STLCC-CPLS;Morrison 12/18/2015 Page 1 Microtome SOP Prepared by: Bob Morrison STLCC , Instrumentation Specialist May 2008, updated Dec 2015 for models at BRDG Historage-Elec Controls Mechanical Controls Leica @ BRDG American Optical @BRDG Caution: Never attempt to hold or grab any component of the machine to stop motion

Transcript of Caution: Never attempt to hold or grab any component of the...

Page 1: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015 Page 1

Microtome SOP

Prepared by: Bob Morrison

STLCC , Instrumentation Specialist

May 2008, updated Dec 2015 for models at BRDG

Historage-Elec Controls

Mechanical Controls

Leica @ BRDG

American Optical @BRDG

Caution: Never attempt to hold or grab

any component of the machine to stop

motion

Page 2: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: American Optical, Model 820, SN 70298

STLCC-CPLS;Morrison 12/18/2015 Page 2

Hotlink to American Optical, Model 820 Reference Manual…..(pdf)

Hotlink to Effective Use and Care of

Microtome…..(LARGE pdf, 93 pgs)

Hotlink to Microtomy and Specimen

Prepartion… Leica manual pdf

Page 3: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: AO820,

Basic

Components/Function

STLCC-CPLS;Morrison 12/18/2015 Page 3

Knife Clamp Lock/unlock:slide carriage to

cut on other area of knife surface

Knife Angle Levers: Lock/unlock to

adjust cutting angle

Knife Angle Position: Screw to adjust

angle after unlocking with levers

Flywheel Lock: move forward to

release for cutting motion, aft to lock

Specimen Mount/Clamp:

Page 4: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: American

Optical, SN, Lube Schedule

STLCC-CPLS;Morrison 12/18/2015 Page 4

Page 5: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: Leica, Sliding; Model SM 2000R

STLCC-CPLS;Morrison 12/18/2015 Page 5

Hotlink to Leica Model SM2000R, Reference Manual…..(pdf)

Page 6: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome; Leica,

Operation Controls

STLCC-CPLS;Morrison 12/18/2015 Page 6

Page 7: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: Leica, Safety Features, “read before use”

STLCC-CPLS;Morrison 12/18/2015 Page 7

Page 8: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome:

Leica,

Specimen

Clamp,

Knife

Holder

STLCC-CPLS;Morrison 12/18/2015 Page 8

Page 9: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: Leica,

Operating the

Instrument

STLCC-CPLS;Morrison 12/18/2015 Page 9

Page 10: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome:

Leica,

Trimming,

Sectioning

STLCC-CPLS;Morrison 12/18/2015 Page 10

Page 11: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

Microtome: Leica, Accessories, Disposable Blade Holder,

Waste tray

STLCC-CPLS;Morrison 12/18/2015 Page 11

Page 12: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015Page 12

Microtome : General Setup and Operating Instructions

1. Prepare specimen in paraffin or desired mounting material

2. Before plugging in the device, mount specimen and the knife and adjust per page 4

3. Plug in the device and set electronic controls per page 5

4. Release the cutting flywheel by moving the locking lever up, page 3

5. Engage the flywheel motion by moving the lever on the flywheel clockwise (cw), page 3

6. Engage knife platform motion by rotating the platform lock lever clockwise (significant force is required), page 3

7. Cancel or stop using the electronic buttons ONLY. Never attempt to hold or grab any component of the machine to stop motion, page 5.

Page 13: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015 Page 13

Microtome; Mechanical Controls, Major

Flywheel; Specimen Motion (up, down,

lock, unlock) rotate to move specimen

above knife area

Rate of Cut, 0-200 mm/sec

(speed of up/down motion) CW to

increase.

Free Flywheel lever ; down to lock, up to

allow cutting motion

Knife Adjustments (see next slide)

Knife Mount Lock; CW to clamp knife

for motion and slicing

Speciment Adjust (see next slide)

Page 14: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015 Page 14

Microtome; Mechanical Controls, Knife (K) and Specimen (S)

(DO NOT PLUG IN DEVICE AT THIS TIME)

K3. Set height of knife, screw CCW, then slide bar

K1. Release Levers; set angle of knife mount

K2. Release Levers; set angle of knife mount

S1. Release to remove entire block if needed

S2. Clamp specimen to block, CW

S3: Specimen side control lock

Page 15: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015 Page 15

Microtome; Historange with Electronic Controls

2. Select what to monitor; stroke,

feed used, feed remaining

1. Plug in device

3. Reset to counters to zero

4. Increase/decrease desired

specimen slice thickness

7. Push both to cancel

slicing

5a. Start a Single cut stroke

5b. Start Multiple cut strokes

6. Must hold down

button to move knife

forward to obtain slices

Hotlink to Historange model LKB 2218 Manual…..(pdf)

Page 16: Caution: Never attempt to hold or grab any component of the …users.stlcc.edu/departments/fvbio/Microtome.pdf · Microtome : General Setup and Operating Instructions 1. Prepare specimen

STLCC-CPLS;Morrison 12/18/2015Page 16

Microtome: Other Protocols or Notes

• Future home of other or more details protocols….