Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the...

19
11/4/2015 1 { Basics of Ultrasound Physics Department of Anesthesiology and Perioperative Medicine Division of Cardiothoracic Anesthesiology and Critical Care Medical College of Georgia, Georgia Regents University Manuel Castresana, MD, FCCM, FACA o Sound is the vibration of a physical medium o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The resulting tissue vibrations consist of areas of compression and rarefaction resembling a sine wave Vibrations

Transcript of Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the...

Page 1: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

1

{

Basics of Ultrasound Physics

Department of Anesthesiology and Perioperative Medicine

Division of Cardiothoracic Anesthesiology and Critical Care Medical College of Georgia, Georgia Regents University

Manuel Castresana, MD, FCCM, FACA

o Sound is the vibration of a physical medium

o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations

o The resulting tissue vibrations consist of areas of compression and rarefaction resembling a sine wave

Vibrations

Page 2: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

2

o US is sound with frequencies greater than those of the audible range for humans (20,000 Hz)

o Medical ultrasound waves have frequencies of 2–10 MHz

o Sound waves are mechanical waves created in the transducer by back and forth displacement

Physics

o US transducers send out sound waves then “listen” for returning echoes

o Amplitude of a sound wave represents its peak pressure

o Frequency and wavelength: Sound waves are characterized by their frequency (expressed in cycle per second Hz) and wavelength

o Propagation velocity determine by the medium (soft tissue 1,540 m/s)

Physics

Page 3: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

3

Sound Wave

A

mp

lit

ud

e

Wavelength

Velocity

0 .5 microseconds

Frequency = 4 cycles/0.5 = 8 MHz

o Acoustic impedance determines the amount of sound waves transmitted and reflected by tissues

o Reflection occurs when the ultrasound beam hits two tissues (areas) with different acoustic impedances

o Large differences in impedances inhibit useful information

Physics

Page 4: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

4

o The propagation of sound waves through the body is affected by the different tissues encountered

o It can result in signal

o Reflection

o Refraction (angle, acoustic impedance)

o Scattering (small or irregular structures)

o Attenuation (dispersion, absorption)

Interactions

Interactions of Sound and Tissue

Transesophageal Echocardiography. 2nd ed. Perrino and Reeves

Page 5: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

5

o Hyperechoic

o Structure reflects most sound waves

o Structure appears white on screen

Terms

o Anechoic

o Structure allows most sound waves through

o Structure appears black on screen

Terms

Page 6: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

6

o Lower frequency

Transducers

o Lower spatial resolution

o Deeper penetration

o Higher frequency

Transducers

o Better spatial resolution

o Reduced penetration

Page 7: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

7

o The higher the frequency, the better the resolution

o The better the resolution, the better your ability to distinguish objects from each other

Transducers

o High frequency US – provides a very detailed image of superficial structures to a depth of approximately of 5 cm

o Lower frequency US – capable of reaching into deeper structures but provides less detailed images

Transducers

Page 8: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

8

o Low 2-5 MHz o Image depth: Deep

o Attenuation: Low

o Image quality: Lower

o High 5-10 MHz o Image depth: Shallow

o Attenuation: High

o Image quality: Higher

Frequency

o Linear probe: 4-10 MHz, good near field images, poor penetration (vessels)

o Curved probe: 2-5 MHz, wider field of view (abdomen)

o Sector probe: 1.5-4 MHz, small footprint (cardiac)

Probe Selection

Page 9: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

9

o Transverse (short axis) – a cross-section, provides information about structures adjacent to the vessel of interest

o Longitudinal (long axis) – depicts structures anterior and posterior to the vessel of interest

Planes and Views

o 2D (two dimensional): Standard imaging mode

o M (motion) mode: Exchanges 2D image quality for high temporal resolution along a single line

o 3D: Uses ECG-gated acquisition of 2D images over a wide range of angles, reconstructs the image after acquiring the data

Modes: B mode

Page 10: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

10

o PW: Displays velocity spectrum at a point (limit by velocity)

o CW: Displays velocity spectrum at a line (not limited by high velocity)

o Color

o The strength of the Doppler signal is related to the velocity and the angle (zero)

Modes: Doppler

Doppler

Page 11: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

11

M Mode

Color Doppler

Notch of the transducer

should match the dot

Page 12: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

12

o Gain – Degree of amplification of the returning sound

o Increasing the gain increases the strength of the returning echoes and results in a lighter image

o Decreasing the gain does the opposite

Knobology

Knobology Too much gain

Page 13: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

13

Too little gain

Knobology

Optimal gain

Knobology

Page 14: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

14

o Depth

o Each frequency has a range of depth of penetration

o Decrease the depth to visualize superficial structures

o May need to increase the depth of penetration to visualize larger organs

Knobology

o Zoom

o Can place zoom box on a portion of a frozen image to enlarge that portion of the image

o May lose some resolution because pixels are enlarged

Knobology

Page 15: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

15

o Resolution

o Attenuation

o Propagation

o Doppler

Artifacts

o Acoustic shadowing: When US reaches an structure with very high acoustic impedance it will not penetrate any further

o Reverberation: If the US reaches two reflectors, it might reflect multiple times, like a candle standing between two mirrors

o Mirror image: Areas of high acoustic impedance may serve as an acoustic mirror deflecting the bean to the side

o Refraction artifact: Side by side with the anatomical structure

Common Artifacts

Page 16: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

16

Reverberation

Mirror Image

Page 17: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

17

o Dynamic guidance: Localization and insertion image guided, more precise, difficult to maintain sterility

o Static guidance: Localization and marking, cannulation is not image guided

US Guidance

o Pressure

o Alignment

o Rotation

o Tilt

Probe Adjustments

Page 18: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

18

Page 19: Basics of Ultrasound Physics - PAWS · Basics of Ultrasound Physics ... o In ultrasound, the transducer is a mechanical vibrator – the transducer creates tissue vibrations o The

11/4/2015

19

o Choose the correct probe

o Identify landmarks

o Probe orientation

o Maintain for 2D 90 for Doppler 0° Adjust gain and depth (and focus)

o Make adjustments to the probe (PART)

Key Points

Thank You