Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods...

2
Application Highlights A Flame Ionization Detector (FID) to detect the C1 through C12 n-paraffins to a lower detection limit of 1 ppm, except for trace peaks eluting on the tail of a major component. A Thermal Conductivity Detector (TCD) to detect air composite, carbon dioxide, C1 through C5 paraffins with an initial C6+ composite backflush to detector. System configured to meet Gas Processors Association Methods 2177, 2261, 2186, and 2286. Approximate analysis time is 30 minutes. Optional Configurations Detailed hydrocarbon analysis of extended natural gas TCD/FID/FPD or TCD/FID/SCD for extended natural gas with trace sulfur analysis TCD/TCD/FID for extended natural gas with helium or hydrogen Application 325-00 Agilent Extended Natural Gas Analyzer For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. Technical Overview 1 Methane 2 Ethane 3 Propane 4 Isobutane 5 n-Butane 6 Isopentane 7 n-Pentane 8 Hexane 9 Benzene 10 nC7+ 11 Toluene 12 nC8+ 13 Ethylbenzene 14 m/p-Xylene 15 o-Xylene 16 nC9+ 17 nC10+ 18 nC11+ 19 nC12+ FID output TCD output 1 2 2 11 3 3 4 4 1 5 5 6 6 7 8 7 8 9 10 9 10 12 13 14 15 16 17 18 19 1 C6+ backflush 2 Air composite 3 Methane 4 Carbon dioxide 5 Ethane 6 Propane 7 Isobutane 8 n-Butane 9 Isopentane 10 n-Pentane

Transcript of Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods...

Page 1: Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods 2177, 2261, 2186, and 2286.

Application Highlights• A Flame Ionization Detector (FID) to detect the C1

through C12 n-paraffins to a lower detection limit of1 ppm, except for trace peaks eluting on the tail of amajor component.

• A Thermal Conductivity Detector (TCD) to detect aircomposite, carbon dioxide, C1 through C5 paraffins withan initial C6+ composite backflush to detector.

• System configured to meet Gas Processors AssociationMethods 2177, 2261, 2186, and 2286.

• Approximate analysis time is 30 minutes.

Optional Configurations• Detailed hydrocarbon analysis of extended natural gas

• TCD/FID/FPD or TCD/FID/SCD for extended naturalgas with trace sulfur analysis

• TCD/TCD/FID for extended natural gas with helium orhydrogen

Application 325-00

Agilent Extended Natural Gas Analyzer

For More Information

For more information on our products and services, visit ourWeb site at www.agilent.com/chem.

Technical Overview

1 Methane2 Ethane3 Propane4 Isobutane5 n-Butane6 Isopentane7 n-Pentane8 Hexane9 Benzene10 nC7+

11 Toluene12 nC8+13 Ethylbenzene14 m/p-Xylene15 o-Xylene16 nC9+17 nC10+18 nC11+19 nC12+

FID output

TCD output

1

2

2

11

3

3

4

41

5

5

6

6

7

8

7

8

910

9 10

12

13

14

15

16

17 18 19

1 C6+ backflush2 Air composite3 Methane4 Carbon dioxide5 Ethane6 Propane7 Isobutane8 n-Butane9 Isopentane10 n-Pentane

Page 2: Application 325-00 Agilent Extended Natural Gas Analyzer · Application Highlights ... Methods 2177, 2261, 2186, and 2286.

Agilent shall not be liable for errors contained herein or for incidental or consequen-tial damages in connection with the furnishing, performance, or use of thismaterial.

Information, descriptions, and specifications in this publication are subject to changewithout notice.

© Agilent Technologies, Inc. 2002

Printed in the USADecember 6, 20025988-6736EN

www.agilent.com/chem

1 Methane2 Ethane3 Propane4 Isobutane5 n-Butane6 Isopentane7 n-Pentane8 Hexane9 Benzene10 nC7+

11 Toluene12 nC8+13 Ethylbenzene14 m/p-Xylene15 o-Xylene16 nC9+17 nC10+18 nC11+19 nC12+

FID output

TCD output

1

2

2

11

3

3

4

41

5

5

6

6

7

8

7

8

910

9 10

12

13

14

15

16

17 18 19

1 C6+ backflush2 Air composite3 Methane4 Carbon dioxide5 Ethane6 Propane7 Isobutane8 n-Butane9 Isopentane10 n-Pentane

FID and TCD output from the Agilent Extended Natural Gas Analyzer.