DR. SIDHARTH YADAVDEPT. OF ORTHOPAEDICSNKPSIMS
![Page 2: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/2.jpg)
Carnochan (1840) was the first surgeon, who thought that hip joint could also be replaced artificially.
A wooden block was installed between the damaged ends of a hip joint in New York by Carnochan.
Dr Marius N Smith-Petersen introduced the mould arthroplasty (1925).
![Page 3: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/3.jpg)
The original design was ball-shaped hollow hemisphere of glass which could fit over the ball of the hip joint.
The objective was to stimulate cartilage regeneration on both sides of the moulded glass joint.
He used a reactive synovial like membrane.
![Page 4: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/4.jpg)
The glass could not withstand the stresses of walking and quickly failed.
This led to use of other materials, such as Viscaloid (a celluloid derivative, 1925), Pyrex (1933), Bakelite (1939), and later that year, an alloy of Cobalt-Chromium is called Vitallium (1936).
This Vitallium material was very strong and resistant to corrosion, and continued to be employed in various prostheses since that time.
![Page 5: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/5.jpg)
In 1938 PHILIP WILIE performed the first total joint replacement. He used stainless steel.
In 1950, SEVEN KIAER used dental scrylic cement in orthopaedics by using it to bind a plastic proesthesis to bone.
In 1951, McKEE & WATESON- FARRAR used thompson type femoral component & a metal acetabular cup. Both were of chrome cobalt alloy.
![Page 6: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/6.jpg)
In 1962 UHMWPE was introduced.
In 1967 MULLER introduced FIRST GENERATION METAL ON METAL ARTHROPLASTY.
In 1970, Pierre Boutin was the first to introduce a ceramic-on-ceramic hip arthroplasty
![Page 7: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/7.jpg)
• Ball and socket joint
• Femoral head or ball is at the end of thighbone or femur.
• Socket or acetabulum holds the ball.
![Page 8: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/8.jpg)
Acetabular component - consists of two components
Cup - usually made of titanium
Liner - can be medical grade plastic, metal or ceramic
Cup
Liner
![Page 9: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/9.jpg)
The femoral component (stem and neck portion), is made of metal.
Femoral head is made either of metal or ceramic.
Stem is shaped to fit into the bone and support the new joint.
Neck
Stem
Head
![Page 10: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/10.jpg)
• Range of Motion• Stability• Wear resistance• Lifestyle• Gender• Age• Weight• Severity of Osteoarthritis disease
![Page 11: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/11.jpg)
Patients may have: Young and active lifestyle Rheumatoid arthritis Osteoarthritis Post-traumatic arthritis Collagen disorders Avascular necrosis Non-union of femoral fractures
![Page 12: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/12.jpg)
Inappropriate for patients who recently or currently have: Joint sepsis (infection) Insufficient bone stock Marked atrophy (weakness) Upper femur deformity Skeletal immaturity
![Page 13: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/13.jpg)
The bearing is the union of the ball and socket.
Bearings are made from a variety of commercially available materials.
![Page 14: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/14.jpg)
• Metal-on-Polyethylene (poly)– “traditional” bearing– Poly cup with a metal ball
• Metal-on-Metal– All metal ball and cup
• Ceramic-on-Polyethylene (poly)– Poly cup with a ceramic ball
• Ceramic-on-Ceramic– Ceramic cup and ball
![Page 15: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/15.jpg)
Poly acts as a cushion for the joint
Shell made of titanium alloy
Cup made of medical grade cross-linked polyethylene
Long clinical history since 1963
![Page 16: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/16.jpg)
All bearing surfaces wear with time
Cross Linked Polyethylene exhibits 86% wear reduction vs. traditional polyethylene in mechanical tests
80
50
10
20
30
40
50
60
70
80
90
100
Wea
r R
ate
Wear Rates: Crosslinked vs. Traditional PolyWear Rates: Crosslinked vs. Traditional Poly
Traditional Polyethylene
Cross Linked Polyethylene from
DePuy
![Page 17: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/17.jpg)
Ball and liner made of high-standard, surgical-grade cobalt chromium alloys
Available since 1960s
![Page 18: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/18.jpg)
Wear rates up to 100 times less than metal-on-poly bearings in mechanical tests
80
0.10
10
20
30
40
50
60
70
80
90
100
Wea
r R
ate
(mm
3/m
illio
n c
ycle
s)
Traditional Polyethylene Ultamet Metal on Metal Articulation from DePuy
Wear Rate: Metal vs. PolyethyleneWear Rate: Metal vs. Polyethylene
![Page 19: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/19.jpg)
Increased ion concentration in patients.
Cobalt is associated with various cardiac ,neurological & endocrine symptoms.
Long term exposure can cause cancer.
Hypersensitivity
One patient has been found with chromium & cobalt in CSF.
![Page 20: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/20.jpg)
Alumina ceramics was introduced as a bearing surface in 1971 as an alternative to metal on poly.
Since then many advances has been made in ceramics :-
- Alumina that was introduced in 1970 (1st generation ) had low density & a very coarse micro structure.
![Page 21: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/21.jpg)
- SECOND GENERATION (1980’s) :- had reduced microstructure & grain size.
- THIRD GENERATION (1990’s) :- had improved mechenical strength , more reduced microstructure grain size & manufactured by HOT ISOSTATIC PRESSURING.
- FOURTH GENERATION ( 2000’s) :- new alumina matrix composite.
![Page 22: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/22.jpg)
Hardness Hydrophilic Inertness Resistant to scratch High biocompatibility
These characteristic’s make less wear than metal.
![Page 23: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/23.jpg)
At equal partical levels alumina is less toxic then metal.
Types :- - Alumina (Al2O3) - Zirconia ( ZrO2 )
fracture toughness of ZIRCONIA is twice then alumina.
A new compound matrix composite was developed & early results have shown hopeful in both COC & COP in laboratory.
![Page 24: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/24.jpg)
Cup made of cross-linked poly
Ball made of Delta Ceramic Composite
zirconia aluminum ceramic
![Page 25: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/25.jpg)
Highly Cross Linked Polyethylene exhibits 86% wear reduction over standard poly liners in mechanical test
Delta Ceramic Ball resistant to chipping or cracking 97% tougher than traditional
alumina ceramics in mechanical tests
Smoother femoral ball decreases wear potential
4.3
8.5
0
1
2
3
4
5
6
7
8
9
Delta Conventional Alumina Ceramics
Fracture toughness: Delta
vs. Alumina
![Page 26: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/26.jpg)
Ball and cup made of alumina ceramic
Shorter clinical history
Not toxic to the body
![Page 27: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/27.jpg)
Ceramic femoral heads offer:
Excellent compatibility within the body
Good mechanical performance
Very hard and scratch-resistant
Chemical and temperature stability
![Page 28: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/28.jpg)
100-200 times less wear rate than other materials in mechanical tests
Resistance to deformation and surface scratching
Less particle debris around joint available to get into the body
![Page 29: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/29.jpg)
OONISHI & COWORKER’S reported head penetration for ceramic & metal:-
- Alumina :- 0.1 mm/yr.- Metal :- 0.25mm/yr.
SCHULLER MART reported :-- 0.03mm/yr for alumina- 0.10 mm/yr for metal ( after 9-11 yr of
follow up)
![Page 30: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/30.jpg)
Brittle Low fracture toughness Squeaking noice heard with activity.
![Page 31: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/31.jpg)
Received 15 November 2005; revised 12 July 2006; accepted 13 July 2006Published online 13 October 2006 in Wiley InterScience (www.interscience.wiley.com). DOI: 10.1002/jbm.b.30691
![Page 32: Bearing surfaces](https://reader035.fdocuments.net/reader035/viewer/2022081419/556c936fd8b42a7a738b5005/html5/thumbnails/32.jpg)