Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020 SCIENCE ADVANCES | RESEARCH ARTICLE 1 of 8 APPLIED PHYSICS Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometers Rui Zhang 1,2 *, Wei Xiao 1 *, Yudong Ding 1 , Yulong Feng 1 , Xiang Peng 1 , Liang Shen 3 , Chenxi Sun 1 , Teng Wu 1† , Yulong Wu 1 , Yucheng Yang 1 , Zhaoyu Zheng 1 , Xiangzhi Zhang 1 , Jingbiao Chen 1 , Hong Guo 1† Understanding the relationship between brain activity and specific mental function is important for medical diagnosis of brain symptoms, such as epilepsy. Magnetoencephalography (MEG), which uses an array of high- sensitivity magnetometers to record magnetic field signals generated from neural currents occurring naturally in the brain, is a noninvasive method for locating the brain activities. The MEG is normally performed in a mag- netically shielded room. Here, we introduce an unshielded MEG system based on optically pumped atomic magnetometers. We build an atomic magnetic gradiometer, together with feedback methods, to reduce the environment magnetic field noise. We successfully observe the alpha rhythm signals related to closed eyes and clear auditory evoked field signals in unshielded Earth’s field. Combined with improvements in the min- iaturization of the atomic magnetometer, our method is promising to realize a practical wearable and movable unshielded MEG system and bring new insights into medical diagnosis of brain symptoms. INTRODUCTION Magnetoencephalography (MEG) is a functional neuroimaging tech- nique for mapping human brain activities using an array of magneto- meters to detect magnetic fields produced by neural currents (13). Compared with its electric signal detection–based counterpart, i.e., scalp electroencephalography (EEG) (45), MEG has a higher spatial resolution due to the nearly homogeneous magnetic permeability of biological tissues (6). MEG is a promising noninvasive method for locating brain activities compared with intracranial EEG, an invasive method using electrodes placed directly on the exposed surface of the brain to record electrical signals (7). MEG has various applications in neuroscience (8), cognitive science (9), and especially in the diagnosis of brain symptoms such as epilepsy presurgical evaluation (10), Alzheimer’s disease (11), Parkinson’s disease (12), traumatic brain injury (13), and multiple sclerosis (14). The MEG system typically uses two different kinds of magne- tometers as the magnetic sensors, the superconducting quantum interference devices (SQUIDs) (1517) or the optically pumped atomic magnetometers (OPMs) (1825). For the commercial MEG systems, it is the SQUID sensors that are widely used. Although the SQUID sensor has a sufficient sensitivity to detect the weak magnetic field signals originating from brain activities, the cryogenic operational condition demands that the sensor geometry has to be fixed in a liquid helium dewar, which makes the measuring helmet of the commercial MEG system costly and inflexible. The associated immovability of the measuring helmet limits the applications of MEG in cases where head motion is needed. A general tendency within the MEG community is replacing SQUID with noncryogenic high-sensitivity OPM as the magnetic sensor (1825), considering that the sensitivity of the OPM sensor has been comparable with that of the most sensitive SQUID sensor and the miniaturization has become much improved (2627). A wearable MEG system based on an array of OPM sensors, which allows head movement and customization of the measuring helmet, has already shown similar performances in both the spatial resolution and sensitivity in com- parison to a commercial MEG system based on SQUID (25). There are several advantages of using OPM sensor in MEG, for example, noncryogenic, movable, custom-design friendly, and less expensive. These advantages make it possible to construct a wearable and movable MEG system for noninvasive long-term observations of the patients with brain symptoms, such as epilepsy, to capture the infrequent epileptiform events (28). Such a task is conventionally tackled by the scalp EEG, considering that the MEG system based on SQUID is not suitable for long-term observations (29). The per- formance of the EEG in locating the brain symptoms is limited by the nonuniform resistive properties of biological tissues. From this aspect, using OPM as the MEG sensor is important for some clinical applications, especially in the noninvasive surgical treatment of ep- ilepsy that demands a precise localization of the epileptogenic focus. Almost all the MEG measurements based on OPM sensors are performed in magnetically shielded rooms (MSRs) for suppressing the ambient magnetic field noise and providing a nearly zero mag- netic field circumstance therein. The application of MSR improves the signal-to-noise ratio (SNR) of the MEG signal, whereas it limits the potential advantages of using OPM-based MEG in clinical ap- plications and for long-term observations (1028). For example, the crowded inner space of MSR makes the long-term observations less efficient. There are several important advantages to eliminating the MSR, e.g., reducing cost, ease of siting of OPM-based MEG systems, and a more naturalistic environment. The unshielded MEG system also makes it possible to be combined with the low-field magnetic resonance imaging system (16) and provides a convenient way for the coregistration in locating and imaging the brain activities. Thus, it is of great demand that the OPM-based MEG measurements can be performed in unshielded Earth’s field. 1 State Key Laboratory of Advanced Optical Communication Systems and Networks, Department of Electronics, and Center for Quantum Information Technology, Peking University, Beijing 100871, China. 2 College of Liberal Arts and Sciences, and Interdisciplinary Center for Quantum Information, National University of Defense Technology, Changsha, Hunan 410073, China. 3 State Key Laboratory of Information Photonics and Optical Communications, Beijing University of Posts and Telecom- munications, Beijing 100876, China. *These authors contributed equally to this work. †Corresponding author. Email: [email protected], [email protected] (T.W.); [email protected] (H.G.) Copyright © 2020 The Authors, some rights reserved; exclusive licensee American Association for the Advancement of Science. No claim to original U.S. Government Works. Distributed under a Creative Commons Attribution NonCommercial License 4.0 (CC BY-NC). on August 27, 2020 http://advances.sciencemag.org/ Downloaded from

Transcript of APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic...

Page 1: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

1 of 8

A P P L I E D P H Y S I C S

Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometersRui Zhang1,2*, Wei Xiao1*, Yudong Ding1, Yulong Feng1, Xiang Peng1, Liang Shen3, Chenxi Sun1, Teng Wu1†, Yulong Wu1, Yucheng Yang1, Zhaoyu Zheng1, Xiangzhi Zhang1, Jingbiao Chen1, Hong Guo1†

Understanding the relationship between brain activity and specific mental function is important for medical diagnosis of brain symptoms, such as epilepsy. Magnetoencephalography (MEG), which uses an array of high- sensitivity magnetometers to record magnetic field signals generated from neural currents occurring naturally in the brain, is a noninvasive method for locating the brain activities. The MEG is normally performed in a mag-netically shielded room. Here, we introduce an unshielded MEG system based on optically pumped atomic magnetometers. We build an atomic magnetic gradiometer, together with feedback methods, to reduce the environment magnetic field noise. We successfully observe the alpha rhythm signals related to closed eyes and clear auditory evoked field signals in unshielded Earth’s field. Combined with improvements in the min-iaturization of the atomic magnetometer, our method is promising to realize a practical wearable and movable unshielded MEG system and bring new insights into medical diagnosis of brain symptoms.

INTRODUCTIONMagnetoencephalography (MEG) is a functional neuroimaging tech­nique for mapping human brain activities using an array of magneto­meters to detect magnetic fields produced by neural currents (1–3). Compared with its electric signal detection–based counterpart, i.e., scalp electroencephalography (EEG) (4, 5), MEG has a higher spatial resolution due to the nearly homogeneous magnetic permeability of biological tissues (6). MEG is a promising noninvasive method for locating brain activities compared with intracranial EEG, an invasive method using electrodes placed directly on the exposed surface of the brain to record electrical signals (7). MEG has various applications in neuroscience (8), cognitive science (9), and especially in the diagnosis of brain symptoms such as epilepsy presurgical evaluation (10), Alzheimer’s disease (11), Parkinson’s disease (12), traumatic brain injury (13), and multiple sclerosis (14).

The MEG system typically uses two different kinds of magne­tometers as the magnetic sensors, the superconducting quantum interference devices (SQUIDs) (15–17) or the optically pumped atomic magnetometers (OPMs) (18–25). For the commercial MEG systems, it is the SQUID sensors that are widely used. Although the SQUID sensor has a sufficient sensitivity to detect the weak magnetic field signals originating from brain activities, the cryogenic operational condition demands that the sensor geometry has to be fixed in a liquid helium dewar, which makes the measuring helmet of the commercial MEG system costly and inflexible. The associated immovability of the measuring helmet limits the applications of MEG in cases where head motion is needed. A general tendency within the MEG community is replacing SQUID with noncryogenic

high­sensitivity OPM as the magnetic sensor (18–25), considering that the sensitivity of the OPM sensor has been comparable with that of the most sensitive SQUID sensor and the miniaturization has become much improved (26, 27). A wearable MEG system based on an array of OPM sensors, which allows head movement and customization of the measuring helmet, has already shown similar performances in both the spatial resolution and sensitivity in com­parison to a commercial MEG system based on SQUID (25).

There are several advantages of using OPM sensor in MEG, for example, noncryogenic, movable, custom­design friendly, and less expensive. These advantages make it possible to construct a wearable and movable MEG system for noninvasive long­term observations of the patients with brain symptoms, such as epilepsy, to capture the infrequent epileptiform events (28). Such a task is conventionally tackled by the scalp EEG, considering that the MEG system based on SQUID is not suitable for long­term observations (29). The per­formance of the EEG in locating the brain symptoms is limited by the nonuniform resistive properties of biological tissues. From this aspect, using OPM as the MEG sensor is important for some clinical applications, especially in the noninvasive surgical treatment of ep­ilepsy that demands a precise localization of the epileptogenic focus.

Almost all the MEG measurements based on OPM sensors are performed in magnetically shielded rooms (MSRs) for suppressing the ambient magnetic field noise and providing a nearly zero mag­netic field circumstance therein. The application of MSR improves the signal­to­noise ratio (SNR) of the MEG signal, whereas it limits the potential advantages of using OPM­based MEG in clinical ap­plications and for long­term observations (10, 28). For example, the crowded inner space of MSR makes the long­term observations less efficient. There are several important advantages to eliminating the MSR, e.g., reducing cost, ease of siting of OPM­based MEG systems, and a more naturalistic environment. The unshielded MEG system also makes it possible to be combined with the low­field magnetic resonance imaging system (16) and provides a convenient way for the coregistration in locating and imaging the brain activities. Thus, it is of great demand that the OPM­based MEG measurements can be performed in unshielded Earth’s field.

1State Key Laboratory of Advanced Optical Communication Systems and Networks, Department of Electronics, and Center for Quantum Information Technology, Peking University, Beijing 100871, China. 2College of Liberal Arts and Sciences, and Interdisciplinary Center for Quantum Information, National University of Defense Technology, Changsha, Hunan 410073, China. 3State Key Laboratory of Information Photonics and Optical Communications, Beijing University of Posts and Telecom-munications, Beijing 100876, China.*These authors contributed equally to this work.†Corresponding author. Email: [email protected], [email protected] (T.W.); [email protected] (H.G.)

Copyright © 2020 The Authors, some rights reserved; exclusive licensee American Association for the Advancement of Science. No claim to original U.S. Government Works. Distributed under a Creative Commons Attribution NonCommercial License 4.0 (CC BY-NC).

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 2: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

2 of 8

Here, we develop and test an OPM­based MEG system operat­ing in unshielded Earth’s field. The OPM that we are using is based on amplitude­modulated nonlinear magneto­optical rotation (AM­NMOR; section S1), which is different from the spin­exchange relaxation­free (SERF) magnetometer (30–32), a conventionally used OPM sensor for MEG. The SERF magnetometer has the advantages of high degree of miniaturization and portability and is one of the most sensitive magnetic sensors operating near­zero magnetic field. As a comparison, the AM­NMOR magnetometer operates in a much wider range of magnetic field and could cover the Earth’s field (33, 34). We use two AM­NMOR cesium magnetometers and perform gradiometric detection, together with feedback methods, to reduce the magnetic field noise. The atomic magnetic gradiometer has a magnetic field gradient noise floor (sensitivity) of ~4 fT/cm Hz1/2 (1 fT = 10−15 T) under the unshielded condition. On the basis of the atomic magnetic gradiometer, we successfully observe the alpha rhythm signal related to open and closed eyes and a clear auditory evoked magnetic field (AEF) signal. We also fully charac­terize the technical issues to be solved and the corresponding im­provements to be implemented for realizing a practical wearable and movable MEG system. We recently learned that Limes et al. (35) also reported successful MEG measurements in unshielded Earth’s field. Compared with the AM­NMOR magnetometer that we use, the researchers demonstrate a configuration of atomic magnetom­eter operating in the pulsed mode. Through using microfabricated multipass vapor cells and custom electronics, they realize a compact system and reduce the power consuming to 5 W, making a quite no­table step forward.

RESULTSIn this section, we describe the measurement results of our MEG system in unshielded Earth’s field. First, we give a brief description on the unshielded MEG system and the performance in suppressing the common­mode magnetic field noise. Second, we present two typical MEG measurements. One is for the spontaneous alpha rhythm signal related to open and closed eyes. The other one is for the AEF signal. All the experiments are performed under the approval of the local Institutional Review Board and with all the participants providing written informed consent.

MEG system in unshielded Earth’s fieldThe overall layout of the real system is shown in Fig. 1A. We use two sets of three orthogonal coils. The outer layer is for adjusting the bias magnetic field strength, while the inner layer is for compensating the magnetic field fluctuations. The OPM sensors are placed at the center of the coils. The person is lying on a wooden table with the head resting on the OPM sensors. Figure 1 (B and C) shows the de­tails of the alpha rhythm signal and AEF signal detection experiments, respectively. The horizontal components of the Earth’s field are com­pensated to be within ±10 nT (1 nT = 10−9 T) around zero so that the remaining magnetic field is along the vertical direction, which is the sensitive axis of the OPM sensor. It is important to have the large magnetic field along the sensitive axis of the sensor to reduce the dead­zone effect. Technical details of the OPM sensor are described in Materials and Methods (36). The field strength B0 along the vertical direction can be set to any value (within ±10 nT) below 105 nT by adjusting the currents applied to the outer layer of the coils (the strength of the Earth’s field is ~54,500 nT in Beijing). Given that the

bandwidth of our OPM sensor is determined by the spin precession frequency, which is proportional to the strength of the magnetic field, we set the minimum magnetic field to be around 5000 nT.

The distance between the top and the bottom OPM sensors is 6 cm, and the two sensors are distant from the surrounding magnetic substances. The environment magnetic noise is taken as the common­ mode noise for the two sensors. Furthermore, since the head of the participant is placed more closely to the top OPM sensor, the ampli­tude of the MEG signal detected by the top sensor is almost an order of magnitude larger than that of the bottom one (section S2). On the basis of this consideration, we use the two OPM sensors to realize an atomic magnetic gradiometer (37), through which the measured results of the two sensors are subtracted so as to enhance the SNR of the MEG signal.

The ability of suppressing the common­mode magnetic noise, or the common­mode noise rejection ratio (CMRR), plays a critical role in determining the performance of the magnetic gradiometer. To get a higher CMRR, the responses of the two OPMs with respect to the same magnetic field signal should be as closed as possible. Figure 1 (D and E) shows that the measured frequency responses of the two OPMs are almost identical. Both of the two OPMs are working in the closed­loop mode, which provides a larger bandwidth and better performance in reducing the common­mode noise com­pared with the open­loop mode (37, 38). The 3­dB bandwidth of the two OPMs is both ~950 Hz, which is primarily determined by the parameter settings of the proportional integral derivative (PID) feedback system. To characterize the ability of rejecting the common­ mode magnetic field noise and the CMRR of the gradiometer, we apply a sinusoidal magnetic field to each of the two OPM sensors and measure the residual signal amplitude after subtracting the mea­surement results from the two OPMs. The orange triangle in Fig. 1F shows the measured CMRR of the gradiometer at different frequencies. The CMRR is larger than 1000 for frequencies lower than 30 Hz and then decreases with the increased frequency while still larger than 400 for frequencies between 30 and 100 Hz. The decrease in the CMRR at higher frequencies is primarily due to the nonidentical frequency responses of the two OPMs. The CMRR could be improved by increasing the bandwidths of the two OPMs (section S1). Details of the procedures for measuring the frequency responses of the two OPMs and the CMRR of the atomic gradiometer are described in Materials and Methods.

The magnetic field noise for an unshielded magnetically noisy environment, which is surrounded with ferromagnetic substances, could be much larger than tens of pT/Hz1/2 for the range of frequen­cies where the MEG signals occur. Therefore, even using a magnetic gradiometer with a CMRR of 1000, the residual magnetic noise floor of the atomic gradiometer is still larger than most of the MEG signals. The conventional method for suppressing the environment magnetic noise is using passive shielding. In our system, instead, the environ­ment magnetic noise is monitored with the bottom OPM sensor and then compensated by automatically controlling the current applied to the coils along the vertical direction (see Materials and Methods for details). Compared with magnetic field noise along the vertical direction, our OPM sensors are not sensitive to magnetic field noise along the horizontal directions as these field components add in quadrature to the dominant vertical magnetic field component. The blue square in Fig. 1F shows the result of CMRR using this noise compensation method, indicating that the magnetic noise is sup­pressed by ~800 times for frequencies below 10 Hz. The residual

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 3: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

3 of 8

noise is determined by the intrinsic noise level of the OPM sensor. For higher frequencies, the CMRR decreases due to the reduction of the gain of the error signal for field stabilization and can be im­proved by increasing the bandwidths of the two OPMs and optimiz­ing the parameter settings of PID controller for field stabilization (section S3).

In general, the combination of the two methods, i.e., the atomic gradiometer and the active field stabilization, gives an expected overall magnetic noise suppression ratio up to 106 for frequencies smaller than 10 Hz (see the green circles in Fig. 1F) and is comparable to a value expected from a passive MSR with seven mu­metal layers (39). These results indicate that our magnetic gradiometer, together with the field stabilization, is capable of measuring the weak MEG signals even in a noisy unshielded Earth’s field. It should be noted that the overall noise suppression ratio shown in Fig. 1F is a direct

multiplication of the CMRRs of the two individual noise suppression mechanisms. This is valid for conditions where the ambient mag­netic field noise is along the sensitive axis of the sensor and is large and identical for the two sensors, while the field generated inside the coils for field stabilization is homogeneous.

Spontaneous alpha rhythm signalThe alpha rhythm, or the alpha wave, is a pattern of slow brain wave with a typical frequency ranging from 7 to 13 Hz and arises from the synchronous and coherent electrical activity of thalamic pacemaker cells in humans (40). It is normally considered to represent the activity of the visual cortex in an idle state and/or the areas of the cortex not in use. For example, the alpha rhythm signal for a normal person is stronger with closed eyes and becomes suppressed with open eyes. The amplitude of the alpha rhythm signal is relative

OPM1OPM2

A B

C

Nonmag. earphone

OPM1OPM2

Am

plit

ud

e (a

rb. u

nit

)

1

0.1

Frequency (Hz)1 10 100 1000

Frequency (Hz)1 10 100 1000

Ph

ase

shif

t (°

) 0

−90

−180

−270

OPM1 OPM2

OPM1 OPM2

D

E

F

−3-dB level

CombinedField stabilizationGradiometer

106

105

104

103

102

10

1

CM

RR

Frequency (Hz)10 100 10001

Fig. 1. Unshielded MEG system using OPMs. (A) The overall arrangement of the practical unshielded MEG system. The OPMs are placed at the center of the two sets of three orthogonal coils. The diameter of the coils for the outer layer is 3.6 m, and for the inner layer is 1.5 m. (B and C) Detailed arrangements for detecting the alpha rhythm signal (B) and AEF signal (C). (D) Amplitude responses of the two OPMs. The orange dashed line shows the −3-dB level relative to the signal amplitude at 1 Hz. arb. unit, arbitrary units. (E) Phase responses of the two OPMs. (F) The green circle shows the ability of the unshielded MEG system to reduce the common-mode magnetic field noise under different frequencies and is a combined effect from the atomic gradiometer (orange triangle) and field stabilization (blue square). (Photo credit: Wei Xiao, Rui Zhang, and Hong Guo, Peking University, Beijing, China.)

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 4: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

4 of 8

strong among all the signals detected with EEG and MEG (40). On the basis of these characteristics, at first, we measure the alpha rhythm signal of a normal person in unshielded Earth’s field to test the performance of our MEG system.

For the alpha rhythm signal measurements, the person is asked to put his head, typically, his inion, right on top of the OPM sensor (see Fig. 1B) to have the strongest signal measured. The cesium vapor cell is kept at room temperature and has no heating issue for the person. Details of the measurement procedures are described in Materials and Methods. With no person present, the green line (region) in Fig. 2A shows that the noise floor of the atomic gradio­meter is 20 fT/Hz1/2 for frequencies between 1 and 30 Hz (magnetic field gradient noise level of ~4 fT/cm Hz1/2 with a 6­cm baseline). The noise of the atomic gradiometer for frequencies larger than 40 Hz (not shown in Fig. 2A) comes from the intrinsic noise of the OPM sensors and is not reduced with the gradiometric detection. Com­pared with the sensor operating in the open­loop mode, the sensor operating in the closed­loop is known to have enhanced bandwidth,

however, have no improved SNR, and results in an increase in the sensor noise outside the original magnetic resonance line width (section S3). The noise level slightly increases when the person is present, which may come from the mechanical vibrations. The person is asked to keep his eyes closed and open, both for 180 s. The blue region in Fig. 2A shows the recorded spectrum when the person closes his eyes. As a comparison, the spectrum obtained with open eyes is also plotted, shown as the yellow region in Fig. 2A. We could observe an alpha rhythm signal around 10 Hz when the person keeps his eyes closed, including a suppression in the signal amplitude when the eyes are open. Besides the dominant alpha rhythm signal around 10 Hz, there is also a small peak around 20 Hz, which is possibly due to the second harmonic of the alpha rhythm signal (41).

The person is also asked to repeat closing and opening his eyes every 30 s, and the magnetic field signal is recorded in time domain, shown in Fig. 2B. There is a clear change in the amplitude of the magnetic field signal every 30 s. When the eyes open, the peak­to­peak variation of the magnetic field signal is ~0.5 pT and increases

Frequency (Hz)5 10 15 20 25 30 35 40

Sp

ectr

al d

ensi

ty (

pT

/Hz1/

2 )

0

0.05

0.10 No person

With person, eyes closed

With person, eyes open

A

Alpha rhythm signal

Mag

net

ic f

ield

(p

T)

0

0.5

1.0

−0.5

−1.0

B Eyes closed

Time (s)0 50 100 150 200 250 300

Time (s)0 50 100 150 200 250 300

Fre

qu

ency

(H

z)

5

20

10

15

0 0.16Magnetic field (pT/Hz1/2)

Eyes open

C

Fig. 2. Spontaneous alpha rhythm signal. (A) The green region shows the measured spectral density of the atomic gradiometer with no person present and indicates a noise floor of 20 fT/Hz1/2 for frequencies between 5 and 30 Hz. When the person is present and asked to keep his eyes open, the measured result is shown in the yellow region. When the person closes his eyes, a clear alpha rhythm signal around 10 Hz is observed, see the blue region. (B) The time domain signal acquired when the person is asked to repeat closing and opening his eyes every 30 s. (C) The spectrogram for the acquired 300-s data shown in (B).

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 5: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

5 of 8

to ~1 pT with closed eyes. Figure 2C shows the spectrogram of the acquired 300­s data, of which the frequency range is 5 to 20 Hz. All the measurements are performed in unshielded Earth’s field, with a bias field of ~20,000 nT. The same measurements are also performed under a bias field of ~50,000 nT, under which the recorded signal is almost the same as in Fig. 2 (section S6). For even larger magnetic field, the nonlinear Zeeman effect tends to broaden the line width of the OPM magnetic resonance signal and degrades the performance of the OPM sensor. We can replace cesium with potassium to construct the OPM sensor for large magnetic field. Potassium has a much larger nonlinear Zeeman effect than cesium and, thus, makes the different peaks of the magnetic resonance signal well separated for larger magnetic field. One of these separated peaks can be used for magnetic field measurement.

Auditory evoked signalThe AEF signal is also measured in unshielded Earth’s field. The AEF signal originates from the auditory evoked electric potential generated within the brain while an auditory stimulus is presented (42). The AEF signal has characteristics such as latency and positive and negative peaks, with an amplitude much smaller than that of the spontaneous MEG signal, such as the alpha waves. Measuring the AEF signal is of great importance in understanding, examining, and determining the auditory ability level of an individual, especially, for predicting and diagnosing the hearing loss in newborns, infants, and multiple handicaps.

We measure the AEF signals for three healthy male individuals aged between 22 and 25 years old (see details of the measurement procedures in Materials and Methods). During the measurement, each person is lying on his right side so as to place the temporal lobe right above the OPM sensor (see Fig. 1C). To prevent the heating on the sensor from the face and drifts in the OPM signal amplitude for long­term measurements, we place a silicone pad with a thickness

of 2 mm between the head and the sensor. For each person, we apply 480 auditory stimuli. The time interval between each stimulus is randomly distributed between 0.7 and 1.7 s to prevent the person from being adapted to the stimulus. Each stimulus lasts for 100 ms and is applied automatically to the left ear of the person via plastic tubes and a nonmagnetic earpiece (see the red component in Fig. 1C). For all the 480 auditory stimuli, the frequency of the tune is stan­dard and is 440 Hz. Figure 3A shows the measured AEF signals in the time domain for the three individuals. Each line is averaged from the 480 measured results of all the stimuli with a standard frequency. An AEF signal occurs and can be seen ~100 ms after the stimulus. Such a signal is called M100 and is the most prominent magnetic signature of the cerebral auditory response. Although the AEF sig­nal waveforms of the three persons are similar, the signal strengths are quite different. This is because, for different person, the source of the AEF signal has different depth relative to the scalp (sensor). Figure 3B shows the map of all the 480 measured results for person no. 2. It is shown that for all the measurements, our system is capable of measuring the AEF signal in a stable state.

DISCUSSIONIn this work, we report the MEG results in unshielded Earth’s field with a prototype atomic gradiometer. We show that our unshielded MEG system is capable of measuring the spontaneous alpha rhythm signal and the AEF signal for normal persons. The results prove that the unshielded MEG system based on OPM sensor has sufficient sensitivity for various MEG applications. This is promising to realize a wearable and movable MEG system, which is important for normal clinical applications and for long­term observations and medical diagnosis of brain symptoms. To achieve this goal, several technical issues and challenges need to be considered and solved. These include, but may be not limited to, the suppression of the magnetic

Time (s)−0.2 −0.1 0.0 0.1 0.2 0.3 0.4 0.5

Time (s)−0.2 −0.1 0.0 0.1 0.2 0.3 0.4 0.5

A

B

Mag

net

ic f

ied

(fT

)

0

50

100

−100

−50

150

−150

Tri

al #

200

300

400

100

0

Person no.1Person no.2Person no.3

−200

−100

0

100

200

Mag

netic field

(fT)

Auditory stimulusstart time

Fig. 3. AEF signal. (A) The measured AEF signals in the time domain for three individuals. The auditory stimulus has a frequency of 440 Hz and is repeated for 480 times. The purple dashed arrow shows the start time of the stimulus. The AEF signal occurs ~100 ms after the stimulus. (B) The map of the 480 measured results for person no.2.

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 6: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

6 of 8

gradient noise, the miniaturization of the magnetic sensors for multichannel operation, and the calibrations of the systematic errors from movement of the person.

For the first issue, the atomic gradiometer that we use can sup­press the common­mode magnetic field noise, while it has no effect on suppressing the variations of the magnetic gradient between the two sensors. Thus, the variation of the magnetic gradient is mixed into the detected signal and distorts the MEG signal. This issue can be solved with using more reference magnetometers to actively stabilize the magnetic gradient and to form a higher­order magnetic gradiometer (sections S2 and S4).

Since one of the potential applications of the MEG is to localize the neuronal activity, an array of OPM sensors is needed. The AM­NMOR OPM is suitable for constructing an array given that there is no cross­talk effect between the sensors. Currently, a dominant issue limiting the possibility of a whole­head array is the size of the device. To reduce the size of the device, we need to use a minia­turized vapor cell, with a typical size smaller than 10 mm. Although the relaxation rate of a paraffin­coated cesium cell with a diameter of 3 mm is comparable to that used in our experiment (43), decreasing the size of the vapor cell would markedly decrease the number of atoms that involves in the magnetic sensing, which, in turn, degrades the sensitivity of the magnetometer. This issue can be tackled by heating the vapor cell to increase the atomic number density. Heat­ing paraffin­coated cell with a normal size from 20° to 95°C is experimentally shown to be achievable and is expected to increase the atomic number density by 400 times (44). This is possible to make the magnetic sensitivity of the OPM sensor on the order of femtotesla, which is sufficient for the MEG measurements.

The third issue comes from the movement of the person, which could bring systematic errors in the measured results. First, the move­ment of the person changes the relative direction between the MEG signal and the bias geomagnetic field and, in turn, affects the per­formance of source location, since it is the projection of the MEG signal along the direction of the bias geomagnetic field to be mea­sured. Second, the change in the relative direction between the sensitive axis of the magnetometer and the geomagnetic field also generates heading error in the magnetometer (45). For example, the heading error is on the order of nanotesla when the cesium atomic magnetometer is rotating. To solve these issues, the head movement of the person should be constantly monitored and taken as input parameters for data after processing (46). Besides, the heading error of the atomic magnetometer can be suppressed by either modulating the polarization of the pump beam between left and right circularly polarized at the Larmor frequency (47) or using linearly polarized laser beams or even using other atomic elements, e.g., potassium to construct the OPM sensor.

MATERIALS AND METHODSOptically pumped atomic magnetometerThe magnetometer is based on a pump­probe scheme to polarize the cesium atomic spins and monitor the Larmor precession. The pump laser is circularly polarized and has a 1/e2 diameter of ~2.7 mm. The central frequency of the pump laser is locked to the D1 transi­tion line (from 6 2S1/2 F = 3 to 6 2P1/2 F′ = 4, where F and F′ are the total angular momentum numbers) with dichroic atomic vapor laser lock. The amplitude of the pump beam is modulated with an acoustic­optical modulator at the Larmor frequency. The modula­

tion duty cycle is 20%. The averaged power of the pump laser is ~50 W. The probe laser is linearly polarized and has a 1/e2 diameter of ~1.1 mm. The power of the probe laser is ~50 W. The central frequency of the probe laser is positively detuned by ~400 MHz from the D2 transition line (from 6 2S1/2 F = 4 to 6 2P3/2 F′ = 5). The cesium vapor cell is antirelaxation­coated and is kept at room tem­perature, with a typical magnetic resonance line width of ~5 Hz. The diameter and the length of the cylinder vapor cell are both 25 mm. The vapor cell, including all the optical components, such as the polarizers, wave plates, mirrors, and the Wollaston prism, is mounted in a three­dimensional printed structure, which has a size of 5 cm by 24 cm by 27 cm. The pump and probe laser beams are coupled into the magnetic sensor with optical fibers. The transmitted probe laser from the vapor cell is fiber­coupled to the sensor and then de­tected with a balanced photodetector. The output signal from the bal­anced photodetector is demodulated with a lock­in amplifier (Stanford Research Systems, SR865A, LIA), from which the in­phase compo­nent amplitude is proportional to the difference between the Larmor frequency and the modulation frequency of the pump laser.

Frequency response of the atomic magnetometerTo measure the frequency responses of the magnetometers, we use a pair of Helmholtz coils, which has a diameter of 30 cm and is connected in series with a resistance of 500 ohm and is driven with a signal generator (Keysight, 53230A), to generate a sinusoidal magnetic field with an amplitude of 3.7 nT. The readouts of the two magnetometers, together with the sinusoidal signal from the signal generator, are recorded simultaneously with a data acquisition card (National Instruments, USB6363) at a sampling rate of 40 kSa/s. The readouts are demodulated at the frequency of the applied sinusoidal signal to extract the amplitude and phase shift relative to the sinusoidal signal. The frequency responses of the two magnetometers are thus obtained by scanning the modulation frequency from 1 to 2010 Hz and repeating the above measurement procedures. Similarly, we get the amplitude­frequency characteristics of the gradiometer by de­modulating the difference of the two OPM sensors’ readouts at the frequency of the applied sinusoidal signal. The frequency dependence of the CMRR of the gradiometer is thus obtained by dividing the amplitude­frequency characteristic of OPM1 by that of the gradiometer.

Magnetic field stabilizationThe magnetic noise is compensated using OPM2 and a pair of vertical coils. We use OPM2 as a reference magnetometer to monitor the magnetic field fluctuations. The measured signal from OPM2 is fed into a PID controller (Stanford Research Systems, SIM960), from which the output signal is used to control the current added into the coils. To characterize the ability of rejecting the common­mode mag­netic field noise of the field stabilization, we monitor the residual magnetic noise with OPM1. We use the same coils as those for mea­suring the frequency response to add a white magnetic field noise with a bandwidth of 200 Hz and an amplitude of 58 pT/Hz1/2 to both of the two OPMs and record the readout of OPM1 at a sampling rate of 40 kSa/s. We divide the noise spectrum density of OPM1 obtained under two different conditions, i.e., without and with field stabilization, to derive the CMRR of the field stabilization.

Measurement of the spontaneous alpha rhythm signalEach measurement takes 180 s. To avoid the problem of synchroni­zation, the person is asked to open (close) his eyes before recording

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 7: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

7 of 8

the data and keep his eyes open (closed) until the recording is fin­ished. The multifunction input/output device USB6363 (National Instruments) is used for data acquisition and is controlled with a LabVIEW program. To further confirm that the peak in Fig. 2A around 10 Hz is related to closing eyes, the participant is required to repeat closing and opening his eyes every 30 s to modulate the alpha rhythm signal. We set a 30­s timer around the unshielded MEG system and generate a tinkling sound every 30 s. The person then opens or closes his eyes once hearing this sound.

Measurement of the AEF signalFor the AEF signal measurement, we need to know the precise time duration between the AEF signal and the auditory stimulus. To do so, we use the MATLAB program to generate auditory stimuli sig­nals in advance and save the sound file to the computer. The time interval between each stimulus is randomly distributed between 0.7 and 1.7 s, and each stimulus is a 440­Hz sinusoidal wave with a time duration of 100 ms. Then, we use the LabVIEW program to play the generated sound file and record the measured magnetic field data at the same time. For each measurement, the time uncertainty between playing the sound and recording the data is less than 1 ms, which has negligible effects for the AEF signal measurement. The trans­mission of the sound between the loudspeaker and the person makes relatively large but constant time delay, which is ~10 ms and can be compensated with proper data processing. The measured data are filtered with a 0.5­ to ∼30­Hz band­pass filter.

SUPPLEMENTARY MATERIALSSupplementary material for this article is available at http://advances.sciencemag.org/cgi/content/full/6/24/eaba8792/DC1

REFERENCES AND NOTES 1. D. Cohen, Magnetoencephalography: Evidence of magnetic fields produced by

alpha-rhythm currents. Science 161, 784–786 (1968). 2. D. Cohen, Magnetoencephalography: Detection of the brain’s electrical activity

with a superconducting magnetometer. Science 175, 664–666 (1972). 3. S. Baillet, Magnetoencephalography for brain electrophysiology and imaging.

Nat. Neurosci. 20, 327–339 (2017). 4. D. Cohen, B. N. Cuffin, Demonstration of useful differences between

magnetoencephalogram and electroencephalogram. Electroencephalogr. Clin. Neurophysiol. 56, 38–51 (1983).

5. J. P. Wikswo Jr., A. Gevins, S. J. Williamson, The future of the EEG and MEG. Electroencephalogr. Clin. Neurophysiol. 87, 1–9 (1993).

6. P. Hansen, M. Kringelbach, R Salmelin, MEG: An Introduction to Methods (Oxford Univ. Press, 2010).

7. J. P. Lachaux, D. Rudrauf, P. Kahaneb, Intracranial EEG and human brain mapping. J. Physiol. Paris 97, 613–628 (2003).

8. F. L. da Silva, EEG and MEG: Relevance to neuroscience. Neuron 80, 1112–1128 (2013). 9. J. R. King, F. Faugeras, A. Gramfort, A. Schurger, I. El Karoui, J. D. Sitt, B. Rohaut,

C. Wacongne, E. Labyt, T. Bekinschtein, L. Cohen, L. Naccache, S. Dehaene, Single-trial decoding of auditory novelty responses facilitates the detection of residual consciousness. Neuroimage 83, 726–738 (2013).

10. H. Otsubo, A. Ochi, I. Elliott, S. H. Chuang, J. T. Rutka, V. Jay, M. Aung, D. F. Sobel, O. C. Snead III, MEG predicts epileptic zone in lesional extrahippocampal epilepsy: 12 pediatric surgery cases. Epilepsia 42, 1523–1530 (2001).

11. D. López-Sanz, R. Bruña, P. Garcés, C. Camara, N. Serrano, I. C. Rodríguez-Rojo, M. L. Delgado, M. Montenegro, R. López-Higes, M. Yus, F. Maestú, Alpha band disruption in the AD-continuum starts in the subjective cognitive decline stage: A MEG study. Sci. Rep. 6, 37685 (2016).

12. E. Heinrichs Graham, T. W. Wilson, P. M. Santamaria, S. K. Heithoff, D. Torres-Russotto, J. A. L. Hutter-Saunders, K. A. Estes, J. L. Meza, R. L. Mosley, H. E. Gendelman, Neuromagnetic evidence of abnormal movement-related beta desynchronization in parkinson's disease. Cereb. Cortex 24, 2669–2678 (2013).

13. A. Alhourani, T. A. Wozny, D. Krishnaswamy, S. Pathak, S. A. Walls, A. S. Ghuman, D. N. Krieger, D. O. Okonkwo, R. Mark Richardson, A. Niranjan, Magnetoencephalography-

based identification of functional connectivity network disruption following mild traumatic brain injury. J. Neurophysiol. 116, 1840–1847 (2016).

14. P. Tewarie, M. M. Schoonheim, D. I. Schouten, C. H. Polman, L. J. Balk, B. M. J. Uitdehaag, J. J. G. Geurts, A. Hillebrand, F. Barkhof, C. J. Stam, Functional brain networks: Linking thalamic atrophy to clinical disability in multiple sclerosis, a multimodal fMRI and MEG study. Hum. Brain Mapping 36, 603–618 (2015).

15. D. Brenner, S. J. Williamson, L. Kaufman, Visually evoked magnetic fields of the human brain. Science 190, 480–482 (1975).

16. V. S. Zotev, A. N. Matlashov, P. L. Volegov, I. M. Savukov, M. A. Espy, J. C. Mosher, J. J. Gomez, R. H. Kraus Jr., Microtesla MRI of the human brain combined with MEG. J. Magn. Reson. 194, 115–120 (2008).

17. Y. Seki, T. Miyashita, A. Kandori, A. Maki, H. Koizumi, Simultaneous measurement of neuronal activity and cortical hemodynamics by unshielded magnetoencephalography and near-infrared spectroscopy. J. Biomed. Opt. 17, 107001 (2012).

18. H. Xia, A. B.-A. Baranga, D. Hoffman, M. V. Romalis, Magnetoencephalography with an atomic magnetometer. Appl. Phys. Lett. 89, 211104 (2006).

19. T. H. Sander, J. Preusser, R. Mhaskar, J. Kitching, L. Trahms, S. Knappe, Magnetoencephalography with a chip-scale atomic magnetometer. Biomed. Opt. Express 3, 981–990 (2012).

20. V. K. Shah, R. T. Wakai, A compact, high performance atomic magnetometer for biomedical applications. Phys. Med. Biol. 58, 8153–8161 (2013).

21. C. N. Johnson, P. D. D. Schwindt, M. Weisend, Multi-sensor magnetoencephalography with atomic magnetometers. Phys. Med. Biol. 58, 6065–6077 (2013).

22. K. Kim, S. Begus, H. Xia, S.-K. Lee, V. Jazbinsek, Z. Trontelj, M. V. Romalis, Multi-channel atomic magnetometer for magnetoencephalography: A configuration study. Neuroimage 89, 143–151 (2014).

23. K. Kamada, D. Sato, Y. Ito, H. Natsukawa, K. Okano, N. Mizutani, T. Kobayashi, Human magnetoencephalogram measurements using newly developed compact module of high-sensitivity atomic magnetometer. Jpn. J. Appl. Phys. 54, 026601 (2015).

24. E. Boto, S. S. Meyer, V. Shah, O. Alem, S. Knappe, P. Kruger, T. Mark Fromhold, M. Lim, P. M. Glover, P. G. Morris, R. Bowtell, G. R. Barnes, M. J. Brookes, A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers. Neuroimage 149, 404–414 (2017).

25. E. Boto, N. Holmes, J. Leggett, G. Roberts, V. Shah, S. S. Meyer, L. D. Muñoz, K. J. Mullinger, T. M. Tierney, S. Bestmann, G. R. Barnes, R. Bowtell, M. J. Brookes, Moving magnetoencephalography towards real-world applications with a wearable system. Nature 555, 657–661 (2018).

26. I. K. Kominis, T. W. Kornack, J. C. Allred, M. V. Romalis, A subfemtotesla multichannel atomic magnetometer. Nature 422, 596–599 (2003).

27. V. Shah, S. Knappe, P. D. D. Schwindt, J. Kitching, Subpicotesla atomic magnetometry with a microfabricated vapor cell. Nat. Photonics. 1, 649–652 (2007).

28. O. Alem, A. M. Benison, D. S. Barth, J. Kitching, S. Knappe, Magnetoencephalography of epilepsy with a microfabricated atomic magnetrode. J. Neurosci. 34, 14324–14327 (2014).

29. C. Baumgartner, E. Pataraia, G. Lindinger, L. Deecke, Magnetoencephalography in focal epilepsy. Epilepsia 41, S39–S47 (2000).

30. W. Happer, H. Tang, Spin-exchange shift and narrowing of magnetic-resonance lines in optically pumped alkali vapors. Phys. Rev. Lett. 31, 273–276 (1973).

31. J. C. Allred, R. N. Lyman, T. W. Kornack, M. V. Romalis, High-sensitivity atomic magnetometer unaffected by spin-exchange relaxation. Phys. Rev. Lett. 89, 130801 (2002).

32. S. J. Seltzer, Developments in alkali-metal atomic magnetometry, thesis, Princeton University (2008).

33. W. Gawlik, L. Krzemień, S. Pustelny, D. Sangla, J. Zachorowski, M. Graf, A. O. Sushkov, D. Budker, Nonlinear magneto-optical rotation with amplitude-modulated light. Appl. Phys. Lett. 88, 131108 (2006).

34. D. Budker, D. F. J. Kimball, Optical Magnetometry (Cambridge Univ. Press, 2013). 35. M. E. Limes, E. L. Foley, T. W. Kornack, S. Caliga, S. McBride, A. Braun, W. Lee, V. G. Lucivero,

M. V. Romalis, Total-field atomic gradiometer for unshielded portable magnetoencephalography, arXiv:2001.03534 [physics.med-ph] (7 January 2020).

36. R. Zhang, B. Pang, Y. Ding, W. Li, J. Chen, X. Peng, H. Guo, 3D-printed sensor for unshielded scalar magnetometry based on nonlinear magneto-optical rotation with amplitude modulated light, in 2019 IEEE International Frequency Control Symposium (IFCS) (IEEE, 2019), pp. 14–18.

37. D. Sheng, A. R. Perry, S. P. Krzyzewski, S. Geller, J. Kitching, S. Knappe, A microfabricated optically-pumped magnetic gradiometer. Appl. Phys. Lett. 110, 031106 (2017).

38. R. Zhang, B. Pang, W. Li, Y. Yang, J. Chen, X. Peng, H. Guo, Frequency response of a close-loop bell-bloom magnetometer, in 2018 IEEE International Frequency Control Symposium (IFCS) (IEEE, 2018), pp. 21–24.

39. J. Bork, H.-D. Hahlbohm, R. Klein, A. Schnabel, The 8-layered magnetically shielded room of the PTB: Design and construction, in Biomag2000: Proceedings of the 12th International Conference on Biomagnetism (University of Technology, 2001).

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 8: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

Zhang et al., Sci. Adv. 2020; 6 : eaba8792 12 June 2020

S C I E N C E A D V A N C E S | R E S E A R C H A R T I C L E

8 of 8

40. M. Hämäläinen, R. Hari, R. J. Ilmoniemi, J. Knuutila, O. V. Lounasmaa, Magnetoencephalography-theory, instrumentation, and applications to noninvasive studies of the working human brain. Rev. Mod. Phys. 65, 413 (1993).

41. F. H. Lopes da Silva, J. P. Pijn, D. Velis, P. C. G. Nijssen, Alpha rhythms: Noise, dynamics and models. Int. J. Psychophysiol. 26, 237–249 (1997).

42. R. Hari, K. Aittoniemi, M. L. Järvinen, T. Katila, T. Varpula, Auditory evoked transient and sustained magnetic fields of the human brain. Localization of neural generators. Exp. Brain Res. 40, 237–240 (1980).

43. M. V. Balabas, D. Budker, J. Kitching, P. D. D. Schwindt, J. E. Stalnaker, Magnetometry with millimeter-scale antirelaxation-coated alkali-metal vapor cells. J. Opt. Soc. Am. B 23, 1001–1006 (2006).

44. W. Li, M. Balabas, X. Peng, S. Pustelny, A. Wickenbrock, H. Guo, D. Budker, Characterization of high-temperature performance of cesium vapor cells with anti-relaxation coating. J. Appl. Phys. 121, 063104 (2017).

45. E. B. Alexandrov, Recent progress in optically pumped magnetometers. Phys. Scripta T105, 27–30 (2003).

46. S. Supek, C. J. Aine, Magnetoencephalography: From Signals to Dynamic Cortical Networks (Springer International Publishing, 2019).

47. A. Ben-Kish, M. V. Romalis, Dead-zone-free atomic magnetometry with simultaneous excitation of orientation and alignment resonances. Phys. Rev. Lett. 105, 193601 (2010).

Acknowledgment: We acknowledge W. Zhang from the National Institute of Metrology of China for assistance with the initial early-stage experimental efforts. Funding: This work is supported by the National Natural Science Foundation of China (grants nos. 61571018, 61531003, and 91436210), the National Science Fund for Distinguished Young Scholars of China (grant no. 61225003), and National Hi-Tech Research and Development (863) Program. T.W. acknowledges the support from the start-up funding for young researchers of the Peking University. Author contributions: R.Z. developed the atomic gradiometer and the field

stabilization system, designed and performed the experiments, analyzed the data, and wrote the manuscript. W.X. designed and performed the experiments, analyzed the data, and wrote the manuscript. Y.D. helped to develop the atomic gradiometer and the field stabilization system and performed the experiments. Y.F. devised the experimental protocol, performed experiments, analyzed the data, and proofread and edited the manuscript. X.P. proposed the experimental concept and proofread and edited the manuscript. L.S. helped to develop the field stabilization and performed the experiments. C.S. devised the experimental protocol, performed experiments, analyzed the data, and proofread and edited the manuscript. T.W. contributed to the design of the experiment and wrote the manuscript. Y.W. performed the experiments. Y.Y. helped to develop the atomic gradiometer and performed the experiments. Z.Z. helped to develop the atomic gradiometer and performed the experiments. X.Z. performed the experiments. J.C. proposed the experimental concept, supervised the project, and proofread and edited the manuscript. H.G. provided the overall management of the project, contributed to the design of the experiment, and proofread and edited the manuscript. All authors contributed with discussions and to the final form of the manuscript. Competing interests: The authors declare that they have no competing interests. Data and materials availability: All data needed to evaluate the conclusions in the paper are present in the paper and/or the Supplementary Materials. Additional data related to this paper may be requested to the corresponding authors.

Submitted 13 January 2020Accepted 1 May 2020Published 12 June 202010.1126/sciadv.aba8792

Citation: R. Zhang, W. Xiao, Y. Ding, Y. Feng, X. Peng, L. Shen, C. Sun, T. Wu, Y. Wu, Y. Yang, Z. Zheng, X. Zhang, J. Chen, H. Guo, Recording brain activities in unshielded Earth’s field with optically pumped atomic magnetometers. Sci. Adv. 6, eaba8792 (2020).

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from

Page 9: APPLIED PHYSICS Copyright © 2020 Recording brain activities in … · most sensitive magnetic sensors operating near zero magnetic field. As a comparison, the AM NMOR magnetometer

magnetometersRecording brain activities in unshielded Earth's field with optically pumped atomic

Yang, Zhaoyu Zheng, Xiangzhi Zhang, Jingbiao Chen and Hong GuoRui Zhang, Wei Xiao, Yudong Ding, Yulong Feng, Xiang Peng, Liang Shen, Chenxi Sun, Teng Wu, Yulong Wu, Yucheng

DOI: 10.1126/sciadv.aba8792 (24), eaba8792.6Sci Adv 

ARTICLE TOOLS http://advances.sciencemag.org/content/6/24/eaba8792

MATERIALSSUPPLEMENTARY http://advances.sciencemag.org/content/suppl/2020/06/08/6.24.eaba8792.DC1

REFERENCES

http://advances.sciencemag.org/content/6/24/eaba8792#BIBLThis article cites 39 articles, 4 of which you can access for free

PERMISSIONS http://www.sciencemag.org/help/reprints-and-permissions

Terms of ServiceUse of this article is subject to the

is a registered trademark of AAAS.Science AdvancesYork Avenue NW, Washington, DC 20005. The title (ISSN 2375-2548) is published by the American Association for the Advancement of Science, 1200 NewScience Advances

License 4.0 (CC BY-NC).Science. No claim to original U.S. Government Works. Distributed under a Creative Commons Attribution NonCommercial Copyright © 2020 The Authors, some rights reserved; exclusive licensee American Association for the Advancement of

on August 27, 2020

http://advances.sciencemag.org/

Dow

nloaded from