AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight...

6
AUTOCLAVE STEAM STERILIZERS (S CLASS BENCH TOP) Head Office Phone (02) 9604 3911 Head Office Fax (02) 9725 1706 Email [email protected] Web www.thermoline.com.au DEC.2014 Image: HS-1607SD Hanshin Bench Top Steam Sterilizer Small and easy to operate Three preset sterilization cycles Adjustable cycle presets Rapid heat up times Option to attach printer Automatic safety door lock Audible and visual alarms

Transcript of AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight...

Page 1: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

AUTOCLAVESTEAM STERILIZERS(S CLASS BENCH TOP)

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Image: HS-1607SD

Hanshin Bench Top Steam Sterilizer

• Small and easy to operate• Three preset sterilization cycles• Adjustable cycle presets• Rapid heat up times• Option to attach printer• Automatic safety door lock• Audible and visual alarms

Page 2: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

SPECIFICATIONS

PRODUCT DETAILS

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Model HS-1607SD

External Dimensions (WxDxH mm) 380 x 534 x 315

Chamber Dimensions (dia x D mm) 160 x 377

Capacity (litres) 7

Reservoir (litres) 1.5

Trays (WxDxH mm) 2 trays 120 x 320 x 18

Air Filter HEPA 99.997% for 0.3micron particles

Power 2000W / 8 Amps

Sterilization Agent Saturated Steam of Max 2.1 Bar Pressure

Air Removal Gravity (positive pressure pulsing)

Control Fully automated microprocessor

Exposure Temperature 121°C to 135°C

Weight (kg) 28

The HS-1607SD is a small, easy to operate, bench top steam sterilizer with a 7 litre capacity stainless steel chamber. This model is ideal for small laboratories or health care facilities such as dental surgeries which may have instruments or other items requiring sterilization. This model features five pre-set sterilization cycles which can be easily selected by pressing the ‘prog’ button located on the front panel.

Each cycle is performed automatically (heating, sterilizing and venting) and a signal sounds to alert the user that the cycle has completed. The front panel has an LED display to show the current working temperature and error code number should an error occur during operation. User safety is ensured with a door lock system, safety pressure valve and an overheat protection device.

Thermoline Scientific have been manufacturing and distributing high quality laboratory and scientific testing equipment since 1970. Over this time, Thermoline has grown to be a leading brand in the science industry, with our equipment being used in small and large Hospitals, Universities and Research Laboratories across Australia and the Asia Pacific region.

We’re proud to say that we are 100% Australian owned and operated.

Safety Devices

Automatic lid lock which prevents the door from opening while the chamber is still hot

Cycle Number 1 2 3 4 5

Load Type Unwrapped Wrapped Unwrapped Wrapped Liquid

Exposure Temperature 134°C 121°C

Exposure Time (min) 3 15 20 30 30

Dry Time (min) - 30 20 30 -

Cooling Time (min) - - - - 60

Page 3: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

AUTOCLAVESTEAM STERILIZERS(S CLASS BENCH TOP)

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Image: HS-3041SD

Hanshin Bench Top Steam Sterilizer

• Two models available - 21.5L and 40.5L• Five preset sterilization cycles• Adjustable cycle presets• Rapid heat up times• Option to attach printer• Automatic safety door lock• Audible and visual alarms• TGA Approval Number ARTG No. 112473

Certified Class S Steam Sterilizer(Gravity air removal and vacuum drying system)S class sterilizers are capable of sterilizing non-wrapped solid instruments and liquid products. These sterilizers also have a special purpose cycle program for this process.This sterilizer has obtained a CE certificate by passing all tests regulated by EU standards.

Page 4: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

SPECIFICATIONS

PRODUCT DETAILS

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Model HS-2321SD HS-3041SD

External Dimensions (WxDxH mm) 513 x 625 x 442 615 x 757 x 505

Chamber Dimensions (WxDxH mm) 225 x 425 x 225 300 x 450 x 300

Capacity (litres) 21.5 40.5

Reservoir (litres) 5 6

Tray (WxDxH mm) 2 trays - 216 x 405 x 44 2 trays - 291 x 44 x 432

Pressure Gauge Dial Type, 760mmHg, -4bar

Power 1600W / 7 Amps 2000W / 8 Amps

Sterilization Agent Saturated Steam of Max 2.1 Bar Pressure

Air Removal Gravity

Control Fully automated microprocessor

Exposure Temperature 121°C to 134°C

Air Filter HEPA 99.997% for 0.3micron particles

Weight (kg) 55 77

The two larger S Class models available are the HS-2321SD and HS-3041SD. These bench top steam sterilizer have a 21.5L and 40.5L capacity stainless steel chamber, holding two stainless steel trays for the easy storage of items needing to be sterilized. These units are very well engineered and have been approved by the TGA for safe use throughout Australia.

An easy to use control panel located on the front of the unit allows users to change programs, view the drying phase, turn the unit on and off as well as indicators to show the cycle progress and water level. Five pre-set cycle programs are also supplied which have commonly used cycle times and temperatures.

These models have a safety door lock which prevents the door from being opened during operation and prevents cycles from starting should the door be left open. An inbuilt safety valve helps protect the chamber should it exceed the standard pressure rating.

Thermoline Scientific have been manufacturing and distributing high quality laboratory and scientific testing equipment since 1970. Over this time, Thermoline has grown to be a leading brand in the science industry, with our equipment being used in small and large Hospitals, Universities and Research Laboratories across Australia and the Asia Pacific region.

We’re proud to say that we are 100% Australian owned and operated.

Safety Devices

Automatic lid lock which prevents the door from opening while the chamber is still hot

Cycle Number 1 2 3 4 5

Load Type Unwrapped Wrapped Unwrapped Wrapped Liquid

Exposure Temperature 134°C 121°C

Exposure Time (min) 5 15 20 30 30

Dry Time (min) 15 30 -

Page 5: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

AUTOCLAVESTEAM STERILIZERS(B CLASS BENCH TOP)

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Hanshin Bench Top Steam Sterilizer

• Two models available - 21.5L and 40.5L• Rectangle vessle with two trays• Five preset sterilization cycles• Adjustable cycle presets• Large LED display• Rapid heat up times• Printer fitted standard• Automatic safety door lock• Audible and visual alarms• TGA Approval Number ARTG No. 112473

Certified Class B Steam Sterilizer(Pre-vacuum air removal and vacuum drying)B class sterilizers are suitable for use with medical items such as wrapped or non-wrapped solid instruments, hollow type instruments, porous products and liquid products (regardless of package and load type).These models provide effective sterilization of “hard to sterilize” items such as hollow tubes, endoscopes and sponges. A powerful air removal system ensures Hollow Load Tests are completed in accordance with strict European Standards.

Image: HS-3041BD

Page 6: AUTOCLAVE STEAM STERILIZERS - Thermoline · Air Filter HEPA 99.997% for 0.3micron particles Weight (kg) 55 77 The two larger S Class models available are the HS-2321SD and HS-3041SD.

SPECIFICATIONS

PRODUCT DETAILS

Head Office Phone (02) 9604 3911Head Office Fax (02) 9725 1706

Email [email protected] www.thermoline.com.au

DEC

.2014

Model HS-2321BD HS-3041BD

External Dimensions (WxDxH mm) 513 x 625 x 442 615 x 757 x 505

Chamber Dimensions (WxDxH mm) 225 x 425 x 225 300 x 450 x 300

Capacity (litres) 21.5 40.5

Reservoir (litres) 5 6

Tray (WxDxH mm) 2 trays 216 x 405 x 44 2 trays 291 x 432 x 44

Air Filter HEPA 99.997% for 0.3micron particles

Power 2300 W / 9Amps 3100W / 13Amps (15 Amp plug fitted)

Sterilization Agent Saturated Steam of Max 2.1 Bar Pressure

Air Removal Vacuum / Gravity

Control Fully automated microprocessor

Exposure Temperature 121°C to 135°C

Weight (kg) 56 81

The BD series autoclave is designed for economical efficiency but can still sterilize all your medical instruments and devices efficiently. The process firstly removes all air from the chamber then filling with steam which allows the steam to then penetrate and sterilize effectively. The sterilizing cycles have a unique drying process so the load is dried under vacuum and ensures the entire process complies with international standard procedures.

This model is equipped with a printer which prints out all important reference data. The two models avaialble are 21.5L and a large 40.5 litre capacity chamber manufactured from 316 stainless steel, holding two stainless steel trays. Five pre-set cycle programs plus the Bowie-Dick/Helix Test are available all selectable via the front controller.

This model has been approved by the TGA for safe use throughout Australia.

Thermoline Scientific have been manufacturing and distributing high quality laboratory and scientific testing equipment since 1970. Over this time, Thermoline has grown to be a leading brand in the science industry, with our equipment being used in small and large Hospitals, Universities and Research Laboratories across Australia and the Asia Pacific region.

We’re proud to say that we are 100% Australian owned and operated.

Safety Devices

Automatic lid lock which prevents the door from opening while the chamber is still hot

Cycle Number 1 2 3 4 5 Bowie- Dick Vacuum Leak

Load Type Unwrapped Wrapped Unwrapped Wrapped Liquids Test Pack n/a

Exposure Temperature 134°C 121°C 134°C Dwell: 5minTest: 10minExposure Time (min) 5 15 20 30 30 3.5

Dry Time (min) 15 30 None 5

Cooling Time (min) - 60 -