coding example 1 : No option is selected.With horizontal installation, Pipe size 50 mm, 1 MPa class and flow range code “10”“CDT3210-050-10AL-A”

Coding example 2 : All possible options are selected.With vertical installation, Pipe size 100 mm, 1 MPa class, flow range code “05”, A welding socket for flowmeter installationattached, Degreasing required, A feed line temperature sensor attached, and A welding socket for temperature sensor wellattached.“CDT3410-100-05AL-A/CFS/DEGPT3/PTS”

Coding example 3 : Piping specifications are different from the ones in above model code.With horizontal installation, Pipe size 80 mm, Pipe material SUS 304TPD, 1MPa class, flow range code “20”, A feed linetemperature sensor attached, and A welding socket for temperature sensor well attached.“CDT3210-080-20AL-A/PT2/PTS/Z”

When piping specifications for example, are different from the ones in model code as shown in the Coding example 3, your requirements fallinto the additional functions. Likewise if the sockets or three-way sockets provided by customers require different length of Pitot tube from ourstandard, it also falls into the additional functions.Accordingly, the last letter of model code other than the contents in above will be “/Z”.

MODEL CODE Please specify also INITIAL PAREMETER SETTING CODE on page 9

Blank

/Z

Carbon steel pipe for pressure service STPG (JIS G 3454 Sch 40)

Description RemarksInstall-ationtype

Horizontal type

Vertical type

1 MPa class

2 MPa class

20 mm

25 mm

32 mm

40 mm

50 mm

65 mm

80 mm

100 mm

125 mm

150 mm

200 mm

250 mm

300 mm

350 mm

400 mm

450 mm

Standard (See TYPICAL INSTALLATION METHODS on page 7)

Carbon steel pipe for ordinary piping SGP JIS G 3452Pressure rating

Pipe size Nominal diameter for installation pipe

Full scale range Flow rate range Select a code from the table "STANDARD FULL SCALE RANGE."

Indicator type

Application

Version

Fixed code

For liquid

Version symbol

2

4

CDT3 0

10

20

-

-020

-025

-032

-040

-050

-065

-080

-100

-125

-150

-200

-250

-300

-350

-400

-450

-

-05

-10

-20

-50

A

A

L

L

-A

-A

/

Option *5

Additional functions

/CFT

/CFS

/DEG

/PL

/PT1

/PT2

/PT3

/PTS

Three-way socket for flow-meter installation (CFTW)Welding socket for flow-meter installation (CFSW)

Degreasing (Standard)

PVC lining pipe

Feed line temperature sensor for 20 to 40 mmFeed line temperature sensor for 50 to 80 mmFeed line temperature sensor for 100 to 450 mmWelding socket fortemperature sensor well

Not provided

Provided

Applicable for 1 MPa class, 20 to 32 mm measuring pipe size. Only the material of FCMB is available.Nominal size 3/4 inch. Only the material of SS 400 is available.

Material of O-ring : NBR The procedure and grade of degreasing comply with Tokyo Keiso's standard. Applicable for all measuring pipe sizes.

Inform us L, D dimensions in the drawing on page 7.

RTD Pt100 Ω, 3 wire, JIS grade A with a protecting tees.

RTD Pt100 Ω, 3 wire, JIS grade A with a protection well of its connection size R1/2 inchRTD Pt100 Ω, 3 wire, JIS grade A with a protection well of its connection size R1/2 inchNominal size 1/2 inch. Only the material of SS 400 is available. Applicable for nominal sizes 50 to 450 mm of measuring pipe.No need to specify if your requirements are described by using above coding.Other requirements including specifications of pipe and the length of pitot tube other than above.

Page 6: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E6 TOKYO KEISO CO., LTD.

OPTION

See following option item list to check if your requirements are available or not.

Option item list for SGP and STPG pipes

Optional fittings for flowmeter installation

Optional feed line temperature sensor and its connection fittings

20

25

32

40

50

65

80

100

125

150

200

250

300

350

400

450

Thr

ee-w

ay

sock

et

Pipe sizemm

Standard installationmethod of flowmeter

Option

For installation of flowmeter

A feed line temperature sensor and its connection fittings

SG

P, S

TP

G p

ipes

3/43/43/4

113/4

1-1/41-1/43/4

3/4 inch socket

3/4 inch socket

3/4 inch socket

CFT (Size 20 mm)

CFT (Size 25 mm)

CFT (Size 32 mm)

CFS (3/4 inch Socket)

CFS (3/4 inch Socket)

CFS (3/4 inch Socket)

PT1 (Protection tees with 20 mm in size)

PT1 (Protection tees with 25 mm in size)

PT1 (Protection tees with 32 mm in size)

PT1 (Protection tees with 40 mm in size)

PT2 (Protection well connection size R1/2)

PTS (1/2 inch socket for protection well connection)PT3 (Protection well

connection size R1/2)

Remarks

The three-way sockets (CFT) for flow-meter installation are available only for 1 MPa class and material of FCMB.The three-way sockets (CFT) for 2 MPa class and other materials than FCMB are supplied by customers.

The optional protection tees (PT1) are available only for 1 MPa class and ma-terial of FCMB.Consult us for the protection tees for 2 MPa class and other materials than FCMB.The sockets for flowmeter installation (CFS) and sockets for protection well connection (PTS) are available only for the material SS400. Those materi-als other than SS400 are supplied by customers.

Note : Consult Tokyo Keiso Co., Ltd. for such specifications as PVC lining pipe, piping specifications designated by customers, feed line temperature sensor (resistance temperature detector, RTD) and its protection well.

Three-way socket for flowmeter installation (Model : CFTW)Option code : CFT

Sizemm

Dimensions of three-way socket

Applicable pipe size : 20 to 32 mmApplicable piping class : 1 MPaApplicable material : FCMB

Welding socket for flowmeter installation (Model : CFSW) Option code : CFS

Applicable pipe size : 40 to 450 mmApplicable piping class : 1 MPa, 2 MPaApplicable material : SS400

Rc3/4

B 2-d

AA

20

25

32

d A B

Rc3/4

Rc1

Rc1-1/4

33

35

36

33

36

4139

Rp3/4

Ø34

.5

Feed line temperature sensor and its protection tees (Option code : PT1 for 20 to 40 mm)

Extension wirePTFE insulated and fluorocarbon resin sheathed

Heat shrinkable tubeSilicone rubber

SheathSUS304

G1/4

Resistance temperature detector (RTD)

Temperature sensor : RTD Pt 100Ω, 3 wire, JIS grade A

Applicable pipe size : 20 to 40 mmApplicable piping class : 1 MPaApplicable material of protection tees : FCMB

Protection wellSUS304

G1/4

Protection teesFCMB

Dimensions of protection teesSizemm

5 30 20 10

2000 (150)

(59)

ø6.

4

A

2-d

(B)

A

20

25

32

40

d A B

Rc3/4

Rc1

Rc1-1/4

Rc1-1/2

28

30

33

34

50

51

53

55

Page 7: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E TOKYO KEISO CO., LTD. 7

Feed line temperature sensor and its protection well (Option code : PT2 for 50 to 80 mm, PT3 for 100 to 450 mm)

Extension wirePTFE insulated and fluoro-carbon resin sheathed

Heat shrinkabletubeSilicone rubber

SheathSUS304

Resistancetemperaturedetector(RTD)

Temperature sensor : RTD Pt 100 Ω, 3 wire, JIS grade A

Applicable pipe size : 50 to 450 mmApplicable piping class : 1 MPa, 2 MPaApplicable material : SS 400

Tube fitting Ø6.4SUS316

ProtectionwellSUS304

Welding socket for above protection well(Option code : PTS for 50 to 450 mm)

5 30 20

2000

10PT2: (150) PT3: (170)

ø6.

4

HE

X 2

2.2

R1/2

ø7

ø10

12 19

3PT2: (75) PT3: (100)

PT2: (106) PT3: (131)

Rp1/2

37

ø28OPTION PVC LINING

Option code : PL for 32 to 150 mm

The full scale ranges are based on the flow rate when the flowruns through the inside diameter of the pipe lined with PVC. Thepipe consists of a tee joint of the Eslo coat LX made by SekisuiChemical Co.’s Eslon®.If you use the tee joints with the different inside diameter fromabove, the extra calibration is required.For further consultation inform us of the dimension “D” in thefollowing drawing.

When the main pipe size is between 32 mm and 150 mm, provideeither one of the following tee joint and bushing assembly tomake the connection size 20 mm:— tee joint with 25 mm in branch line + A bushing with 25 mm x20 mm— tee joint with larger than 25 mm in branch line + One or morebushings in series to make the final connection size 20 mmFor further consultation inform us of the dimension “L” in thefollowing drawing.

Without these dimensions Tokyo Keiso Co., Ltd makesthem comply with our standard practice as shown in the followingtable. The full scale flow rate will be set as the values in the tableon page 2.

Consult Tokyo Keiso Co., Ltd. for the additional equipment suchas the feed line temperature sensor (RTD) and protection well forits installation or else.

STANDARD ACCESSORY Protection cover for indicator

Avoid installing this instrument at the places exposed to directsunlight or rain as much as possible. If it is inevitable, providewith a protection cover above indicator.

OTHER PIPES Other types of piping configuration are also available. Feel free to contact Tokyo Keiso Co., Ltd. Inform us your piping

specifications and following L and D dimensions.

TYPICAL INSTALLATION METHODS Install the flowmeter on the three-way socket for the pipe sizes

from 20 to 32 mm as shown in the following drawing.Weld a socket on the straight pipe and install the flowmeter onthe socket for the pipe sizes 40 mm or more.You can install the flowmeter either on horizontal or vertical pipefor model CDT32.

Rc3/4

øD

L

Bushing(R1Rc 3/4)

to 32mm

Pipesizemm

Dimensions based on ourstandard practices

Applicable pipe size : 32 to 150 mmApplicable piping class : 1 MP

øD

L

D L

32

40

50

65

80

100

125

150

29.5

34.7

46.2

59.7

70.9

95.2

119.7

142

62.5

65.5

71.5

80.5

88.5

103.5

117.5

130.5

Protection cover made of polypropylene

Pipe size CDT32For horizontal and vertical piping

CDT34For horizontal piping

Three-waysocket

Straight pipe

Socket

20 to 32mm

40 to 450mm

Page 8: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E8 TOKYO KEISO CO., LTD.

FLOW MEASUREMENT CONFIGURATION

Install the CDT3000 on return line. The CDT3000 installed on feed line gives a calculation error and results in wrong measurement.

PARAMETER SETTING AND ADJUSTMENT

The key operation on the front panel enables the setting changes.• Mode change : Selectable from “Automatic change-over cooling or heating”, “Cooling fixed mode”, “Heating fixed mode”• Low cut-off of flow rate : Can be set at any value between 0 to 100% F.S. (Standard setting : 7%)• Indication filter : Selectable from 0, 2, 4, 8, 16, 32 seconds in moving average. Applicable for flow rate/thermal flow, feed

and return temperature.(Standard setting : 4 seconds for flow rate/thermal flow, 4 seconds for feed and return temperature)

• Maximum temperature difference : Can be set at any value between 0.5 to 80.0°C. (Standard setting : 20°C)• Zero adjustment : Available for flow rate/differential pressure, temperature difference.

Current output : Flow ratePulse output : CalorieAlarm output : Measuring error

Flow rate measurement : QReturn line measured temperature : TR

Power supply : 24 V DCFeed line temperature

Heat source

Feed line

Return line

Heat duty

Feed line temperature sensorPt 100 Ω, 3 wire, JIS grade AFeed line measured temperature : TS

Automatic change-over from heating to cooling mode and vice versa

From heating to cooling mode

Conversion coefficient : Calorie is calculated and indicated as k= 4.186 MJ/°Cm3.

Conversion coefficient : Calorie is calculated and indicated as k= 4.123 MJ/°Cm3.

Conversion coefficient : Calorie is calculated and indicated as k= 4.186 MJ/°Cm3.

Conversion coefficient : Calorie is calculated and indicated as k= 4.123 MJ/°Cm3.

Feed line temperature °C*6

From cooling to heating mode

10 20 25 30 40

The thermal flow is determined by the following equation.Thermal flow F = Q TS—TR k = Q T k [MJ/h]

Flow rate : Q [m3/h] : Measured by CDT3000Feed line temperature : TS [°C] : Measured by a RTD supplied by customersReturn line temperature : TR [°C] : Measured by CDT3000Calorie conversion coefficient : k [MJ/°Cm3] : At cooling mode 4.186/ At heating mode 4.123

Note : When T 0.5°C, the thermal flow automatically becomes 0 MJ/h due to the low cut-off function of temperature difference.

*6 Operation at 25°C or lower is cooling mode, and 30°C or higher is heating mode according to our standard practice. The different specifiedvalues override these values. At start up just after the installation of CDT3000 when feed line temperature is between 25 to 30°C, the thermal flow is calculated and indicated as k= 4.186 MJ/°Cm3.Even at this stage, when T0.5°C, the thermal flow becomes 0 MJ/h due to the low cut-off function.

Page 9: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E TOKYO KEISO CO., LTD. 9

INITIAL PARAMETER SETTING CODE (Please specify also MODEL CODE on page 5)

*7 Setting of Temperature at cooling and heating mode, Current output required or not, Pulse output required or not, No alarm output designation, Back light required or not can not be set by the indication key operation. However, if current output or pulse output is added, the output items can be also changed by indication key operation.*8 Align the flowmeter in the same flow direction as indicated. The change of flow direction after start-up requires the stoppage of liquid flow.*9 Designate the values of full scale range and maximum thermal flow if other than standard ranges are required , referring to the "INDICATION SETTING LIST" while selecting code "Z".*10 If you have any other special setting requirements other than above, feel free to contact us

Parameters in shaded area are set initially as standard.

Description Setting range

Full scale range

Maximum thermal flow

Thermal flow multipliers

Standard

Designate values *9

Standard

Designate the values *9

Standard

MJ/h

10 MJ/h

100 MJ/h

GJ/h

Standard

Designate it between the values indicated on the right

Standard

Designate it between the values indicated on the right

Standard

Designate it between the values indicated on the right

Not required

Flow rate (Standard)

Thermal flow

Temperature difference

Feed line temperature

Return line temperature

Differential pressure

Not required

Totalized volume

Totalized calorie (Standard)

Bottom to top

Left to right (Standard)

Right to left

Top to bottom

Version symbol

Not required

Fixed operation mode

Low cut-off value

Differential pressure filter

Feed line temperature filter

Return line temperature filter

Back light *7

OUT 2 (No alarm output) *7

Others *10

See "STANDARD SETTING LIST(1), (2)."

See "INDICATION SETTING LIST."

See "STANDARD SETTING LIST(1), (2)."

See "INDICATION SETTING LIST."

See "STANDARD SETTING LIST(1), (2)."

See "INDICATION SETTING LIST."

Setting of temperature at cooling mode *7

25°C15 to 24°C only integer

30°C31 to 40°C only integer

20°C21 to 60°C only integer

Outputted to indication range 0 to 100% F.S.

Outputted to indication range 0 to 100% F.S.

Outputted to maximum set temperature difference

Outputted to -10 to 120°COutputted to -10 to 120°COutputted to differential pressure range

100 c/h or more and less than 1000 c/h

100 c/h or more and less than 1000 c/h

Cooling or heating mode

Between 0 to 50% only integer

0s, 2s, 4s (Standard), 8s, 16s, 32s

0s, 2s, 4s (Standard), 8s, 16s, 32s

0s, 2s, 4s (Standard), 8s, 16s, 32s

Invalid

Designate if any

Blank

Setting of temperature at heating mode *7

Setting of maximum temperature difference

Current output *7

OUT 1 Output (pulse output) *7

Flow direction *8

Version

Other special setting

CDT3 - - -A

1

Z1

Z

1

2

3

4

5

-25

-30

20

-0

-1

-2

-3

-4

-5

-6

0

1

2

1

6

7

8-A

/Z

Page 10: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E10 TOKYO KEISO CO., LTD.

STANDARD SETTING LIST (1)

INDICATION SETTING LIST

0.2 to 1.999 m3/h

2 to 19.99 m3/h

20 to 199.9 m3/h

200 to 1999 m3/h

0.200 to 1.999

2.00 to 19.99

20.0 to 199.9

200 to 1999

Full scale range(Maximum flow rate)

Indication rangeand decimal place

Indicationunit

only m3/h

Note : You can not set full scale range of 2000 m3/h or more.

0.2 to 1.999 MJ/h

2 to 19.99 MJ/h

20 to 199.9 MJ/h

200 to 1999 MJ/h

2000 to 19990 MJ/h

20000 to 199900 MJ/h

200000 to 1999000 MJ/h

0.200 to 1.999

2.00 to 19.99

20.0 to 199.9

200 to 1999

200 to 1999

200 to 1999

200 to 1999

Maximum thermal flow range

Indication range and decimal place

Indication unit and multipliers

Thermal flow multiplier code

Note : You can not set maximum thermal flow range of 2000 GJ/h or more.

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

100 MJ/h

GJ/h

2

3

4 5

Standard full scale range (maximum flow rate) and Maximum thermal flow indication setting list when using 1 MPa class, 20 to 32 mm three-way socket and 40 to 450 mm SGP pipe

1 MPa classFlow rate range code : 05

Pipe sizemm

Pipe inside dia-meter

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Flow rate range code : 10Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Flow rate range code : 20

20

25

32

40

50

65

80

100

125

150

200

250

300

350

400

450

26

34

43

41.6

52.9

67.9

80.7

105.3

130.8

155.2

204.7

254.2

304.7

339.8

390.6

441.4

3.00

4.50

8.50

9.00

15.00

25.0

35.0

60.0

90.0

130.0

240

350

500

650

850

1100

m3/hfixed

m3/hfixed

m3/hfixed

260

380

720

760

1300

210

300

510

760

1100

210

300

420

550

720

930

MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

4.00

6.50

12.00

13.00

20.0

35.0

50.0

85.0

130.0

190.0

340

500

750

900

1200

1500

340

550

1100

1100

1700

300

420

720

1100

1600

290

420

630

760

1100

1300

6.00

9.00

17.00

18.00

30.0

50.0

75.0

120.0

190.0

270

480

720

1000

1300

1700

510

760

1500

1600

260

420

630

1100

1600

230

410

610

840

1100

1500

CDT310--05AL-A CDT310--10AL-A CDT310--20AL-A

MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

Standard full scale range (maximum flow rate) and Maximum thermal flow indication setting list when using 2 MPa class, 20 to 32 mm three-way socket and 40 to 450 mm STPG pipe

2 MPa class20 to 32 mm Flow rate range code : 1040 to 450 mm Flow rate range code : 05

Pipe sizemm

Pipe inside dia-meter

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

20 to 32 mm Flow rate range code : 2040 to 450 mm Flow rate range code : 10

20 to 32 mm Flow rate range code : 5040 to 450 mm Flow rate range code : 20

20

25

32

40

50

65

80

100

125

150

200

250

300

350

400

450

21.4

27.2

35.5

41.2

52.7

65.9

78.1

102.3

126.6

151

199.9

248.8

297.9

333.4

381

428.6

3.00

4.50

9.00

9.00

15.00

24.0

35.0

55.0

90.0

130.0

230

350

500

600

800

1000

m3/hfixed

m3/hfixed

m3/hfixed

260

380

760

760

1300

210

300

470

760

1100

200

300

420

510

670

840

MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

4.00

6.50

13.00

12.50

20.0

34.0

48.0

80.0

130.0

180.0

320

500

700

900

1150

1500

340

550

1100

1100

1700

290

410

670

1100

1600

270

420

590

760

1000

1300

6.00

9.00

16.00

18.00

30.0

48.0

70.0

110.0

180.0

260

460

700

1000

1250

1650

510

760

1400

1600

260

410

590

1000

1600

220

390

590

840

1100

1400

CDT320--10AL-ACDT320--05AL-A

CDT320--20AL-ACDT320--10AL-A

CDT320--50AL-ACDT320--20AL-A

MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

100 MJ/h

Page 11: CDT3000 Series - TOKYO KEISO · CDT3000 Series Multi-Digital ... Applicable pipe size ranges ... Easy installation by welding a 20 mm, 3/4 inch socket on exist ing piping for insertion

CDT3000 series Calorie Monitor

TG-F1056-2E TOKYO KEISO CO., LTD. 11

Head Office : Shiba Toho Building, 1 – 7 – 24 Shibakoen, Minato-ku, Tokyo 105 – 8558

Tel : +81-3 – 3431 – 1625 (KEY) ; Fax : +81-3 – 3433 – 4922

e-mail : [email protected] ; URL : http://www.tokyokeiso.co.jp

* Specification is subject to change without notice.

STANDARD SETTING LIST (2)

Standard full scale range (maximum flow rate) and Maximum thermal flow indication setting list when using 1 MPa class, 32 to 150 mm lining pipe

Standard multipliers for totalization list

PVC liningpipe Flow rate range code : 05

Pipe sizemm

Pipe inside dia-meter

Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Flow rate range code : 10Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Flow rate range code : 20Full scale range (maximum flow rate)

Maximumthermal flow

Indication value

Indicationunit

Indication value

Indicationunit

Full scale range(maximum flow rate) m3/h Multipliers for totalized flow

Maximum thermal flow range MJ/h

Multipliers for total-ized flow Maximum count number range

1000c/h to 1999c/h(Not recommended)

1 to 9.99

10 to 99.9

100 to 999

1000 to 1999

100 to 999

1000 to 9990

10000 to 99900

100000 to 999000

1000000 to 1999000

100c/h to 999c/h

0.01 m3

0.1 m3

1 m3 (Indicated as m3)

10 m3

1 MJ (Indicated as MJ)

10 MJ

100 MJ

1000 MJ

1000 MJ

32

40

50

65

80

100

125

150

29.5

34.7

46.2

59.7

70.9

95.2

119.7

142

5.00

6.50

11.00

20.0

30.0

50.0

85.0

120.0

m3/hfixed

m3/hfixed

m3/hfixed

420

550

930

1700

260

420

720

1100

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

MJ/h

MJ/h

MJ/h

MJ/h

10 MJ/h

10 MJ/h

10 MJ/h

100 MJ/h

7.00

9.00

16.00

28.0

40.0

75.0

120.0

170.0

590

760

1400

240

340

630

1100

1500

10.00

13.00

22.0

40.0

60.0

100.0

170.0

240

840

1100

1900

340

510

840

1500

210

CDT310--05AL-A/PL CDT310--10AL-A/PL CDT310--20AL-A/PL