Audio03 audiometry

37
Basic Principles of Sound The Audiometer Test Environment Patient’s Role Clinician’s Role Air Conduction Audiometry Bone Conduction Audiometry Audiogram Interpretation Masking Pure Tone Audiometry

Transcript of Audio03 audiometry

• Basic Principles of Sound

• The Audiometer

• Test Environment

• Patient’s Role

• Clinician’s Role

• Air Conduction Audiometry

• Bone Conduction Audiometry

• Audiogram Interpretation

• Masking

Pure Tone Audiometry

• Sound Waves and Propagation of Sound

• Frequency (Hz)

• Resonance

• Intensity (The Decibel or dB)

Basics of Sound

• Generic Audiometer

Audiometer

• Clinical and Portable Audiometer

Audiometer

• Earphones

Audiometer

• Insert Earphone

Audiometer

• Bone Conduction Vibrator

Audiometer

• Computerized Audiometer

Audiometer

• Sound Treated Booth

Test Environment

• Circumaural Enclosures

Test Environment

• Quiet Room

Test Environment

• Hand Raising

• Signal Button

• Verbal Response

• False Positives and False Negatives

Patient’s Role

• Instructions

• Patient’s Position

• Placement of Earphones

• Test Procedures for Screening

• Test Procedures for Pure Tone Thresholds

Clinician’s Role

• Instructions• What are they listening for

• How to respond

• Verify they understand instructions

Clinician’s Role

• Position of Client - Adults and Children

• Earphone Placement

Clinician’s Role

• Screening Test Procedures (adults)• Instructions

• Test Frequencies

• Disposition of Failures

Clinician’s Role

• Pure Tone Threshold Procedures• Self-test

• Place earphones on patient

• Test better ear first

• Order of test frequencies

• Test other ear

• Use ASHA protocol for threshold testing.

Clinician’s Role

• Pure Tone Average (PTA)• Three tone average

• Two tone average

• Other methods

Clinician’s Role

• Degree of Hearing Loss based on PTA• None

• Slight

• Mild

• Moderate

• Moderately Severe

• Severe

• Profound

Clinician’s Role

• Role of Bone Conduction• Purpose

• Mastoid Placement

• Forehead Placement

Clinician’s Role

• Basic Audiogram

Audiograms

• Symbols

Audiograms

• Type

• Degree

• Configuration

Audiograms

• Type of Loss - Conductive

Audiograms

• Type of Loss - Sensorineural

Audiograms

• Type of Loss - Mixed

Audiograms

• Configuration of Loss - Flat

Audiograms

• Configuration of Loss - Sloping (falling)

Audiograms

• Configuration of Loss - Rising

Audiograms

• Configuration of Loss - Tent

Audiograms

• Configuration of Loss - Mid-frequency or Cookie-bite

Audiograms

• Cross Hearing and Interaural Attenutation

• Masking Defined

• Rules

• Types of Masking Noise

Masking

• Cross Hearing

• Interaural Attenuation• Definition

• Variables

• transducers

• frequency

• individual variability

Masking

• Definition• Introduction of noise in NTE for the purpose of

eliminating cross-hearing.

Masking

• Rule• Apply masking to NTE whenever the AC of the TE

exceeds the BC for the NT cochlea by the amount of the minimum IA values.

• Minimum IA values• Supraaural phones = 40 dB

• Insert phones = 70 dB

• Bone conduction = 0 dB

Masking

• Types of Masking Noise• White Noise

• Narrow Band Noise

Masking

Summary