Cad Cam Construction of Surgical Imolant for Craniofaical Implant
-
Upload
sorabh-jain -
Category
Documents
-
view
18 -
download
0
Transcript of Cad Cam Construction of Surgical Imolant for Craniofaical Implant
Computer-aided design andmanufacturing construction of a surgicaltemplate for craniofacial implantpositioning to support a definitive nasalprosthesis
Leonardo CioccaMassimiliano FantiniFrancesca De CrescenzioFranco PersianiRoberto Scotti
Authors’ affiliations:Leonardo Ciocca, Maxillo-Facial Prosthodontics,Section of Prosthodontics, Department of Oral Science,Alma Mater Studiorum University of Bologna, Bologna,ItalyMassimiliano Fantini, Francesca De Crescenzio,Virtual Reality and Simulation Laboratory, IIEngineering Faculty, Alma Mater StudiorumUniversity of Bologna, Forlı, ItalyFranco Persiani, II Engineering Faculty, Alma MaterStudiorum University of Bologna, Forlı, ItalyRoberto Scotti, Oral and Maxillo-Facial Rehabilitation,Section of Prosthodontics, Department of Oral Science,Alma Mater Studiorum University of Bologna, Bologna,Italy
Corresponding author:Dr Leonardo CioccaVia S. Vitale 5940125 BolognaItaly.Tel.: þ 39 051 208 8145Fax: þ 39 051 22 5208e-mail: [email protected]
Key words: CAD–CAM, craniofacial implants, nasal prosthesis, rapid prototyping, virtual surgery
Abstract:
Aim: To design a surgical template to guide the insertion of craniofacial implants for nasal prosthesis
retention.
Materials and methods: The planning of the implant position was obtained using software for virtual
surgery; the positions were transferred to a free-form computer-aided design modeling software and
used to design the surgical guides. A rapid prototyping system was used to 3D-print a three-part
template: a helmet to support the others, a starting guide to mark the skin before flap elevation, and a
surgical guide for bone drilling. An accuracy evaluation between the planned and the placed final
position of each implant was carried out by measuring the inclination of the axis of the implant
(angular deviation) and the position of the apex of the implant (deviation at apex).
Results: The implant in the glabella differed in angulation by 7.781, while the two implants in the
premaxilla differed by 1.86 and 4.551, respectively. The deviation values at the apex of the implants
with respect to the planned position were 1.17 mm for the implant in the glabella and 2.81 and
3.39 mm, respectively, for those implanted in the maxilla.
Conclusions: The protocol presented in this article may represent a viable way to position craniofacial
implants for supporting nasal prostheses.
Computer-aided design and manufacturing
(CAD–CAM) technology is developing rapidly
in the field of maxillofacial prosthetics. In recent
articles, we described protocols for elaborating
and rapidly prototyping molds for auricular pros-
theses for a patient who required ablative surgery
of the external ear for tumor removal (Ciocca et
al. 2007a, 2007b) and for a patient affected by
Treacher Collins syndrome (Ciocca et al. 2010a,
2010b). However, these previous reports did not
discuss the problem of identifying the bone avail-
able for craniofacial implant insertion. Osseoin-
tegrated implants have various advantages over
either adhesive or spectacle-retained devices for
the reconstruction of an ablated nose (Ciocca
et al. 2010a, 2010b). They provide better retention
of the prosthesis, so that it is properly positioned
and the patient can wear it more confidently. The
prosthesis can be made thinner, with feathered
edges that blend with the skin, which offers the
patient improved esthetics. Preoperative planning
with the maxillofacial surgeon and the prostho-
dontist is vital for optimal outcomes: today, the
virtual planning of the craniofacial implant inser-
tion and the rapid prototyping of the surgical
template requires the collaboration of the CAD–
CAM specialist engineer. Many articles have
described computerized technology without con-
sidering this important step and the procedure
required for the correct positioning of craniofacial
implants within the external volume of an ear or
nasal prosthesis (Nusinov & Gay 1980; Manko-
vich et al. 1986; Girod et al. 1995; Beumer et al.
1998; Coward et al. 1999; Penkner et al. 1999;
Runte et al. 2002; Cheah et al. 2003a, 2003b;
Hecker 2003; Kubon & Anderson 2003; Lemon
et al. 2003; Reitemeier et al. 2004; Mardini et al.
2005).
Typically, the available bone in the glabella and
the premaxilla is the major factor that determines
implant position for retention of nasal prosthesis.
Diagnosis and treatment planning are multi-dis-
ciplinary, and the use of new CAD–CAM tech-
nologies may improve implant-supported
prosthetic rehabilitation. Borderline patients
may be studied during the diagnostic phase using
Date:Accepted 5 August 2010
To cite this article:Ciocca L, Fantini M, De Crescenzio F, Persiani F, Scotti R.Computer-aided design and manufacturing construction of asurgical template for craniofacial implant positioning tosupport a definitive nasal prosthesis.Clin. Oral Impl. Res. 22, 2011; 850–856.doi: 10.1111/j.1600-0501.2010.02066.x
850 c� 2010 John Wiley & Sons A/S
the virtual simulation of the surgery, avoiding
surgical over-treatments or reducing more aggres-
sive surgeries (Van Steenberghe et al. 2005; Balshi
et al. 2006; Lal et al. 2006; Rosenfeld et al. 2006).
However, no attempt has been made to guide
the position of craniofacial implants using a
prosthetic virtual simulation regarding the final
rehabilitation of the nose or a rapidly prototyped
surgical template derived from this process. This
article describes the computer-aided design and
rapid prototyping of surgical template for the
prosthetically guided insertion of craniofacial
implants.
Material and methods
A 58-year-old man presenting with a total loss of
the nose due to a gunshot was scheduled for a
definitive nasal prosthesis, anchored on osseoin-
tegrated craniofacial implants (Fig. 1).
CT scan elaboration with NobelGuide
The CT data were uploaded into NobelGuide
software (Nobel Biocare, Kloten, Switzerland)
and elaborated to plan the implant surgery in
the nasal region, where a sufficient quantity of
available bone was present. Two implants were
positioned in the premaxillary area in the nasal
floor and one in the glabellar region. The length
of the implants was 11.5 mm and the diameter
was 3.75 mm. After positioning the implants, the
frontal, lateral, and upper orthographic views
(with and without the skeletal region of interest)
were collected as JPG images (Fig. 2).
CT and laser scanner data integration
The CT data were uploaded into Amira 3.1.1
software (Mercury Computer Systems, Chelms-
ford, MA, USA) and elaborated to reconstruct the
3D digital model of the skull surface by setting
the same threshold value used in NobelGuide
(276 Hounsfield Unit [HU]). The 3D digital
model of the skin surface was also obtained by
setting a suitable threshold value. Both models
were achieved semi automatically by threshold-
based segmentation, contour extraction, and sur-
face reconstruction. This process is particularly
useful for distinguishing between soft tissues and
skeletal structures. Moreover, to augment the
obtained region of the face, CT data were inte-
grated with laser scanner data that had been
collected previously for designing and manufac-
turing the eyeglasses-supported provisional nasal
prosthesis. In that instance, a laser scanner
(NextEngine Desktop 3D Scanner; NextEngine,
Santa Monica, CA, USA) was used to acquire the
facial skin surface from five (left, right, frontal,
upper, lower) different perspectives, covering a
wider area of the face with respect to CT data
acquisition focusing just on the nasal defect. As
usual in reverse engineering post-processing, the
five scans have been carefully aligned and merged
to obtain the final digital model of the patient’s
entire face. Skin surfaces from CT and laser
scanner data were both imported into Rapidform
XOS2 (INUS Technology, Seoul, Korea) and re-
gistration process was carried out for data inte-
gration into a single coordinate system. Three
pairs of corresponding reference points were se-
lected on both skin surfaces (CT and laser scan-
ner) for an initial rough alignment. When
Fig. 2. Frontal, lateral, and upper views (with and without the skeletal region of interest) by NobelGuide.
Fig. 1. Initial anatomy.
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
c� 2010 John Wiley & Sons A/S 851 | Clin. Oral Impl. Res. 22, 2011 / 850–856
performing this operation, the first selected shell
(skin surface from laser scanner data) was moved
to the second selected shell (skin surface from
CT data). The refinement of the alignment was
performed semiautomatically using iterative
closest points (ICP)-based fine registration. To
evaluate the accuracy of the registration process,
a surface deviation analysis between CT and
laser scanner data was performed, yielding a
distance mean value of o1 mm. The color
map visualization of the surface deviation ana-
lysis is shown in Fig. 3: the regions with higher
deviation were localized around the mouth and
the residual left nasal ala, probably due to
different facial expressions during the two data
acquisition sessions. After the registration pro-
cess, the skin surface acquired from the laser
scanner was simply used to replace the skin
surface reconstructed from CT data integration
between skull surface from CT and skin surface
from laser scanner (Fig. 4).
Virtual planning transfer
In Rhino 3.0 (Robert McNeel & Associates,
Seattle, WA, USA), the frontal, lateral, and upper
images, without the skeletal region of interest,
collected by NobelGuide, were imported as back-
ground bitmaps in the corresponding views
(scaled to match each other and located in space,
so they all lined up). In each of the three views,
background bitmaps were used as a guide to trace
over the construction lines for the axis of the
three implants. The Crv2View command (curves
from two views) was used to create the axis of the
three implants in 3D space by selecting the
corresponding construction lines in two views.
The implants were modeled as cylinders (dia-
meter 3.75 mm and height 11.5 mm) and placed
in 3D space according to the relative axis and the
background bitmaps (Fig. 5a). For this kind of
construction, just two views are necessary, but
using three views avoid eventually problems due
to overlapping of reference features in the images.
Digital models of skull surfaces from CTand skin
surfaces from laser scanner were also imported for
referencing in 3D space. After replacing back-
ground bitmaps with the images, with the skele-
tal region of interest, collected by NobelGuide,
both digital models were moved, overlapping the
skull surface with respect to the new background
in each view (Fig. 5b).
Designing the template
Once all the models had been imported into
Rhino, a template with surgical guides for im-
plants placement, as planned previously, was
designed. It was developed in three parts: a
main template for referencing on the patient’s
head and two interchangeable overhanging surgi-
cal guides. The main template was designed as a
customized helmet by the offset (5 mm) of the
frontal–upper part of the 3D digital model of the
scanned face to ensure a correct matching with
the head of the patient. A dovetail joint was
added in the front for connecting the two surgical
guides, both provided with guide cylinders. The
first surgical guide was designed just to mark the
skin corresponding to the implants’ axis before
surgically cutting the soft tissues, while the
second one was developed to guide the drilling
of the bone for implant placement (Fig. 6).
Rapid prototyping of the template
The helmet template and the two interchange-
able overhanging surgical guides were directly
manufactured using a 3D soluble support tech-
nology rapid prototyping system (Stratasys, Eden
Prairie, MN, USA). The working principle is
based on fused deposition modeling by acryloni-
trile butadiene styrene plastic material (ABS
P400) and soluble support material to sustain
the prototype under construction. By this pro-
cess, prototypes are built up layer by layer (thick-
ness 0.254 mm) with two available filling
options: solid and sparse. In the first case, each
section of the model is completely filled with
ABS material. In the second one, the interior part
of the model is replaced with a honeycomb
structure. Solid fills are stronger and heavier,
while sparse fills are weaker and lighter, saving
material and speeding up the build process.
Therefore, the digital models were exported in
solid to layer format and directly prototyped in a
single work session, choosing the sparse fill
option for the helmet template and the solid fill
option for the overhanging guides to obtain
stronger elements. The process was completed
by washing the models in an agitation system
with a hot soapy water bath to remove all the
support material for hands-free model completion
(Fig. 7). Table 1 shows the amount and cost of the
ABS and the support material consumed.
Fig. 4. Skull surface from CT data (left), skin and skull surfaces from CT data (center), and integration between the skull
surface from CT data and skin surface from laser scanner data (right).
Fig. 3. Skin surface from CT data (left), laser scanner data (center), and surface deviation analysis between CT and laser
scanner data (right).
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
852 | Clin. Oral Impl. Res. 22, 2011 / 850–856 c� 2010 John Wiley & Sons A/S
Clinical procedure
The surgical template was tested on the patient,
and the insertion procedure was checked in rela-
tion to the dimensions of the drill and the hand
piece before surgical intervention (Fig. 8). The
patient underwent general anesthesia for the
surgery: before starting, the position of the surgi-
cal template was marked onto the skin with a
skin pencil to facilitate repositioning during the
second phase of the surgery. Then, the insertion
landmark points were pointed on the skin and
deeper, with a needle and dermographic ink, as a
guide for flap elevation (Fig. 9a, b). A plastic
surgery was executed to eliminate the left nasal
ala, still present as residual structure from the
first emergency surgery. When the skull bone
was exposed, the second surgical template was
fixed to the helmet and it was repositioned onto
the head according to the lines previously
marked. The surgical pilot drill was guided by
the holes of the templates and a pin was inserted
to test the inclination and the position with
respect to the available bone. The insertion of
implants (Branemark System RP TiUnite, Nobel
Biocare) was performed according to the conven-
tional protocol (Fig. 9c, d), and the flaps were
sutured covering the screw taps of the implants
(Fig. 9e).
Results
A CT scan was performed after the surgery to
verify the accuracy of the surgical protocol with
respect to the CAD–CAM planned design. In Fig.
Fig. 6. Design of the surgical template for implant positioning: helmet template (a),
starting guide to mark the skin (b), surgical guide to drill the bone in the glabellar
region (c), and surgical guide to drill the bone in the premaxillary area (d).
Fig. 7. Rapid prototyping of the surgical template for implant positioning: helmet template, starting guide to
mark the skin, and surgical guides to drill the bone (in the glabellar region and in the premaxillary area).
Table 1. Material cost
Surgical template elements ABSmaterial(cm3)
Supportmaterial(cm3)
Buildingtime
Cost ofmaterial(h)
Helmet template 98.45 116.30 14 h53 min
51.12
Surgical guide to mark the skin 8.82 4.60 1 h 20 min 4.29Surgical guide to drill the bone in theglabellar region
7.67 3.70 1 h 2 min 3.82
Surgical guide to drill the bone in thepremaxillary area
9.68 5.82 1 h 46 min 4.78
Total 124.62 130.42 19 h 1 min 64.01
Fig. 5. Virtual planning transfer (frontal and lateral views): implant modeling, according to background bitmaps (a) and skull
positioning, according to background bitmaps (b).
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
c� 2010 John Wiley & Sons A/S 853 | Clin. Oral Impl. Res. 22, 2011 / 850–856
10 are shown the frontal and lateral views of the
radiographic control after implants placement to
support the nasal prosthesis. The presence of
previously placed implants for dental prosthesis
and the distribution of bullets due to the gunshot
can also be observed.
The post-operative CT data were uploaded into
Amira and elaborated to reconstruct the digital
model of the skull surface by setting the same
threshold value used previously (276HU), and the
3D digital models of the inserted implants were
reconstructed by setting a suitable threshold value.
The digital models were imported in Rapid-
form XOS2 and the registration process was
carried out by selecting three pairs of correspond-
ing reference points on both skull surfaces (pre-
and postsurgical intervention) for initial rough
alignment. The refinement of the alignment was
performed using ICP registration. The process for
pre- and postdata integration into the same co-
ordinate system allowed comparison of the
planned position with the real position of the
implants in Rhino environment (Fig. 11).
The accuracy between planned and placed was
quantitatively evaluated by measuring two para-
meters: the inclination of the axis of the implant
(angular deviation) and the position of the apex of
the implant (deviation at apex; Table 2). Because
the axes of planned and placed implants are repre-
sented by two straight lines in space that do not lie
in a plane, the angular deviation is evaluated as the
minor angle between two skew lines, defined as
either of the angles between any two lines parallel
to them and passing through a point in space.
The implant in the glabella differed in angula-
tion by 7.781, while the two implants in the
premaxilla differed by 1.86 and 4.551, respec-
tively. Consistent with literature data (Van
Steenberghe et al. 2003; Di Giacomo et al.
2005; Ozan et al. 2009) for accuracy when using
the CAD–CAM system during implant surgery,
the angle deviation was acceptable in terms of
safety and prosthetic implications. The deviation
values at the apex of the implants with respect to
the planned position were 1.17 mm for the im-
plant in the glabella and 2.81 and 3.39 mm,
respectively, for those implanted in the maxilla.
The position of the implants resulted in the mean
values of the data from the literature.
Discussion
Different computer-aided surgery systems har-
ness the advantages of optimal 3D diagnosis
and software-based planning by accurately trans-
ferring the virtual implant positions to the corre-
sponding anatomical patient’s sites. Since 1997,
different approaches for computer-assisted
implant planning have been available for oral
Fig. 8. Try-in of the surgical template on the patient and check of the insertion procedure.
Fig. 9. Surgery: initial landmarking (a, b), implant positioning (c, d), and the flap suture (e).
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
854 | Clin. Oral Impl. Res. 22, 2011 / 850–856 c� 2010 John Wiley & Sons A/S
implants (Ewers et al. 2004; Mupparapu & Singer
2004), but none have been useful for the insertion
of craniofacial implants. Moreover, although
several studies have been published on the accu-
racy of these models (Tal & Moses 1991;
DelBalso et al. 1994; White et al. 2001; Hatcher
et al. 2003), no data are available on the accuracy
of craniofacial implant positioning with respect to
the planned position.
This study presented a new protocol of con-
structing surgical templates for craniofacial im-
plants. Starting from a 3D system (NobelGuide,
Nobel Biocare), planning of the ideal position of
three craniofacial implants was carried out. The
spatial position of each implant, represented in
the NobelGuide, was transferred into CAD soft-
ware that allowed projection of the surgical tem-
plate. When the helmet and the inserts were
prototyped and sterilized, the patient was sched-
uled for surgery.
The main innovation in such a procedure was
the integration of CTand laser scanner data as the
starting point for the design of the surgical guide.
Coordinate systems integration from multimodal
devices has already been carried out for the gen-
eration of a craniofacial database (Suwardhi et al.
2005), but not for the purpose described in this
paper. To reduce the patient’s exposure to X-rays,
CT scanning should be focused on the effective
region of interest, while the completely safe 3D
laser scanning may cover a much wider area. Thus,
the CT scan, focused on the nasal defect, was
integrated with 3D laser scanning of the entire face
and head surface to design the customized helmet.
The main problem of such a surgical guide was
the stability: the helmet was designed on a rigid
and fixed frontal surface of the patient, while the
skin is resilient and mobile. Even if the glabella
landmark can be readily detected, minor pressure
on the template may dislocate it in a wrong
position. Taking care in positioning it with respect
to the three landmarks (two supra-ocular bone
arches and glabella) with no pressure on the skin,
the template position was accurately marked on
the frontal and lateral skin of the skull, so as to
reproduce the same position each time the tem-
plate was used. However, a bone pin retention
system will also be necessary in the future for
better stabilization of the template.
As result of this main problem, the implant in
the glabellar region was affected by an error in the
inclination in spite of a very good placement
position. The two implants in the premaxillary
area in the nasal floor were more internally
displaced with respect to the planned place due
to a slightly incorrect repositioning of the helmet
during the two-step drilling surgery. Moreover,
they were both inserted into the bone tissue less
in depth than planned.
The main advantages of this protocol are the use
of a CAD–CAM system to guide the implant
surgery and to project the implant position accord-
ing to prosthetic options. The first advantage
allows the surgeon to accurately plan the flap and
the plastic surgery: in the case reported here, for
example, a reduction in the thickness of the mu-
codermal flap in the floor of the nose was necessary
to obtain a correct emerging profile of the healing/
prosthetic abutments. The second advantage allows
the prosthodontist to take into account the pros-
thetic issues in terms of the inclination for a better
impression and accessibility for home hygienic
maintenance care around implants.
The main disadvantages are that it is consum-
ing and expensive due to the elaboration of the
surgical template. Moreover, the rapid prototyp-
ing equipment (software and 3D printer) may
represent a barrier.
Conclusions
The protocol presented here simplifies the im-
plant surgery for the insertion of craniofacial
implants to support a nasal prosthesis. The sur-
gical template can be rapidly manufactured using
CAD–CAM technology in combination with
other systems for virtual surgery. This protocol
facilitates a more accurate positioning of cranio-
facial implants than unguided surgery.
Fig. 10. Radiographic control after implants placement to support the nasal prosthesis (indicated by black arrows): (a) frontal
view; (b) lateral view. In the radiography can also be noticed the presence of previously placed implants for dental prosthesis
and the distribution of bullets due to the gunshot.
Fig. 11. Comparison between the planned (continuous line axis) and placed (dashed line axis) implant position to
quantitatively evaluate the accuracy: (a) frontal view; (b) lateral view.
Table 2. Values of accuracy measurements
Site Deviationat apex (mm)
Angulardeviation (1)
Glabella 1.17 7.78Maxilla 1 2.81 1.86Maxilla 2 3.39 4.55
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
c� 2010 John Wiley & Sons A/S 855 | Clin. Oral Impl. Res. 22, 2011 / 850–856
References
Balshi, S.F., Wolfinger, G.J. & Balshi, T.J. (2006) Surgical
planning and prosthesis construction using computed
tomography, CAD/CAM technology, and the internet
for immediate loading of dental implants. Journal of
Esthetic Restorative Dentistry 18: 312–323.
Beumer, J., Curtis, T.A. & Marunick, M.T. (1998)
Maxillofacial Rehabilitation: Prosthetic and Surgical
Consideration. 1st edition, 377–453. St. Louis:
Ishiyaku EuroAmerica.
Cheah, C.M., Chua, C.K., Tan, K.H. & Teo, C.K.
(2003a) Integration of laser surface digitizing with
CAD/CAM techniques for developing facial prosthe-
sis. Part I: design and fabrication of prosthesis replicas.
International Journal of Prosthodontics 16: 435–441.
Cheah, C.M., Chua, C.K., Tan, K.H. & Teo, C.K.
(2003b) Integration of laser surface digitizing with
CAD/CAM techniques for developing facial prosthe-
sis. Part II: development of molding techniques for
casting prosthetic parts. International Journal of
Prosthodontics 16: 543–548.
Ciocca, L., De Crescenzio, F., Fantini, M. & Scotti, R.
(2010a) CAD/CAM bilateral ear prostheses construc-
tion for Treacher Collins syndrome patients using
laser scanning and rapid prototyping. Computers
Methods in Biomechanics and Biomedical Engineer-
ing 13: 379–386.
Ciocca, L., Fantini, M., Marchetti, C., Scotti, R. &
Monaco, C. (2010b) Immediate facial rehabilitation in
cancer patients using CAD–CAM and rapid prototyp-
ing technology: a pilot study. Supportive Care in
Cancer 18: 723–728.
Ciocca, L., Maremonti, P., Bianchi, B. & Scotti, R.
(2007a) Maxillofacial rehabilitation after rhinectomy
using two different treatment options: clinical reports.
Journal of Oral Rehabilitation 34: 311–315.
Ciocca, L., Mingucci, R., Gassino, G. & Scotti, R. (2007b)
CAD/CAM ear model and virtual construction of the
mold. Journal of Prosthetic Dentistry 98: 339–343.
Coward, T.J., Watson, R.M. & Wilkinson, I.C. (1999)
Fabrication of a wax ear by rapid-process modeling
using stereolithography. International Journal of
Prosthodontics 12: 20–27.
DelBalso, A.M., Greiner, F.G. & Licata, M. (1994) Role
of diagnostic imaging in evaluation of the dental
implant patient. Radiographics 14: 699–719.
Di Giacomo, G.A., Cury, P.R., de Araujo, N.S., Sen-
dyk, W.R. & Sendyk, C.L. (2005) Clinical application
of stereolithographic surgical guides for implant place-
ment: preliminary results. Journal of Periodontology
76: 503–507.
Ewers, R., Schicho, K., Truppe, M., Seemann, R.,
Reichwein, A., Figl, M. & Wagner, A. (2004) Com-
puter-aided navigation in dental implantology: 7 years
of clinical experience. International Journal of Oral
and Maxillofacial Surgery 62: 329–334.
Girod, S., Keeve, E. & Girod, B. (1995) Advances in
interactive craniofacial surgery planning by 3D simu-
lation and visualization. International Journal of Oral
and Maxillofacial Surgery 24: 120–125.
Hatcher, D.C., Dial, C. & Mayorga, C. (2003) Cone beam
CT for pre-surgical assessment of implant sites. Journal
of the California Dental Association 31: 825–833.
Hecker, D.M. (2003) Maxillofacial rehabilitation of a
large facial defect resulting from an arteriovenous
malformation utilizing a two-piece prosthesis. Journal
Prosthetic Dentistry 89: 109–113.
Kubon, T.M. & Anderson, J.D. (2003) An implant-
retained auricular impression technique to minimize
soft tissue distortion. Journal of Prosthetic Dentistry
89: 97–101.
Lal, K., White, G.S., Morea, D.N. & Wright, R.F. (2006)
Use of stereolithographic templates for surgical and
prosthodontic implant planning and placement. Part
I. The concept. Journal of Prosthodontics 15: 51–58.
Lemon, J.C., Okay, D.J., Powers, J.M., Martin, J.W. &
Chambers, M.S. (2003) Facial moulage: the effect of a
retarder on compressive strength and working and
setting times of irreversible hydrocolloid impression
material. Journal of Prosthetic Dentistry 90:
276–281.
Mankovich, N.J., Curtis, D.A., Kagawa, T. & Beumer, J.
III (1986) Comparison of computer-based fabrication of
alloplastic cranial implants with conventional techni-
ques. Journal of Prosthetic Dentistry 55: 60–69.
Mardini, M.A., Ercoli, C. & Grasser, G.N. (2005) A
technique to produce a mirror-image wax pattern of
an ear using rapid prototyping technology. Journal of
Prosthetic Dentistry 94: 1955–1958.
Mupparapu, M. & Singer, S.R. (2004) Implant imaging
for the dentist. Journal of the Canadian Dental
Association 70: 32–37.
Nusinov, N.S. & Gay, W.D. (1980) A method for
obtaining the reverse image of an ear. Journal of
Prosthetic Dentistry 44: 68–71.
Ozan, O., Turkyilmaz, I., Ersan Ersoy, E., McGlumphy,
E.A. & Rosenstiel, S.F. (2009) Clinical accuracy of 3
different types of computed tomography derived
stereolithographic surgical guides in implant place-
ment. Journal of Oral & Maxillofacial Surgery 67:
394–401.
Penkner, K., Santler, G., Mayer, W., Pierer, G. &
Lorenzoni, M. (1999) Fabricating auricular prosthesis
using three-dimensional soft tissue model itself. Jour-
nal of Prosthetic Dentistry 82: 482–484.
Reitemeier, B., Notni, G., Heinze, M., Schone, C.,
Schmidt, A. & Fichtner, D. (2004) Optical modeling
of extraoral defects. Journal of Prosthetic Dentistry
91: 80–84.
Rosenfeld, A.L., Mandelaris, G.A. & Tardieu, P.B.
(2006) Prosthetically directed implant placement
using computer software to ensure precise placement
and predictable prosthetic outcomes. Part 3: stereo-
lithographic drilling guides that do not require bone
exposure and the immediate delivery of teeth. Inter-
national Journal of Periodontics and Restorative
Dentistry 26: 493–499.
Runte, C., Dirksen, D., Delere, H., Thomas, C., Runte,
B., Meyer, U., Von Bally, G. & Bollmann, F. (2002)
Optical data acquisition for computer-assisted design
of facial prostheses. International Journal of Prostho-
dontics 15: 129–132.
Suwardhi, D., Setan, H., Chong, A.K. & Rajion, Z.
(2005) Coordinate systems integration for craniofacial
database from multimodal devices. Proceedings ITB
Engineering Science 37: 1–14.
Tal, H. & Moses, O. (1991) A comparison of panoramic
radiography with computed tomography in the plan-
ning of implant surgery. Dentomaxillofacial Radi-
olology 20: 40–42.
Van Steenberghe, D., Glauser, R., Blomback, U., An-
dersson, M., Schutyser, F., Pettersson, A. & Wendel-
hag, I. (2005) A computed tomographic scanderived
customized surgical template and fixed prosthesis for
flapless surgery and immediate loading of implants in
fully edentulous maxillae: a prospective multicenter
study. Clinical Implant Dentistry and Related Re-
search 7 (Suppl. 1): S111–S120.
Van Steenberghe, D., Malevez, C., Van Cleynenbreugel,
J., Serhal, C.B., Dhoore, E., Schutyser, F., Suetens, P.
& Jacobs, R. (2003) Accuracy of drilling guides
for transfer from three-dimensional CT-based planning
to placement of zygoma implants in human cadavers.
Clinical Oral Implants Research 14: 131–136.
White, S.C., Heslop, E.W., Hollender, L.G., Mosier,
K.M., Ruprecht, A. & Shrout, M.K. (2001) Para-
meters of radiologic care: an official report of
the American Academy of Oral and Maxillofacial
Radiology. Oral Surgery, Oral Medicine, Oral
Pathology, Oral Radiology and Endodontics 91:
498–511.
Ciocca et al �CAD–CAM construction of a surgical template for craniofacial implant positioning
856 | Clin. Oral Impl. Res. 22, 2011 / 850–856 c� 2010 John Wiley & Sons A/S
Copyright of Clinical Oral Implants Research is the property of Wiley-Blackwell and its content may not be
copied or emailed to multiple sites or posted to a listserv without the copyright holder's express written
permission. However, users may print, download, or email articles for individual use.