Krishna Persaud, SCEAS, The University of Manchester, UK Biosensors Krishna Persaud

Transcript of Biosensors - ISOCS

Page 1: Biosensors - ISOCS

Krishna Persaud, SCEAS, The University of Manchester, UK

Biosensors Krishna Persaud

Page 2: Biosensors - ISOCS

• Biomolecules and their uses

• An Overview

Inspiration from Nature

Krishna Persaud, SCEAS, The University of Manchester, UK 2

Page 3: Biosensors - ISOCS

Pit Viper

Krishna Persaud, SCEAS, The University of Manchester, UK 3

Thermoreceptors

Page 4: Biosensors - ISOCS

From Runaway, by Michael Crichton.

Published by Tri-Star Pictures in 1985

Sniffing Robot

4

Page 5: Biosensors - ISOCS

Father of the Biosensor

Professor Leland C Clark Jnr

1918–2005 5 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 6: Biosensors - ISOCS

Components of a Biosensor

Detector

6

Page 7: Biosensors - ISOCS

1916 First report on immobilization of proteins : adsorption of invertase on activated charcoal

1922 First glass pH electrode

1956 Clark published his definitive paper on the oxygen electrode.

1962 First description of a biosensor: an amperometric enzyme electrodre for glucose (Clark)

1969 Guilbault and Montalvo – First potentiometric biosensor:urease immobilized on an ammonia electrode to detect urea

1970 Bergveld – ion selective Field Effect Transistor (ISFET)

1975 Lubbers and Opitz described a fibre-optic sensor with immobilised indicator to measure carbon dioxide or oxygen.

History of Biosensors

7 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 8: Biosensors - ISOCS

1975 First commercial biosensor ( Yellow springs

Instruments glucose biosensor)

1975 First microbe based biosensor, First immunosensor

1976 First bedside artificial pancreas (Miles)

1980 First fibre optic pH sensor for in vivo blood gases (Peterson)

1982 First fibre optic-based biosensor for glucose

1983 First surface plasmon resonance (SPR) immunosensor

1984 First mediated amperometric biosensor: ferrocene used with glucose oxidase for glucose detection

History of Biosensors

8 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 9: Biosensors - ISOCS

1987 Blood-glucose biosensor launched by MediSense ExacTech

1990 SPR based biosensor by Pharmacia BIACore

1992 Hand held blood biosensor by i-STAT

1996 Launching of Glucocard

1998 Blood glucose biosensor launch by LifeScan FastTake

1998 Roche Diagnostics by Merger of Roche and Boehringer mannheim

2000-Current Quantum dots, nanoparticles, nanowires, nanotubes, etc

History of Biosensors

9 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 10: Biosensors - ISOCS

1. LINEARITY Linearity of the sensor should be high

forthe detection of high substrate

concentration.

2. SENSITIVITY Value of the electrode response per

substrate concentration.

3. SELECTIVITY Chemicals Interference must be

minimised for obtaining the correct

result.

4.RESPONSE TIME Time necessary for having 95%

of the response.

Basic Characteristics of a

Biosensor

10 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 11: Biosensors - ISOCS

Biomolecules

11

Peptides, polypeptides (chain of

amino acids)

Proteins

E.g. enzymes, antibodies

Nucleic acids

E.g. ribonucleic acid (uracil

not thymine),

deoxyribonucleic acid

Polysaccharides

E.g. cellulose

Lignin (plant cells)

Page 12: Biosensors - ISOCS

Key-Lock Mechanism – Selectivity

12

Enzyme to catalyse reaction

Antibody to neutralise bacteria and virus

Page 13: Biosensors - ISOCS

Electrochemical biosensors

13

Blood glucose sensor

Diabetes common in

elderly people/obese

Millions sold since

1993

Disposable strips

Sold by Abbott et al.

Page 14: Biosensors - ISOCS

Electrochemical biosensors

14

Based on electrochemistry

Use enzymatic redox reaction

Plastic strips with polymers

and enzyme

Millions sold since 1993

2 electrode cell

Page 15: Biosensors - ISOCS

Optical biosensors – I

15

Fluorescent dyes

Page 16: Biosensors - ISOCS

Optical biosensors – Labelled antibodies

16

Inject antigen into host

(chicken, duck, mouse, rat,

goat, horse, sheep ..) and

obtain anti-body

Large molecule 150kDa

Label anti-body (e.g. anti-

goat) with fluorescent dye

Used for biochemical assays

Like ELISA

Page 17: Biosensors - ISOCS

Competitive ELISA

17

Enzyme-linked immuno

sorbent array

Many welled micro-titer

plate

Page 18: Biosensors - ISOCS

Optical fibre biosensors - II

18

Surface plasmon resonance

sensor

Page 19: Biosensors - ISOCS

Receptors as biosensors

19

7TM receptor/GPCR

Target protein

Measure conformal

change or cell pathway

Page 21: Biosensors - ISOCS

Cell based biosensors

21

Cytometers for

Red blood cells

Platelets

White blood cells

Bacteria

SAW resonators

Cantilever beams

1 ul blood = 5M erythrocytes, 500k thrombocytes, 10k leukocytes

Page 22: Biosensors - ISOCS

Biosensor

Analyte

Sample handling/ preparation

Detection

Signal

Analysis

Response

Biosensor Components

Page 23: Biosensors - ISOCS

1. The Analyte (What do you want to detect)

Molecule - Protein, toxin, peptide, vitamin, sugar,

metal ion

2. Sample handling (How to deliver the analyte to the sensitive region?)

(Micro) fluidics - Concentration increase/decrease),

Filtration/selection

Biosensor

Page 24: Biosensors - ISOCS

4. Signal

(How do you know there was a detection)

3. Detection/Recognition

(How do you specifically recognize the analyte?)

Biosensor

24 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 25: Biosensors - ISOCS

Example of biosensors

Pregnancy test

Detects the hCG protein in urine.

Glucose monitoring device (for diabetes patients)

Monitors the glucose level in the blood.

Page 26: Biosensors - ISOCS

Example of biosensors

Infectous disease biosensor

from RBS

Old time coal miners’ biosensor

Page 27: Biosensors - ISOCS

electrode

substrate product

Enzyme

Apply voltage Measure current proportional

to concentration of substrate

Principle of Electrochemical Biosensors

Page 28: Biosensors - ISOCS

E, V

time

E-t waveform

potentiostat

Electrochemical cell

counter

working electrode

N2

inlet

Protein film

reference

insulator electrode

material

Equipment for developing electrochemical biosensors

Cyclic

voltammetry

28 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 29: Biosensors - ISOCS

-2

-1.5

-1

-0.5

0

0.5

1

1.5

2

-0.8-0.6-0.4-0.200.20.40.60.8

I ,

A

E, V vs SCE

Cyclic voltammogram (CV) at 100 mV s-1

and 25 oC of Mycobacterium

Tuberculosis KatG catalase-peroxidase in a thin film of

dimyristoylphosphatidylcholine on basal plane PG electrode, in anaerobic pH 6.0

buffer.

Oxidation

Of FeII

Reduction

Of FeIII Reversible

Peaks for

Direct electron

transfer

Cyclic Voltammetry

Page 30: Biosensors - ISOCS

Electrode

enzyme

A lipid-enzyme film

Page 31: Biosensors - ISOCS

Catalytic enzyme electrochemistry

a basis for biosensor - glucose oxidase

oxidation

Fc mediator

Fc + glucose

+ enzyme

I = f [glucose]

A. Cass, G. Davis, G. D. Francis, H. O. A. Hill, W. J. Aston, I. J. Higgins, E. V. Plotkin, L. D. L.

Scott, A. P. F. Turner, Anal. Chem. 56, 667-671 (1984).

Mediator shuttles

Electrons between

Enzyme and electrode

Page 32: Biosensors - ISOCS

Scheme 2

Glucose + GO(FAD) + 2 H+ gluconolactone + GO(FADH2) (1)

GO(FADH2) + 2 Fc+ GO(FAD) + 2 Fc + 2 H

+ (4)

Fc Fc+ + 2 e

- (at electrode) (5)

Mechanism for catalytic oxidation of glucose

With Glucose oxidase (GO) and Fc mediator

Signal can also be measured by amperometry:

Hold const. E where oxidation occurs, measure I vs time

Fc = ferrocenecarboxylate

Page 33: Biosensors - ISOCS

Commercial Glucose Sensors

• Biggest biosensor success story!

• Diabetic patients monitor blood glucose

at home

• First made by Medisense (early 1990s),

now 5 or more commercial test systems

• Rapid analysis from single drop of blood

• Enzyme-electrochemical device on a

slide

Page 34: Biosensors - ISOCS

Patient Diabetes Management

Insulin secretion by pancreas regulated

by blood glucose, 4.4 to 6.6 mM normal

In diabetes, regulation breaks down

Wide swings of glucose levels

Glucose tests tell patient how much

insulin to administer

34 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 35: Biosensors - ISOCS

• Most sensors use enzyme called glucose oxidase (GO)

• Most sensors are constructed on electrodes, and use a

mediator to carry electrons from enzyme to GO

Fc = mediator, ferrocene, an iron complex

These reactions occur in the sensor:

Fc

Fc+ + e- (measured)

GOR + 2 Fc + --> GOox + 2 Fc

GOox + glucose --> GOR + gluconolactone

Reach and Wilson, Anal. Chem. 64, 381A (1992)

G. Ramsay, Commercial Biosensors, J. Wiley, 1998.

Page 36: Biosensors - ISOCS

Glucose biosensor test strips (~$0.40-0.80 ea.)

Meter

Read glucose

Dry coating of GO + Fc

Patient adds drop of blood,

then inserts slide into meter

Output:

amperometry I

t

Patient reads glucose level on meter

e’s

electrodes

Page 37: Biosensors - ISOCS

Research on glucose sensors

Non-invasive biosensors - skin, saliva

Implantable glucose sensors to

accompany artificial pancreas -

feedback control of insulin supply

Record is 3-4 weeks for implantable

sensor in humans

37 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 38: Biosensors - ISOCS

Other biosensors

Cholesterol - based on cholesterol oxidase

Antigen-antibody sensors - toxic substances, pathogenic bacteria

Small molecules and ions in living things: H+, K+, Na+, CO2, H2O2

DNA hybridization and damage

Micro or nanoarrays, optical abs or fluor.

38 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 39: Biosensors - ISOCS

Negative

surfacePolycation soln.,

then wash

+ + + + + + + + + + + +

soln. of negative protein

then wash

+ + + + + + + + + + + +

+ + + + + + + + + + + +

+ + + + + + + + + + + +

Repeat steps for desired

number of layers

Protein

layer

Polycation layersProtein

layer

Polycation soln.,

then wash

Figure 19

Layer by layer

Film construction:

Page 40: Biosensors - ISOCS

PSS layer

SPAN layer

Enzyme

layer

Detection of hydrogen peroxide Conductive polymers efficiently wire

peroxidase enzymes to graphite

Xin Yu, G. A. Sotzing, F. Papadimitrakopoulos, J. F. Rusling,

Highly Efficient Wiring of Enzymes to Electrodes by Ultrathin

Conductive Polyion Underlayers: Enhanced Catalytic

Response to Hydrogen Peroxide, Anal. Chem., 2003, 75,

4565-4571.

e’s

(sulfonated polyaniline)

Page 41: Biosensors - ISOCS

Horseradish Peroxidase (HRP)

100nm

50nm

Tapping mode atomic force microscopy (AFM) image

of HRP film

Page 42: Biosensors - ISOCS

O2

H2O2

PFeIII PFeII PFeII-O2

2e-, 2H+

H2O2 + PFeII•PFeIV=O

active oxidant

+e-

-e-

PFeIII + H2O + O2

H2O2

Possible reduced species in red

Electrochemical Response of Peroxidases

Page 43: Biosensors - ISOCS

-10

0

10

20

30

40

50

60

-0.8-0.6-0.4-0.200.2

I,A

E, V vs SCE

with SPAN

a

0

0.5

2

4

6

7.5

M H20

2

Catalytic reduction of H2O2 by peroxidase films

Catalytic cycles increase current

FeIII/FeII

reduction

Page 44: Biosensors - ISOCS

0

0.5

1

0 100 200 300 400

I,

A

t, s

with PAPSA

without PAPSA

Rotating electrode amperometry at 0 V

HRP, 50 nmol H2O2 additions

span

No span

reduction

Page 45: Biosensors - ISOCS

0

0.2

0.4

0.6

0.8

1

1.2

0 0.1 0.2 0.3 0.4 0.5 0.6

I,

A

[H2O

2], M

PAPSA/HRPPAPSA/Mb

Mb

HRP

Rotating electrode amperometry at 0 V

Sensitivity much higher with conductive polymer (SPAN);

Electrically wires all the protein to electrode

Span/HRP

Span/Mb

Page 46: Biosensors - ISOCS

Part 2

Immobilisation Techniques

Krishna Persaud, SCEAS, The University of Manchester, UK 46

Page 47: Biosensors - ISOCS

47 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 48: Biosensors - ISOCS

• Enzymes

• Antibodies

• DNA

• Receptors

• Organelles

• Microorganisms

• Animal and plant cells or tissues

Biological Sensing Elements

48 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 49: Biosensors - ISOCS

• Specificity

• Storage

• Operational and environmental stability

• Analyte to be detected, - chemical

compounds, antigens, microbes,

hormones, nucleic acids, or subjective

parameters such as smell and taste.

Choice of Biomaterial

49 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 50: Biosensors - ISOCS

• Thermal stability

• Ability to act in highly acidic, alkaline,

hydrophobic (organic solvent), or oxidizing

environments

• Extremophiles such as the thermophiles,

alkalophiles, and halophiles have gained

importance in the production of enzymes

for use in biosensors and other

applications

Environmental Conditions

50 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 51: Biosensors - ISOCS

Purified enzymes have been most

commonly used in the construction of

biosensors.

The major advantage of using a pure

enzyme is its high analytical specificity.

Disadvantages: High cost, poor stability

Enzymes

51 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 52: Biosensors - ISOCS

• Immobilization of whole cells has been

shown to be a better alternative to

immobilization of purified enzymes

• Avoids the lengthy and expensive

operations of enzyme purification

• Preserves the enzyme in its natural

environment, thus protecting it from

inactivation either during immobilization or

reuse

Whole Cells

52 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 53: Biosensors - ISOCS

• Permeabilized using physical, chemical,

and enzymatic approaches.

• The most common technique is the use of

organic solvent or detergents.

• Removes some of the lipids from the cell

membranes, thus creating minute pores,

allowing the free diffusion of small

molecular weight substrates/products

across the cell membrane

Cell Permeabilisation

53 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 54: Biosensors - ISOCS

• Permeabilization renders the cell

nonviable and also empties it of most of

the small molecular weight cofactors

• Minimises the unwanted side reactions

• Economic source of intracellular enzymes

for simple biosensor applications that do

not require cofactor regeneration or

metabolic respiration, such as glucose

oxidase, amino acid oxidase, and urease

54 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 55: Biosensors - ISOCS

• Immobilized viable cells in the fabrication

of biosensors

• Substrate assimilation capacity

• Respiratory metabolic activity - BOD

• Inhibition of microbial respiration -

Pollutants

• Genetic modification - organic and

pesticide contamination – luciferase

reaction

Immobilised viable cells

55 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 56: Biosensors - ISOCS

Source: http://www.whatsnextnetwork.com/technology/media/cell_adhesion.jpg

Whole Cell Sensors

56 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 57: Biosensors - ISOCS

Whole Cell Sensors

Harness normal genetic processes

May detect dozens of pathogens

Modifiable/customizable

Reports bioavailability

Temperature/pH sensitive

Short shelf-life

57 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 58: Biosensors - ISOCS

• Adsorption

• Entrapment

• Covalent binding

• Crosslinking

• Combination of all these techniques

Immobilisation techniques

58 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 59: Biosensors - ISOCS

• Passive trapping of cells into the pores of

membranes made up of cellulose or other

synthetic materials

• Retain the cells or enzymes in the close

proximity of the transducer surface using

dialysis membrane

• Entrapment in a variety of synthetic or

natural polymeric gels

59 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 60: Biosensors - ISOCS

Polyvinyl alcohol (PVA) is one of the most

widely studied polymers, because it can

form membranes, fibres, and so on.

Enzymes have been immobilized in these

membranes either by entrapment,

covalent binding, crosslinking, freezing

and thawing, γ-irradiation,

photocrosslinking, or entrapment followed

by crosslinking

Poly Vinyl Alcohol

60 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 61: Biosensors - ISOCS

• Membrane discs of PVA-glutaraldehyde

containing free carbonyl groups have been

prepared for the binding of biomaterials

through their free amine groups

• Modified polyvinyl chloride membranes

polyacrylonitrile membranes and albumin-

poly(ethylene glycol) hydrogel

• Albumin-poly(ethylene glycol) hydrogels –

highly biocompatible

61 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 62: Biosensors - ISOCS

• Biospecific reversible immobilization using

lectins or hydrophobic surfaces can also

be used for the introduction of biologic

catalysts into analytical systems

• Recently, novel sol-gel synthetic

techniques have been developed to

immobilize biologically active molecules in

stable, optically transparent, porous silica

glass matrix

62 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 63: Biosensors - ISOCS

• Nonviable cell preparations have been

immobilized in radiation polymerized

acrylamide.

• The major advantage of γ-ray

polymerization against chemical

polymerization is that the polymerization

can be carried out even under frozen

conditions.

63 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 64: Biosensors - ISOCS

Biosensor Based on Immobilized Indicator Cells

A B-lymphocyte cell line was encapsulated in a collagen gel matrix (Banerjee et al., 2007).

This assay measures alkaline phosphatase or lactate dehydrogenase released by cells infected with pathogens or exposed to different toxins.

The system was tested using different strains of Listeria, listeriolysin O, and enterotoxins from Bacillus species.

Banerjee et al., Laboratory Investigation

(2007) 1-11

64 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 65: Biosensors - ISOCS

The cryo-SEM images of pathogen- or toxin-induced damage of

Ped-2E9 cells in collagen gel matrix.

FROM:

A novel and simple cell-based detection system with a collagen-encapsulated B-lymphocyte cell line as a

biosensor for rapid detection of pathogens and toxins (2007)

Pratik Banerjee, Dominik Lenz, Joseph Paul Robinson, Jenna L Rickus and Arun K Bhunia

Laboratory Investigation (2007) 1-11

65 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 66: Biosensors - ISOCS

Cell Viability – acridine orange

66 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 67: Biosensors - ISOCS

67 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 68: Biosensors - ISOCS

68 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 69: Biosensors - ISOCS

69 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 70: Biosensors - ISOCS

70 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 71: Biosensors - ISOCS

71 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 72: Biosensors - ISOCS

72 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 73: Biosensors - ISOCS

73

Neuron Based Biosensors

Definition: a biosensor that uses living neural cells to detect substance of interest

Why neuron based biosensors?

Key advantage: a single neuron-based sensor can potentially detect a vast number of chemical and biological agents

A healthy neuron generates voltage pulses (“action potentals”) spontaneously on the membrane of the axon.

Changes in environment (presence of chemicals or biological agents) modulate the neuron’s electrical activity.

Neuron exhibits a unique electrical response to particular agents

Image source: http://www.cic-caracas.org/departments/science/Topic11.php

Neuron Based Biosensor

73 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 74: Biosensors - ISOCS

74

Review: Detection

Favored method of detection is Microelectrode Arrays (MEAs).

Electrodes fabricated on surface of device

Monitor signal externally; doesn’t damage cells

Neural signals typically in the range of 100s of uVpp

Many working neuron-based sensors utilize MEAs

Much research focused on improving control of neural growth on MEAs. Photo Source: Nam, et al., “Gold-coated microelectrode

array with thiol linked self-assembled monolayers for

engineering neuronal cultures.” IEEE Transactions on

Biomedical Engineering, 51:1 (2004) 158-165.

74 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 75: Biosensors - ISOCS

75

Analysis of Neural Response

MEA recordings of neural

activity

a) Spontaneous activity

b) Cyclothiazide

c) MK-801

d) NBQX

Source: Chiappalone et a. “Networks of neurons coupled to microelectrode

arrays: a neuronal sensory system for pharmacological applications”

Biosensor and Bioelectronics, 18:5-6 (2003), 627-634

Time domain analysis Characterize response for various

substances • Amplitude

• Duration of burst

• Time interval between bursts

75 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 76: Biosensors - ISOCS

76

Analysis of Neural Response

Neural response to ethanol

Neural response to hydrogen peroxide

Source: Prasad et al. “Neurons as sensors: individual and cascaded

chemical sensing” Biosensor and Bioelectronics, 19:12 (2004), 1599-1610

Frequency domain analysis

Example: Prasad et al. 2004: Examine and

characterize frequency components of neural

response for particular substances

76 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 77: Biosensors - ISOCS

77

Challenges

Controlling interaction of living neuron to device.

Ideal of 1:1 association of neurons to electrodes is

difficult to achieve

Affects signal-to-noise ratio

Affects reproducibility and repeatability of response

Long term maintenance of cells in vitro

Stability of device (corrosion, biofouling, etc)

77 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 78: Biosensors - ISOCS

78

Review: Cell Patterning Techniques

Physical Immobilization

Topographical Patterning

Chemical Patterning

Dielectrophoresis

Source: James, et al., 2004.

Source: Craighead, et al., 1998.

Source: Tooker, et al., 2004.

Source: Prasad, et al., 2003.

Goal is to enhance detectibility of action potentials by patterning neurons over electrodes

78 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 79: Biosensors - ISOCS

79

Cell Patterning Using SAMs SAMs form a single layer of molecules on a

substrate.

Advantages: Creates a biocompatible membrane like

microenvironment

• Supporting structure for growth

• Directs growth

Relatively easy to create

Long term stability

Customizable

Many Types of SAMs

Recent research has focused on using thiols on gold substrates

Self Assembled Monolayers

79 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 80: Biosensors - ISOCS

80

Thiol-based SAMs

Structure:

Alkane chain, typically with

10-20 methylene units

Head group with a strong

preferential adsorption to

the substrate used. Eg:

Thiol (-SH) head groups

and Au(111) substrates

Tail group gives the SAM

its functionality

Source: “Self Assembled Monolayers”

http://www.ifm.liu.se/applphys/ftir/sams.html

80 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 81: Biosensors - ISOCS

81

Thiols on Au(111)

Thiol head group bonds

to the threefold hollow

site on gold surface.

Van der Waals forces

between alkane chain

causes them to lie at

30 degree angle Source: “Self Assembled Monolayers”

http://www.ifm.liu.se/applphys/ftir/sams.html

Commonly used SAMs:

MUA: 11-mercaptoundecanoic acid

11-AUT: 11-amino-1-undecanethiol

81 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 82: Biosensors - ISOCS

82

Nam et al. 2004

Contribution: Coated microelectrode arrays with gold in order to use alkanethiol-based SAM techniques

Techninque:

Coat MEAs with 50-80A of gold

Immerse in MUA solution for 2 hours to create SAM

Expose SAM to other compounds to produce layer of NHS esters

Use uCP to apply poly-D-lysine. Stable PDL layer created by covalent linking to SAM layer

Unstamped areas covered with chemical that inhibits cell growth

Source: Nam, et al., “Gold-coated microelectrode array

with thiol linked self-assembled monolayers for engineering

neuronal cultures.” IEEE Transactions on Biomedical

Engineering, 51:1 (2004) 158-165.

82 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 83: Biosensors - ISOCS

83

Nam et al. 2004

Results: Demonstrated cell viability on PDL

linked gold surface

Good resolution stamped 100 x 100um grid pattern of 10um line width

Cells complied to pattern for > 2 weeks

Recording of spontaneous neural activity to verify cell activity.

Enhanced amplitudes up to 500uVpp (100-200uVpp typical)

Gold MEAs were not reusable Source: Nam, et al., “Gold-coated microelectrode array

with thiol linked self-assembled monolayers for engineering

neuronal cultures.” IEEE Transactions on Biomedical

Engineering, 51:1 (2004) 158-165.

83 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 84: Biosensors - ISOCS

84

Nam et al. 2006

Updated process different SAM 3-glycidoxypropyl

trimethoxysilane (3-GPS)

uCP for protein pattern stamping

Results Neurons complied to patterns for 2-3

weeks

Spontaneous neural activity recorded:

• Note SAM increased impedance by factor of 2-3

• Mean SNR of 6.5 at 2 weeks

• mean amplitude of extracellular spikes was 25uVpp at 7 DIV and 50uVpp at 20-24 DIV.

• Background noise 2.9uVpp

400ms

Source: Nam et al. “Epoxy-silane linking of biomolecules is simple

and effective for patterning neuronal cultures.” Biosensors and

Bioelectronics 22 (2006) 589–597

84 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 85: Biosensors - ISOCS

85

Palyvoda et al. 2007

Technique:

Create gold electrodes

Immerse in 11-AUT solution to create SAM

Studied effect of pad size on neural guidance

Contribution: used SAM to support and guide neural growth directly

No intermediate protien layer, e.g. polylysine (which is difficult to pattern, nonphysiological, toxic under some conditions)

Image of neurons on 50x50um SAM coated gold electrode

Source: Palyvoda, et. al., “Culturing neuron cells on

electrode with self-assembly monolayer”

Biosensor and Bioelectronics, 22 (2007) 2346-2350.

85 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 86: Biosensors - ISOCS

86

Palyvoda et al. 2007

Results 50x50um pad size comes

close to single neuron immobilization, with error.

Source: Palyvoda, et. al., “Culturing neuron cells on

electrode with self-assembly monolayer”

Biosensor and Bioelectronics, 22 (2007) 2346-2350.

86 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 87: Biosensors - ISOCS

• Example Applications

• Cell Based Biosensors

Applications

87 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 88: Biosensors - ISOCS

Potential Applications

• Clinical diagnostics

• Food and agricultural processes

• Environmental (air, soil, and water) monitoring

• Detection of warfare agents.

88

Page 89: Biosensors - ISOCS

Food Analysis Study of biomolecules and their interaction Drug Development Crime detection Medical diagnosis (both clinical and laboratory use)

Environmental field monitoring Quality control Industrial Process Control Detection systems for biological warfare agents Manufacturing of pharmaceuticals and replacement organs

Application of Biosensors

89 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 90: Biosensors - ISOCS

Cell based biosensors

90

Cytometers for

Red blood cells

Platelets

White blood cells

Bacteria

SAW resonators

Cantilever beams

Page 91: Biosensors - ISOCS

Quorum Sensing

Secretion of Signalling molecules that are

autoinducers

Other cells have receptors that bind to the

molecule

The result is gene transcription is mediated

The more cells – the more autoinducer –

more gene transcription – up regulation of

genes

91 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 92: Biosensors - ISOCS

Quorum sensing

Credit: Bonnie Bassler, et al., Princeton U. 92

Page 93: Biosensors - ISOCS

Quorum Sensing Circuit

93

Page 94: Biosensors - ISOCS

Microprocessor

Chemical and

Biochemical

Processes

Cell-based Biosensor

Electro-

mechanical

Processes

Synthetic biology?

94

Page 95: Biosensors - ISOCS

Band dector (of chemical concentrations)

Credit: S. Basu, et al., Princeton U.

95

Page 96: Biosensors - ISOCS

Credit: M.L. Simpson, M. L., et al. , Oak Ridge National Lab

Compounds Detected

• Napthalene

• TCE

• Tolulene

• Benzene

• Xylene

Sensitivity: <50 ppb

Cells glow when detect toxins. Luminescence is detected by an integrated photosensor circuit.

Cells

Photodetector Chip

Biochip for toxin detection

96

Page 97: Biosensors - ISOCS

97

Ca++

Analytes

Surface

Immunoglobulin

Signaling cascade

Aequorin

Data analysis

Cell-Based

Biosensor

Krishna Persaud, SCEAS, The University of Manchester, UK

Page 98: Biosensors - ISOCS

Current Biosensor Technologies

• Bioreporters

o using cells or other organisms to detect relevant stimuli utilizes inherently rapid response times of cellular signaling easy to use reporting system needs to be established in advance organisms need to be maintained in their testing environment

Bomb-Sniffing Plants Colorado State University

98

Page 99: Biosensors - ISOCS

Phenol Detection

DmpR (Dimethyl phenol regulatory protein),

an NtrC-like regulatory protein for the

phenol degradation of Pseudomonas sp.

strain CF600

Strategy - modifying the phenol detection

capacity of DmpR by using mutagenic

PCR

99 Krishna Persaud, SCEAS, The University of Manchester, UK

Page 100: Biosensors - ISOCS

Schematic illustration of sensing principle of whole cell biosensor (pRLuc42R) to detect phenol at

glance.

Gupta S, Saxena M, Saini N, Mahmooduzzafar , et al. (2012) An Effective Strategy for a Whole-Cell Biosensor Based on Putative

Effector Interaction Site of the Regulatory DmpR Protein. PLoS ONE 7(8): e43527. doi:10.1371/journal.pone.0043527

http://www.plosone.org/article/info:doi/10.1371/journal.pone.0043527

Page 101: Biosensors - ISOCS

Linearity curve is presented as phenol concentration v/s NL value of luminescence.

Gupta S, Saxena M, Saini N, Mahmooduzzafar , et al. (2012) An Effective Strategy for a Whole-Cell Biosensor Based on Putative

Effector Interaction Site of the Regulatory DmpR Protein. PLoS ONE 7(8): e43527. doi:10.1371/journal.pone.0043527

http://www.plosone.org/article/info:doi/10.1371/journal.pone.0043527

Page 102: Biosensors - ISOCS

• Many types of biosensors

• Many different transduction principles

• Future is bright

• Cell Based Biosensors – Can be widely used in environmental testing

• Genetically modified organisms can be tailored to detect specific analytes

• Both DEAD Cells as well as LIVE cells can be used as viable biosensors – dependent on the application

Conclusions