Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie Professor University of Wisconsin Co-Founder and Chairman of the Board TomoTherapy Inc.

Transcript of Applications of Modern Radiotherapy...

Page 1: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Applications of Modern Radiotherapy Systems

Thomas Rockwell Mackie

ProfessorUniversity of Wisconsin

Co-Founder and Chairman of the Board

TomoTherapy Inc.

Page 2: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Financial Disclosure

I am a founder and Chairman of TomoTherapy Inc. (Madison, WI) which is participating in the commercial development of helical tomotherapy.

Page 3: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Acknowledgements

John SchreinerJerry BattistaTim HolmesGustavo OliveraWeiguo LuKatja Langen

Paul KeallDavid ShepardCedric YuThomas BortfeldAn LiuChet Ramsey

Page 4: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Outline

• Setup correction for interfraction motion

• Off-line adaptive techniques

• On-line adaptive strategies

• Modern Delivery methods

• Treatment Planning Issues

• Delivery time

• Novel clinical applications

• Clinical impact

Page 5: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Courtesy of John Schreiner, Kingston Regional Cancer Centre, Ontario

Dose Sculpting and Painting

2-D Planning

IMRT

3-D

Conformal

Dose

Painting

Page 6: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Why 3D Image-Guided Radiotherapy (IGRT)?

• Eventually, most curative radiotherapy will be IMRT, even many palliative treatments, e.g., re-treatments.

• All IMRT should be image-guided:– IMRT is justified by sparing critical tissues (conformal avoidance) which produces higher dose gradients.

– IGRT enables higher gradients to be delivered safely and effectively.

– IGRT enables a smaller setup margins to be defined.• In some radiotherapy sites, e.g., prostate, IGRT may be more important than IMRT.

• 2D imaging is inadequate to obtain volume information.

Page 7: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

“If you can’t see it , you can’t hit it

If you can’t hit it, you can’t cure it”

From Jerry Battista

Page 8: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

1 second 1 minute 1 hour 1 day 1 week

Intra-fractional time scale

Inter-fractional time scale

time

respiratory,

cardiac

motion digestive

system

motion

bowel/

bladder

filling

random/

systematic

setup errors

tumor

growth

and

shrinkage

weight

gain and

loss

Radiotherapy Time Scales

Page 9: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Is Daily Imaging Necessary?Translational Setup Error by Disease Site

0.0

1.0

2.0

3.0

4.0

5.0

6.0

7.0

0 1 2 3 4 5

Disease Site

Error (mm)

Rotational Setup Error by Disease Site

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

0 1 2 3 4 5

Disease Site

Error (degrees)

∆ systematic

● random

lateral

longitudinal

vertical

Brain H&N Lung Prostate Brain H&N Lung Prostate

∆ systematic

● random

roll

• Translational corrections smaller for brain and H&N than for lung and prostate treatments

• Rotational corrections greater for brain and H&N than for lung and prostate

• Shows that when we image, we do make shifts. Does this improve outcomes? Can we use this to image less?

Page 10: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Residual Errors For Imaging Protocols

Error As a Function of % Days Image Guidance Used

0

20

40

60

80

100

0 20 40 60 80 100

% Image Guidance

% Times Error

Occurs

Error > 3 mm

Error > 5 mm

Error > 10 mm

Every second day

Weekly

NeverEvery day

Adapted from Zeidan et al, Int. J. Radiat. Oncol. Biol. Phys. 2007; 67:670-677

Head and Neck

Small residualuncertainty

Page 11: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Off-line Adaptive Techniques

• Dose verification and replanning

• Dose reconstruction

• Deformable registration of contours and dose distributions

• Dose trending

Page 12: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Off-Line Adaptive3-D Imaging

OptimizedPlanning

Pre-DeliveryImaging

RadiotherapyTreatment

DoseReconstruction

Setup Adjustment

DeformableDose

Registration

Page 13: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Quantitative Imaging forAdaptive Therapy

• Quantitative images are required for many

Adaptive Therapy Processes:

– Delivery Verification.

– Dose Reconstruction.

– Deformable Dose Registration.

– Re-optimization.

• All of these can be achieved with

tomotherapy’s CT images.

Page 14: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Patient courtesy of Tim Holmes, St. Agnes Cancer Center Baltimore, MD

Re-calculated plan with shifted target

Dose Verification – Recomputing Dose

Page 15: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Planed Dose to the PTV

Delivered Dose to the PTV

Planned Max Cord Dose

Max Cord Dose with Existing Plan

Per-fraction DVH

Re-calculated plan with shifted target

Clinical Adaptive Planning

Patient courtesy of Tim Holmes St. Agnes Cancer Center Baltimore, MD

Page 16: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Patient courtesy of Tim Holmes, St. Agnes Cancer Center Baltimore, MD

Planed Dose to the PTV

Delivered Dose to the PTV

Planned Max Cord Dose

Max Cord Dose with Existing Plan

PerPer--fraction fraction DVHDVH

Revised Plan Revised Plan DVH for DVH for remaining remaining fractionsfractions

Replanning

Revised plan for remaining Revised plan for remaining

fractionsfractions

Max Cord Dose with Revised Plan

Page 17: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Original Planning CT

Original planning CT

Reference CT

Daily CT Daily CT mapped

to Reference CT

Page 18: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Original planning CT

Reference CT

Daily CT Daily CT mapped

to Reference CT

Page 19: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Original planning CT

Reference CT

Daily CT Daily CT mapped

to Reference CT

Page 20: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Accuracy of Automatic H+N MVCT contours

kVCT contours: MVCT contours:

Parotids: Visual inspection of automatic

contours

Page 21: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

MVCT to kVCTDeformable Image Registration

kVCT:

PTV

Cord

MVCT:

PTV

Cord

Page 22: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Accuracy of Automatic H+N MVCT Contours

Spinal Cord: Automatic vs. manual contours

Compare dose-based end points: Dmax, Dmean

3 patients:

Dmax %difference: 1.1 + 3.5%

Dmean %difference: 0.1 + 2.5 %

Langen et al., AAPM, 2006

Dmean

0.6

0.8

1.0

0 5 10 15 20 25 30 35Treatment fraction

Dose (Gy)

Automatic Manual

Dmax

1.0

1.2

1.4

0 5 10 15 20 25 30 35Treatment fraction

Dose (Gy)

Automatic Manual

Page 23: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Trending

1st fraction

Page 24: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Cont. to 35 fractions

Trending

Page 25: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Re-plan at 20 fractions

Trending

Page 26: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

On-line Adaptive Techniques

• Detection of intrafraction motion

• Methods of motion management• Gating

• Tracking

• Delivery modification

Page 27: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Lung Motion Dynamics

The motion can be complex

Motion close

to diaphragm

Motion mid

lung

Motion tumor

at center

Motion lower

chest

Page 28: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Lung Motion Dynamics

Page 29: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Navotek RealTrack System• Fine radioactive wire is implanted in the patient.• Can see the wire on IGRT systems.• Can track wire position in real time.

Page 30: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Calypso System• Electromagnetic transponders are implanted in the patient.• Can see the transponder on IGRT systems.• Can track transponder position in real time.

Page 31: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Degradation of Alignment Quality after Initial Setup for Prostate Cancer

0

10

20

30

40

50

60

0 5 10 15Time (minutes)

Precentageof observations

Percentage of displacements

observed

over time after initial alignment

>3 mm

>5 mm

>7 mm

>10 mm

17 patients

551 Calypso

sessions

Mean: 32

tracks

per patient

Courtesy of Katja Langen, M.D. Anderson Cancer Center, Orlando, FL

Page 32: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

MRI-Guided Systems for Real Time Imaging

View Ray

AlbertaUtrecht

Page 33: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Motion Management

• Margin: Put an ITV on a CTV

• Breath holding: Either active or passive

• Gating: Not very efficient

• Tracking: Move the patient or move the beam

• Delivery Modification: Can also handle non-rigid motion

Page 34: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

The Breathing Cycle

Inhalation Inhalation

Exhalation

Gating Paradigm

Assumes no motion

in gated portion

Loss of Efficiency

Page 35: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Motion Management (4D Planning and Delivery)

Treating while the patient breathes is more accurate as compared to tracking, and saves time as compared to gating.

Works in Progress

Page 36: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Nemo Demo

Motion SurrogateNemo

Static Tank

Page 37: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Fish Represents a Moving Tumor

Page 38: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

How Necessary is Motion Management?Magnitude of Motion

Tumor Size

< 5 mm > 5mm < 1cm > 1 cm

SmallFraction, Including Margin, of OAR Volume

Not Needed Not Needed

Not Needed Not Needed PerhapsNeeded

LikelyNeeded

PerhapsNeeded

MediumFraction of OAR Volume

LargeFraction of OAR Volume

Not Needed

Not Needed

Page 39: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Why IMRT Is Needed

• 5 IMRT beams is more conformal than 5 conformally-shaped beams.

– If uniform fields were sufficient, the pencil beam weights for each beam would be identical.

54

12

3

54

12

3

Page 40: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Multileaf Collimators (MLC’s)

Siemens Varian NOMOS

• Conventional MLC’s were designed for field shaping

and have limitations when used for IMRT.

BinaryConventional

• Binary (off-on) MLC’s are designed for IMRT and are

the easiest to model and provide high modulation.

Page 41: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Segmental MLC IMRT

• “Step and shoot”.

• Deliver multiple MLC apertures within a field to apply the intensity in a paint-by-number fashion.

• May be a straightforward technique for “forward” optimization.

• Little change in paradigm involved in field boundary verification using portal imaging.

• May be relatively time consuming if field delivery is verified in exactly the same way.

• Only discrete intensity levels can be delivered.

Page 42: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Segmental MLC IMRT

A set of leaf sequences to deliver

an intensity modulated field.Conventional MLC’s have been designed

for field shaping not IMRT.

Page 43: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Dynamic MLC IMRT

• Pairs of MLC leaves are in continuous movement across the field with the intensity at a point equal to the total exposure time of the leaf pair above it.

• Most efficient delivery for modest modulation of intensities.

• High spatial variation of intensities are difficult.

• Continuous intensity levels.

• Difficult to verify with conventional techniques as anatomic details blend into the continuous intensity levels.

Page 44: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Dynamic MLC Leaf Motion

From Paul Keall, Stanford University

Page 45: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Intensity Modulated Arc Therapy (IMAT)

• Collimator leaves move dynamically as the gantry rotates.

• Beams delivered from all coplanar directions. • Requires multiple arc deliveries to achieve intensity modulation.

• Provided field length is not too long, no couch translations are necessary.

• Proposed by Cedric Yu and implemented by Wilfried DeNeve (Ghent, Belgium).

• Single arc IMAT is branded “VMAT” or “RapidArc” and also changes the dose rate during the rotation to achieve some limited modulation.

Page 46: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

IMAT

Field shape changes dynamically during rotation.Needs multiple rotations.

Page 47: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

IMAT Intensity Levels

i is the number of non-zero intensity levels.n is the number of rotations.

3

n = 2

Two Rotation IMAT

71i

n = 3 n = 1

Three Rotation IMAT

One Rotation

IMAT

2 1ni = −

Following Equation is from Cedric Yu:

Page 48: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

3 SeparateRotations with

Different IntensitiesPer Rotation

1 2 3 4 5 6 7

Example of 3 IMAT Rotations

Yields 7 Unique Non-Zero Intensity Levels

From David ShepardAnd Cedric Yu

Page 49: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Tomotherapy

• Tomotherapy is intensity modulated rotational therapy with a fan beam of radiation and is analogous to a CT scanner.

• It utilizes a binary collimator to provide the modulation.

• Serial tomotherapy was first form of IMRT

• In helical tomotherapy, the gantry and couch move simultaneously.

Page 50: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

TomoTherapy HI-ART Unit

Linac ShownWithout Shielding

Pulse FormingNetwork andModulator

DetectorBeam StopHigh VoltagePower Supply

ControlComputer

Magnetron

Data Acquisition System

Circulator

Gun Board

Page 51: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

How the Intensity is Modulated with Tomotherapy

Open

Closed

For Low Intensity ALeaf Is Open for a Short Time DuringThe Projection

For High Intensity ALeaf Is Open for a Long Time During The Projection

One Rotation is Divided into Angular SegmentsCalled a Projection

Binary leaves were specifically designed for IMRT.

Page 52: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Sample Sinogram

Collimator Position

Angle

A delivery sinogramis a representationof the energy fluence delivered to thepatient. The energyfluence distributionis the realization ofthe sinogram andtakes into accountphoton attenuation.

Page 53: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

A Sinogram Specifies the Relation Between Collimators and Voxels

Collimator Index

GantryAngle

Page 54: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Helical Delivery Sinogram

Example with 13 Rotations

GantryAngle

Collimator Index

Darker Is HigherIntensity or LongerOpening Time

AvoidanceOf NormalTissue

The delivery sinogramspecifies the intensity(or leaf opening time) as a function ofgantry angle.

Page 55: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

IMRT Requires Optimization

Hard Constraints• cannot be violated

• may not lead to a feasible solution

Soft Constraints• constraints may be violated

• find optimal intensity profile

• may lead to a local minimum

Page 56: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Beamlet-Based Optimized Planning

• Two-step approach to treatment planning:

1. Fluence map optimization – Delivery constraints ignored

2. Leaf sequencing – Accounts for delivery constraints

• Employed by nearly all vendors.

• Corvus (NOMOS)• Eclipse (Varian) • XiO (CMS)• Pinnacle (Philips)• Oncentra (Nucletron)• Hi-Art (TomoTherapy)

Page 57: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Field Divided into a Grid of Beamlets

From Cedric Yu, U. of Maryland

Page 58: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

2

1

1

1

1

1

2

2

2

1

1

1

2

1

1

2

1

1

2

2

1

3

21 1 32

3 12

3

2 1

3 1 11

11 11

21111

1

2

1

1

2

2

21111 2

Optimized Fluence Map

From Cedric Yu, U. of Maryland

Page 59: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Leaf Sequencing

2

1

1

1

1

1

2

2

2

1

1

1

2

1

1

2

1

1

2

2

1

3

21 1 32

3 12

3

2 1

3 1 11

11 11

21111

1

2

1

1

2

2

21111 2

Optimized Fluence Map

Deliverable Apertures

From Cedric Yu, U. of Maryland

Page 60: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Beamlet IMRT Approach for Conventional MLC IMRT

Clinical Objectives, Constraints

Intensity Maps

MLC Segments1

654

32

From Cedric Yu, U. of Maryland

Page 61: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Aperture-Based IMRT

Clinical Objectives, Constraints

MLC Segments1

654

32

Shepard, Earl, Li, Naqvi, Yu

“Direct aperture optimization”

Med. Phys. 29(6):1007-1018, 2002

From Cedric Yu, U. of Maryland

Page 62: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Delivery Type Dictates IdealOptimization Type

X

IMAT

XX√Aperture

√√XBeamlet

Tomotherapy

(Binary MLC)

DMLC (Dynamic MLC)

SMLC

(Step and Shoot)

Delivery Type

Ideal Optimization Type

Page 63: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Inadequateangularsampling

Inadequateintensitysampling.

Inadequatespatial (pixel)resolution.

Page 64: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Spatial and Angular ResolutionAngular sampling interval, ∆Φdominates image resolution at the peripheryof an axial image.

Pixel size, ∆ldominates imageresolution at thecenter. ∆l

∆Φ

FOV

Page 65: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Angular Sampling Required for CT

FOV is the field of view.

∆l is the spatial resolution.

1260 views

FOV = 40 cm

∆l = 0.1 cm

Large FOV

630 viewsRequirement

FOV = 20 cm

∆l = 0.1 cm

Small FOV

# / oFOV

Viewsl

ππππ

∆∆∆∆====360

Page 66: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Angular Sampling Required for Radiotherapy

FOV is the field of view (max field width)

∆l is the spatial resolution (collimator

resolution)

51 views

FOV = 40 cm

∆l = 0.6 cm

Tomotherapy

25 viewsRequirement

FOV = 15 cm

∆l = 0.5 cm

Conv. IMRT

# / o FOVViews

l

ππππ

∆∆∆∆====360

Page 67: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Brahme‘s Classic 1982 Paper on Rotational IMRT

Phantom

Target OAR

Dose

Rotating

Source

From Tomas Bortfeld

Fluence

Page 68: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Sufficient Modulation and Beam Numbers Are Needed to “Construct Dose”

1 Beam 5 Beams 11 Beams

17 Beams 25 Beams 51 Beams

mm mm mm

mmmmmm

Notice the High Degree of ModulationRequired Even if Rotation is Used

Page 69: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Discretization Error

i is the number of non-zero intensity levels.N is the number of beam directions.

2.2%

i = 1

N = 180

Single Arc

VMAT

0.04%2.2%1 S.D. Error

i = 100

N = 51

i = 5

N = 7

Helical Tomotherapy

Step and Shoot

IMRT

σσσσ ====1

12 i N

Following Equation is from Imaging Theory:

Page 70: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Conventional IMRT Delivery Time Analysis

From Sha Chang, UNC

Average Tomo

Page 71: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Tomo’s Time In Room and Treatment Time

Ave = 5 min

Treatment Time = Beam on Time = Time for Treatment Irradiation

Ave = 17 min

• Other studies have put the average beam on time at 7 minutes for tomotherapy

• Depends on the amount of modulation

• Shorter tumors like prostate have shorter beam on time

• Longer tumors like cranial spinal have longer beam on times

Page 72: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Re-treatments, using tomotherapy for patients not eligible for conventional photon radiation therapy due to cord tolerance.

Patients courtesy of UAB

Re-Treatments

Page 73: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Complex Abdominal/Pelvic

Heidelberg University Clinic

Page 74: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

CraniospinalTomotherapyTomotherapy TraditionalTraditional

Page 75: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Craniospinal

TomotherapyTomotherapy

TraditionalTraditional

Page 76: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Total Marrow

Irradiation

(TMI)

From Dr. An Liu,City of Hope, Duarte CA

Conformal Avoidance of:BrainThyroidLungsLiverKidneysSmall Bowel

Page 77: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Strategy for Conformal Avoidance Radiotherapy

• Use generous treatment volumes.• Outline normal sensitive tissues and concentrate on avoiding them.

• Use image-guidance to assure that the normal tissues are being avoided.

• Conformal avoidance radiotherapy is the complement of conformal radiotherapy.

• If you can’t see it you can’t avoid it.• If you can’t avoid it you can’t spare it.

Page 78: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

With Better Avoidance of Normal Tissue is Hypofraction Possible?

• In prostate CA, the tumor may repair even better than the normal tissues.

• In lung CA, rapid proliferation reduces the treatment control probability as the treatment is extended in duration.

• Provided better avoidance of sensitive tissues is maintained, fewer fractions of higher dose/fraction will provide both better tumor control and be less expense to deliver.

• Carefulness can be cost effective.

Page 79: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Image-Guided Radiotherapy of the Future

• Image-based staging of the primary and regional field.

– Determine hypoxic and highly proliferative regions using bioimaging and paint in higher dose.

– Conformally avoid sensitive structures in the regional field.

• IMRT with 3-D image verification.

– Less fraction quantity – greater fraction quality.

– Adaptive radiotherapy to provide patient-specific QA of the whole course of therapy.

Page 80: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Image-Guided Radiotherapy of the Future (Cont.)

• Image-based monitoring of outcome.

– e.g., PET scans for regional or metastatic development using a priori information.

• Aggressive treatment of recurrences or distant metastases using conformal avoidance to spare critical structures.

– Better QA of first treatment will allow safer retreatments.

– “Weeding the garden” with image-guided radiotherapy and prevent spread with chemotherapy and immunotherapy.

Page 81: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Oligometastases or “Weeding the Garden”

• Following definitive radiotherapy with local control we often have metastatic progression.

• Chemotherapy (analogous to pre-emergent herbicides) is known to be effective against 100 to 1000 cell tumorlets.

• With PET it is possible to infer the presence of tumorletswith 100,000 to 1,000,000 labeled cells.

• Perform PET scan followups to catch the emergent tumorlets.

• Weed with conformal avoidance hypofractionatedradiotherapy before they can seed more metastases.

• Keep careful track of the cumulative dose delivered so the process can be repeated several times if necessary.

Page 82: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Courtesy of Chet Ramsey, Thompson Cancer Survival Center

Planned Using PET-CT

PET-CT Scans

Treating Multiple Metastases Determined From PET Scans

Tomotherapy Treatment Plan

Page 83: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Visualization Gap

102 103 104 105 106 107 108 109

External Beam

Radiotherapy

Effective

MR/CT Tumor

Visualization

PET Tumor VisualizationChemotherapy

Effective

Tumor Cell Density

(cells/cm3)

1010

Targeted Agents

Effective

101

Visualization Gap

Page 84: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Implications of the Visualization Gap

• Systemic agents are effective for tiny tumorlets.

• Larger tumorlets may shrink so that they are not visible but they are likely to return.

• Systemic agents are more effective when no tumorletsare visible, i.e., used as a prophylaxis agent.

• Radiation therapy is effective for much larger tumors

• If imaging systems were sensitive for smaller tumors, radiation therapy could be used systemically.

• Systemic agents should aim for higher cell kill.

• If the tumor size range of systemic agents and radiation therapy could overlap, cancer could be made a chronic disease.

Page 85: Applications of Modern Radiotherapy Systemsmpss.iop.org/Summer_School/2009/Programme2009/MackieModernpho… · Applications of Modern Radiotherapy Systems Thomas Rockwell Mackie ...

Conclusions• Setup corrections can correct for translations and

rotations.

• Adaptive therapy accounts for changes in the patient.

• Motion management can accommodate for breathing

and organ filling.

• IMRT and rotational therapy will dominate curative

treatments

• The type of optimization depends on the type of

delivery.

• IMRT takes more time than conventional radiotherapy

• Treatments not possible with conventional

radiotherapy are being done.

• Conformal avoidance enables hypofractionated

treatments.