Calibration and Measurement Capability of Gas Analysis

download Calibration and Measurement Capability of Gas Analysis

of 15

description

j

Transcript of Calibration and Measurement Capability of Gas Analysis

  • 1Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 1

    Calibration and Measurement Capability of Gas Analysis

    Calibration and Measurement Calibration and Measurement Capability of Gas AnalysisCapability of Gas Analysis

    ByBy

    Ratirat Ratirat SinweeruthaiSinweeruthai

    Gas Analysis Laboratory

    Department of Chemical Metrology and Biometry

    National Institute of Metrology (Thailand)

    May 25, 2009

    NIMT, Thailand

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 2

    Gas Analysis LaboratoryGas Analysis LaboratoryGas Analysis Laboratory

    As of May 2009

  • 2Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 3

    Gas Analysis SectionGas Analysis SectionGas Analysis Section

    GC : TCD, FID

    SO2 Analyzer

    FT-IR

    Nox Analyzer

    O2 Analyzer

    CO Analyzer

    ND-IR Analyzer

    Micro GC

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 4

    Gas Weighing SectionGas Weighing SectionGas Weighing Section

  • 3Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 5

    Gas Filling SectionGas Filling SectionGas Filling Section

    Cleaning System for Gas CylinderCleaning System for Gas Cylinder Gas Filling System Gas Filling System

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 6

    11stst stage of traceability scheme on calibration at NIMTstage of traceability scheme on calibration at NIMTSI

    CERI

    Japan Gas

    Company

    NIMT

    JapaneseUsers

    Thai Gas Companiesor Users

    NMIJ

    primary

    Secondary

    calibration

    calibration

    Certification(Relative Uncertainty1%2%)

    Transfer

    Secondary

    Working

    High-purity gasCRM

    comparison

    (Firstly, for O2/N2 mixtures around 20% and CO/N2 mixtures 5 ppm 100 ppm)

    jcss Certification, within 1%

    JCSS Certification, 1%2% Working

    calibration

    Accreditation of ISO/IEC 17025

    Secondary

    Working

    Working

  • 4Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 7

    Certified Standard Gas and CalibrationCertified Standard Gas and CalibrationCertified Standard Gas and Calibration

    Uncertainty : within 1%

    CR

    M

    WR

    M

    WR

    M

    RM

    O2 Analyzer

    CO Analyzer

    High Accuracy Low Accuracy

    Uncertainty : within 1%

    It is developing the method for reducing the uncertainty

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 8

    Method for Certified Standard GasMethod for Certified Standard GasMethod for Certified Standard Gas

    Uncertainty : within 1%

    CR

    M

    QC

    cyl

    ind

    er

    WR

    M(s

    am

    ple

    )

    QC

    cyl

    ind

    er

    QC

    cyl

    ind

    er

  • 5Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 9

    Accreditation ScopeFor Certify and Calibration

    Accreditation ScopeAccreditation ScopeFor Certify and CalibrationFor Certify and Calibration

    Calibration O2 Analyzer and Certified O2 in N2

    Range: 0.05 - 0.30 mol/mol

    Uncertainty (k=2) : 0.25 % relative

    WR

    M

    N2

    99

    .99

    5%

    O2 Analyzer

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 10

    Accreditation ScopeFor Certify and Calibration

    Accreditation ScopeAccreditation ScopeFor Certify and CalibrationFor Certify and Calibration

    Calibration CO Analyzer and Certified CO in N2

    Range: 5 - 25 mol/molUncertainty (k=2) : 0.80 % relative

    Range: 25 - 100 mol/molUncertainty (k=2) : 0.45 % relative

    WR

    M

    N2

    99

    .99

    5%

    CO Analyzer

  • 6Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 11

    Quality Systems that need to be fulfilled according to Gas Analysis Laboratory

    Quality Systems that need to be fulfilled Quality Systems that need to be fulfilled according to Gas Analysis Laboratoryaccording to Gas Analysis Laboratory

    Quality System ISO/IEC 17025:2005

    Gas Analysis Production according to ISO 6143

    Analysis SectionAnalysis SectionAnalysis Section

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 12

    Service Route for O2 in N2 Service Route for OService Route for O22 in Nin N2 2

    First Stage:First Stage:Service Type: Calibration and Certified Gas Mixture (CRM)Range: 0.05 mol/mol to 0.30 mol/mol Analysis Technique: Paramagnetic Oxygen AnalyzerService Time: April 2008Certified Reference Materials: CERI (u(x)= 0.2%rel)Bilateral Comparison: NMIJ, August 2008 (Draft B)Accreditation Body: IA Japan (In Process)

    Degrees of equivalence plot

  • 7Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 13

    Service Route for O2 in N2Service Route for OService Route for O22 in Nin N22Second Stage:Second Stage:

    Service Type: Preparation Primary Reference Mixture (PRM)Certified Gas Mixture (CRM)Calibration

    Range: 100 mol/mol to 0.30 mol/molAnalysis Technique: GC-TCD & Paramagnetic Oxygen AnalyzerService Time: within 2009

    Third Stage:Third Stage:Service Type: Preparation Primary Reference Mixture (PRM)

    Certified Gas Mixture (CRM)Calibration

    Range: 1 mol/mol to 100 mol/mol Analysis Technique: GC-PDHIDComparison : NMiService Time: May 2010

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 14

    Service Route for CO2 in N2 Service Route for COService Route for CO22 in Nin N2 2 First Stage:First Stage:

    Service Type: Preparation Primary Reference Mixture (PRM)Certified Gas Mixture (CRM) Calibration

    Range: 100 mol/mol to 0.14 mol/mol Analysis Technique: GC-FID with MethanizerService Time: 2010

    Second Stage:Second Stage:Service Type: Preparation Primary Reference Mixture (PRM)

    Certified Gas Mixture (CRM)Calibration

    Range: 1 mol/mol to 100 mol/mol Analysis Technique: GC-FID with MethanizerComparison : NMiService Time: 2011

  • 8Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 15

    Service Route for CO in N2Service Route for CO in NService Route for CO in N22 First Stage:

    Service Type: Calibration and Certified Gas Mixture (CRM)Range: 5 mol/mol to 100 mol/molAnalysis Technique: ND-IR & GC-FID with MethanizerService Time: April 2008Inter-comparison: CCQM

    Second Stage:Service Type : Preparation Primary Reference Mixture (PRM)

    Certified Gas Mixture (CRM)Calibration

    Range: 100 mol/mol to 0.14 mol/mol Analysis Technique: ND-IR & GC-FID with MethanizerBilateral Comparison: NMiService: 2011

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 16

    First Stage: Inter-comparison CO in N2

    Range: 5 mol/molAnalysis Technique: ND-IR Period: May 2008Certified Reference Material: CERIOrganization: CCQM Status: Draft AAccreditation Body: IA Japan Status: In Process

    NIMT

    Degrees of equivalence plotGC-FID ND-IR FT-IR

  • 9Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 17

    Calibration ServicesCalibration Services CO AnalyzerCO Analyzer OO22 AnalyzerAnalyzer

    Certified Reference MaterialsCertified Reference Materials CO in NCO in N22 OO2 2 in Nin N22

    Training CourseTraining Course Planning to open in 2009Planning to open in 2009

    Calibration, Certification and TrainingCalibration, Certification and TrainingCalibration, Certification and Training

    Primary reference materials Primary reference materials O2 in N2 in range of 1 % to 25 % O2 in N2 in range of 1 % to 25 %

    Planning to sell within 2009Planning to sell within 2009

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 18

    22ndnd stage of traceability scheme on calibration at NIMTstage of traceability scheme on calibration at NIMT(planned)(planned)SI

    Japan Gas

    Company

    NIMT

    JapaneseUsers

    Thai Users

    NMIJ

    Preparation/Calibration

    Secondary

    Secondary

    Working

    High-purityCRM

    Thai Gas

    Company

    Working

    Primary Gas prepared by NIMT

    calibration

    CERI primary

    Secondary

    calibrationcomparison

    jcss Certification, within 1%

    JCSS Certification, 1%2%

    Secondary

    NIMT Certification

    Gas Manufacture Certification

    Transfer

    calibration

    Accreditation of ISO Guide 34

    WorkingWorking

    comparison

    NMIs

  • 10

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 19

    GAS Reference MaterialsGAS Reference MaterialsGAS Reference Materials

    Vehicle emission gas

    Stack emission gas

    Plan to produce in 2009-2012

    Natural Gas

    Ethanol in N2

    Plan to produce in 201x

    Plan to produce in 2011-201x

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 20

    Quality Systems that need to be fulfilled according to Gas Analysis Laboratory

    Quality Systems that need to be fulfilled Quality Systems that need to be fulfilled according to Gas Analysis Laboratoryaccording to Gas Analysis Laboratory

    Quality System ISO/IEC 17025:2005

    Gas Analysis Production according to ISO 6142 and ISO 6143

    Production of Reference Material according to ISO Guide 34

  • 11

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 21

    33rdrd stage of traceability scheme on calibration at NIMTstage of traceability scheme on calibration at NIMT(planned)(planned)SI

    CERI

    Japanese Gas

    Company

    NIMT

    JapaneseUsers

    Thai Users

    NMIJ

    primary

    Secondary

    calibration

    Secondary

    Secondary

    Working

    High-purity gasCRM

    Thai Gas

    CompanyWorking

    primary

    calibration

    SI

    High-purity gas, the purity (impurities) is determined by NIMT

    comparison

    comparison

    jcss Certification, within 1%

    JCSS Certification, 1%2%

    NIMT Certification

    Gas Manufacture Certification

    calibration

    Preparation/Calibration

    Accreditation of ISO Guide 34

    WorkingWorking

    Secondary

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 22

    Mile-stoneMileMile--stonestone

    -2008 ?Calibration using jcss secondary mixturesCertify Working standard gas (ISO/IEC 17025)

    -2010 ?Accreditation of ISO Guide 34Preparation of Primary/Secondary gas mixtures by NIMT

    -201x ? NIMT prepare High-purity gas, Primary/Secondary gas mixtures.

    1st stage

    2nd stage

    3rd stage

  • 12

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 23

    AcknowledgementAcknowledgementAcknowledgement

    How to certified gases accurately.

    How to calibrate gas analyzers accurately

    How to prepare PRM by using gravimetric method for binary mixture.

    How to estimate uncertainty following ISO 6143

    How to manage the quality system ISO/IEC 17025

    National Metrology Institute of Japan (NMIJ)

    National Metrology Institute of Japan National Metrology Institute of Japan (NMIJ)(NMIJ)

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 24

  • 13

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 25

    NMIJ FacilitiesNMIJ FacilitiesNMIJ Facilities

    Gravimetric Method

    ( ISO 6142 )

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 26

    AcknowledgementAcknowledgementAcknowledgement

    How to prepare PRM by using gravimetric method (ISO 6142). Binary Mixture Multi component mixture

    Natural gas mixture Emission gas mixture

    Ethanol gas mixture How to analysis impurity in purity gases or reagent.

    FT-IR GC-PDHID Moisture Analyzer (Cavity ring down technique) Karl Ficher

    Safety principle for preparing PRM Principle for dynamic dilution mixing ISO 6145

    Dutch Metrology Institute (VSL)Dutch Metrology Institute (VSL)Dutch Metrology Institute (VSL)

  • 14

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 27

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 28

    PRM by VSLPRM by VSLPRM by VSL

  • 15

    Chemical Metrology and Biometry Department, National Institute of Metrology (Thailand) 29

    VSL FacilitiesVSL FacilitiesVSL Facilities

    Gravimetric Method

    ( ISO 6142 )

    Dynamic Volumetric Method

    ( ISO 6145 )