Bureau of Drug and Narcotic
Mechanical Calibration
Temperature control Calibrated thermometer 37 + 0.5 ˚C
Distance between the inside bottom of the vesseland the bottom of the basket or paddle is maintained at 25 + 2 mm during the test use a spacing gauge or ball
![Page 2: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/2.jpg)
USP Performance Verification Test (PVT)
Bureau of Drug and Narcotic
• Provide experimental proof that the dissolution test assembly is suitable for dissolution testing
• Confirms the acceptable operation of the equipment
• The result must conform to range given in the certificate for the specification lot of reference material.
• PVT is a prerequisite for a proper performance qualification of the dissolution assembly.
![Page 3: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/3.jpg)
Bureau of Drug and Narcotic
Reference Standards for PVT
![Page 4: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/4.jpg)
Bureau of Drug and Narcotic
Prednisone USP RS
Check • Current lot• Validity
![Page 5: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/5.jpg)
Bureau of Drug and Narcotic
Prednisone USP RS
Certificate Label
![Page 6: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/6.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Check • Current lot• Validity
![Page 7: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/7.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Certificate & Label
![Page 8: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/8.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Store in a dry place. Store at controlled room temperature Not exceeding 25 ºC
![Page 9: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/9.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Certificate
![Page 10: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/10.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Dissolution medium:- Water: DI - Preparing the medium “Deaeration”
![Page 11: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/11.jpg)
USP Deaeration Technique- Heat water while stirring gently to about 41 – 45 C-Filter under vacuum through 0.45 um porosity filter into a filtering flask with a stirring device-Seal the flask, continue to apply vacuum while stirring for 5 min-Vacuum should be < 100 mbar-Temperature of medium not fall below 37 C
![Page 12: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/12.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Procedure: 499 g of deaerated water (500 ml) 37 C each apparatus at 50-rpm 30 minutes uv at 242 nm alcohol < 5% of total volume of standard solution (If used)
![Page 13: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/13.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
• Single - Stage Test• Two – Stage Test
Result: • GM • %CV
![Page 14: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/14.jpg)
Bureau of Drug and Narcotic
Prednisone Tablets RS
Futility Factor
![Page 15: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/15.jpg)
Bureau of Drug and Narcotic
• Deaerated water• Weight and carefully introduce dissolution medium to the vessel 499 ml (500 g)• Allow media in each vessel to reach 37 + 0.5 ºC and use immediately.• Always handle tablets with - gloves (not cotton) - forceps or tweezers which will not scratch or damage the surface of the tablets• Examine all 6 tablets, do not use chipped, cracked or Capped tablets• No need to weigh Tablets RS beforehand (Weight is for information and investigation purposes)
Steps for PVT
![Page 16: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/16.jpg)
Bureau of Drug and Narcotic
• Prepare to drop tablets.• For Apparatus 2 Paddle - Drop tablets into non-rotating apparatus - Tablets must settle to the bottom of the vessel before rotation of the shaft begins.• Record time• Samples should remain at the bottom center of the vessel • Visually inspect and record any unusual observations• Withdraw sample at 30 minutes + 36 seconds (+2%) and filter immediately• Measure the amount of prednisone dissolved at 242 nm in comparison with a solution of known concentration of USP Prednisone RS• Calculation and conclusion
Steps for PVT
![Page 17: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/17.jpg)
Bureau of Drug and Narcotic
www.usp.orgSearch for PVT calculation tool
Select“Begin using the tool”
PVT Calculation Toolkit
![Page 18: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/18.jpg)
Bureau of Drug and Narcotic
Select prednisone tablet RS lot
PVT Calculation Toolkit
![Page 19: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/19.jpg)
Bureau of Drug and Narcotic
PVT Calculation Toolkit
![Page 20: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/20.jpg)
Bureau of Drug and Narcotic
Select apparatus
PVT Calculation Toolkit
![Page 21: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/21.jpg)
Bureau of Drug and Narcotic
Select configuration of dissolution
PVT Calculation Toolkit
![Page 22: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/22.jpg)
Bureau of Drug and Narcotic
Select testing procedure1-stage or2-stage
PVT Calculation Toolkit
![Page 23: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/23.jpg)
Bureau of Drug and Narcotic
Submit data
PVT Calculation Toolkit
![Page 24: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/24.jpg)
Bureau of Drug and Narcotic
Information - Date - Operator - ID - Note
PVT Calculation Toolkit
![Page 25: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/25.jpg)
Bureau of Drug and Narcotic
Data of6 vessels
with 3 decimals
PVT Calculation Toolkit
![Page 26: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/26.jpg)
Bureau of Drug and Narcotic
Submit data
PVT Calculation Toolkit
![Page 27: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/27.jpg)
Bureau of Drug and Narcotic
Result & Report
PVT Calculation Toolkit
![Page 28: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/28.jpg)
Recommendations for PVT
Bureau of Drug and Narcotic
- Designated vessels and shaft for each position in dissolution instrument- Rigorous mechanical calibration- Proper deaeration of dissolution medium- Test should be started as soon as possible after media deaeration- Filter samples immediately after withdrawing samples from each vessel
![Page 29: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/29.jpg)
Prednisone Tablets RS
Bureau of Drug and Narcotic
Tablet should be stored in original container, according to description on label Remove tablets from package immediately prior to testing Do not put tablet RS next to water bath while waiting for bath to equilibrate Avoid exposure to excess humidity Prednisone Tablets RS will give low results Do not dry Prednisone Tablets RS in an oven No need to weigh Tablets RS beforehand
![Page 30: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/30.jpg)
Sources of factors influence dissolution
Bureau of Drug and Narcotic
Note: instrument and analyst are qualified
![Page 31: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/31.jpg)
Vibration
Bureau of Drug and Narcotic
USP specification: “ No part of the assembly, including the environment in which the assembly is placed, contributes significant motion, agitation, or vibration beyond that due to the smoothly rotating stirring element.”
Source of vibration: * Apparatus - Tension or dirt on the drive chain or belt - Worn parts - Alignment of belt guide plates - Surging of individual spindles - Turbulence in the water bath
![Page 32: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/32.jpg)
Laboratory Environment
Bureau of Drug and Narcotic
Has anything changed in the lab environment? - New equipment added to the immediate area - Other changes (e.g. construction)
Vibration transfer - Be careful how many testers put on the same bench - Is bench suitable? - What else is on the bench or nearby? shaker, mixer, centrifuge - consider use of vibration damping pads
![Page 33: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/33.jpg)
Medium
Bureau of Drug and Narcotic
Using deaerated media - use immediately after deaeration - transfer media carefully to minimize re-aeration of the media
Air bubbles can attach to particles - decreasing dissolution by reducing surface area - Increasing dissolution by # lifting and moving particles # bubble clogging the meshes of the basket and retaining particles in a zone of higher shear stress compare to the vessel bottom
![Page 34: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/34.jpg)
Filtration
Bureau of Drug and Narcotic
Removal of undissolved particles i.e. drug or excipients in the sample solution
Filtration should occur concomitantly with or immediately after sampling
Unfiltered sample can continue drug dissolution before filtration
Use Individual, clean syringes and filters for sampling each vessel, each time
![Page 35: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/35.jpg)
Filtration
Bureau of Drug and Narcotic
Precautions:
- Prewetting may be necessary - Adsorption of the drug onto the filter has to be evaluated
- Chemical compatibility and extractable impurities should be considered by the selection of the proper (inert) filter material
- Important to perform filter check/validation*** PVDF membrane filter is recommended for PVT
![Page 36: Bureau of Drug and Narcotic Mechanical Calibration Temperature control Calibrated thermometer 37 + 0.5 ˚ C Distance between the inside bottom of the.](https://reader035.fdocuments.net/reader035/viewer/2022070401/56649f1f5503460f94c375c7/html5/thumbnails/36.jpg)
When to perform PVT ?
Bureau of Drug and Narcotic
Installation of new apparatus Movement or relocation of apparatus Equipment maintenance - Adjustment of drive mechanism - Replacement of drive belt - Replacement of circulator - Replacement of vessels, baskets, paddle etc Routine test schedule If only one apparatus, basket or paddle, is used, then only that apparatus needs to be tested.