Beam Instrumentation & Beam Diagnostics CAS 2003 Rhodri Jones [Hermann Schmickler] (CERN)

Transcript of Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall...

Page 1: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam Instrumentation &

Beam Diagnostics

CAS 2003

Rhodri Jones [Hermann Schmickler] (CERN)

Page 2: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Write-up and Hand-Outs• The two lectures on beam instrumentation and

diagnostics contain a lot of visual information (movies etc) which are not easy to reproduce in hand-outs.

• On the internet a similar lecture from CAS2002 can be downloaded from the url:http://sl-div.web.cern.ch/sl-div-bi/CAS%20/lecture/

• In about 2 weeks the lectures of 2003 will be available from the same url.

Page 3: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

A Starter

A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation of particle beams with the precision required todiagnose, tune, operate and improve the accelerators and associated transfer lines.

This means that Beam Instrumentation combines the disciplines ofaccelerator physics, mechanical engineering, electronic engineering and software engineering.

In Short: One of the most fascinating fields of work I can imagine

Page 4: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Starter Example:Setup for Head-tail observations

StraightStriplineCoupler

Beam PipeBeam

Hybrid

VMEAcquisitionvia GPIB

Sum

Difference

Bunch SynchronousTrigger

GPIB linkUNIX

User Interface

Fast (2GS/s per channel)Digital Oscilloscope

SPS Tunnel

Page 5: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

…snapshot of movies to follow...

time

Time evolution ofhead-tail phase-shift

I ∆xSimulated betatronoscillation of bunch

2 cursors marking time of head- and tail sample

Page 6: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Head-Tail Principle

Page 7: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Head-Tail Principle

Page 8: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Deducible Beam Parameter: Chromaticity(…more of this tomorrow in the diagnostics part)

Qs-1 = 230 turns

Page 9: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Outline for Today• Focus on “What and How we Measure”

& the technologies involved

• Introduction→ What do we mean by “Beam Instrumentation”→ What instruments are involved

• Beam Instrumentation Selection→ Beam Position Measurement→ Beam Loss Measurement→ Beam Intensity Measurement→ Luminosity Measurement

Page 10: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Introduction

→ The “eyes” of the machine operators • i.e the instruments that observe beam behaviour

• What beam parameters do we measure?→ Beam Position

• Horizontal and vertical all around the ring• Corrected using orbit corrector magnets (dipoles)

→ Beam Loss all around the ring• Especially important for superconducting machines

→ Beam Intensity (& lifetime measurement for a collider)• Circulating current and bunch-by-bunch charge

→ Beam size• Transverse and longitudinal distribution

→ Collision rate / Luminosity (for colliders)

• What do we mean by beam instrumentation?

Page 11: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

More Measurements

• Machine Tunes

• Machine Chromaticities

QFQF QF

QDQD

SFSF

SF SD SD

Characteristic Frequencyof the Magnet Lattice

Produced by the strength of theQuadrupole magnets

Achromatic incident light

Lens

Focal length isenergy dependent

Optics Analogy: Spread in the Machine Tune due to Particle Energy SpreadControlled by Sextupole magnets

Page 12: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Typical Instruments• Beam Position

→ electrostatic or electromagnetic pick-ups and related electronics• Beam Loss

→ ionisation chambers or pin diodes• Beam Intensity

→ beam current transformers• Beam Size (transverse)

in diagnostics section of tomorrow• Beam Size (longitudinal)

→ RF pick-ups or synchrotron light• Luminosity

→ ionisation chambers or semiconductors• Machine Tunes and Chromacitities

in diagnostics section of tomorrow

Page 13: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

LHC BPM System - General Layout

100 Warm, Button Electrode BPMsButtons Recuperated from LEP

24 Directional Couplers& 44 Enlarged ButtonBPMs for theInteraction Regions

~ 24 Special BPMs forTransverse Damper, Tune

& Chromaticity Measurements

922 Button Electrode BPMs (24mm)for the

Main Arcs & Dispersion Suppressors

Total of 1158 BPMs for the LHCand its Transfer Lines

44 Warm BPMs -24mm Button Electrodes

Page 14: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Arc BPM - SSS Layout

Page 15: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

LHC Arc Type BPM (String 2)

Page 16: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Interaction Region BPMs

StriplineElectrode

Liquid HeliumCapillaries

Q2 Coupler

• Directional stripline couplers→ Outputs signals only from the upstream port→ Can distinguish between counter-rotating beams

Page 17: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Electromagnetic (Directional) coupler

• Is a transmission line (strip line) which couples to the transverse electromagnetic (TEM) beam field

Zt ∞= 60 ln[(r+h)/r] ≡ Z0*[a/2π(r+h)]

• Z0 is the characteristic impedance• a, r, h, l are the mechanical dimensions• t = l/c is the propagation time in the coupler

• Two termination ports→ Upstream: usually used to acquire signal.

• Same signal seen whether Downstream port is open, shorted or terminated on Z0

→ Downstream: 3 cases• Upstream terminated on Z0 ⇒ no signal• Downstream open ⇒ delayed signal• Upstream short circuit ⇒ delayed & reversed signal

r

a

h

UpstreamDownstream

Beam

Page 18: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Coupler frequency & time response

→ Sinusodal amplitude response•Maximum signal for f = 1/4*t •Zero signal for f = 1/2*t

→ Time domain:•Bipolar pulse

-40

-35

-30

-25

-20

-15

-10

-5

0

0 1000 2000 3000 4000MHzdB

Coupler 15 cm long

Pulse width = 1.6 ns / I = 1e9p/b

-300

-200

-100

0

100

200

300

0 2 4 6 8 10ns

mU

30 cm 50 cm75 cm

Pulse width = 4.8 ns / I = 1e9p/b

-80

-60

-40

-20

0

20

40

60

80

0 5 10 15 20ns

mU

15 cm 30 cm 50 cm

Page 19: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Electrostatic PU (Button)

dr

Referenceorbit

Ib

Ce

RlZt

V to V

• Variant of electrostatic PU• Low cost ⇒ most popular

Zt∞ ≡ d / 2πrCe

Low frequency cut-offT = RlCe (few hundreds MHz)

Page 20: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Button Frequency & Time Response

Button; Ce = 8 pF / Rl = 50 Ohm

-35

-30

-25

-20

-15

-10

-5

010 100 1000MHzdB

N = 1:1N = 2:1N = 4:1

Button; Ce = 8 pF / Zt = 1.1 Ω / Q = 1e9p/b

-60

-40

-20

0

20

40

60

80

0 5 10 15ns

mU

W = 5.0 nsW = 2.0 nsW = 1.0 ns

• Frequency domain:→ Impedance transformers

improve the low frequency levels at the expense of the high frequency

• Time domain:→ Differentiated pulse→ Amplitude is exponentially

bunch length dependent

Page 21: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

What does a real (LHC) electrostatic button monitor look like?

Page 22: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Processing system families

From G.Vismara: BIW 2000; highly recommended

Legend:

/ Single channel

Wide Band

Narrow band

NormalizerProcessor

ActiveCircuitry

Heterodyne POS = (A-B) SynchronousDetection

AGCon Σ

MPX

PassiveNormaliz.

POS = [log(A/B)] = [log(A)-log(B)]

DifferentialAmplifier

Logarithm. Amplifiers

IndividualTreatment

Limiter,∆t to Ampl.

Amplitudeto Time POS = [A/B]

POS = [ATN(A/B)] Amplitudeto Phase

.Limiter,φ to Ampl.

POS = ∆ / ΣHeterodyneHybrid∆ / Σ

HomodyneDetection

POS = ∆ / Σ or= (A-B)/(A+B)

Sample,Track,Integr. & Hold

Switch. gainAmplifier

ElectrodesA, B

Page 23: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Criteria for Electronics Choice so called “Processor Electronics”

• Accuracy: mechanical and electromagnetic errors, electronic components

• Resolution• Stability over time• Sensitivity and Dynamic Range• Acquisition (Measurement) Time• Linearity (over aperture & intensity)• Radiation tolerance

Page 24: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

LINEARITY Comparison

Transfer Function

-1

-0.5

0

0.5

1

-1 -0.5 0 0.5 1

Normalized Position (U)

Computed Position (U)

∆/ΣAtn(a/b)loga-logb

Page 25: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Simple Example:Sampling of Individual Detector Signals

LPFilter

A

B

LPFilter

ADC

ADC

S&H

S&H

Position=

(A-B) / (A+B)

A

B

Dynamic range!

Page 26: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

A Real System:LHC Amplitude to Time Normaliser Schematics

INPUT OUTPUT

A A

BB

T1 = 1.5 ns

T1 = 1.5 ns

Page 27: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Wide Band Time Normaliser

-2.5

-2.0

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

Time [ns]

Ampl

itude

A

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

Ampl

itude

B

1.5ns

A

B

B + 1.5ns

A + (B + 1.5ns)A B

Beam

Page 28: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Wide Band Time Normaliser

-2.5

-2.0

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

Time [ns]

Ampl

itude

A

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

Ampl

itude

B

1.5ns

A

B

A + 1.5ns

B + (A + 1.5ns)

A + (B + 1.5ns)A B

∆t depends on position

Page 29: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Wide Band Time Normaliser

-2.5

-2.0

-1.5

-1.0

-0.5

0.0

0.5

1.0

1.5

Time [ns]

Ampl

itude

A

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

Ampl

itude

B

A B

A+(B+1.5ns)

B+(A+1.5ns)+10ns

System outputInterval = 10 ± 1.5ns

Page 30: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Amplitude to Time Normaliser Evaluation

Advantages Limitations• Reduced number of

channels (x2)• No need for gain selection• Input dynamic > 50 dB• Signal dynamic independent

on the number of bunches• ~10 dB compression of the

position dynamic (recombination)

• Auto-trigger• Reduced N° of bits at

equivalent resolution (Normalisation)

• Mainly reserved for bunched beams

• Tight time adjustment

• Propagation delay stability and switching time uncertainty are the limiting performance factors

• No Intensity information

Page 31: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

LHC Beam Position System Layout

CAL. and TESTGENERATOR ADC

LOWPASS

FILTER

CALIBRATOR

PICK-UP

50 ΩCABLE

ΣIntensity

Measurement

Trigger

AutoTrigger

50 ΩCABLE

LOWPASS

FILTERCALIBRATOR

NORMALISER INTEGRATOROPTICAL

LINK

TUNNEL SURFACE

DAB

Page 32: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

WBTN - Linearity v Intensity

-5%

-4%

-3%

-2%

-1%

0%

1%

2%

3%

4%

5%

1E+08 1E+09 1E+10 1E+11 1E+12Number of Charges per Bunch

Perc

enta

ge E

rror w

.r.t.

Hal

f Rad

ius

[%]

Linearity - High SensitivityLinearity - Low SensitivityNoise - High SensitivityNoise - Low sensitivity

Pilot Nominal Ultimate

For LHC Arc BPMs 1% ~ 130µm

Page 33: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

WBTN - Linearity v Position

-8%

-6%

-4%

-2%

0%

2%

4%

6%

8%

-1 -0.5 0 0.5 1Normalised Position

Perc

enta

ge E

rror w

.r.t.

Hal

f Rad

ius

[%]

Calibration PointMeasured ValueCalibrated ValueLinearised Value

For LHC Arc BPMs 1% ~ 130µm

Page 34: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

WBTN - Radiation Issues

• The Front-end Electronics for the Arc BPMs will be located under the main quadrupoles→ can expect to see a dose of some 12Gy/year

• Tests in the SPS-TCC2 area during 2000 showed that use of DIGITAL components in the tunnel should be avoided→ Most memories and FPGA’s too easily corrupted→ Qualification of components long & difficult

Page 35: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

WBTN - Radiation Issues

• In 2001 : Fibre-Optic Link added to LHC BPM system→ only the minimum of analogue electronics kept in tunnel→ all sensitive digital electronics located on the surface→ allows easy access to most of the acquisition system

• Cost of large scale fibre-optic installation compensated by→ elimination of 13km of expensive low loss coaxial cable→ reduction in number of acquisition crates→ no bunch synchronous timing required in the tunnel

Page 36: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

WBTN - Radiation Test Results2001 Test results of the

very front-end WBTN cardwith Fibre-Optic Link

Initial performance

6ps rms jitter

After 650Gy no significantdeterioration in the performanceis visible

6.5ps rmsjitter

Page 37: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The LHC BPM Acquisition System

70MHz Low Pass FiltersSupplied by TRIUMF (Canada)

Single-Mode Fibre-Optic Link

VME basedDigital Acquisition Board

TRIUMF (Canada)(2 x 12bit 40MHz Acq)

WBTN Mezzanine Card(10bit digitisation at 40MHz)

1310nm Diode Laser Transmitter

Very Front-End WBTN Card

Tunnel

Surface

Page 38: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Operational Prototype Results in 2001System extensively used in SPS for electron cloud & instability studies.

Page 39: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Typical Instruments• Beam Position

→ electrostatic or electromagnetic pick-ups and related electronics• Beam Loss

→ ionisation chambers or pin diodes• Beam Intensity

→ beam current transformers• Beam Size (transverse)

in diagnostics section of tomorrow• Beam Size (longitudinal)

→ RF pick-ups or synchrotron light• Luminosity

→ ionisation chambers or semiconductors• Machine Tunes and Chromacitities

in diagnostics section of tomorrow

Page 40: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The LHC Beam Loss SystemRole of the BLM system:

1. Protect the LHC from damage

2. Dump the beam to avoid magnet quenches

3. Diagnostic tool to improve the performance of the LHC

• Acquisition requirements:→ Calculation of quench level equivalent chamber signal

– Electric currents from 600 pA to 300 µA→ A dump should be requested at 50% of the quench level

– i.e. from 300 pA to 150 µA→ Extend dynamic range for sufficient sensitivity at low losses

– Measuring current from 60 pA to 300 µA→ Arc BLM acquisition rate not faster than one turn (89 µs)

– Fastest total loss is ~ 6 turns & will be detected by special BLMs.

Page 41: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Structure of the BLM Readout Chain

→ transforms particle losses into an electric current→ 6 per quadrupole (3 for each LHC ring) ⇒ ~3000 monitors

• Analogue Front-End→ measures current and transmits data from Tunnel ⇒ Surface

• Dump Controller→ processes data and interfaces to the beam interlock system

• Ionisation Chamber

Analog Signals

max. 400 m

Digital Signals

max. 2 kmAnalog

Front end

DumpController

IonizationChamber

SurfaceTunnel

Page 42: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Quench Level EquivalentChamber Current

1.E-11

1.E-10

1.E-09

1.E-08

1.E-07

1.E-06

1.E-05

1.E-04

1.E-03

1.E-05 1.E-04 1.E-03 1.E-02 1.E-01 1.E+00 1.E+01 1.E+02 1.E+03loss duration [s]

Cha

mbe

r cur

rent

[A]

450 GeV max450 GeV min7 TeV max7 TeV min

One turn

300 µA

600 pA

60 pA

Page 43: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Charge-Balanced Converter

Vtr

V-

C

One-shotDT

Tresholdcomparator

Integrator

Referencecurrent source fout

Iref

iin(t)

VTr

∆va

∆TT

t

v(t)

va(t)

TIifref

in

∆=

iin(t)iin(t) + Iref

Page 44: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Current-Frequency Characteristics

Frequency versus input current with linear error (with 400 m)

1.E-02

1.E-01

1.E+00

1.E+01

1.E+02

1.E+03

1.E+04

1.E+05

1.E+06

1.E+07

1.E-11 1.E-10 1.E-09 1.E-08 1.E-07 1.E-06 1.E-05 1.E-04 1.E-03input current [A]

outp

ut fr

eque

ncy

[Hz]

-10%

-9%

-8%

-7%

-6%

-5%

-4%

-3%

-2%

-1%

0%

1%

Frequency [Hz]flin_corr [Hz]Lin Error [%]

Page 45: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

The Typical Instruments• Beam Position

→ electrostatic or electromagnetic pick-ups and related electronics• Beam Loss

→ ionisation chambers or pin diodes• Beam Intensity

→ beam current transformers• Beam Size (transverse)

in diagnostics section of tomorrow• Beam Size (longitudinal)

→ RF pick-ups or synchrotron light• Luminosity

→ ionisation chambers or semiconductors• Machine Tunes and Chromacitities

in diagnostics section of tomorrow

Page 46: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Fast Beam Current Transformer

• Installed in the SPS and LHC transfer lines• LHC fast BCT will be a scaled version• Capable of 40MHz bunch by bunch measurement• Dynamic range to cover 5×109 to 1.7 × 1011 cpb

Page 47: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Fast Beam Current Transformer

BEAM

ImageCurrent

Ceramic Gap

80nm Ti Coating⇒20Ω to damp anycavity resonances

1:40 PassiveTransformer

Calibration winding

• 500MHz Bandwidth• Low droop (< 0.2%/µs)

Page 48: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Acquisition Electronics

→ Designed by the Laboratoire de Physique Corpusculaire, Clermont-Ferrand for use in the LHCb Preshower Detector.

→ Uses interleaved, 20MHz integrators and sample & hold circuitry to give 40MHz data.

• Digital Acquisition→ PMC size Mezzanine card developed by CERN & contains

• Fast integrator chip• 12bit, 40MHz ADC• Timing provided by the TTCbi module, part of the Timing,

Trigger & Control system developed for the LHC experiments

Mezzanine sits on the same VME 40MHz Data Acquisition Board developed for the LHC Beam Position System (TRIUMF, Canada)

• Analogue Acquisition based on a fast integrator chip

Page 49: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Acquisition Electronics

Track& Hold

Track& Hold

MUXDifferentialInput fromFast BCT

20MHz Clock Creation

40MHz TTC Input from TTCbi

Fast Integrator Chip

12-bitADC

To VMEDigital

AcquisitionBoard

Page 50: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Acquisition Electronics

FBCT Signal after 200m of Cable

Integrator Output

25ns

Data taken on LHC type beams at the CERN-SPS (2002)

Page 51: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Results from the CERN-SPS (2002)

1 11 21 31 41 51 61 71 81 91 101

111

121

131

141

151

161

171

181

191

201

211

221

231

241

251

261

271

281

291

301

311

321

331

341

351

361

371

381

3911 11 21 31 41 51 61 71 81 91 101

111

121

131

141

151

161

171

181

191

201

211

221

231

241

251

261

271

281

291

301

311

321

331

341

351

361

371

381

3911 11 21 31 41 51 61 71 81 91 101

111

121

131

141

151

161

171

181

191

201

211

221

231

241

251

261

271

281

291

301

311

321

331

341

351

361

371

381

3911 11 21 31 41 51 61 71 81 91 101

111

121

131

141

151

161

171

181

191

201

211

221

231

241

251

261

271

281

291

301

311

321

331

341

351

361

371

381

3911 11 21 31 41 51 61 71 81 91 101

111

121

131

141

151

161

171

181

191

201

211

221

231

241

251

261

271

281

291

301

311

321

331

341

351

361

371

381

391

Bad RF Capture of a single SPS LHC Batch (72 bunches)

Page 52: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

• Beam Position→ electrostatic or electromagnetic pick-ups and related electronics

• Beam Loss→ ionisation chambers or pin diodes

• Beam Intensity→ beam current transformers

• Beam Size (transverse)in diagnostics section of tomorrow

• Beam Size (longitudinal)→ RF pick-ups or synchrotron light

• Luminosity→ ionisation chambers or semiconductors

• Machine Tunes and Chromacititiesin diagnostics section of tomorrow

The Typical Instruments

Page 53: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Luminosity & Beam-Beam Tune Shift

• Luminosity

• Normalized emittance

• Beam-beam tune shift

• To maximize L and minimize the stored energy, increase N to the tune shift limit, choose large M and small β*

)LHC(006.04Nr

4MNfL

N

pbb

*

2*

N

2*

2

rev

≤πε

=ν∆

βσ

γ=ε

πσ=

*

bbrev

MNfLβ

ν∆γ=∴

Page 54: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Nominal locations of the front quadrupole(TAS) and neutral (TAN) absorbers

• The TAS absorbs forward collision products that have escaped the beam tube in front of Q1 (mostly charged pions and photons)

• The TAN absorbs forward neutral collision products (mostly neutrons and photons) and is placed in front of the outer beam separation dipole D2

Page 55: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

TAN Power Deposition (W/kgm)

• Peak power density of 1-10 W.kg-1.m-1 (location of ionization chamber)

• A 3m radiation hard cable will allow electronics to be located in a region withpower density < 10-5 W.kg-1.m-1 (100 Gy/year for nominal operation)

Page 56: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

LHC Luminosity Measurement

• Capable of 40MHz acquisition• Has to withstand high radiation dose: ~108 Gy/year

→ estimated 1018 Neutrons/cm2 over its lifetime (20yrs LHC operation)→ estimated 1016 Protons/cm2 over its lifetime (20yrs LHC operation)

• No maintenance

Requirements:

Candidates:• Ionisation Chambers

→ developed by LBL– Good radiation hardness– Difficult to get working at 40MHz

• CdTe detectors→ developed by CERN in collaboration with LETI (Grenoble)

– Fulfills 40MHz requirement– Not yet proven for the highest levels of radiation

Page 57: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Polycrystalline CdTe Detectors• Experience at CERN with CdTe X-RAY detector

→ running in LEP for beam emittance measurement→ was used up to the end with total dose 1014 Gray

• Advantages→ large number of e- created per MIP (~5 × Diamond)→ very fast response time→ simple construction

Page 58: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

CdTe Detectors – Test Set-upBBiiaass VVoollttaaggee

5500 oohhmm ccaabbllee AAA=== 888000

DigitalOscilloscope

DDeetteeccttoorr

Amplifier Output

AveragedAmplifier Output

Output Histogram

~10ns

Sr90Source

Page 59: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Polycrystalline CdTe Detectors

Comparison of charge collection(normalized for a thickness of 470 microns)

0

10000

20000

30000

40000

50000

60000

70000

0 100 200 300 400 500 600 700Bias voltage

det 253 / 2 GhZ linear Amplifierdet172 / integrating amplifier

Cha

rge

colle

cted

(e- p

er d

etec

ted

Sr90

dec

ay)

Page 60: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Irradiation Test Results• CERN-SPS (2001)

→ Irradiation test up to 1015 neutrons/cm2

T ime response of CdT e samplesbe fore irradia tion: sam ple 172R E F

afte r irrad ia tion sam ple 172N (10^ 15 n /cm 2)

-9

-7

-5

-3

-1

-10 -5 0 5 10 15 20 25 30 35 40 ns

Out

put s

igna

l (A.

U)

172R E F172N

Page 61: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

Irradiation Test Results• Triga type reactor (Ljubljana,Slovenia - 2002)

→ Irradiation steps• 1013 neutrons/cm2

• 1015 neutrons/cm2

• 1016 neutrons/cm2 activation of all set-up– next step 1018 neutrons/cm2 (2003)

Carrier lifetime measurementCdTe detector n:417 (326 microns /400 Volts)

-1.00

-0.50

0.00-10 -5 0 5 10 15 20 25 30

Time in nanosecond

Ampl

itude

of t

he s

igna

l (A.

U)lin

ear s

cale

After 10^16 n/cm2before irradiation

Page 62: Beam Instrumentation Beam Diagnostics - CERN...A Beam Diagnostics and Instrumentation activity shall design, build, maintain and improve the diagnostic instruments that allow the observation

Beam InstrumentationCAS 2003 Rhodri Jones (CERN - AB/BDI)

We have seen a wide variety of technologies and applications

Tomorrow we will see how to use the instruments to run and optimize

accelerators=

Accelerator Diagnostics

http://sl-div.web.cern.ch/sl-div-bi/CAS%20/lecture/