Removing and Replacing the Transducer Battery
PAD
PAD Avalon CTS Transducers M2725A, M2726A & M2727A Instruction Sheet
CAUTION: NEVER immerse a transducer in liquid if the battery drawer is open or has been removed. Check the condition of the drawer before refitting, and verify that it is fully closed before use.
x1 x1
Use only this tool to open drawer
M2720-64001 Battery Replacement Kit
x1
Centralinsertion point(unmarked)
Old battery
New battery
Old battery
Silicone grease inside gasket
continued > >
Correct use Incorrect use Recycle according tolocal laws
*M2720-9061C*
Printed in GermanyAugust 2005Part Number M2720-9061C, 12NC 4512 610 11401
© Copyright 1995-2005 Koninklijke Philips Electronics N.V.All Rights Reserved.Philips Medizin Systeme Boeblingen GmbHHewlett-Packard-Str.271034 Boeblingen, Germany
0366
Wipe off excess silicone grease.
10
Informaton for Silcone Grease
Manufacturer:
ELBEL. Böwing GmbHDelkenheimer Str. 1865719 Hofheim a. Ts.Germany
Product: Elbesil BM Silicone Grease
Lot number: Impressed on the fold at the bottom of the tube
Minimum shelf life: See sticker on battery exchange kit packaging