Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

37
Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung

Transcript of Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Page 1: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Backscatterfor EPID dosimetry

Tuesday Seminar

Wonmo Sung

Page 2: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Contents

• Introduction• Literature Review• SNUH clinic implementation• Conclusion

Page 3: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Electronic Portal Imaging Devices (EPID)

: Amorphous silicon photodiode

1) 2)

Introduction

1) For patient positioning

2) For DOSIMETRY!

-> PDIP (Portal Dose Image Prediction)

Page 4: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

However…

non-uniform support arm backscatter ?

+ How about backscatter from bunker?

Page 5: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Literature ReviewArm Backscatter• Physical approach

lead sheet btw arm & EPID (2010)

→ Varian BSS-EPID development (2012)• Software approach

backscatter kernel (2010) → (2013)

+ Bunker backscatter (2012)

All of above research were conducted in the same research group (PI: Peter B. Greer, University of Newcastle, Australia)

Page 6: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Backscatter(Rowshanfarzad et al, MP, 2010)

Devices: aS500 & E-arm(same physical components as the aS1000)

For 40x30 cm2 field(EPID + arm) – (EPID) = Max 6%

EPID+ARM EPID

Page 7: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Physical Approach

- lead sheet- BSS-EPID

Page 8: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Physical Approach - lead sheet(Rowshanfarzad et al, PMB, 2010)

Optimal lead sheet : 2 mm(Backscatter reduction vs. weight)

6 MV 18 MV

Lead 2 mm

0.1% 0.2%

w/o lead 2.3% 1.3%

For 30x30 cm2 field, at CAX

Page 9: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Symmetry improvement of in-line profiles

Contrast-to-noise ratio & resolution are decreased. But not significant reduction in image quality

Physical Approach - lead sheet(Rowshanfarzad et al, PMB, 2010)

Page 10: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

The displacement of the EPID central pixel was measured during a 360◦ gantry rotation with and without lead which was 1 pixel different.

Physical Approach - lead sheet(Rowshanfarzad et al, PMB, 2010)

Page 11: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

• Varian developed BackScatter-Shielded (BSS)-EPID.

• BSS-EPID incorporates a layer of lead shielding beneath the active area of EPID

• Compare BSS-EPID + arm vs. BSS-EPID

Physical Approach – Varian BSS-EPID(King et al, MP, 2012)

• For >99% pixels, 0.5% change• At CAX, maximum 0.2% effect on the central axis

Page 12: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Physical Approach – Varian BSS-EPID(King et al, MP, 2012)

• Use BSS-EPID to reconstruct dose in water for IMRT fields• Compare with MapCheck

2 cm depth, 2%/2 mm gammaBlack dots: failed

Page 13: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Software Approach

- backscatter kernel1) field size independent2) field size dependent

3) dose conversion

- just correct measurement(Varian updated)

Page 14: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Software approach – backscatter kernel (Rowshanfarzad et al, MP, 2010)

Rowshanfarzad developed simple backscatter kernel using measurements & Monte Carlo.

EPID+ARM EPID

Deconvolution+Gaussian fit

Page 15: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Gamma evaluation (2%/2 mm) for 20 × 20 cm2 field size

BEFORE AFTER

Software approach – backscatter kernel(Rowshanfarzad et al, MP, 2010)

Page 16: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

2%/2 mm Gamma evaluation, H&N IMRT Per FieldAvg. 2% improvement with backscatter kernel

Software approach – backscatter kernel (Rowshanfarzad et al, MP, 2010)

Page 17: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

FS dependent kernel…(Berry et al, MP, 2010)Kernel for dose reconstruction…(Wang et al., MP, 2013)

Anyway, better kernel development so far.

Software approach – backscatter kernel

Page 18: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

So far, we talked about…Energy fluence-convolution PDIP (Peter B. Greer, MP 36 (2009) 547-555)~ similar to CCC1) energy fluence (open beam+MLC leakage+head scatter) from Pinnacle2) several kernels

However, we use…

Varian simple convolution PDIP (Ann Van Esch, Green 71 (2004) 223-234)

1) No energy spcetra condisderation

2) Single kernel to model 1) radiation source distribution 2) head scatter on the incident energy fluence, 3) effect of EPID scatter, 4) optical glare on the dose deposition in the EPID, 5) MLC leaf end transmission,…

TERMA

Page 19: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Varian PDIP algorithm (Ann Van Esch)

Page 20: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Ann accepted three major drawbacks of her algorithm + 1D correction

1) 40x40cm2 diagonal profile does not describe 40 x 30 cm2 field edge.2) No backscatter correction3) No spectral consideration (for head scatter, MLC…)

Software approach – Just correct measurement!(Vinall et al, BJR, 2010)

She suggested 2D correction method to measurement dataPlease note that this is correction to measurement! Not to prediction as Peter did.Therefore…this method still possess problem.

Page 21: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Software approach – Just correct measurement!(Vinall et al, BJR, 2010)

R-arm, 40 × 30 cm2 field,in-line profile,

Predicted : Calculation

Official : Measurement + 1D-correction

2D no BS : Measurement + 2D-correction + backscatter cor-rection

2D with BS : Measurement + 2D-correction + no backscatter correc-tion

Page 22: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Software approach – Just correct measurement!(Vinall et al, BJR, 2010)

R-arm, 10 × 10 cm2 field,in-line profile,

Predicted : Calculation

Official : Measurement + 1D-correction

2D no BS : Measurement + 2D-correction + backscatter cor-rection

2D with BS : Measurement + 2D-correction + no backscatter correc-tion

Page 23: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Bunker backscatter

Page 24: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Bunker Backscatter(Rowshanfarzad et al, JACMP, 2012)

1) Portable brick wall experiment2) Response variation with various gantry angle (MatriXX: output variation)Distance btw isocenter – left wall:370, right wall:385, floor:130, ceiling 145 cm3) “Worst case scenario” : SDD=150+phantom, relative differences within 1.0%

1)

Page 25: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

1) Portable brick wall experiment2) Response variation with various gantry angle (MatriXX: output variation)Distance btw isocenter – left wall:370, right wall:385, floor:130, ceiling 145 cm3) “Worst case scenario” : SDD=150+phantom, relative differences within 1.0%

2)

Bunker Backscatter(Rowshanfarzad et al, JACMP, 2012)

Page 26: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

1) Portable brick wall experiment2) Response variation with various gantry angle (MatriXX: output variation)Distance btw isocenter – left wall:370, right wall:385, floor:130, ceiling 145 cm3) “Worst case scenario” : SDD=150+phantom, relative differences within 1.0%

3)

Bunker Backscatter(Rowshanfarzad et al, JACMP, 2012)

Page 27: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

SNUH clinic implementation

Page 28: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

• 방사선치료임상실습(805.611 Practicum in Clinical Radiation Therapy)

Instructor : 예성준

• 10-11 Weeks : three clinic-related topics

Lecturer : 김정인

① OSLD sensitivity trends : 박소연 , 박종인 ② MatriXX commissioning : 이재기 , 이현석 ③ PDIP commissioning : 성원모 , 허태민

Page 29: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Varian PDIP commissioning procedure

Beam data import

EPID calibra-tion

Imager Dark & Flood field calibration

Dosimetry calibration

Profile Correction

Dmax 1D diagonal profile (w2CAD)

Output Factor measurement

Actual Fluence measurement

Optimal & Actual fluence

Intensity profileCopy from PBC model

Page 30: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Varian Update (CTB-PV-887-A, Oct. 15th, 2012 released)

(a) The pre-configured PDIP algorithm for import into ‘Beam Configuration’.

(b) A file for 2D profile correction including backscatter correction of the support arm imported during ‘Dosimetry Calibration’ on the treatment console.

Beam data import

EPID calibra-tion

Imager Dark & Flood field calibration

Dosimetry calibration

Profile Correction

Dmax 1D diagonal profile (w2CAD)

Output Factor measurement

Actual Fluence measurement

Intensity profileCopy from PBC model

(b) 2D profile includ-ing backscatter cor-

rection

Optimal & Actual fluence(a) Golden Beam Data

Page 31: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Varian Update (CTB-PV-887-A, Oct. 15th, 2012 released)

Using Golden beam data means…

Golden EPID output factor & Golden kernel

<My opinion - limitation>

1) Machines specific characteristics

Ex. Same EPID? Same Linac?

Page 32: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Varian Update (CTB-PV-887-A, Oct. 15th, 2012 released)

2D profile correction to measurement

including backscatter provided by Varian

G

<My opinion - limitation>

1) field-size dependency on backscatter

2) Imager at off-axis

Page 33: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Possible choices

Prediction

1D-corrected Measured Beam Data

Golden Beam

Measurement

2D-corrected Measured Beam Data

1D-corrected Measurement

2D-corrected Measurement

BESTRESULTS

??

1D-corrected Measurement

Varian Previous Approach

Varian Updated Approach

P1

P2

PG

M1

M2

Page 34: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Preliminary resultsMachine : Varian Trilogy 6X, aS1000, E-arm

Varian provided plan

Relative dose normalized at isocenter

ROI selection : Threshold of 5%

Gamma Pass rate (%, 3 mm / 3% criteria, Local)

  P1 - M1 P1 - M2 PG - M1 PG - M2

IMRT1 84.1 82.2 91.5 93.5

IMRT2 91.5 94.3 91.7 94.2

VMAT1 72.2 76 65.6 69.5

VMAT2 95.3 91.7 98.7 97.9

Average 85.775 86.05 86.875 88.775

So far, “Varian previous approach” was the worstand “Varian updated approach” was the best. How-ever, NOT ALWAYS.

Page 35: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Future plan

Re-Commissioning 1D-corrected measured Beam Data Commissioning 2D-corrected measured Beam Data

Select 10 H&N clinical cases

Mapcheck measurements to verify Trilogy TPS parameters (?) EPID measurements for 1D-corrected & 2D-corrected 10 H&N

Page 36: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Conclusions

Further improvement in the agreement between the model prediction and EPID measurements is required.

More investigation is necessary for SNUH clinic im-plementation.

Page 37: Backscatter for EPID dosimetry Tuesday Seminar Wonmo Sung.

Thanks for your listening!Any questions?