Beam Induced Movement and DDDs

Javier Vargas

Centro Nacional de Biotecnología-CSIC

[email protected]

Page 2: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The past…

Page 3: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The past…

Photographic film

Directly detects electrons

Good signal-to-noise performance (SNR)

Slow and laborious (films must be scanned, not possible automation)

Page 4: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The past…

Photographic film

Directly detects electrons

Good signal-to-noise performance (SNR)

Slow and laborious (films must be scanned, not possible automation)

CCD detectors

Suitable for automation.

Detect photons not electrons, it is needed an scintillator (bad spatial resolution).

Bad signal-to-noise performance.

Page 5: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Direct-electron Detectors Devices (DDDs)

Directly detects electrons.

Suitable for automation.

Better signal-to-noise ratio than film.

Page 6: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Page 7: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

•The amount of additional noise added by a detector is measured by its detective quantum efficiency (DQE)

•Ideally a detector would not add any noise and so have a DQE of 1 but all real detectors have values less than 1.

Page 8: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Direct-electron Detectors Devices (DDDs)

Directly detects electrons.

Suitable for automation.

Better signal-to-noise ratio than film.

Very fast acquisition (17-100s frames/second)

We can “see” how the particles behaves as a function of the dose!

Page 9: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Direct-electron Detectors Devices (DDDs)

Directly detects electrons.

Suitable for automation.

Better signal-to-noise ratio than film.

Very fast acquisition (17-100s frames/second)

We can “see” how the particles behaves as a function of the dose!

The particles move !!! Beam Induced Movement

Page 10: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Page 11: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Page 12: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Brilot, A.F., Chen, J.Z., Cheng, A., Pan, J., Harrison, S.C., Potter, C.S., Carragher, B., Henderson, R., and Grigorieff, N. (2012). Beam-induced motion of vitrified specimen on holey carbon film. J Struct Biol 177, 630-637.

Page 13: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

The present…

Global correction [Li 2013] By frame correction [Bai 2013]

Line track correction [Scheres 2014]

Page 14: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Li et al., 2013•The alignment method consists in a pure drift correction approach with some local correction dividing each sub-frame area into 3 × 3 regions, and determining the local motion of each region.

Fast, converts each movie to an aligned micrograph: Simple connection with standard image processing workflows

Not the best option for pronounced local BIM movements

[Li 2013]

Page 15: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Bai et al., 2013 16 Frame average

8 Frame average

6 Frame average

4 Frame average

2 Frame average

Page 16: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Bai et al., 2013

Page 17: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Scheres, 2014

Page 18: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Accurate in correcting for local and global in-plane

BIM movements

Optical Flow

Fast and “standard” (converts each

movie to an aligned micrograph)

New Graphical Tool to characterize BIM

pattern

Abrishami et al., JSB, 2015

Page 19: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Abrishami et al., JSB, 2015

Page 20: Beam Induced Movement and DDDs Javier Vargas Centro Nacional de Biotecnología-CSIC jvargas@cnb.csic.es.

Abrishami et al., JSB, 2015